You are on page 1of 8

Health Devices IPM System

INSPECTION AND PREVENTIVE MAINTENANCE


__________________________________________________________________________________________

Procedure No. 458-20010301 (Combined)


__________________________________________________________________________________________

Critical Care Ventilators


Used For:
Ventilators, Intensive Care [17-429]
Ventilators, Intensive Care, Neonatal/Pediatric [14-361]
Ventilators, Transport [18-098]

______________________________________________________________________________________

Also Called: Respirators

Commonly Used In: Critical care units, general medical/surgical units, emergency departments

Scope: Applies to all ventilators except jet ventilators, negative-pressure ventilators, portable
ventilators (see Procedure 471), and anesthesia unit ventilators (see Procedure 461)

Risk Level: High

Type Interval Time Required

Major 6 months _____ hours


Minor Not Applicable

Notes:
Interval Note: Inspection and preventive maintenance intervals should be scheduled according to the
manufacturer's recommendations, which may be related to hours of use. However, units should have
a major inspection at least every six months. Pre-use checks should be performed by a respiratory
therapist or respiratory equipment technician.
_______________________________________________________________________________________

Overview
safe operation. All these controls allow the
Mechanical ventilators are used to compensate clinician to provide better patient management,
for deficiencies in normal breathing. A ventilator even for patients with serious respiratory
may aid or augment spontaneous breathing or impairments.
may completely regulate a prescribed breathing
pattern for patients who cannot breathe for A mechanical ventilator is composed of four basic
themselves. Most modern ventilators use subsystems: the ventilator and its controls;
positive-pressure inflation of the lungs to monitors and alarms; gas supply; and patient
accomplish these functions. circuit (which includes the breathing circuit and
may include a humidifier and nebulizer). Each
The ventilator provides direct control of the subsystem requires its own inspection and
patient's ventilatory variables, as well as other preventive maintenance procedures.
variables (e.g., the concentration of inspired
oxygen), and the limits on certain variables for
Citations
Health Devices Inspection and Preventive Maintenance System 2001 ECRI. All Rights Reserved.
Page 1 of 8
Procedure No. 458-20010301 Critical Care Ventilators (Combined)

(from Health Devices if not otherwise procedures or frequencies recommended by the


noted) manufacturer.

Inadequate pressure relief in infant ventilators Many microprocessor-controlled ventilators have


[Hazard], 1983 Apr;12:150-1. self-diagnostic programs. When the ventilator's
Leaving ventilator-dependent patients hardware (e.g., solenoid valves, transducers) is
unattended [Hazard], 1986 Apr;15:102-3. checked by its own software, manual inspection
Infant ventilators [Evaluation], 1986 tasks can be eliminated.
Aug;15:219-46.
Remote alarms for ventilators and other life- Manufacturers' recommended procedures for
support equipment, 1986 Dec;15:323-4. inspection and preventive maintenance of
Microprocessor-controlled third-generation mechanical ventilators vary in both methods and
critical care ventilators [Evaluation], 1989 required accuracy. In addition, ventilation
Feb;18:59-83. modes, controls, and algorithms for calculated
Intensive care ventilators [Evaluation], 1998 variables vary greatly according to manufacturer
Sep-Oct;27:308+. and model. This procedure provides the basic
Minimum requirements for ventilator testing framework for complete ventilator inspection and
[Guidance article], 1998 Sep-Oct;27:363+. preventive maintenance. Manufacturers'
Mis-set ventilator alarms can be lethal [Hazard] recommended procedures should be added
1999, Apr;28:165. where appropriate. Several tasks, particularly
Intensive care ventilators [Update evaluation] Operational Modes, Controls, and Monitored
2000 Jul-Aug;29:247+. Parameters and Alarms, should be customized to
make a procedure that is model-specific.
Test Apparatus, Supplies,
Parts Qualitative Tasks
 Ground resistance ohmmeter (Major,
Acceptance) Chassis/Housing. Check for shipping
damage; report any damage to the
 Leakage current meter or electrical safety manufacturer, shipper, or service organization,
analyzer (Major, Acceptance) and arrange for repair or replacement.

