Professional Documents
Culture Documents
1. INTRODUCTION
This document is a guide dedicated to intermediate or advanced users of the MS 5800 & MultiView tube inspection IRIS system. It
explains the different defect responses that can be observed by an IRIS technician or analyst. Setup and analysis guidelines are
given for every situation.
This document is written for users having at the minimum a basic understanding of Ultrasound NDT technology and the IRIS
system.
This document shows several “ray tracing” schematics. These “ray tracing” are arrows showing the direction of the focalized
ultrasound beam emitted by the IRIS transducer. The arrow itself shows the center of the beam, where the ultrasound waves are the
strongest.
When the returned echo arrow points towards the turbine mirror, the signal is at the maximum possible. This results as a clear
image on the B-scan and C-scan views.
On the other hand, when the inside (or outside) surfaces present an important angle (due to a defect), the ultrasound beam can be
deviated away from the turbine mirror. On drawings, the arrow will point in a direction different than the mirror. This does not mean
NO signal gets to the mirror; ultrasound waves are more like a “front” and there will still be signal reaching the mirror, but with
(much) lower amplitude.
Reflection misses
the turbine mirror
due to the higher
Reflection is angle of the ID
directed towards surface: signal
the turbine loss is likely to
mirror: maximum occur.
signal amplitude is
obtained.
Turbine mirror
(facing the page)
Tube section
Figure 1: Example of internal defects reflecting the ultrasound beam at different angles. When the
reflected angle is more important, the signal amplitude is lower and signal loss is likely to occur.
IRIS-ADVANCED-GUIDE.doc Page 1 / 19
Olympus NDT Canada
505, boul. du Parc-Technologique Tel. : (418) 872-1155
Québec (Québec) G1P 4S9 Fax. : (418) 872-5431
Canada Web site : www.olympusndt.com
3. PROBE CENTERING
IRIS turbines should always be kept centered to help uniform detection around the circumference. The probe centering can not be
kept perfect in the real conditions for obvious reasons, but care should be taken to obtain the best centering conditions. Sometimes
the centering device shaft can be bent, or its threads, and should be fixed or replaced.
The drawings below illustrate why the IRIS probes are so sensitive to centering. Basically, signal loss occurs when the ID surface
reflects the ultrasound beam at a high angle, resulting in lower amplitude echoes to reach the turbine mirror.
A C
IRIS-ADVANCED-GUIDE.doc Page 2 / 19
Olympus NDT Canada
505, boul. du Parc-Technologique Tel. : (418) 872-1155
Québec (Québec) G1P 4S9 Fax. : (418) 872-5431
Canada Web site : www.olympusndt.com
C: Mirror at
maximum D: Signal loss
distance from due to ID surface
ID surface being reflecting
UT beam at
higher angle
4. TURBINE SELECTION
The turbine diameter, by itself, has no effect on the signals. However, a bigger diameter turbine also include a bigger diameter
mirror, which results as greater chances of having more echo signal back. Whether the cause for angled echo reflection is a defect,
or just off-centering, the bigger mirror always help to obtain clearer B-scan and C-scan views.
The following example show the effect of probe off-centering has on a smaller mirror turbine versus a bigger size mirror:
Maximum signal
amplitude is easier
caught with a bigger
mirror
Figure 4: A 17 mm turbine mirror (left) catches more signal amplitude than the
12 mm turbine (right) when the probe is not properly centered.
How to choose turbine size then? Always use the 17 mm model when possible. The 12 mm should be used only when the biggest
turbine can not be inserted inside the tube.
IRIS-ADVANCED-GUIDE.doc Page 3 / 19
Olympus NDT Canada
505, boul. du Parc-Technologique Tel. : (418) 872-1155
Québec (Québec) G1P 4S9 Fax. : (418) 872-5431
Canada Web site : www.olympusndt.com
5. OD ROUND-BOTTOM PIT
While OD pitting is easily detected with the IRIS method, signal loss is sometimes observed on the edges of the
defects. This is more likely to occur on deeper and sharper pits. This is easier understood when loking at the
“maximum amplitude” UT ray tracing as the turbine revolves inside the tube:
Round-bottomed
OD pit
Maximum ID and
OD echoes
Turbine 45°
mirror, face view
Figure 5: Maximum Ultrasound wave intensity is obtained at the deepest point of the defect (0°).
ID echo
OD echo
Figure 6 : Both the ID and OD echoes are received with very good signal amplitude.
