Professional Documents
Culture Documents
SMCS - 0761
The 6V6000 Dial Bore Gauge Group permits measurements to be taken in diameter
ranges of 30.0 to 50.8 mm (1.181 to 2.000"). The gauge has a depth range of 8.5 to 150
mm (.33 to 5.9") and is marked in intervals of .002 mm (.00008").
NOTICE
4. Ball Tool
6. Hex Wrenches
7. Storage Box
https://sis2.cat.com/#/detail?
keyword=smhs8037&tab=3&serialNumber=&infoType=15&serviceMediaNumber=SMHS8037&serviceIeSystemControlNumber=SMHS80370001
(1) The following procedure shows how to set dial bore gauge (1) using quick-setting
wrench (2) and setting fixture (3). Quick-setting wrench (2) should not be used as a
setting standard. However, the wrench does give a means of presetting the gauge to
within approximately 0.05 mm (.002") of the standard size. The wrench also makes it
easier to adjust the gauge points.
(2) Before setting the gauge, learn the following characteristics of quick setting wrench
(2) and the gauge point and lock nut assembly on dial bore gauge (1).
(3) The lock nut drive slot in wrench (2) has been machined off center so there is a small
lug (A) and large lug (B). The gauge point drive slot has also been machined this way to
make small lug (C) and large lug (D). Large lug (B) is also in alignment with micrometer
scale (E) on the wrench.
(4) On the gauge point and lock nut assembly, the flats on both the lock nut and the end
of the gauge point have been machined off center. This makes large flat (F) and small flat
(G) on the gauge point and large flat (H) and small flat (J) on the lock nut. IMPORTANT:
The lock nut must always be installed with large chamfer (K) toward the gauge head.
(5) Find the correct gauge point using the table below. For example, if the bore to be
measured has a diameter of 34.4 mm (1.3543"), the gauge point from vial no. 1 should
be used.
https://sis2.cat.com/#/detail?
keyword=smhs8037&tab=3&serialNumber=&infoType=15&serviceMediaNumber=SMHS8037&serviceIeSystemControlNumber=SMHS80370001
NOTE: Although the gauge point in vial no. 4 has a range up to 50.8 mm (2.0000"), the
setting fixture can only be calibrated to 50.0 mm (1.9685").
(6) Install the gauge point in dial bore gauge (1) as follows:
a) Slide adjusting knob (4) all the way out to the maximum open position. Hold wrench (2)
so the micrometer scale is up as shown. With large flat (H) of lock nut (5) up, put gauge
point (6) in position in wrench (2).
b) Slide adjusting knob (4) inward while turning it slowly until the drive slot in the wrench
engages flats (F) and (G) of the gauge point. Hold the wrench at a small angle, with
gauge point (6) upward. Turn gauge point (6) into dial bore gauge (1) until lock nut (5)
comes in contact with the head of the dial bore gauge. Do not tighten the lock nut.
c) While holding lock nut (5) in this position, turn adjusting knob (4) to the desired setting.
Keep a light inward pressure on knob (4) while making this setting. Always come to the
final setting while turning knob (4) clockwise [as seen from same end of wrench as knob
(4)].
d) Tighten lock nut (5) by turning locking knob (7) and adjusting knob (4) at the same
time while keeping just enough tension on the adjusting knob to hold the micrometer
setting.
NOTE: After the final calibration of the dial bore gauge, the "0" point of the indicator hand
may not be in the most desired reading position. To change this position, put wrench (2)
on the gauge point and look at the micrometer reading. Loosen lock nut (5) and turn
adjusting knob (4) until the micrometer reading changes the same amount as the
indicator hand is to be moved.
https://sis2.cat.com/#/detail?
keyword=smhs8037&tab=3&serialNumber=&infoType=15&serviceMediaNumber=SMHS8037&serviceIeSystemControlNumber=SMHS80370001
(7) Use nest (8) from the 6V6030 Setting Fixture Group to calibrate the 6V6000 Dial Bore
Gauge. Nest (8) has a diameter range of 30.0 to 50.8 mm (1.181" to 2.000"). Put nest (8)
in position on the setting fixture group as shown. Make sure the spindle lock is not
engaged. Turn micrometer spindle (9) counterclockwise to a point approximately 0.5 mm
(0.020") more than the size being set.
(8) Carefully put gauge (1) in nest (8) with the gauge point toward the micrometer spindle
and keyway (10) in alignment with key (11) in the nest. Turn spindle (9) in toward the
gauge until the micrometer reads the exact size being set. Tighten the spindle lock. The
spindle should now be in contact with the gauge point. IMPORTANT: Always set the
micrometer by turning the spindle toward the gauge. If the spindle is turned in past the
desired setting, turn the spindle out and then back in again to the correct setting.
(9) Twist nest (8) in the horizontal plane until indicator hand (12) shows the minimum or
"most minus (-)" reading. The indicator hand should be within approximately 30° of the 12
o'clock position on the dial. If it is not, use the quick setting wrench to adjust the position
of the gauge point as shown in step 6 on page 7.