 Lung simulator with adjustable Check that the ventilator is suitably constructed
compliance/ventilator tester (Major, to withstand normal hospital use and abuse. For
Acceptance) instance, a unit with venting on the top of the
housing or poorly protected or sealed controls
 Pressure gauge or meter with 2 cm H2O and indicators may be prone to fluid entry. (Such
design deficiencies should usually be recognized
resolution from -20 to +120 cm H2O (Major,
during prepurchase evaluation. However, if any
Acceptance) are evident, discuss corrective action with the
manufacturer. If not correctable, warn users or
 Oxygen analyzer (Major, Acceptance) take other preventive measures.)

 Various breathing circuit adapters (Major, Examine the exterior of the ventilator for
Acceptance) cleanliness and general physical condition.
Ensure that plastic housings are intact, that all
 Syringe (1.5 L) (Major, Acceptance) assembly hardware (e.g., screws, fasteners) is
present and tight. (Acceptance)
 Items required for manufacturer's tasks
(Major, Acceptance)
Chassis/Housing. Examine the exterior of
the ventilator for cleanliness and general
physical condition. Be sure that plastic housings
Procedure are intact, that all hardware is present and tight,
and that there are no signs of spilled liquids or
Before beginning an inspection, carefully read other serious abuse. (Major)
this procedure and the manufacturer's
instruction and service manuals; be sure that Mount/Fasteners. Ensure that the
you understand how to operate the equipment, assembly and weight distribution is stable and
the significance of each control and indicator, that the ventilator will not tip over when pushed
and the alarm capabilities. Also review special or when a caster is jammed on an obstacle (e.g.,
inspection and preventive maintenance line cord, threshold), as may occur during
transport. If the ventilator is designed to rest on

Health Devices Inspection and Preventive Maintenance System 2001 ECRI. All Rights Reserved.
Page 2 of 8
Procedure No. 458-20010301 Critical Care Ventilators (Combined)

a shelf, ensure that it has nonslip legs or length of 10 ft [3 m] is suitable for most
supports. applications, although 18 ft [5.5 m] has been
suggested for operating room equipment.)
If the ventilator is mounted on a stand or cart,
examine the security of the mount. Check the The cord should be of suitable quality and
mounting security of all components or attached current-carrying capacity. Hard Service (SO, ST,
monitors. (Acceptance) or STO), Junior Hard Service (SJO, SJT, or
SJTO), or an equivalent-quality cord should be
Mount/Fasteners. If the ventilator is used.
mounted on a stand or cart, examine the
condition of the mount. If it is attached to a wall Verify that the ventilator has adequate
or rests on a shelf, check the security of this protection against power loss (e.g., from
attachment. Check the mounting security of all accidental disconnection of a detachable power
components or attached monitors. (Major) cord, disconnection of the power cord from the
wall, or depleted battery if a battery-powered
device is not plugged in). Equipment having a
Casters/Brakes. Verify that the correct
detachable power cord should also have
casters have been supplied with the ventilator
adequate capture devices, cleats, or channels to
(e.g., size, correct swivel). (ECRI recommends 5
hold the cord in place. If these are absent,
in [12.7 cm] diameter casters for mobile
request that the supplier provide suitable means
ventilators to reduce shock to the unit and to
of securing the cord. Verify that the ventilator
minimize the effort required to roll the ventilator
has adequate alarms or indicators for line-power
across elevator thresholds and other uneven
loss and battery depletion and an adequate
surfaces.) Verify caster and brake operation.
battery-charging indicator. (Acceptance)
(Acceptance)

