You are on page 1of 2

Status

Passed
Service Required
Removed From Use

Health Devices IPM System


Mobile High-Efficiency-Filter Air Cleaners IPM No. 475-20010301 (Acceptance)

Facility: Dept:
WO No.: Date: Inspector:
Mfr: Model: SN:
CTRL No: Loc: IPM Time: _____.____hr

SYSTEM COMPONENTS
CTRL No. Description

SPECIAL PRECAUTIONS

Because air filtration units can harbor potentially infectious organisms, technicians should
follow universal precautions for infection control (e.g., wearing respirators approved by the
National Institute for Occupational Safety and Health [NIOSH], wearing gloves), especially
during prefilter or HEPA filter replacement or any other internal work on these units. Rubbing
or shaking filters can aerosolize trapped dust and thereby create an infection risk. Some units
incorporate special features or designs to contain the filter and minimize the potential for
spread of contamination. Used filters should be treated as infectious waste. Most filters are
disposable; reusable filters can be disinfected with a soaking spray of phenolic disinfectant or
undiluted bleach. (For more details on infection control, see IPM Safety.)

The internal surfaces of the mobile air filtration unit should be disinfected before internal
cleaning or maintenance. Refer to the manufacturer's recommendations for unit disinfection
procedures; most recommend gas disinfection using paraformaldehyde vapors. This type of
disinfection is best done outdoors.

If the mobile air filtration unit contains UVGI lamps, eye protection may be needed for some
procedures. Otherwise, ensure that the lamps are not activated whenever they can cause
personnel exposure.

TEST APPARATUS
Control/Serial No.
Leakage current meter or electrical safety analyzer
Ground resistance ohmmeter
Airflow-measuring instrument (e.g., flow hood, anemometer, pitot
tube) and appropriate ductwork
Differential pressure gauge capable of measuring 2 inches of water
in 0.01-inch increments
Particle counter capable of measuring 0.3 μm particles (optional;
required for CDC testing)

SUPPLIES
NIOSH-approved respirators

Health Devices Inspection and Preventive Maintenance System 2001 ECRI. All Rights Reserved.
Page 1 of 2
Procedure No. 475-20010301 Mobile High-Efficiency-Filter Air Cleaners (Acceptance)
WO No.:____________

SUPPLIES
Gloves
Eye protection (from UVGI lamps)
Leak test solution

QUALITATIVE TASKS
Pass Fail Pass Fail
Chassis/Housing Controls/Switches
Mount/Fasteners Fan/Blower
Casters/Brakes Indicators/Displays
AC Plug Alarms/Interlocks
Line Cord Labeling
Strain Reliefs Battery
Circuit Breaker/Fuse UVGI Lamps (if present)
Tubes/Hoses Area of Use
Filters HEPA Seal

QUANTITATIVE TASKS
Set/Indicated Measured Pass Fail
Grounding Resistance (0.5 )
Chassis Leakage Current (300 µA)
Airflow (80% of manufacturer specification)

NOTES

Health Devices Inspection and Preventive Maintenance System 2001 ECRI. All Rights Reserved.
Page 2 of 2