IRIS-ADVANCED-GUIDE.doc Page 4 / 19
Olympus NDT Canada
505, boul. du Parc-Technologique Tel. : (418) 872-1155
Québec (Québec) G1P 4S9 Fax. : (418) 872-5431
Canada Web site : www.olympusndt.com
Total or partial
reflection ID echo
OD echo
Figure 7 : OD echo not reaching the mirror with maximum amplitude. Signal loss could occur.
Figure 8 : Maximum amplitude is resumed when beam is oriented outside the defect area.
IRIS-ADVANCED-GUIDE.doc Page 5 / 19
Olympus NDT Canada
505, boul. du Parc-Technologique Tel. : (418) 872-1155
Québec (Québec) G1P 4S9 Fax. : (418) 872-5431
Canada Web site : www.olympusndt.com
In RED: areas of
lower sensitivity.
Loss of OD signal
on edges of defect
is frequent with
deeper OD pitting
Clean, defect-free
ID signal
Analysis:
Signal loss on the edges of the OD pitting is not a problem, as the deeper part of the defect (the most important
part) presents a “flatter” surface that reflects the ultrasound beam with good amplitude.
In fact the presence of signal loss – as on the above C-scan image – can be used in analysis along with the red
color as an indication of deeper flaws.
- The WT (wall thickness) C-scan view is recommended.
- Always measure the deepest part of the defect, and with pits it is likely to be the middle part. Always analyze
B-scan data and use the manual sizing if required.
- Double back wall echo detection is rare unless the pitting is really severe.
IRIS-ADVANCED-GUIDE.doc Page 6 / 19
Olympus NDT Canada
505, boul. du Parc-Technologique Tel. : (418) 872-1155
Québec (Québec) G1P 4S9 Fax. : (418) 872-5431
Canada Web site : www.olympusndt.com
6. ID ROUND-BOTTOM PIT
Actual pitting in Carbon Steel often show a semi-spherical section. The sizing of internal pitting is always more
difficult than of OD defects, for there is likely to have more signal loss from both interface echoes.
ID and OD echoes
Figure 11: An ID pit deepest point is easily measured with IRIS as the
ultrasound beam is oriented perpendicular to both surfaces.
OD echo
OD echo
ID echo
Figure 12: Signal is still very good when the ultrasound beam is directed at very low angle.
IRIS-ADVANCED-GUIDE.doc Page 7 / 19
Olympus NDT Canada
505, boul. du Parc-Technologique Tel. : (418) 872-1155
Québec (Québec) G1P 4S9 Fax. : (418) 872-5431
Canada Web site : www.olympusndt.com
OD echo – longer
time of flight
ID echo
Figure 13: A small rotation of the mirror causes a drastic attenuation of the ID echo.
The OD echo will travel longer time of flight as well.
Figure 14: The ID echo is composed of a beam containing reflections from several directions. This
example show the ID echo can be detected with a longer time of flight.
IRIS-ADVANCED-GUIDE.doc Page 8 / 19
Olympus NDT Canada
505, boul. du Parc-Technologique Tel. : (418) 872-1155
Québec (Québec) G1P 4S9 Fax. : (418) 872-5431
Canada Web site : www.olympusndt.com
Figure 15: ID pitting will cause loss of signal at the edge of the defect.
Figure 16: Internal pitting will appear very similar to OD pitting on the WT C-scan.
IRIS-ADVANCED-GUIDE.doc Page 9 / 19
Olympus NDT Canada
505, boul. du Parc-Technologique Tel. : (418) 872-1155
Québec (Québec) G1P 4S9 Fax. : (418) 872-5431
Canada Web site : www.olympusndt.com
Figure 17: The ID nature of pitting becomes obvious with IRIS, but when looking at the ID C-scan view.
Analysis:
Signal loss on the edges of the ID pitting is not a problem; the deeper part of the defect, as OD pitting, will act as
a reflector to the ultrasound beam.
The signal loss on the edge of the pit appears as a black surrounding around defects in the C-scan view. This can
actually help to locate the pits quicker.
- The ID (internal diameter) C-scan view is highly recommended. Make sure to optimize color palette if needed,
but in most cases the “Setup Wizard” preset is very good.
- The WT C-scan will show most of ID pitting; however the color palette might not always reflect the true depth
of the pitting, as double-back wall detection is frequent (see above image).
