You are on page 1of 3

Determinants of Milk Flow Through Nipple Units

Role of Hole Size and Nipple Thickness


Oommen P. Mathew, MD

\s=b\ The aim of the present study was to fed with nipple units designed for pre¬ MATERIALS AND METHODS
elucidate the role of hole size and thick- term infants. However, the functional
ness in determining milk flow through nip- Nipple units commonly used in the feeding
superiority of these types of nipples of newborn infants formed the basis of this
ple units during bottle feeding. Commonly over other types for these infants has
used standard nipple units (SMA single\x=req-\ study. These nipple units were manufac¬
not been adequately documented. For tured and/or distributed by Mead-Johnson
hole, Enfamil single-hole, and Twist-on)
for term and preterm infants, as well as instance, the assumption has been made (Evansville, Ind), Ross Laboratories (Co¬
that preterm infants generate less pres¬ lumbus, Ohio), and Wyeth (Philadelphia,
Nuk-type nipple units (SMA Nuk, Enfamil
Natural, and Nuk) were tested. The size of sure per suck and the preterm nipple Pa), and consisted of standard nipple units
the nipple hole and wall thickness were units are therefore designed to deliver for term and preterm infants as well as the
determined for each nipple unit. Airflow higher flow compared with nipple units Nuk-type units. The general characteristics
of these nipple units have been reported
was measured by forcing pressurized air designed for term infants. However, it previously.1
through the feed hole. Simulated sucks is not clear whether there is any advan¬
The following tests and measurements
were used to measure the milk flow. A
marked variability in airflow and milk flow
tage in increasing the milk flow in pre¬ were conducted on each nipple: the size of the
term infants. On the contrary, higher nipple hole, the thickness at the tip of the
was observed within and among the vari-
milk flow may produce some adverse nipple, the airflow, and the milk flow. The
ous types of nipple units studied. Within
effects by altering the breathing pat¬ size of the nipple hole was calculated using a
each type of nipple unit, both milk flow and
airflow measurements correlated well tern significantly. profile projector (Mitutoyo Manufacturing
with hole size. The thickness of the nipple The milk flow through nipple units Co). The diameter of the hole was measured
units contributed minimally to the ob- can be altered by a number of factors: along the and y axes. The average of the
the characteristics of the nipple, the ri¬ two values obtained was taken as the diame¬
served variability. We conclude that differ-
ter of the feed hole. The wall thickness was
ences in hole size primarily account for gidity of the container, and the sucking determined using a Magna-Mike thickness
the observed variability in milk flow. This pressure generated by the infant.1"3
finding may be clinically important in that gauge (Automatic Inspection Device, Ine).
Since the pressure generated by the in¬ The wall thickness was measured by using a
rapid milk flow can lead to apnea and bra- fant varies from suck to suck and from small steel ball on the inside of the nipple and
dycardia in some preterm infants. The infant to infant, we have developed a a magnetic probe on the outside of the nip¬
above observations imply that design
changes are necessary to reduce the vari- system with which we can evaluate the ple's bulb. The gauge senses the magnetic
ability of milk flow within each nipple type. different nipple units at a constant suck¬ field strength between the ball and the tip of
ing frequency and pressure. ' Since rigid the probe. Since the magnetic field varies in a
Moreover, milk-flow measurements made known manner with the distance between
using a simple mechanical system and containers are used, the differences in
airflow measurements used by the indus- milk flow can be attributed solely to the ball and probe, the wall thickness can be
differences in nipple characteristics. calculated accurately with a microprocessor.
try are equally sensitive to evaluate nipple The airflow was measured by inserting the
flow. Recently, we reported a marked vari¬
(AJDC. 1990;144:222-224) nipple into an orifice and then inserting a
ability in milk flow within the same type plunger into the nipple forcing pressurized
and among the different types of nipple air (42 psi) through the feed hole. The flow
number of different types of nipple units. ' We speculated that the size of the rate of the air passing through the nipple was
units are available for use in neo¬
nipple hole is the most important vari¬ measured with an airflow meter.
nates; some are specially designed for able in determining milk flow; the thick¬ The milk flow through the nipple was mea¬
premature infants, whereas others are ness of the nipple unit may also contrib¬ sured as reported previously.1 Briefly, the
designed to resemble the human nipple. ute to variability. nipple to be tested was firmly secured on a
At present, most preterm infants are The aim of the present study was to bottle containing ready-to-feed formula (En¬
establish the role of these two variables famil, 281 kJ/dL), inverted, and placed on a
in determining milk flow. An additional rigid container. A negative pressure pulse of
Accepted for publication September 18,1989. -120 cm H20 at 40 cycles per minute was
From the Department of Pediatrics, University objective of the study was to compare then applied to the rigid container with a
of Texas Medical Branch, Galveston. the milk-flow determinations of the sim¬ breast pump. The number of sucks required
Reprint reqests to Perinatal Pediatrics, Univer-
sity of Texas Medical Branch, Galveston, TX 77550 ple mechanical system with airflow to empty a bottle with 120 mL of formula was
(Dr Mathew). measurements, the industry standard. determined and was inversely proportional

