3. Clinical Purpose/Description Clinical Purpose/Description: A compact, light weight device used in emergency, trauma and ICU departments as well as operating theaters for conventional radiography 4. Technical Specification: Mobile stand Compact and light weight design with mobile castors for easy maneuverability Fully counterbalanced with compact flat detector Backup battery with continues operating time of 2hrs/100 exposures Dose area measurement system Automatic exposure control KV, mA increase & decrease Effective braking system for parking, transport and emergency. Articulated arm for maximum positioning flexibility in any patient position. Emergency stop button Standby and exposure release switch Ready and X-Ray on switch with Indicators All cables shall be concealed in the arm system Storage facility for flat panel detector and other supplies Digital display of kV, mAS and/or mA and time X-ray generator: High Frequency generator without put power 25Kw kV range: 40-125Kvp mA: 20 mA to 320 mA mAs: max. 280 per exposure Exposure time: for patient exposure should be ≤1s and for tube capacity should be up to 5s Self diagnostic programme with Indicators like tube over heating/high voltage, power supply not ready, broken filament error, earth fault error, etc.. X-ray tube and collimator: Output should match the output of the generator Dual focus rotating anode with spot size less than 1.3mm Tube focal spot selection Switch Total filtration: minimum 2 mm Al Tube voltage: 40-125Kvp Anode heat capacity: 300KHU Anode cooling rate: 40KHU/min. Tube horizontal movement: 45 cm or more Tube vertical movement: 100 cm or more Adjustable multi leaf collimator with SID laser localizer Collimator lamp on switch Detector System: The detector should be solid state flat detector or latest technology with cesium iodide scintillator. Detector active size: 35cm x 43cm or more. Pixel size: 148 um or less. Detector Quantum Efficiency (D.Q.E): 65% @ Zero LP/mm or more. With at least two user selectable input fields Active matrix size: 2.8k x 2.3k Image acquisition and image processing: Shall have integrated 17" high brightness LCD/TFT color button/touch screen monitor. The digital workstation should be based on the latest high speed processors of at least 32 bit. Patient data management- Electronic record with name, date, anatomy, etc.. Automatic digital brightness and contrast control for optimal image quality Image rotation, reversal (left/right), and up/down on last image hold Annotation (text, trace, arrow, line, rectangle, circle in images) Measurements (length, distance and angles in images) Image post processing features: zoom & pan, edge enhancement, contrast and brightness, etc.. Connectivity: Integrated data interface via LAN/Wi-Fi enables to connect into the hospital network. Storage of 3000 images on hard disc Integrated facility documentation with DVD/CD, USB and DVD recording DICOM: System should be ready DICOM interface and have networking capability DICOM print DICOM store enables image transfer to DICOM compliant workstations for off-line processing of images, store images/runs on CD-R to PACS system and to PC DICOM Modality for communications with HIS/RIS system DICOM Modality Performed Procedure Steps (DMPPS) 5. System Configuration Accessories, Spares, Consumables and other components: Main unit with x-ray generator, x-ray tube----1 Flat panel detector----1 All software packages shall be provided as standard A CD-R/W based long term archiving with envelope----10,000 Whole body lead aprons with 0.5mm, 0.35mm, 0.25mm lead thickness (or equivalent) ----2 of each Thyroid shield ------3 Gonad shield small, medium, large size----2 of each Lead goggles -----3 Phantoms and meters for the quality control and calibration of the various components (x ray, medical displays) of the system shall be provided All standard accessories and parts required to operate the equipment, including all standard tools and cleaning and lubrication materials, to be included in the offer. 6. Operating Environment: The unit shall be capable of operating continuously in ambient temperature of: +10ᵒC to +40ᵒC Relative Humidity: <85% 7. Utility Requirements: Power input: 220V/ 50Hz 8. Standards & Safety Requirements: Shall meet IEC-60601(Or Equivalent ) General Requirements of electrical Safety 9. Installation/Training/Commissioning: The supplier must provide installation, and commissioning of the device The supplier must provide sufficient technical and end user training on site. 10. Warranty/ After sales service: The supplier must be provide minimum of Two years warranty including Labour, X-ray tube, detectors and spare parts and accessories from the date of installation as well as after sale services. The supplier must provide extended warranty from 3rd year to 5th year inclusive of labour, spare parts, accessories, detector and X Ray tube. 10. Warranty and After Sale service The supplier must provide minimum of Two year basic warranty including labor and spare part from the date of commissioning. After basic warranty the supplier must agree for after sales service 11. Documentation: User, and service manual in English 12. Packaging and Labeling: Packing of all the goods clearly marked and securely packed. Each goods will be further packed in separate package with all its standard accessories of distinct identification and numbers consecutively. Additional packing and labeling requirements should bear in each package Each item with all accessories /spare part shall be configured and packed in one unit.