- Rely on B-scan view to understand how the pitting should be measured. Manual measurement is common:
Sizing is performed
between the deepest
(middle) portion of
the pit and the
estimated back wall
Estimated back wall
surface (dotted red)
- As actual pitting show uneven or dirtier surface, this is likely to further attenuate echoes, in particular the OD
echo which might be lost in some cases, or detected with the double of its time of flight as the DAC amplifies
the 2nd reflection. This is explained at the end of the document.
IRIS-ADVANCED-GUIDE.doc Page 10 / 19
Olympus NDT Canada
505, boul. du Parc-Technologique Tel. : (418) 872-1155
Québec (Québec) G1P 4S9 Fax. : (418) 872-5431
Canada Web site : www.olympusndt.com
ID C-scan view
Correct sizing of ID pit
using the manual
measurement
Figure 19: Air fin cooler inspected with IRIS. The ID C-scan is shown
as the fins cause the OD echoes to be weakened.
When sizing internal defects, it is often required to use the “manual” measurement (blue cursor) as per the above
image. The loss of OD signal prevents from using the automatic measurement.
Always measure the “remaining wall” value.
IRIS-ADVANCED-GUIDE.doc Page 11 / 19
Olympus NDT Canada
505, boul. du Parc-Technologique Tel. : (418) 872-1155
Québec (Québec) G1P 4S9 Fax. : (418) 872-5431
Canada Web site : www.olympusndt.com
Severe ID damage
The next image shows a real tube suffering severe internal corrosion. Such condition cause inevitable attenuation
of the echoes, both front wall (ID) and back wall (OD). The OD echo amplitude suffers from two sources of
attenuation: 1) internal surface condition that causes attenuation by absorption and 2) diffraction from edges of
internal defects.
Figure 20: Severe ID damage attenuates echoes, causing loss of back wall signal.
The need to machine artificial defects at the lower cost often results as defects presenting sharp edges that are
rarely found in real damage condition. The sharp edges, or angled surfaces, are the ennemy of IRIS echoes, as
they are directed outside the turbine mirror. The following examples show the Ultrasound “maximum amplitude”
ray tracing for some common calibration defects.
Figure 21: Signal loss or attenuation on edges of an OD “flat mill” calibration defect
IRIS-ADVANCED-GUIDE.doc Page 12 / 19
Olympus NDT Canada
505, boul. du Parc-Technologique Tel. : (418) 872-1155
Québec (Québec) G1P 4S9 Fax. : (418) 872-5431
Canada Web site : www.olympusndt.com
Loss of OD echo in
the regions in RED on
the previous image
Signal loss can normally be improved by increasing GAIN 1 and 2 simultaneously by 2 or 4 dB.
Signal loss is
likely to occur
IRIS-ADVANCED-GUIDE.doc Page 13 / 19
Olympus NDT Canada
505, boul. du Parc-Technologique Tel. : (418) 872-1155
Québec (Québec) G1P 4S9 Fax. : (418) 872-5431
Canada Web site : www.olympusndt.com
The response to such defect can be improved, but it is important to keep in mind this kind of defect does NOT
represent real pitting condition. In other words, optimizing for such defect is a waste of time as real internal
pitting, due to their rounder shape, are much easier to detect.
Here are example parameters that were changed in order to obtain clearer ID echo, such as on the next image:
- 20 MHz / 1” focal transducer, instead of a 15 MHz, to obtain more resolution at the surface.
- Slightly higher GAIN 1 and 2 parameters
- Note: The GAIN 1 can be drastically increased (> 10 dB) to obtain a full ID reading. However, again, this is no
good setup for the real condition as the OD reading will be “false-triggering” due to high signals amplitude.
The ID reading is
more constant with
higher gains and
higher frequency
transducer.
Figure 25: Same flat bottomed hole, but with optimized setup.
IRIS-ADVANCED-GUIDE.doc Page 14 / 19
Olympus NDT Canada
505, boul. du Parc-Technologique Tel. : (418) 872-1155
Québec (Québec) G1P 4S9 Fax. : (418) 872-5431
Canada Web site : www.olympusndt.com
Figure 26: Drilled holes are the worst calibration defects for IRIS.
As seen on the next image, most of the data is missing from the defect. Only the middle section – and fortunately
the deepest – can be measured.
Drilled holes should be avoided at all times with IRIS. The best pit shape to be used are still round bottomed holes,
especially necessary when they are located on the ID surface.
