EPX-Series
Diagnostic X-ray Systems
Service Manual
1
EPX-SERISE SERVICE MANUAL Rev. E
2
EPX-SERISE SERVICE MANUAL Rev. E
SERVICE CHAPTER
1. CALIBRATION------------------------------------------------------------------------------------------- 7
4. COLLIMATOR-------------------------------------------------------------------------------------------33
5. TROUBLE SHOOTING-------------------------------------------------------------------------------35
REVISON HISTORY
3
EPX-SERISE SERVICE MANUAL Rev. E
4
EPX-SERISE SERVICE MANUAL Rev. E
5
EPX-SERISE SERVICE MANUAL Rev. E
Service chapter
This manual is designed to ensure correct use and operation of EPX model diagnostic x-ray units.
Please read all the lines thoroughly before you use the equipment.
Incorrect use and operation exceeding described conditions in this manual may cause damage of the
equipment and shorten its life. Particular attention must be paid to all the warnings, cautions and notes
incorporated herein.
This equipment should be used only by the legally qualified persons and practitioners.
Regarding the AC cable and software, only the ones developed and supplied by Ecotron Inc. should be
used.
Only trained maintenance staff may remove the covers of the x-ray system.
Do not modify the equipment at your discretion and in case any modification is required unavoidably, ask
the help of Ecotron or its authorized dealer for the service.
Information for repair will be provided by Ecotron. (Circuit diagram, Component part list, descriptions,
Calibration instructions, or other information)
No responsibility is taken by Ecotron for any infringement of patients or other rights of the third parties
which many results from the use of this manual.
NOTE: If the output calibration is needed after periodical inspections or repairs, please observe following
rules and methods described in this manual.
CAUTION
The ionizing radiation is dangerous for the operator if the following safety measures
are not strictly observed.
NOTE
6
EPX-SERISE SERVICE MANUAL Rev. E
1. Calibration
1.1 EPX-F1600
1.1.1 Line Power Switch
7
EPX-SERISE SERVICE MANUAL Rev. E
Where, C00~C09 are calibration code of tube current at AC110V model and C10~C19 are calibration
code of tube current at AC220V model.
C00 or C10 is calibration code of tube current (30mA at 40kV~50kV) at AC110V or AC220V and 30.0
is calibration data of tube current (30mA at 40kV~50kV).
8
EPX-SERISE SERVICE MANUAL Rev. E
b. After it is adjusted, push the STORE button during 3sec(see operation sheet).
c. DIP SWITCH #4 on the PCB OP BOARD should be set to OFF position and the expose X-ray.
e. If it is not 6V, set the calibration data again.
f. DIP SWITCH #4 on the OP BOARD PCB is set to ON position.
g. Then push the kV UP switch (see operation sheet).
h. The display should indicate as follows,
Where, C01 or C11 is calibration code of tube current (25mA at 40kV~50kV) and 25.0 is calibration
data of tube current (25mA at 40kV~50kV).
9
EPX-SERISE SERVICE MANUAL Rev. E
Where, P00~P07 are calibration code of filament preheat current at AC110V, P08~P15 are calibration
code of filament preheat current at AC220V, P00 or P08 is Filament preheat calibration code (Preheat
calibration code of 30mA), and 1.50 or 1.40 is Filament preheat calibration data (Preheat calibration
data of 30mA).
g. Adjust data by using the mAs Up-Down switch (see operation sheet).
h. After it is adjusted, push the STORE button during 3sec(see operation sheet).
i. DIP SWITCH #4 on the PCB OP BOARD is set to OFF position then expose X-ray.
j. If overshooting or undershooting occurs in the waveform, set the calibration data again.
k. Then push the KV UP switch (see operation sheet).
l. The display should indicate as follows,
Where, P01 or P09 is Filament preheat calibration code (Preheat calibration code of 25mA) and
1.45 or 1.35 is Filament preheat calibration data (Preheat calibration data of 25mA).
m. Adjust data by using the mAs Up-Down switch on the Control Panel.
n. After it is adjusted, push the Store switch during 3sec(see operation sheet).
o. DIP SWITCH #4 on the PCB OP BOARD is set to OFF position then exposure X-ray.
p. If overshooting or undershooting occurs in the waveform, set the calibration data again.
