Operating manual
Copyright © 2020 Eppendorf AG, Germany. All rights reserved, including graphics and images. No part of
this publication may be reproduced without the prior permission of the copyright owner.
Eppendorf® and the Eppendorf Brand Design are registered trademarks of Eppendorf AG, Germany.
Registered trademarks and protected trademarks are not marked in all cases with ® or ™ in this manual.
Table of contents
1 Operating instructions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7
1.1 Using this manual . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7
1.2 Danger symbols and danger levels . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7
1.2.1 Danger symbols . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7
1.2.2 Danger levels. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7
1.3 Symbols used . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7
1.4 Abbreviations used . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8
2 Safety. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9
2.1 Intended use . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9
2.2 User profile . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9
2.3 Information on product liability . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9
2.4 Application limits . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10
2.4.1 Declaration concerning the ATEX directive (2014/34/EU) . . . . . . . . . . . . . . . . . . . . . . 10
2.5 Warnings for intended use . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10
2.5.1 Personal injury or damage to device. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10
2.5.2 Incorrect handling of the centrifuge . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 12
2.5.3 Incorrect handling of the rotors . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 13
2.5.4 Extreme strain on the centrifugation tubes . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 14
2.6 Safety instructions on the device . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 15
3 Product description . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 17
3.1 Product overview . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 17
3.1.1 Centrifuge 5702. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 17
3.1.2 Centrifuge 5702 R / RH . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 18
3.2 Delivery package . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 19
3.3 Features. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 19
3.4 Name plate . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 21
4 Installation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 23
4.1 Selecting the location . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 23
4.2 Preparing installation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 24
4.3 Installing the instrument . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 25
5 Operation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 27
5.1 Operating controls . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 27
5.2 Switching on the centrifuge . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 30
5.3 Replacing the rotor . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 30
5.3.1 Inserting the rotor . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 30
5.3.2 Removing the rotor . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 30
5.4 Loading a fixed-angle rotor. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 31
5.5 Loading a swing-bucket rotor . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 32
5.5.1 Inserting the bucket in the swing-bucket rotor. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 32
5.5.2 Performing an imbalance calibration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 32
5.5.3 Loading buckets symmetrically. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 33
5.6 Closing the centrifuge lid . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 35
Table of contents
4 Centrifuge 5702/5702 R/5702 RH
English (EN)
5.7 Centrifugation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 35
5.7.1 Centrifugation with time setting . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 35
5.7.2 Centrifuging in continuous operation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 37
5.7.3 Short run centrifugation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 37
5.7.4 Setting the soft ramp. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 38
5.7.5 Setting the start of the time counting (At set rpm function). . . . . . . . . . . . . . . . . . . . . 38
5.7.6 Calculating the speed of centrifugation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 39
5.8 Centrifuge 5702 R, Centrifuge 5702 RH: Heating and cooling. . . . . . . . . . . . . . . . . . . . . . . . . . 39
5.8.1 Setting the temperature . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 39
5.8.2 Temperature display . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 40
5.8.3 Temperature monitoring . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 40
5.8.4 Temperature control run FastTemp. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 40
5.8.5 Continuous cooling . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 41
5.8.6 Centrifuge 5702 RH: Temperature profiles . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 42
5.9 Aerosol-tight centrifugation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 43
5.9.1 Aerosol-tight centrifugation in a fixed-angle rotor . . . . . . . . . . . . . . . . . . . . . . . . . . . . 43
5.10 Switching off the centrifuge . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 44
6 Device settings . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 45
6.1 Changing the operating state . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 45
6.2 Key lock . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 45
6.2.1 Centrifuge 5702 R, Centrifuge 5702 RH: Securing the program against changes . . . . 46
6.2.2 Centrifuge 5702: Displaying the key lock status. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 46
6.3 Speakers . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 46
6.3.1 Display the status of the speakers. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 47
7 Programs . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 49
7.1 Creating and storing a program . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 49
7.2 Saving the current settings as a program . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 49
7.3 Calling up a program . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 50
7.4 Editing programs . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 50
7.5 Deleting programs . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 50
7.6 Exiting the program . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 50
8 Maintenance . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 51
8.1 Service. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 51
8.2 Preparing cleaning/disinfection . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 51
8.3 Cleaning/disinfection . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 52
8.3.1 Cleaning and disinfecting the device . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 53
8.3.2 Cleaning and disinfecting the rotor . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 53
8.4 Additional care instructions for refrigerated centrifuges. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 54
8.5 Cleaning glass breakage . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 55
8.6 Replacing the fuses . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 55
8.7 Decontamination before shipment . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 56
9 Troubleshooting . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 57
9.1 General errors . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 57
9.2 Error messages . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 58
9.3 Emergency release . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 61
Table of contents
Centrifuge 5702/5702 R/5702 RH 5
English (EN)
11 Technical data . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 65
11.1 Power supply. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 65
11.2 Ambient conditions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 65
11.3 Weight/dimensions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 66
11.4 Noise level. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 66
11.5 Application parameters . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 67
11.6 Acceleration and deceleration times. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 68
11.7 Service life for accessories . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 69
13 Ordering information . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 79
13.1 Rotor A-4-38 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 79
13.1.1 Rotor A-4-38 with round buckets . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 79
13.1.2 Rotor A-4-38 with rectangular buckets . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 80
13.2 Rotor A-8-17 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 80
13.3 Rotor F-45-24-11. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 81
13.4 Rotor F-35-30-17. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 81
13.5 Rotor F-45-18-17-Cryo . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 82
13.6 Fuses . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 82
Certificates . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 83
Table of contents
6 Centrifuge 5702/5702 R/5702 RH
English (EN)
Operating instructions
Centrifuge 5702/5702 R/5702 RH 7
English (EN)
1 Operating instructions
1.1 Using this manual
Read this operating manual completely before using the device for the first time. Observe the
instructions for use of the accessories where applicable.