Figure 27: A drilled-like internal hole. Only a few pixels in the middle of the defect indicate its depth.
IRIS-ADVANCED-GUIDE.doc Page 15 / 19
Olympus NDT Canada
505, boul. du Parc-Technologique Tel. : (418) 872-1155
Québec (Québec) G1P 4S9 Fax. : (418) 872-5431
Canada Web site : www.olympusndt.com
MultiView already includes all necessary tools for proper ID or OD defect analysis and sizing. Following are some
tips on recommended operations/commands and others that should be avoided.
Setup Wizard:
The “Setup Wizard” does a great job of presetting many parameters in order to have the IRIS “up and running”
with almost no adjustments made, as long as the proper parameters are entered in the Wizard boxes. ALL IRIS
parameters are preset, and that also includes color palettes, gains, blanking, etc., so if modifications are required,
they should be minor.
Main window:
Color palette
Defect on ID side
Defect on
OD side
Automatic % wall
loss reading
(top) and manual
reading (bottom)
Figure 28: The main MultiView window with all useful commands.
UT settings:
The GAIN 1, GAIN 2 and Blanking are the only 3 parameters that should be adjusted. All others are advanced
parameters that can improve signal quality by 10% only, at the best case. For most cases adjusting the advanced
parameters is a waste of time, while having the proper transducer of clearing bubbles should be the priority.
IRIS-ADVANCED-GUIDE.doc Page 16 / 19
Olympus NDT Canada
505, boul. du Parc-Technologique Tel. : (418) 872-1155
Québec (Québec) G1P 4S9 Fax. : (418) 872-5431
Canada Web site : www.olympusndt.com
Echoes are
synchronized on
the ID signal for
easier setup
Inspection TAB
Adjust material
velocity if WT
reading is not
accurate
Change Trigger
LEVEL sign only;
NEVER change
the LEVEL value
Figure 32: Only the LEVEL SIGN (+/-) should be changed in order to obtain a smoother WT reading. But
again, use of the “Setup Wizard” is recommended as this is automatically preset.
IRIS-ADVANCED-GUIDE.doc Page 17 / 19
Olympus NDT Canada
505, boul. du Parc-Technologique Tel. : (418) 872-1155
Québec (Québec) G1P 4S9 Fax. : (418) 872-5431
Canada Web site : www.olympusndt.com
The ECHO FINDER tab is used for setting up the sync pin gate (“target pin”) and “probe delay” parameters, if ever
required.
Refer to the “IRIS SETUP GUIDE” document for more details.
Figure 34: The OPTIONS tab should be left unchecked at all times, as
these parameters were developed for non-IRIS applications.
IRIS-ADVANCED-GUIDE.doc Page 18 / 19
Olympus NDT Canada
505, boul. du Parc-Technologique Tel. : (418) 872-1155
Québec (Québec) G1P 4S9 Fax. : (418) 872-5431
Canada Web site : www.olympusndt.com
9. DOUBLE BACK-WALL
Deeper OD defects or wall thinning can sometimes cause a strange effect called the “double back wall” reading.
This occurs as the first reflection of the OD surface does not get detected by the system, because it is either too
weak or too early in time (deep defect).
As the sound keeps bouncing between the ID and OD surfaces, there are multiple reflections of the back wall. The
system has a DAC (increasing gain over time) so the multiple reflections (repetitions) often get amplified with more
amplitude than the first reflection. This results as a detection of the double back wall echo, as can be seen on the
A-scan image below:
Detection occurs at
First back wall the second reflection,
reflection is too causing the WT value
weak to be to display as double of
detected the remaining wall
Figure 35: Double back wall detection example as the first reflection is too weak
(low signal amplitude) to be detected.
Figure 36: The double back wall reflection is detected, as the first reflection was either too weak or too
early (deep defect) to be properly detected. Data interpretation with help of the manual measurement
is necessary to accurately measure defect depth.
This phenomenon, even tough seeming discouraging, does NOT prevent the user from sizing detects. First, the B-
scan has to be analyzed for understanding the OD profile and locate where the double back wall measurement
occurred. As the user must measure the deeper section of the defect, it is likely that the double – or sometimes
even triple – reflections will be present at this location. This is easily overrun by using the manual cursor and
dividing the B-scan thickness in 2 (double reflection) or even in 3 (rare case of triple reflection).
IRIS-ADVANCED-GUIDE.doc Page 19 / 19