Casters/Brakes. If the ventilator moves on Line Cord. Inspect the cord for signs of
damage. If damaged, replace the entire cord or,
casters, check their condition. Verify that they
if the damage is near one end, cut out the
turn and swivel, as appropriate, and look for
defective portion. Be sure to wire a new power
accumulations of lint and thread around the
cord or plug with the correct polarity. Also, check
casters. Check the operation of brakes and
line cords of battery chargers. (Major)
swivel locks, if the ventilator is so equipped.
(Major)
Strain Reliefs. Examine the strain reliefs at
both ends of the line cord. Be sure that they hold
AC Plug. A solidly constructed, good quality
the cord securely. (Major, Acceptance)
plug with adequate strain relief is acceptable, but
the use of a Hospital Grade plug (identifiable by
a green dot and/or labeling) will eliminate Circuit Breaker/Fuse. If the ventilator
guesswork and ensure a plug of acceptable has a switch-type circuit breaker, check that it
construction quality. Right-angle plugs are moves freely. If the ventilator is protected by an
unacceptable for devices that are moved external fuse, verify that the fuse type is labeled
frequently. A good quality two-prong plug is and that all fuses and spares are the proper
acceptable for double-insulated devices. Replace current rating and type. If the value and type are
the plug or have the supplier replace it if it is not not labeled, check the manual for the proper
Hospital Grade or otherwise suitable. Hospital current rating and type and permanently mark
Grade molded plugs are acceptable. this information on the ventilator housing near
the fuse holder.
Examine the AC power plug for damage. Attempt
to wiggle the blades to determine if they are Especially for critical or life-support devices,
secure. Shake nonmolded plugs and listen for verify that accessory outlets have independent
rattles that could indicate loose screws. overcurrent protection (fuse or circuit breaker)
(Acceptance) so that a short in a device plugged into the
accessory outlet or an accessory overload will
not disable the primary device. If this is not
AC Plug. Examine the AC power plug for
available, then consider labeling the primary
damage. Attempt to wiggle the blades to check
device to clearly indicate where the ventilator's
that they are secure. Shake the plug and listen
fuse or circuit breaker is located, and/or install a
for rattles that could indicate loose screws. If
fused Hospital Grade (or similar quality) plug on
any damage is suspected, open the plug and
any commonly used accessories that are not
inspect it. (Major)
already provided with suitable overcurrent
protection. (Acceptance)
Line Cord. Ensure that the line cord is long
enough for the ventilator's intended application;
an extension cord should not be required. (A

Health Devices Inspection and Preventive Maintenance System 2001 ECRI. All Rights Reserved.
Page 3 of 8
Procedure No. 458-20010301 Critical Care Ventilators (Combined)

Circuit Breaker/Fuse. If the ventilator Filters. Check the condition of gas filters.
has a switch-type circuit breaker, check that it Check for corrosion residue indicative of liquid,
moves freely. If the ventilator is protected by an gaseous, or solid particle contaminants in the
external fuse, check its value and type against gas supply; advise appropriate personnel if
that marked on the chassis and ensure that a found. Clean or replace if appropriate, and
spare is provided. (Major) indicate this on the inspection form. (Major)

