EPX-SERISE SERVICE MANUAL Rev.

EPX-Series
Diagnostic X-ray Systems

Service Manual

Ecotron Co., Ltd


Made by Korea
#504~505Ho, Hanshin IT Tower II, 47, Digital-ro 9-gil, Geumcheon-Gu, Seoul, Korea
Obelis S.A
Avenue de Tervuren, 34, bte 44 1040 Brussels, Belgium / E-mail:mail@obelis.net
Representative: Mr.Gideon ELKAYAM / TEL:32.2.732.59.54, FAX:32.2.732.60.03

D/N: ECO-SM-EPX-001 Rev. E

1
EPX-SERISE SERVICE MANUAL Rev. E

(This page intentionally left blank)

2
EPX-SERISE SERVICE MANUAL Rev. E

SERVICE CHAPTER

1. CALIBRATION------------------------------------------------------------------------------------------- 7

2. DIP SWITCH SETTING--------------------------------------------------------------------------------31

3 .HARD-WIRED hand switch control---------------------------------------------------------------32

4. COLLIMATOR-------------------------------------------------------------------------------------------33

5. TROUBLE SHOOTING-------------------------------------------------------------------------------35

6. SERVICE CHECK LIST-------------------------------------------------------------------------------37

REVISON HISTORY

3
EPX-SERISE SERVICE MANUAL Rev. E

Revision Number Date Description

A SEP 11, 2013 First Edition

B SEP 30, 2014 European Authorized Representative Change

C JAN 12, 2015 Manual Updates

D SEP 01, 2015 Manual Updates

E JUN 21, 2016 Manual Updates

4
EPX-SERISE SERVICE MANUAL Rev. E

(This page intentionally left blank)

5
EPX-SERISE SERVICE MANUAL Rev. E

Service chapter
This manual is designed to ensure correct use and operation of EPX model diagnostic x-ray units.
Please read all the lines thoroughly before you use the equipment.

Incorrect use and operation exceeding described conditions in this manual may cause damage of the
equipment and shorten its life. Particular attention must be paid to all the warnings, cautions and notes
incorporated herein.

This equipment should be used only by the legally qualified persons and practitioners.

Regarding the AC cable and software, only the ones developed and supplied by Ecotron Inc. should be
used.

Only trained maintenance staff may remove the covers of the x-ray system.

Do not modify the equipment at your discretion and in case any modification is required unavoidably, ask
the help of Ecotron or its authorized dealer for the service.

Information for repair will be provided by Ecotron. (Circuit diagram, Component part list, descriptions,
Calibration instructions, or other information)

No responsibility is taken by Ecotron for any infringement of patients or other rights of the third parties
which many results from the use of this manual.

Always keep the manual at hand for your reference.

NOTE: If the output calibration is needed after periodical inspections or repairs, please observe following
rules and methods described in this manual.

CAUTION
The ionizing radiation is dangerous for the operator if the following safety measures
are not strictly observed.

NOTE

● Assembly Information: No Assembly is required.

6
EPX-SERISE SERVICE MANUAL Rev. E

1. Calibration
1.1 EPX-F1600
1.1.1 Line Power Switch

Line switch should be at off position.


Connect power line cord to the power outlet.
Turn on the line switch.
The FND on the control panel should illuminate.

1.1.2 Calibration Procedure


1.1.2.1 Tube voltage confirmation and calibration.
a. Turn off the line switch.
b. Detach the control panel from the unit to access the PCB OP BOARD.
c. Connect the channel one probe of the oscilloscope to TP8 (IP F/B).
d. Connect the channel two probe of the oscilloscope to TP11 (EP F/B).
e. Connect the ground to TP13 or TP14 (GND) on the OP board.
f. Turn on the line switch.
g. Set the kV and the mAs at 40KV and 5mAs(30mA), respectively.
h. Then observe the KV waveform of the oscilloscope channel 2.
i. If it is not 2V, adjust KV by turning VR3 on the OP board.

j. Set the kV and the mAs at 90KV and 5mAs(16mA), respectively.


k. Then observe the KV waveform of the oscilloscope channel 2.
l. If it is not 4.5V, adjust KV by turning VR3 on the OP board.

7
EPX-SERISE SERVICE MANUAL Rev. E

1.1.2.2 Tube current confirmation and calibration.

a. Turn off the line switch.


b. Detach the control panel from the unit to access the PCB OP BOARD.
c. Connect the channel one (1) probe of the oscilloscope to TP8 (IP F/B).
d. Connect the channel two (2) probe of the oscilloscope to TP11 (EP F/B).
e. Connect the ground to TP13 or TP14 (GND) on the OP board.
f. Turn on the line switch.
g. Set the kV and the mAs at 40kV and 5mAs(30mA), respectively.
h. Then observe the mA waveform of the oscilloscope channel 1.
i. If it is not 6V, adjust mA using the calibration data on the OP BOARD.
j. DIP SWITCH #4 on the OP BOARD PCB is set to ON position.

Then the display should indicate as follows,

Where, C00~C09 are calibration code of tube current at AC110V model and C10~C19 are calibration
code of tube current at AC220V model.
C00 or C10 is calibration code of tube current (30mA at 40kV~50kV) at AC110V or AC220V and 30.0
is calibration data of tube current (30mA at 40kV~50kV).

a. Adjust mA by using the mAs Up-Down switch (see operation sheet).

8
EPX-SERISE SERVICE MANUAL Rev. E

b. After it is adjusted, push the STORE button during 3sec(see operation sheet).
c. DIP SWITCH #4 on the PCB OP BOARD should be set to OFF position and the expose X-ray.
e. If it is not 6V, set the calibration data again.
f. DIP SWITCH #4 on the OP BOARD PCB is set to ON position.
g. Then push the kV UP switch (see operation sheet).
h. The display should indicate as follows,

Where, C01 or C11 is calibration code of tube current (25mA at 40kV~50kV) and 25.0 is calibration
data of tube current (25mA at 40kV~50kV).

i. Adjust mA by using the mAs Up-Down switch (see operation sheet).


j. After it is adjusted, push the STORE switch during 3sec(see operation sheet).
k. DIP SWITCH #4 on the PCB OP BOARD is set to OFF position then expose X-ray.
l. If it is not 5V, set the calibration data again.

Filament preheat calibration


This is preheating filament for making tube current.
The overshooting or the undershooting may occur from the tube current if the Preheat calibration data
is set by mistake.
Changing the calibration data is possible in accordance with the tube condition.

a. Turn off the line switch.


b. Detach the control panel from the unit to access the PCB OP BOARD.
c. DIP SWITCH #4 on the PCB OP BOARD is set to ON position.
d. Turn on the line switch.
f. Push APR S/W ① and then the display should indicate as follows,

9
EPX-SERISE SERVICE MANUAL Rev. E

Where, P00~P07 are calibration code of filament preheat current at AC110V, P08~P15 are calibration
code of filament preheat current at AC220V, P00 or P08 is Filament preheat calibration code (Preheat
calibration code of 30mA), and 1.50 or 1.40 is Filament preheat calibration data (Preheat calibration
data of 30mA).

g. Adjust data by using the mAs Up-Down switch (see operation sheet).
h. After it is adjusted, push the STORE button during 3sec(see operation sheet).
i. DIP SWITCH #4 on the PCB OP BOARD is set to OFF position then expose X-ray.
j. If overshooting or undershooting occurs in the waveform, set the calibration data again.
k. Then push the KV UP switch (see operation sheet).
l. The display should indicate as follows,

Where, P01 or P09 is Filament preheat calibration code (Preheat calibration code of 25mA) and
1.45 or 1.35 is Filament preheat calibration data (Preheat calibration data of 25mA).

m. Adjust data by using the mAs Up-Down switch on the Control Panel.
n. After it is adjusted, push the Store switch during 3sec(see operation sheet).
o. DIP SWITCH #4 on the PCB OP BOARD is set to OFF position then exposure X-ray.
p. If overshooting or undershooting occurs in the waveform, set the calibration data again.

10
EPX-SERISE SERVICE MANUAL Rev. E

1.2 EPX-F2400
1.2.1 Line Power Switch

Line switch should be at off position.


Connect power line cord to the power outlet.
Turn on the line switch.
The FND on the control panel should illuminate.

1.2.2 Calibration Procedure


1.2.2.1 Tube voltage confirmation and calibration.
a. Turn off the line switch.
b. Detach the control panel from the unit to access the PCB OP BOARD.
c. Connect the channel one probe of the oscilloscope to TP8 (IP F/B).
d. Connect the channel two probe of the oscilloscope to TP11 (EP F/B).
e. Connect the ground to TP13 or TP14 (GND) on the OP board.
f. Turn on the line switch.
g. Set the kV and the mAs at 40KV and 5mAs(40mA), respectively.
h. Then observe the KV waveform of the oscilloscope channel 2.
i. If it is not 2V, adjust KV by turning VR3 on the OP board.

j. Set the kV and the mAs at 100KV and 5mAs(20mA), respectively.


k. Then observe the KV waveform of the oscilloscope channel 2.
l. If it is not 5V, adjust KV by turning VR3 on the OP board.

11
EPX-SERISE SERVICE MANUAL Rev. E

1.2.2.2 Tube current confirmation and calibration.

a. Turn off the line switch.


b. Detach the control panel from the unit to access the PCB OP BOARD.
c. Connect the channel one (1) probe of the oscilloscope to TP8 (IP F/B).
d. Connect the channel two (2) probe of the oscilloscope to TP11 (EP F/B).
e. Connect the ground to TP13 or TP14 (GND) on the OP board.
f. Turn on the line switch.
g. Set the kV and the mAs at 40kV and 5mAs(40mA), respectively.
h. Then observe the mA waveform of the oscilloscope channel 1.
i. If it is not 6V, adjust mA using the calibration data on the OP BOARD.
j. DIP SWITCH #4 on the OP BOARD PCB is set to ON position.

Then the display should indicate as follows,

Where, C00~C07 are calibration code of tube current at AC110V model and C08~C15 are calibration
code of tube current at AC220V model.
C00 or C08 is calibration code of tube current (40mA at 40kV~60kV) at AC110V or AC220V and 40.0
is calibration data of tube current (40mA at 40kV~60kV).