10
EPX-SERISE SERVICE MANUAL Rev. E
1.2 EPX-F2400
1.2.1 Line Power Switch
11
EPX-SERISE SERVICE MANUAL Rev. E
Where, C00~C07 are calibration code of tube current at AC110V model and C08~C15 are calibration
code of tube current at AC220V model.
C00 or C08 is calibration code of tube current (40mA at 40kV~60kV) at AC110V or AC220V and 40.0
is calibration data of tube current (40mA at 40kV~60kV).
12
EPX-SERISE SERVICE MANUAL Rev. E
Where, C01 or C09 is calibration code of tube current (35mA at 40kV~60kV) and 35.0 is calibration
data of tube current (35mA at 40kV~60kV).
13
EPX-SERISE SERVICE MANUAL Rev. E
Where, P00~P04 are calibration code of filament preheat current at AC110V, P05~P09 are calibration
code of filament preheat current at AC220V, P00 or P05 is Filament preheat calibration code (Preheat
calibration code of 40mA), and 0.70 or 0.72 is Filament preheat calibration data (Preheat calibration
data of 40mA).
g. Adjust data by using the mAs Up-Down switch (see operation sheet).
h. After it is adjusted, push the STORE button during 3sec(see operation sheet).
i. DIP SWITCH #4 on the PCB OP BOARD is set to OFF position then expose X-ray.
j. If overshooting or undershooting occurs in the waveform, set the calibration data again.
k. Then push the KV UP switch (see operation sheet).
l. The display should indicate as follows,
Where, P01 or P06 is Filament preheat calibration code (Preheat calibration code of 35mA) and
0.67 or 0.69 is Filament preheat calibration data (Preheat calibration data of 35mA).
m. Adjust data by using the mAs Up-Down switch on the Control Panel.
n. After it is adjusted, push the Store switch during 3sec(see operation sheet).
o. DIP SWITCH #4 on the PCB OP BOARD is set to OFF position then exposure X-ray.
p. If overshooting or undershooting occurs in the waveform, set the calibration data again.
14
EPX-SERISE SERVICE MANUAL Rev. E
1.3 EPX-F2800
1.3.1 Line Power Switch
15
EPX-SERISE SERVICE MANUAL Rev. E
Where, C00~C15 are calibration code of tube current at AC110V model and C16~C31 are calibration
code of tube current at AC220V model.
C00 or C16 is calibration code of tube current (40mA at 40kV~60kV) at AC110V or AC220V and 40.0
is calibration data of tube current (40mA at 40kV~60kV).
16
EPX-SERISE SERVICE MANUAL Rev. E
Where, C01 or C17 is calibration code of tube current (30mA at 61kV~70kV) and 30.0 is calibration
data of tube current (30mA at 61kV~70kV).
17
EPX-SERISE SERVICE MANUAL Rev. E
where, P00~P06 are calibration code of filament preheat current at AC110V, P07~P13 are calibration
code of filament preheat current at AC220V, P00 or P07 is Filament preheat calibration code (Preheat
calibration code of 35mA), and 0.70 or 0.70 is Filament preheat calibration data (Preheat calibration
data of 35mA).
g. Adjust data by using the mAs Up-Down switch (see operation sheet).
h. After it is adjusted, push the STORE button during 3sec(see operation sheet).
i. DIP SWITCH #4 on the PCB OP BOARD is set to OFF position then expose X-ray.
j. If overshooting or undershooting occurs in the waveform, set the calibration data again.
k. Then push the KV UP switch (see operation sheet).
l. The display should indicate as follows,
Where, P01 or P08 is Filament preheat calibration code (Preheat calibration code of 30mA) and 0.67
or 0.67 is Filament preheat calibration data (Preheat calibration data of 30mA).
m. Adjust data by using the mAs Up-Down switch on the Control Panel.
n. After it is adjusted, push the Store switch during 3sec(see operation sheet).
o. DIP SWITCH #4 on the PCB OP BOARD is set to OFF position then exposure X-ray.
p. If overshooting or undershooting occurs in the waveform, set the calibration data again.
18
EPX-SERISE SERVICE MANUAL Rev. E
1.4 EPX-F3200
1.4.1 Line Power Switch
19
EPX-SERISE SERVICE MANUAL Rev. E
Where, C00~C07 are calibration code of tube current at AC110V model and C08~C15 are calibration
code of tube current at AC220V model. C00 or C08 is calibration code of tube current (60mA at
40kV~50kV) at AC110V and 60.0 is calibration data of tube current (60mA at 40kV~50kV).