Tubes/Hoses. Check the condition of all Controls/Switches. Verify that setup


tubing and hoses. They should be clean and parameters accessible through hidden or service
intact. (Major, Acceptance) menus are correctly set for the appropriate
application and are consistent with other units of
Cables. Inspect any cables and their strain the same model. Instruction and service manuals
reliefs for general condition. (Acceptance) may contain instructions regarding such menus.
If they do not, contact the manufacturer. Discuss
appropriate settings with the department head
Cables. Inspect any cables and their strain
and users. If alarm capabilities are included, see
reliefs for general condition. Carefully examine
the Acceptance test for Alarms.
cables to detect breaks in the insulation and to
ensure that they are securely gripped in the
Examine all controls and switches for physical
connectors at each end, which will prevent
condition, secure mounting, and correct motion.
rotation or other strain. Where appropriate,
If a control has fixed-limit stops, check for
verify that there are no intermittent faults by
proper alignment, as well as positive stopping.
flexing cables near each end and looking for
(Acceptance)
erratic operation or by using an ohmmeter.
(Major)
Controls/Switches. Before changing any
controls or alarm limits, check their positions. If
Fittings/Connectors. Verify appropriate
any settings appear inordinate (e.g., alarm limits
connectors are supplied if connection to other
at the ends of their range), consider the
hospital equipment or systems is required.
possibility of inappropriate clinical use or of
Ventilators that connect to the central piped
incipient device failure. Investigate questionable
medical gas system should have the matching
control settings on a home care ventilator.
DISS or quick-connect fitting for the appropriate
Consult with the patient's physician to determine
gas. Verify that suitable connectors are supplied
correct settings. The patient or caregiver should
with the ventilator so that adapters are not
receive additional training, if required. Record
required.
the settings of those controls that should be
returned to their original positions following the
Examine all gas and liquid fittings and
inspection.
connectors, as well as all electrical cable
connectors, for general condition. Electrical
Examine all controls and switches for physical
contacts should be straight, clean, and bright.
condition, secure mounting, and correct motion.
Gas and liquid fittings should be tight and should
Check that control knobs have not slipped on
not leak. If keyed connectors are used (e.g., pin-
their shafts. Where a control should operate
indexed gas connectors), ensure that no pins are
against fixed-limit stops, check for proper
missing and that the keying is correct. Keying
alignment, as well as positive stopping. Check
pins should be securely seated in "blind" holes so
membrane switches for damage (e.g., from
that they cannot be forced in farther.
fingernails, pens). During the inspection, be sure
(Acceptance)
to check that each control and switch performs
its proper function. (Major)
Fittings/Connectors. Examine all gas
fittings and connectors for general condition. Gas
Pressure-Relief Mechanism. Check the
fittings should be tight and should not leak.
proper operation of the pressure-relief
Verify that keyed connectors (e.g., pin-indexed
mechanism, if so equipped, by occluding the
gas connectors) are used where appropriate,
breathing circuit and observing the resulting
that all pins are in place and secure, and that
peak pressure on the unit's pressure indicator.
keying is correct. Connectors to hospital central
Verify that pressure is vented and does not
piped medical gas systems should have the
exceed manufacturer specifications. (Major,
appropriate DISS or quick-connect fitting to
Acceptance)
eliminate the need for adapters. (Major)

Filters. Ensure that all gas filters are supplied. Fan/Compressor. Check the physical
condition and proper operation of these
Check their condition. (Acceptance)
components, if so-equipped. Check for automatic
activation of the compressor when the piped gas

Health Devices Inspection and Preventive Maintenance System 2001 ECRI. All Rights Reserved.
Page 4 of 8
Procedure No. 458-20010301 Critical Care Ventilators (Combined)