12
EPX-SERISE SERVICE MANUAL Rev. E

a. Adjust mA by using the mAs Up-Down switch (see operation sheet).


b. After it is adjusted, push the STORE button during 3sec(see operation sheet).
c. DIP SWITCH #4 on the PCB OP BOARD should be set to OFF position and the expose X-ray.
e. If it is not 6V, set the calibration data again.
f. DIP SWITCH #4 on the OP BOARD PCB is set to ON position.
g. Then push the kV UP switch (see operation sheet).
h. The display should indicate as follows,

Where, C01 or C09 is calibration code of tube current (35mA at 40kV~60kV) and 35.0 is calibration
data of tube current (35mA at 40kV~60kV).

i. Adjust mA by using the mAs Up-Down switch (see operation sheet).


j. After it is adjusted, push the STORE switch during 3sec(see operation sheet).
k. DIP SWITCH #4 on the PCB OP BOARD is set to OFF position then expose X-ray.
l. If it is not 5.25V, set the calibration data again.

Filament preheat calibration


This is preheating filament for making tube current.
The overshooting or the undershooting may occur from the tube current if the Preheat calibration data
is set by mistake.
Changing the calibration data is possible in accordance with the tube condition.

a. Turn off the line switch.


b. Detach the control panel from the unit to access the PCB OP BOARD.
c. DIP SWITCH #4 on the PCB OP BOARD is set to ON position.
d. Turn on the line switch.
f. Push APR S/W ① and then the display should indicate as follows,

13
EPX-SERISE SERVICE MANUAL Rev. E

Where, P00~P04 are calibration code of filament preheat current at AC110V, P05~P09 are calibration
code of filament preheat current at AC220V, P00 or P05 is Filament preheat calibration code (Preheat
calibration code of 40mA), and 0.70 or 0.72 is Filament preheat calibration data (Preheat calibration
data of 40mA).

g. Adjust data by using the mAs Up-Down switch (see operation sheet).
h. After it is adjusted, push the STORE button during 3sec(see operation sheet).
i. DIP SWITCH #4 on the PCB OP BOARD is set to OFF position then expose X-ray.
j. If overshooting or undershooting occurs in the waveform, set the calibration data again.
k. Then push the KV UP switch (see operation sheet).
l. The display should indicate as follows,

Where, P01 or P06 is Filament preheat calibration code (Preheat calibration code of 35mA) and
0.67 or 0.69 is Filament preheat calibration data (Preheat calibration data of 35mA).

m. Adjust data by using the mAs Up-Down switch on the Control Panel.
n. After it is adjusted, push the Store switch during 3sec(see operation sheet).
o. DIP SWITCH #4 on the PCB OP BOARD is set to OFF position then exposure X-ray.
p. If overshooting or undershooting occurs in the waveform, set the calibration data again.

14
EPX-SERISE SERVICE MANUAL Rev. E

1.3 EPX-F2800
1.3.1 Line Power Switch

Line switch should be at off position.


Connect power line cord to the power outlet.
Turn on the line switch.
The FND on the control panel should illuminate.

1.3.2 Calibration Procedure


1.3.2.1 Tube voltage confirmation and calibration.

a. Turn off the line switch.


b. Detach the control panel from the unit to access the PCB OP BOARD.
c. Connect the channel one probe of the oscilloscope to TP8 (IP F/B).
d. Connect the channel two probe of the oscilloscope to TP11 (EP F/B).
e. Connect the ground to TP13 or TP14(GND) on the OP board.
f. Turn on the line switch.
g. Set the kV and the mAs at 40KV and 5mAs(40mA), respectively.
h. Then observe the KV waveform of the oscilloscope channel 2.
i. If it is not 2V, adjust KV by turning VR3 on the OP board.

j. Set the kV and the mAs at 120KV and 5mAs(16mA), respectively.


k. Then observe the KV waveform of the oscilloscope channel 2.
l. If it is not 6V, adjust KV by turning VR3 on the OP board.

15
EPX-SERISE SERVICE MANUAL Rev. E

1.3.2.2 Tube current confirmation and calibration.

a. Turn off the line switch.


b. Detach the control panel from the unit to access the PCB OP BOARD.
c. Connect the channel one (1) probe of the oscilloscope to TP8 (IP F/B).
d. Connect the channel two (2) probe of the oscilloscope to TP11 (EP F/B).
e. Connect the ground to TP13 or TP14 on the OP board.
f. Turn on the line switch.
g. Set the kV and the mAs at 40kV and 5mAs(40mA), respectively.
h. Then observe the mA waveform of the oscilloscope channel 1.
i. If it is not 6V, adjust mA using the calibration data on the OP BOARD.
j. DIP SWITCH #4 on the OP BOARD PCB is set to ON position.

Then the display should indicate as follows,

Where, C00~C15 are calibration code of tube current at AC110V model and C16~C31 are calibration
code of tube current at AC220V model.
C00 or C16 is calibration code of tube current (40mA at 40kV~60kV) at AC110V or AC220V and 40.0
is calibration data of tube current (40mA at 40kV~60kV).

16
EPX-SERISE SERVICE MANUAL Rev. E

a. Adjust mA by using the mAs Up-Down switch (see operation sheet).


b. After it is adjusted, push the STORE button during 3sec(see operation sheet).
c. DIP SWITCH #4 on the PCB OP BOARD should be set to OFF position and the expose X-ray.
e. If it is not 6V, set the calibration data again.
f. DIP SWITCH #4 on the OP BOARD PCB is set to ON position.
g. Then push the kV UP switch (see operation sheet).
h. The display should indicate as follows,

Where, C01 or C17 is calibration code of tube current (30mA at 61kV~70kV) and 30.0 is calibration
data of tube current (30mA at 61kV~70kV).

i. Adjust mA by using the mAs Up-Down switch (see operation sheet).


j. After it is adjusted, push the STORE switch during 3sec(see operation sheet).
k. DIP SWITCH #4 on the PCB OP BOARD is set to OFF position then expose X-ray.
l. If it is not 5.15V, set the calibration data again.

Filament preheat calibration


This is preheating filament for making tube current.
The overshooting or the undershooting may occur from the tube current if the Preheat calibration data
is set by mistake.
Changing the calibration data is possible in accordance with the tube condition.

a. Turn off the line switch.


b. Detach the control panel from the unit to access the PCB OP BOARD.
c. DIP SWITCH #4 on the PCB OP BOARD is set to ON position.
d. Turn on the line switch.
f. Push APR S/W ① and then the display should indicate as follows,

17
EPX-SERISE SERVICE MANUAL Rev. E

where, P00~P06 are calibration code of filament preheat current at AC110V, P07~P13 are calibration
code of filament preheat current at AC220V, P00 or P07 is Filament preheat calibration code (Preheat
calibration code of 35mA), and 0.70 or 0.70 is Filament preheat calibration data (Preheat calibration
data of 35mA).

g. Adjust data by using the mAs Up-Down switch (see operation sheet).
h. After it is adjusted, push the STORE button during 3sec(see operation sheet).
i. DIP SWITCH #4 on the PCB OP BOARD is set to OFF position then expose X-ray.
j. If overshooting or undershooting occurs in the waveform, set the calibration data again.
k. Then push the KV UP switch (see operation sheet).
l. The display should indicate as follows,

Where, P01 or P08 is Filament preheat calibration code (Preheat calibration code of 30mA) and 0.67
or 0.67 is Filament preheat calibration data (Preheat calibration data of 30mA).

m. Adjust data by using the mAs Up-Down switch on the Control Panel.
n. After it is adjusted, push the Store switch during 3sec(see operation sheet).
o. DIP SWITCH #4 on the PCB OP BOARD is set to OFF position then exposure X-ray.
p. If overshooting or undershooting occurs in the waveform, set the calibration data again.

18
EPX-SERISE SERVICE MANUAL Rev. E

1.4 EPX-F3200
1.4.1 Line Power Switch

Line switch should be at off position.


Connect power line cord to the power outlet.
Turn on the line switch.
The FND on the control panel should illuminate.

1.4.2 Calibration Procedure


1.4.2.1 Tube voltage confirmation and calibration.

a. Turn off the line switch.


b. Detach the control panel from the unit to access the PCB OP BOARD.
c. Connect the channel one probe of the oscilloscope to TP8 (IP F/B).
d. Connect the channel two probe of the oscilloscope to TP11 (EP F/B).
e. Connect the ground to TP13 or TP14 on the OP board.
f. Turn on the line switch.
g. Set the kV and the mAs at 40KV and 5mAs(60mA), respectively.
h. Then observe the KV waveform of the oscilloscope channel 2.
i. If it is not 2V, adjust KV by turning VR3 on the OP board.

j. Set the kV and the mAs at 100KV and 5mAs(30mA), respectively.


k. Then observe the KV waveform of the oscilloscope channel 2.
l. If it is not 5V, adjust KV by turning VR3 on the OP board.

19
EPX-SERISE SERVICE MANUAL Rev. E

1.4.2.2 Tube current confirmation and calibration.

a. Turn off the line switch.


b. Detach the control panel from the unit to access the PCB OP BOARD.
c. Connect the channel one (1) probe of the oscilloscope to TP8 (IP F/B).
d. Connect the channel two (2) probe of the oscilloscope to TP11 (EP F/B).
e. Connect the ground to TP13 or TP14 or on the OP board.
f. Turn on the line switch.
g. Set the kV and the mAs at 40kV and 5mAs(60mA), respectively.
h. Then observe the mA waveform of the oscilloscope channel 1.
i. If it is not 6.0 V, adjust mA using the calibration data on the OP BOARD.
j. DIP SWITCH #4 on the OP BOARD PCB is set to ON position.

Then the display should indicate as follows,

Where, C00~C07 are calibration code of tube current at AC110V model and C08~C15 are calibration
code of tube current at AC220V model. C00 or C08 is calibration code of tube current (60mA at
40kV~50kV) at AC110V and 60.0 is calibration data of tube current (60mA at 40kV~50kV).

a. Adjust mA by using the mAs Up-Down switch (see operation sheet).


b. After it is adjusted, push the STORE button during 3sec(see operation sheet).

20
EPX-SERISE SERVICE MANUAL Rev. E

c. DIP SWITCH #4 on the PCB OP BOARD should be set to OFF position and the expose X-ray.
e. If it is not 6.0V, set the calibration data again.
f. DIP SWITCH #4 on the OP BOARD PCB is set to ON position.
g. Then push the kV UP switch (see operation sheet).
h. The display should indicate as follows,

Where, C01 or C09 is calibration code of tube current (50mA at 51kV~60kV) and 50.0 is calibration
data of tube current (50mA at 51kV~60kV).

i. Adjust mA by using the mAs Up-Down switch (see operation sheet).


j. After it is adjusted, push the STORE switch during 3sec(see operation sheet).
k. DIP SWITCH #4 on the PCB OP BOARD is set to OFF position then expose X-ray.
l. If it is not 5.0V, set the calibration data again.

