Professional Documents
Culture Documents
1. Practical Workshop
The workshop presents a primed Aquarius in recirculation mode. Each group will be supported to
practically connect and run a brief therapy simulation. 10 minute discussion on Troubleshooting -
common alarms (high and low priority), teaching appropriate observation skills and responses.
Assess vascular catheter for Check patient access. Restart the blood pump.
occlusions/clots. Reposition the patient.
Access is against the vessel
wall.
Reposition the catheter and/or
line.
Swap the lumens over.
Consider aspirating the lumens.
Consider pulling the catheter
back.
Consider a new catheter.
Review anti-coagulation.
Rotate the blood pump ¼
anti-clockwise.
Access line is kinked/clamped. Remove the kink/unclamp the Restart the blood pump.
line
Tubing set is kinked close to Check access line, including Restart the blood pump.
access or pre-filter pressure access and pre-filter sensors, for
sensors. kinks or occlusions.
High Access Pressure Consider anticoagulant and Restart the blood pump.
(Access pressure exceeeds replacement fluid review.
alarm limit of Check patient access.
between -50 and +200mmHg) Reposition the patient.
Access is against the vessel wall.
Reposition the catheter and/or line.
Consider aspirating or swapping the
lumens.
Consider pulling the catheter back.
Consider a new catheter.
Review anti-coagulation.
Remove the kink/unclamp the line
Low return pressure Blood flow rate is too low. Consider increasing the blood Restart the blood pump.
(Return pressure is below the pump speed.
lower alarm limit of 20 to The blood pump has stopped.
200mmHg) Restart the blood pump.
Return line is occluded or clotted. Reposition the line. Restart the blood pump.
Consider swapping the lumens
Drip chamber is clotting. Consider anticoagulant and Restart the blood pump.
replacement fluid administration
review.
Ensure adequate blood level in the
drip chamber.
Prepare to end the therapy.
.
Faulty return pressure transducer. Possible faulty transducer Prepare to end treatment
High pre filter pressure Check filter for signs of clotting Access line is “kicking” Restart the blood pump.
(The pre filter pressure exceeds Visible clots in the filter
the upper alarm limit of Visible clots in return chamber
400mmHg) Check clotting
Consider Increasing pre-dilution
3 consecutive high pre filter
pressure alarms, consider ending
the treatment.
Air detected Air or micro-foam visible in the drip Remove air from the drip chamber To remove air from the line:
chamber or Return line. and Return line.
Step1. Attach syringe to the
The return line is not properly Ensure Return line tubing is not top of the return chamber
positioned in the air detector. “pinched” within the air detector. after carefully releasing the
pressure from the line.
The blood level is too low in the drip Raise the blood level in the drip Step 2. Press the “clamp
chamber. chamber. key” to open the return line
clamp.
Air detector sensors are dirty. Clean the sensors.
Marks on the return line. Ensure line is not pinched within the
air detector.
Check all connections, including
access, filter and Heparin Step 3. Remove all air from
Ensure all connections are secure the return chamber with the
connections for sources of air leaks.
and patent. syringe.
Ensure both seals on the Accusol Step 4. Confirm the tubing
bag/s are broken, bag/s are spiked, is correctly inserted into the
all clamps are open, lines are not air detector
trapped or kinked. Step 5. If level in the drip
chamber is correct and the
The air detector is faulty. bubbles are out of the
tubing, press the “clamp
key” to close the return line
clamp.
Step 6. Resume treatment
by pressing the blood pump
key.
Step 7. Monitor and control
the level in the drip
chamber
Balance alarm Counted balance alarm: All clamps are open. Restart the treatment
Check substitution/dialysate
line The patient’s fluid balance Lines and bags are hanging
deviates more than 50 g in adult freely.
treatment or 20 g in pediatric
Or
treatment for more than 15s. Lines and bags are not kinked,
The deviation is less than 120 g clamped or blocked.
Balance alarm
Check filtration/effluent line The deviation could not be Bag connections are correct Note: If 5 counted
compensated during TFL balance alarms occurred
Bags and lines are not resting on during the 20 minute
Review UF variation in More the cart frame window period it will not
screen. be possible to resume
Bags are not swinging. the treatment. Follow the
onscreen instructions to
Lines on the manifold are all end the treatment.
unclamped.
If no further alarms are
Accusol bags are spiked. shown for a period of 20
minutes, the balance
Seals are broken on the Accusol alarm counter will reset to
bags. zero.
Filter membrane is damaged / ruptured. Filtrate / plasma contains blood. Select Options and End
(Note: Ultrafiltrate typically is a treatment, follow the on-
rose colour) screen instructions to wash
Check BLD % reading in MORE back and end the treatment.
screen.
Consider doing a urine dipstick to
confirm the presence of blood.
Discontinue treatment and
change circuit.
High TMP TMP has risen slowly – filter is Reduce post-dilution flow rate and Restart the treatment
(-50 to +450mmHg) clogging. increase pre-dilution flow rate.
Review anti-coagulation.