supply pressure falls below operating pressure. confirm that it does, in fact, charge the battery.
(Acceptance) Be sure that the battery is recharged or charging
when the inspection is complete. When it is
Fan/Compressor. Check the physical necessary to replace a battery, label it with the
condition and proper operation of these date. (Major)
components if so-equipped. Check for automatic
activation of the compressor when the piped gas Indicators/Displays. Confirm the
supply pressure falls below operating pressure. operation of all lights, indicators, meters,
Clean or replace fan and/or compressor filters gauges, and visual displays on the ventilator and
and lubricate as required, according to the charger (if so equipped). Be sure that all
manufacturer's instructions. (Major) segments of a digital display function. Record
reading of an hour meter, if present. (Major,
Power Continuity. [Delete this task if the Acceptance)
ventilator operates on battery power.]
Alarms. Verify that configurable alarm
To ensure constant ventilation when line power features are appropriately set and consistent
is interrupted, ventilators should automatically among all ventilators. It should not be possible
switch to battery operation. If the unit is not for critical alarms to be turned off, silenced, or
equipped to do this, it should be connected to an defeated without adequate warning to the
uninterruptible power supply (UPS). A UPS, operator or automatic alarm reactivation after a
however, is inconvenient, can be lost or short delay. Such deficiencies should usually be
overlooked, and must be accounted for in recognized during prepurchase evaluation.
inventory control and equipment maintenance. A However, if any are found, review the
UPS should be periodically checked to ensure justification for purchasing this ventilator and
proper operation. Because of the added expense discuss corrective action with the manufacturer.
and the other disadvantages inherent in using a (Alarm features may be optional or
UPS, ventilators should be purchased with programmable.) If no remedy is available, a user
internal batteries when possible. (Acceptance) training program should be instituted to reduce
the risk of incorrect use. A warning label on the
Battery/Charger. Determine the ventilator or a poster in the area of use may be
replacement interval for all batteries and appropriate. If the ventilator has an alarm-
document the interval(s) in an equipment control silence feature, check the method of reset (i.e.,
system or a hard copy file for the ventilator. Be manual or automatic) against the manufacturer's
sure to include batteries/cells for clocks and/or specifications. (Acceptance)
memory logs. Disconnect the battery and
determine that the ventilator still operates on Audible Signals. Verify that alarms are
line power. loud, distinctive, and/or bright enough to be
noticed in the environment in which the
Operate the ventilator on battery power for ventilator will normally be used. If a remote
several minutes to verify that the battery is alarm-indicator is required, verify that it is
charged and can hold a charge. Activate the available and functioning. Audible alarm-volume
battery test function, if so equipped. Check the controls should not allow the alarm to be turned
condition of the battery charger, and verify that off or lowered to an indiscernible volume.
battery charge indicators function. Provide users (Acceptance)
with instructions and/or checklist procedure to
ensure adequate battery charging and Operational Modes. [Note: This task
performance. (Acceptance) should be customized into specific tasks for each
mode, which will create corresponding task titles
Battery/Charger. Inspect the physical on the inspection form.]
condition of batteries and battery connectors if
readily accessible. Check operation of battery- Verify proper operation of the following modes
operated power-loss alarms, if so equipped. found on most ventilators:
Operate the ventilator on battery power for
several minutes to check that the battery is Control
charged and can hold a charge. (The inspection Assist/control
can be carried out on battery power to help Synchronized intermittent mandatory ventilation
confirm adequate battery capacity.) Check (SIMV)
battery condition by activating the battery test Pressure support
function or measuring the output voltage; for Continuous positive airway pressure/positive
lead-acid batteries, measure the specific gravity end-exhalation pressure (CPAP/PEEP) (Major,
and check the fluid level. Check the condition of Acceptance)
the battery charger and, to the extent possible,

Health Devices Inspection and Preventive Maintenance System 2001 ECRI. All Rights Reserved.
Page 5 of 8
Procedure No. 458-20010301 Critical Care Ventilators (Combined)