This operating manual is part of the product. Please keep it in a place that is easily accessible.
Enclose this operating manual when transferring the device to third parties.
The current version of the operating manual for all available languages can be found on our webpage
www.eppendorf.com/manuals.
The safety instructions in this manual have the following danger symbols and danger levels:
Depiction Meaning
1. Actions in the specified order
2.
Actions without a specified order
• List
Text Display or software texts
Additional information
Operating instructions
8 Centrifuge 5702/5702 R/5702 RH
English (EN)
PCR
Polymerase Chain Reaction
rcf
Relative centrifugal force : g-force in m/s2
rpm
Revolutions per minute
UV
Ultraviolet radiation
Safety
Centrifuge 5702/5702 R/5702 RH 9
English (EN)
2 Safety
2.1 Intended use
The Centrifuge 5702/5702 R/5702 RH is designed for separating liquid substance mixtures with different
densities, in particular, for processing and analyzing samples from the human body in in-vitro diagnostic
applications to ensure that the in-vitro diagnostic device can be used according to its intended purpose.
This centrifuge including components is an in-vitro diagnostic device according to Directive 98/79/EC of
the European Parliament and the Council dated October 27, 1998.
Eppendorf centrifuges are exclusively intended for indoor use by trained specialists.
The device and accessories may only be operated by trained and skilled personnel.
Before using the device, read the operating manual and the instructions for use of the accessories carefully
and familiarize yourself with the device's mode of operation.
In the following cases, the designated protection of the device may be affected. The liability for any
resulting damage or personal injury is then transferred to the owner:
Do not operate the device in areas where explosive substances are handled.
Do not use this device to process any explosive or highly reactive substances.
Do not use this device to process any substances which may generate an explosive
atmosphere.
Due to its design and the environmental conditions inside the device, the Centrifuge 5702/5702 R/5702 RH
is not suitable for use in a potentially explosive atmosphere.
The device may only be used in a safe environment, such as in the open environment of a ventilated
laboratory or a fume hood. The use of substances that may contribute to a potentially explosive atmosphere
is not permitted. The final decision on the risks associated with the use of such substances lies with the
user.
Only switch on the device if the device and mains/power cord are undamaged.
Only operate devices which have been installed or repaired properly.
In case of danger, disconnect the device from the mains/power supply voltage. Disconnect
the mains/power plug from the device or the earth/grounded socket. Use the isolating
device intended for this purpose (e.g. the emergency switch in the laboratory).
Only connect the device to voltage sources which correspond with the electrical
requirements on the name plate.
Only use earth/grounded sockets with a protective earth (PE) conductor.
Only use the mains/power cord supplied.
Safety
Centrifuge 5702/5702 R/5702 RH 11
English (EN)
When handling infectious liquids and pathogenic germs, observe the national regulations,
the biosafety level of your laboratory, the material safety data sheets, and the
manufacturer's application notes.
Use aerosol-tight sealing systems for the centrifugation of these substances.
When working with pathogenic germs which belong to a higher risk group, more than one
aerosol-tight bioseal must be used.
Wear your personal protective equipment.
For comprehensive regulations about handling germs or biological material of risk group II
or higher, please refer to the "Laboratory Biosafety Manual" (source: World Health
Organization, Laboratory Biosafety Manual, the current edition).
Do not reach between the device and centrifuge lid when opening or closing the centrifuge
lid.
Do not reach into the locking mechanism of the centrifuge lid.
Open the centrifuge lid fully to ensure that the centrifuge lid cannot slam shut.
Wait for the rotor to stop before activating the emergency release.
To check, look through the monitoring glass in the centrifuge lid.
After installing the device, wait for at least 3 h. Only then connect the device to the mains/
power line.
NOTICE! Damage from knocking against or moving the device during operation.
If the rotor hits against the rotor chamber wall, this will cause considerable damage to the
device and rotor.
WARNING! Risk of injury from improperly attached rotors and rotor lids.
Only centrifuge with the rotor and rotor lid firmly tightened.