TMP has risen rapidly – filtrate line Check blood flow / exchange ratio.
or bags clamped or kinked. Increase blood flow rate accordingly. .
High TMP from the start of therapy Unclamp or remove kink from line.
Low TMP Filtrate pump runs slower than the Modify blood flow rate and/or fluid Restart the treatment
treatment pumps. exchange.
The filtrate line is closed between Check filtrate line is not kinked,
the filter/kidney and waste/effluent clamped or twisted. .
bag/s. Line is not “pinched” within the pump
housing.
Pressure test disabled Air detection system does not detect Make sure that the return line Restart the pressure test.
“free air” in the tubing. does not contain air.
(The clamp and pressure test is disabled.)
Make sure that the return line is
properly installed in the air
detector and that the air
detector is pushed in.
No fluid chamber detected The degassing chamber (ADU) is not Ensure the ADU chamber is Restart the treatment
inserted. sitting on the pressure
switch. (Black and red
switch)
Check Degassing
Chamber The ADU worked for more than 25s Check all clamps on the To remove fluid from the sensor
(Position: Short line towards the without detecting fluid in the chamber. substitution/dialysate line line (long line with green
back and long line at the front.) are unclamped. clamp):
The hydrophobic filter on the ADU Check the Step1: Clamp the line.
sensor line is wet. substitution/dialysate line is
(Measured pressure < –ve 300 mmHg). not kinked. Step2: Disconnect line from the
sensor and hold the line up.
The system detects a positive pressure If using a manifold ensure
higher than +30 mmHg. lines are not clamped or Step3: Attach a 10ml syringe,
kinked. unclamp the green clamp and
Fluid is detected in the ADU sensor push air into the line until the
line. Check if the frangible pins line is clear of any water.
on the bags are broken.
Step4: Clamp the line again,
Ensure both seals on the remove the syringe and attach
Accusol bag are broken. line to the sensor.
Change The substitution/dialysate bag/s are Change the bag/s. Restart the treatment
substitution/dialysate bag nearly empty (approx. 150mls left per bag)
OR
The filtrate bag has reached maximum
Change the bag/s
Change filtrate/effluent bag permissible weight
Check transducer The pressure domes have not detected Ensure the domes are Restart the treatment
connections any pressure change for 15 seconds. securely connected.
IMPORTANT: do not
remove any pressure
sensors during treatment.
If domes are in place,
consider increasing blood
pump or treatment pump
rates.
Turnover Failure Fluid leaking from bags. Ensure all connections are Restart the treatment.
(Balance pump speeds are securely attached and
consistently higher or lower than patent.
the programmed speed for more Incorrect line installation
than 20 consecutive minutes to Ensure lines are not kinked,
ensure accurate fluid delivery.) clamps are unclamped, and
lines are not twisted.
Incorrect bag installation.
Bag/s is not spiked, spike or
frangible pin is blocking the
flow of fluid, bag/s is not
hanging freely, or bags are
Pump calibration out of range swinging.
Heater cools down Balance system has stopped for more Ensure the balance system Active cool down management
than 15 s and the heater plate is active. may take up to 10 minutes. The
temperature is above 43°C. treatment will restart
The treatment is paused automatically.
until the temperature
returns to normal.
System Test Failure Ensure both pump doors are closed Close the door/s. Press the blood pump to restart
the self test.
Heparin Pump The plunger is incorrectly positioned. During Self test/Start up, If the error cannot be reset, put
(Actual values of control and ensure no line set is system out of operation and
protective systems are outside of installed notify technical support.
their limits.)
During treatment:
1. Check the heparin line is
not clamped.
2. Go to "Options" and
"Change Syringe", following
the on-screen text. Note: It
is NOT necessary to
remove the syringe during
this process
3. If problem persists,
program the pump to 0,
clamp the line and remove
the syringe.
Heparin driver is making a grinding Ensure Heparin line is not
sound clamped.
No Bag No bag/s on the waste/filtrate scale. Place bag/s on the scale. Restart the treatment pumps
(Less than 45g are detected per
scale.) Bag weighs less than 45g. Use a bag weighing more
than 45g. Note: It is not
recommended to re-use
used/empty Accusol bags.
Lines or the manifold is touching the
frame of the machine Ensure lines and the
manifold is not touching the
frame of the machine.
Master key transfer A key press longer than 60 seconds The alarms are to notify the Clear the alarms by pressing
was detected user that an issue the blood pump key.
Keyboard Failure occurred).
Short disturbances in the
communication between the master
Operation Mode
and the controller CPU.
Blood Flow Failure The number of revolutions of the pump Check the blood flow rate Press the blood pump to restart
exceeds or falls below the alarm limits the treatment.
by ± 5%. Check blood pump tubing is
not twisted/pinched within
the pump housing
Notes:
Pre – course exercise: Please make notes on one of the following subjects
Consider the ‘state of the art’ for anticoagulation in your PICU: How do we choose Heparin,
No Anticoagulation or an alternative modality?
Citrate and Predilution, what we could learn from experience in Adult ICU, and where might
this lead our patient care?
The decision tree is a visual aid to map your thoughts