(10) If the indicator hand is within 30° of the 12 o'clock position, loosen screw (13) and
move the dial so the zero on the scale is in alignment with the indicator hand. Tighten
screw (13). Check to make sure the gauge has not moved during this setting. If the
gauge has not moved, it is now set to size. If the gauge has moved, again follow steps 9
and 10.
https://sis2.cat.com/#/detail?
keyword=smhs8037&tab=3&serialNumber=&infoType=15&serviceMediaNumber=SMHS8037&serviceIeSystemControlNumber=SMHS80370001
(1) Put the 3/32" hex wrench (1) into the socket of the centralizer adjusting cam (2). Turn
the wrench and cam until centralizer points (3) are fully retracted. With the wrench in the
socket, put the gauge head approximately 9 mm (.35") into the bore being measured.
(2) With the gauge in a vertical position as shown, turn centralizer adjusting cam (2) until
the centralizer points are firmly in contact with the wall of the bore.
NOTE: To retract the centralizer points and indicating finger (4), push in on retractor lever
(5).
(1) When centralizer points (1) become worn, they must be adjusted to keep accuracy.
Always adjust both points at the same time. Do this by putting the long leg of 3/32" hex
wrench (2) through both centralizer points, and turning both points at the same time until
new wear surfaces are in contact with the wall of the bore.
(1) If necessary, the indicator housing can be turned up to 90° in relation to the gauge
head using the following procedure. Put the 3/32" hex wrench (1) into access hole (A) for
the locking screw. Loosen the locking screw 1/2 no more than turn (180°). Turn indicator
housing (2) to the desired position. Push gauge head (3) and indicator housing (2) toward
https://sis2.cat.com/#/detail?
keyword=smhs8037&tab=3&serialNumber=&infoType=15&serviceMediaNumber=SMHS8037&serviceIeSystemControlNumber=SMHS80370001
each other as far as possible, and lock them into position by tightening the locking screw
[while this is being done, the indicator needle should move clockwise to approximately
the 8 o'clock position].
NOTE: Counter hand (4) shows which turn or revolution indicator hand (5) is on. When
the dial bore gauge is correctly assembled and is at rest out of the bore, indicator hand
(5) should be between the 7 o'clock and 9 o'clock positions with counter hand (4) just to
the right of zero.
(1) When the indicating finger ball becomes flat or worn, a replacement can be installed,
or the ball can be turned to a new position. Put a clean cloth under the gauge during
replacement of the ball. Put indicating finger clamp (1) in position over indicating finger
(2) as shown. Clamp (1) prevents movement of finger (2), which makes it easier to
remove the ball and install a replacement.
(2) Push the blade of ball tool (3) under the ball through the slot in the indicating finger
which intersects the ball socket. Pry up on the ball until it comes out of the socket. Ball (4)
will be held between the blade and the spring clip of ball tool (3).
https://sis2.cat.com/#/detail?
keyword=smhs8037&tab=3&serialNumber=&infoType=15&serviceMediaNumber=SMHS8037&serviceIeSystemControlNumber=SMHS80370001
(3) Turn ball (4) to a new position while it is being held in the ball tool [or remove ball (4)
and install a new ball in the ball tool]. Push ball (4) back into the socket with ball tool (3).
Make sure the ball is fully in the socket by pushing firmly against the ball with the back of
ball tool (3).
(4) Slide the ball tool out of the indicating finger. Remove indicating finger clamp (1).
Check the calibration of the gauge; see SETTING THE DIAL BORE GAUGE, starting
with step 7 on page 9.
(1) When the gauge point ball becomes flat or worn, a replacement ball can be installed,
or the ball can be turned to a new position using the following procedure. Remove gauge
point (1) from the dial bore gauge. Put a clean cloth under the gauge point during ball
replacement.
(2) Put the blade of ball tool (2) in position in the slot under the ball. Put your thumb over
the ball, and slowly pry up with the ball tool until the ball comes out.
(3) Turn ball (3) to a new position while it is being held in the ball tool [or remove ball (3)
and install a new ball in the ball tool]. Push ball (3) back into gauge point (1) with tool (2).
Make sure the ball is fully in the gauge point. Check the calibration of the gauge; see
SETTING THE DIAL BORE GAUGE, starting with step 7 on page 9.
https://sis2.cat.com/#/detail?
keyword=smhs8037&tab=3&serialNumber=&infoType=15&serviceMediaNumber=SMHS8037&serviceIeSystemControlNumber=SMHS80370001
(2) Install gauge point No. 1 [diameter range 30.0 to 35.2 mm (1.1811 to 1.3858")] in the
dial bore gauge. Use quick setting wrench (1) to set the gauge at 30.2 mm (1.189"); see
step 6 on page 7 of the topic, SETTING THE DIAL BORE GAUGE. Put dial bore gauge
(2) in position in the master ring. If wrench (1) is calibrated correctly, indicator hand (3)
should be within 30° of the 12 o'clock position on the dial. If the indicator hand is not
within this range, go to the next step.
(3) Put gauge (2) in master ring (4), and set the gauge to read zero. The scale should be
set so the zero position is at 12 o'clock. Put wrench (1) in position on gauge (2), and
loosen adjusting knob set screw (5) with a 5/64" hex wrench. Turn adjusting knob (6) so
the micrometer scale reads exactly 30.2. Tighten set screw (5) with knob (6) at the 30.2
reading. Check gauge (2) in master ring (4) to make sure the gauge is still set at 30.2 mm
(1.189").
PPI-00076472
2020/10/07
09:07:10+07:00
SMHS80370001
© 2020 Caterpillar Inc.
https://sis2.cat.com/#/detail?
keyword=smhs8037&tab=3&serialNumber=&infoType=15&serviceMediaNumber=SMHS8037&serviceIeSystemControlNumber=SMHS80370001