20
EPX-SERISE SERVICE MANUAL Rev. E
c. DIP SWITCH #4 on the PCB OP BOARD should be set to OFF position and the expose X-ray.
e. If it is not 6.0V, set the calibration data again.
f. DIP SWITCH #4 on the OP BOARD PCB is set to ON position.
g. Then push the kV UP switch (see operation sheet).
h. The display should indicate as follows,
Where, C01 or C09 is calibration code of tube current (50mA at 51kV~60kV) and 50.0 is calibration
data of tube current (50mA at 51kV~60kV).
21
EPX-SERISE SERVICE MANUAL Rev. E
Where, P00~P05 are calibration code of filament preheat current at AC110V, P06~P11 are calibration
code of filament preheat current at AC220V, P00 or P06 is Filament preheat calibration code (Preheat
calibration code of 60mA), and 1.65 or 1.65 is Filament preheat calibration data (Preheat calibration
data of 60mA).
g. Adjust data by using the mAs Up-Down switch (see operation sheet).
h. After it is adjusted, push the STORE button during 3sec(see operation sheet).
i. DIP SWITCH #4 on the PCB OP BOARD is set to OFF position then expose X-ray.
j. If overshooting or undershooting occurs in the waveform, set the calibration data again.
k. Then push the KV UP switch (see operation sheet).
l. The display should indicate as follows,
where, P01 or P07 is Filament preheat calibration code (Preheat calibration code of 50mA) and 1.60
or 1.60 is Filament preheat calibration data (Preheat calibration data of 50mA).
m. Adjust data by using the mAs Up-Down switch on the Control Panel.
n. After it is adjusted, push the Store switch during 3sec(see operation sheet).
o. DIP SWITCH #4 on the PCB OP BOARD is set to OFF position then exposure X-ray.
p. If overshooting or undershooting occurs in the waveform, set the calibration data again.
22
EPX-SERISE SERVICE MANUAL Rev. E
1.5 EPX-F4000
1.5.1 Line Power Switch
23
EPX-SERISE SERVICE MANUAL Rev. E
Where, C00~C14 are calibration code of tube current at AC110V model and C15~C29 are calibration
code of tube current at AC220V model. C00 or C15 is calibration code of tube current (80mA at 50kV)
at 80.0 is calibration data of tube current (80mA at 50kV).
24
EPX-SERISE SERVICE MANUAL Rev. E
c. DIP SWITCH #4 on the PCB OP BOARD should be set to OFF position and the expose X-ray.
e. If it is not 4V, set the calibration data again.
f. DIP SWITCH #4 on the OP BOARD PCB is set to ON position.
g. Then push the kV UP switch (see operation sheet).
h. The display should indicate as follows,
Where, C01 or C16 is calibration code of tube current (70mA at 50kV) and 70.0 is calibration data of
tube current (70mA at 50kV).
25
EPX-SERISE SERVICE MANUAL Rev. E
Where, P00~P09 are calibration code of filament preheat current at AC110V, P10~P19 are calibration
code of filament preheat current at AC220V, P00 or P10 is Filament preheat calibration code (Preheat
calibration code of 80mA), and 2.50 or 2.50 is Filament preheat calibration data (Preheat calibration
data of 80mA).
g. Adjust data by using the mAs Up-Down switch (see operation sheet).
h. After it is adjusted, push the STORE button during 3sec(see operation sheet).
i. DIP SWITCH #4 on the PCB OP BOARD is set to OFF position then expose X-ray.
j. If overshooting or undershooting occurs in the waveform, set the calibration data again.
k. Then push the KV UP switch (see operation sheet).
l. The display should indicate as follows,
Where, P01 or P11 is Filament preheat calibration code (Preheat calibration code of 70mA) and 2.20
or 2.20 is Filament preheat calibration data (Preheat calibration data of 70mA).
m. Adjust data by using the mAs Up-Down switch on the Control Panel.
n. After it is adjusted, push the Store switch during 3sec(see operation sheet).
o. DIP SWITCH #4 on the PCB OP BOARD is set to OFF position then exposure X-ray.
p. If overshooting or undershooting occurs in the waveform, set the calibration data again.
26
EPX-SERISE SERVICE MANUAL Rev. E
1.6 EPX-F5000
1.6.1 Line Power Switch
27
EPX-SERISE SERVICE MANUAL Rev. E
Where, C00~C20 are calibration code of tube current at AC110V model and C21~C41 are calibration
code of tube current at AC220V model. C00 or C21 is calibration code of tube current (100mA at
50kV) at A0.0(=100) is calibration data of tube current (100mA at 50kV).