If unusual noises occur when the centrifuge starts, the rotor or the rotor lid may not be
attached properly. Stop the centrifugation immediately.
Avoid using aggressive chemicals such as strong and weak alkalis, strong acids, solutions
with mercury ions, copper ions and other heavy metal ions, halogenated hydrocarbons,
concentrated saline solutions and phenol.
If it is contaminated by aggressive chemicals, clean the rotor and especially the rotor bores
immediately using a neutral cleaning agent.
Due to the manufacturing process, color variations may occur on PTFE coated rotors.
These color variations do not affect the service life or resistance to chemicals.
Always tighten the rotor with the enclosed Upper side of the device,
rotor key. under the centrifuge lid.
3 Product description
3.1 Product overview
3.1.1 Centrifuge 5702
Abb. 3-1: Centrifuge 5702
1 2 3
speed top
press rpm/s
time
press start/stop
open
short
ge 5702
Centrifu
6 5 4 7
1 2 3
prog
speed
press rpm/rcf
time
press start/stop
temp
open
fast
short temp
8
R
ge 5702
Centrifu
6 7 5 4
3.3 Features
The low-speed centrifuge family 5702 has been especially developed for cell culture laboratories and
clinical research laboratories with a low to medium throughput. Due to their compact design these
centrifuges fit on nearly all types of lab bench and their quiet operation improves your work environment.
They can be operated with six different rotor options to accommodate nearly all types of tubes.
Product features
• Maximum speed: 3000 × g (4 400 rpm)
• Very compact footprint – fits on any lab bench
• Very quiet operation for a better work environment
• Low height of the device makes loading and unloading of samples easier
• SOFT braking function for slow acceleration and braking. Optimized for cell separation using gradient
centrifugation
• The At set rpm function starts the timer when the selected speed has been reached; for reproducible
centrifugation
• Key lock prevents unintentional adjustments
• Stainless steel rotor chamber is rust-proof and easy to clean
• Electronic imbalance detection for maximum safety
Product description
20 Centrifuge 5702/5702 R/5702 RH
English (EN)
1 2 3 4
14 Made in 11
Germany
13 12
Mark for waste electrical and electronic equipment in accordance with the EN
50419 standard in accordance with directive 2012/19/EU (WEEE) of the
European Union
RoHS mark in accordance with standard SJ/T 11364, Marking for the restriction
of the use of hazardous substances in electrical and electronic products, People's
Republic of China
EAC marking for medical devices; conformity with the technical regulations of
the Eurasian Economic Union
MED
Conformity with the relevant directives for the Eurasian Economic Union
Installation
Centrifuge 5702/5702 R/5702 RH 23
English (EN)
4 Installation
4.1 Selecting the location
Only connect the device to voltage sources which correspond with the electrical
requirements on the name plate.
Only use earth/grounded sockets with a protective earth (PE) conductor.
Only use the mains/power cord supplied.
NOTICE! If an error occurs, objects in the immediate vicinity of the device may become
damaged.
Do not install the device near heat sources (e.g. heating, drying cabinet).
Do not expose the device to direct sunlight.
Ensure unobstructed air circulation. Maintain a clearance of at least 30 cm around all
ventilation gaps.
The mains/power switch and the disconnecting device of the mains/power line must be easily
accessible during operation (e.g. a residual current circuit breaker).
Installation
24 Centrifuge 5702/5702 R/5702 RH
English (EN)
Do not use this device near strong electromagnetic sources (e.g. unshielded high frequency sources) as
they could impede proper functioning of the device.
Transport and lift the device with an adequate number of helpers only.
Use a transport aid for transporting the device.
The device may only be stored and transported in its original packing.
Retain the original packing, transport straps, packing material and transport securing
devices.
Do not cut up the transport straps.
Only connect the device to voltage sources which correspond with the electrical
requirements on the name plate.
Only use earth/grounded sockets with a protective earth (PE) conductor.
Only use the mains/power cord supplied.
After installing the device, wait for at least 3 h. Only then connect the device to the mains/
power line.
NOTICE! Centrifuge 5702 R, Centrifuge 5702 RH: Compressor damage after improper
transport.
Prerequisites
• The device has been prepared in accordance with the operating manual.
• The device has adapted to the ambient temperature (waiting time 3 h).
• The compressor is ready for operation (waiting time 4 h).
1. Connect the mains/power cord of the centrifuge to the mains/power supply.
2. Switch on the centrifuge using the mains/power switch.
• The standby key lights up green.
• The display is active.
5 Operation
5.1 Operating controls
Abb. 5-1: Centrifuge 5702 operating controls
time speed
press start/stop press rpm/rcf
short open
6 5 4 3 2
1 2 3 4
1 prog 2
time speed
press start/stop press rpm/rcf
temp
1
fast
short open
temp
10 9 8 7 6 5
min rpm
sec rcf
1 2 3 4
min C rpm
sec rcf
1 2 3 4 5 6 7 8
Only Centrifuge 5702 R: When setting the soft ramp, the symbol only appears on the
display as of serial number 03556 and higher.
