PRACTICAL 2:
DETERMINATION OF VISCOSITY OF FOOD PRODUCTS
USING A VISCOMETER.
INTRODUCTION
OBJECTIVE
1
PROCEDURE
2
RESULT
Viscometer Model: LV
Spindle#:1
Speed (rpm)
Sample
6 (x10) 12 (x5) 30 (x2) 60 (x1)
Red 1 325.0 242.5 154.0 100.0
Horse 2 330.0 237.5 153.0 100.0
Viscosity Average ± Std Dev 327.5 ± 2.5 240.0 ± 2.5 153.5 ± 0.5 100.0 ± 0.0
Viscometer Model: LV
Speed (rpm) : 6
Spindle#
Sample
1 (x10) 2 (x50) 3 (x200) 4 (x1000)
3
CALCULATION
1. Effect of speed on viscosity
Viscosity of sample using spindle 1
Viscosity (centipoise) = Dial reading × Factor
Speed
6.0 12.0 30.0 60.0
(rpm)
Viscosity 327.5 x10 240.0 x 5 153.5 x 2 100.0 x 1
(Cp) = 375 = 1200 = 307 = 100.0
1000
800
600
400
200
0
6 (x10) 12 (x5) 30 (x2) 60 (x1)
4
2. Effect of spindle number on viscosity
Speed = 6 rpm
Table 3.0 Viscosity of RED HORSE
Spindle # 1 2 3 4
Viscosity 327.5 x 10 325.0 x 50 300.0 200 x
x 200 1000
(Cp) = 3275 = 16250
= 60000 = 200000
Spindle # 1 2 3 4
Viscosity 715.0 x 10 762.5 x 50 850.0 x 950.0 x
200 1000
(Cp) = 7150 = 38125
= 170000 = 950000
5
DISCUSSION
Viscosity testing is one of the most important methods used to check the quality of
materials. Various industries rely on viscosity checks of their products to produce a product
with consistent texture. There are numerous instruments available to the food industry to
measure viscosity for quality control and thus ensure that products made are of consistent
quality. Bostwick consistometer is one of the simplest instrument that is used in the food
industry, which determines the consistency of a food by measuring the distance it flows under
its own weight. Typical food products measured include tomato ketchup, jams, etc. (Sahi, S.
2018). In this experiment, rotational viscometer which is LV Brookfield Viscometer was used
to measure the viscosity of food product.
From the result,in table 2.0, the viscosity of starjuice are 7150 cP, 38125 cP, 170000
cP, and 950000 cP for spindle #1, 2, 3 and 4, respectively. Meanwhile, in table 2.1, the
viscosity of redhorse are 3275 cP, 16250cP, 60000 cP, and 200000 cP for spindle #1,2,3 and
4 respectively. This results shows that starjuice has higher viscosity the redhorse and it is
indicated that starjuice is more concentrated than starjuice. This is due to the increasing of
spindle size and rotational speed of starjuice which then lead to greater resistence. The result
can be seen as shown in figure 3.2. According to literature review, the viscosity of standard
fluid is 12,257 cP. Its accuracy is ± 1% of 12,257 or ± 122.57 cP. Therefore, any viscosity
reading between 11,634.4 and 12,879.6 cP indicates that the viscometer is operating correctly.
Any reading outside these limits may indicate a viscometer problem (EJB, 2013).
Nevertheless, the value in the result obtained may be affected by some errors. There
are different problems, and most important of them include inconstant temperature during
viscosity measurements, phase changes and large particles in fluid. The temperature should
be tightly controlled during conducting the experiment because the viscosity of food can be
highly affected by temperature (Kassim. A. Abas,et al, 2010). The speed of the rotation and
6
the amount of time the spindle is rotating are other considerations when performing viscometer
testing. Furthermore, during calibration, the spindle also must rotate at least five (5) times
before viscosity reading is taken.
For a better reproducibility, the same spindle and speed combination should be used
for multiple tests. If the viscosity of the sample is unknown, the viscosity is tested by taking
the smallest spindle first and replacing it in ascending order by the next larger spindle until a
valid measurement result is achieved. To obtain a valid measurement the torque value must
be between 10 % and 100 %. If the torque value is higher than 100 %, a smaller spindle has
to be used. If the torque value is lower than 10 %, a bigger spindle has to be used. The higher
the torque value the better the accuracy is, since the accuracy of the measuring system
depends on the full scale range (EJB, 2013)
CONCLUSION
In conclusion, the viscosity of the sample had been determined based on different
speed and spindle, and the application of viscometer was also determined. Starjuice has
higher viscosity than Redhorase therefore it is more concentrated than Redhorse. The lowest
viscosity range can be covered by measuring with the biggest spindle at maximum speed. The
highest viscosity range can be covered by measuring with the smallest spindle at the lowest
speed.
REFERENCES
E.JB. (November 18, 2013). Brookfield Dial Viscometer Operating Instructions, pp:85-150
7
Sahi, S. (2018, June 19). Viscosity measurements in food products and manufacturing - New
Food. Retrieved November 02, 2020, from
https://www.newfoodmagazine.com/article/15042/viscosity-measurements-food-
products/
Vessele, M. (2019, June 28). Viscosity Testing's Role in Maintaining High-Quality Food
Production. Retrieved November 02, 2020, from https://www.coleparmer.com/tech-
article/viscosity-in-food-manufacturing