28
EPX-SERISE SERVICE MANUAL Rev. E
b. After it is adjusted, push the STORE button during 3sec(see operation sheet).
c. DIP SWITCH #4 on the PCB OP BOARD should be set to OFF position and the expose X-ray.
e. If it is not 5.1V, set the calibration data again.
f. DIP SWITCH #4 on the OP BOARD PCB is set to ON position.
g. Then push the kV UP switch (see operation sheet).
h. The display should indicate as follows,
Where, C01 or C22 is calibration code of tube current (80mA at 50kV) and 80.0 is calibration data of
tube current (80mA at 50kV).
29
EPX-SERISE SERVICE MANUAL Rev. E
Where, P00~P08 are calibration code of filament preheat current at AC110V, P09~P15 are calibration
code of filament preheat current at AC220V, P00 or P09 is Filament preheat calibration code (Preheat
calibration code of 100mA), and 2.40 or 2.40 is Filament preheat calibration data (Preheat calibration
data of 100mA).
g. Adjust data by using the mAs Up-Down switch (see operation sheet).
h. After it is adjusted, push the STORE button during 3sec(see operation sheet).
i. DIP SWITCH #4 on the PCB OP BOARD is set to OFF position then expose X-ray.
j. If overshooting or undershooting occurs in the waveform, set the calibration data again.
k. Then push the KV UP switch (see operation sheet).
l. The display should indicate as follows,
Where, P01 or P10 is Filament preheat calibration code (Preheat calibration code of 80mA) and 2.25
or 2.25 is Filament preheat calibration data (Preheat calibration data of 80mA).
m. Adjust data by using the mAs Up-Down switch on the Control Panel.
n. After it is adjusted, push the Store switch during 3sec(see operation sheet).
o. DIP SWITCH #4 on the PCB OP BOARD is set to OFF position then exposure X-ray.
p. If overshooting or undershooting occurs in the waveform, set the calibration data again.
30
EPX-SERISE SERVICE MANUAL Rev. E
For normal operation, only the DIP SW#1 must be set to ON position.
31
EPX-SERISE SERVICE MANUAL Rev. E
SWITCH 1
SWITCH 2
a. Press and hold the Switch #1 for longer than 1 second, which will get the “Prep or
Ready” status on, then press Switch #2 to get the “Exposure”.
b. Press and hold the Switch #1 and Switch #2 simultaneously for longer than 1 second.
This action will result in “Prep or Ready” status and “Exposure” at the same time.
32
EPX-SERISE SERVICE MANUAL Rev. E
4. Collimator
a. Halogen Lamp * 1
b. Laser pointer * 2
c. Horizontal and vertical lighting range control
i. Block diagram
ii. Layout
a. External Layout
33
EPX-SERISE SERVICE MANUAL Rev. E
b. Internal Layout
c. Knob Layout
34
EPX-SERISE SERVICE MANUAL Rev. E
5. Trouble shooting
If you have any error message when you are using portable x-ray, please check below service
check list
Error
DESCRIPTION INDICATION
Code
2 THERMAL ERROR (INTERLOCK) Excess from 60℃
10 STANDBY FIL FB ERROR Excess from 200mA (10% of Max. Current, 2A)
35
EPX-SERISE SERVICE MANUAL Rev. E
7 Ready SW ERR
a. Replace the hand switch
8 Exposure SW ERR
15 Run EP FB low ERR a. If error repeats at the condition of 40kV 0.4mAs EXP, check the
followings by sequence.
16 Run IP FB High ERR 1. Check TP in the power input of the OP board
2. Check the Reference voltage value
17 Run EP FB High ERR 3. If the reference voltage is normal, replace the FET board
4. If replacing FET board does not solve the problem, the replace the
18 Run IP FB low ERR HV tank.
30 KV_OVP ERR a. If error pops up right after the unit power is turned on, then replace
the OP board.