Filament preheat calibration


This is preheating filament for making tube current.
The overshooting or the undershooting may occur from the tube current if the Preheat calibration
data is set by mistake.
Changing the calibration data is possible in accordance with the tube condition.
a. Turn off the line switch.
b. Detach the control panel from the unit to access the PCB OP BOARD.
c. DIP SWITCH #4 on the PCB OP BOARD is set to ON position.
d. Turn on the line switch.
f. Push APR S/W ① and then the display should indicate as follows,

21
EPX-SERISE SERVICE MANUAL Rev. E

Where, P00~P05 are calibration code of filament preheat current at AC110V, P06~P11 are calibration
code of filament preheat current at AC220V, P00 or P06 is Filament preheat calibration code (Preheat
calibration code of 60mA), and 1.65 or 1.65 is Filament preheat calibration data (Preheat calibration
data of 60mA).

g. Adjust data by using the mAs Up-Down switch (see operation sheet).
h. After it is adjusted, push the STORE button during 3sec(see operation sheet).
i. DIP SWITCH #4 on the PCB OP BOARD is set to OFF position then expose X-ray.
j. If overshooting or undershooting occurs in the waveform, set the calibration data again.
k. Then push the KV UP switch (see operation sheet).
l. The display should indicate as follows,

where, P01 or P07 is Filament preheat calibration code (Preheat calibration code of 50mA) and 1.60
or 1.60 is Filament preheat calibration data (Preheat calibration data of 50mA).

m. Adjust data by using the mAs Up-Down switch on the Control Panel.
n. After it is adjusted, push the Store switch during 3sec(see operation sheet).
o. DIP SWITCH #4 on the PCB OP BOARD is set to OFF position then exposure X-ray.
p. If overshooting or undershooting occurs in the waveform, set the calibration data again.

22
EPX-SERISE SERVICE MANUAL Rev. E

1.5 EPX-F4000
1.5.1 Line Power Switch

Line switch should be at off position.


Connect power line cord to the power outlet.
Turn on the line switch.
The FND on the control panel should illuminate.

1.5.2 Calibration Procedure


1.5.2.1 Tube voltage confirmation and calibration.

a. Turn off the line switch.


b. Detach the control panel from the unit to access the PCB OP BOARD.
c. Connect the channel one probe of the oscilloscope to TP8 (IP F/B).
d. Connect the channel two probe of the oscilloscope to TP11 (EP F/B).
e. Connect the ground to TP13 or TP14 on the OP board.
f. Turn on the line switch.
g. Set the kV and the mAs at 50KV and 5mAs(80mA), respectively.
h. Then observe the KV waveform of the oscilloscope channel 2.
i. If it is not 2.5V, adjust KV by turning VR3 on the OP board.

j. Set the kV and the mAs at 110KV and 5mAs(25mA), respectively.


k. Then observe the KV waveform of the oscilloscope channel 2.
l. If it is not 5.5V, adjust KV by turning VR3 on the OP board.

23
EPX-SERISE SERVICE MANUAL Rev. E

1.5.2.2 Tube current confirmation and calibration.

a. Turn off the line switch.


b. Detach the control panel from the unit to access the PCB OP BOARD.
c. Connect the channel one (1) probe of the oscilloscope to TP8 (IP F/B).
d. Connect the channel two (2) probe of the oscilloscope to TP11 (EP F/B).
e. Connect the ground to TP13 or TP14 on the OP board.
f. Turn on the line switch.
g. Set the kV and the mAs at 50kV and 5mAs(80mA), respectively.
h. Then observe the mA waveform of the oscilloscope channel 1.
i. If it is not 5.44V, adjust mA using the calibration data on the OP BOARD.
j. DIP SWITCH #4 on the OP BOARD PCB is set to ON position.

Then the display should indicate as follows,

Where, C00~C14 are calibration code of tube current at AC110V model and C15~C29 are calibration
code of tube current at AC220V model. C00 or C15 is calibration code of tube current (80mA at 50kV)
at 80.0 is calibration data of tube current (80mA at 50kV).

a. Adjust mA by using the mAs Up-Down switch (see operation sheet).


b. After it is adjusted, push the STORE button during 3sec(see operation sheet).

24
EPX-SERISE SERVICE MANUAL Rev. E

c. DIP SWITCH #4 on the PCB OP BOARD should be set to OFF position and the expose X-ray.
e. If it is not 4V, set the calibration data again.
f. DIP SWITCH #4 on the OP BOARD PCB is set to ON position.
g. Then push the kV UP switch (see operation sheet).
h. The display should indicate as follows,

Where, C01 or C16 is calibration code of tube current (70mA at 50kV) and 70.0 is calibration data of
tube current (70mA at 50kV).

i. Adjust mA by using the mAs Up-Down switch (see operation sheet).


j. After it is adjusted, push the STORE switch during 3sec(see operation sheet).
k. DIP SWITCH #4 on the PCB OP BOARD is set to OFF position then expose X-ray.
l. If it is not 3.0V, set the calibration data again.

Filament preheat calibration


This is preheating filament for making tube current.
The overshooting or the undershooting may occur from the tube current if the Preheat calibration
data is set by mistake.
Changing the calibration data is possible in accordance with the tube condition.
a. Turn off the line switch.
b. Detach the control panel from the unit to access the PCB OP BOARD.
c. DIP SWITCH #4 on the PCB OP BOARD is set to ON position.
d. Turn on the line switch.
f. Push APR S/W ① and then the display should indicate as follows,

25
EPX-SERISE SERVICE MANUAL Rev. E

Where, P00~P09 are calibration code of filament preheat current at AC110V, P10~P19 are calibration
code of filament preheat current at AC220V, P00 or P10 is Filament preheat calibration code (Preheat
calibration code of 80mA), and 2.50 or 2.50 is Filament preheat calibration data (Preheat calibration
data of 80mA).

g. Adjust data by using the mAs Up-Down switch (see operation sheet).
h. After it is adjusted, push the STORE button during 3sec(see operation sheet).
i. DIP SWITCH #4 on the PCB OP BOARD is set to OFF position then expose X-ray.
j. If overshooting or undershooting occurs in the waveform, set the calibration data again.
k. Then push the KV UP switch (see operation sheet).
l. The display should indicate as follows,

Where, P01 or P11 is Filament preheat calibration code (Preheat calibration code of 70mA) and 2.20
or 2.20 is Filament preheat calibration data (Preheat calibration data of 70mA).

m. Adjust data by using the mAs Up-Down switch on the Control Panel.
n. After it is adjusted, push the Store switch during 3sec(see operation sheet).
o. DIP SWITCH #4 on the PCB OP BOARD is set to OFF position then exposure X-ray.
p. If overshooting or undershooting occurs in the waveform, set the calibration data again.

26
EPX-SERISE SERVICE MANUAL Rev. E

1.6 EPX-F5000
1.6.1 Line Power Switch

Line switch should be at off position.


Connect power line cord to the power outlet.
Turn on the line switch.
The FND on the control panel should illuminate.

1.6.2 Calibration Procedure


1.6.2.1 Tube voltage confirmation and calibration.

a. Turn off the line switch.


b. Detach the control panel from the unit to access the PCB OP BOARD.
c. Connect the channel one probe of the oscilloscope to TP8 (IP F/B).
d. Connect the channel two probe of the oscilloscope to TP11 (EP F/B).
e. Connect the ground to TP13 or TP14 or on the OP board.
f. Turn on the line switch.
g. Set the kV and the mAs at 40KV and 5mAs(100mA), respectively.
h. Then observe the KV waveform of the oscilloscope channel 2.
i. If it is not 2V, adjust KV by turning VR3 on the OP board.

j. Set the kV and the mAs at 110KV and 5mAs(40mA), respectively.


k. Then observe the KV waveform of the oscilloscope channel 2.
l. If it is not 5.5V, adjust KV by turning VR3 on the OP board.

27
EPX-SERISE SERVICE MANUAL Rev. E

1.6.2.2 Tube current confirmation and calibration.

a. Turn off the line switch.


b. Detach the control panel from the unit to access the PCB OP BOARD.
c. Connect the channel one (1) probe of the oscilloscope to TP8 (IP F/B).
d. Connect the channel two (2) probe of the oscilloscope to TP11 (EP F/B).
e. Connect the ground to TP13 or TP14 or on the OP board.
f. Turn on the line switch.
g. Set the kV and the mAs at 40kV and 5mAs(100mA), respectively.
h. Then observe the mA waveform of the oscilloscope channel 1.
i. If it is not 5.1V, adjust mA using the calibration data on the OP BOARD.
j. DIP SWITCH #4 on the OP BOARD PCB is set to ON position.

Then the display should indicate as follows,

Where, C00~C20 are calibration code of tube current at AC110V model and C21~C41 are calibration
code of tube current at AC220V model. C00 or C21 is calibration code of tube current (100mA at
50kV) at A0.0(=100) is calibration data of tube current (100mA at 50kV).

a. Adjust mA by using the mAs Up-Down switch (see operation sheet).

28
EPX-SERISE SERVICE MANUAL Rev. E

b. After it is adjusted, push the STORE button during 3sec(see operation sheet).
c. DIP SWITCH #4 on the PCB OP BOARD should be set to OFF position and the expose X-ray.
e. If it is not 5.1V, set the calibration data again.
f. DIP SWITCH #4 on the OP BOARD PCB is set to ON position.
g. Then push the kV UP switch (see operation sheet).
h. The display should indicate as follows,

Where, C01 or C22 is calibration code of tube current (80mA at 50kV) and 80.0 is calibration data of
tube current (80mA at 50kV).

i. Adjust mA by using the mAs Up-Down switch (see operation sheet).


j. After it is adjusted, push the STORE switch during 3sec(see operation sheet).
k. DIP SWITCH #4 on the PCB OP BOARD is set to OFF position then expose X-ray.
l. If it is not 4.08V, set the calibration data again.

Filament preheat calibration


This is preheating filament for making tube current.
The overshooting or the undershooting may occur from the tube current if the Preheat calibration data
is set by mistake.
Changing the calibration data is possible in accordance with the tube condition.

a. Turn off the line switch.


b. Detach the control panel from the unit to access the PCB OP BOARD.
c. DIP SWITCH #4 on the PCB OP BOARD is set to ON position.
d. Turn on the line switch.
f. Push APR S/W ① and then the display should indicate as follows,

29
EPX-SERISE SERVICE MANUAL Rev. E

Where, P00~P08 are calibration code of filament preheat current at AC110V, P09~P15 are calibration
code of filament preheat current at AC220V, P00 or P09 is Filament preheat calibration code (Preheat
calibration code of 100mA), and 2.40 or 2.40 is Filament preheat calibration data (Preheat calibration
data of 100mA).

g. Adjust data by using the mAs Up-Down switch (see operation sheet).
h. After it is adjusted, push the STORE button during 3sec(see operation sheet).
i. DIP SWITCH #4 on the PCB OP BOARD is set to OFF position then expose X-ray.
j. If overshooting or undershooting occurs in the waveform, set the calibration data again.
k. Then push the KV UP switch (see operation sheet).
l. The display should indicate as follows,

Where, P01 or P10 is Filament preheat calibration code (Preheat calibration code of 80mA) and 2.25
or 2.25 is Filament preheat calibration data (Preheat calibration data of 80mA).

m. Adjust data by using the mAs Up-Down switch on the Control Panel.
n. After it is adjusted, push the Store switch during 3sec(see operation sheet).
o. DIP SWITCH #4 on the PCB OP BOARD is set to OFF position then exposure X-ray.
p. If overshooting or undershooting occurs in the waveform, set the calibration data again.

