Service Troubleshooting Guide

Radiology Solutions Services

Document ID: 68274899


NX
Type 4406

Overview of Error Codes on NX workstation

► Purpose of this document

This document gives an overview of all error codes which may be present
on the NX workstation

► Changes compared to previous revision

The following modifications have been implemented:


Added
• SDRMESS (Error messages from Smart DR (DR 600, DR 100s))
• SSCMESS (Error messages from Sedecal Softconsole)
• THXMESS (Error messages from Technix (DR100e))

DOCUMENT CONTROL NOTE:


The controlled version of this document is available from the Agfa Medimg Library. Any printed copy of this document is uncontrolled.

Edition 1, Revision 2 Document_standard_e_template_v06 Page 1 of 1


11-2020 Copyright © 2020 Agfa NV Agfa Company Confidential
APCMESS
(Error messages from Agfa Positioner Control (DR 400, DR 600))

Status 2020-10-28
Error Code Summary Message Instruction
APCMESS68 Grid warning A grid is required. Please insert grid.
APCMESS69 Grid Warning No grid is required. Please remove grid.
APCMESS71 Grid warning SID is not suited for the inserted grid.
APCMESS73 Connection Error No communication between NX Workstation and Please check connection.
the modality.
APCMESS74 License Error There are no matching licenses found. Please check the license file.
APCMESS75 Software version error The firmware version of the modality is not Please contact your FSE.
compatible with the version of APC. No
communication possible.
APCMESS77 Bucky error Please insert detector.
APCMESS78 Bucky error Please insert cassette.
APCMESS79 Bucky error Please change cassette.
APCMESS82 Grid error Please check if grid is correctly inserted.
APCMESS83 Collimator error Collimator encountered a problem. Please check collimator. Restart if problem
persists.
APCMESS84 Motion warning Travel limit reached. Automatic movement was Please re-position.
stopped.
APCMESS85 Collision warning Tube Head Collision possible. Move system manually to desired position.
APCMESS89 Modality error System encountered a problem: {0}. Restart system and call service if problem
persists.
APCMESS90 Modality initializing Modality initializing
APCMESS91 No connection with THC No connection with THC. No verify possible.
APCMESS96 Error setting autoposition Invalid parameters for automatic motion Please call sevice if the problem persists.
APCMESS97 Automatic motion warning Automatic motion currently not possible Please move system manually to desired
position.
APCMESS99 Automatic motion warning Automatic motion currently not possible Please move tube head up until this message
disappears and restart automatic motion.
APCMESS100 Automatic motion warning Automatic motion currently not possible Please move tube head down until this message
disappears and restart automatic motion.

APCMESS101 Automatic motion warning Automatic motion currently not possible Please move tube head left until this message
disappears and restart automatic motion.
APCMESS102 Automatic motion warning Automatic motion currently not possible Please move tube head right until this message
disappears and restart automatic motion.

APCMESS103 Automatic motion warning Automatic motion currently not possible Please move tube head back until this message
disappears and restart automatic motion.

APCMESS104 Automatic motion warning Automatic motion currently not possible Please move tube head forward until this
message disappears and restart automatic
motion.
APCMESS105 Automatic motion warning System encountered a problem during automatic Please move system manually to desired
motion. Target not reached. position.
APCMESS108 Loading configuration from file Some configuration settings have not yet been Do you wish to continue loading the configuration
saved. Loading a new configuration file will file and lose the unsaved settings?
dismiss these setting.
APCMESS109 Bucky drawer error Please close bucky drawer
APCMESS111 Centering warning Automatic centering currently not possible. Please center tube manually.
APCMESS112 Collimator Initialization error Collimator encountered a problem. Please ensure that collimator is in automatic
mode and restart system.
APCMESS113 Table Collision warning Problem moving the table. Please check if objects are under the table top.
APCMESS114 SID not reachable SID not reachable. Please lower table.
APCMESS117 Modality error Emergency stop button is pressed. Please release button and restart system.
APCMESS118 Automatic motion warning Target position cannot be reached. Please position system manually.
APCMESS119 Detector Format cannot be detected Format of inserted cassette/detector cannot be Please set collimation area manually.
detected by ACSS in selected Bucky or battery of
detector is not inserted.
APCMESS121 Modality error An issue with the suspension rope has occurred, Call service immediately.
it is no longer allowed to operate the system.

APCMESS122 Wall stand warning Requested movement not possible due to


possible collision by the overhead handle.
APCMESS123 Position not reachable Target position cannot be reached. Please push table top aside.
APCMESS124 Remote Control warning Remote control encountered a problem and is not Please restart system. If error persists, please call
usable any longer. service (error {0}).
APCMESS125 Position not reachable SID is not reachable. Please lower wall stand bucky.
APCMESS126 Unintended Movement Unintended movement of Tube or Bucky. Please confirm and call service if the problem
persists.
APCMESS127 Table and Wall not operational Neither table nor wall stand are operational, free Please call service.
exams are possible.
APCMESS128 Table not operational The table is not operational, free exams and wall Please call service.
stand exams are possible.

Edition 1, Revision 2
11-2020 Livelink ID: 68274899 APCMESS
APCMESS
(Error messages from Agfa Positioner Control (DR 400, DR 600))

Status 2020-10-28
Error Code Summary Message Instruction
APCMESS129 Wall stand not operational The Wall stand is not operational, free exams and Please call service.
table exams are possible.
APCMESS130 Grid Error System is unable to determine the grid type. Please check the grid. Contact service if the
problem persists.
APCMESS131 License Error A matching license has been found, but there is a This installation will be aborted.
license version mismatch.
APCMESS132 Seismic Sensor Board defective Earthquake detection control is out of operation Please call service.
due to defective SSB board.
APCMESS133 System detected an earthquake System detected an earthquake. Restart the system
APCMESS134 System detected an earthquake System detected an earthquake. Only manual Please call service.
positioning possible.
APCMESS135 DTS positioning warning Start and/or end position of Digital Tomosynthesis Please re-position.
Exam not valid or tube head collision possible.

APCMESS136 [DTS] Positioning error System encountered a problem during automatic Please restart Tomosynthesis sequence.
motion. Target not reached.
APCMESS137 [DTS] Move to initial position Please move to the initial position
APCMESS138 Autoposition modality error One or more Free or Parking-Cleaning positions Adjust the Tube Reference of the relevant
has {0} set as Tube Reference. positions before disabling the modality.
APCMESS139 Autoposition modality warning The Tube Reference of the Free positions has Check the configuration of the X-Y-Z axes for
been adjusted. these positions.
APCMESS140 DTS positioning warning DTS aborted. Table adjustment detected after
start of DTS workflow.
APCMESS141 Foot pedal lock is activated Foot pedal lock is activated. Please check the foot pedal lock switch.
APCMESS142 Bucky Movement detected Bucky movement detected. Please check positioning.

Edition 1, Revision 2
11-2020 Livelink ID: 68274899 APCMESS
ASCMESS
(Error messages from Agfa SoftConsole)

Status 2020-10-29
Error Code Summary Message Instruction
ASCMESS1 Installation not allowed There are no matching licenses found. This
installation will be aborted
ASCMESS2 Version not supported A matching license has been found, but there is a
license version mismatch. This installation will be
aborted
ASCMESS3 No user permission You do not have enough priviliges to install this
product.
Administrator priviliges are needed.
This installation will be cancelled.
ASCMESS10 Press Move button until initial position has been Press Move button until initial position has been Press Move button until initial position has been
reached. reached. reached.
ASCMESS11 Please adjust collimation field width if required Please adjust collimation field width if required Please adjust collimation field width if required

Edition 1, Revision 2
11-2020 Livelink ID: 68274899 ASCMESS
AutoQC2MESS
(Error messages from AutoQC2)

Status 2020-10-29
Error Code Summary Message Instruction
MESS1 Are you sure to remove exposure <exposure- Are you sure to remove exposure <exposure-
type> ? type> ?
MESS2 This exposure is used for different tests. Are you This exposure is used for different tests. Are you
sure t... sure t...
MESS3 No Data Available. No Data Available.
MESS4 No exposure selected. No exposure selected.
MESS5 Do you want to save the modified data? Do you want to save the modified data?
MESS6 Incorrect screen resolution. Please contact your Incorrect screen resolution. Please contact your
system a... system a...
MESS7 (nr-of-exposures) exposures waiting for (nr-of-exposures) exposures waiting for
processing, maxim... processing, maxim...
MESS10 Are you sure to remove the selected report(s)? Are you sure to remove the selected report(s)?
MESS12 Not possible to save the data. One or more Not possible to save the data. One or more
values are out... values are out...
MESS13 No reports found. No reports found.
MESS14 No history data found. No history data found.
MESS16 Please check serial number or remove tube with Please check serial number or remove tube with
number <tu... number <tu...
MESS17 No tube selected. Please select a tube first. No tube selected. Please select a tube first.
MESS18 Illegal Tube Serial Number. Illegal Tube Serial Number.
MESS19 Are you sure to remove X-ray Tube <Tubenr> ? Are you sure to remove X-ray Tube <Tubenr> ?
MESS20 Online help is not installed on this system. Online help is not installed on this system.
MESS24 Some of the images appear not to be flat Some of the images appear not to be flat
fields.Please ... fields.Please ...
MESS25 <file name> not found.The file '<file name>': could <file name> not found.The file '<file name>': could
not... not...
MESS26 Image analysis failed. Image analysis failed.
MESS27 Report file (<filename>) creation failed.Please Report file (<filename>) creation failed.Please
contact... contact...
MESS28 Export data via file (<filename>) creation failed. Export data via file (<filename>) creation failed.
MESS29 No Report Selected No Report Selected
MESS30 This Report contains no Exposures. This Report contains no Exposures.
MESS31 Sorry, no data available to export. Sorry, no data available to export.
MESS32 File (<file name>) could not be found. File (<file name>) could not be found.
MESS33 Not possible to generate the results. Some of the Not possible to generate the results. Some of the
images ... images ...
MESS34 License in Grace Period.Grace Period remaining License in Grace Period.Grace Period remaining
time: <t... time: <t...
MESS35 No license available.Check logging, call service. No license available.Check logging, call service.

MESS39 Not possible to generate the results. Some of the Not possible to generate the results. Some of the
images ... images ...
MESS40 No tube available. No tube available.
MESS41 License Error License Error
MESS42 Grace Period Expired Grace Period Expired
MESS43 Not possible to save data, empty fields are not Not possible to save data, empty fields are not
allowed allowed
MESS44 DICOM Import failed.Contact your service DICOM Import failed.Contact your service
engineer. engineer.
MESS45 Invalid input. Invalid input.
MESS46 Database already upgraded with latest version. Database already upgraded with latest version.

Edition 1, Revision 2
11-2020 Livelink ID: 68274899 AutoQC2MESS
CDIMESS
(Error messages from Canon-Detector (DX-D30, DX-D35))

Error Code Title Message Instruction Argument


CDIMESS1 Installation Error No installation possible The maximum amount of installed panels is
reached. \nTo install this panel, de-install another
panel first
CDIMESS7 Link Message Panel {0} is successfully linked Panel name
CDIMESS9 Link Error Linking Failed. The panel is not installed. To link this panel, be sure it is installed first.
CDIMESS20 Calibration step succeeded with a dose of {0}% actual dose
\n(Dose should be between 80% and 120%)
CDIMESS21 Calibration step failed with a dose of {0}% \n(Dose actual dose
should be between 80% and 120%). Use lower
dose.
CDIMESS22 Calibration step failed with a dose of {0}% \n(Dose actual dose
should be between 80% and 120%). Use higher
dose.
CDIMESS23 Verifying calibration. Step {0} of {1}. stepnumber
CDIMESS23 Verifying calibration. Step {0} of {1}. stepcount
CDIMESS25 Calibration verification result:\n{0} Verification failed, please recalibrate.\nIf this Each TestName: TestResult
problem persists, contact your system
administrator.
CDIMESS26 Connection not stable enough. Please check the connection of panel {0} panel name
CDIMESS27 Connection error There is no connection with panel {0}. Please check the panel connection. panel name
CDIMESS28 Battery of panel {0} not fully charged. Please replace the battery with a fully charged panel name
battery.
CDIMESS29 Current firmware version of panel {0} is equal to Current firmware version is {1}.\nUpdate firmware panel name
update firmware version. version is {2}.\nDo you want to continue?

CDIMESS29 Current firmware version of panel {0} is equal to Current firmware version is {1}.\nUpdate firmware current firmware version
update firmware version. version is {2}.\nDo you want to continue?

CDIMESS29 Current firmware version of panel {0} is equal to Current firmware version is {1}.\nUpdate firmware update firmware version
update firmware version. version is {2}.\nDo you want to continue?

CDIMESS30 Current firmware version of panel {0} is higher Current firmware version is {1}\nUpdate firmware panel name
than update firmware version. version is {2}\nDo you want to continue?

CDIMESS30 Current firmware version of panel {0} is higher Current firmware version is {1}\nUpdate firmware current firmware version
than update firmware version. version is {2}\nDo you want to continue?

CDIMESS30 Current firmware version of panel {0} is higher Current firmware version is {1}\nUpdate firmware update firmware version
than update firmware version. version is {2}\nDo you want to continue?

CDIMESS31 Re-boot and re-link Please perform following actions:\n1. Re-boot panel name
panel {0}\n2. Wait until the panel's LEDs have
following state:\nPower = Stable; Ready = Off;
Link = Off \n3. Re-link panel {0} to this
workstation.
CDIMESS32 No connection. The firmware update can not be verified. Please check the connection with panel {0}. panel name
CDIMESS33 Updated firmware version of panel {0} is not the Updated firmware version is {1}. \nExpected panel name
expected update firmware version firmware version is {2}.
CDIMESS33 Updated firmware version of panel {0} is not the Updated firmware version is {1}. \nExpected updated firmware version
expected update firmware version firmware version is {2}.
CDIMESS33 Updated firmware version of panel {0} is not the Updated firmware version is {1}. \nExpected expected firmware version
expected update firmware version firmware version is {2}.
CDIMESS34 Firmware Update of panel {0} failed panel name
CDIMESS35 Calibration Mode {0} of {1}\n{2}\nTake exposure Number of calibration mode
{3} of {4}
CDIMESS35 Calibration Mode {0} of {1}\n{2}\nTake exposure total number of calibration modes
{3} of {4}
CDIMESS35 Calibration Mode {0} of {1}\n{2}\nTake exposure Calibration Mode Name
{3} of {4}
CDIMESS35 Calibration Mode {0} of {1}\n{2}\nTake exposure number of exposure
{3} of {4}
CDIMESS35 Calibration Mode {0} of {1}\n{2}\nTake exposure total number of exposures
{3} of {4}
CDIMESS36 Update Succeeded
CDIMESS37 Dead Pixel Map Update of panel {0} failed panel name
CDIMESS38 Uniformity Test failed \nPlease check collimation.
The panel should be fully covered.\nPlease check
beam path for objects.
CDIMESS39 Panel {0} is not yet calibrated on this system Panel Name
\nPlease make sure that there are no blocking
objects on the panel. \nPlease make sure that the
collimated area fully covers the panel.
CDIMESS40 Take exposure to perform a Uniformity Check
\nSee "Calibration Key User Manual" for
exposure parameters
CDIMESS41 Test Finished Panel {0} connected Panel Name
CDIMESS42 Test Finished Panel {0} not connected Panel Name
CDIMESS43 Successfully Linked panel {0} with serial number Panel Name
{1}
CDIMESS43 Successfully Linked panel {0} with serial number Serial Number
{1}
CDIMESS44 The filename should not be empty
CDIMESS45 The filename does not have the correct file
extension

Edition 1, Revision 2
11-2020 Livelink ID: 68274899 CDIMESS
CDIMESS
(Error messages from Canon-Detector (DX-D30, DX-D35))

Error Code Title Message Instruction Argument


CDIMESS46 The file does not exist
CDIMESS47 The name of the file is not specified
CDIMESS48 The dead pixel file does not belong to the
selected panel
CDIMESS49 The filename is not a valid firmware
CDIMESS50 The filename does not contain a valid version
number
CDIMESS51 NX Conflict Can't start DX-D30C link tool while other NX
software is running
CDIMESS52 Panel connection error Panel Disconnected
CDIMESS53 Panel {0} went into auto-sleep mode. Calibration panel name
failed \nTo calibrate panel {0}, re-start calibration
procedure.
CDIMESS54 License Error License error: license {0} for {1} not found. Please check the license file. TECHNICAL-LICENSENAME
CDIMESS54 License Error License error: license {0} for {1} not found. Please check the license file. COMPONENT/APPLICATION
CDIMESS55 Updating the XIF-box The XIF-box is updated with the new IP address To enable the new IP-address of the XIF-box, the New XIF IP Address
succeeded {0} XIF-box needs to perform a power cycle.\nPlease
check the connectivity after the power cycle.

CDIMESS56 XIF-box not found. The XIF-box is not found on current IP-address Please check if:\n- The XIF-box is connected.\n- Current XIF IP Address
{0}. The IP-address of the XIF-box is {0}
CDIMESS57 Updating the XIF-box failed Updating the XIF-box with IP-address {0} failed. Please try again. New XIF IP Address
CDIMESS58 Test Finished XIF-box connected
CDIMESS59 Test Finished XIF-box not connected
CDIMESS60 Uniformity test failed.\nThe average pixel value is Please check collimation. The panel should be average pixel value percentage
{0}%.\n(Average pixel value should be between fully covered.\nPlease check beam path for
{1}% and {2}%). objects.\nPlease check the "Calibration Key User
Manual" for exposure parameters.
CDIMESS60 Uniformity test failed.\nThe average pixel value is Please check collimation. The panel should be lower limit average pixel value
{0}%.\n(Average pixel value should be between fully covered.\nPlease check beam path for percentage
{1}% and {2}%). objects.\nPlease check the "Calibration Key User
Manual" for exposure parameters.
CDIMESS60 Uniformity test failed.\nThe average pixel value is Please check collimation. The panel should be upper limit average pixel value
{0}%.\n(Average pixel value should be between fully covered.\nPlease check beam path for percentage
{1}% and {2}%). objects.\nPlease check the "Calibration Key User
Manual" for exposure parameters.
CDIMESS61 Incorrect Detector Type The selected detector type {0} for panel {1} is Check the detector type of panel {1} in the CDI selected detector type
incorrect. Panel Installation Tool.
CDIMESS61 Incorrect Detector Type The selected detector type {0} for panel {1} is Check the detector type of panel {1} in the CDI panel name
incorrect. Panel Installation Tool.
CDIMESS62 No Service Rights {0} is only available for Service accounts Tool name
CDIMESS63 The current panel {0} is not in the bucky. Please put panel {0} in the bucky and click "OK" to panel name
continue.
CDIMESS65 Calibration Mode {0} of {1}\n{2}\nTake exposure Number of calibration mode
{3} of {4}\nThis Calibration Mode should be done
at {5} fps.
CDIMESS65 Calibration Mode {0} of {1}\n{2}\nTake exposure total number of calibration modes
{3} of {4}\nThis Calibration Mode should be done
at {5} fps.
CDIMESS65 Calibration Mode {0} of {1}\n{2}\nTake exposure Calibration Mode Name
{3} of {4}\nThis Calibration Mode should be done
at {5} fps.
CDIMESS65 Calibration Mode {0} of {1}\n{2}\nTake exposure number of exposure
{3} of {4}\nThis Calibration Mode should be done
at {5} fps.
CDIMESS65 Calibration Mode {0} of {1}\n{2}\nTake exposure total number of exposures
{3} of {4}\nThis Calibration Mode should be done
at {5} fps.
CDIMESS65 Calibration Mode {0} of {1}\n{2}\nTake exposure Frames Per second for current
{3} of {4}\nThis Calibration Mode should be done calibration mode
at {5} fps.
CDIMESS66 Test Finished Power Box connected
CDIMESS67 Test Finished Power Box not connected
CDIMESS68 Updating the Power Box The Power Box is updated with the new IP To enable the new IP-address of the Power Box , New Power Box IP Address
succeeded address {0} the Power Box needs to perform a power
cycle.\nPlease check the connectivity after the
power cycle.
CDIMESS69 Power Box not found. The Power Box is not found on current IP-address Please check if:\n- The Power Box is connected.\n- Current Power Box IP Address
{0}. The IP-address of the Power Box is {0}

CDIMESS70 Updating the Power Box failed Updating the Power Box with IP-address {0} Please try again. New Power Box IP Address
failed.
CDIMESS71 Cooling unit failed Calibration failed because a cooling unit error Please check your cooling unit, then restart the
calibration.\nIf the problem persist, please contact
your Service Engineer
CDIMESS73 The cooling unit is not working or is detached. Please check the cooling unit.

Edition 1, Revision 2
11-2020 Livelink ID: 68274899 CDIMESS
CDIMESS
(Error messages from Canon-Detector (DX-D30, DX-D35))

Error Code Title Message Instruction Argument


CDIMESS74 Calibration step failed with a dose of {0}% \n(Dose actual dose
should be between 80% and 120%). Use lower
dose.\n\nCalibration Mode {1} of {2}\n{3}\nTake
exposure {4} of {5}\nThis Calibration Mode should
be done at {6} fps.
CDIMESS74 Calibration step failed with a dose of {0}% \n(Dose Number of calibration mode
should be between 80% and 120%). Use lower
dose.\n\nCalibration Mode {1} of {2}\n{3}\nTake
exposure {4} of {5}\nThis Calibration Mode should
be done at {6} fps.
CDIMESS74 Calibration step failed with a dose of {0}% \n(Dose total number of calibration modes
should be between 80% and 120%). Use lower
dose.\n\nCalibration Mode {1} of {2}\n{3}\nTake
exposure {4} of {5}\nThis Calibration Mode should
be done at {6} fps.
CDIMESS74 Calibration step failed with a dose of {0}% \n(Dose Calibration Mode Name
should be between 80% and 120%). Use lower
dose.\n\nCalibration Mode {1} of {2}\n{3}\nTake
exposure {4} of {5}\nThis Calibration Mode should
be done at {6} fps.
CDIMESS74 Calibration step failed with a dose of {0}% \n(Dose number of exposure
should be between 80% and 120%). Use lower
dose.\n\nCalibration Mode {1} of {2}\n{3}\nTake
exposure {4} of {5}\nThis Calibration Mode should
be done at {6} fps.
CDIMESS74 Calibration step failed with a dose of {0}% \n(Dose total number of exposures
should be between 80% and 120%). Use lower
dose.\n\nCalibration Mode {1} of {2}\n{3}\nTake
exposure {4} of {5}\nThis Calibration Mode should
be done at {6} fps.
CDIMESS74 Calibration step failed with a dose of {0}% \n(Dose Frames Per second for current
should be between 80% and 120%). Use lower calibration mode
dose.\n\nCalibration Mode {1} of {2}\n{3}\nTake
exposure {4} of {5}\nThis Calibration Mode should
be done at {6} fps.
CDIMESS75 Calibration step failed with a dose of {0}% \n(Dose actual dose
should be between 80% and 120%). Use higher
dose.\n\nCalibration Mode {1} of {2}\n{3}\nTake
exposure {4} of {5}\nThis Calibration Mode should
be done at {6} fps.
CDIMESS75 Calibration step failed with a dose of {0}% \n(Dose Number of calibration mode
should be between 80% and 120%). Use higher
dose.\n\nCalibration Mode {1} of {2}\n{3}\nTake
exposure {4} of {5}\nThis Calibration Mode should
be done at {6} fps.
CDIMESS75 Calibration step failed with a dose of {0}% \n(Dose total number of calibration modes
should be between 80% and 120%). Use higher
dose.\n\nCalibration Mode {1} of {2}\n{3}\nTake
exposure {4} of {5}\nThis Calibration Mode should
be done at {6} fps.
CDIMESS75 Calibration step failed with a dose of {0}% \n(Dose Calibration Mode Name
should be between 80% and 120%). Use higher
dose.\n\nCalibration Mode {1} of {2}\n{3}\nTake
exposure {4} of {5}\nThis Calibration Mode should
be done at {6} fps.
CDIMESS75 Calibration step failed with a dose of {0}% \n(Dose number of exposure
should be between 80% and 120%). Use higher
dose.\n\nCalibration Mode {1} of {2}\n{3}\nTake
exposure {4} of {5}\nThis Calibration Mode should
be done at {6} fps.
CDIMESS75 Calibration step failed with a dose of {0}% \n(Dose total number of exposures
should be between 80% and 120%). Use higher
dose.\n\nCalibration Mode {1} of {2}\n{3}\nTake
exposure {4} of {5}\nThis Calibration Mode should
be done at {6} fps.
CDIMESS75 Calibration step failed with a dose of {0}% \n(Dose Frames Per second for current
should be between 80% and 120%). Use higher calibration mode
dose.\n\nCalibration Mode {1} of {2}\n{3}\nTake
exposure {4} of {5}\nThis Calibration Mode should
be done at {6} fps.
CDIMESS375 Error DR-detector in location {0} having serial number Please contact your service engineer
{1} received error : {2}.
CDIMESS376 Error A DR Detector Error ({0}) occured for DR-detector If the problem persists, please contact your
in location {1} having serial number {2}. Service Engineer.