For devices with serial numbers < 03556 (see Setting the soft ramp on p. 38).
Operation
30 Centrifuge 5702/5702 R/5702 RH
English (EN)
Prerequisites
• The device has been installed in accordance with the operating manual.
1. Switch the centrifuge on using the mains/power switch.
2. Press the Standby key, if required.
The display shows the parameters of the last run.
3. Press the open key to open the closed centrifuge lid.
1. Turn the rotor nut counterclockwise using the rotor key supplied.
2. Remove rotor by lifting it vertically.
Operation
Centrifuge 5702/5702 R/5702 RH 31
English (EN)
1. Check maximum load (adapter, vessel, and contents) for each rotor bore.
2. Load rotors and adapters only with the tubes intended for them.
3. To ensure symmetrical loading, insert sets of two tubes in opposite bores.
Tubes located opposite each other must be of the same type and contain the same filling quantity.
To minimize weight differences between filled sample tubes, we recommend taring with a balance. This
will reduce wear of the drive and operating noise.
Operation
32 Centrifuge 5702/5702 R/5702 RH
English (EN)
Prerequisites
• The combination of rotor, bucket, adapter and tube has been approved by Eppendorf.
• Buckets that are located opposite each other belong to the same weight class. The weight class is
engraved in the sides of the groove: e.g., 68.
• The grooves of the buckets are clean and slightly coated with pivot grease.
1. Check the maximum load (adapter, tube, and contents) for each bucket. Check the length of the tubes.
On each rotor there is an indication of the weight that a fully loaded bucket must not exceed.
2. Insert the bucket into the rotor. Load the rotor symmetrically.
All rotor positions must be equipped with buckets.
Only place buckets with the same weight class in opposite locations.
3. Check to see if all buckets are completely hung and can freely swing out.
If you use tubes or plates for the first time, perform a swing-out test.
To check whether the buckets can swing out, you can perform a manual swing-out test. The maximum
speed of centrifugation is 1000 rpm.
The loading shown on the right-hand side is incorrect as it places an uneven load on the pegs of the rotor.
To reduce vibrations and noise, load all buckets of the swing-buckets rotor equally.
Operation
34 Centrifuge 5702/5702 R/5702 RH
English (EN)
If rotor lids or caps have come into contact with organic solvents, clean them immediately.
Regularly check the rotor lids and caps for damage and cracks.
Immediately replace rotor lids or caps that show cracks or milky stains.
If you are using the cap, do not centrifuge any substances which contain phenol or
chloroform.
1 Sealing ring
Do not reach between the device and centrifuge lid when opening or closing the centrifuge
lid.
Do not reach into the locking mechanism of the centrifuge lid.
Open the centrifuge lid fully to ensure that the centrifuge lid cannot slam shut.
5.7 Centrifugation
Prerequisites:
• The centrifuge is switched on.
• The rotor has been inserted and attached correctly.
• The rotor has been loaded correctly.
• Buckets can swing out freely.
• The centrifuge lid is closed.
WARNING! Risk of injury from improperly attached rotors and rotor lids.
Only centrifuge with the rotor and rotor lid firmly tightened.
If unusual noises occur when the centrifuge starts, the rotor or the rotor lid may not be
attached properly. Stop the centrifugation immediately.
6. To maintain the temperature in the rotor chamber, the centrifuge lid of Centrifuge 5702 R and
Centrifuge 5702 RH remains closed. Press the open key to open the lid.
The centrifuge lid of Centrifuge 5702 opens automatically.
The short spin centrifugation runs as long as the short key is pressed. The centrifugation is performed at
maximum rotational speed of the rotor.
1. Centrifuge 5702 R, Centrifuge 5702 RH: Use the temp arrow keys to set the temperature.
2. To start the short run centrifugation, press and hold the short key.
3. Release the short key to end short run centrifugation.
During the braking process, the centrifugation time flashes on the display.
4. To maintain the temperature in the rotor chamber, the centrifuge lid ofCentrifuge 5702 R and Centrifuge
5702 RH remains closed. Press the open key to open the lid.
The centrifuge lid of Centrifuge 5702 opens automatically.
Operation
38 Centrifuge 5702/5702 R/5702 RH
English (EN)
For the Centrifuge 5702/5702 R/5702 RH, 2 settings for soft ramps are available. Use the soft ramps for
sensitive applications.
Prerequisites
• The centrifuge lid is open.
1. To check which soft ramps are set, briefly press the short key.
The display shows the set soft ramps.
Soft ramps Centrifuge 5702 Centrifuge 5702 R Centrifuge 5702 R Centrifuge 5702 RH
serial number < 03556 serial number > 03556
Fast br on br on No symbol No symbol
Slow br OF br OF
5.7.5 Setting the start of the time counting (At set rpm function)
Prerequisites
• The centrifuge lid is open.