30
EPX-SERISE SERVICE MANUAL Rev. E

2. DIP Switch setting

DIP Switches are used for following setting.

SW #1: ERROR Enable (ON)/Disable (OFF)


SW #2: Reserved
SW #3: Initialize memories (reserved)
SW #4: Calibration mode (ON)/Normal mode (Off)

For normal operation, only the DIP SW#1 must be set to ON position.

31
EPX-SERISE SERVICE MANUAL Rev. E

3. Hard-wired hand switch control


3.1 Introduction
The Exposure Hand Switch consists of a two step hand switch.
The first (1) stage of the hand switch is called the “Preparation (Prep or READY)” stage.
The second (2) stage of the hand switch is called the “Exposure (Exp or X-Ray Out)“ stage.

SWITCH 1
SWITCH 2

3.2 Exposure process


There are two ways to use the hand switch to get the exposure.

a. Press and hold the Switch #1 for longer than 1 second, which will get the “Prep or
Ready” status on, then press Switch #2 to get the “Exposure”.

b. Press and hold the Switch #1 and Switch #2 simultaneously for longer than 1 second.
This action will result in “Prep or Ready” status and “Exposure” at the same time.

32
EPX-SERISE SERVICE MANUAL Rev. E

4. Collimator

a. Halogen Lamp * 1
b. Laser pointer * 2
c. Horizontal and vertical lighting range control

i. Block diagram

ii. Layout
a. External Layout

33
EPX-SERISE SERVICE MANUAL Rev. E

b. Internal Layout

c. Knob Layout

34
EPX-SERISE SERVICE MANUAL Rev. E

5. Trouble shooting
If you have any error message when you are using portable x-ray, please check below service
check list

5.1 EPX-F1600/F2400/F2800/F3200/F4000/F5000 Error Code Table

Error
DESCRIPTION INDICATION
Code
2 THERMAL ERROR (INTERLOCK) Excess from 60℃

7 READY SW ERR Ready Switch Malfunction

8 EXPOSURE SW ERROR Exposure Switch Malfunction

10 STANDBY FIL FB ERROR Excess from 200mA (10% of Max. Current, 2A)

11 STANDBY EP FB ERROR Excess from 12kV (10% of Max. Voltage, 120kV)

12 STANDBY IP FB ERROR Excess from 3.5mA (10% of Max. Current, 35mA)

14 RUN FIL FB ERROR Out of Range from Reference ±10%

RUN EP FB or AC POWER LOW Out of Range from Reference -10% or AC Power


15
ERROR Line poor

16 RUN IP FB HIGH ERROR Out of Range from Reference +10%

17 RUN EP FB HIGH ERROR Out of Range from Reference +10%

18 RUN IP FB LOW ERROR Out of Range from Reference -10%

28 PULSE BY PULSE ERROR FET Board Fault (Hardware)

29 FET-OTP ERROR FET board overheat

30 KV_OVP ERROR kV Over Voltage (Hardware)

31 FIL_OC ERROR FIL Over Current (Hardware)

32 IP_OCP ERROR mA Over Current (Hardware)

35
EPX-SERISE SERVICE MANUAL Rev. E

5.2 Action to correct the error of EPX-Series

Error Codes Check points and Action

7 Ready SW ERR
a. Replace the hand switch
8 Exposure SW ERR

10 Standby Fil FB Error

a. If error repeats at the condition of 40kV 0.4mAs EXP, replace the OP


11 Standby EP FB Error
board
12 Standby IP FB Error

14 Run Fil FB Error

15 Run EP FB low ERR a. If error repeats at the condition of 40kV 0.4mAs EXP, check the
followings by sequence.
16 Run IP FB High ERR 1. Check TP in the power input of the OP board
2. Check the Reference voltage value
17 Run EP FB High ERR 3. If the reference voltage is normal, replace the FET board
4. If replacing FET board does not solve the problem, the replace the
18 Run IP FB low ERR HV tank.

28 Pulse by Pulse ERR

30 KV_OVP ERR a. If error pops up right after the unit power is turned on, then replace
the OP board.
b. If error repeats at the condition of 40kV 0.4mAs EXP,
31 FIL_OC ERR FIL
1. Replace the FET board
2. If replacing FET board does not solve the problem, the replace the
32 IP_OCP ERR HV tank.

Warning Codes
Wait until the temperature inside the monotank comes down to normal
Thermal ERR excess
2 (Once the temperature drop to normal, the Warning Message will
from 60°C
disappear)
Wait until the temperature inside the FET board comes down to normal
FET-OTP FET board
29 (Once the temperature drop to normal, the Warning Message will
overheat
disappear)

36
EPX-SERISE SERVICE MANUAL Rev. E

6. Service Check List


6.1 EPX-F1600

Please fill in the blank.

kV/mAs condition(When
Product S/N Error code
error is occurred)

※ Open upper cover of portable x-ray below picture.

6.1.1 Voltage check


1) Turn on AC switch

TP26 TP24 TP25 TP6


VDC Range 4.8V~5.0V 11.5V~12.5V -11.5V~-12.5V +3V~3.6V
Pass Fail Pass Fail Pass Fail Pass Fail
Pass/Fail

Measured Voltage
If it is fail case
6.1.2. kV, mA waveform check

37
EPX-SERISE SERVICE MANUAL Rev. E

1) Turn off AC switch


2) Please connect probe to EP_FB (kV feedback_TP11), IP_FB (mA feedback_TP8) test point on operation
board below picture.
3)Turn on AC switch.

4) 40kV/5mAs waveform
Please set kV, mAs below table and take exposure
Exposure condition : 40kV/5mAs
CH 1: kV 2V/div CH 2 : mA 2V/div 40msec/div

Setting value EP_FB IP_FB


Ref voltage 40kV 2V 6V
Pass/Fail Pass Fail Pass Fail
Measured kV feedback voltage

If kV, mA waveforms are fail, please attach your kV, mA waveform

5) 50kV/5mAs waveform

38
EPX-SERISE SERVICE MANUAL Rev. E

Please set kV, mAs below table and take exposure


Exposure condition : 50kV/5mAs
CH 1: kV 2V/div CH 2 : mA 2V/div 40msec/div

Setting value EP_FB IP_FB


Ref voltage 50kV 2.5V 6V
Pass/Fail Pass Fail Pass Fail
Measured kV feedback voltage

If kV,mA waveforms are fail, please attach your kV, mA waveform

6) 60kV/5mAs waveform
Please set kV, mAs below table and take exposure
Exposure condition : 60kV/5mAs
CH 1: kV 2V/div CH 2 : mA 2V/div 40msec/div

Setting value EP_FB IP_FB


Ref voltage 60kV 3V 5V
Pass/Fail Pass Fail Pass Fail
Measured kV feedback voltage

If kV,mA waveforms are fail, please attach your kV, mA waveform


7) 70kV/5mAs waveform

39
EPX-SERISE SERVICE MANUAL Rev. E

Please set kV, mAs below table and take exposure


Exposure condition : 70kV/5mAs
CH 1: kV 2V/div CH 2 : mA 2V/div 40msec/div

Setting value EP_FB IP_FB


Ref voltage 70kV 3.5V 4V
Pass/Fail Pass Fail Pass Fail
Measured kV feedback voltage

If kV,mA waveforms are fail, please attach your kV, mA waveform

8) 80kV/5mAs waveform
Please set kV, mAs below table and take exposure
Exposure condition : 80kV/5mAs
CH 1: kV 2V/div CH 2 : mA 2V/div 100msec/div

Setting value EP_FB IP_FB


Ref voltage 80kV 4V 3.6V
Pass/Fail Pass Fail Pass Fail
Measured kV feedback voltage

If kV,mA waveforms are fail, please attach your kV, mA waveform

9) 90kV/5mAs waveform

40
EPX-SERISE SERVICE MANUAL Rev. E

Please set kV, mAs below table and take exposure


Exposure condition : 90kV/5mAs
CH 1: kV 2V/div CH 2 : mA 2V/div 100msec/div

Setting value EP_FB IP_FB


Ref voltage 90kV 4.5V 3.1V
Pass/Fail Pass Fail Pass Fail
Measured kV feedback voltage

If kV,mA waveforms are fail, please attach your kV, mA waveform

6.1.3. Waveform of poor AC power condition


If AC power is under 100V or 200VAC, you have below abnormal waveform. Error 015,016,017,018 is
occurred by poor AC power.

6.1.4 kV, mA PWM frequency check

41
EPX-SERISE SERVICE MANUAL Rev. E

1) Turn off AC swtich


2) Please connect probe to 200kHz(TP15), 140kHz(TP16) test point on operation board below picture.
3) Turn on AC switch.

4) kV / mA PWM frequency

CH 1: kV 2V/div CH 2 : mA 2V/div 4μsec/div

CH CH 1 mA PWM CH 2 kV PWM
Frequency 200kHz 140kHz
Pass/Fail Pass Fail Pass Fail
Measured frequency

6.1.5 mA calibration data check.

42
EPX-SERISE SERVICE MANUAL Rev. E

DIP Switches are used for following setting.


SW #1: ERROR Enable (ON)/Disable (OFF)
SW #2: Reserved
SW #3: Initialize memories (reserved)
SW #4: Calibration mode (ON)/Normal mode (Off)
For normal operation, only the DIP SW#1 must be set to ON position.
DIP SWITCH #4 on the OP BOARD PCB is set to ON position.
Then the display should indicate as follows,

3) Turn on AC switch.

Calibration code

Input data

Up/Down switches increase and decrease C00~C09(AC110V), C10~C19(AC220V)


Please check Input data.

Input data and Tube current should be the same.

43
EPX-SERISE SERVICE MANUAL Rev. E

Tube Current
Tube Voltage mAs Tube Current INPUT DATA
Parameter
AC110V AC220V KV MIN MAX mA mA

C00 C10 40~50 (1) 0.4 20 30.0 30.0

C01 C11 40~50 (2) 25 32 25.0 25.0

C02 C12 51~60 (1) 0.4 20 25.0 25.0

C03 C13 51~60 (2) 25 32 20.0 20.0

C04 C14 61~70 (1) 0.4 20 20.0 20.0

C05 C15 61~70 (2) 25 32 15.0 15.0

C06 C16 71~80 (1) 0.4 20 18 18

C07 C17 71~80 (2) 25 32 14 14

C08 C18 81~90 (1) 0.4 20 16 16

C09 C19 81~90 (2) 25 32 12 12

4) If mA calibration is done, Please dip switch 4 set to OFF position.

6.1.6 Filament Preheat Calibration Code and Data check

1) Turn off AC switch


2) Dip switch on op board

DIP Switches are used for following setting.


SW #1: ERROR Enable (ON)/Disable (OFF)
SW #2: Reserved
SW #3: Initialize memories (reserved)
SW #4: Calibration mode (ON)/Normal mode (Off)
For normal operation, only the DIP SW#1 must be set to ON position.

44
EPX-SERISE SERVICE MANUAL Rev. E

DIP SWITCH #4 on the OP BOARD PCB is set to ON position.


Then the display should indicate as follows,

3) Turn on AC switch.
4) In mA calibration mode, Press the number 1 switch as shown in figure. Mode is converted.

P_Calibration code

Input data

Up/Down switches increase and decrease P00~P07(AC110V), P08~P15(AC220V)


Please check Input data.
Input data and Tube current should be the same.

Fill in the blank below table because P_DATA is different for each product.

Model Current(mA) P_Mode(110V) P_Mode(220V) P_DATA


30 P00 P08
25 P01 P09
20 P02 P10
18 P03 P11
1.6kW
16 P04 P12
15 P05 P13
14 P06 P14
12 P07 P15

5) If Preheat calibration is done, Please dip switch 4 set to OFF position.

45
EPX-SERISE SERVICE MANUAL Rev. E

6.2 EPX-F2400

Please fill in the blank.

kV/mAs condition(When
Product S/N Error code
error is occurred)

※ Open upper cover of portable x-ray below picture.

6.2.1 Voltage check


1) Turn on AC switch

TP26 TP24 TP25 TP6


VDC Range 4.8V~5.0V 11.5V~12.5V -11.5V~-12.5V +3V~3.6V
Pass Fail Pass Fail Pass Fail Pass Fail
Pass/Fail

Measured Voltage
If it is fail case

46
EPX-SERISE SERVICE MANUAL Rev. E

6.2.2. kV, mA waveform check


1) Turn off AC switch
2) Please connect probe to EP_FB (kV feedback_TP11), IP_FB (mA feedback_TP8) test point on operation
board below picture.
3)Turn on AC switch.

4) 40kV/5mAs waveform
Please set kV, mAs below table and take exposure
Exposure condition : 40kV/5mAs
CH 1: kV 2V/div CH 2 : mA 2V/div 40msec/div

Setting value EP_FB IP_FB


Ref voltage 40kV 2V 6V
Pass/Fail Pass Fail Pass Fail
Measured kV feedback voltage

If kV, mA waveforms are fail, please attach your kV, mA waveform

47
EPX-SERISE SERVICE MANUAL Rev. E

5) 50kV/5mAs waveform
Please set kV, mAs below table and take exposure
Exposure condition : 50kV/5mAs
CH 1: kV 2V/div CH 2 : mA 2V/div 40msec/div

Setting value EP_FB IP_FB


Ref voltage 50kV 2.5V 6V
Pass/Fail Pass Fail Pass Fail
Measured kV feedback voltage

If kV,mA waveforms are fail, please attach your kV, mA waveform

6) 60kV/5mAs waveform
Please set kV, mAs below table and take exposure
Exposure condition : 60kV/5mAs
CH 1: kV 2V/div CH 2 : mA 2V/div 40msec/div

Setting value EP_FB IP_FB


Ref voltage 60kV 3V 6V
Pass/Fail Pass Fail Pass Fail
Measured kV feedback voltage

If kV,mA waveforms are fail, please attach your kV, mA waveform

48
EPX-SERISE SERVICE MANUAL Rev. E

7) 70kV/5mAs waveform
Please set kV, mAs below table and take exposure
Exposure condition : 70kV/5mAs
CH 1: kV 2V/div CH 2 : mA 2V/div 40msec/div

Setting value EP_FB IP_FB


Ref voltage 70kV 3.5V 4.5V
Pass/Fail Pass Fail Pass Fail
Measured kV feedback voltage

If kV,mA waveforms are fail, please attach your kV, mA waveform

8) 80kV/5mAs waveform
Please set kV, mAs below table and take exposure
Exposure condition : 80kV/5mAs
CH 1: kV 2V/div CH 2 : mA 2V/div 40msec/div

Setting value EP_FB IP_FB


Ref voltage 80kV 4V 4.5V
Pass/Fail Pass Fail Pass Fail
Measured kV feedback voltage

If kV,mA waveforms are fail, please attach your kV, mA waveform

49
EPX-SERISE SERVICE MANUAL Rev. E

9) 90kV/5mAs waveform
Please set kV, mAs below table and take exposure
Exposure condition : 90kV/5mAs
CH 1: kV 2V/div CH 2 : mA 2V/div 40msec/div

Setting value EP_FB IP_FB


Ref voltage 90kV 4.5V 3.75V
Pass/Fail Pass Fail Pass Fail
Measured kV feedback voltage

If kV,mA waveforms are fail, please attach your kV, mA waveform

10) 100kV/5mAs waveform


Please set kV, mAs below table and take exposure
Exposure condition : 100kV/5mAs
CH 1: kV 2V/div CH 2 : mA 2V/div 40msec/div

Setting value EP_FB IP_FB


Ref voltage 100kV 5V 3V
Pass/Fail Pass Fail Pass Fail
Measured kV feedback voltage

If kV,mA waveforms are fail, please attach your kV,mA waveform

50
EPX-SERISE SERVICE MANUAL Rev. E

6.2.3. Waveform of poor AC power condition


If AC power is under 100V or 200VAC, you have below abnormal waveform. Error 015,016,017,018 is
occurred by poor AC power.

6.2.4 kV, mA PWM frequency check


1) Turn off AC swtich
2) Please connect probe to 200kHz(TP15), 140kHz(TP16) test point on operation board below picture.
3) Turn on AC switch.

51
EPX-SERISE SERVICE MANUAL Rev. E

4) kV / mA PWM frequency

CH 1: kV 2V/div CH 2 : mA 2V/div 4μsec/div

CH CH 1 mA PWM CH 2 kV PWM
Frequency 200kHz 140kHz
Pass/Fail Pass Fail Pass Fail
Measured frequency

6.2.5 mA calibration data check.

DIP Switches are used for following setting.


SW #1: ERROR Enable (ON)/Disable (OFF)
SW #2: Reserved
SW #3: Initialize memories (reserved)
SW #4: Calibration mode (ON)/Normal mode (Off)
For normal operation, only the DIP SW#1 must be set to ON position.
DIP SWITCH #4 on the OP BOARD PCB is set to ON position.
Then the display should indicate as follows,

52
EPX-SERISE SERVICE MANUAL Rev. E

3) Turn on AC switch.

Calibration code

Input data

Up/Down switches increase and decrease C00~C07(AC110V), C08~C15(AC220V)


Please check Input data.

Input data and Tube current should be the same.

Tube Current
Tube Voltage mAs Tube Current INPUT DATA
Parameter
AC110V AC220V KV MIN MAX mA mA

C00 C08 40~60(1) 0.4 64 40.0 40.0


C01 C09 40~60(2) 80 100 35.0 30.0
C02 C10 61~70(1) 0.4 100 30.0 25.0
C03 C11 71~80(1) 0.4 32 30.0 30.0
C04 C12 71~80(2) 40 80 25.0 25.0
C05 C13 81~90(1) 0.4 40 25.0 25.0
C06 C14 81~90(2) 50 80 20.0 20.0
C07 C15 91~100(1) 0.4 80 20.0 20.0

4) If mA calibration is done, Please dip switch 4 set to OFF position.

53
EPX-SERISE SERVICE MANUAL Rev. E

6.2.6 Filament Preheat Calibration Code and Data check

1) Turn off AC switch


2) Dip switch on op board

DIP Switches are used for following setting.


SW #1: ERROR Enable (ON)/Disable (OFF)
SW #2: Reserved
SW #3: Initialize memories (reserved)
SW #4: Calibration mode (ON)/Normal mode (Off)
For normal operation, only the DIP SW#1 must be set to ON position.
DIP SWITCH #4 on the OP BOARD PCB is set to ON position.
Then the display should indicate as follows,

3) Turn on AC switch.
4) In mA calibration mode, Press the number 1 switch as shown in figure. Mode is converted.

P_Calibration code

Input data

Up/Down switches increase and decrease P00~P04(AC110V), P05~P09(AC220V)


Please check Input data.
Input data and Tube current should be the same.

54
EPX-SERISE SERVICE MANUAL Rev. E

Fill in the blank below table because P_DATA is different for each product.

Model Current(mA) P_Mode(110V) P_Mode(220V) P_DATA(A)


40 P00 P05
35 P01 P06
2.4kW 30 P02 P07
25 P03 P08
20 P04 P09

5) If Preheat calibration is done, Please dip switch 4 set to OFF position.

55
EPX-SERISE SERVICE MANUAL Rev. E

6.3 EPX-F2800

Please fill in the blank.

kV/mAs condition(When
Product S/N Error code
error is occurred)

※ Open upper cover of portable x-ray below picture.

6.3.1 Voltage check


1) Turn on AC switch

TP26 TP24 TP25 TP6


VDC Range 4.8V~5.0V 11.5V~12.5V -11.5V~-12.5V +3V~3.6V
Pass Fail Pass Fail Pass Fail Pass Fail
Pass/Fail

Measured Voltage
If it is fail case

56
EPX-SERISE SERVICE MANUAL Rev. E

6.3.2. kV, mA waveform check


1) Turn off AC switch
2) Please connect probe to EP_FB (kV feedback_TP11), IP_FB (mA feedback_TP8) test point on operation
board below picture.
3)Turn on AC switch.