CDIMESS2000 Calibration step skipped.


CDIMESS2001 Calibration step failed, please remove grid and try
again.
CDIMESS2002 Error when loading panel Error when loading panel Please make sure the panel is installed and
configured correctly.
CDIMESS2003 Checking battery and wireless connection...
CDIMESS2004 Updating dead pixel file...
CDIMESS2005 Updating firmware...
CDIMESS2006 Verifying firmware update...
CDIMESS2007 Shutting down the update process...
CDIMESS2008 Please select the firmware file to update the panel

Edition 1, Revision 2
11-2020 Livelink ID: 68274899 CDIMESS
CDIMESS
(Error messages from Canon-Detector (DX-D30, DX-D35))

Error Code Title Message Instruction Argument


CDIMESS2009 Please select the dead pixel file to update the
panel
CDIMESS2010 Select Firmware File
CDIMESS2011 Select Dead Pixel file
CDIMESS2012 Please verify the specified configuration and
\npress the next button to start the update
process.
CDIMESS2013 Selected Configuration:
CDIMESS2014 Panel to update:
CDIMESS2015 Select the panel to update
CDIMESS2016 Firmware file:
CDIMESS2017 Dead Pixel file:
CDIMESS2018 Confirm Settings
CDIMESS2019 Select Panel
CDIMESS2020 Dynamic exposure error It is not possible to make a dynamic exposure with Please make sure to use a RF panel for dynamic
panel {0} exposures.

Edition 1, Revision 2
11-2020 Livelink ID: 68274899 CDIMESS
CPIMESS
(Error messages from CPI-Generator (DR 800))

Status 2020-10-29
Error Code Summary Message Instruction
CPIMESS1 License Error There are no matching licenses found. Please check the license file.
CPIMESS2 Generator Error Generator encountered a problem: <ErrorID>. Press OK to reset the generator. Please call
service if the problem persists.
CPIMESS3 Generator Warning Generator Message: <ErrorID>
CPIMESS4 Generator Connection Error The connection with the generator cannot be Please make sure that the ethernet connection
established. between the NX Workstation and the Generator is
properly plugged in and restart the NX
Workstation. Please call service if the problem
persists.
CPIMESS5 License Error A matching license has been found, but there is a This installation will be aborted.
license version mismatch.
CPIMESS6 DAP Not Ready DAP device is in warm-up state. Please allow the DAP chamber to warm up.
CPIMESS7 DAP Device Error DAP device encountered an Error : <ErrorID>. Please reset the system and call service if the
problem persists.
CPIMESS8 AEC Device Error AEC device encountered an Error. Please reset the system and call service if the
problem persists.
CPIMESS9 AEC Backup Time AEC exposure exceeded the allowed backup
time.
CPIMESS10 AEC max mAs AEC exposure exceeded the allowed max mAs.

CPIMESS11 Tomo Device Error Tomo device encountered an Error. Please reset the system and call service if the
problem persists.
CPIMESS12 Tomo Backup Time Tomo exposure exceeded the programmed
backup time.
CPIMESS13 Generator High Temperature Generator detected a high internal temperature. Please allow the generator to cool down. Please
call service if the problem persists.
CPIMESS14 Generator Cycle Limit Generator exceeded its calculated duty cycle Please allow the generator to cool down. Please
limit. call service if the problem persists.
CPIMESS15 Generator Corrupted Data Generator encountered corrupted data. Please load the factory defaults and call service if
the problem persists.
CPIMESS16 Generator COMM Error Generator encountered internal COMM error: Please check the generator documentation and
<ErrorID>. call service if the problem persists.
CPIMESS17 Generator License Generator detected no or invalid license. Please contact service.
CPIMESS18 Feature Lock Feature locked due to license issue. Please contact service.
CPIMESS19 Manual Abort Exposure was manually aborted by the operator.

CPIMESS20 Generator Prep Timeout Generator has encountered a Prep timeout.


CPIMESS21 Generator heat warning Generator tube encountered a heat warning. Please allow the tube to cool down before making
further exposures
CPIMESS22 Generator heat limit Generator tube exceeded its heat limit. Please wait for the tube to cool down.
CPIMESS23 Exposure not allowed Exposure not allowed due to system state. Pleas check all system components and contact
service if the problem persists.
CPIMESS24 mAs not reached Configured mAs could not be reached within the Please check the exposure configuration and
allocated time. contact service if the problem persists.
CPIMESS25 Parameter limit reached Parameter limit reached: <Parameter>
CPIMESS26 Invalid parameter change Invalid Parameter Change: <Parameter>
CPIMESS27 kV During Standby kV is detected during standby. Please shut down the generator and contact
service
CPIMESS28 AEC Dose out of range AEC Dose out of Range. Exposure parameters
could not be set
CPIMESS29 DAP Device Not Responding DAP Device is not responding. Please call service if the problem persists.

Edition 1, Revision 2
11-2020 Livelink ID: 68274899 CPIMESS
DAP_MESS
(Error messages from DAP-Meter)

Status 2019-09-26
Error Code Summary Message Instruction
DAP_MESS1 DAP is disconnected The DAP is disconnected from the NX Please check the connection between the NX
Workstation or is damaged. Workstation and the DAP meter. Replace if
necessary.
DAP_MESS2 DAP chamber is not responding The DAP console has no connection with the Please check the power and indication lights of
measuring chamber the measuring chamber. Contact your FSE if the
problem persists.
DAP_MESS3 DAP Failure The DAP-meter generated an error. Please check the display of the DAP-meter for the
correct error code and check the manual for
corrective actions.
DAP_MESS4 No valid license No valid license found! DAP connection will not Please check the license file.
work.
DAP_MESS5 Invalid License Version Invalid license version found! DAP connection will Please check the license file.
not work.
DAP_MESS6 No Valid License No valid license found! DAP connection will not Please check the license file.
work.
DAP_MESS7 Invalid License Version Invalid license version found! DAP connection will Please check the license file.
not work.
DAP_MESS8 COM port Error Access is denied to all COM ports. Please check if other applications are blocking the
COM ports. Contact your FSE if the problem
persists.

Edition 1, Revision 2
11-2020 Livelink ID: 68274899 DAP_MESS
DMIMESS
(Error messages from DR Retrofit Panels)

Status 2020-10-29
Error Code Summary Message Instruction
DMIMESS1 License Error License error: license <TECHNICAL- Please check the license file.
LICENSENAME> for
<COMPONENT/APPLICATION> not found.
DMIMESS2 <Software name + version> Setup A matching license has been found, but there is a
license version mismatch./nThis installation will be
aborted.
DMIMESS3 <Software name + version> Setup There are no matching licenses found. This
installation will be aborted.
DMIMESS4 <Software name + version> Setup You do not have enough privileges to install this
product.\nAdministrator privileges are
needed.\nThis installation will be cancelled.
DMIMESS5 SW Conflict Can't start <Tool Name> while other NX software
is running.
DMIMESS6 Software Conflict Cannot start <Tool Name> while another instance
of <Tool Name> is running.
DMIMESS7 Save Changes Do you want to save your changes?
DMIMESS8 Temperature Warning DR Detector <detector name>'s temperature is
high (&gt; <warning high temperature> degrees
Celsius).\nImage quality might be impacted.
DMIMESS9 Temperature Error DR Detector <detector name>'s temperature is The DR Detector is in sleep mode to cool down.
too high (&gt; <error high temperature> degrees
Celsius).
DMIMESS10 Warming up DR Detector <detector name> is warming When the DR Detector is warmed-up, normal
up.\nDR Detector's internal temperature = <DR operation is possible.
Detector's temperature> degrees Celsius.
DMIMESS11 DR Detector Sleep Mode The DR Detector is going into sleep within To continue working, please re-click the
<duration untill panel goes into sleep> minutes. thumbnail.

DMIMESS12 Configuration Error We detected missing map files! Would you like to install them from a CD /
directory?
DMIMESS13 Temperature Error DR Detector <detector name>'s temperature is The DR Detector is in aging mode to warm up.
too low (&lt; <error lowtemperature> degrees
Celsius).
DMIMESS14 Firmware upgrade instruction A web browser will be launched, which will allow
you to upgrade the detectors firmware.
DMIMESS15 Data missing Following data is missing:\n<list of missing
maps>\n\nPlease call service.
DMIMESS16 No Connection No connection with panel <panel's name>
DMIMESS17 Connection error DR Detector <detector name> was disconnected
and the software could not re-connect. Please
check cables and restart NX, DR Detector and
the SSU-box completely.
DMIMESS18 Waiting for Connection... This might take up to 2 minutes on initial
start.\nMake sure your network settings are
properly configured
DMIMESS19 New panel detected We detected a new panel, would you like to install
it?
DMIMESS20 Install Maps from CD For this next step, please insert your panel's
installation CD with correct Serial Number
Label.\nIf the data is on another device (eg. USB)
please insert it.\n\nClick Ok to continue.
DMIMESS21 install Maps from CD Reading device...\n<status information>
DMIMESS22 Verifying Serial Number Please wait a moment until the verification is
complete...\nMake sure the panel is connected.
The IP Ping indicator should be green.

DMIMESS23 Installation Succesful Serial Number is correct!


DMIMESS24 Retrieve logs Downloading logs from panel... (<log
number>/<total number of log>)
DMIMESS25 New panel detected... We detected a new panel, would you like to install
it?\nMark that this will overwrite the current
installed panel!
DMIMESS27 Retrieve Logs Logging is successfully retrieved.
DMIMESS28 No Administrator Rights <Tool name> is only available for Administrator
accounts.
DMIMESS29 Configuration error The connected DR Detector is not configured.
Please connect a configured panel.\nIf the
problem persists, please call Service.
DMIMESS30 Check beam path Please check that:\nThere are no unwanted
objects in the beam path\nCompression paddle is
removed\nThe dedicated PMMA block is placed
on the bucky\nMo-Mo filter is used\n\nNote that
there must be minimal 1 minute between each
exposure.
DMIMESS31 Warming up - Calibraiton DR Detector <detector name> is warming When the DR Detector is warmed-up, calibration
up.\nDR Detector's internal temperature = <DR flow will continue.\nPlease wait...
Detector's temperature> degrees Celsius.

Edition 1, Revision 2
11-2020 Livelink ID: 68274899 DMIMESS
DMIMESS
(Error messages from DR Retrofit Panels)

Status 2020-10-29
Error Code Summary Message Instruction
DMIMESS32 Temperature Warning - Calibration DR Detector <detector name>'s temperature is
high (&gt; <warning high temperature> degrees
Celsius).\nImage quality might be
impacted.\nCalibration flow ends.
DMIMESS33 Temperature Error - Calibration DR Detector <detector name>'s temperature is
too high (&gt; <error high temperature> degrees
Celsius).\nThe DR Detector is in sleep mode to
cool down.\nCalibration flow ends.
DMIMESS34 Calibration: Dose message Set kV: <kV>, mAs: <mAs>.
DMIMESS35 Calibration: Exposure Take exposure <imagenumber> of <number of
acquisitons per dose setting>.
DMIMESS37 Calibration: Processing Data Processing data...
DMIMESS40 Calibration: Average pixel value too low Calibration step failed. \nAverage pixel value of Use higher mAs.\nPlease check the "Calibration
<aveage pixel value>. \nTarget average pixel Key User Manual" for exposure parameters.
value is <target average pixel value>
DMIMESS41 Calibration: Average pixel value too high Calibration step failed. \nAverage pixel value of Use lower mAs.\nPlease check the "Calibration
<aveage pixel value>. \nTarget average pixel Key User Manual" for exposure parameters.
value is <target average pixel value>
DMIMESS43 Calibration: Calibration failed - generic Calibration failed with error: <error>. \nPress
continue to restart calibration.
DMIMESS44 Calibration: Time-out A time-out error occurred.\nCalibration failed.
DMIMESS45 Task executing A task is still executing, closing this tool may
cause unexpected behaviour or damage. Do you
want to close the tool?
DMIMESS46 Wrong Firmware The firmware on panel <panel name> is not Please update the firmware to a compatible
compatible with the installed DMI Software version
<version number>.
DMIMESS47 Wrong Firmware The firmware on panel <panel name> is not Please update the firmware to a compatible
compatible with the installed DMI Software version
<version number>.
DMIMESS48 Uniformity failed and ROI is full The detector surface uniformity test failed. Please inspect the image to find a cause or check
the beam path for disruptive objects.\nFilename:
<filename>\nFolder: <folder>\nCan the problem
be fixed, to redo the exposure and the uniformity
check?
DMIMESS49 Uniformity failed and ROI is made smaller The detector surface uniformity test failed.\n\nA
smaller area will be used to check the uniformity
again.
DMIMESS50 Uniformity test with smaller ROI succeeded The smaller area of the detector surface passed
the uniformity test.\n\nThe calibration continues
and the uniformity check will be performed on the
next exposures on a smaller area.

DMIMESS51 Uniformity test with smaller ROI did not succeed The smaller area of the detector surface did not Please inform the modality vendor to inspect the
resolve the uniformity test. images you can find here:\nFilename:
<filename>\nFolder: <folderpath>\Do you agree
to skip this uniformity check and continue with the
calibration procedure?
DMIMESS52 Uniformity test disabled The uniformity test failed and is disabled.\n\nBe
aware that the calibration continues and no
uniformity check will be performed on the next
exposures.
DMIMESS53 Uniformity failed end Uniformity test failed.
DMIMESS54 Wait after previous exposure Please wait: <time left> s
DMIMESS55 Incorrect Power Option The Power Option, 'Put the computer to sleep' is
not set to Never!\n(Current setting is <current
setting>).\n Please, change this setting to Never
and reboot the system.

Edition 1, Revision 2
11-2020 Livelink ID: 68274899 DMIMESS
GGCMESS
(Communication component for a variety of non-integrated retrofit generator types)

Status 2019-09-26
Error Code Summary Message Instruction
GGCMESS1 License Error There are no matching licenses found. Please check the license file.
GGCMESS2 License Error A matching license has been found, but there is a Please check the license file.
license version mismatch.
GGCMESS3 Generator Connection Error The connection with the generator cannot be Please make sure that the connection between
established. the NX Workstation and the Generator is properly
plugged in and restart the NX Workstation. Please
call service if the problem persists.

Edition 1, Revision 2
11-2020 Livelink ID: 68274899 GGCMESS
GMPMESS
(Error messages from Positioner component for communication with GMM DR 800)

Status 2020-10-29
Error Code Summary Message Instruction
GMPMESS4 License Error There are no matching licenses found. Please check the license file.
GMPMESS5 Positioner Error Positioner Hardware Failure: <ID> <Category>. Please contact service if the problem persists.
GMPMESS6 Connection Error The connection with the positioner cannot be Please check the connection between positioner
established. and NX Workstation. Please contact service if the
problem persists.
GMPMESS7 License Error A matching license has been found, but there is a This installation will be aborted.
license version mismatch.
GMPMESS8 Collimator Error Collimator encountered a problem. Please check collimator. Restart if problem
persists.
GMPMESS9 Service tool Running Another service tool is running. Please close all service tools to proceed.
GMPMESS10 NX Backend Running NX Backend is running. Please stop NX completely to proceed.
GMPMESS11 Invalid Position The configured position is not valid. It is not
possible to set the target position.
GMPMESS12 No connection to modality There is no connection with the modality.
Autopositions cannot be verified.
GMPMESS13 Invalid File The selected file is not a valid Autopositions file. Please select a valid file.
GMPMESS14 Invalid Autopositions The configuration contains invalid autopositions.
Activation is not possible.
GMPMESS15 Travel Limit warning Travel limit reached. Movement was stopped.
GMPMESS16 Tube not pointing towards detector The tube is not pointing towards the detector. Please adjust tube.
GMPMESS17 Automatic motion warning System encountered a problem during automatic Please move system manually to desired
motion. position.
GMPMESS18 Invalid Position Invalid parameters for automatic motion. Please call sevice if the problem persists.
GMPMESS19 Position Reached Target Position Reached
GMPMESS21 Movement not enabled Movement is not enabled.
GMPMESS23 FLFS ROI Limit FLFS ROI Limit Reached
GMPMESS24 Ceiling Tube not in Parking Position The ceiling tube is not in parking position, table
movement is not possible.
GMPMESS25 [DTS] Move to initial position Please move to the initial position
GMPMESS26 [DTS] Wrong Bucky Position Bucky is in an invalid position to start the DTS Please reposition the bucky
sequence.
GMPMESS28 Grid Warning A grid is required. Please insert grid.
GMPMESS29 Grid Warning No grid is required. Please remove grid.
GMPMESS30 Grid Warning SID is not suited for the inserted grid.
GMPMESS31 Grid Error Please check if grid is correctly inserted.
GMPMESS32 Grid Error System is unable to determine the grid type. Please check the grid. Contact service if the
problem persists.

Edition 1, Revision 2
11-2020 Livelink ID: 68274899 GMPMESS
IRIMESS
(Error messages from Innolux panels (DR14e, DR17e))

Status 2020-10-29
Error Code Summary Message Instruction
IRIMESS2 License Error License error: license <TECHNICAL- Please check the license file.
LICENSENAME> for
<COMPONENT/APPLICATION> not found.
IRIMESS3 <Software name + version> Setup A matching license has been found, but there is a
license version mismatch./nThis installation will be
aborted.
IRIMESS4 <Software name + version> Setup There are no matching licenses found. This
installation will be aborted.
IRIMESS5 <Software name + version> Setup You do not have enough privileges to install this
product./n
Administrator privileges are needed./n
This installation will be canceled.
IRIMESS6 SW Conflict Can't start <Tool Name> while other NX software STOP NX completely before launching <Tool
is running. Name>.
IRIMESS7 Software Conflict Cannot start IRI Panel Installation Tool while
another instance of IRI Panel Installation Tool is
running.
IRIMESS8 FTP Server Error The FTP Server is not active or configured Check if:\nThe FTP Server is running\nUser "root"
incorrectly. has access to the server\nUser "dr-id1200" has
access to the server.
IRIMESS9 NIC Error The Network Interface Card (NIC) is incorrect Check if:\nThe name of the NIC is "Eth1"\nThe
configured. NIC has only one IP address (192.168.xxx.10).
IRIMESS10 Firewall Error The Windows Firewall is incorrect configurerd Check if:\The Windows Firewall is off, or\nFTP
connection is allowed by the firewall.
IRIMESS11 No Administrator Rights <Tool name> is only available for Administrator
accounts.
IRIMESS12 Select path with MSD Please select path with Machine Specific Data of
the panel (eg. CD).\n(Do not connect the panel
yet.)
IRIMESS13 Select path with MSD: Selected path is not Selected path is not correct. Please select path with Machine Specific Data of
correct. the panel (eg. CD).\n(Do not connect the panel
yet.)
IRIMESS14 Select path with MSD: Panel already installed The panel <panel ID> is already installed. Please select path with Machine Specific Data of
the panel (eg. CD).\n(Do not connect the panel
yet.)
IRIMESS15 Registering panel Registering panel <panel ID>...
IRIMESS16 Please reboot panel Please reboot panel <panel ID> and connect it via
cable.
IRIMESS18 Checking IP addresses Checking the IP-addresses of panel <Panel ID>.

IRIMESS19 Checking IP addresses: IP conflict Checking the IP-addresses of panel <Panel ID>. The IP-address <IP address of panel> of panel
<to be installed Panel ID> is already used for
panel <other panel ID>.\nClick next to change IP-
address of panel <to be installed Panel ID>.
IRIMESS20 Installing MSD Installing Machine Specific Data...\n(This takes <Instruction>
several minutes).
IRIMESS21 No other panels connected Looking for connected panels...\n(Only panel Please disable or shut down the following panel
<panel ID> should be connected.) (s): <panel ID>.
IRIMESS22 No other panels connected: other panel Looking for connected panels...\n(Only panel Please disconnect or shut down following
connected. <panel ID> should be connected.) panel(s): <List of IDs of wrongly connected
panels>.
IRIMESS24 Updating configuration files Updating configuration files...
IRIMESS25 Installation Canceled: Disconnect panel for Installation of panel <panel ID>
rollback canceled.\nPlease disconnect the
panel.\nRollback of the panel <panel ID>...
IRIMESS26 Installation Canceled: Rollback completed Installation of panel <panel ID>
canceled.\nRollback completed.\nClose the
window.
IRIMESS27 Panel installation completed Panel <panel ID> is successfully added to the
configuration.\nPlease re-boot the panel and
close the window.
IRIMESS28 Unsaved settings Please save the changes in the IRI Panel
Installation Tool before adding a panel.
IRIMESS29 Remove panel? Are you sure that you want to remove panel Please shut down the panel...
<panel ID>?
IRIMESS30 Connect all panels Please connect all installed panels via wire to
apply changes.
IRIMESS31 Save Changes Do you want to save changes?
IRIMESS33 Ping result Panel <name> (<IP Address>) is not connected.

IRIMESS35 Retrieving InfoCounterReportFile Not possible to retrieve Info Counter Report. Please connect panel and try again.
IRIMESS38 Detector <detector ID> Detector <Detector Name> initialized and ready
for use.
IRIMESS39 Temperature Warning DR Detector <detector name>'s temperature is
high (&gt; <warning high temperature> degrees
Celsius).