The g-force that is shown on the display is standardized to the A-4-38 rotor with conical tubes 15 mL
without adapter. If you use other rotors and adapters, other g-forces are reached.
The maximum g-force and maximum radius for the respective rotors and adapters can be
found in chapter "Rotors, tubes, adapters"(see Rotors, tubes and adapters on p. 71).
To calculate the g-force, use the following formula in accordance with DIN 58 970:
• rcf = 1.118 · 10-5 · n2 · rmax
– rcf: g-force
– n: rotational speed in rpm
– rmax: maximum centrifugation radius in cm
Example 1
• The adapter for HPLC vials in the F-45-18-17-Cryo rotor has a maximum radius of 8.3 cm.
• At 6 142 rpm, a max. g-force of 3 500 × g is reached.
Example 2
• The 100 mL adapter has a maximum radius of 13.5 cm.
• At 4,000 rpm, a maximum g-force of 2,415 × g is reached.
With Centrifuge 5702 R, the rotor chamber can be cooled. With Centrifuge 5702 RH, the rotor chamber can
be heated and cooled.
The temperature that can actually be reached depends on the rotor and the set rotational
speed.
In case of a rotor stop (continuous cooling), cooling takes longer than during centrifugation or
a temperature control run.
At higher ambient temperatures, a brief fan noise is possible until the set temperature has
been reached. Fan noise indicates a heavy cooling performance.
At ambient temperatures of < 18 °C, a running-in period of approx. 1 h is required for proper
operation.
Prerequisites
• The centrifuge is switched on.
1. Use the temp arrow keys to set the target temperature.
2. Set the centrifugation time and speed of centrifugation.
3. To start the centrifugation, press the time rotary knob.
Temperature during rotor stop: The display alternately shows the set temperature
(long) and the actual temperature (short)
Temperature display during centrifugation: actual temperature
After the set temperature has been reached, the centrifuge reacts to temperature deviations during
centrifugation as follows:
With the FastTemp function, you can immediately start a temperature run without samples, at rotor-specific
or temperature-specific speeds. This will quickly bring the rotor chamber, including rotor and adapter, up
to the set temperature.
Prerequisites
• The centrifuge is switched on.
• Rotor, rotor lid and adapter are correctly mounted.
• The centrifuge lid is closed.
1. Set the speed of centrifugation for the following run.
2. Use the arrow keys to set the target temperature.
3. Press the fast temp key.
Operation
Centrifuge 5702/5702 R/5702 RH 41
English (EN)
If you perform a temperature control run and close the buckets with aerosol-tight caps, negative pressure
builds up in the buckets. After the temperature control run, the caps cannot be removed.
The continuous cooling keeps the set temperature in the rotor chamber during rotor stop.
• During continuous cooling, the set temperature is displayed.
• Irrespective of the set temperature, the temperature does not go below 4 °C to prevent the rotor
chamber from freezing and to prevent condensation.
• In case of a rotor stop, temperature control takes longer than during centrifugation or a temperature
control run.
• Continuous cooling stops after 8 h.
Prerequisites
• The centrifuge is switched on.
• The centrifuge lid is closed.
• The set temperature is lower than the ambient temperature.
1. Continuous cooling starts automatically.
Operation
42 Centrifuge 5702/5702 R/5702 RH
English (EN)
The Centrifuge 5702 RH has a controlled heating and refrigeration system. This makes it possible to exactly
temper sensitive samples.
A special temperature profile is stored in the software for each rotor. The temperature profile defines the
speed at which the rotor performs the FastTemp temperature control run. The aim is to temper the rotor
chamber and the rotor to the set temperature in the shortest possible time. The temperature is to be
maintained, with small tolerances.
Prerequisites
• The centrifuge lid is open.
1. Press the fast temp key.
The last selected temperature profile is displayed.
Display Rotor
ro F 35 Rotor F-35-30-17
ro F 24 Rotor F-45-24-11
ro F 18 Rotor F-45-18-17-Cryo
ro A4 rE Rotor A-4-38 with rectangular buckets
ro A4 ro Rotor A-4-38 with round buckets
ro A8 Rotor A-8-17
ro AL L Temperature profile for all rotors
2. Use the arrow keys to select the temperature profile for the rotor used.
The selected temperature profile is adopted after 3 s. The display shows the default values again.
Operation
Centrifuge 5702/5702 R/5702 RH 43
English (EN)
WARNING! Damage to health due to limited aerosol tightness with an incorrect rotor/
rotor lid combination.
Aerosol-tight centrifugation is guaranteed only if the rotors and rotor lids intended for this
purpose are used. The designation of aerosol-tight fixed-angle rotors always starts with FA.
The aerosol-tight rotors and rotor lids of this centrifuge are additionally marked with a red
ring on the rotor and a red rotor lid screw.
Always use rotors and rotor lids marked aerosol-tight together for aerosol-tight
centrifugation. The details specifying in which centrifuge the aerosol-tight rotors and rotor
lids may be used can be found on the rotor and on the top of the rotor lid.