Labeling. Check that all necessary placards, and exposed (unpainted and not anodized) metal
labels, and instruction cards are present and on the chassis. We recommend a maximum of
legible. (Major, Acceptance) 0.5 . If the system is modular or composed of
separate components, verify grounding of the
mainframe and each module or component.
Accessories. Verify that all necessary
(Major, Acceptance)
features and accessories have been supplied with
the ventilator. Verify that the breathing circuit
and filters are compatible with the ventilator Chassis Leakage Current. Note: Some
according to the manufacturer's devices (especially devices incorporating a
recommendations. Check for leaks in the microprocessor, motor, or compressor) may be
breathing circuit, ensuring that fittings, adapters, damaged by switching polarity while the device
and other components (e.g., exhalation valves, is on. If you perform reverse polarity testing,
H-valves, PEEP valves, water traps, nebulizers) turn off the ventilator until the motor stops or for
are properly assembled and functioning at least 10 sec before switching polarity.
correctly. Although reversed polarity testing is not
required, it may be advisable on a ventilator of
At least one copy each (two are generally questionable quality or on ventilators used in the
preferred) of the instruction and service home.
manuals, including schematics, should be
shipped with the unit and filed in the central With the polarity of the power line normal and
equipment file. A copy of the operator's manual the equipment ground wire disconnected,
should be kept with the ventilator and read by all measure chassis leakage current with the
operators before the ventilator is put in use. ventilator operating in all normal modes,
(Acceptance) including on, standby, and off. If the ventilator
has a UPS, readings should be taken with the
UPS connected. Record the maximum leakage
Accessories. Confirm the presence and
current; it should not exceed 300 A for patient-
check the condition of accessories. Verify that
care equipment used in a patient-care vicinity
the breathing circuit and filters are compatible
[NFPA 99 7-5.1.3.5(a)].
with the ventilator according to the
manufacturer's recommendations. Check for
The measurements should be made with all
leaks in the breathing circuit, ensuring that
accessories that are normally powered from the
fittings, adapters, and other components (e.g.,
same line cord connected and turned on. This
exhalation valves, H-valves, PEEP valves, water
applies to devices that are plugged into
traps, nebulizers) are properly assembled and
accessory outlets on the device and to devices
functioning correctly.
that are plugged into a multiple-outlet strip so
that the devices are grounded through a single
Oxygen-air proportioner. See Procedure 444.
line or extension cord.
(Major)
Leakage current must be measured with the
Compressor. Test according to the ventilator powered by a conventional (grounded)
manufacturer's recommendations. (Major, power system, even if it is normally used in an
Acceptance) area with isolated power. If the ventilator has a
special plug (e.g., explosion proof), a
Gas Cylinders, Gauges, and corresponding adapter is required.
Regulators (for transport
ventilators). Verify that these components Though confirmation of grounding integrity
provides reasonable assurance of safety for
are present, securely mounted, and in good
devices with permanent redundant grounding
condition and that there is adequate gas supply.
(e.g., a bedside monitor grounded through the
(Major, Acceptance)
line cord and its central station connection),
NFPA 99 calls for measurement of chassis
Humidifiers. See Procedure 431. (Major, leakage current with the redundant ground intact
Acceptance) [NFPA 99 7-5.1.3.5]. Where practical, verify that
these devices meet appropriate chassis leakage
current requirements before installation or
Quantitative Tasks connection to ground during acceptance testing.

Be alert for leakage current of the ventilator in


Grounding Resistance. Using an
the off mode that is greater than about 30 A
ohmmeter, electrical safety analyzer, or
and is greater than or equal to the leakage
multimeter with good resolution of fractional
current in the on mode. Although this may be
ohms, measure and record the resistance
normal and proper for the ventilator, it may
between the grounding pin of the power cord

Health Devices Inspection and Preventive Maintenance System 2001 ECRI. All Rights Reserved.
Page 6 of 8
Procedure No. 458-20010301 Critical Care Ventilators (Combined)

indicate that the on/off switch is incorrectly volume from the product of the test lung and
wired in the neutral (instead of the hot) line. breathing circuit compliance [C] and the PIP [V =
Incorrect switch wiring poses a risk to service C  PIP]. To determine the compliance, deliver a
personnel who believe that the power is set volume [same volume that the ventilator is
disconnected when the switch is off. Check the set to deliver] to the breathing circuit and test
wiring, or contact the manufacturer. lung with a large syringe, and record the
(Acceptance) resultant change in pressure at the inlet of the
test lung. The compliance is the delivered
Chassis Leakage Current. Measure volume divided by the recorded pressure.) The
chassis leakage current to ground with the measured volume should be within 10% of the
grounding conductor of plug-connected set tidal volume.
equipment temporarily opened. Operate the
ventilator in all normal modes, including on, Record the number of breaths delivered during a
standby, and off, and record the maximum 1-minute period. Verify that the measured rate is
leakage current. If the ventilator has a UPS, within 1 breath/min of the set respiration rate
readings should be taken with the UPS (may be ±2 breaths/min at high set rates).
connected. (Major, Acceptance)