Edition 1, Revision 2
11-2020 Livelink ID: 68274899 IRIMESS
IRIMESS
(Error messages from Innolux panels (DR14e, DR17e))

Status 2020-10-29
Error Code Summary Message Instruction
IRIMESS40 Temperature Warning DR Detector <detector name>'s temperature is
low (&lt; <warning low temperature> degrees
Celsius).
IRIMESS41 Temperature Error DR Detector <detector name>'s temperature is The DR Detector is shutted down.
too high (&gt; <warning temperature> degrees
Celsius).
IRIMESS42 Temperature Error DR Detector <detector name>'s temperature is The DR Detector is shutted down.
too low (&lt; <warning low temperature> degrees
Celsius).
IRIMESS43 Shock Error DR detector <Panel Name> received a shock. Please be aware that this might impact the image
quality. If this occurs a re-calibration might be
needed. If the problem persists, please contact
your Service Technician.
IRIMESS45 Crack Error A crack is detected in DR detector <Panel Please be aware that this might impact the image
Name>. quality. If this occurs a re-calibration might be
needed. If the problem persists, please contact
your Service Technician.
IRIMESS46 Check beam path Please check that there are no unwanted objects
in the beam path.
IRIMESS47 Calibration: Calibration Mode started Starting Calibration Mode: <Calibration Mode>.
IRIMESS48 Calibration: Dose message Set dose to <dose>µGy, kV: <kV>, mAs: <mAs>,
maximum 200 ms\nmAs value (SID =120) is
indicative and can vary to reach the indicated
dose.\nThe average SID for normal operation
should be set.\n\nAfter you set the dose, expose
the panel.
IRIMESS49 Calibration: Exposure made Exposure is made...
IRIMESS50 Calibration: Take exposure Take exposure <exposure number> of <total
number of exposure during this calibration step>
for <Calibration step>.
IRIMESS51 Calibration: Updating Calibration files Updating calibration files for <Calibration Step>...

IRIMESS52 Calibration: Detector Connection DR Detector Connection <Detector name> Check if the DR Detector is powered on. If not,
Failed. please power on the DR detector.\nCheck if the
DR detector is connected to this workstation.\nIf
not, link the DR detector to this workstation (using
a cable).
IRIMESS53 Calibration: Average pixel value too low Calibration step failed. \nAverage pixel value of Use higher dose.
<aveage pixel value>. \nTarget average pixel
value is <target average pixel value>.
IRIMESS54 Calibration: Average pixel value too high Calibration step failed. \nAverage pixel value of Use lower dose.
<average pixel value>. \nTarget average pixel
value is <target average pixel value>.
IRIMESS55 Calibration: Uniformity Test failed. Uniformity Test failed. Check collimation: The panel should be fully
covered.\nPlease check beam path for objects.
IRIMESS56 Calibration: Calibration failed - generic Calibration step failed with reason: <reason>. Press continue to restart calibration mode
<calibration mode>.
IRIMESS58 Recovery Image Connection loss with detector <Panel Restore connection.
Name>.\nAttention: If you select another
thumbnail, the image will be lost.
IRIMESS59 DR Detector Error DR Detector <Panel ID> encountered a problem: Please call service if the problem persists.
<Error ID>\nThe DR Detector will be shut down.

IRIMESS60 Configuration Error A configuration error occured: <ErrorInfo> Please call service if the problem persists.
IRIMESS61 Hardware Error The Hardware encountered a problem: <ErrorID>. The system will shut down. Please call service if
the problem persists.
IRIMESS62 DR Detector Error DR Detector <panel ID> encountered a problem: Please re-boot the DR Detector. Please call
<ErrorID>. service if the problem persists.
IRIMESS63 Software Error The DR Detector Software encountered a Please reboot the Software. Please call service if
problem: <ErrorID>. the problem persists.
IRIMESS64 Software Error DR Detector Software encountered a problem: Please re-boot the Software. Please perform a
<ErrorID>. calibration if the problem persists.
IRIMESS65 DR Detector Connection Error The connection with the DR Detector <panel ID> Please make sure that the DR Detector <panel
cannot be established. ID> is powered on and connected to the system.
Please call service if the problem persists.

IRIMESS66 Correction Data Load Error DR Detector Software encountered a correction Please perform a calibration if the problem
data load error: <ErrorID>. persists or call service if the problem persists.

Edition 1, Revision 2
11-2020 Livelink ID: 68274899 IRIMESS
IRIMESS
(Error messages from Innolux panels (DR14e, DR17e))

Status 2020-10-29
Error Code Summary Message Instruction
IRIMESS67 Configuration Parameter Error A parameter configuration error occured: Please call service if the problem persists.
<ErrorID>.
IRIMESS68 Temperature Warning The battery in DR Detector <detector name>'s Check the ambient temperature.\nWait longer
temperature is high (&gt; <warning high between exposures to lower the temperature in
temperature> degrees Celsius).\nImage quality the DR Detector.
might be impacted.
IRIMESS69 Temperature Warning The battery in DR Detector <detector name>'s Check the ambient temperature.
temperature is low (&lt; <warning low
temperature> degrees Celsius).\nImage quality
might be impacted.
IRIMESS70 Temperature Error The battery temperature in DR Detector <detector Remove the battery from the DR Detector\n
name> is too high (&gt; <error high temperature> If the removed battery remains hot, replace it with
degrees Celsius).\nImage quality might be a new one, as the battery is defective.\nIf the
impacted. temperature of the removed battery lowers, check
the cables in the DR Detector.
IRIMESS71 Temperature Error The battery temperature in DR Detector <detector Check the ambient temperature.
name> is too low (&lt; <error low temperature>
degrees Celsius).\nImage quality might be
impacted.
IRIMESS73 Connection Error Please connect panel <panel ID> via cable.
IRIMESS74 Firmware Status Panel <Panel ID> is already up-to-date.\nNo need
for upgrade.
IRIMESS75 Firmware Status Panel <panel ID> has a firmware
mismatch.\nWould you like to proceed?
IRIMESS76 Firmware Status The upgrade will start after this final confirmation. Make sure the panel <panel ID> has enough
battery or is powered via cable.\nSelect OK to
start.
IRIMESS77 Preparing Upgrade Firmware upgrade starting...\nPlease do not
disconnect the panel!
IRIMESS78 Doing Upgrade Upgrading...\nPlease do not disconnect the panel!

IRIMESS79 Finalizing Upgrade Firmware upgrade finalizing...\nPlease do not


disconnect the panel!
IRIMESS80 Firmware Upgrade Error Firmware upgrade interrupted...\nPlease re-do the
firmware upgrade.
IRIMESS81 Configuration Error A configuration mismatch is detected. Please make sure that the Trigger Mode is
configured the same for each IRI DR Detector in
the NX Config Tool.
IRIMESS82 Panel connected wireless Panel <connected panel ID> is connected Please connect via cable.
wireless instead of via cable.
IRIMESS83 Re-boot all panels Please re-boot all panels.
IRIMESS84 Calibration: Remove filter Remove the filter.
IRIMESS85 (Re-)Calibrate panel By adding panel <panel ID> to this configuration, Please (re-)calibrate panel <panel ID> on the
the calibration files are reset to factory default. main NX, after completing installation.

IRIMESS86 Calibration: Preparing for next calibration step. Preparing for the next calibration step.\nThis
might take several minutes.
IRIMESS87 Bad Wifi connection Warning: the quality of the Wifi connection with
the detector is not good.\n
It is strongly advised to improve the Wifi or the
calibration sequence may fail.
IRIMESS88 Test result of image transfer for wireless The average transfer time is <average ms> ms,
connection (Optimal wireless transfer time is less than
<optimal ms> ms).
IRIMESS89 Test result of image transfer for wired connection The average transfer time is <number of ms> ms.

IRIMESS90 Checking connection quality Checking connection quality, please wait ...
IRIMESS91 PC clock tampered The clock of the PC has been turned back with Please restore PC clock to correct date and time
more than 24 hours. in order to work with the EPS panel.
IRIMESS92 Serial number not found on dongle The serial number of the EPS panel <EPS panel Check your dongle configuration on the Agfa EPS
serial number> is not found on the dongle. portal and correct it. If problem persists, call
service.
IRIMESS93 EPS dongle not attached to PC The dongle containing the EPS license is not Please attach the dongle.
attached to the NX workstation.
IRIMESS94 EPS license is expired The EPS license for panel <panel serial number> Renew the EPS license on the Agfa EPS portal.
has expired
IRIMESS96 EPS license about to expire The EPS license for panel <panel serial number> Please renew the EPS license on the Agfa EPS
will expire on <date> at <time>. portal.
IRIMESS97 No combination with EPS panel allowed EPS-panel cannot be combined with other panels. Check configuration. Make sure only 1 EPS-panel
is configured.
IRIMESS99 Exposure time during calibration It is not allowed to expose longer than Please change the exposure time.
200ms.\nThe <Calibration type> shall be
resumed.

Edition 1, Revision 2
11-2020 Livelink ID: 68274899 IRIMESS
IRIMESS
(Error messages from Innolux panels (DR14e, DR17e))

Status 2020-10-29
Error Code Summary Message Instruction
IRIMESS101 Recovery Image The image did not arrive in the expected time Attention: If you select another thumbnail, the
frame.(probally due to bad wifi) image will be lost.
IRIMESS102 Updating the panel settings Updating the panel settings for panel
<SerialNumber>
IRIMESS103 Panel installation completed Panel {0} is successfully added to the
configuration.
IRIMESS104 Changing IP Address IP Addresses (subnet) shall only be changed Make sure panel <panel ID> has enough battery
when a conflict occurs with the hospital network. or is powered via cable.

IRIMESS105 Changing IP Address with multiple panels Panels with a different subnet are detected. All Make sure panel <panel ID> has enough battery
panels must be configured for the same or is powered via cable.\nChose the proposed IP
subnet.\nProposed New IP Address <IP Address.
Address>.
IRIMESS106 The IP Address of panel <panel ID> is changed The IP Address of panel <panel ID> is changed. Please reboot panel <panel ID> to activate the
new IP Address <IP Address>.
IRIMESS107 Changing IP Address of Network Adapter Eth 1 Please change the IP Address of the Network Select OK to confirm and restart this application.
Adapter Eth1 to <Selected New IP
Address>\nThe Network and Sharing Center has
started.
IRIMESS108 Upgrade confirmation Please, reboot panel <panel ID>.
IRIMESS109 Configuration Error Power Options are not correctly Please check Power Options configuration. If the
configured.\nSystem may never go to sleep. problem persists, call service.
IRIMESS110 Performing Offset Calibration An offset calibration is running for panel <panel
name>. This might take a few seconds.

Edition 1, Revision 2
11-2020 Livelink ID: 68274899 IRIMESS
MESS-AllProjects
(Error messages from NX, DX-D 300, DX-D 600, all Sedecal Products in general)

Status 2020-10-29
Error Code Summary Message Instruction
MESS1 Invalid parameter range Invalid parameter range
MESS1 No queues have been defined yet. No queues have been defined yet.
MESS10 Generator Communication Failure Generator Communication Failure
MESS10 Are you sure you want to clear the selected Are you sure you want to clear the selected
job(s)? job(s)?
MESS100 [FLFS] FLFS sequence aborted FLFS sequence aborted
MESS101 [FLFS] Please insert panel in portrait in bucky Please insert the panel in portrait or rotate the
bucky.
MESS102 Error, name not unique. Error, name not unique.
MESS102 [FLFS] Move to initial position Press Move button until initial position has been
reached.
MESS103 The installed digitizer device type does not belong The installed digitizer device type does not belong
to the same family as the already configured to the same family as the already configured
digitizers. Please select a dig digitizers. Please select a dig
MESS104 Invalid configuration Error initializing devices, please check your NX
configuration
MESS104 Info: Fast ID has been switched on. The write Info: Fast ID has been switched on. The write
destination... destination...
MESS105 Error, the connection with the medical printer Error, the connection with the medical printer
device could not be established. device could not be established.
MESS105 Collimator in Manual mode Please set the collimator to Automatic mode by Contact service if turning the collimator key does
turning the collimator key. not resolve the issue.
MESS106 Calibration Warning The calibration of DR Detector "<ID>" will expire Please recalibrate the DR Detector.
in <number> days.
MESS106 Are you sure you want to delete the printer profile Are you sure you want to delete the printer profile
and its related settings? and its related settings?
MESS107 Calibration Expired The calibration of DR Detector "<ID>" has Please recalibrate the DR Detector.
expired.
MESS107 Error, the file could not be opened or read Error, the file could not be opened or read
MESS108 File could not be used to import requested device File could not be used to import requested device
model. model.
MESS108 Please insert panel in landscape in bucky Please insert the panel in landscape or rotate the
bucky.
MESS109 Acquisition not possible Acquisition is temporarily not possible "<Error Please wait for the system to recover.
Code>".
MESS109 Error, the connection with the RIS could not be Error, the connection with the RIS could not be
established. established.
MESS11 Workstation Communication Error Workstation Communication Error
MESS110 Post Exposure Parameters <parameters>
MESS111 Max Fluo Duration Reached The fluoroscopy exam has reached its maximum
duration.
MESS112 [DTS] Move to initial position Please move to the initial position
MESS112 Error, invalid move target selected for the Error, invalid move target selected for the
selected Exposure type. This Exam Group selected Exposure type. This Exam Group
already contains an Exposure type with the s already contains an Exposure type with the s
MESS113 [DTS] DTS Sequence completed DTS Sequence completed
MESS113 Do you want to remove the selected Exam Do you want to remove the selected Exam
Group? Group?
MESS114 [DTS] DTS sequence aborted DTS sequence aborted
MESS114 Do you want to remove the selected Exposure Do you want to remove the selected Exposure
type? type?
MESS115 Warm-up warning message Make sure that the collimator blades are fully
closed and that no patient is present in the
room.\nTo avoid radiation on a panel, remove the
panel, turn the tube away from the panel, or cover
the panel with a lead apron.
MESS115 Do you want to remove the selected Exposure Do you want to remove the selected Exposure
Group? Group?
MESS116 Warm-up Procedure completed Warm-up Procedure completed
MESS116 Exposure Group <name of the Exposure Group> Exposure Group <name of the Exposure Group>
has become empty. has become empty.
MESS117 Warm-up Procedure aborted Warm-up Procedure aborted
MESS117 The previous replacements have not been The previous replacements have not been
committed yet. Please confirm or cancel the committed yet. Please confirm or cancel the
previous replacements first. previous replacements first.
MESS118 Abort Workflow Please check panel connection (wifi) or\nPress
abort to end the workflow with image loss.
Reconstruction might not be possible.

Edition 1, Revision 2
11-2020 Livelink ID: 68274899 MESS-AllProjects
MESS-AllProjects
(Error messages from NX, DX-D 300, DX-D 600, all Sedecal Products in general)

Status 2020-10-29
Error Code Summary Message Instruction
MESS118 Error, the file has been modified outside the tool. Error, the file has been modified outside the tool.
Load aborted! Load aborted!
MESS119 Title = Export Acquired Dose RecordsMessage = Make sure to select an accessible location and
The file ... retry the export of Acquired Dose Records
MESS12 Software console is already running. Software console is already running.
MESS12 No valid license found ! Generator connection will No valid license found! Generator connection will
not work not work.
MESS120 The selected Exposure Type is used in the The selected Exposure Type is used in the
layouts of one or more Exam Groups and layouts of one or more Exam Groups and
Exposure Groups. By this operation the layouts Exposure Groups. By this operation the layouts

MESS121 The selected Exposure Group is used in the The selected Exposure Group is used in the
layout of one or more Exam Groups. This layout of one or more Exam Groups. This
operation will affect these layouts, do you w operation will affect these layouts, do you w
MESS122 The selected Exposure Group is also configured The selected Exposure Group is also configured
as the Eme... as the Eme...
MESS123 The selected Exposure Type is used as the The selected Exposure Type is used as the
emergency Exposure Type for one or more emergency Exposure Type for one or more
Digitizers. Removal of this Exposure Type will Digitizers. Removal of this Exposure Type will
MESS124 Some configuration settings have not yet been Some configuration settings have not yet been
saved. Loading a new configuration file will saved. Loading a new configuration file will
dismiss these settings. Do you wish t dismiss these settings. Do you wish t
MESS125 The device type has already been installed on the The device type has already been installed on the
system, do you want to overwrite it ? system, do you want to overwrite it ?
MESS126 Error, the workspace does not contain all Error, the workspace does not contain all
mandatory configuration settings. mandatory configuration settings.
MESS127 Error, Fast-ID is enabled, but there are no Error, Fast-ID is enabled, but there are no
digitizers defined digitizers defined
MESS128 Warning, not all images of the printsheet Warning, not all images of the printsheet
templates in Exposure Group <name> have an templates in Exposure Group <name> have an
Exposure type assigned. Exposure type assigned.
MESS129 Cassette size <size> has not been linked to a film Cassette size <size> has not been linked to a film
size. ... size. ...
MESS13 Shutdown canceled by Workstation Shutdown canceled by Workstation
MESS130 Warning, Exposure group <xxx> contains no Warning, Exposure group <xxx> contains no
Exposure types for Age group <xxx>. Exposure types for Age group <xxx>.
MESS131 Error, both ID-tablet and Fast ID are configured. Error, both ID-tablet and Fast ID are configured.
There may be only one write destination. There may be only one write destination.
MESS132 Error, IDIS workflow cannot be combined with Error, IDIS workflow cannot be combined with
automatic ca... automatic ca...
MESS133 Warning, Exam Group <xxx> is empty for Age Warning, Exam Group <xxx> is empty for Age
group <xxx>. group <xxx>.
MESS134 No default film size available. Please define a No default film size available. Please define a
default f... default f...
MESS135 Film size <film size value> is not linked with a Film size <film size value> is not linked with a
default ... default ...
MESS136 Title = Licenses OverviewMessage = The Title = Licenses OverviewMessage = The
following demo-l... following demo-l...
MESS137 Title = MUSICA 2 Demo LicenseMessage = Your Title = MUSICA 2 Demo LicenseMessage = Your
Musica2 dem... Musica2 dem...
MESS138 Title = X-Ray Device connection errorMessage = Title = X-Ray Device connection errorMessage =
There wa... There wa...
MESS139 The configuration tool has detected that the The configuration tool has detected that the
system is st... system is st...
MESS14 Warning, the actual SID is not suited for the Warning, the actual SID is not suited for the
applied grid. ( xx cm < SID < xx cm) applied grid. ( xx cm < SID < xx cm)
MESS14 There are still active jobs waiting in the print There are still active jobs waiting in the print
queue queue
MESS140 Title = Delete ImagesMessage = The selected Title = Delete ImagesMessage = The selected
images coul... images coul...
MESS141 Title = StartupMessage = Invalid NX license or Title = StartupMessage = Invalid NX license or
wrong do... wrong do...
MESS142 Title = Read CassetteMessage = The cassette Title = Read CassetteMessage = The cassette
could not b... could not b...
MESS143 Title = ID Tablet connectionMessage = A Title = ID Tablet connectionMessage = A
connection with... connection with...
MESS144 Title = Read CassetteMessage = Cassette not Title = Read CassetteMessage = Cassette not
identified ... identified ...
MESS145 Title = Start NXMessage = Incorrect screen Incorrect screen resolution Please contact your system administrator
resolution....
MESS146 Error, The following devices were removed Error, The following devices were removed
archive/printer... archive/printer...
MESS147 Title = Import ImagesMessage = Error reading Title = Import ImagesMessage = Error reading
CD. No ima... CD. No ima...

Edition 1, Revision 2
11-2020 Livelink ID: 68274899 MESS-AllProjects
MESS-AllProjects
(Error messages from NX, DX-D 300, DX-D 600, all Sedecal Products in general)

Status 2020-10-29
Error Code Summary Message Instruction
MESS148 Title = Transfer ImagesMessage = Too many Title = Transfer ImagesMessage = Too many
open exams. T... open exams. T...
MESS149 Title = Digitizer connectionMessage = An Title = Digitizer connectionMessage = An
unlicensed dig... unlicensed dig...
MESS15 Warning, The grid is -not- inserted although not Warning, The grid is -not- inserted although not
requested by the modality workstation. requested by the modality workstation.
MESS152 Title = Message = Access the Add Image dialog Title = Message = Access the Add Image dialog
to resolv... to resolv...
MESS153 Title = Message = No image selected. Please Title = Message = No image selected. Please
select an i... select an i...
MESS154 Title = Message = You can fill out this image cell Title = Message = You can fill out this image cell
as f... as f...
MESS155 Title = Message = Drag and drop an image from Title = Message = Drag and drop an image from
the image... the image...
MESS156 Title = Message = The image is not yet assigned Title = Message = The image is not yet assigned
to a sh... to a sh...
MESS157 Title = Message = The image is not yet assigned Title = Message = The image is not yet assigned
to a sh... to a sh...
MESS158 Title = Message = IDInstruction = Icon = Buttons Title = Message = IDInstruction = Icon = Buttons
... ...
MESS159 Title = Message = No thumbnail defined. Drag an Title = Message = No thumbnail defined. Drag an
image f... image f...
MESS16 Error, The grid is not correctly positioned Error, The grid is not correctly positioned
MESS16 Select on console the DR-detector position with Select on console the DR-detector position with Select on console the DR-detector position with
correct exposure settings correct exposure settings correct exposure settings.
MESS160 Title = Message = Date should be in format Title = Message = Date should be in format
{0}Instruc... {0}Instruc...
MESS161 Title = Message = Incorrect valueInstruction = Title = Message = Incorrect valueInstruction =
Icon... Icon...
MESS162 Title = Export imagesMessage = Invalid folder Title = Export imagesMessage = Invalid folder
selected:... selected:...
MESS163 Title = Export ImagesMessage = Unknown error Title = Export ImagesMessage = Unknown error
occurred d... occurred d...
MESS164 Title = Delete ImagesMessage = System is System is running out of disk space Please contact your system administrator
running out of...
MESS165 Title = NX LicensesMessage = License error: your Title = NX LicensesMessage = License error: your
grace ... grace ...
MESS166 Title = Message = NX is already Title = Message = NX is already
runningInstruction = ... runningInstruction = ...
MESS168 Title = Message = No Digitiser Configured ! NX Title = Message = No Digitiser Configured ! NX
can not ... can not ...
MESS169 Title = Import imagesMessage = No CD Title = Import imagesMessage = No CD
availableInstruc... availableInstruc...
MESS17 warning - not possible to set "attribute" to "value" warning - not possible to set "attribute" to "value"

MESS17 DR-detector successfully calibrated. DR-detector successfully calibrated.