4) 40kV/5mAs waveform
Please set kV, mAs below table and take exposure
Exposure condition : 40kV/5mAs
CH 1: kV 2V/div CH 2 : mA 2V/div 40msec/div

Setting value EP_FB IP_FB


Ref voltage 40kV 2V 6V
Pass/Fail Pass Fail Pass Fail
Measured kV feedback voltage

If kV, mA waveforms are fail, please attach your kV, mA waveform

57
EPX-SERISE SERVICE MANUAL Rev. E

5) 50kV/5mAs waveform
Please set kV, mAs below table and take exposure
Exposure condition : 50kV/5mAs
CH 1: kV 2V/div CH 2 : mA 2V/div 40msec/div

Setting value EP_FB IP_FB


Ref voltage 50kV 2.5V 6V
Pass/Fail Pass Fail Pass Fail
Measured kV feedback voltage

If kV,mA waveforms are fail, please attach your kV, mA waveform

6) 60kV/5mAs waveform
Please set kV, mAs below table and take exposure
Exposure condition : 60kV/5mAs
CH 1: kV 2V/div CH 2 : mA 2V/div 40msec/div

Setting value EP_FB IP_FB


Ref voltage 60kV 3V 6V
Pass/Fail Pass Fail Pass Fail
Measured kV feedback voltage

If kV,mA waveforms are fail, please attach your kV, mA waveform

58
EPX-SERISE SERVICE MANUAL Rev. E

7) 70kV/5mAs waveform
Please set kV, mAs below table and take exposure
Exposure condition : 70kV/5mAs
CH 1: kV 2V/div CH 2 : mA 2V/div 40msec/div

Setting value EP_FB IP_FB


Ref voltage 70kV 3.5V 4.5V
Pass/Fail Pass Fail Pass Fail
Measured kV feedback voltage

If kV,mA waveforms are fail, please attach your kV, mA waveform

8) 80kV/5mAs waveform
Please set kV, mAs below table and take exposure
Exposure condition : 80kV/5mAs
CH 1: kV 2V/div CH 2 : mA 2V/div 40msec/div

Setting value EP_FB IP_FB


Ref voltage 80kV 4V 5.25V
Pass/Fail Pass Fail Pass Fail
Measured kV feedback voltage

If kV,mA waveforms are fail, please attach your kV, mA waveform

59
EPX-SERISE SERVICE MANUAL Rev. E

9) 90kV/5mAs waveform
Please set kV, mAs below table and take exposure
Exposure condition : 90kV/5mAs
CH 1: kV 2V/div CH 2 : mA 2V/div 40msec/div

Setting value EP_FB IP_FB


Ref voltage 90kV 4.5V 4.5V
Pass/Fail Pass Fail Pass Fail
Measured kV feedback voltage

If kV,mA waveforms are fail, please attach your kV, mA waveform

10) 100kV/5mAs waveform


Please set kV, mAs below table and take exposure
Exposure condition : 100kV/5mAs
CH 1: kV 2V/div CH 2 : mA 2V/div 40msec/div

Setting value EP_FB IP_FB


Ref voltage 100kV 5V 3.75V
Pass/Fail Pass Fail Pass Fail
Measured kV feedback voltage

If kV,mA waveforms are fail, please attach your kV,mA waveform

60
EPX-SERISE SERVICE MANUAL Rev. E

11) 110kV/5mAs waveform


Please set kV, mAs below table and take exposure
Exposure condition : 110kV/5mAs
CH 1: kV 2V/div CH 2 : mA 2V/div 40msec/div

Setting value EP_FB IP_FB


Ref voltage 110kV 5.5V 2.7V
Pass/Fail Pass Fail Pass Fail
Measured kV feedback voltage

If kV,mA waveforms are fail, please attach your kV, mA waveform

12) 120kV/5mAs waveform


Please set kV, mAs below table and take exposure
Exposure condition : 120kV/5mAs
CH 1: kV 2V/div CH 2 : mA 2V/div 100msec/div

Setting value EP_FB IP_FB


Ref voltage 120kV 6V 2.4V
Pass/Fail Pass Fail Pass Fail
Measured kV feedback voltage

If kV,mA waveforms are fail, please attach your kV,mA waveform

61
EPX-SERISE SERVICE MANUAL Rev. E

6.3.3. Waveform of poor AC power condition


If AC power is under 100V or 200VAC, you have below abnormal waveform. Error 015,016,017,018 is
occurred by poor AC power.

6.3.4 kV, mA PWM frequency check


1) Turn off AC swtich
2) Please connect probe to 200kHz(TP15), 140kHz(TP16) test point on operation board below picture.
3) Turn on AC switch.

62
EPX-SERISE SERVICE MANUAL Rev. E

4) kV / mA PWM frequency

CH 1: kV 2V/div CH 2 : mA 2V/div 4μsec/div

CH CH 1 mA PWM CH 2 kV PWM
Frequency 200kHz 140kHz
Pass/Fail Pass Fail Pass Fail
Measured frequency

6.3.5 mA calibration data check.

DIP Switches are used for following setting.


SW #1: ERROR Enable (ON)/Disable (OFF)
SW #2: Reserved
SW #3: Initialize memories (reserved)
SW #4: Calibration mode (ON)/Normal mode (Off)
For normal operation, only the DIP SW#1 must be set to ON position.
DIP SWITCH #4 on the OP BOARD PCB is set to ON position.
Then the display should indicate as follows,

63
EPX-SERISE SERVICE MANUAL Rev. E

3) Turn on AC switch.

Calibration code

Input data

Up/Down switches increase and decrease C00~C015(AC110V), C16~C31(AC220V)


Please check Input data.

Input data and Tube current should be the same.

Tube Current
Tube Voltage mAs Tube Current INPUT DATA
Parameter
AC110V AC220V KV MIN MAX mA mA
C00 C16 40~60 0.4 100 40.0 40.0
C01 C17 61~70(1) 0.4 64 30.0 30.0
C02 C18 61~70(2) 80 100 25.0 25.0
C03 C19 71~80(1) 0.4 10 35.0 35.0
C04 C20 71~80(2) 13 32 30.0 30.0
C05 C21 71~80(3) 40 80 25.0 25.0
C06 C22 81~90(1) 0.4 10 30.0 30.0
C07 C23 81~90(2) 13 40 25.0 25.0
C08 C24 81~90(3) 50 80 20.0 20.0
C09 C25 91~100(1) 0.4 13 25.0 25.0
C10 C26 91~100(2) 16 50 20.0 20.0
C11 C27 91~100(3) 64 80 16.0 16.0
C12 C28 101~110(1) 0.4 32 18.0 18.0
C13 C29 101~110(2) 40 64 16.0 16.0
C14 C30 111~120(1) 0.4 20 16.0 16.0
C15 C31 111~120(2) 25 50 14.0 14.0

4) If mA calibration is done, Please dip switch 4 set to OFF position.

64
EPX-SERISE SERVICE MANUAL Rev. E

6.3.6 Filament Preheat Calibration Code and Data check

1) Turn off AC switch


2) Dip switch on op board

DIP Switches are used for following setting.


SW #1: ERROR Enable (ON)/Disable (OFF)
SW #2: Reserved
SW #3: Initialize memories (reserved)
SW #4: Calibration mode (ON)/Normal mode (Off)
For normal operation, only the DIP SW#1 must be set to ON position.
DIP SWITCH #4 on the OP BOARD PCB is set to ON position.
Then the display should indicate as follows,

3) Turn on AC switch.
4) In mA calibration mode, Press the number 1 switch as shown in figure. Mode is converted.

P_Calibration code

Input data

Up/Down switches increase and decrease P00~P06(AC110V), P07~P13(AC220V)


Please check Input data.
Input data and Tube current should be the same.

65
EPX-SERISE SERVICE MANUAL Rev. E

Fill in the blank below table because P_DATA is different for each product.

Model Current(mA) P_Mode(110V) P_Mode(220V) P_DATA(A)


40 P00 P08
35 P01 P09
30 P02 P10
25 P03 P11
2.8kW
20 P04 P12
18 P05 P13
16 P06 P14
14 P07 P15

5) If Preheat calibration is done, Please dip switch 4 set to OFF position.

66
EPX-SERISE SERVICE MANUAL Rev. E

6.4 EPX-F3200

Please fill in the blank.

kV/mAs condition(When
Product S/N Error code
error is occurred)

※ Open upper cover of portable x-ray below picture.

6.4.1 Voltage check


1) Turn on AC switch

TP26 TP24 TP25 TP6

VDC Range 4.8V~5.0V 11.5V~12.5V -11.5V~-12.5V +3V~3.6V


Pass Fail Pass Fail Pass Fail Pass Fail
Pass/Fail

Measured Voltage
If it is fail case

67
EPX-SERISE SERVICE MANUAL Rev. E

6.4.2. kV, mA waveform check


1) Turn off AC switch
2) Please connect probe to EP_FB (kV feedback_TP11), IP_FB (mA feedback_TP8) test point on operation
board below picture.
3)Turn on AC switch.

4) 40kV/5mAs waveform
Please set kV, mAs below table and take exposure
Exposure condition : 40kV/5mAs
CH 1: kV 2V/div CH 2 : mA 2V/div 20msec/div

Setting value EP_FB IP_FB


Ref voltage 40kV 2V 6V
Pass/Fail Pass Fail Pass Fail
Measured kV feedback voltage

If kV, mA waveforms are fail, please attach your kV, mA waveform

68
EPX-SERISE SERVICE MANUAL Rev. E

5) 50kV/5mAs waveform
Please set kV, mAs below table and take exposure
Exposure condition : 50kV/5mAs
CH 1: kV 2V/div CH 2 : mA 2V/div 20msec/div

Setting value EP_FB IP_FB


Ref voltage 50kV 2.5V 6V
Pass/Fail Pass Fail Pass Fail
Measured kV feedback voltage

If kV,mA waveforms are fail, please attach your kV, mA waveform

6) 60kV/5mAs waveform
Please set kV, mAs below table and take exposure
Exposure condition : 60kV/5mAs
CH 1: kV 2V/div CH 2 : mA 2V/div 20msec/div

Setting value EP_FB IP_FB


Ref voltage 60kV 3V 5V
Pass/Fail Pass Fail Pass Fail
Measured kV feedback voltage

If kV,mA waveforms are fail, please attach your kV, mA waveform

69
EPX-SERISE SERVICE MANUAL Rev. E

7) 70kV/5mAs waveform
Please set kV, mAs below table and take exposure
Exposure condition : 70kV/5mAs
CH 1: kV 2V/div CH 2 : mA 2V/div 40msec/div

Setting value EP_FB IP_FB


Ref voltage 70kV 3.5V 4V
Pass/Fail Pass Fail Pass Fail
Measured kV feedback voltage

If kV,mA waveforms are fail, please attach your kV, mA waveform

8) 80kV/5mAs waveform
Please set kV, mAs below table and take exposure
Exposure condition : 80kV/5mAs
CH 1: kV 2V/div CH 2 : mA 2V/div 40msec/div

Setting value EP_FB IP_FB


Ref voltage 80kV 4V 4V
Pass/Fail Pass Fail Pass Fail
Measured kV feedback voltage

If kV,mA waveforms are fail, please attach your kV, mA waveform

70
EPX-SERISE SERVICE MANUAL Rev. E

9) 90kV/5mAs waveform
Please set kV, mAs below table and take exposure
Exposure condition : 90kV/5mAs
CH 1: kV 2V/div CH 2 : mA 2V/div 40msec/div

Setting value EP_FB IP_FB


Ref voltage 90kV 4.5V 3V
Pass/Fail Pass Fail Pass Fail
Measured kV feedback voltage

If kV,mA waveforms are fail, please attach your kV, mA waveform

10) 100kV/5mAs waveform


Please set kV, mAs below table and take exposure
Exposure condition : 100kV/5mAs
CH 1: kV 2V/div CH 2 : mA 2V/div 40msec/div

Setting value EP_FB IP_FB


Ref voltage 100kV 5V 3V
Pass/Fail Pass Fail Pass Fail
Measured kV feedback voltage

If kV,mA waveforms are fail, please attach your kV,mA waveform

71
EPX-SERISE SERVICE MANUAL Rev. E

6.4.3. Waveform of poor AC power condition


If AC power is under 100V or 200VAC, you have below abnormal waveform. Error 015,016,017,018 is
occurred by poor AC power.