The measurements should be made with all Monitored Parameters and Alarms.
accessories normally powered from the same line [Note: This task should be customized into
cord connected and turned on and off. This specific measurement tasks for each monitored
includes other equipment that is plugged into the parameter and alarm. This will create
primary device's accessory receptacles, as well corresponding data fields on the inspection
as equipment plugged into a multiple-outlet strip form.]
so that all are grounded through a single line or
extension cord. Chassis leakage current to Verify proper operation and accuracy of
ground should not exceed 300 A. (Major) monitored parameters and their associated
alarms, which may include:
Controls. [Note: This task should be
customized into specific measurement tasks for Breathing rate
each control function. This will create Inspiratory time
corresponding data fields on the inspection Peak inspiratory pressure (PIP)
form.] Peak inspiratory flow
Positive end-expiratory pressure (PEEP)
Verify proper operation and accuracy of Mean airway pressure (MAP)
ventilation controls, which may include: Volume (both tidal and minute volume)
Fraction of inspired oxygen (FIO2); see Oxygen
Tidal volume Analyzer Procedure 417
Respiration rate Temperature of inspired air
Inspiratory time
Expiratory time Attach the ventilator to a lung simulator or
Inspiratory:expiratory (I:E) ratio ventilator tester and verify that displayed values
Flow are within 10% of simulated values.
Trigger sensitivity
100% oxygen delivery Induce alarm conditions to activate audible and
Sigh visual alarms. Verify alarm messages on
Manual breath displays. If the ventilator has an alarm-silence
Waveshape feature, check the method of reset (i.e., manual
or automatic) against the manufacturer's
Verify the operation and accuarcy of ventilation specifications. Confirm appropriate volume, as
controls and indicators. These tests are typically well as the operation of a volume control, if so
performed by attaching the ventilator to a lung equipped. If audible alarms have been silenced
simulator or ventilator tester and comparing or the volume set too low, alert clinical staff to
measured values to settings on the ventilator. the importance of keeping alarms at the
Follow manufacturer recommendations for appropriate level. If the ventilator is used with a
appropriate ventilator settings at which to verify remote alarm, verify that the alarm and indicator
proper operation and accuracy. Measured values function properly. Alarm activation (e.g., for high
should typically be within 10% of settings. PIP, low pressure, low FIO2) should also be
within 10% of set values. (Major, Acceptance)
Record the tidal volume measured by the volume
monitor at the outlet of the exhalation valve or
the PEEP valve, if used. (Volume can also be
confirmed by calculating the delivered tidal
Preventive Maintenance
Health Devices Inspection and Preventive Maintenance System 2001 ECRI. All Rights Reserved.
Page 7 of 8
Procedure No. 458-20010301 Critical Care Ventilators (Combined)

Clean. Clean the exterior, interior, and


components if needed. (Major)

Calibrate. Calibrate according to the


manufacturer's instructions. (Major)

Replace. Replace components according to


the manufacturer's instructions. (Major)

Before Returning to Use


Ensure that all controls are set properly. Set
alarms loud enough to alert personnel in the
area in which the ventilator will be used. Other
controls should be in their normal pre-use
positions. If the ventilator is being used at home,
ensure that controls are set correctly before it is
returned to the patient.

Attach a caution tag in a prominent position so


that the user will be aware that control settings
may have been changed.

Recharge battery-powered ventilators, or equip


them with fresh batteries, if needed.

Health Devices Inspection and Preventive Maintenance System 2001 ECRI. All Rights Reserved.
Page 8 of 8

You might also like