b. If error repeats at the condition of 40kV 0.4mAs EXP,
31 FIL_OC ERR FIL
1. Replace the FET board
2. If replacing FET board does not solve the problem, the replace the
32 IP_OCP ERR HV tank.
Warning Codes
Wait until the temperature inside the monotank comes down to normal
Thermal ERR excess
2 (Once the temperature drop to normal, the Warning Message will
from 60°C
disappear)
Wait until the temperature inside the FET board comes down to normal
FET-OTP FET board
29 (Once the temperature drop to normal, the Warning Message will
overheat
disappear)
36
EPX-SERISE SERVICE MANUAL Rev. E
kV/mAs condition(When
Product S/N Error code
error is occurred)
Measured Voltage
If it is fail case
6.1.2. kV, mA waveform check
37
EPX-SERISE SERVICE MANUAL Rev. E
4) 40kV/5mAs waveform
Please set kV, mAs below table and take exposure
Exposure condition : 40kV/5mAs
CH 1: kV 2V/div CH 2 : mA 2V/div 40msec/div
5) 50kV/5mAs waveform
38
EPX-SERISE SERVICE MANUAL Rev. E
6) 60kV/5mAs waveform
Please set kV, mAs below table and take exposure
Exposure condition : 60kV/5mAs
CH 1: kV 2V/div CH 2 : mA 2V/div 40msec/div
39
EPX-SERISE SERVICE MANUAL Rev. E
8) 80kV/5mAs waveform
Please set kV, mAs below table and take exposure
Exposure condition : 80kV/5mAs
CH 1: kV 2V/div CH 2 : mA 2V/div 100msec/div
9) 90kV/5mAs waveform
40
EPX-SERISE SERVICE MANUAL Rev. E
41
EPX-SERISE SERVICE MANUAL Rev. E
4) kV / mA PWM frequency
CH CH 1 mA PWM CH 2 kV PWM
Frequency 200kHz 140kHz
Pass/Fail Pass Fail Pass Fail
Measured frequency
42
EPX-SERISE SERVICE MANUAL Rev. E
3) Turn on AC switch.
Calibration code
Input data
43
EPX-SERISE SERVICE MANUAL Rev. E
Tube Current
Tube Voltage mAs Tube Current INPUT DATA
Parameter
AC110V AC220V KV MIN MAX mA mA
44
EPX-SERISE SERVICE MANUAL Rev. E
3) Turn on AC switch.
4) In mA calibration mode, Press the number 1 switch as shown in figure. Mode is converted.
P_Calibration code
Input data
Fill in the blank below table because P_DATA is different for each product.
45
EPX-SERISE SERVICE MANUAL Rev. E
6.2 EPX-F2400
kV/mAs condition(When
Product S/N Error code
error is occurred)
Measured Voltage
If it is fail case
46
EPX-SERISE SERVICE MANUAL Rev. E
4) 40kV/5mAs waveform
Please set kV, mAs below table and take exposure
Exposure condition : 40kV/5mAs
CH 1: kV 2V/div CH 2 : mA 2V/div 40msec/div
47
EPX-SERISE SERVICE MANUAL Rev. E
5) 50kV/5mAs waveform
Please set kV, mAs below table and take exposure
Exposure condition : 50kV/5mAs
CH 1: kV 2V/div CH 2 : mA 2V/div 40msec/div
6) 60kV/5mAs waveform
Please set kV, mAs below table and take exposure
Exposure condition : 60kV/5mAs
CH 1: kV 2V/div CH 2 : mA 2V/div 40msec/div
48
EPX-SERISE SERVICE MANUAL Rev. E
7) 70kV/5mAs waveform
Please set kV, mAs below table and take exposure
Exposure condition : 70kV/5mAs
CH 1: kV 2V/div CH 2 : mA 2V/div 40msec/div
8) 80kV/5mAs waveform
Please set kV, mAs below table and take exposure
Exposure condition : 80kV/5mAs
CH 1: kV 2V/div CH 2 : mA 2V/div 40msec/div
49
EPX-SERISE SERVICE MANUAL Rev. E
9) 90kV/5mAs waveform
Please set kV, mAs below table and take exposure
Exposure condition : 90kV/5mAs
CH 1: kV 2V/div CH 2 : mA 2V/div 40msec/div
50
EPX-SERISE SERVICE MANUAL Rev. E
51
EPX-SERISE SERVICE MANUAL Rev. E
4) kV / mA PWM frequency
CH CH 1 mA PWM CH 2 kV PWM
Frequency 200kHz 140kHz
Pass/Fail Pass Fail Pass Fail
Measured frequency
52
EPX-SERISE SERVICE MANUAL Rev. E
3) Turn on AC switch.
Calibration code
Input data
Tube Current
Tube Voltage mAs Tube Current INPUT DATA
Parameter
AC110V AC220V KV MIN MAX mA mA
53
EPX-SERISE SERVICE MANUAL Rev. E
3) Turn on AC switch.
4) In mA calibration mode, Press the number 1 switch as shown in figure. Mode is converted.
P_Calibration code
Input data
54
EPX-SERISE SERVICE MANUAL Rev. E
Fill in the blank below table because P_DATA is different for each product.
55
EPX-SERISE SERVICE MANUAL Rev. E
6.3 EPX-F2800
kV/mAs condition(When
Product S/N Error code
error is occurred)
Measured Voltage
If it is fail case
56
EPX-SERISE SERVICE MANUAL Rev. E
4) 40kV/5mAs waveform
Please set kV, mAs below table and take exposure
Exposure condition : 40kV/5mAs
CH 1: kV 2V/div CH 2 : mA 2V/div 40msec/div
57
EPX-SERISE SERVICE MANUAL Rev. E
5) 50kV/5mAs waveform
Please set kV, mAs below table and take exposure
Exposure condition : 50kV/5mAs
CH 1: kV 2V/div CH 2 : mA 2V/div 40msec/div
6) 60kV/5mAs waveform
Please set kV, mAs below table and take exposure
Exposure condition : 60kV/5mAs
CH 1: kV 2V/div CH 2 : mA 2V/div 40msec/div
58
EPX-SERISE SERVICE MANUAL Rev. E
7) 70kV/5mAs waveform
Please set kV, mAs below table and take exposure
Exposure condition : 70kV/5mAs
CH 1: kV 2V/div CH 2 : mA 2V/div 40msec/div
8) 80kV/5mAs waveform
Please set kV, mAs below table and take exposure
Exposure condition : 80kV/5mAs
CH 1: kV 2V/div CH 2 : mA 2V/div 40msec/div
59
EPX-SERISE SERVICE MANUAL Rev. E
9) 90kV/5mAs waveform
Please set kV, mAs below table and take exposure
Exposure condition : 90kV/5mAs
CH 1: kV 2V/div CH 2 : mA 2V/div 40msec/div
60
EPX-SERISE SERVICE MANUAL Rev. E
61
EPX-SERISE SERVICE MANUAL Rev. E
62
EPX-SERISE SERVICE MANUAL Rev. E
4) kV / mA PWM frequency
CH CH 1 mA PWM CH 2 kV PWM
Frequency 200kHz 140kHz
Pass/Fail Pass Fail Pass Fail
Measured frequency
63
EPX-SERISE SERVICE MANUAL Rev. E
3) Turn on AC switch.
Calibration code
Input data
Tube Current
Tube Voltage mAs Tube Current INPUT DATA
Parameter
AC110V AC220V KV MIN MAX mA mA
C00 C16 40~60 0.4 100 40.0 40.0
C01 C17 61~70(1) 0.4 64 30.0 30.0
C02 C18 61~70(2) 80 100 25.0 25.0
C03 C19 71~80(1) 0.4 10 35.0 35.0
C04 C20 71~80(2) 13 32 30.0 30.0
C05 C21 71~80(3) 40 80 25.0 25.0
C06 C22 81~90(1) 0.4 10 30.0 30.0
C07 C23 81~90(2) 13 40 25.0 25.0
C08 C24 81~90(3) 50 80 20.0 20.0
C09 C25 91~100(1) 0.4 13 25.0 25.0
C10 C26 91~100(2) 16 50 20.0 20.0
C11 C27 91~100(3) 64 80 16.0 16.0
C12 C28 101~110(1) 0.4 32 18.0 18.0
C13 C29 101~110(2) 40 64 16.0 16.0
C14 C30 111~120(1) 0.4 20 16.0 16.0
C15 C31 111~120(2) 25 50 14.0 14.0
64
EPX-SERISE SERVICE MANUAL Rev. E
3) Turn on AC switch.
4) In mA calibration mode, Press the number 1 switch as shown in figure. Mode is converted.
P_Calibration code
Input data
65
EPX-SERISE SERVICE MANUAL Rev. E
Fill in the blank below table because P_DATA is different for each product.