Downloaded From: http://archpedi.jamanetwork.com/ by a Missouri S & T User on 06/22/2015


to milk flow. Twenty nipple units were evalu¬
ated in each category. Comparison of Various Types of Nipple Units Used in the Neonatal Period
Stepwise linear regression analysis was Mean +
SD
used to correlate the hole size and wall thick¬
ness to airflow and milk flow rates. A value of Hole Size, Thickness, Airflow,
P<.05 was considered significant. Nipple Units cm cm m Us Milk Flow*
Term Infants
RESULTS Standard
Nipple Units for Term Infants SMA single-hole 0.043 ±0.003 0.077 ±0.012 30.7 + 2.8 777 ±106
Enfamil single-hole 0.036 ±0.004 0.084 ±0.011 24.4 ±3.6 1068 ±185
The results obtained with various Twist-on 0.031 ±0.008 0.127±0.015 5.1 ±2.7 627±268
nipple units are summarized in the Ta¬ Nuk-type
ble. In general, Nuk-type nipples had SMA 0.069 ±0.004 0.132 ±0.015 52.1 ±3.5 135±11
larger feed holes compared with stan¬ Enfamil Natural 0.040 ±0.003 0.116 ±0.020 25.9 ±3.3 342 ±56
dard nipple units for term infants. Nuk 0.055±0.004 0.110±0.012 42.3±3.5 183±49
Among the nipple units for term infants, Preterm Infants
the largest percentage variation in hole Standard
size was observed with the Twist-on SMA single-hole 0.039 ±0.005 0.131 ±0.015 23.8 ±5.3 827 ±180
nipple unit for term infants, made by Enfamil single-hole 0.054 ±0.005 0.115±0.016 43.1 ±2.9 320 ±37
Ross Laboratories. Among the stan¬ Twist-on 0.032 + 0.007 0.111 ±0.022 28.8±10.3 705 ±480
dard types, the greatest thickness at *
Values represent the number of simulated sucks to empty a bottle with 120 mL of formula.
the tip was observed with the Twist-on
nipple unit, and among the Nuk types,
the greatest thickness was observed
with the SMA unit. The SMA single-
hole unit had the largest percentage
variation in thickness among the stan¬
dard units, and the Enfamil Natural had
the largest percentage variation in
thickness among the Nuk-type nipple
units (Table).
Greater hole size was associated with
higher airflow and milk flow (indicated
by a smaller number of simulated
sucks). Regression analysis revealed a
highly significant (P<.001) correlation
between hole size and airflow as well as
between hole size and milk flow for stan¬
dard nipple units (Fig 1). This correla¬
tion was not significantly improved by
the addition of thickness. As for the Fig 1.—Standard nipple unit for term infants. The circle indicates SMA single-hole unit; the
square, Enfamil single-hole unit; and the triangle, Twist-on unit. Open symbols indicate airflow;
standard nipple unit, a highly signifi¬ solid symbols, milk flow. The higher the number of sucks, the lower the milk flow.
cant (P<.005) correlation existed be¬
tween hole size and airflow as well as
between hole size and milk flow for SMA
Nuk units and Enfamil Natural units,
whereas the correlation was less signifi¬
cant (P<.05) for the Nuk nipple distrib¬
uted through Ross Laboratories (Fig 2).
Thickness significantly improved this
correlation with milk flow only with the
SMA Nuk nipple unit.
Nipple Units for Preterm Infants
These nipple units had larger feed
holes compared with their counterparts
for term infants for Enfamil and smaller
feed holes for SMA nipple units (Table).
Both airflow and milk flow were corre¬
Fig 2.—Nuk-type nipple units. Correlations between airflow and hole size and milk flow and hole
spondingly higher with Enfamil and size are shown. Triangles indicate the Nuk nipple distributed through Ross Laboratory. See Fig
lower with SMA units for premature 1 for definitions of other symbols.