Only use aerosol-tight rotor lids in combination with the rotors that are specified on the
rotor lid.
Only use aerosol-tight buckets with the corresponding caps.
Check the integrity of the seals of the aerosol-tight rotor lids or caps before each use.
Only use aerosol-tight rotor lids or caps if the seals are undamaged and clean.
Do not exceed temperatures of 121 °C or a time of more than 20 min. while autoclaving.
After each proper autoclaving process (121 °C, 20 min.), coat the threads of the rotor lid
screw with a thin layer of pivot grease (order no. Int. 5810 350.050, North America
022634330).
Replace aerosol-tight rotor lids without replaceable seals after 50 autoclaving cycles.
Only the seal of aerosol-tight rotor lids with exchangeable seals (e.g. QuickLock rotor lids)
must be replaced after 50 autoclaving cycles.
Replace aerosol-tight rotor caps after 50 autoclaving cycles.
Never store aerosol-tight rotors or buckets closed.
The aerosol tightness of rotors, rotor lids, buckets and caps has been tested and certified
according to Annex AA of IEC 61010-2-020.
6 Device settings
6.1 Changing the operating state
The centrifuge has 2 operating states, the ready state and standby mode. You can actively change between
the two operating states.
Prerequisites
• The centrifuge is not performing a run.
Press the Standby key to change the operating state.
The operating state changes. The Standby key changes color.
If the key lock is activated, the following centrifugation parameters cannot be changed:
• Centrifugation time
• Centrifuge 5702 R, Centrifuge 5702 RH: Temperature
• g-force or rotational speed
• Soft ramps
• Status of the At set rpm function
The following settings can be changed despite the key lock being activated:
• Start and stop centrifugation. Press the time rotary knob.
• Set the unit for the speed of centrifugation [rpm/ rcf]. Press the speed rotary knob.
Activating the key lock
Prerequisites
• The centrifuge lid is open.
Press the short and open keys simultaneously for > 5 s.
The centrifugation parameters cannot be changed.
Device settings
46 Centrifuge 5702/5702 R/5702 RH
English (EN)
Prerequisites
• The centrifuge lid is open.
Press the short and open keys simultaneously for > 5 s.
The centrifugation parameters cannot be changed.
6.2.1 Centrifuge 5702 R, Centrifuge 5702 RH: Securing the program against changes
On the Centrifuge 5702 R and the Centrifuge 5702 RH, the key lock status is shown on the display.
Prerequisites
• The centrifuge lid is open.
Shortly press the short and open keys simultaneously.
The display shows the current status of the key lock.
6.3 Speakers
Prerequisites
• The centrifuge lid is open.
Press the open key for > 2 s.
The speakers are switched on.
Switching off the speakers
Prerequisites
• The centrifuge lid is open.
Press the open key for > 2 s.
Device settings
Centrifuge 5702/5702 R/5702 RH 47
English (EN)
This function is only available for the Centrifuge 5702. On the Centrifuge 5702 R and the Centrifuge 5702
RH the status of the speakers is shown on the display.
Prerequisites
• The centrifuge lid is open.
Press the open key briefly.
The display shows the status of the speakers.
Device settings
48 Centrifuge 5702/5702 R/5702 RH
English (EN)
Programs
Centrifuge 5702/5702 R/5702 RH 49
English (EN)
7 Programs
You can store 2 programs each on the Centrifuge 5702 R and on the Centrifuge 5702 RH.
Prerequisites
• Rotor stop.
1. Use the rotary knob time to set the centrifugation time.
2. Use the temp arrow keys to set the temperature.
3. Use the rotary knob speed to set the speed of centrifugation.
4. Set the start of time counting (function At set rpm). Press rotary knob time > 2 s to do this.
5. Press the short key > 5 s to set the soft ramp.
6. Select the program slot. Press the prog 1 key or the prog 2 key > 2 s.
• A signal sounds.
• The program key does not flash any more. The program key lights up blue.
• The parameters of the program are saved.
Prerequisites
• Rotor stop.
Press the prog 1 key or the prog 2 key > 2 s.
• A signal sounds.
• The program key lights up blue.
• The parameters of the program are saved.
Programs
50 Centrifuge 5702/5702 R/5702 RH
English (EN)
Prerequisites
• Rotor stop.
Press the prog 1 key or prog 2 key to call up a program.
• The program key lights up blue.
• The display shows the parameters of the program.
If a program is loaded, the parameters cannot be changed. The following text appears in the
display - Pr 1 for program 1 or Pr 2 for program 2.
Prerequisites
• The program has been created and stored.
• The program has been called up. The program key prog 1 or prog 2 lights up blue.
1. Press the key again on which the program has been stored.
The program key does not light up any more.
The display shows the centrifugation parameters.
The centrifugation parameters can be adjusted.
2. Change the centrifugation parameters.
3. Store the program on the old program slot. Press the prog 1 key or prog 2 > 2 s to do this.
• A signal sounds.