MESS17 Are you sure you want to remove the selected Are you sure you want to remove the selected
printer profile? printer profile?
MESS171 Warning, all entries will be removed from the Warning, all entries will be removed from the
current exa... current exa...
MESS172 Title = Fast IDMessage = No Fast ID Connection Title = Fast IDMessage = No Fast ID Connection
with Dig... with Dig...
MESS173 Title = Print imageMessage = The sheet(s) cannot Title = Print imageMessage = The sheet(s) cannot
be pri... be pri...
MESS174 Title = NX LicensesMessage = The paediatric Title = NX LicensesMessage = The paediatric
license has... license has...
MESS175 Title = Receive ImageMessage = An incoming Title = Receive ImageMessage = An incoming
DICOM Object... DICOM Object...
MESS176 When a role is deleted, all users assigned to this When a role is deleted, all users assigned to this
role l... role l...
MESS177 The last defined user is about to be deleted. Do The last defined user is about to be deleted. Do
you want... you want...
MESS178 Error, archive automation destination does not Error, archive automation destination does not
exist exist
MESS179 Error, archive storage commit is not allowed Error, archive storage commit is not allowed
MESS18 Door Open error. The X-ray room door is open Door Open error. The X-ray room door is open
when the X-ray equipment is in use. when the X-ray equipment is in use.
MESS18 Wrong calibration can cause non diagnostic Wrong calibration can cause non diagnostic
image quality. image quality.
MESS180 Error, emergency exposure of digitiser has no Error, emergency exposure of digitiser has no
protocolcode protocolcode
MESS181 Error, the emergency settings of digitizer has no Error, the emergency settings of digitizer has no
valid e... valid e...
MESS182 Error, the ExamGroupView of the emergency Error, the ExamGroupView of the emergency
settings does n... settings does n...
MESS183 Error, the ExposureGroup of the emergency Error, the ExposureGroup of the emergency
settings does n... settings does n...

Edition 1, Revision 2
11-2020 Livelink ID: 68274899 MESS-AllProjects
MESS-AllProjects
(Error messages from NX, DX-D 300, DX-D 600, all Sedecal Products in general)

Status 2020-10-29
Error Code Summary Message Instruction
MESS184 Error, There must be at least one exposuretype Error, There must be at least one exposuretype
OR exposur... OR exposur...
MESS185 Error, the printerprofile of the exposuretype in Error, the printerprofile of the exposuretype in
examgrou... examgrou...
MESS186 Error, the printerprofile of the exposuretype in Error, the printerprofile of the exposuretype in
examgrou... examgrou...
MESS187 Error, the exposuretype in examgroup of Error, the exposuretype in examgroup of
agegroup has an u... agegroup has an u...
MESS188 Error, the emergency has no protocolcode Error, the emergency has no protocolcode
MESS189 Error, the device can not be deleted, because Error, the device can not be deleted, because
there are s... there are s...
MESS19 No Communication No Communication
MESS19 All archived images must be manually archived All archived images must be manually archived
again to see them appear under the new again to see them appear under the new
accession number in the PACS accession number in the PACS
MESS190 Error, there are no speedclasses selected; at Error, there are no speedclasses selected; at
least one m... least one m...
MESS191 Error, there are no digitisers defined; at least one Error, there are no digitisers defined; at least one
must... must...
MESS192 Error, there are too many Stratus digitisers; only Error, there are too many Stratus digitisers; only
one al... one al...
MESS193 Error, there must be at least one exam tree Error, there must be at least one exam tree
defined with ... defined with ...
MESS194 Error, FastID preview AE title and port must be Error, FastID preview AE title and port must be
specified specified
MESS195 Error, ID stationname, processing stationname Error, ID stationname, processing stationname
and portnum... and portnum...
MESS196 Error, the emergency examgroup is no valid Error, the emergency examgroup is no valid
examgroup examgroup
MESS197 Error, the emergency exposuregroup is no valid Error, the emergency exposuregroup is no valid
exposuregroup exposuregroup
MESS198 Error, a printerprofile in exposuregroup contains a Error, a printerprofile in exposuregroup contains a
not e... not e...
MESS199 Error, the secure DICOM connection is enabled, Error, the secure DICOM connection is enabled,
while the ... while the ...
MESS2 FLFS Sequence Completed FLFS Sequence Completed
MESS20 No AEC configured in this generator. Technique No AEC configured in this generator. Technique
rejected rejected
MESS20 Invalid parameter range Invalid parameter range
MESS200 Error, the local ports in the NX workstation have Error, the local ports in the NX workstation have
an overlap an overlap
MESS201 Error, a portnumber is zero Error, a portnumber is zero
MESS202 Error, FastID is disabled, so there may be no Error, FastID is disabled, so there may be no
digitisers ... digitisers ...
MESS203 Error, FastID is enabled, so there must be (only) Error, FastID is enabled, so there must be (only)
one dig... one dig...
MESS204 Error, FastID workstation is disabled, so there Error, FastID workstation is disabled, so there
must be a... must be a...
MESS205 Error, FastID is enabled, but too many digitisers Error, FastID is enabled, but too many digitisers
have Fa... have Fa...
MESS206 Title = Musica PreProcessing failedMessage = Title = Musica PreProcessing failedMessage =
Musica Pre... Musica Pre...
MESS207 Title = Critical problemMessage = The NX Server Title = Critical problemMessage = The NX Server
has enc... has enc...
MESS208 Title = Musica PreProcessing failedMessage = Title = Musica PreProcessing failedMessage =
The system... The system...
MESS21 Required value is missing Required value is missing.
MESS21 Energy saving behaviour Energy saving behaviour
MESS21 FLFS Sequence Completed FLFS Sequence Completed
MESS21 Title = Message = This image has already been Title = Message = This image has already been
sent to a... sent to a...
MESS210 Not all devices could be deleted since some Not all devices could be deleted since some
devices still... devices still...
MESS211 An invalid archive SOP Class has been selected An invalid archive SOP Class has been selected
for the de... for the de...
MESS212 The modality settings have been changed. The modality settings have been changed.
Images that have... Images that have...
MESS213 Prior Viewing: The View Priors configuration is Prior Viewing: The View Priors configuration is
incorrect incorrect
MESS214 Image not available. No valid pixel file found. Image not available. No valid pixel file found.
MESS215 The selected examination has been closed and is The selected examination has been closed and is
not available anymore on this system. not available anymore on this system.

Edition 1, Revision 2
11-2020 Livelink ID: 68274899 MESS-AllProjects
MESS-AllProjects
(Error messages from NX, DX-D 300, DX-D 600, all Sedecal Products in general)

Status 2020-10-29
Error Code Summary Message Instruction
MESS216 Failed to send generator parameters: Model Failed to send generator parameters: Model
Mismatch Mismatch
MESS217 Failed to send generator parameters: Xray Device Failed to send generator parameters: Xray Device
Interface Startup Failure Interface Startup Failure
MESS218 Some expsore types do not have valid Musica2/3 Some expsore types do not have valid Musica2/3
processing parameters. These exposure types processing parameters. These exposure types
will not be available in NX will not be available in NX
MESS219 Mammo XRG parameters Received Mammo XRG parameters Received
MESS22 The name specified has already been used. The name specified has already been used. Please stipulate another name.
Please stipulate another name.
MESS22 Empty battery - shutting down system Empty battery - shutting down system
MESS22 FLFS Sequence Aborted FLFS Sequence Aborted
MESS22 Title = Close ExamMessage = When closing this Title = Close ExamMessage = When closing this
exam, any... exam, any...
MESS220 XRG connection error XRG connection error
MESS221 Waiting for Data Waiting for Data
MESS222 An unexpected error has occured while loading An unexpected error has occured while loading
the configuration interface the configuration interface
MESS223 Value can contain only <x> characters Value can contain only <x> characters
MESS224 Value must be between <x> and <y> Value must be between <x> and <y>
MESS225 Date can not be in the future Date can not be in the future
MESS226 Image Scanning Failure Image Scanning Failure
MESS227 Image Erasure Failure Image Erasure Failure
MESS228 21 exam groups have the 'Show exam in UI' flag 21 exam groups have the 'Show exam in UI' flag
set. set.
MESS229 30 exposures have the 'Show exposure in UI' flag 30 exposures have the 'Show exposure in UI' flag
set. set.
MESS23 Spatial charge. Please, decrease mA value Spatial charge. Please, decrease mA value
MESS23 Value out of range Value <value> is out of range [min,max].
MESS23 Settings Accepted Settings Accepted
MESS230 Title = Export Repeat/Reject RecordsMessage = The Repeat/Reject Records have been exported
The Repea... successfully
MESS231 Person Name can only be composed out of Person Name can only be composed out of
different elements different elements
MESS232 The CD drive is busy. Retry when the CD drive is The CD drive is busy. Retry when the CD drive is
no longe... no longe...
MESS233 Drag and drop an image from the image overview Drag and drop an image from the image overview
into the i... into the i...
MESS234 The digitizer went off-line or in error state. Please The digitizer went off-line or in error state. Please
verify whether the digitiser is properly connected verify whether the digitiser is properly connected
and powered on. and powered on.
MESS236 The digitizer is back on-line. Please re-identify the The digitizer is back on-line. Please re-identify the
cassette if it is not scanned automatically. cassette if it is not scanned automatically.
MESS237 The digitizer went off-line during a scan cycle. The digitizer went off-line during a scan cycle.
This is a critical error. If the digitiser was switched This is a critical error. If the digitiser was switched
off the image may not b off the image may not b
MESS24 Spatial charge. Please, increase kV value or Spatial charge. Please, increase kV value or
exposure time exposure time
MESS24 Invalid value Invalid value: <value>
MESS24 Collimator failed Collimator failed FLFS sequence aborted
MESS240 Import Images: The import is aborted. Unable to Import Images: The import is aborted. Unable to
import import
MESS241 Unable to import the tecnical image. Please close Unable to import the tecnical image. Please close
some op... some op...
MESS242 Error reading drive X Error reading drive X
MESS243 unable to write a (mammo / not mammo) unable to write a (mammo / not mammo)
exposure on a (not ... exposure on a (not ...
MESS244 Impossible to send a "non mammo" image in Impossible to send a "non mammo" image in
DICOM MG format DICOM MG format
MESS246 Source and target session must not be the same Source and target session must not be the same

MESS247 This view modifier code is not allowed for this This view modifier code is not allowed for this
exposure ... exposure ...
MESS248 The template exposure is not present anymore in The template exposure is not present anymore in
the exam tree the exam tree
MESS249 Error, this view code is not allowed for mammo Error, this view code is not allowed for mammo
exposures exposures

Edition 1, Revision 2
11-2020 Livelink ID: 68274899 MESS-AllProjects
MESS-AllProjects
(Error messages from NX, DX-D 300, DX-D 600, all Sedecal Products in general)

Status 2020-10-29
Error Code Summary Message Instruction
MESS25 No DR-detectors configured, No DR-detectors configured.
MESS25 FLFS sequence completed FLFS sequence completed
MESS25 Please insert the panel in portrait in the bucky Please insert the panel in portrait in the bucky
MESS25 Title = Cassette handlingMessage = The Title = Cassette handlingMessage = The
requested action... requested action...
MESS250 Warning, The flag "send with same SOP instance Warning, The flag "send with same SOP instance
UID" is no... UID" is no...
MESS251 The selected SPS can not be retrieved from the The selected SPS can not be retrieved from the
RIS anymore RIS anymore
MESS252 License error License error
MESS253 License error: your grace period has expired. License error: your grace period has expired.
Please cont... Please cont...
MESS254 All configuration data will be lost after the All configuration data will be lost after the
installatio... installatio...
MESS255 Not for diagnosis Not for diagnosis
MESS256 Are you sure you want to restore the image and Are you sure you want to restore the image and
sheet text... sheet text...
MESS257 Maximum number of exposures reached (2000) Maximum number of exposures reached (2000)
MESS258 Error, check the exposure settings, modality Error, check the exposure settings, modality
settings and... settings and...
MESS259 No valid NX license. NX will be closed. No valid NX license. NX will be closed.
MESS26 Generator parameter <parameter name> out of Generator parameter <parameter name> out of
range range
MESS26 Receptor has not been replaced or inserted Receptor has not been replaced or inserted
MESS26 Title = Cassette handlingMessage = Unable to Title = Cassette handlingMessage = Unable to
write cass... write cass...
MESS260 Maximum number of protocol codes (2000) Maximum number of protocol codes (2000)
reached. The file... reached. The file...
MESS261 Maximum number of protocol codes reached Maximum number of protocol codes reached
(2000). (2000).
MESS262 Select an image from the image overview and Select an image from the image overview and
click here to insert click here to insert
MESS263 Edit exposure: Unable to save the data. Probably Edit exposure: Unable to save the data. Probably
caused by external activities, e.g. ... caused by external activities, e.g. ...
MESS264 The connection with room "room name" is lost. The connection with room "room name" is lost.
The exam wi... The exam wi...
MESS266 Connection Error Connection Error
MESS268 Warning, Platinum Musica2/3 packages Warning, Platinum Musica2/3 packages
(Skeleton, Abdomen, Chest) must be used in all (Skeleton, Abdomen, Chest) must be used in all
available age groups available age groups
MESS27 Room not ready Room not ready
MESS27 Please insert the panel in landscape in the bucky Please insert the panel in landscape in the bucky

MESS270 Error, the file version is more recent then the Error, the file version is more recent then the
version of the Config Tool version of the Config Tool
MESS271 All configuration and image data will be lost after All configuration and image data will be lost after
the i... the i...
MESS272 No NX workstations configured ! NX can not be No NX workstations configured ! NX can not be
started. Pl... started. Pl...
MESS273 Warning, this procotol code is larger than 16 Warning, this procotol code is larger than 16
characters.... characters....
MESS274 There are x assigned protocol codes larger than There are x assigned protocol codes larger than
16 charac... 16 charac...
MESS275 Not supported DICOM protocol codes detected : Not supported DICOM protocol codes detected :
number of c... number of c...
MESS276 Not all changes are applied on the partial Not all changes are applied on the partial
exposures : ex... exposures : ex...
MESS277 Exposure type "xx" - film size "Default" can not be Exposure type "xx" - film size "Default" can not be
set f... set f...
MESS278 The requested action has not been performed. The requested action has not been performed.
MESS279 Nx has encountered an internal error Nx has encountered an internal error
MESS28 Acquisition failed Acquisition failed
MESS28 Title = Close ExamMessage = Some print sheets Title = Close ExamMessage = Some print sheets
are not c... are not c...
MESS280 Supporting services not yet ready Supporting services not yet ready
MESS281 Target exam group equals the current exam Target exam group equals the current exam
group group

Edition 1, Revision 2
11-2020 Livelink ID: 68274899 MESS-AllProjects
MESS-AllProjects
(Error messages from NX, DX-D 300, DX-D 600, all Sedecal Products in general)

Status 2020-10-29
Error Code Summary Message Instruction
MESS282 Error, missing bitmap file for bitmap cell on sheet Error, missing bitmap file for bitmap cell on sheet
text box : "file name" text box : "file name"
MESS283 The inserted cassette could not be identified. The inserted cassette could not be identified.
Please identify the cassette manually in Please identify the cassette manually in
examination view. examination view.
MESS284 wrong ID tablet version number wrong ID tablet version number
MESS285 It is impossible to upgrade from an "type" to an It is impossible to upgrade from an "type" to an
"type". ... "type". ...
MESS286 This will stop NX. Any ongoing operations will be This will stop NX. Any ongoing operations will be
aborted... aborted...
MESS287 Service warning. Enhanced digitiser service mode Service warning. Enhanced digitiser service mode
is currently switched on. As soon as you're done is currently switched on. As soon as you're done
with your analysis, switch off with your analysis, switch off
MESS288 Service warning. Enhanced archive service mode Service warning. Enhanced archive service mode
is currently switched on. As soon as you're done is currently switched on. As soon as you're done
with your analysis, switch off t with your analysis, switch off t
MESS289 Digitizer CPF has been uploaded successfully. Digitizer CPF has been uploaded successfully.
Please activate the CPF on the digitizer to Please activate the CPF on the digitizer to
complete the configuration. complete the configuration.
MESS29 NX backend is running. NX backend is running. Stop NX completely to proceed.
MESS29 Modality positioning failure Modality positioning failure
MESS290 An error has occured during the download of the An error has occured during the download of the
CPF. Please restart the procedure. CPF. Please restart the procedure.
MESS291 An error has occured during the upload of the An error has occured during the upload of the
CPF. Please restart the procedure. CPF. Please restart the procedure.
MESS292 Digitizer CPF is being modified by another NX Digitizer CPF is being modified by another NX
workstation. workstation.
MESS293 Identical AE Title Identical AE Title
MESS294 Identical AE Title, Different IP Address Identical AE Title, Different IP Address
MESS295 An error occured while starting the external An error occured while starting the external
application. Please contact your system application. Please contact your system
administrator. administrator.
MESS296 Empty name is not allowed Empty name is not allowed
MESS297 Drag an image from the image overview and drop Drag an image from the image overview and drop
it here or select an image from the image it here or select an image from the image
overview and click here to insert overview and click here to insert
MESS298 Not enough disk space available on configured Not enough disk space available on configured
location location
MESS299 The selected file is not a valid NX Repeat/Reject The selected file is not a valid NX Repeat/Reject
Statistics XML file Statistics XML file
MESS3 FLFS Sequence Aborted FLFS Sequence Aborted
MESS30 FLFS sequence aborted FLFS sequence aborted
MESS30 Title = Print imageMessage = The print sheet is Title = Print imageMessage = The print sheet is
not com... not com...
MESS30 Set the collimator settings manually before Set the collimator settings manually before Make sure the table top is properly centered
exposing DR Detector: exposed area = 46 x 46 exposing DR Detector: exposed area = 46 x 46 allowing full coverage of the detectors surface.
cm, no pre-filter cm, no pre-filter.
MESS300 Export Automation buffer almost full Export Automation buffer almost full
MESS301 Export Automation buffer is full. The closed exam Export Automation buffer is full. The closed exam
will not be copied to the Export Automation buffer. will not be copied to the Export Automation buffer.

MESS302 Conflicting regional settings detected. Conflicting regional settings detected.


MESS305 System startup failure: The labeling configuration System startup failure: The labeling configuration
has been tampered with. has been tampered with.
MESS306 System start-up error: The system has not been System start-up error: The system has not been
activated. activated.
MESS307 Error, package license {0} is present but disabled Error, package license {0} is present but disabled

MESS308 Invalid destination has been selected. Invalid destination has been selected.
MESS309 Title = Export Repeat/Reject RecordsMessage = No report created because no Repeat/Reject
No report... Records are available
MESS31 Unable to add detector. Invalid file. Unable to add detector. Invalid file.
MESS31 Parameter values accepted Parameter values accepted
MESS31 Title = Message = A rejected image cannot be Title = Message = A rejected image cannot be
assigned t... assigned t...
MESS310 The system has encountered an unexpected The system has encountered an unexpected
error. error.
MESS311 Error, non default user taste configured for Error, non default user taste configured for
MAMMO packag... MAMMO packag...
MESS312 Error, the musica 2 mammo T2 package is used Error, the musica 2 mammo T2 package is used
in a CR1B c... in a CR1B c...
MESS314 No digitzer license available. The image will be No digitzer license available. The image will be
routed back to the routing station. routed back to the routing station.
MESS315 The digitizer is back online. The digitizer is back online.

Edition 1, Revision 2
11-2020 Livelink ID: 68274899 MESS-AllProjects
MESS-AllProjects
(Error messages from NX, DX-D 300, DX-D 600, all Sedecal Products in general)

Status 2020-10-29
Error Code Summary Message Instruction
MESS316 Warning: In normal working conditions, it is not Warning: In normal working conditions, it is not
advisabl... advisabl...
MESS318 Warning : Your hardware configuration does not Warning : Your hardware configuration does not
meet the s... meet the s...
MESS319 Error, an incompatible dose statistic parameter Error, an incompatible dose statistic parameter
has been selected. has been selected.
MESS32 Sending exposure parameters failed Sending exposure parameters failed: Please select thumbnail again.
<generator/collimator> error.
MESS32 Positioner parameter <parameter name> value Positioner parameter <parameter name> value
out of range out of range
MESS321 The digitizer is offline. Please make sure the The digitizer is offline. Please make sure the
digitiser is turned on and started up. digitiser is turned on and started up.
MESS322 Another instance is already running Another instance is already running
MESS323 Invalid character 'x' used Invalid character 'x' used
MESS324 Warning: there are system diagnosis exposure Warning: there are system diagnosis exposure
routed to MU... routed to MU...
MESS325 The digitizer is occupied with an ongoing image The digitizer is occupied with an ongoing image
acquisition. Please wait until the digitizer is ready acquisition. Please wait until the digitizer is ready
and try again. and try again.
MESS326 The digitizer is off-line or in error state. Please The digitizer is off-line or in error state. Please
make sure the digitizer is switched on and started make sure the digitizer is switched on and started
up. up.
MESS327 Connection with the digitizer has been broken. Connection with the digitizer has been broken.
Please contact Service. Please restart NX. Please contact Service. Please restart NX.
MESS328 Unspecified error while accessing the digitizer. Unspecified error while accessing the digitizer.
Please contact Service. Please restart NX. Please contact Service. Please restart NX.
MESS329 The digitizer still contains a cassette. Please The digitizer still contains a cassette. Please
remove it and try again. remove it and try again.
MESS33 Millenium Shutdown Millenium Shutdown
MESS33 Connection speed The connection speed to DR Detector "<ID>" is
too low.
MESS330 Unable to send system diagnosis images to Unable to send system diagnosis images to
MUSPAX archive MUSPAX archive
MESS331 Warning! The inserted medium (DVD+RW) may Warning! The inserted medium (DVD+RW) may Please make sure the medium is empty before
contain data already. contain data. continuing
MESS332 File version is not supported anymore! File version is not supported anymore!
MESS333 For CPF generation: Error, emergency exposure For CPF generation: Error, emergency exposure
of digitize... of digitize...
MESS334 Another logged on user is still using NX on this Another logged on user is still using NX on this
system. NX can be used by at most one user at a system. NX can be used by at most one user at a
time. Try again after the other time. Try again after the other
MESS335 No NX Office Viewer license is available on [NX No NX Office Viewer license is available on [NX
workstati... workstati...
MESS336 The maximum allowed NX Office Viewer The maximum allowed NX Office Viewer
connections to [NX w... connections to [NX w...
MESS337 The connection with [NX workstation] is lost. This The connection with [NX workstation] is lost. This
system... system...
MESS339 NX Office Viewer Error NX Office Viewer Error
MESS340 Activate settings has failed. The configuration Activate settings has failed. The configuration
settings of the NX workstation have not been settings of the NX workstation have not been
updated. updated.
MESS341 Invalid password. Please enter the correct Invalid password. Please enter the correct
password. password.
MESS342 Password successfully changed. Password successfully changed.
MESS343 Invalid master password. Please enter the correct Invalid master password. Please enter the correct
master password. master password.
MESS344 The entered passwords are not identical. Please The entered passwords are not identical. Please
enter identical passwords. enter identical passwords.
MESS345 Export busy Export busy
MESS347 NX Office Viewer: A new version of the NX Office NX Office Viewer: A new version of the NX Office
Viewer is available from NX workstation <NX Viewer is available from NX workstation <NX
workstation>. The installation can workstation>. The installation can
MESS349 A not supported screen resolution has been A not supported screen resolution has been
detected. The minimum supported screen detected. The minimum supported screen
resolution is 1024x768. Please adjust your scr resolution is 1024x768. Please adjust your scr
MESS35 A DR Detector Error occured. Calibration DR- A DR Detector Error occurred. Calibration DR-
Detector failed. Detector failed.
MESS351 NX Workstation problem: No NX workstation was NX Workstation problem: No NX workstation was
found on the network. The NX Office Viewer can found on the network. The NX Office Viewer can
not be started. not be started.
MESS352 The password has not been changed. The password has not been changed.
MESS353 An internal error occurred while changing the An internal error occurred while changing the
password. Your password has not been changed! password. Your password has not been changed!
Please try again. Please try again.
MESS356 NX Office Viewer: An error has occurred while NX Office Viewer: An error has occurred while
downloading the NX Office Viewer installer from downloading the NX Office Viewer installer from
NX workstation <NX workstation>. NX workstation <NX workstation>.