66
EPX-SERISE SERVICE MANUAL Rev. E
6.4 EPX-F3200
kV/mAs condition(When
Product S/N Error code
error is occurred)
Measured Voltage
If it is fail case
67
EPX-SERISE SERVICE MANUAL Rev. E
4) 40kV/5mAs waveform
Please set kV, mAs below table and take exposure
Exposure condition : 40kV/5mAs
CH 1: kV 2V/div CH 2 : mA 2V/div 20msec/div
68
EPX-SERISE SERVICE MANUAL Rev. E
5) 50kV/5mAs waveform
Please set kV, mAs below table and take exposure
Exposure condition : 50kV/5mAs
CH 1: kV 2V/div CH 2 : mA 2V/div 20msec/div
6) 60kV/5mAs waveform
Please set kV, mAs below table and take exposure
Exposure condition : 60kV/5mAs
CH 1: kV 2V/div CH 2 : mA 2V/div 20msec/div
69
EPX-SERISE SERVICE MANUAL Rev. E
7) 70kV/5mAs waveform
Please set kV, mAs below table and take exposure
Exposure condition : 70kV/5mAs
CH 1: kV 2V/div CH 2 : mA 2V/div 40msec/div
8) 80kV/5mAs waveform
Please set kV, mAs below table and take exposure
Exposure condition : 80kV/5mAs
CH 1: kV 2V/div CH 2 : mA 2V/div 40msec/div
70
EPX-SERISE SERVICE MANUAL Rev. E
9) 90kV/5mAs waveform
Please set kV, mAs below table and take exposure
Exposure condition : 90kV/5mAs
CH 1: kV 2V/div CH 2 : mA 2V/div 40msec/div
71
EPX-SERISE SERVICE MANUAL Rev. E
72
EPX-SERISE SERVICE MANUAL Rev. E
4) kV / mA PWM frequency
CH CH 1 mA PWM CH 2 kV PWM
Frequency 200kHz 140kHz
Pass/Fail Pass Fail Pass Fail
Measured frequency
73
EPX-SERISE SERVICE MANUAL Rev. E
3) Turn on AC switch.
Calibration code
Input data
Tube Current
Tube Voltage mAs Tube Current INPUT DATA
Parameter
AC110V AC220V KV MIN MAX mA mA
74
EPX-SERISE SERVICE MANUAL Rev. E
3) Turn on AC switch.
4) In mA calibration mode, Press the number 1 switch as shown in figure. Mode is converted.
P_Calibration code
Input data
75
EPX-SERISE SERVICE MANUAL Rev. E
Fill in the blank below table because P_DATA is different for each product.
76
EPX-SERISE SERVICE MANUAL Rev. E
6.5 EPX-F4000
kV/mAs condition(When
Product S/N Error code
error is occurred)
Measured Voltage
If it is fail case
77
EPX-SERISE SERVICE MANUAL Rev. E
4) 50kV/5mAs waveform
Please set kV, mAs below table and take exposure
Exposure condition : 50kV/5mAs
CH 1: kV 2V/div CH 2 : mA 2V/div 20msec/div
78
EPX-SERISE SERVICE MANUAL Rev. E
6) 60kV/5mAs waveform
Please set kV, mAs below table and take exposure
Exposure condition : 60kV/5mAs
CH 1: kV 2V/div CH 2 : mA 2V/div 20msec/div
7) 70kV/5mAs waveform
Please set kV, mAs below table and take exposure
Exposure condition : 70kV/5mAs
CH 1: kV 2V/div CH 2 : mA 2V/div 20msec/div
79
EPX-SERISE SERVICE MANUAL Rev. E
8) 80kV/5mAs waveform
Please set kV, mAs below table and take exposure
Exposure condition : 80kV/5mAs
CH 1: kV 2V/div CH 2 : mA 2V/div 20msec/div
9) 90kV/5mAs waveform
Please set kV, mAs below table and take exposure
Exposure condition : 90kV/5mAs
CH 1: kV 2V/div CH 2 : mA 2V/div 40msec/div
80
EPX-SERISE SERVICE MANUAL Rev. E
81
EPX-SERISE SERVICE MANUAL Rev. E
82
EPX-SERISE SERVICE MANUAL Rev. E
4) kV / mA PWM frequency
CH CH 1 mA PWM CH 2 kV PWM
Frequency 200kHz 140kHz
Pass/Fail Pass Fail Pass Fail
Measured frequency
83
EPX-SERISE SERVICE MANUAL Rev. E
3) Turn on AC switch.