• The program key does not flash any more. The program key lights up in blue.
• The parameters of the program are saved.
Prerequisites
• The program has been called up. The program key prog 1 or prog 2 lights up blue.
1. Press the prog 1key or the prog 2 key to exit the program.
• The program key does not light up any more.
• The display shows the centrifugation parameters.
• The centrifugation parameters can be adjusted.
Maintenance
Centrifuge 5702/5702 R/5702 RH 51
English (EN)
8 Maintenance
8.1 Service
Have the centrifuge's electrical safety, especially the paths for the protective connections,
checked every 12 months by trained and skilled personnel.
We recommend to have the centrifuge and the associated rotors checked by Technical Service during a
service at least every 12 months. Please note the country-specific regulations.
Clean all accessible surfaces of the device and the accessories at least weekly and when contaminated.
Clean the rotor regularly. This way the rotor is protected and the durability is prolonged.
Furthermore, observe the notes on decontamination (see Decontamination before shipment on p. 56)
when the device is sent to the authorized Technical Service for repairs.
The procedure described in the following chapter applies to the cleaning as well as to the disinfection or
decontamination. The table below describes the steps required on top of this:
Cleaning Disinfecting/decontamination
1. Use a mild cleaning fluid to clean the accessible 1. Choose the disinfection method which
surfaces of the device and the accessories. corresponds to the legal regulations and
2. Carry out the cleaning as described in the guidelines in place for your range of application.
following chapter. For example, use alcohol (ethanol, isopropanol)
or alcohol-based disinfectants.
2. Carry out the disinfection or decontamination as
described in the following chapter.
3. Then clean the device and the accessories.
If you have any further questions regarding the cleaning and disinfection or decontamination
or regarding the cleaning fluid to be used, contact the Eppendorf AG Application Support. The
contact details are provided on the back of this manual.
Maintenance
52 Centrifuge 5702/5702 R/5702 RH
English (EN)
8.3 Cleaning/disinfection
Switch off the device and disconnect it from the mains/power line before starting cleaning
or disinfection.
Do not allow any liquids to penetrate the inside of the housing.
Do not perform a spray clean/spray disinfection on the housing.
Only reconnect the device to the mains/power line when it is completely dry, both inside
and outside.
Do not use any aggressive chemicals on the device or its accessories, such as strong and
weak bases, strong acids, acetone, formaldehyde, halogenated hydrocarbons or phenol.
If the device has been contaminated by aggressive chemicals, clean it immediately using a
mild cleaning agent.
Do not use any corrosive cleaning agents, aggressive solvents or abrasive polishes.
Do not incubate the accessories in aggressive cleaning agents or disinfectants for longer
periods.
Do not use UV, beta, gamma, or any other high-energy radiation for disinfection.
Avoid storage in areas with strong UV radiation.
Autoclaving
Fixed-angle rotors and adapters can be autoclaved (121 °C, 20 min).
Rotor crosses of swing-bucket rotors cannot be autoclaved.
After a maximum of 50 autoclaving cycles, the aerosol-tight caps must be exchanged.
Maintenance
Centrifuge 5702/5702 R/5702 RH 53
English (EN)
Cleaning agent:
• alcohol 70% (ethanol, isopropanol)
• mild neutral cleaning agent
• lint-free cloth
1. Open the lid.
2. Switch off the device and disconnect it from the voltage supply.
3. Remove the rotor.
4. Clean and disinfect all accessible surfaces of the device, including the mains/power cord, using a damp
cloth and the cleaning agents.
5. Thoroughly clean the rubber seal of the rotor chamber with water.
6. Let the rubber seal dry off.
7. Rub the dry rubber seal with glycerol or talcum. This prevents the rubber seal from becoming brittle.
Other components of the device, such as the motor shaft and rotor cone, must not be lubricated.
8. Clean the motor shaft with a soft, dry, lint-free cloth.
9. Check the motor shaft for damage.
10. Check the device for corrosion and damage.
11. Leave the centrifuge lid open when the device is not in use.
12. Only connect the device to the mains/power supply if it is completely dry on the inside and outside.
1. Inspect the rotor and accessories for damage and corrosion. Do not use damaged rotors or accessories.
2. Clean and disinfect the rotors and accessories with the recommended cleaning agents.
3. Clean and disinfect the rotor bores with a bottle brush.
4. Rinse the rotors and accessories thoroughly with distilled water. Rinse the rotor bores of fixed-angle
rotors particularly thoroughly.
Do not immerse the rotor in liquid as liquid can get trapped inside the cavities.
5. Place the rotors and accessories on a towel to dry. Place the fixed-angle rotors with the rotor bores
facing down so the bores can dry.
