Revision B Revised issue for TB No. SSD-1400-14 issued on 8-Jul.-2002.

No. SSD-1400-14B Page: 1/1


TECHNICAL BULLETIN Date: 7-Nov.-2002

CATEGORY PARTS REQUIRED


Top urgent Attached
As soon as possible Separately sent to you
Next time you visit the customer Prepare at your side
When the symptom appears Apply to ALOKA office for your territory when necessary
Information Sales item. Order to the sales person in charge of your country

TITLE: Solution of the abnormal working of a mechanical probe when SCU-1400C was installed.

MODEL AFFECTED: SSD-1400

SERIAL NUMBERS AFFECTED: All that have been shipped

PURPOSE: This is to inform the abnormal working of a mechanical probe when SCU-1400C was installed
into SSD-1400 and measure to solve this.

SYMPTOM: The ultrasound image of a mechanical probe does not appear when turning the system on while
the transducer element inside of a mechanical probe is scanning very quickly.
SOLUTIONS:
1. Disconnect a cable of “CO-PSC-128-L-07C5” from a connector of “J414” on EP400800EG and
before.
2. Stick a caution label of “Do not connect a P414 to J414” to a connector of P414 to prevent
misconnection trouble.

Note: This problem was solved on EP400800EH and later. In case where EP400800EH is
incorporated in SCU-1400C, a cable of “CO-PSC-128-L-07C5” must be connected
to J414 on EP400800EH.
The serial number of SCU-1400C which EP400800EH was incorporated are as
follows.
S/N. M00322, M00338, M00339, M00391~M00397, M00411~

ITEMS OF REQUIRED PARTS:


None
Rev. B added
ATTACHED DOCUMENTS:
1 sheet.
ESTIMATED LABOR TIME:
0.5 hour
SPECIAL TOOLS AND TEST EQUIPMENT REQUIRED:
Phillips-type screw driver: +3mm and Service manual.
ESSENTIAL CONDITION:
None.
Approved by
Rev.B

- How to revise the Technical Bulletin -


♦ Replace all pages of SSD-1400-14 with all pages of SSD-
1400-14B. General Manager
♦ Replace all pages of attached documents with new one. Medical Ultrasound Engineering Dept.

Written by
The replaced pages are no longer valid and should be
discarded.
ALOKA CO.,LTD. Rev.B

① ケーブル(CO-PSC-128-L-07C5)の
(J414)をオプション PC 板(EP4008*)の
コネクタから外す。

① Disconnect a cable of “CO-PSC-128-L-07C5”


from a connector of “J414” on EP400800**.

J414 CO-PSC-128-L-0705

注意事項
修理作業時、誤って接続しないよう P414 のコネクタに「P414 は接続不要」の
注意書きを貼る。

NOTICE
Stick a caution label of “Do not connect P414 to J414” to a connector of P414 to
prevent misconnection trouble.
Rev. B added

注 意
EP400800EH より対策されましたので、リビジョンが EH の基板の場合は、J414 へは
ケーブルを接続してください。
尚、EP400800EH が使用されている SCU-1400 の S/N は、以下の通りです。
CAUTION
This problem was solved on EP400800EH and later. In case where EP400800EH is
incorporated in SCU-1400C, a cable of “CO-PSC-128-L-07C5” must be connected to
J414 on EP400800EH.
The serial number of SCU-1400C which EP400800EH was incorporated are as
follows.
S/N. M00322, M00338, M00339, M00391~M00397, M00411~

− 1 −