Edition 1, Revision 2
11-2020 Livelink ID: 68274899 MESS-AllProjects
MESS-AllProjects
(Error messages from NX, DX-D 300, DX-D 600, all Sedecal Products in general)

Status 2020-10-29
Error Code Summary Message Instruction
MESS357 DICOM Digitizer communication error. Image DICOM Digitizer communication error. Image
receipt will be affected. The problem might be receipt will be affected. The problem might be
caused by a tcp port conflict. Please caused by a tcp port conflict. Please
MESS358 Title = Export Acquired Dose RecordsMessage = The acquired dose records have been exported
The Acqui... successfully
MESS360 You do not have the permission to save in this You do not have the permission to save in this
directory. directory.
MESS361 The file name is invalid. The file name is invalid.
MESS362 Title = Transfer ImagesMessage = The source Title = Transfer ImagesMessage = The source
and target ... and target ...
MESS363 Title = Transfer ImagesMessage = The source Title = Transfer ImagesMessage = The source
examination... examination...
MESS364 Title = Transfer ImagesMessage = The target Title = Transfer ImagesMessage = The target
examination... examination...
MESS365 Delete device: Are you sure you want to delete? Delete device: Are you sure you want to delete?
MESS366 Incorrect name representation Incorrect name representation
MESS367 Title = CalibrationMessage = Calibration distance Calibration distance must be between {0} cm and Please re-enter the calibration value
must ... {1} cm
MESS368 Calibration: Calibration DR-detector(s) for Calibration: Calibration DR-detector(s) for
mode{Z} expired mode{Z} expired
MESS369 Calibration: Calibration DR-detector(s) for Calibration: Calibration DR-detector(s) for
mode{Z} will expire in {7, ...} day(s) mode{Z} will expire in {7, ...} day(s)
MESS371 Error, non licensed digitiser(s) is/are configured Error, non licensed digitiser(s) is/are configured
MESS372 Error, XRay generator configuration parameter: { Error, XRay generator configuration parameter: {
XRayGeneratorDevice attributtname}, { error XRayGeneratorDevice attributtname}, { error
explanation } explanation }
MESS373 Error, Enabled DR_WORKFLOW license implies Error, Enabled DR_WORKFLOW license implies
enabled licens... enabled licens...
MESS374 DR-detector in location X having serial number Y DR-detector in location X having serial number Y
has exce... has exce...
MESS375 DR-detector in location X having serial number Y DR-detector in location X having serial number Y
received... received...
MESS376 Generator/detector errors Generator/detector errors
MESS377 Startup error: Supporting services not yet ready Startup error: Supporting services not yet ready
MESS378 Initialize Cassette: The plate type code is invalid Initialize Cassette: The plate type code is invalid
MESS379 No generator configured. No generator configured.
MESS38 Title = Print SheetMessage = The sheet cannot Title = Print SheetMessage = The sheet cannot
be printe... be printe...
MESS380 Device already shut down! Device already shut down!
MESS381 Device is in 'shutdown for maintenance' mode. Device is in 'shutdown for maintenance' mode.
MESS382 Device is starting up. Device is starting up.
MESS383 Device is in an unknown state. Device is in an unknown state.
MESS384 Error opening COM port. Generator connection is Error opening COM port. Generator connection is
aborted. aborted.
MESS385 Device is in error state. Device is in error state.
MESS386 Digitizer: Type Number mismatch Digitizer: Type Number mismatch
MESS387 Title = Receive imageMessage = An incoming Title = Receive imageMessage = An incoming
DICOM Object... DICOM Object...
MESS388 Failed to send collimator parameters: Failed to send collimator parameters:
Communication Failure Communication Failure
MESS389 Not supported to configure FLFS exposure(s) withNot supported to configure FLFS exposure(s) with
DR receptor type DR receptor type
MESS39 Title = Archive imageMessage = The image has Title = Archive imageMessage = The image has
already be... already be...
MESS39 Preparing calibration Preparing calibration. This could take upto <...> Set X-Ray Parameter according to DR Detector
seconds. Please wait. Calibration Key User.
MESS390 Failed to send U-arm parameters: Failed to send U-arm parameters:
Communication Failure Communication Failure
MESS391 Error, the recovery procedure of X-ray device Error, the recovery procedure of X-ray device
<name> has ... <name> has ...
MESS392 Error, the recovery procedure of X-ray device Error, the recovery procedure of X-ray device
<name> has ... <name> has ...
MESS393 Error, the recovery procedure of X-ray device Error, the recovery procedure of X-ray device
<name> has ... <name> has ...
MESS394 Error, license missing to support DR pannel Error, license missing to support DR pannel
configuration configuration
MESS395 DR: The acquired image couldn't be persisted The acquired image could not be persisted Please verify whether there is enough free disk
space
MESS396 Error, maximum 1 Musica MCE Detection Engine Error, maximum 1 Musica MCE Detection Engine
can be confi... can be confi...
MESS397 Error, Musica MCE server side AE title and port Error, Musica MCE server side AE title and port
must be s... must be s...

Edition 1, Revision 2
11-2020 Livelink ID: 68274899 MESS-AllProjects
MESS-AllProjects
(Error messages from NX, DX-D 300, DX-D 600, all Sedecal Products in general)

Status 2020-10-29
Error Code Summary Message Instruction
MESS398 Loading configuration file: Specified file contains Loading configuration file: Specified file contains
unlicensed version features unlicensed version features
MESS399 Warning, Musica MCE processing is not Warning, Musica MCE processing is not
supported in a CR1B configuration supported in a CR1B configuration
MESS4 Settings Accepted Settings Accepted
MESS40 Make next exposure Make exposure <index> of <NumExposures>.
MESS40 Title = Identify CassetteMessage = This cassette Title = Identify CassetteMessage = This cassette
alread... alread...
MESS400 Error, for the configuration of a Musica MCE Error, for the configuration of a Musica MCE
Engine Archive, the Musica2 Mammo MCE Engine Archive, the Musica2 Mammo MCE
License must be enabled. License must be enabled.
MESS401 Digitizer: Error, speed class free digitiser(s) Digitizer: Error, speed class free digitiser(s)
cannot be configured in LGM dose registering cannot be configured in LGM dose registering
mode mode
MESS402 Context = Verification config toolTitle = Error: DR workflow and DR Mammo workflow Change dose registering mode to EI via
WorkflowMes... license not supported in LGM dose registering Activation Utility
mode
MESS403 Digitizer: Wrong model file or to old version of Digitizer: Wrong model file or to old version of
model file for selected digitizer model file for selected digitizer
MESS404 Digitizer: Only 1 locally attached ASAP digitizer Digitizer: Only 1 locally attached ASAP digitizer
can be configured can be configured
MESS405 Digitizer: A remotely attached ASAP digitizer can Digitizer: A remotely attached ASAP digitizer can
only be combined with a locally attached ASAP only be combined with a locally attached ASAP
digitizer of the same product fa digitizer of the same product fa
MESS406 Digitizer: Only 1 remotely attached ASAP digitizer Digitizer: Only 1 remotely attached ASAP digitizer
can be configured can be configured
MESS407 Digitizer: Only a local and remote ASAP digitizer Digitizer: Only a local and remote ASAP digitizer
of the same product family can be combined of the same product family can be combined

MESS408 Digitizer: Warning, driver not installed for {0} Digitizer: Warning, driver not installed for {0}
digitizer digitizer
MESS409 van spetDigitizer: Warning, IPAddress not set for van spetDigitizer: Warning, IPAddress not set for
locally attached ASAP digitizer locally attached ASAP digitizer
MESS41 Title = Idenitfy CassetteMessage = There is no Title = Idenitfy CassetteMessage = There is no
image se... image se...
MESS410 Error, Maximum of Custom markers exceeded Error, Maximum of Custom markers exceeded
due to missing NX_NEW_FEATURES_3_0 due to missing NX_NEW_FEATURES_3_0
license license
MESS411 Error, Maximum of RIS devices exceeded due to Error, Maximum of RIS devices exceeded due to
missing NX_NEW_FEATURES_3_0 license missing NX_NEW_FEATURES_3_0 license
MESS412 Error, Maximum of Export destinations exceeded Error, Maximum of Export destinations exceeded
due to missing NX_NEW_FEATURES_3_0 due to missing NX_NEW_FEATURES_3_0
license license
MESS413 Digitizer: A newer model file for digitizer {digitizer Digitizer: A newer model file for digitizer {digitizer
name} is available in the database or in the name} is available in the database or in the
digitizer models folder digitizer models folder
MESS414 Collimator communication error with CAN-BUS Collimator communication error with CAN-BUS
MESS415 Possible failure in transfer of collimator settings Possible failure in transfer of collimator settings
for the currect exposure. for the currect exposure.
MESS416 Configured collimator values could not be set. Configured collimator values could not be set.
Current collimator settings are: Current collimator settings are:
<height>,<width>,<prefiltration> <height>,<width>,<prefiltration>
MESS417 DR-detector in location X having serial number Y DR-detector in location X having serial number Y
has exce... has exce...
MESS418 License error License error
MESS419 Digitizer error: Digitizer does not support Digitizer error: Digitizer does not support
speedclass free operation. Cassette not scanned. speedclass free operation. Cassette not scanned.

MESS42 Title = Message = The patient data cannot be Title = Message = The patient data cannot be
edited any... edited any...
MESS420 Name Representation differs between system Name Representation differs between system
<value> and import file <value>. <value> and import file <value>.
MESS421 The archive did not accept the GSPS SOP Class, The archive did not accept the GSPS SOP Class,
therefor the PS-object could not be archived therefor the PS-object could not be archived

MESS422 Please insert a blank disk Please insert a blank disk


MESS423 Patient Photo: Not a valid Patient Photo Patient Photo: Not a valid Patient Photo
MESS424 Memory Warning: Low memory conditions Low memory conditions observed Please reboot the system
observed
MESS425 Invalid license or wrong password Invalid license or wrong password
MESS426 Service protection expired or enter correct Service protection expired or enter correct
password password
MESS427 Protection error or wrong password Protection error or wrong password

Edition 1, Revision 2
11-2020 Livelink ID: 68274899 MESS-AllProjects
MESS-AllProjects
(Error messages from NX, DX-D 300, DX-D 600, all Sedecal Products in general)

Status 2020-10-29
Error Code Summary Message Instruction
MESS428 Protection of service tools: Error, device type Protection of service tools: Error, device type
number mismatch number mismatch
MESS429 No dongle found No dongle found
MESS43 Detector-battery is low Battery-status of DR Detector "<ID>" is Low. Please replace the battery with a fully charged.
MESS430 License already expired License already expired
MESS431 Permanent storage not accessible Permanent storage not accessible
MESS436 Protection of service tools: Error, protection of Protection of service tools: Error, protection of
service tools has expired service tools has expired
MESS437 Protection of service tools: Warning, protection of Protection of service tools: Warning, protection of
service tools has expired. This is your last login service tools has expired. This is your last login

MESS438 Protection of service tools: Wrong dongle found Wrong dongle found
MESS439 Protection of service tools: Error, no valid ALF file Protection of service tools: Error, no valid ALF file
found found
MESS44 Title = Message = The query could not be Title = Message = The query could not be
completed. Ple... completed. Ple...
MESS44 Detector-battery is empty Battery-status of DR Detector "<ID>" is Empty. Please replace the battery with a fully charged.
MESS440 Protection of service tools: Dongle has been Dongle has been changed. Restart your action
changed. Restart your action
MESS443 Error, Musica2/3 pediatric license is present but Error, Musica2/3 pediatric license is present but
disabled disabled
MESS444 Export Images, No CD/DVD drive available Export Images, No CD/DVD drive available
MESS445 Image plate gain calibration: Error, no license Image plate gain calibration: Error, no license
available to use this image plate available to use this image plate
MESS446 Image plate gain calibration: Error, no valid image Image plate gain calibration: Error, no valid image
plate gain calibration file found plate gain calibration file found
MESS447 Image plate gain calibration: Error, this image Image plate gain calibration: Error, this image
plate is not allowed plate is not allowed
MESS448 This product cannot be installed on this Windows This product cannot be installed on this Windows
version. version.
MESS449 Reading cassette information: Error, cassette not Reading cassette information: Error, cassette not
compatible with current NX version compatible with current NX version
MESS45 Title = Message = There are no matching Title = Message = There are no matching
accession numbe... accession numbe...
MESS450 Dongle not suitable for protection of service tools Dongle not suitable for protection of service tools
MESS451 Deactivation code expired or enter correct Deactivation code expired or enter correct
deactivation code deactivation code
MESS452 Error, "IDIS1 Connection" cannot be combined Error, "IDIS1 Connection" cannot be combined
with "XRG Connection for Mammo" or "Musica2 with "XRG Connection for Mammo" or "Musica2
Mammo" Mammo"
MESS453 Identify cassette: Cassette couldn't be identified Identify cassette: Cassette couldn't be identified

MESS454 Digitizer: Digitizer connection problem Digitizer: Digitizer connection problem


MESS455 Digitizer: Timeout occurred. The command has Digitizer: Timeout occurred. The command has
been aborted. been aborted.
MESS456 Digitizer: The digitizer cannot process the Digitizer: The digitizer cannot process the
command. command.
MESS457 Digitizer: The operation was cancelled on the Digitizer: The operation was cancelled on the
digitizer. digitizer.
MESS458 Digitizer: Timeout occurred. The command has Digitizer: Timeout occurred. The command has
been aborted. been aborted.
MESS459 Pre-processing: No gain calibration possible Pre-processing: No gain calibration possible
MESS46 Title = Open ExamMessage = The exam cannot Title = Open ExamMessage = The exam cannot
be opened be... be opened be...
MESS460 Digitizer: Some digitizers do not support Digitizer: Some digitizers do not support
Processing Station Names longer than 15 Processing Station Names longer than 15
characters characters
MESS461 Pre-processing: Image is not gain calibrated Pre-processing: Image is not gain calibrated
MESS462 Title = Receive DR imageMessage = A recovery A recovery image has been received and stored Select the recovery study in the patient switch,
image has ... in study {0}. Orientation can be wrong! apply the correct exposure type to the image and
transfer it to the correct patient.
MESS463 Service Warning: Enhanced DR Detector service Service Warning: Enhanced DR Detector service
mode is currently switched on. mode is currently switched on.
MESS464 Recovery Configuration: The selected Exposure Recovery Configuration: The selected Exposure
Type is used in a Recovery Exposure Type Type is used in a Recovery Exposure Type
MESS465 Recovery Configuration: Error, no exposure type Recovery Configuration: Error, no exposure type
defined for the <GenRad / Mammo> recovery defined for the <GenRad / Mammo> recovery
procedure procedure

Edition 1, Revision 2
11-2020 Livelink ID: 68274899 MESS-AllProjects
MESS-AllProjects
(Error messages from NX, DX-D 300, DX-D 600, all Sedecal Products in general)

Status 2020-10-29
Error Code Summary Message Instruction
MESS466 Recovery Configuration: Error, no protocol code Recovery Configuration: Error, no protocol code
defined for the <GenRad / Mammo> recovery defined for the <GenRad / Mammo> recovery
exposure settings exposure settings
MESS467 Recovery Configuration: Error, no valid exposure Recovery Configuration: Error, no valid exposure
type for the <GenRad / Mammo> recovery type for the <GenRad / Mammo> recovery
exposure settings exposure settings
MESS468 Error: duplicate AETitle, SOPClass combination Error: duplicate AETitle, SOPClass combination
found found
MESS469 Receive image: Unable to find a compatible Receive image: Unable to find a compatible
exposure type for a recovered image exposure type for a recovered image
MESS47 Auto stitch failed for patient [XXX] with patient ID Auto stitch failed for patient [XXX] with patient ID
[XXX] [XXX]
MESS470 Title = Receive imageMessage = An emergency An emergency image has been received and Select the emergency study in the patient switch,
image has b... stored in study {0}. Unable to find a compatible apply correct exposure type to the image and
emergency exposure type for the image. transfer it to the correct patient.
Orientation can be wrong!
MESS471 Receive image: Unable to find a compatible Receive image: Unable to find a compatible
exposure type for an emergency image exposure type for an emergency image
MESS472 Export Automation: No media found Export Automation: No media found
MESS473 Could not write to the disk: Please insert a new Could not write to the disk: Please insert a new
disk disk
MESS474 Warning, a non-licensed exam type is defined for Warning, a non-licensed exam type is defined for
a digitizer emergency button of digitizer <digitizer a digitizer emergency button of digitizer <digitizer
name> name>
MESS475 Warning, the exam type <GenRad or Mammo> is Warning, the exam type <GenRad or Mammo> is
not licensed and should not be used for a not licensed and should not be used for a
recovery exposure recovery exposure
MESS478 DR Full Leg Full Spine: Stitching failed: <failure DR Full Leg Full Spine: Stitching failed: <failure
reason> reason>
MESS482 DR Full Leg Full Spine: Full Leg Full Spine DR Full Leg Full Spine: Full Leg Full Spine
sequence failed: <failure reason> sequence failed: <failure reason>
MESS486 DR Full Leg Full Spine: Parameter validation error DR Full Leg Full Spine: Parameter validation error

MESS487 Import: All or some imports failed because the Import: All or some imports failed because the
export version is more recent than the installed export version is more recent than the installed
NX version NX version
MESS488 DR Full Leg Full Spine: Anatomic stitching failed, DR Full Leg Full Spine: Anatomic stitching failed,
system will try to stitch in Grid mode system will try to stitch in Grid mode
MESS489 Warning: the selected exposure type is used as Warning: the selected exposure type is used as
the emergency exposure type for the the emergency exposure type for the
XRayGenerator device XRayGenerator device
MESS490 Digitizer: Some digitizers support a maximum of Digitizer: Some digitizers support a maximum of
20 Processing Stations. If you add this one, this 20 Processing Stations. If you add this one, this
limit will be exceeded. limit will be exceeded.
MESS491 Title = Archive Image InformationMessage = The The Images could not be archived for destination Contact your system administrator
Images c... <destination name>
MESS492 Title = Archive Image / Dose InformationMessage The Images and/or Dose Information could not be Contact your system administrator
= The I... archived for destination <destination name>
MESS493 Title = Archive Dose InformationMessage = The The Dose Information could not be archived for Contact your system administrator
Dose Info... destination <destination name>
MESS494 Monitor: The resolution of the configured monitor Monitor: The resolution of the configured monitor
does not match with the resolution of the does not match with the resolution of the
physically attached monitor physically attached monitor
MESS495 Error, not all exposure types in the exposure Error, not all exposure types in the exposure
group are conform the automated DR fullscreen group are conform the automated DR fullscreen
workflow workflow
MESS496 Edit exposure: This station does not allow the Edit exposure: This station does not allow the
editing of some exposure parameters as long as editing of some exposure parameters as long as
the image has not been acquired the image has not been acquired
MESS497 Edit Exposure: A connection to the ID Tablet Edit Exposure: A connection to the ID Tablet
could not be made could not be made
MESS498 Edit Exposure: Re-identification of the cassette Edit Exposure: Re-identification of the cassette
failed! failed!
MESS499 Edit Exposure: The cassette was not correctly re- Edit Exposure: The cassette was not correctly re-
identified identified
MESS5 Collimator failed Collimator failed
MESS50 Title = Add PatientMessage = No new patients Title = Add PatientMessage = No new patients
can be add... can be add...
MESS500 Edit Exposure: Wrong cassette. The cassette is Edit Exposure: Wrong cassette. The cassette is
not identified for the current exposure not identified for the current exposure
MESS501 Edit Exposure: Changes cannot be applied Edit Exposure: Changes cannot be applied
without cassette re-identification without cassette re-identification
MESS502 Error, AE title of MCE archive must be different Error, AE title of MCE archive must be different
from AE title of Workstation MCE settings from AE title of Workstation MCE settings
MESS503 DR Full Leg Full Spine: Full Leg Full Spine Full Leg Full Spine sequence aborted
sequence aborted

Edition 1, Revision 2
11-2020 Livelink ID: 68274899 MESS-AllProjects
MESS-AllProjects
(Error messages from NX, DX-D 300, DX-D 600, all Sedecal Products in general)

Status 2020-10-29
Error Code Summary Message Instruction
MESS504 Auto Backup: Backup Failed! Check logfiles for Auto Backup: Backup Failed! Check logfiles for
details. details.
MESS505 Error, there is no XRay device defined; at least Error, there is no XRay device defined; at least
one must be defined when the IDIS license is one must be defined when the IDIS license is
enabled enabled
MESS506 Warning: The Extended Dose Reporting has Warning: The Extended Dose Reporting has
encountered an unexpected error encountered an unexpected error
MESS507 Prior Viewing: Prior viewing is not possible when Prior Viewing: Prior viewing is not possible when
Patient ID is empty Patient ID is empty
MESS508 Power failure detected: Shutdown of NX activated Power failure detected: Shutdown of NX activated
due to power failure due to power failure
MESS509 Error, enabled Multi Frame Imaging license Error, enabled Multi Frame Imaging license
requires an enabled DR_Workflow license requires an enabled DR_Workflow license
MESS510 Warning, Fast-ID is enabled, so only one digitiser Warning, Fast-ID is enabled, so only one digitiser
is supported is supported
MESS511 Central Monitoring System: The name Central Monitoring System: The name
representation of the Central Monitoring System representation of the Central Monitoring System
differs from one of the NX workstations differs from one of the NX workstations
MESS512 UI Configuration: You configured Scan Resolution UI Configuration: You configured Scan Resolution
in the View Image details but none of the in the View Image details but none of the
configured digitizers supports changi configured digitizers supports changi
MESS513 UI Configuration: You configured Scan Resolution UI Configuration: You configured Scan Resolution
in the Edit Image details but this is not supported in the Edit Image details but this is not supported
on Central Monitoring Syste on Central Monitoring Syste
MESS514 Identify Cassette: This image cannot be used as Identify Cassette: This image cannot be used as
the source for identification the source for identification
MESS515 Close Exam: A fluo sequence or a rapid Close Exam: A fluo sequence or a rapid
sequence shall be saved automatically sequence shall be saved automatically
MESS516 Error, no initial RIS device configured Error, no initial RIS device configured
MESS517 Error, archive {0} has no configuration for Error, archive {0} has no configuration for
dynamic images dynamic images
MESS518 Warning, no archive defined for supporting Multi Warning, no archive defined for supporting Multi
Frame Imaging Frame Imaging
MESS519 Error, XRF image storage not configured for Error, XRF image storage not configured for
export automation export automation
MESS52 Title = Receive ImageMessage = There is not There is not enough free space on the disk to To continue, please contact your system
enough free... receive the digitised images administrator
MESS520 Error, XRF image storage not configured for Error, XRF image storage not configured for
DICOM export {0} DICOM export {0}
MESS521 Error, license to support Musica3 is missing for Error, license to support Musica3 is missing for
package {0} package {0}
MESS522 Edit exposure: Changing the parameters "view Edit exposure: Changing the parameters "view
position", "image laterality", "exposure type" shall position", "image laterality", "exposure type" shall
be executed on all images ... be executed on all images ...
MESS523 Warning, changing the acquisition mode to Warning, changing the acquisition mode to
dynamic will remove the exposure type from all dynamic will remove the exposure type from all
print sheets print sheets
MESS524 Warning, license is missing for assigned package Warning, license is missing for assigned package
{0} {0}
MESS525 Warning, missing Musica2/3 processing Warning, missing Musica2/3 processing
parameters parameters
MESS526 Warning, Musica3 is not enabled for package {0} Warning, Musica3 is not enabled for package {0}

MESS527 Error, package {0} must be licensed Error, package {0} must be licensed
MESS528 Error, package {0} must use Musica3 Error, package {0} must use Musica3
MESS529 Partial Import: All Modality Settings will be cleared Partial Import: All Modality Settings will be cleared
because there is no XRG device configured on because there is no XRG device configured on
NX NX
MESS530 Delete XRG device: Are you sure you want to Delete XRG device: Are you sure you want to
delete the device? All modality settings will be delete the device? All modality settings will be
cleared too. cleared too.
MESS531 Internal error: NX is unable to show the image. Internal error: NX is unable to show the image.
The image may still be available in the refused The image may still be available in the refused
folder. folder.
MESS532 XRG attribute value conversion: Do you agree to XRG attribute value conversion: Do you agree to
apply following conversions on the imported apply following conversions on the imported
modality settings? modality settings?
MESS533 Partial Import: All Modality Settings will be cleared Partial Import: All Modality Settings will be cleared
because the imported modality settings are because the imported modality settings are
incompatible with the ... incompatible with the ...
MESS534 DES processing: DES processing failed DES processing: DES processing failed
MESS535 Add exam tree: Exam tree added containing Add exam tree: Exam tree added containing
exposure settings and modality settings exposure settings and modality settings
MESS537 Title = Message = Maintenance interval expired. Title = Message = Maintenance interval expired.
Please ... Please ...
MESS538 Identify Cassette / Copy Exposure: No valid Identify Cassette / Copy Exposure: No valid
ExposureTypeUid found. No exposure data ExposureTypeUid found. No exposure data
available. available.