6. Clean the rotor cone with a soft, dry, lint-free cloth. Do not lubricate the rotor cone.
7. Inspect the rotor cone for damage.
8. Place the dry rotor onto the motor shaft.
9. Tighten the rotor nut firmly by turning it clockwise with the rotor key.
10. If necessary, equip the fixed-angle rotor with the cleaned adapters.
11. Equip the swing-bucket rotor with the cleaned buckets and adapters.
Maintenance
54 Centrifuge 5702/5702 R/5702 RH
English (EN)
Empty and clean the condensation water tray regularly and especially after liquid spillage in the rotor
chamber. Pull out the condensation water tray at the front right under the device.
Empty and clean the condensation water tray regularly.Pull out the condensation water tray to the left
under the device.
Regularly clear the rotor chamber of ice formations via defrosting, either by leaving the centrifuge lid
open or carrying out a brief temperature control run at approx. 30°C.
To relieve the gas springs in the centrifuge lid, leave the centrifuge lid open when not in use for a long
period.
Residual moisture can escape.
Leave the centrifuge lid open when not in use for a long period.
Residual moisture can escape.The lid spring is relieved.
Wipe up condensation water in the rotor chamber. Use a soft absorbent cloth for this.
No later than every 6 months, remove any dust deposits from the ventilation gaps of the centrifuge
using a brush or swab.First switch off the centrifuge and remove the mains/power plug.
Maintenance
Centrifuge 5702/5702 R/5702 RH 55
English (EN)
When using glass tubes there is a risk of glass breakage in the rotor chamber. The resulting glass splinters
are swirled around in the rotor chamber during centrifugation and have a sandblasting effect on the rotor
and accessories. The smallest glass particles become lodged in the rubber parts (e.g., the motor guide, the
rotor chamber seal, and the rubber mats of adapters).
1. Switch off the device and isolate it from the mains/power supply.
2. Pull the fuse holder out of the device.
3. Replace the fuses.
Maintenance
56 Centrifuge 5702/5702 R/5702 RH
English (EN)
If you are shipping the device to the authorized Technical Service for repairs or to your authorized dealer
for disposal please note the following:
9 Troubleshooting
If you cannot remedy an error with the recommended measures, please contact your local Eppendorf
partner. The contact address can be found on the Internet at www.eppendorf.com.
If the centrifuge lid cannot be opened, you can activate the emergency release manually.
Wait for the rotor to stop before activating the emergency release.
To check, look through the monitoring glass in the centrifuge lid.
The emergency release consists of a cord with a plastic knob. The emergency release is located in the
bottom panel at the front right device foot.
Transport and lift the device with an adequate number of helpers only.
Use a transport aid for transporting the device.
10.2 Storage
10.3 Disposal
If the product needs to be disposed of, the relevant legal regulations must be observed.
Information on the disposal of electrical and electronic devices in the European Community:
Within the European Community, the disposal of electrical devices is regulated by national regulations
based on EU Directive 2012/19/EU pertaining to waste electrical and electronic equipment (WEEE).
According to these regulations, any devices supplied after August 13, 2005, in the business-to-business
sphere, to which this product is assigned, may no longer be disposed of in municipal or domestic waste. To
document this, they have been marked with the following marking:
Because disposal regulations may differ from one country to another within the EU, please contact your
supplier if necessary.
Technical data
Centrifuge 5702/5702 R/5702 RH 65
English (EN)
11 Technical data
11.1 Power supply
11.3 Weight/dimensions
The noise level was measured according to (DIN EN ISO 3745) frontally in a sound measuring room with
accuracy class 1 at a distance of 1 m from the device and at lab bench height.
The following table shows the approximate acceleration and deceleration times according to DIN 58 970 for
the rotors of the Centrifuge 5702/5702 R/5702 RH. All values given are reference values. Fluctuations may
occur depending on the condition of the device and the load.
Do not use accessories that have exceeded their maximum service life.
Eppendorf states the maximum service life of the rotors and accessories both in years and in the maximum
number of cycles. The decisive factor for the service life is which case occurs first, usually this is the
number of years in operation.
Each centrifugation run during which the rotor is accelerated and braked is counted as a cycle independent
of the speed and the duration of the centrifugation run.
All other rotors and rotor lids can be used during the entire service life of the centrifuge if the following
conditions are met:
• proper use
• recommended maintenance
• undamaged condition
The date of manufacture is stamped on the rotors and buckets in the format 03/15 or 03/2015 (= March
2015). On the inside of the plastic-rotor lids and aerosol-tight caps, the date of manufacture is stamped in
the form of a clock .
Only use rotors that are intended for use with the corresponding centrifuge.
Please note the manufacturer’s information on the centrifugation resistance of the sample tubes used
(maximum g-force).
Rotor A-4-38 Round bucket Aerosol-tight cap Max. load per bucket 190 g
5702 722.006 5702 721.000 (adapter, tube and contents):
5702 761.001
13 Ordering information
13.1 Rotor A-4-38
13.1.1 Rotor A-4-38 with round buckets
13.6 Fuses
Order no. Description
(International)
Fuse
5425 351.003 2.5 A T (230 V), set of 2
5425 353.006 5 A T (100V/120 V), UL, set of 2
5703 851.136 6.3 AT (100 V), 2 pieces