Calibration code
Input data
Tube Current
Tube Voltage mAs Tube Current INPUT DATA
Parameter
AC 110V AC 220V KV MIN MAX mA mA
84
EPX-SERISE SERVICE MANUAL Rev. E
3) Turn on AC switch.
4) In mA calibration mode, Press the number 1 switch as shown in figure. Mode is converted.
P_Calibration code
Input data
85
EPX-SERISE SERVICE MANUAL Rev. E
Fill in the blank below table because P_DATA is different for each product.
86
EPX-SERISE SERVICE MANUAL Rev. E
6.6 EPX-F5000
kV/mAs condition(When
Product S/N Error code
error is occurred)
Measured Voltage
If it is fail case
87
EPX-SERISE SERVICE MANUAL Rev. E
4) 40kV/5mAs waveform
Please set kV, mAs below table and take exposure
Exposure condition : 40kV/5mAs
CH 1: kV 2V/div CH 2 : mA 2V/div 20msec/div
88
EPX-SERISE SERVICE MANUAL Rev. E
5) 50kV/5mAs waveform
Please set kV, mAs below table and take exposure
Exposure condition : 50kV/5mAs
CH 1: kV 2V/div CH 2 : mA 2V/div 20msec/div
6) 60kV/5mAs waveform
Please set kV, mAs below table and take exposure
Exposure condition : 60kV/5mAs
CH 1: kV 2V/div CH 2 : mA 2V/div 20msec/div
89
EPX-SERISE SERVICE MANUAL Rev. E
7) 70kV/5mAs waveform
Please set kV, mAs below table and take exposure
Exposure condition : 70kV/5mAs
CH 1: kV 2V/div CH 2 : mA 2V/div 20msec/div
8) 80kV/5mAs waveform
Please set kV, mAs below table and take exposure
Exposure condition : 80kV/5mAs
CH 1: kV 2V/div CH 2 : mA 2V/div 20msec/div
90
EPX-SERISE SERVICE MANUAL Rev. E
9) 90kV/5mAs waveform
Please set kV, mAs below table and take exposure
Exposure condition : 90kV/5mAs
CH 1: kV 2V/div CH 2 : mA 2V/div 40msec/div
91
EPX-SERISE SERVICE MANUAL Rev. E
92
EPX-SERISE SERVICE MANUAL Rev. E
4) kV / mA PWM frequency
CH CH 1 mA PWM CH 2 kV PWM
Frequency 200kHz 140kHz
Pass/Fail Pass Fail Pass Fail
Measured frequency
93
EPX-SERISE SERVICE MANUAL Rev. E
3) Turn on AC switch.
Calibration code
Input data
94
EPX-SERISE SERVICE MANUAL Rev. E
Tube Current
Tube Voltage mAs Tube Current INPUT DATA
Parameter
AC110V AC220V KV MIN MAX mA mA
C00 C21 50(1) 0.4 10 100 100
C01 C22 50(2) 13 32 80 80
C02 C23 50(3) 40 100 70 70
C03 C24 51~60(1) 0.4 10 80 80
C04 C25 51~60(2) 13 32 70 70
C05 C26 51~60(3) 40 100 60 60
C06 C27 61~70(1) 0.4 10 70 70
C07 C28 61~70(2) 13 32 60 60
C08 C29 61~70(3) 40 100 50 50
C09 C30 71~80(1) 0.4 10 60 60
C10 C31 71~80(2) 13 32 50 50
C11 C32 71~80(3) 40 100 40 40
C12 C33 81~90(1) 0.4 10 50 50
C13 C34 81~90(2) 13 32 40 40
C14 C35 81~90(3) 40 100 30 30
C15 C36 91~100(1) 0.4 10 45 45
C16 C37 91~100(2) 13 32 40 40
C17 C38 91~100(3) 40 100 30 30
C18 C39 101~110(1) 0.4 10 40 40
C19 C40 101~110 (2) 13 32 30 30
C20 C41 101~110 (3) 40 100 20 20
95
EPX-SERISE SERVICE MANUAL Rev. E
3) Turn on AC switch.
4) In mA calibration mode, Press the number 1 switch as shown in figure. Mode is converted.
P_Calibration code
Input data
96
EPX-SERISE SERVICE MANUAL Rev. E
Fill in the blank below table because P_DATA is different for each product.
80 P01 P10
70 P02 P11
60 P03 P12
45 P05 P14
40 P06 P15
30 P07 P16
25 P08 P17
97