Downloaded From: http://archpedi.jamanetwork.com/ by a Missouri S & T User on 06/22/2015


infants. Again, regression analysis
showed highly significant (P<.001) cor¬
relations between hole size and airflow
as well as between hole size and milk
flow (Fig 3). No significant correlation
between thickness and airflow or milk
flow was observed with any of the nipple
units.
COMMENT
Using simulated sucks, we recently
reported a great variability in milk flow
within and among the various types of
nipple units available for use in new¬
borns.1 Results of the present study Fig 3.—Standard nipple units for preterm infants. See Fig 1 for definitions of symbols.
confirm and extend this observation.
The number of simulated sucks fell
within the 2 SDs reported previously.1 of the nipple units studied, there was a signed for premature infants? If so, how
The marked variability observed in poor correlation between nipple thick¬ should these differ from nipple units de¬
both studies underscores the variability ness and airflow or milk flow. Hence, signed for term infants? How much vari¬
in milk flow seen with these nipple wall thickness at the tip, at least in the ability in flow characteristics is accept¬
units. As a result of the earlier study, ranges studied, does not appear to be an able among nipple units designed for a
we hypothesized that differences in the important determinant of milk flow. specific group of infants, such as term or
size of the nipple hole probably account¬ Even though it is clear that differ¬ preterm infants?
ed for most of the observed variability in ences in hole size are primarily responsi¬ The present study documents that
milk flow and we suggested that rapid ble for the variation in milk flow, the the wide variability in flow characteris¬
milk flow may lead to apnea and brady¬ various tests used did not reveal the tics of the nipple units is due primarily
cardia in some preterm infants.1 The reason for this variability. One likely to differences in the size of feed holes
present results clearly document that explanation is the methods used in mak¬ and suggests the need for design
milk flow and airflow correlate well with ing the feed hole. The feed holes are changes to reduce the observed var¬
the size of the nipple hole. This correla¬ drilled mechanically, a method that iability.
tion was observed with all the different would be expected to produce signifi¬
types of nipple units studied. The differ¬ cant variation in hole size. Innovations
ence in thickness (within the ranges in this technique, such as the use of The author thanks Curtis P. Cowen and Mead-
Johnson Nutritional Division for their technical as¬
studied) did not significantly improve laser, may help reduce this variation in sistance and Linda Bodiford for secretarial
the correlation between hole size and hole size. Preliminary data on laser-cut assistance.
milk flow, except with the SMA Nuk nipple units indeed support this conten¬
nipple. tion (O.P.M., unpublished data, 1989).
Our results indicate that airflow and A decrease in ventilation during nip¬ References
simulated sucks are both equally good in ple feeding is well documented in both
evaluating functional characteristics of term and preterm infants.4,5 Among pre¬
the nipple (Figs 1 through 3). However, term infants, the reduction in ventila¬ 1. Mathew OP. Nipple unit for newborn infants:
a functional comparison. Pediatrics. 1988;81:688\x=req-\
simulated sucks employ relatively sim¬ tion is greater in the more premature 691.
ple methods compared with the airflow groups.5 Even apnea and bradycardia 2. Duara S, Rufenbergh L, Bancalari E. Vary-
ing feeding container rigidity: effect on intake and
measurements used by the industry. may occur in some of these infants.6 Im¬ ventilation in preterm infants. Pediatr Res.
Moreover, since simulated suck evalu¬ maturity of the central nervous system 1985;19:340.
3. Mathew OP. Regulation of breathing pattern
ates milk flow rather than airflow, and presumably is responsible for the above during feeding: role of suck, swallow and nutrients.
since it employs negative pressure observation. Alternatively, the high In: Mathew OP, Sant' Ambrogio G, eds. Respira-
pulse rather than constant positive milk flow present in these studies may tory Function of the Upper Airway. New York,
NY: Marcel Dekker Inc; 1988:535-560.
pressure, this form of evaluation resem¬ have contributed, at least in part, to the
4. Mathew OP, Clark ML, Pronske ML, Luna-
bles more closely the physiological con¬ adverse effects mentioned above. If this Solarzno HG, Peterson MD. Breathing pattern and
ditions observed during bottle feeding. assumption is true, the occurrence of ventilation during oral feeding in term infants J
Pediatr. 1985;106:810-813.
We did not measure directly the dis- apnea and bradycardia can be reduced 5. Shivpuri CR, Martin RJ, Carlo WA, Fanaroff
tensibility of the nipple units. Instead, significantly by decreasing the milk AA. Decreased ventilation in preterm infants dur-
we chose to measure the wall thickness flow. ing oral feedings. J Pediatr. 1983;103:285-289.
6. Mathew OP. Respiratory control during nip-
at the tip of the nipple, which is an indi¬ Our results raise several questions. ple feeding in preterm infants Pediatr Pulmonol.
rect measure of distensibility. For most Do we need nipple units specially de- 1988;5:220-224.

Downloaded From: http://archpedi.jamanetwork.com/ by a Missouri S & T User on 06/22/2015

You might also like