34086-001 Version A
11-3
6. Clear the alarm by pressing the Alarm Reset button. The audible alarm silences and the display changes to
VENT CHECK.
34086-001 Version A
11-4
Alarm Test
Use the Alarm Test to verify that the audible alarm is working correctly.
34086-001 Version A
11-5
34086-001 Version A
11-6
Display Test
Use the Display Test to verify that the ventilator displays are working correctly.
NOTE
The External Power LED is tested and verified when the AC adapter is connected to the ventilator (see
“Chapter 11:–Ventilator Checkout Tests”).
34086-001 Version A
11-7
34086-001 Version A
11-8
Control Test
Use the Control Test to verify that the ventilator buttons and the Scroll knob are working correctly.
34086-001 Version A
11-9
3. Test each control by pressing each button, one at a time. When pressed, verify that the name of the button
pressed displays in the display window. Control names are as shown in the table below.
Control Display
Display Select SELECT
Breath Rate BREATH RATE
Tidal Volume TIDAL VOLUME
Pressure Control PRES CONTROL
Inspiratory Time INSP TIME
Pressure Support PRES SUPPORT
FiO2 (Flush) FIO2
(LTV2 2200 only)
Sensitivity SENSITIVITY
High Pres Limit HI PRESSURE
Low Peak Pres LO PRESSURE
Low Min Vol LOW MIN VOL
Alarm Silence SILENCE
Alarm Reset ALARM RESET
Power/Standby ON / STNDBY
Volume & Pressure MODE VOL/PRS
Assist/Control & SIMV/CPAP MODE A/C S/C
Insp / Exp Hold IE HOLD
Manual Breath MANUAL BRTH
Low Pressure O2 Source LOW PRES O2
(LTV2 2200 only)
Control Lock CONTROL LOCK
Scroll knob rotate Left ROTATE LEFT
Scroll knob rotate Right ROTATE RIGHT
PEEP PEEP
Battery Check BATT STATUS
4. Test the Scroll knob by turning it clockwise and counterclockwise. Verify that the direction of rotation
displays in the display window.
5. To exit the control test, press the Menu/Select button again and the next menu item displays.
If the Control Test fails, see “Chapter 15:–Troubleshooting” for more information.
34086-001 Version A
11-10
Leak Test
Use the Leak Test to test the patient circuit for leaks.
Leak Test is available upon powering on the ventilator (if Leak Test Query is enabled in the Queries
section of the Extended Features Menu) or with Leak Test in the Vent Check section of the Extended
Features.
34086-001 Version A
11-11
10. To exit the Leak Test, press the Menu/Select button. The ventilator will restart and present the LeakTest
YES message again. If the previous leak test passed, turn the Scroll knob to LeakTest NO and press the
Menu/Select button to continue with startup.
If the ventilator fails the Leak Test, see “Chapter 15:–Troubleshooting” for more information.
34086-001 Version A
11-12
Exit
To exit the vent check mode and return to normal ventilation mode at any point proceed as follows:
34086-001 Version A