6.4.4 kV, mA PWM frequency check


1) Turn off AC swtich
2) Please connect probe to 200kHz(TP15), 140kHz(TP16) test point on operation board below picture.
3) Turn on AC switch.

72
EPX-SERISE SERVICE MANUAL Rev. E

4) kV / mA PWM frequency

CH 1: kV 2V/div CH 2 : mA 2V/div 4μsec/div

CH CH 1 mA PWM CH 2 kV PWM
Frequency 200kHz 140kHz
Pass/Fail Pass Fail Pass Fail
Measured frequency

6.4.5 mA calibration data check.

DIP Switches are used for following setting.


SW #1: ERROR Enable (ON)/Disable (OFF)
SW #2: Reserved
SW #3: Initialize memories (reserved)
SW #4: Calibration mode (ON)/Normal mode (Off)
For normal operation, only the DIP SW#1 must be set to ON position.
DIP SWITCH #4 on the OP BOARD PCB is set to ON position.
Then the display should indicate as follows,

73
EPX-SERISE SERVICE MANUAL Rev. E

3) Turn on AC switch.

Calibration code

Input data

Up/Down switches increase and decrease C00~C07(AC110V), C08~C15(AC220V)


Please check Input data.

Input data and Tube current should be the same.

Tube Current
Tube Voltage mAs Tube Current INPUT DATA
Parameter
AC110V AC220V KV MIN MAX mA mA

C00 C08 40~50 0.4 100 60.0 60.0


C01 C09 51~60 0.4 80 50.0 50.0
C02 C10 61~70 0.4 100 40.0 40.0
C03 C11 71~80(1) 0.4 32 40.0 40.0
C04 C12 71~80(2) 40 80 35.0 35.0
C05 C13 81~90 0.4 80 30.0 30.0
C06 C14 91~100(1) 0.4 50 30.0 30.0
C07 C15 91~100(2) 64 80 25.0 25.0

4) If mA calibration is done, Please dip switch 4 set to OFF position.

74
EPX-SERISE SERVICE MANUAL Rev. E

6.4.6 Filament Preheat Calibration Code and Data check

1) Turn off AC switch


2) Dip switch on op board

DIP Switches are used for following setting.


SW #1: ERROR Enable (ON)/Disable (OFF)
SW #2: Reserved
SW #3: Initialize memories (reserved)
SW #4: Calibration mode (ON)/Normal mode (Off)
For normal operation, only the DIP SW#1 must be set to ON position.
DIP SWITCH #4 on the OP BOARD PCB is set to ON position.
Then the display should indicate as follows,

3) Turn on AC switch.
4) In mA calibration mode, Press the number 1 switch as shown in figure. Mode is converted.

P_Calibration code

Input data

Up/Down switches increase and decrease P00~P05(AC110V), P06~P11(AC220V)


Please check Input data.
Input data and Tube current should be the same.

75
EPX-SERISE SERVICE MANUAL Rev. E

Fill in the blank below table because P_DATA is different for each product.

Model Current(mA) P_Mode(110V) P_Mode(220V) P_DATA(A)


60 P00 P06
50 P01 P07
40 P02 P08
3.2kW
35 P03 P09
30 P04 P10
25 P05 P11

5) If Preheat calibration is done, Please dip switch 4 set to OFF position.

76
EPX-SERISE SERVICE MANUAL Rev. E

6.5 EPX-F4000

Please fill in the blank.

kV/mAs condition(When
Product S/N Error code
error is occurred)

※ Open upper cover of portable x-ray below picture.

6.5.1 Voltage check


1) Turn on AC switch

TP26 TP24 TP25 TP6


VDC Range 4.8V~5.0V 11.5V~12.5V -11.5V~-12.5V +3V~3.6V
Pass Fail Pass Fail Pass Fail Pass Fail
Pass/Fail

Measured Voltage
If it is fail case

77
EPX-SERISE SERVICE MANUAL Rev. E

6.5.2. kV, mA waveform check


1) Turn off AC switch
2) Please connect probe to EP_FB (kV feedback_TP11), IP_FB (mA feedback_TP8) test point on operation
board below picture.
3)Turn on AC switch.

4) 50kV/5mAs waveform
Please set kV, mAs below table and take exposure
Exposure condition : 50kV/5mAs
CH 1: kV 2V/div CH 2 : mA 2V/div 20msec/div

Setting value EP_FB IP_FB


Ref voltage 50kV 2.5V 5.44V
Pass/Fail Pass Fail Pass Fail
Measured kV feedback voltage

If kV,mA waveforms are fail, please attach your kV, mA waveform

78
EPX-SERISE SERVICE MANUAL Rev. E

6) 60kV/5mAs waveform
Please set kV, mAs below table and take exposure
Exposure condition : 60kV/5mAs
CH 1: kV 2V/div CH 2 : mA 2V/div 20msec/div

Setting value EP_FB IP_FB


Ref voltage 60kV 3V 4.08V
Pass/Fail Pass Fail Pass Fail
Measured kV feedback voltage

If kV,mA waveforms are fail, please attach your kV, mA waveform

7) 70kV/5mAs waveform
Please set kV, mAs below table and take exposure
Exposure condition : 70kV/5mAs
CH 1: kV 2V/div CH 2 : mA 2V/div 20msec/div

Setting value EP_FB IP_FB


Ref voltage 70kV 3.5V 3.4V
Pass/Fail Pass Fail Pass Fail
Measured kV feedback voltage

If kV,mA waveforms are fail, please attach your kV, mA waveform

79
EPX-SERISE SERVICE MANUAL Rev. E

8) 80kV/5mAs waveform
Please set kV, mAs below table and take exposure
Exposure condition : 80kV/5mAs
CH 1: kV 2V/div CH 2 : mA 2V/div 20msec/div

Setting value EP_FB IP_FB


Ref voltage 80kV 4V 3.4V
Pass/Fail Pass Fail Pass Fail
Measured kV feedback voltage

If kV,mA waveforms are fail, please attach your kV, mA waveform

9) 90kV/5mAs waveform
Please set kV, mAs below table and take exposure
Exposure condition : 90kV/5mAs
CH 1: kV 2V/div CH 2 : mA 2V/div 40msec/div

Setting value EP_FB IP_FB


Ref voltage 90kV 4.5V 2.72V
Pass/Fail Pass Fail Pass Fail
Measured kV feedback voltage

If kV,mA waveforms are fail, please attach your kV, mA waveform

80
EPX-SERISE SERVICE MANUAL Rev. E

10) 100kV/5mAs waveform


Please set kV, mAs below table and take exposure
Exposure condition : 100kV/5mAs
CH 1: kV 2V/div CH 2 : mA 2V/div 40msec/div

Setting value EP_FB IP_FB


Ref voltage 100kV 5V 2.38V
Pass/Fail Pass Fail Pass Fail
Measured kV feedback voltage

If kV,mA waveforms are fail, please attach your kV,mA waveform

11) 110kV/5mAs waveform


Please set kV, mAs below table and take exposure
Exposure condition : 110kV/5mAs
CH 1: kV 2V/div CH 2 : mA 2V/div 40msec/div

Setting value EP_FB IP_FB


Ref voltage 110kV 5.5V 2.04V
Pass/Fail Pass Fail Pass Fail
Measured kV feedback voltage

If kV,mA waveforms are fail, please attach your kV,mA waveform

81
EPX-SERISE SERVICE MANUAL Rev. E

6.5.3. Waveform of poor AC power condition


If AC power is under 100V or 200VAC, you have below abnormal waveform. Error 015,016,017,018 is
occurred by poor AC power.

6.5.4 kV, mA PWM frequency check


1) Turn off AC swtich
2) Please connect probe to 200kHz(TP15), 140kHz(TP16) test point on operation board below picture.
3) Turn on AC switch.

82
EPX-SERISE SERVICE MANUAL Rev. E

4) kV / mA PWM frequency

CH 1: kV 2V/div CH 2 : mA 2V/div 4μsec/div

CH CH 1 mA PWM CH 2 kV PWM
Frequency 200kHz 140kHz
Pass/Fail Pass Fail Pass Fail
Measured frequency

6.5.5 mA calibration data check.

DIP Switches are used for following setting.


SW #1: ERROR Enable (ON)/Disable (OFF)
SW #2: Reserved
SW #3: Initialize memories (reserved)
SW #4: Calibration mode (ON)/Normal mode (Off)
For normal operation, only the DIP SW#1 must be set to ON position.
DIP SWITCH #4 on the OP BOARD PCB is set to ON position.
Then the display should indicate as follows,

83
EPX-SERISE SERVICE MANUAL Rev. E

3) Turn on AC switch.

Calibration code

Input data

Up/Down switches increase and decrease C00~C014(AC110V), C15~C29(AC220V)


Please check Input data.

Input data and Tube current should be the same.