Edition 1, Revision 2
11-2020 Livelink ID: 68274899 MESS-AllProjects
MESS-AllProjects
(Error messages from NX, DX-D 300, DX-D 600, all Sedecal Products in general)

Status 2020-10-29
Error Code Summary Message Instruction
MESS539 Maintenance: Start new maintenance period Maintenance: Start new maintenance period
MESS54 Title = Queuing ErrorMessage = Job execution Title = Queuing ErrorMessage = Job execution
failureI... failureI...
MESS540 Verification of settings: Dual Energy exposures Dual Energy exposures are not supported by the
are not supported by the configured modality. configured modality. No Dual Energy exposures
shall be available in NX.
MESS541 Verification of settings: Tomosynthesis exposures Tomosynthesis exposures are not supported by
are not supported by the configured modality. the configured modality. No Tomosynthesis
exposures shall be available in NX.
MESS542 DTS reconstruction: DTS reconstruction failed DTS processing failed Please refer to the user manual for possible
causes
MESS543 DTS acquisition: DTS sequence has been DTS sequence has been rejected automatically Please refer to the user manual for possible
rejected automatically because it was not because it was not complete causes
complete
MESS544 Error, invalid or unlicensed IP address specified Error, invalid or unlicensed IP address specified
MESS545 Error, incomplete IP address specified Error, incomplete IP address specified
MESS546 Error, hostname contains spaces or is empty Error, hostname contains spaces or is empty
MESS547 Warning, hostname resolves as an IP address Warning, hostname resolves as an IP address
MESS548 Error, IPStack configuration missing or unlicensed Error, IPStack configuration missing or unlicensed

MESS549 Error, no host connection data specified Error, no host connection data specified
MESS55 Invalid configuration The configuration is inconsistent with the Check the logfiles for details and correct the
applicable modelfiles. configuration.
MESS55 Queue xxx of device xxx has stalled Queue xxx of device xxx has stalled
MESS550 Warning, hostname of {0} resolves as an IP Warning, hostname of {0} resolves as an IP
address address
MESS551 Error, maximum 9 action buttons can be Error, maximum 9 action buttons can be
configured for the acquisition screen configured for the acquisition screen
MESS552 Warning, maximum 9 action buttons allowed Warning, maximum 9 action buttons allowed
MESS553 Archive Image / Dose Information: One or more One or more archive jobs in queue <queue Contact your system administrator
archive jobs in queue <queue name> have been name> have been refused by the destination
refused by the destination device device
MESS554 DTS reconstruction: Incomplete DTS sequence. Incomplete DTS sequence. Are you sure you
Are you sure you want to start reconstruction? want to start reconstruction?

MESS555 Import: All XRay Settings will be cleared because All XRay Settings will be cleared because no
no XRDI is installed XRDI is installed
MESS557 Verification of settings: Fluo exposures are not Fluo exposures are not supported by the
supported by the configured modality. configured modality. No Fluo exposures shall be
available in NX.
MESS558 Verification of settings: Dynamic exposures are Dynamic exposures are not supported by the
not supported by the configured modality. configured modality. No Dynamic exposures shall
be available in NX.
MESS559 Enable service protection failed Enable service protection failed
MESS56 Title = Read/Write CassetteMessage = No Title = Read/Write CassetteMessage = No
cassette detect... cassette detect...
MESS56 DR Generator Sync box is disconnected The DR Generator Sync box is disconnected from Check the cable between NX and the DR
the NX Workstation or is damaged. Generator Sync box. Replace if necessary.
MESS560 Error, archive {0} has no configuration for Error, archive {0} has no configuration for
TomoSynthesis images TomoSynthesis images
MESS561 Warning, no archive defined for supporting Warning, no archive defined for supporting
TomoSynthesis Imaging TomoSynthesis Imaging
MESS562 Error, storage for TomoSynthesis images not Error, storage for TomoSynthesis images not
configured for DICOM export {0} configured for DICOM export {0}
MESS563 Warning, no export destination defined for Warning, no export destination defined for
supporting Multi Frame Imaging supporting Multi Frame Imaging
MESS564 Warning, no export destination defined for Warning, no export destination defined for
supporting TomoSynthesis Imaging supporting TomoSynthesis Imaging
MESS565 Error, IPStack configuration of {0} {1} is missing or Error, IPStack configuration of {0} {1} is missing or
unlicensed unlicensed
MESS566 Collimation error: Too many collimation zones Too many collimation zones
MESS567 Collimation error: Collimation failed Collimation failed
MESS57 Title = Digitize CassetteMessage = The cassette Title = Digitize CassetteMessage = The cassette
cannot ... cannot ...
MESS570 DTS reconstruction: Hardware problem with GPU Hardware problem with GPU Please contact your service engineer

MESS571 DTS reconstruction: DTS reconstruction failed DTS reconstruction failed because of missing
because of missing data data
MESS572 Export Acquired Dose Records: No Acquired No Acquired Dose Records available
Dose Records available
MESS573 Freehand Rotate: Proceeding shall undo all Proceeding shall undo all changes on the image.
changes on the image Continue?
MESS574 Freehand Rotate: No license available in room No license available in room <room>. Freehand
<room>. Freehand rotate not possible. rotate not possible.
MESS575 No license available for a configured tool No license available for a configured tool

Edition 1, Revision 2
11-2020 Livelink ID: 68274899 MESS-AllProjects
MESS-AllProjects
(Error messages from NX, DX-D 300, DX-D 600, all Sedecal Products in general)

Status 2020-10-29
Error Code Summary Message Instruction
MESS576 Title = Rearrange images / thumbnailsMessage = One or more images have already been sent to
One or m... the archive, re-ordering is no longer possible
MESS577 Title = Grid annotationMessage = No license No license available in room <room>. Grid
available i... annotation not possible.
MESS578 Title = Message = There must be at least one Title = Message = There must be at least one
exposure t... exposure t...
MESS579 Title = OrthogonMessage = Orthogon cannot be Orthogon cannot be started Contact service if the problem persists
startedI...
MESS58 Portable DR Detector overheated The DR Detector "<ID>" has exceeded the
maximum working temperature.\nCurrent
temperature: <current temp in ºC>ºC\nMaximum
temperature: <max temp in ºC>ºC.\nImaging will
be impossible until the DR Detector has cooled
down sufficiently.
MESS58 Title = MPPS errorMessage = MPPS sending Title = MPPS errorMessage = MPPS sending
failed. The sy... failed. The sy...
MESS580 Title = OrthogonMessage = Orthogon Orthogon measurements failed Contact service if the problem persists
measurements failed...
MESS582 Title = CalibrationMessage = The SID must be The SID must be between {0} cm and {1} cm Please re-enter the SID
between {0...
MESS583 Title = Message = The system is running low on The system is running low on memory resources. Please restart the system at your earliest
memory r... Please restart the system at your earliest convenience
convenience.
MESS584 Title = Message = Emergency or recovery images Emergency or recovery images will be lost
will be... because the required license(s) have expired.
Please contact your system administrator.
Emergency or recovery images will be lost
because the required license(s) have expired.
Please contact your system administrator.
MESS585 Context = Verification config toolTitle = Message Warning: one or more configured external Click here to correct
= W... applications do not exist
MESS586 Title = Message = Critical error during image Critical error during image processing. The NX Use the {0} shortcut on the desktop to restart tthe
processin... Viewer will close. NX Viewer.
MESS588 Context = Verification config toolTitle = Message Error: enabled DR_MAMMO_WORKFLOW
= E... license implies enabled licenses for IDIS1,
MUSICA2, MUSICA2_MAMMO and disabled
licenses for Mammo XRG and Mammo Mixed
Use
MESS589 Context = Verification config toolTitle = Message Error: enabled DR_WORKFLOW and
= E... DR_MAMMO_WORKFLOW licenses imply
IDIS1, MUSICA2, BLACK BORDER,
MUSICA2_MAMMO, Mammo Mixed Use and
disabled license for Mammo XRG
MESS59 Fixed DR Detector overheated The DR Detector "<ID>" has exceeded the
maximum working temperature.\nCurrent
temperature: <current temp in ºC>ºC\nMaximum
temperature: <max temp in ºC>ºC.\nImaging will
be impossible until the DR Detector has cooled
down sufficiently.
MESS590 Context = Verification config toolTitle = Message Error: recovery exposure type for General
= E... Radiology exams is missing for X-ray device {0}
MESS591 Context = Verification config toolTitle = Message Error: recovery exposure type for Mammography
= E... exams is missing for X-ray device device {0}

MESS592 Context = Verification config toolTitle = Message Error, exposure type {0} with MCE requested
= E... cannot be processed with a Musica3 package
MESS593 Context = Verification config toolTitle = Message Error, DR Mammo exposure type {0} cannot have
= E... MCE requested
MESS594 Context = Title = Message = Rendering Rendering configuration is corrupt. Please, try
configuration i... 'revert to original' before contacting your system
administrator.
MESS595 Context = Title = Recovery images There are still recovery image(s) present in the Select the recovery study / studies in the patient
foundMessage = Ther... worklist. Attention: orientation can be wrong! switch, apply correct exposure type to the
image(s) and transfer them to the correct
patient(s).
MESS596 Title = Disk space limitedMessage = The system The system disk is running out of disk space Please contact your system administrator
disk is ...
MESS597 Title = Start NXMessage = Cannot start NX, Cannot start NX, incorrect monitor configuration Please, check whether monitor cable and power
incorrect mo... cable of second monitor are connected. Clicking
OK will stop the Viewer. Restart the Viewer
afterwards.
MESS598 Export Error XML file succesfull exported. Creating HTML was
impossible because of too many records.
MESS599 Error, an NX Workstation AETitle was used as Error, an NX Workstation AETitle was used as
AETitle for a DICOM Viewer AETitle for a DICOM Viewer
MESS6 Please insert the panel in portrait in the bucky Please insert the panel in portrait in the bucky
MESS60 Generator Failure Following generator failure occurred: Please contact your FSE.
"<errordescription> [<errorcode>]"

Edition 1, Revision 2
11-2020 Livelink ID: 68274899 MESS-AllProjects
MESS-AllProjects
(Error messages from NX, DX-D 300, DX-D 600, all Sedecal Products in general)

Status 2020-10-29
Error Code Summary Message Instruction
MESS600 Copy To Clipboard The clipboard is in use by another application Please, try again in a few moments or close the
application which is using the clipboard
MESS601 DR DSA exposures are not supported by the DR DSA exposures are not supported by the Click here to correct
configured modality. No DR DSA exposures shall configured modality. No DR DSA exposures shall
be available in NX. be available in NX.
MESS602 DR RoadMapping exposures are not supported DR RoadMapping exposures are not supported Click here to correct
by the configured modality. No DR RoadMapping by the configured modality. No DR RoadMapping
exposures shall be available in NX. exposures shall be available in NX.

MESS603 Quitting opacity mode Are you sure you want to quit opacity mode?
MESS604 Resetting opacity Are you sure you want to reset opacity?
MESS605 Roadmap mask acquisition Roadmap mask acquisition failed since not all The procedure has to be restarted prior to
mask frames were captured performing a roadmap acquisition
MESS606 Roadmap mask rejection Rejecting the active Roadmap mask will exit the The procedure has to be restarted prior to
procedure performing a roadmap acquisition
MESS61 Start Up Initializing ...
MESS61 Title = Digitise CassetteMessage = Error digitizing Title = Digitise CassetteMessage = Error digitizing
cas... cas...
MESS62 Title = Digitise CassetteMessage = Incorrect Title = Digitise CassetteMessage = Incorrect
digitizer ... digitizer ...
MESS63 Calibration aborted Aborting calibration. Please wait, this can take a
few minutes ...
MESS63 Title = Digitise CassetteMessage = System Title = Digitise CassetteMessage = System
errorInstru... errorInstru...
MESS64 Title = Digitise CassetteMessage = The image Title = Digitise CassetteMessage = The image
cannot be ... cannot be ...
MESS65 Title = Digitise CassetteMessage = Due to a Title = Digitise CassetteMessage = Due to a
network pro... network pro...
MESS66 Title = Message = An incoming Dicom Object was Title = Message = An incoming Dicom Object was
refused ... refused ...
MESS67 Title = Message = An incoming Dicom Object was Title = Message = An incoming Dicom Object was
refused ... refused ...
MESS68 Title = Message = An incoming Dicom Object was Title = Message = An incoming Dicom Object was
refused ... refused ...
MESS69 Title = Message = An incoming Dicom Object was Title = Message = An incoming Dicom Object was
refused ... refused ...
MESS7 Incorrect modality position. Incorrect modality position.
MESS70 Title = Message = An incoming Dicom Object was Title = Message = An incoming Dicom Object was
refused ... refused ...
MESS71 Title = Message = An incoming Dicom Object was Title = Message = An incoming Dicom Object was
refused ... refused ...
MESS72 Title = Message = An incoming Dicom Object was Title = Message = An incoming Dicom Object was
refused ... refused ...
MESS75 Title = Message = The RIS information could not The RIS information could not be retrieved
be retr...
MESS76 Title = Export Repeat/Reject RecordsMessage = The file could not be saved Make sure to select an accessible location and
The file ... retry the exxport of Repeat/Reject Records
MESS77 Title = Message = The images cannot be stitched, Title = Message = The images cannot be stitched,
becaus... becaus...
MESS78 Title = Message = The images could not be Title = Message = The images could not be
stitched. Ple... stitched. Ple...
MESS79 Title = Archive ImageMessage = Storage commit Title = Archive ImageMessage = Storage commit
timeout: ... timeout: ...
MESS8 Are you sure you want to clear the selected jobs? Are you sure you want to clear the selected jobs?

MESS80 Title = Archive ImageMessage = The images The images could not be archived. ({0} Contact your system administrator
could not be ...
MESS81 Title = Message = Please identify the cassette Title = Message = Please identify the cassette
manually... manually...
MESS82 Title = Message = Failed to send generator Title = Message = Failed to send generator
parameters: ... parameters: ...
MESS83 Title = Message = RIS data format not correct. Title = Message = RIS data format not correct.
Some dat... Some dat...
MESS84 Title = Export ImagesMessage = No writable Title = Export ImagesMessage = No writable
medium avail... medium avail...
MESS86 Title = Import ImagesMessage = The DICOMDIR Title = Import ImagesMessage = The DICOMDIR
file is inv... file is inv...
MESS87 Title = Message = RIS query already Title = Message = RIS query already
startedInstructio... startedInstructio...
MESS88 Title = Digitise CassetteMessage = A recovery A recovery image has been received and stored Select the recovery study in the patient switch,
image has... in study {0}. Orientation can be wrong! apply correct exposure type to the image and
transfer it to the correct patient.
MESS89 Title = Message = No permission to start the Title = Message = No permission to start the
applicatio... applicatio...
MESS90 Title = Export ImagesMessage = The selection of Title = Export ImagesMessage = The selection of
images ... images ...

Edition 1, Revision 2
11-2020 Livelink ID: 68274899 MESS-AllProjects
MESS-AllProjects
(Error messages from NX, DX-D 300, DX-D 600, all Sedecal Products in general)

Status 2020-10-29
Error Code Summary Message Instruction
MESS91 Title = Export imagesMessage = Could not write Title = Export imagesMessage = Could not write
to the d... to the d...
MESS93 Possible image loss There is no image available on DR Detector
"<ID>"
MESS93 Error, the specified file could not be created. Error, the specified file could not be created.
MESS94 The specified file already exists, do you want to The specified file already exists, do you want to
overwrite ? overwrite ?
MESS94 DR FLFS sequence aborted DR Full Leg Full Spine sequence aborted: Reduce the ROI and contact service to adapt the
position parameter(s) invalid room configuration.
MESS95 Error, save failed please check that the disk is not Error, save failed please check that the disk is not
full or write protected. full or write protected.
MESS96 Verification has found errors Verification has found errors
MESS97 Your configuration is activated but the backup Your configuration is activated but the backup
creation failed. See log files for details. creation failed. See log files for details.
MESS98 Panel Selection <number of panel in the list> Panels loaded,
select one from the list
MESS98 Error, Activate settings failed. Error, Activate settings failed.
MESS99 [FLFS] FLFS sequence completed FLFS sequence completed
MESS99 Error, the last Activate settings has failed. Error, the last Activate settings has failed.

Edition 1, Revision 2
11-2020 Livelink ID: 68274899 MESS-AllProjects
RFBMESS
(Error messages from Generator Sync Box,
Retrofit Box Firmware, Retrofit Box Controller)
Status 2019-09-26
Error Code Summary Message Instruction
RFBMESS1 "Firmware Update Succeeded" Firmware Update Succeeded
RFBMESS2 Wrong file The selected file is not a firmware file Please select a firmware file
RFBMESS3 "Update failed" Firmware Update Failed Please check log files. Please check the
connection to the DR Generator Sync Box.
RFBMESS4 No info acquired No information received Please check the connection to the DR Generator
Sync Box.
RFBMESS7 NX Conflict Can't start <Tool Name> while other NX software
is running
RFBMESS9 DR Generator Sync Box is disconnected The DR Generator Sync Box is disconnected Check the ethernet-cable between NX and the
from the NX Workstation or is damaged. DR Generator Sync Box. Replace if necessary.
Check network settings of the DR Generator Sync
Box.
RFBMESS10 Logging Retrieved The logging of the DR Generator Sync Box was
successfully retrieved.
RFBMESS11 No Connection No connection with DR Generator Sync Box.
RFBMESS12 Test Finished DR Generator Sync Box connected
RFBMESS13 Test Finished DR Generator Sync Box not connected
RFBMESS14 "Re-booting" Rebooting device...
RFBMESS15 Updating the DR Generator Sync Box succeeded The DR Generator Sync Box is updated with the
new IPv4 address <New DR Generator Sync Box
IP Address>
RFBMESS16 Info Retrieved The Info of the DR Generator Sync Box was
successfully saved.
RFBMESS18 "Please select FW file" Please select the firmware file to update the DR
Generator Sync Box
RFBMESS19 "Check selected FW file" Please verify the specified configuration and
press the next button to start the update process.

RFBMESS20 "Uploading FW. Do not power down" Uploading Firmware file to DR Generator Sync
Box. Do not power down the DR Generator Sync
Box while uploading the firmware!
RFBMESS21 "Executing uploaded file" Executing uploaded file...
RFBMESS22 "FW upgrade check" Checking if firmware is updated...
RFBMESS23 "Upgrade aborted" The upgrade of the DR Generator Sync Box was Please re-start firmware upgrade.
aborted.
RFBMESS24 Update failed Firmware version mismatch between selected file
and file installed on DR Generator Sync Box.

RFBMESS25 No Logging data. No log data available.


RFBMESS26 No Logging data. Retrieving logging data failed.
RFBMESS27 No IP address. Please enter an IP address.
RFBMESS28 No valid IP address. Please enter a valid IP<IP number type> address.

RFBMESS30 DR Generator Sync Box Hardware. Rotary switch inside DR Generator Sync Box is
not in position 0! Please change it.
RFBMESS31 DR Generator Sync Box Firmware. Possible firmware problem with register
<regerister name>\nPlease contact service.
RFBMESS32 Another service tool is still running. Previous service tool detected
RFBMESS34 "The selected file is OK" The selected file is OK. Press next to continue.
RFBMESS35 "Wrong Configuration." Wrong configuration. Exposure button is not
expected to be connected to the Box.
RFBMESS36 License Error There are no matching licenses found. Please check the license file.
RFBMESS37 Software Version Error The firmware version of the GSB is not Please contact your FSE
compatible with the version of RFBC. No
communication possible.
RFBMESS38 Synchronization failed Synchronization between GSB and Panel failed. If the problem persists, please contact your
Exposure can not start. Service Engineer.
RFBMESS39 Exposure aborted Exposure aborted, due to an internal error. If the problem persists, please contact your
<GSBerror> Service Engineer.
RFBMESS40 Exposure not started Exposure not started, due to an internal error. Please try again.\nIf the problem persists, please
<GSBerror> contact your Service Engineer.

