Chapter 12: Appendix

AquaBplus/AquaBplus B2/
Settings Report
AquaBplus HF

Technician's name: Customer/customer no.:

Device type with software version: Serial no.: Inventory no.:

Service report number: Operating hours: Equipment code:

Options installed:
AquaBplus B2 
AquaBplus HF 
AquaBplus UF 
RingBase 
RingUnit / wall-mounted inlet unit  Number of ring mains:
Housing 
Emergency mode 
Set units: a) bar, L/h, °C, m, °dH  a)
b) psi, gal/h, °F, ft, ppm CaCO3  b)

Menu Setting Standard 


System\Start-Stop
Timer program 1  activated  deactivated deactivated
Autostart hrs  Mon  Tue  Wed  Thu  Fri  Sat 
Autostop hrs  Sun 
Timer program 2  activated  deactivated deactivated
Autostart hrs  Mon  Tue  Wed  Thu  Fri  Sat 
Autostop hrs  Sun 
Timer program 3  activated  deactivated deactivated
Autostart hrs  Mon  Tue  Wed  Thu  Fri  Sat 
Autostop hrs  Sun 
Timer program 4  activated  deactivated deactivated
Autostart hrs  Mon  Tue  Wed  Thu  Fri  Sat 
Autostop hrs  Sun 
System\Rinse
Automatic rinse  activated  deactivated activated
Interval h 4h 
Rinse volume L/gal 100 L 
First start time hrs 00:00 hrs 
System\Date and time
Date checked 
Time checked 
Summer / winter time  activated  deactivated deactivated 
System\Language
Set language English 
System\Report
Daily report time hrs 00:00 hrs 
System\Service\Settings\Disinfection
Distribution time min 5 min 
Dwell time min 10 min 
Rinse volume L/gal 200 L 

12-16 Fresenius Medical Care AquaBplus SM-EN 9A-2019


Chapter 12: Appendix

Menu Setting Standard 


System\Service\Settings\Heat disinfection
(Menu only visible if the AquaBplus HF option is present)
Holding time min 30 min 
Consumption time min 30 min 
Over-run min 15 min 
Heat disinfection time max. min 120 min 
Specified min. return temp. °C/°F 80 °C 
Feed temperature max. °C/°F 85 °C 
Timer program 1  activated  deactivated deactivated
Start hrs  Mon  Tue  Wed  Thu  Fri  Sat 
 Sun
Timer program 2  activated  deactivated deactivated
Start hrs  Mon  Tue  Wed  Thu  Fri  Sat 
 Sun
System\Service\Settings\Yield
Yield % 55 % 
System\Service\Settings\Alarm limits
(* = parameter only visible if the AquaBplus B2 option is present)
Permeate conductivity max. µS/cm 40 µS/cm 
Permeate temperature max. °C/°F 32 °C 
Conductivity alarm delay s 300 s 
Soft water conductivity max. µS/cm 1000 µS/cm 
Soft water temperature max. °C/°F 30 °C 
Soft water supply min. L/min / gal/min 25 L/min 
Min. concentrate pressure bar/psi 4 bar 
Max. concentrate pressure bar/psi 15 bar 
Permeate conductivity of stage 1 , max. * µS/cm 50 µS/cm 
Permeate temperature of stage 1 , max. * °C/°F 30 °C 
Concentrate pressure of stage 2 , min. * bar/psi 4 bar 
Concentrate pressure of stage 2 , max. * bar/psi 15 bar 
System\Service\Settings\Ethernet
DHCP  activated  deactivated activated 
IP address . . . – 
Subnet mask . . . – 
Default gateway . . . – 
1st DNS server . . . – 
2nd DNS server . . . – 
System\Service\Configuration
AquaBplus B2  activated  deactivated – 
RingBase  activated  deactivated – 
AquaBplus HF  activated  deactivated – 
Tank control  activated  deactivated – 
Tank & Ring  activated  deactivated – 
HOTfeed  activated  deactivated – 
System capacity L/h / gal/h – 
Ring main length m/ft – 
SN AquaBplus BA S – 
SN AquaBplus B2 BD S – 
SN AquaBplus HF BF S – 
External failure  activated  deactivated – 
Level ext. Failure  high  low – 

Comments:

Date: Signature: Stamp: