Equipment Specification

Equipment Name: X-ray - Mobile, Digital

Facility Type: Tertiary Hospital

Department: Radiology and imaging

Detail Description

1. Generic Name: X ray System

2. GMDN/UMDN Name/ Code: X ray - Mobile, Digital


3. Clinical Purpose/Description
Clinical Purpose/Description: A compact, light weight device used in emergency, trauma and ICU departments as well as
operating theaters for conventional radiography
4. Technical Specification:
Mobile stand
Compact and light weight design with mobile castors for easy maneuverability
Fully counterbalanced with compact flat detector
Backup battery with continues operating time of 2hrs/100 exposures
Dose area measurement system
Automatic exposure control
KV, mA increase & decrease
Effective braking system for parking, transport and emergency.
Articulated arm for maximum positioning flexibility in any patient position.
Emergency stop button
Standby and exposure release switch
Ready and X-Ray on switch with Indicators
All cables shall be concealed in the arm system
Storage facility for flat panel detector and other supplies
Digital display of kV, mAS and/or mA and time
X-ray generator:
High Frequency generator without put power 25Kw
kV range: 40-125Kvp
mA: 20 mA to 320 mA
mAs: max. 280 per exposure
Exposure time: for patient exposure should be ≤1s and for tube capacity should be up to 5s
Self diagnostic programme with Indicators like tube over heating/high voltage, power supply not ready, broken filament error,
earth fault error, etc..
X-ray tube and collimator:
Output should match the output of the generator
Dual focus rotating anode with spot size less than 1.3mm
Tube focal spot selection Switch
Total filtration: minimum 2 mm Al
Tube voltage: 40-125Kvp
Anode heat capacity: 300KHU
Anode cooling rate: 40KHU/min.
Tube horizontal movement: 45 cm or more
Tube vertical movement: 100 cm or more
Adjustable multi leaf collimator with SID laser localizer
Collimator lamp on switch
Detector System:
The detector should be solid state flat detector or latest technology with cesium iodide scintillator.
Detector active size: 35cm x 43cm or more.
Pixel size: 148 um or less.
Detector Quantum Efficiency (D.Q.E): 65% @ Zero LP/mm or more.
With at least two user selectable input fields
Active matrix size: 2.8k x 2.3k
Image acquisition and image processing:
Shall have integrated 17" high brightness LCD/TFT color button/touch screen monitor.
The digital workstation should be based on the latest high speed processors of at least 32 bit.
Patient data management- Electronic record with name, date, anatomy, etc..
Automatic digital brightness and contrast control for optimal image quality
Image rotation, reversal (left/right), and up/down on last image hold
Annotation (text, trace, arrow, line, rectangle, circle in images)
Measurements (length, distance and angles in images)
Image post processing features: zoom & pan, edge enhancement, contrast and brightness, etc..
Connectivity:
Integrated data interface via LAN/Wi-Fi enables to connect into the hospital network.
Storage of 3000 images on hard disc
Integrated facility documentation with DVD/CD, USB and DVD recording
DICOM:
System should be ready DICOM interface and have networking capability
DICOM print
DICOM store enables image transfer to DICOM compliant workstations for off-line processing of images, store images/runs
on CD-R to PACS system and to PC
DICOM Modality for communications with HIS/RIS system
DICOM Modality Performed Procedure Steps (DMPPS)
5. System Configuration Accessories, Spares, Consumables and other components:
Main unit with x-ray generator, x-ray tube----1
Flat panel detector----1
All software packages shall be provided as standard
A CD-R/W based long term archiving with envelope----10,000
Whole body lead aprons with 0.5mm, 0.35mm, 0.25mm lead thickness (or equivalent) ----2 of each
Thyroid shield ------3
Gonad shield small, medium, large size----2 of each
Lead goggles -----3
Phantoms and meters for the quality control and calibration of the various components (x ray, medical displays) of the
system shall be provided
All standard accessories and parts required to operate the equipment, including all standard tools and cleaning and
lubrication materials, to be included in the offer.
6. Operating Environment:
The unit shall be capable of operating continuously in ambient temperature of: +10ᵒC to +40ᵒC
Relative Humidity: <85%
7. Utility Requirements:
Power input: 220V/ 50Hz
8. Standards & Safety Requirements:
Shall meet IEC-60601(Or Equivalent ) General Requirements of electrical Safety
9. Installation/Training/Commissioning:
The supplier must provide installation, and commissioning of the device
The supplier must provide sufficient technical and end user training on site.
10. Warranty/ After sales service:
The supplier must be provide minimum of Two years warranty including Labour, X-ray tube, detectors and spare parts and
accessories from the date of installation as well as after sale services.
The supplier must provide extended warranty from 3rd year to 5th year inclusive of labour, spare parts, accessories,
detector and X Ray tube.
10. Warranty and After Sale service
The supplier must provide minimum of Two year basic warranty including labor and spare part from the date of
commissioning.
After basic warranty the supplier must agree for after sales service
11. Documentation:
User, and service manual in English
12. Packaging and Labeling:
Packing of all the goods clearly marked and securely packed.
Each goods will be further packed in separate package with all its standard accessories of distinct identification and
numbers consecutively.
Additional packing and labeling requirements should bear in each package
Each item with all accessories /spare part shall be configured and packed in one unit.