Edition 1, Revision 2
11-2020 Livelink ID: 68274899 RFBMESS
SDL_MESS
(Error messages from Sedecal Retrofit)

Status 2019-09-26
Error Code Summary Message Instruction
SDL_MESS1 License Error License error: license <TECHNICAL- Please check the license file.
LICENSENAME> for
<COMPONENT/APPLICATION> not found.
SDL_MESS2 Generator connection problem The connection with the generator cannot be Please make sure that the serial cable between
established. the NX Workstation and the Generator is properly
plugged in and restart the NX Workstation. If the
problem persists, please contact your Service
Engineer.
SDL_MESS4 Initalizing Generator initializing
SDL_MESS5 Manual Filter It is mandatory to use a manual prefilter.
SDL_MESS6 Manual Grid It is mandatory to use a grid.
SDL_MESS7 Generator parameters not set Not possible to set generator params, please retry

SDL_MESS8 Generator Message Cannot set mAs for this focal spot. Please check your configuration.

Edition 1, Revision 2
11-2020 Livelink ID: 68274899 SDL_MESS
SDRMESS
(Error messages from Smart DR (DR 600, DR 100s))

Status 2020-10-29
Error Code Summary Message Instruction
SDRMESS1 License Error There are no matching licenses found. Please check the license file.
SDRMESS2 License Error A matching license has been found, but there is a This installation will be aborted.
license version mismatch.
SDRMESS3 Calibration Started Calibration Started
SDRMESS4 Calibration Stopped Calibration Stopped
SDRMESS5 Calibration Point Calibration <Position>_<SID>
SDRMESS6 Calibration out of range Calibration out of range <SID>
SDRMESS7 Camera Health Check cannot be performed Camera Health Check cannot be performed:
<error code>
Camera Health Check cannot be performed:
SDRMESS8 Incorrect Modality Position Modality in wrong position Please point tube towards Wall or Table
Camera Health Check cannot be performed: Please check that the tube is centered
SDRMESS9 Incorrect Position Modality incorrectly positioned perpendicular towards the bucky

Edition 1, Revision 2
11-2020 Livelink ID: 68274899 SDRMESS
SMS_MESS
(Error messages from Siemens (DX-D 500, DR Retrofit))

Status 2019-09-26
Req. ID Summary Message Instruction
SMS_MESS1 License Error License error: license <TECHNICAL- Please check the license file.
LICENSENAME> for
<COMPONENT/APPLICATION> not found.
SMS_MESS2 Generator connection problem The connection with the generator cannot be Please make sure that the serial cable between
established. the NX Workstation and the Generator is properly
plugged in and restart the NX Workstation. If the
problem persists, please contact your Service
Engineer.
SMS_MESS3 Collimator connection problem The connection with the collimator cannot be Please make sure that the cable between the NX
established. Workstation and the collimator is properly plugged
in and restart the NX Workstation. If the problem
persists, please contact your Service Engineer.

Edition 1, Revision 2
11-2020 Livelink ID: 68274899 SMS_MESS
SPMMESS
(Error messages from Spellman Generator (DR 400, DR 600))

Status 2020-10-29
Error Code Summary Message Instruction
SPMMESS1 Fatal generator error A fatal generator error <GeneratorErrorID> Please call your Service Representative and turn
occurred. off the system.
SPMMESS2 Generator error A generator error <GeneratorErrorID> occurred. Please call your Service Representative.
SPMMESS3 Generator error A generator error <GeneratorErrorID> occurred. Please restart the generator and call your Service
Representative if the problem persists.
SPMMESS4 Generator error A generator error <GeneratorErrorID> occurred. Please call your Service Representative if the
problem persists.
SPMMESS5 Prep-timeout Prep-timeout, exposure is not possible. Please release prep-button and try again.
SPMMESS6 Exposure aborted Exposure was aborted by user.
SPMMESS7 Tube too hot The temperature of the tube is above 70°C. Please allow the tube too cool off.\n
Please call your Service Representative if the
problem persists.
SPMMESS8 Wrong ms-value The ms-value is too high. Please change the value and try again.\nPlease
call your Service Representative if the problem
persists.
SPMMESS9 Wrong ms-value The ms-value is too low. Please change the value and try again.\nPlease
call your Service Representative if the problem
persists.
SPMMESS10 Wrong tube position Wrong tube position. Please change the tube position and try again.
\nPlease call your Service Representative if the
problem persists.
SPMMESS11 License Error There are no matching licenses found. Please check the license file.
SPMMESS12 Generator connection problem The connection with the generator cannot be Please make sure that the serial cable between
established. the NX Workstation and the Generator is properly
plugged in and restart the NX Workstation. If the
problem persists, please contact your Service
Engineer.
SPMMESS13 Overload Generator overload. Please switch off the generator and allow cooling
down.\nPlease call your Service Representative if
the problem persists.
SPMMESS15 Dose too low The dose at AEC detected is too low. Please check tube position and exposure
parameters.\nPlease call your Service
Representative if the problem persists.
SPMMESS16 max mAs reached The generator aborted the exposure because Please check the exposure parameters.\nPlease
max mAs was reached. call your Service Representative if the problem
persists.
SPMMESS17 Initializing Generator initializing
SPMMESS22 Generator parameters not set Not possible to set generator params, please retry

SPMMESS23 License Error A matching license has been found, but there is a Please check the license file.
license version mismatch.
SPMMESS24 Parameter limit reached Parameter limit reached: <Parameter>
SPMMESS25 Exposure time too long Too long exposure time detected with AEC Please check your AEC selection and try again.
exposure. Contact Service if the problem persists.

Edition 1, Revision 2
11-2020 Livelink ID: 68274899 SPMMESS
SSCMESS
(Error messages from Sedecal Softconsole)

Status 2020-10-29
Error Code Summary Message Instruction
MESS11 Workstation Communication Error Workstation Communication Error
MESS12 Software console is already running. Software console is already running.
MESS14 Warning, the actual SID is not suited for the Warning, the actual SID is not suited for the
applied grid. ( xx cm < SID < xx cm) applied grid. ( xx cm < SID < xx cm)
MESS15 Warning, The grid is -not- inserted although not Warning, The grid is -not- inserted although not
requested by the modality workstation. requested by the modality workstation.
MESS16 Error, The grid is not correctly positioned Error, The grid is not correctly positioned
MESS18 Door Open error. The X-ray room door is open Door Open error. The X-ray room door is open
when the X-ray equipment is in use. when the X-ray equipment is in use.
MESS19 No Communication No Communication
MESS20 No AEC configured in this generator. Technique No AEC configured in this generator. Technique
rejected rejected
MESS21 Energy saving behaviour Energy saving behaviour
MESS22 Empty battery - shutting down system Empty battery - shutting down system
MESS23 Spatial charge. Please, decrease mA value Spatial charge. Please, decrease mA value
MESS24 Spatial charge. Please, increase kV value or Spatial charge. Please, increase kV value or
exposure time exposure time
MESS25 FLFS sequence completed FLFS sequence completed
MESS26 Generator parameter <parameter name> out of Generator parameter <parameter name> out of
range range
MESS30 FLFS sequence aborted FLFS sequence aborted
MESS31 Parameter values accepted Parameter values accepted
MESS32 Positioner parameter <parameter name> value Positioner parameter <parameter name> value
out of range out of range
MESS33 Millenium Shutdown Millenium Shutdown

Edition 1, Revision 2
11-2020 Livelink ID: 68274899 SSCMESS
TDIMESS
(Error messages from Trixell Dynamic Panel (DR 800))

Status 2020-10-29
Error Code Summary Message Instruction
TDIMESS1 License Error License error: license <TECHNICAL- Please check the license file.
LICENSENAME> for
<COMPONENT/APPLICATION> not found.
TDIMESS3 <Software name + version> Setup A matching license has been found, but there is a
license version mismatch.\n\nThis installation will
be aborted.
TDIMESS4 <Software name + version> Setup There are no matching licenses found.\n\nThis
installation will be aborted.
TDIMESS5 <Software name + version> Setup You do not have enough privileges to install this
product.\n\n Administrator privileges are
needed.\n\n This installation will be cancelled.
TDIMESS6 Calibration: Calibration step started Starting <Calibration Name> for mode <x> of <y>
(PU-Mode <PU-mode>).
TDIMESS7 Calibration: Preparing for exposure Calibrating mode <current number of mode> of Set exposure type to <exposure type>.\nSet dose
<total number of modes> (PU-mode: <PU to <dose>µGy.\n\nkV: <kV>, mA: <mA>\nSet SID
mode>).\n\nPreparing for exposure... = 130.\n(mA value is indicative and may vary to
reach to indicated dose).
TDIMESS8 Calibration: Take Exposure Calibrating mode <current number of mode> of Set exposure type to <exposure type>.\nSet dose
<total number of modes> (PU-mode: <PU to <dose>µGy.\n\nkV: <kV>, mA: <mA>\nSet SID
mode>).\n\nPRESS <EXPOSE BUTTON or = 130.\n(mA value is indicative and may vary to
PEDAL> (Sequence <number of sequence> of reach to indicated dose).
<Total number of sequences>).
TDIMESS9 Calibration: Processing Data Calibrating mode <current number of mode> of So far <time past for current step in seconds>
<total number of modes> (PU-mode: <PU seconds.
mode>).\n\nProcessing Data. Step <number of
current step> of <total numbers of steps>.
TDIMESS10 Calibration: Average pixel value too low Calibrating mode <current number of mode> of Use higher dose for sequence <failed sequence>
<total number of modes> (PU-mode: <PU of <total number of sequences>.
mode>).\n\nCalibration step failed. \nAverage
pixel value of <average pixel value>. \nTarget
average pixel value is <target average pixel
value>.
TDIMESS11 Calibration: Average pixel value too high Calibrating mode <current number of mode> of Use lower dose for sequence <failed sequence>
<total number of modes> (PU-mode: <PU of <total number of sequences>.
mode>).\n\nCalibration step failed. \n
Average pixel value of <average pixel value>.
\nTarget average pixel value is <target average
pixel value>.
TDIMESS12 Calibration: Uniformity Test failed Uniformity Test failed. Please check collimation. The panel should be
fully covered.\n Please check beam path for
objects.
TDIMESS13 Calibration: Calibration failed - generic <Calibration Name> failed.\nReason: <Reason>.

TDIMESS14 Calibration: Do you want to calibrate this mode? Do you want to calibrate mode <x> of <y> (PU- The calibration data of this mode is valid for
mode: <PU mode>)?\n(<Exposure Type> <amount of days> days and <amount of hours>
<sensor area>) hours.
TDIMESS15 Calibration: ECV check results Result of Defect Pixel Map check (PU-Mode <PU- ECV = <ECV value>\nMax. ECV = <max
mode>): ECV>\n<PASSED/FAILED>
TDIMESS16 Calibration: Preparing for exposure Calibrating mode <current number of mode> of Set exposure type to <exposure type>.\nSet dose
<total number of modes> (PU-mode: <PU to <dose>µGy.\n\nkV: <kV>, mAs: <mAs>.\nSet
mode>).\n\nPreparing for exposure... SID = 130.\n(mAs value is indicative and may
vary to reach to indicated dose).

TDIMESS17 Calibration: Take Exposure Calibrating mode <current number of mode> of Set exposure type to <exposure type>.\nSet dose
<total number of modes> (PU-mode: <PU to <dose>µGy.\n\nkV: <kV>, mAs: <mAs>.\nSet
mode>).\n\nPRESS <EXPOSE BUTTON or SID = 130.\n(mAs value is indicative and may
PEDAL> (Sequence <number of sequence> of vary to reach to indicated dose).
<Total number of sequences>).
TDIMESS18 Calibration: Exposing Exposing...
TDIMESS19 Calibration: Stop exposing RELEASE <EXPOSE BUTTON or PEDAL>.
TDIMESS20 Calibration: Processing Data Processing data...
TDIMESS21 Calibration: Panel Temperature Temperature of Panel <Panel> is <Panel Please press button to continue.
temperature> °C.
TDIMESS22 Fatal DR Detector Error DR Detector <panel ID> encountered a fatal Restart sytem and please call Service.
error: <ErrorID>.
TDIMESS25 Frame Grabber DriverTest Result Frame grabber driver is installed correctly.
TDIMESS27 Starting Self-Test Starting Self-Diagnosis...
TDIMESS28 Connection Problem Connection problem, Self-Diagnosis is Check if PU-Box / panel is powered\nCheck if
terminated.\nPlease try again later... network switch is powered\nCheck network
cabling\n
TDIMESS30 Software Conflict Can't start <Tool Name> while other NX software STOP NX completely before launching <Tool
is running. Name>.
TDIMESS31 No Administrator Rights <Tool name> is only available for Administrator
accounts.
TDIMESS32 Software Conflict Cannot start TDI Panel Installation Tool while
another instance of TDI Panel Installation Tool is
running.

Edition 1, Revision 2
11-2020 Livelink ID: 68274899 TDIMESS
TDIMESS
(Error messages from Trixell Dynamic Panel (DR 800))

Status 2020-10-29
Error Code Summary Message Instruction
TDIMESS33 Temperature Error The temperature of the DR detector <panel> is Turn the power of the DR detector off and let it
too high. cool down.
TDIMESS34 Temperature Error The temperature of the DR detector <panel> is Leave the power of the DR detector on to let it
too low. warm up for 4 hours. Make no exposures during
warm up.
TDIMESS35 Panel temp high DR Detector <detector name>'s temperature is
high (&gt; <warning temperature> degrees
Celsius).\nImage quality might be impacted.
TDIMESS36 Calibration:Take RAD exposure PRESS <EXPOSE BUTTON> (Image <number Set exposure type to <exposure type>.\nSet dose
of image> of <Total number of images>). to <dose>µGy.\n\nkV: <kV>, mAs: <mAs>\nSet
SID = 130.\n(mAs value is indicative and may
vary to reach to indicated dose).

TDIMESS37 Calibration Temp Lower Limit DR Detector <detector name>'s temperature is


lower than the Calibration Temperature range (&lt;
<Lower limit Calibration Temperature> degrees
Celsius).\nImage quality might be impacted.

TDIMESS38 Calibration Temp Upper Limit DR Detector <detector name>'s temperature is


higher than the Calibration Temperature range
(&gt; <Upper Limit Calibration Temperature>
degrees Celsius).\nImage quality might be
impacted.
TDIMESS39 Calibration: Preparing for exposure Calibrating mode <current number of mode> of Set exposure type to <exposure type>.\nSet dose
<total number of modes> (PU-mode: <PU to <dose>µGy.\n\nkV: <kV>, mA: <mA>, ms:
mode>).\n\nPreparing for exposure... <ms>\nSet SID = 130.\n(mA value is indicative
and may vary to reach to indicated dose).

TDIMESS40 Calibration: Take Exposure Calibrating mode <current number of mode> of Set exposure type to <exposure type>.\nSet dose
<total number of modes> (PU-mode: <PU to <dose>µGy.\n\nkV: <kV>, mA: <mA>, ms:
mode>).\n\nPRESS <EXPOSE BUTTON or <ms>\nSet SID = 130.\n(mA value is indicative
PEDAL> (Sequence <number of sequence> of and may vary to reach to indicated dose).
<Total number of sequences>).
TDIMESS41 Connection problem Connection problem.\nWorkflow ended. Check the connection between NX Workstation
and PU-box.
TDIMESS42 Settings saved. The settings are saved.
TDIMESS43 Unsaved settings. The settings are not yet saved.\nDo you want to
save settings?
TDIMESS44 Configure PU-box Are you sure that you want to configure the PU-
box?
TDIMESS45 Configure PU-Box Configuring PU-box successfully completed.
TDIMESS46 Configure PU-box Are you sure that you want to clear the
configuration of the PU-box?
TDIMESS47 Configure PU-box Modifying PU-box... Please do not turn of the system.
TDIMESS48 Configure PU-box Clearing the configuration of the PU-box
successfully completed. \nFactory default
configuration and calibration data is loaded.
TDIMESS49 Retrieve Logging/calibration files Retrieving logging/calibration files...
TDIMESS50 Retrieve Logging/calibration files Logging/calibration files are retrieved. Logging/calibration files can be found
under\n<path>.
TDIMESS51 Diagnostic Calibration Check Starting Diagnostic Calibration Check.\nThe Please select panel <panel identifier> and press
Calibration Tool will be started. 'Start Calibration'. For (PU-mode: <PU mode>).

TDIMESS52 Diagnostic Calibration Check: No normal This is not a normal Calibration flow but a Only Service technicians are allowed to perform
Calibration Diagnostic Calibration Check. this flow.\nIf you are not a Service Technician,
please abort this flow and restart.
TDIMESS53 Diagnostic Calibration Check: Retrieving images Retrieving result of the Diagnostic Calibration Please wait...
Check...
TDIMESS54 Diagnostic Calibration Check: Result retrieved. The result of the Diagnostic Calibration Check is The result can be found under\n<path>.
retrieved.
TDIMESS55 Panel Error Error occurred during loading panel modes. Restart system. If problem persists, call service.

Edition 1, Revision 2
11-2020 Livelink ID: 68274899 TDIMESS
THSMESS
(Error messages from Technix (DR100s))

Status 2020-10-29
Error Code Summary Message Instruction
THSMESS1 License Error There are no matching licenses found. Please check the license file.
THSMESS2 License Error A matching license has been found, but there is a This installation will be aborted.
license version mismatch.
THSMESS3 Generator Warning Generator Message: <ErrorID>
THSMESS4 Generator error Generator encountered a problem: <ErrorID>. Restart the system. Call service if the problem
persists.
THSMESS6 System Shutdown System is shutting down.
THSMESS7 Generator error Generator encountered a problem: <ErrorID>. Press OK to continue. If the error appears again,
call Service.
THSMESS8 Temperature warning Tube reached maximum allowed temperature. Please wait for the tube to cool down.
THSMESS9 Generator prep timeout Generator prep timeout. Ensure the exposure switch is pressed in less
then 15s after the preparation switch was
pressed. Call service if the problem persists.
THSMESS10 Bumper Activated Bumper activated. Object hit and stopped movment. Please drive
backwards and away from object.
THSMESS11 Handle not free Movement button pressed during startup of Restart system and ensure no movement buttons
system. are pressed.
THSMESS12 Emergency Button Pressed. Emergency button pressed. Release button and press OK to continue.
THSMESS13 Tube Seasoning. Tube Seasoning required. Press OK to continue. Call Service to perform the
tube seasoning.
THSMESS14 mAs not allowed mAs not allowed for required focus. Change mAs data and repeat the operation.
THSMESS15 Focus not allowed Tube focus not allowed for required data. Change Focus and repeat the operation.
THSMESS16 Generator Connection Error The connection with the generator cannot be Please check the connection between NX
established. Workstation and Generator, and restart the NX
Workstation. Contact service if the problem
persists.
THSMESS17 Manual Abort Exposure was manually aborted by the operator.

THSMESS18 Battery Critical Battery Level Critical. A single exposure can be


made by pressing the Battery Critical icon.
THSMESS19 No Exposures Possible It is not possible to make any further exposures Move the equipment in a place suitable for battery
due to low or empty battery. charging and recharge batteries.
THSMESS20 FLFS not possible Generator Battery level too low to complete FLFS Charge the battery.
workflow. Workflow will be aborted.
THSMESS21 Handswitch Error Handswitch error. Release exposure button. Try again to expose.
If issue persists, call service.
THSMESS22 Tube rotation error Tube anode rotation issue. During preperation phase no anode rotation
detected. If issue persists, call service.
THSMESS23 Filament Error Interrupted filament current. Filament heating current interrupted.
Please try again and/or change focal spot size.
THSMESS24 HV Tank heat error HV Tank reached heat limit. Wait for HV tank to cool down and try again.
THSMESS25 Battery recharge warning The residual charge of the batteries is such that In case you don't proceed with the recharge but
their recharge is allowed in five hours. you continue to work, the recharge of the batteries
in five hours is not guaranteed.
THSMESS26 Low residual charge Low residual charge on batteries. Please fully charge batteries by plugin cable to
mains.

Edition 1, Revision 2
11-2020 Livelink ID: 68274899 THSMESS
THXMESS
(Error messages from Technix (DR100e))

Status 2020-10-29
Error Code Summary Message Instruction
THXMESS3 License Error There are no matching licenses found. Please check the license file.
THXMESS4 License Error A matching license has been found, but there is a Please check the license file.
license version mismatch.
THXMESS5 Generator Connection Error The connection with the generator cannot be Please check the connection between NX
established. Workstation and Generator, and restart the NX
Workstation. Contact service if the problem
persists.
THXMESS6 Device Message The Mobile Unit generated a warning or error Please check the Mobile Unit display and user
message. documentation for more information.

Edition 1, Revision 2
11-2020 Livelink ID: 68274899 THXMESS
TRIMESS
(Error messages from PIXIUM Detector (DR10s, DR14s, Pixium Rad 4343))

Status 2020-10-29
Error Code Summary Message Instruction
TRIMESS19 Save Changes Do you want to save changes?
TRIMESS21 Remove panel? Are you sure that you want to remove panel
<panel name>?
TRIMESS23 No Administrator Rights <Tool name> is only available for Administrator
accounts.
TRIMESS24 Calibration Files not valid Calibration Files not valid for DR detector <Panel Please perform calibration. If the problem
Name>. persists, please contact your Service Engineer.
TRIMESS25 Calibration: Continuing the calibration Continuing the calibration ...
TRIMESS26 Calibration: Prepare User for Exposure Take exposure <imagenumber> of <number of
acquisitons per dose setting> in <amount of
seconds> seconds.
TRIMESS27 Calibration: Take Exposure EXPOSE
TRIMESS28 Calibration: integration started Integration started.
TRIMESS29 Calibration: Integration stopped Integration stopped.
TRIMESS30 Calibration: Calibration Mode started Starting Calibration Mode: <Calibration Mode>.
TRIMESS31 Calibration: Dose message Set dose to <dose>µGy, kV: <kV>, mAs:
<mAs>.\nmAs value (SID =100) is indicative and
can vary to reach the indicated dose.\nThe
average SID for normal operation should be set.
TRIMESS33 Calibration: Average pixel value too low Calibration step failed. \nAverage pixel value of Use higher dose.\nPlease check the "Calibration
<aveage pixel value>. \nTarget average pixel Key User Manual" for exposure parameters.
value is <target average pixel value>.
TRIMESS34 Calibration: Average pixel value too high Calibration step failed. \nAverage pixel value of Use lower dose.\nPlease check the "Calibration
<aveage pixel value>. \nTarget average pixel Key User Manual" for exposure parameters.
value is <target average pixel value>.
TRIMESS38 License Error License error: license <TECHNICAL- Please check the license file.
LICENSENAME> for
<COMPONENT/APPLICATION> not found.
TRIMESS39 <Software name + version> Setup A matching license has been found, but there is a
license version mismatch./nThis installation will be
aborted.
TRIMESS40 <Software name + version> Setup There are no matching licenses found. This
installation will be aborted.
TRIMESS41 <Software name + version> Setup You do not have enough privileges to install this
product./n
Administrator privileges are needed./n
This installation will be canceled.
TRIMESS44 Temperature Error The temperature of the DR detector <panel> is Turn the power of the DR detector off and let it
too high. cool down.
TRIMESS45 Temperature Error The temperature of the DR detector <panel> is Leave the power of the DR detector on to let it
too low. warm up.
TRIMESS46 Calibration: panel temperature Temperature of Panel <Panel> is <Panel Please press button to continue.
temperature> °C.
TRIMESS47 Software Conflict Cannot start TRI Panel Installation Tool while
another instance of TRI Panel Installation Tool is
running.
TRIMESS48 Registering Panel Registering Panel...
TRIMESS49 Connecting to the Panel Connecting to the Panel (takes half a minute)...
TRIMESS50 Starting Self Test Starting Self Test...
TRIMESS56 Self Test Self Test cannot be performed without saving the
Panel settings.\n\nDo you want to save and
continue the Self Test?
TRIMESS57 Get Hardware Get Hardware...
TRIMESS58 Compute ECV Compute ECV...
TRIMESS59 Get voltages Get voltages...
TRIMESS61 Retrieving infoCounterReportFile Retrieve Info Counter Report cannot be
performed without saving the Panel
settings.\n\nDo you want to save and continue
retrieving Info Counter Report?
TRIMESS62 Retrieving infoCounterReportFile Cannot get hardware status, Retrieving
infoCounterReportFile is terminated.
TRIMESS63 Retrieving infoCounterReportFile Cannot compute ECV, Retrieving
infoCounterReportFile is terminated.
TRIMESS64 Retrieving infoCounterReportFile Cannot get voltages, Retrieving
infoCounterReportFile is terminated.
TRIMESS66 Name should have a value. Name should have a value.
TRIMESS67 The panel names should be unique. The panel names should be unique.
TRIMESS68 Sensitivity of panel should have a value. Sensitivity of panel <name> should have a value.