Tube Current
Tube Voltage mAs Tube Current INPUT DATA
Parameter
AC 110V AC 220V KV MIN MAX mA mA

C00 C15 50 (1) 0.4 16 80.0 80.0


C01 C16 50 (2) 20 32 70.0 70.0
C02 C17 50 (3) 40 100 60.0 60.0
C03 C18 51~60 (1) 0.4 32 60.0 60.0
C04 C19 51~60 (2) 40 100 50.0 50.0
C05 C20 61~70 (1) 0.4 32 50.0 50.0
C06 C21 61~70 (2) 40 100 40.0 40.0
C07 C22 71~80 (1) 0.4 10 50.0 45.0
C08 C23 71~80 (2) 13 80 40.0 40.0
C09 C24 81~90 (1) 0.4 32 40.0 35.0
C10 C25 81~90 (2) 40 80 30.0 30.0
C11 C26 91~100 (1) 0.4 32 35.0 30.0
C12 C27 91~100 (2) 40 80 25.0 25.0
C13 C28 101~110 (1) 0.4 32 30.0 25.0
C14 C29 101~110 (2) 40 80 20.0 20.0

4) If mA calibration is done, Please dip switch 4 set to OFF position.

84
EPX-SERISE SERVICE MANUAL Rev. E

6.5.6 Filament Preheat Calibration Code and Data check

1) Turn off AC switch


2) Dip switch on op board

DIP Switches are used for following setting.


SW #1: ERROR Enable (ON)/Disable (OFF)
SW #2: Reserved
SW #3: Initialize memories (reserved)
SW #4: Calibration mode (ON)/Normal mode (Off)
For normal operation, only the DIP SW#1 must be set to ON position.
DIP SWITCH #4 on the OP BOARD PCB is set to ON position.
Then the display should indicate as follows,

3) Turn on AC switch.
4) In mA calibration mode, Press the number 1 switch as shown in figure. Mode is converted.

P_Calibration code

Input data

Up/Down switches increase and decrease P00~P09(AC110V), P10~P19(AC220V)


Please check Input data.
Input data and Tube current should be the same.

85
EPX-SERISE SERVICE MANUAL Rev. E

Fill in the blank below table because P_DATA is different for each product.

Model Current(mA) P_Mode(110V) P_Mode(220V) P_DATA


80 P00 P10
70 P01 P11
60 P02 P12
50 P03 P13
45 P04 P14
4.0KW
40 P05 P15
35 P06 P16
30 P07 P17
25 P08 P18
20 P09 P19

5) If Preheat calibration is done, Please dip switch 4 set to OFF position.

86
EPX-SERISE SERVICE MANUAL Rev. E

6.6 EPX-F5000

Please fill in the blank.

kV/mAs condition(When
Product S/N Error code
error is occurred)

※ Open upper cover of portable x-ray below picture.

6.6.1 Voltage check


1) Turn on AC switch

TP26 TP24 TP25 TP6


VDC Range 4.8V~5.0V 11.5V~12.5V -11.5V~-12.5V +3V~3.6V
Pass Fail Pass Fail Pass Fail Pass Fail
Pass/Fail

Measured Voltage
If it is fail case

87
EPX-SERISE SERVICE MANUAL Rev. E

6.6.2. kV, mA waveform check


1) Turn off AC switch
2) Please connect probe to EP_FB (kV feedback_TP11), IP_FB (mA feedback_TP8) test point on operation
board below picture.
3) Turn on AC switch.

4) 40kV/5mAs waveform
Please set kV, mAs below table and take exposure
Exposure condition : 40kV/5mAs
CH 1: kV 2V/div CH 2 : mA 2V/div 20msec/div

Setting value EP_FB IP_FB


Ref voltage 40kV 2V 5.1V
Pass/Fail Pass Fail Pass Fail
Measured kV feedback voltage

If kV,mA waveforms are fail, please attach your kV, mA waveform

88
EPX-SERISE SERVICE MANUAL Rev. E

5) 50kV/5mAs waveform
Please set kV, mAs below table and take exposure
Exposure condition : 50kV/5mAs
CH 1: kV 2V/div CH 2 : mA 2V/div 20msec/div

Setting value EP_FB IP_FB


Ref voltage 50kV 2.5V 5.1V
Pass/Fail Pass Fail Pass Fail
Measured kV feedback voltage

If kV,mA waveforms are fail, please attach your kV, mA waveform

6) 60kV/5mAs waveform
Please set kV, mAs below table and take exposure
Exposure condition : 60kV/5mAs
CH 1: kV 2V/div CH 2 : mA 2V/div 20msec/div

Setting value EP_FB IP_FB


Ref voltage 60kV 3V 4.08V
Pass/Fail Pass Fail Pass Fail
Measured kV feedback voltage

If kV,mA waveforms are fail, please attach your kV, mA waveform

89
EPX-SERISE SERVICE MANUAL Rev. E

7) 70kV/5mAs waveform
Please set kV, mAs below table and take exposure
Exposure condition : 70kV/5mAs
CH 1: kV 2V/div CH 2 : mA 2V/div 20msec/div

Setting value EP_FB IP_FB


Ref voltage 70kV 3.5V 3.57V
Pass/Fail Pass Fail Pass Fail
Measured kV feedback voltage

If kV,mA waveforms are fail, please attach your kV, mA waveform

8) 80kV/5mAs waveform
Please set kV, mAs below table and take exposure
Exposure condition : 80kV/5mAs
CH 1: kV 2V/div CH 2 : mA 2V/div 20msec/div

Setting value EP_FB IP_FB


Ref voltage 80kV 4V 3.06V
Pass/Fail Pass Fail Pass Fail
Measured kV feedback voltage

If kV,mA waveforms are fail, please attach your kV, mA waveform

90
EPX-SERISE SERVICE MANUAL Rev. E

9) 90kV/5mAs waveform
Please set kV, mAs below table and take exposure
Exposure condition : 90kV/5mAs
CH 1: kV 2V/div CH 2 : mA 2V/div 40msec/div

Setting value EP_FB IP_FB


Ref voltage 90kV 4.5V 2.04V
Pass/Fail Pass Fail Pass Fail
Measured kV feedback voltage

If kV,mA waveforms are fail, please attach your kV, mA waveform

10) 100kV/5mAs waveform


Please set kV, mAs below table and take exposure
Exposure condition : 100kV/5mAs
CH 1: kV 2V/div CH 2 : mA 2V/div 40msec/div

Setting value EP_FB IP_FB


Ref voltage 100kV 5V 1.79V
Pass/Fail Pass Fail Pass Fail
Measured kV feedback voltage

If kV,mA waveforms are fail, please attach your kV,mA waveform

91
EPX-SERISE SERVICE MANUAL Rev. E

11) 110kV/5mAs waveform


Please set kV, mAs below table and take exposure
Exposure condition : 110kV/5mAs
CH 1: kV 2V/div CH 2 : mA 2V/div 40msec/div

Setting value EP_FB IP_FB


Ref voltage 110kV 5.5V 1.53V
Pass/Fail Pass Fail Pass Fail
Measured kV feedback voltage

If kV,mA waveforms are fail, please attach your kV,mA waveform

6.6.3. Waveform of poor AC power condition


If AC power is under 100V or 200VAC, you have below abnormal waveform. Error 015,016,017,018 is
occurred by poor AC power.

92
EPX-SERISE SERVICE MANUAL Rev. E

6.6.4 kV, mA PWM frequency check


1) Turn off AC swtich
2) Please connect probe to 200kHz(TP15), 140kHz(TP16) test point on operation board below picture.
3) Turn on AC switch.

4) kV / mA PWM frequency

CH 1: kV 2V/div CH 2 : mA 2V/div 4μsec/div

CH CH 1 mA PWM CH 2 kV PWM
Frequency 200kHz 140kHz
Pass/Fail Pass Fail Pass Fail
Measured frequency

93
EPX-SERISE SERVICE MANUAL Rev. E

6.6.5 mA calibration data check.

DIP Switches are used for following setting.


SW #1: ERROR Enable (ON)/Disable (OFF)
SW #2: Reserved
SW #3: Initialize memories (reserved)
SW #4: Calibration mode (ON)/Normal mode (Off)
For normal operation, only the DIP SW#1 must be set to ON position.
DIP SWITCH #4 on the OP BOARD PCB is set to ON position.
Then the display should indicate as follows,

3) Turn on AC switch.

Calibration code

Input data

Up/Down switches increase and decrease C00~C20(AC110V), C21~C41(AC220V)


Please check Input data.

94
EPX-SERISE SERVICE MANUAL Rev. E

Input data and Tube current should be the same.

Tube Current
Tube Voltage mAs Tube Current INPUT DATA
Parameter
AC110V AC220V KV MIN MAX mA mA
C00 C21 50(1) 0.4 10 100 100
C01 C22 50(2) 13 32 80 80
C02 C23 50(3) 40 100 70 70
C03 C24 51~60(1) 0.4 10 80 80
C04 C25 51~60(2) 13 32 70 70
C05 C26 51~60(3) 40 100 60 60
C06 C27 61~70(1) 0.4 10 70 70
C07 C28 61~70(2) 13 32 60 60
C08 C29 61~70(3) 40 100 50 50
C09 C30 71~80(1) 0.4 10 60 60
C10 C31 71~80(2) 13 32 50 50
C11 C32 71~80(3) 40 100 40 40
C12 C33 81~90(1) 0.4 10 50 50
C13 C34 81~90(2) 13 32 40 40
C14 C35 81~90(3) 40 100 30 30
C15 C36 91~100(1) 0.4 10 45 45
C16 C37 91~100(2) 13 32 40 40
C17 C38 91~100(3) 40 100 30 30
C18 C39 101~110(1) 0.4 10 40 40
C19 C40 101~110 (2) 13 32 30 30
C20 C41 101~110 (3) 40 100 20 20

4) If mA calibration is done, Please dip switch 4 set to OFF position.

95
EPX-SERISE SERVICE MANUAL Rev. E

6.1.6 Filament Preheat Calibration Code and Data check

1) Turn off AC switch


2) Dip switch on op board

DIP Switches are used for following setting.


SW #1: ERROR Enable (ON)/Disable (OFF)
SW #2: Reserved
SW #3: Initialize memories (reserved)
SW #4: Calibration mode (ON)/Normal mode (Off)
For normal operation, only the DIP SW#1 must be set to ON position.
DIP SWITCH #4 on the OP BOARD PCB is set to ON position.
Then the display should indicate as follows,

3) Turn on AC switch.
4) In mA calibration mode, Press the number 1 switch as shown in figure. Mode is converted.

P_Calibration code

Input data

Up/Down switches increase and decrease P00~P08(AC110V), P09~P17(AC220V)


Please check Input data.
Input data and Tube current should be the same.

96
EPX-SERISE SERVICE MANUAL Rev. E

Fill in the blank below table because P_DATA is different for each product.

Model Current(mA) P_Mode(110V) P_Mode(220V) P_DATA(A)

100 P00 P09

80 P01 P10

70 P02 P11

60 P03 P12

5kW 50 P04 P13

45 P05 P14

40 P06 P15

30 P07 P16

25 P08 P17

5) If Preheat calibration is done, Please dip switch 4 set to OFF position.

97