TRIMESS69 Sensitivity is too low Sensitivity is too low (minimum is <value>).


TRIMESS70 Sensitivity is too high Sensitivity is too high (maximum is <value>).

Edition 1, Revision 2
11-2020 Livelink ID: 68274899 TRIMESS
TRIMESS
(Error messages from PIXIUM Detector (DR10s, DR14s, Pixium Rad 4343))

Status 2020-10-29
Error Code Summary Message Instruction
TRIMESS71 Unexpected sensitivity Unexpected sensitivity. Change it and try again.

TRIMESS72 Panel Type should have a value. Panel Type of panel <name> should have a
value.
TRIMESS73 Check Subnet Subnet mask do not match.
TRIMESS74 No Generator Sync Box installed. No Generator Sync Box installed.
TRIMESS77 No Panel types were found. The configuration is No Panel types were found. The configuration is
corrupt corrupt.
TRIMESS79 File does not exist. File <name> does not exist.
TRIMESS80 Calibration: Take exposure (Selftrigger) Make image <number of dose image> of <total
amount of dose image>.
TRIMESS81 Registering Panel... Registering Panel...
TRIMESS82 Connecting to the Panel (takes half a minute)... Connecting to the Panel (takes half a minute)...
TRIMESS83 Get Hardware... Get Hardware...
TRIMESS84 Compute ECV... Compute ECV...
TRIMESS85 Get voltages... Get voltages...
TRIMESS86 Starting Self Test... Starting Self Test...
TRIMESS90 Panel Type of panel <name> should have a Panel Type of panel <name> should have a
value. value.
TRIMESS91 Invalid IP Address.\nAllowed detector IP Invalid IP Address.\nAllowed detector IP
addresses: 192.1... addresses: 192.168.ww.xx or 10.0.yy.zz,\nand xx
must be lower than 100 while using the IP
addresses 192.168.ww.xx
TRIMESS92 Invalid panel IP Address Invalid panel IP Address.
TRIMESS93 Can only have up to {0} Pixium 4343 (E) panels. Can only have up to <maximum amount of Pixium
4343 (E) panels> Pixium 4343 (E) panels.

TRIMESS94 Can only have up to {0} DR14s and DR10s Can only have up to <maximum amount of Pixium
panels. EZ panels> DR14s and DR10s panels.
TRIMESS95 Panels {0} cannot share the same IP Address {1}. Panels <names of applicable panels> cannot
share the same IP Address <applicable IP
addresses>.
TRIMESS96 Set Panel Parameters Connection problem. Setting panel parameters
through Infra-Red is terminated.
TRIMESS97 Setting new SSID and password.. Setting new SSID and password...
TRIMESS98 Set Panel Parameters Failed setting new SSID and password.
TRIMESS99 Setting new panel IP address and NX IP Setting new panel IP address and NX IP
address...\n address...
TRIMESS100 Set Panel Parameters Failed setting new panel IP address and NX IP
address.
TRIMESS101 Cannot Save NX IP address (<configured NX IP Address>) is
not configured on PC.
TRIMESS102 Mistmatch Serial Number Serial number of the panel (<serial number of the
panel>) mismatches the serial number of the
configuration (<serial number in configuration>).

TRIMESS103 Connection Failure Connection problem. Getting panel parameter


through Infra-Red is terminated.
TRIMESS104 Connection Failure Failed connecting to the panel through Infra-
Red.\n\nCheck:\n - If the panel is powered on.\n -
If the communication port is set correctly.\n - If the
IR transceiver is placed in front of the panel's IR
interface.
TRIMESS105 Duplicated Serial Number Duplicated serial number, Panel (<serial
number>) already configured in another tab.
TRIMESS107 Invalid SSID and PSK Invalid SSID and PSK key.\n\nSSID: Required 5 -
20 charactors\t[A-Z] [a-z] [0-9] [-_]\nPSK:
Required 8 - 20 charactors [A-Z] [a-z] [0-9] [-_]
TRIMESS108 Missing panel IP address Missing panel IP address for <panel serial
number>.
TRIMESS109 Subnets Do Not Match Panel <Panel Serial number>: panel IP address
(<Panel IP Address>) and NX IP address (<NX IP
Address>) are in different subnet.

TRIMESS110 Get Info Succeeded Serial number: <Retrieved serial number>.


TRIMESS111 Missing Serial Number Cannot save without panel serial number.
TRIMESS112 IP Address Conflict IP address <IP address> for Panel <serial
number> is already configured as a local network
connection.
TRIMESS113 Channel Settings Please select at least one channel.
TRIMESS115 Check beam path Please check that there are no unwanted objects
in the beam path.

Edition 1, Revision 2
11-2020 Livelink ID: 68274899 TRIMESS
TRIMESS
(Error messages from PIXIUM Detector (DR10s, DR14s, Pixium Rad 4343))

Status 2020-10-29
Error Code Summary Message Instruction
TRIMESS116 Shock warning DR detector <Panel Name> received a shock.
<Time Stamp>
TRIMESS117 Shock Error DR detector <Panel Name> received an Please be aware that this might impact the image
excessive shock. <Time Stamp> quality. If this occurs a re-calibration might be
needed. If the problem persists, please contact
your Service Technician.
TRIMESS118 Shock information of DR detector Since last connection with NX, DR detector Please be aware that this might impact the image
<panel name> had <amount of shocks> shock(s). quality. If this occurs a re-calibration might be
The last shock occurred on <date & time of last needed. If the problem persists, please contact
shock>. your Service Technician.
TRIMESS119 DR detector detected. Updating Network settings. DR detector detected. Updating Network settings. Please hold the DR Detector in front of the
InfraRed device until the settings are stored.
TRIMESS121 IR Communication interrupted Infrared Communication interrupted.
TRIMESS122 Unknown Device tries to connect via IR. Unknown Device tries to connect via infrared.
TRIMESS123 It is not possible to write Network settings to the It is not possible to write Network settings to the
detec... DR detector while a thumbnail is active.
TRIMESS124 Calibration: Recovery image at start calibration An image is present on DR detector <panel
name>.\nThis will be stored on disk and will be
recovered at the next start-up of NX.
TRIMESS125 Configuration Error Detector type mismatch between TRI tool and NX
configuration tool for detector <panel name>.

TRIMESS126 Get Wifi Status... Get Wifi Status...


TRIMESS127 Retrieving infoCounterReportFile Cannot get Wifi status. Retrieving
infoCounterReportFile is terminated.
TRIMESS128 Invalid Timeout Value Invalid Time Out Sleep and/or Time Out Shut
Down values for Panel {0}.\n\nAcceptable range:
15 min to 99 min.
TRIMESS129 Set Network Parameters on Panel (IR) Set (IR) cannot be performed without saving the
Panel settings.\n\nDo you want to save and
continue on Set (IR)?
TRIMESS133 Panel temp high DR Detector <detector name>'s temperature is
high (&gt; <warning temperature> degrees
Celsius).\nImage quality might be impacted.
TRIMESS135 Calibration Temp Lower Limit DR Detector <detector name>'s temperature is
lower than the Calibration Temperature range (&lt;
<Lower limit Calibration Temperature> degrees
Celsius).\nImage quality might be impacted.

TRIMESS136 Calibration Temp Upper Limit DR Detector <detector name>'s temperature is


higher than the Calibration Temperature range
(&gt; <Upper Limit Calibration Temperature>
degrees Celsius).\nImage quality might be
impacted.
TRIMESS137 Configuration Error In this configuration it is not allowed to use
integration time <used integration time>. Please
select a lower or higher value if possible.

TRIMESS139 Remove all panels from sight Please remove all DR Detectors away from the
InfraRed device.
TRIMESS140 DR Detector <DR Detector name> is in front of Please remove all DR Detectors away from the DR Detector <DR Detector name> detected by
the InfraRed InfraRed device. the InfraRed device.
TRIMESS141 Place DR Detector in front of InfraRed device. Place the DR Detector in front of the InfraRed
device.
TRIMESS142 The DR10s DR14s Registration Tool can not be The DR10s DR14s Registration Tool can not be
used while N... used while NX is not running.
TRIMESS143 The DR10s DR14s Registration Tool can not be The DR10s DR14s Registration Tool can not be
used in curr... used in current configuration.
TRIMESS144 Calibration: Dose message Set dose to <dose>µGy, kV: <kV>, mAs:
<mAs>.\nmAs value (SID =100) is indicative and
can vary to reach the indicated dose.\nThe
average SID for normal operation should be
set.\n\nIf the indicated values are not reachable,
please modify kV to reach the needed dose.

TRIMESS145 Temperature Warning DR Detector <detector name>'s temperature is


lower than the Functional Temperature range (&lt;
<Lower limit Temperature> degrees
Celsius).\nContinuing to operate can impact
Image Quality and Functionality of the DR
Detector and is under responsibility of the
operator.
TRIMESS146 PC clock tampered The clock of the PC has been turned back with Please restore PC clock to correct date and time
more than 24 hours. in order to work with the EPS panel.\nPlease
reboot panel and try again.
TRIMESS147 Serial number not found on dongle The serial number of the EPS panel <EPS panel Check your dongle configuration on the Agfa EPS
serial number> is not found on the dongle. portal and correct it.\nRestart panel and try
again.\nIf problem persists, call service.
TRIMESS148 EPS dongle not attached to PC The dongle containing the EPS license is not Please attach the dongle and restart panel.
attached to the NX workstation.

Edition 1, Revision 2
11-2020 Livelink ID: 68274899 TRIMESS
TRIMESS
(Error messages from PIXIUM Detector (DR10s, DR14s, Pixium Rad 4343))

Status 2020-10-29
Error Code Summary Message Instruction
TRIMESS149 EPS license is expired The EPS license for panel <panel serial number> Renew the EPS license on the Agfa EPS portal
has expired. and reboot panel.
TRIMESS150 No valid EPS license on dongle The EPS license on the dongle for panel <panel Reboot panel to try again. If problem persists,
serial number> is invalid. reprogram dongle on Agfa EPS portal.\nIf
problem persists, call service.
TRIMESS152 EPS license about to expire The EPS license for panel <panel serial number> Please renew the EPS license on the Agfa EPS
will expire within <days> days. portal and reboot panel.
TRIMESS153 No combination with EPS panel allowed EPS panel cannot be combined with other panels. Check configuration. Make sure only 1 EPS-panel
is configured.
TRIMESS154 Check Prep Delay The configured Generator Prepare Delay is <prep If this is not the case, let Service correct it in the
delay> seconds. For a successful calibration, this TRI panel installation tool.
value\n must match the actual prep delay of the
generator.

TRIMESS155 Update of EPS license needed Please reboot the panel in order to update its
EPS license.
TRIMESS156 Detector disconnected Detector <detector name> unexpectedly In case you experience problems with this
disconnected and reconnected again. detector,\n please restart detector and NX
completely.
TRIMESS157 Exposure aborted Exposure aborted (system reacted too slow). Please try again. Call service if problem persist.
Patient was not exposed.
TRIMESS161 Synchronization failed Synchronization with panel failed. Check Wifi communication status. If the problem
persists, please contact your Service Technician.

TRIMESS162 Synchronization failed Synchronization with panel failed. Check connection status. If the problem persists,
please contact your Service Technician.

TRIMESS163 Technical license does not support DTS Panel(s) <panel iD> configured with DTS Please check license.
functionality, but no
NX_TOMOSYNTHESIS_GENRAD license is
present.
TRIMESS164 Firmware does not support DTS Panel(s) <panel iD> configured with DTS Please check firmware.
functionality, but installed firmware does not
support DTS.
TRIMESS165 Combination of EPS and DTS is not possible Panel(s) <panel iD> is an EPS panel. It is
impossible to configure an EPS panel with DTS
functionality.
TRIMESS166 Battery is too low The battery of the DR detector <panel> is too low. Please charge panel.
The DTS exposure can not start.
TRIMESS167 Wifi is too weak The Wifi of the DR detector <panel> is too weak. Please optimize Wifi conditions.\nIf problem
The DTS exposure can not start. persists, call Service.
TRIMESS168 IP Address Conflict: Panel and GSB have the Panel <panel iD list> cannot have same IP
same IP address address as Generator Sync Box.
TRIMESS169 Not possible to make DTS exposure It is not possible to make a DTS exposure when
panel is configured in Selftrigger mode.
TRIMESS170 Make DTS exposure Make a DTS exposure. Keep expose button pressed until all images are
made.
TRIMESS171 IP Address Conflict: DTS panels configured and DTS panels configured and IP address GSB is Please check GSB configuration.
IP address GSB is empty empty.
TRIMESS172 Subnet Conflict: GSB and DTS Panel in different GSB and DTS panel should be configured in the Please configure GSB and DTS panel in the
IP subnet same IP subnet same IP subnet
TRIMESS173 Test Panel Connection Test Panel Connection cannot be performed
without saving the Panel settings.\n\nDo you want
to save and continue the Panel Connection Test?

TRIMESS174 Panel Connection result Connection failed. Panel <name> not connected.

TRIMESS175 Configuration Error Mismatch actual GSB IP address and configured Please run TRI service tool again.
GSB IP address in TRI Tool.
TRIMESS176 Calibration: Average pixel value too low Calibration step failed. \nAverage pixel value of Use higher dose.\nPlease check the "Calibration
<aveage pixel value>. \nTarget average pixel Key User Manual" for exposure
value is <target average pixel value>. parameters.\nMake sure that the correct modality
position is selected on the softconsole.

Edition 1, Revision 2
11-2020 Livelink ID: 68274899 TRIMESS
VDIMESS
(Error messages from ViewWorks Detector (DX-D40, DX-D45, DX-D60))

Status 2020-10-29
Error Code Summary Message Instruction
VDIMESS1 Calibration: Acquiring Dark images Acquiring dark image <number of dark image> of
<Total number of dark images>.
VDIMESS2 Calibration: Preparing for exposure Preparing for exposure...
VDIMESS3 Calibration: Take Exposure Take exposure <number of bright image> of
<Total number of bright images>.
VDIMESS4 Calibration: Processing Data Processing Data. Step <number of current step>
of <total numbers of steps>.
VDIMESS5 Calibration: Average pixel value too low Calibration step failed. \n Use higher dose.\n
Average pixel value of <average pixel value>. Please check the "Calibration Key User Manual"
\nTarget average pixel value is <target average for exposure parameters.
pixel value>.
VDIMESS6 Calibration: Average pixel value too high Calibration step failed. \n Use lower dose.\n
Average pixel value of <average pixel value>. Please check the "Calibration Key User Manual"
\nTarget average pixel value is <target average for exposure parameters.
pixel value>
VDIMESS7 Calibration: Uniformity Test failed Uniformity Test failed. Check collimation: the panel should be fully
covered.\n Check beam path for objects.
VDIMESS8 License Error License error: license <TECHNICAL- Please check the license file.
LICENSENAME> for
<COMPONENT/APPLICATION> not found.
VDIMESS9 Version not supported A matching license has been found, but there is a
license version mismatch.\nThis installation will be
aborted.
VDIMESS10 <Software name + version> Setup There are no matching licenses found. This
installation will be aborted.
VDIMESS11 No user permission You do not have enough privileges to install this
product. Administrator privileges are
needed.\nThis installation will be cancelled.
VDIMESS12 VDI Configuration File Updated The VDI Configuration File was successfully
created.
VDIMESS13 Software Conflict Can't start <Tool Name> while other NX software STOP NX completely before launching <Tool
is running. Name>.
VDIMESS14 Software Conflict Can't start <Tool Name> while other software is
running.
VDIMESS15 Software Conflict Can't start <Tool Name1> while another instance
of <Tool Name2> is running.
VDIMESS16 No Administrator Rights <Tool name> is only available for Administrator
accounts.
VDIMESS17 Calibration Files not valid Calibration Files not valid for panel <Panel Please perform calibration. If the problem
Name>. persists, please contact your Service Enigneer.
VDIMESS18 Logging Retrieved The logging of panel <panel's name> was
successfully retrieved.
VDIMESS19 No Connection No connection with panel <panel's name>.
VDIMESS20 Performing Offset Calibration An offset calibration is running for panel <panel
name>. This might take a few seconds.
VDIMESS21 Wrong integration time The selected integration time of <selected Please contact your Service Responsible to
integration time> does not match the integration correct the configuration.
time of <integration time configured in the VIVIX
tool> configured in the VIVIX tool.
VDIMESS23 "Performing SeflDiagnosis" Panel <Panel Name> is performing a <Progress status>
SelfDiagnosis.
VDIMESS24 "SelfDiagnosis Succeeded" The SelfDiagnosis for Panel <Panel Name> was
successful.
VDIMESS25 "SelfDiagnosis Failed" The SelfDiagnosis for Panel <Panel Name> Please contact your Service
returned errors. Enigneer.\n<feedback of Selfdiagnosis Test>.
VDIMESS26 Calibration: Failed to save calibration data file Failed to save the <Calibration Name> data file. Please contact your Service Engineer.

VDIMESS27 Calibration: Calibration step started Starting <Calibration Name>.


VDIMESS28 Calibration: Calibration failed - generic <Calibration Name> failed.\nReason: <Reason>

VDIMESS29 Calibration: Dose message Make image <number of dose image> of <total Target Average Pixel Value is <Target Average
amount of dose image>. Pixel Value>.
VDIMESS40 "Calibration: Please wait 5 minutes after last Last exposure before starting calibration should Please close the calibration tool and re-start the
exposure" be more than 5 minutes ago. tool after 5 minutes.
VDIMESS41 Wrong Firmware The firmware on panel <panel name> is not Please update the firmware to a compatible
compatible with the installed VDI Software version.
<version number>.
VDIMESS42 Save Changes Do you want to save changes?
VDIMESS43 Saved The PanelSharingConfiguration is successfully To use the PanelSharingConfiguration on other
written to \n<Path + file name> systems, copy this file.
VDIMESS44 Remove room? Are you sure that you want to remove room
<room name>?
VDIMESS46 "Wifi settings for room <RoomName> are written Wifi settings for room <RoomName> are written
successful... successfully to panel <PanelSerialNumber>.

VDIMESS48 "Linking error" There are problems writing Wifi settings for room
<RoomName> to panel <PanelSerialNumber>
\nPlease try again

Edition 1, Revision 2
11-2020 Livelink ID: 68274899 VDIMESS
VDIMESS
(Error messages from ViewWorks Detector (DX-D40, DX-D45, DX-D60))

Status 2020-10-29
Error Code Summary Message Instruction
VDIMESS49 No panels connected No panels connected. If you want to link a panel, please connect the
panel first via wire.
VDIMESS50 Clipping level out of range Clipping level is out of range [0, 16383].
VDIMESS51 Calibration: Detector Connection Failed DR Detector Connection <Detector name> Check if the DR Detector is powered on. If not,
Failed. please power on the DR detector.\nCheck if the
DR detector is connected to this workstation.\nIf
not, link the DR detector to this workstation (and
restart calibration tool).
VDIMESS52 Panel temp high DR Detector <detector name>'s temperature is
high (&gt; <warning temperature> degrees
Celsius).\nImage quality might be impacted.
VDIMESS53 Non-unique room name One or more fields have input validation issues. Room Name: Unique names must be used.
VDIMESS54 Performing SeflDiagnosis Panel <panel name> is performing a
SelfDiagnosis step <Number of current step> of
<Total number of steps>. Result: <Result>
VDIMESS55 Not an AGFA dedicated panel Panel <Panel name> is not an AGFA panel. Can
not continue.
VDIMESS56 Problem Image timeout (probably due to bad WIFI Please improve the WIFI connection.
connection).
VDIMESS57 Calibration: Uniformity Test failed Uniformity Test failed \nPlease see image and Please click 'Yes' to proceed with partial imaging
check beam path for objects.\nWarning: Partial or 'No' to try again.
Calibration is allowed for veterinary licenses, but
the image quality of partially calibrated panel is
not optimal outside the calibrated area. The
minimal ROI to be used for partial calibration is
10x12 cm
VDIMESS61 Clipping level out of range Clipping level is out of range [0, 65535].

Edition 1, Revision 2
11-2020 Livelink ID: 68274899 VDIMESS
VRNMESS
(Error messages from Varian Detector (DX-D10, DX-D20, DX-D Fixed DR Detector))

Status 2020-10-29
Error Code Summary Message Instruction
VRNMESS1 Take calibration exposure no. <exposure> of Take calibration exposure no. <exposure> of
<maxexpo> exposures <maxexpo> exposures
VRNMESS2 Ready to calibrate panel Ready to calibrate panel <panelID> for mode
<modeID>
VRNMESS3 Mode calibrated DR-detector successfully calibrated for mode:
<modeID1>.\nReady to calibrate mode
<modeID2>
VRNMESS5 Backup and dark corrections Backup is created.\nDark corrections in progress.

VRNMESS6 Calibration-Exposure detected Exposure <exposure> detected of <maxexpo>


exposures.
VRNMESS7 Processing started Processing started!
VRNMESS8 Calibration dose too low Calibration failed. Dose too low. Increase dose with <more> percent.
VRNMESS9 Calibration dose too high Calibration failed. Dose too high. Decrease dose with <less> percent.
VRNMESS10 Prep timeout Prep-timeout, no exposure is possible. Please release prep-button; you can expose
again after one second.
VRNMESS11 License Error License error: license <TECHNICAL- Please check the license file.
LICENSENAME> for <COMPONENT
VRNMESS12 No exposure found Calibration failed. Dose too low or not exposed. Increase dose.
VRNMESS13 Exposure timeout during calibration Timeout error during calibration: too much time Re-start calibration.
elapsed before next exposure.

Edition 1, Revision 2
11-2020 Livelink ID: 68274899 VRNMESS