Abstract— This study aims to determine an optimum intensity and type of activity [5] and the detection of falls
estimate for the gravitational vector and vertical acceleration [6], using body-worn tri-axial accelerometers (TA) has thus
profiles using a body-worn tri-axial accelerometer during falls increased dramatically in the last 20 years. Within these
and normal activities of daily living (ADL), validated using a algorithms the measurement or estimation of kinematic
camera based motion analysis system.
quantities such as horizontal and vertical, acceleration,
Five young healthy subjects performed a number of
simulated falls and normal ADL while trunk kinematics were velocity and displacement as well as body segment angle
measured by both an optical motion analysis system and a tri- and rotation measurement, have been made using
axial accelerometer. exclusively accelerometer.
Through low-pass filtering of the trunk tri-axial The total acceleration measured by an accelerometer
accelerometer signal between 1Hz and 2.7Hz using a 1st order during human movement can be described as the sum of the
or higher, Butterworth IIR filter, accurate gravity vector acceleration vector due to linear and rotational movement of
profile can be obtained using the method described here.
the attached body or body segment and the acceleration
Results: a high mean correlation (≥0.83: Coefficient of
Multiple Correlations) and low mean percentage error
vector due to gravity [7]. For the kinematic quantities
(≤2.06m/s2) were found between the vertical acceleration profile mentioned, it is thus desirable to determine an accurate
generated from the tri-axial accelerometer based sensor to gravitational vector component from the TA signal vector,
those from the optical motion capture system. with respect to the local geographic frame.
This proposed system enables optimum gravity vector and Within the accelerometer signal, the frequency spectra of
vertical acceleration profiles to be measured from the trunk the gravitational component overlaps with that of the linear
during falls and normal ADL.
and rotational acceleration component [8].
Previously a number of researchers have estimated that
I. INTRODUCTION
filtering of the TA signal would provide an accurate estimate
Falls only
Oc 1 2 3 4 5 6
fc (Hz) 2.2 2.05 2.15 2.3 2.5 2.75
CMC 0.908 0.909 0.910 0.910 0.910 0.909
delay (s) 0.057 0.110 0.148 0.181 0.206 0.224
Fig. 2. Method for estimating vertical acceleration using the TA signal.
In order to obtain the most appropriate filter cut-off All activities
frequency and order, the RMS and CMC values between the Oc 1 2 3 4 5 6
fc (Hz) 2.2 2.05 2.15 2.25 2.5 2.7
vertical acceleration signal from the optical MA system CMC 0.843 0.844 0.844 0.844 0.844 0.844
(av,MA) and those produced using the method described with delay (s) 0.057 0.110 0.148 0.185 0.206 0.228
the TA signal (av,TA) were calculated for cut-off frequencies Table 1: Filter cut-off frequencies (fc) and signal delays for the
(fc) from 0.1Hz to 5Hz with a 0.05Hz interval and filter maximum CMC values for; ADL only, Falls only and All activities.
order (Oc) from 1 to 6. The combination of frequency (fc) Thus it is clear that a wide range of values between
and order (Oc) that gave the smallest RMS and a CMC 0.85Hz (Oc=1) and 2.75Hz (Oc=6) can be used depending
closest to 1, provided the optimal low-pass filter parameters. on the type of activity being measured if the technique
described previously is used. A summary of the minimum
IV. RESULTS CMC values and maximum RMS values for different
A comparison between the vertical acceleration profiles selected ranges of fc and Oc is presented in Table III.
TABLE III
from the TA sensor and the MA system was made using ADL only Falls only All activities
RMS error and CMC values; these are presented in Table I Oc range 1-6 1-6 1-6
and II, respectively, a summary is presented in Table III. fc range 0.85-2.5Hz 1-2.75Hz 1-2.7Hz
CMC ≥0.7906 ≥0.895 ≥0.8373
From the minimum recorded RMS error values from RMS ≤1.24m/s/s ≤3.083m/s/s ≤2.058
orders 1 to 6, the optimum filter cut-off frequencies ranged Table 1: CMC minimum and RMS maximum values for filter cut-off
from 0.85Hz to 1.45Hz for ADL only, 1.85Hz to 1Hz for frequency ranges (fc range) and filter order ranges (Oc range).
Falls only and from 1.85Hz to 1Hz for all activities which An illustration of the vertical acceleration produced from
includes both falls and ADLs. The Lowest RMS error for both the optical motion analysis system and that produced
ADL only was produced using an fc from 1.1Hz to 1.45Hz using the TA signal in presented in Fig. 3. The av,TA was
7898
produced using an fc of 1.85Hz, which represents an average the gravitational vector and vertical acceleration profile from
filter cut-off frequency for the ranges presented in table III. a TA signal from falls and normal ADL separately and in
On Oc of 2 was chosen as higher orders introduce larger combination, validated using an optical MA system.
delays and increased computation requirements if
implemented in an embedded system. VII. ACKNOWLEDGMENTS
We are deeply indebted to all participants who so willing
V. DISCUSSION AND CONCLUSION gave of their time. Also to the projects CAALYX (IST-
This study aimed to determine an accurate method, 2006-045215) [19] and to eCAALYX for their support [19].
including filter cut-off frequencies and order, for
determining an estimate of the gravitational vector and REFERENCES
vertical acceleration profile from a TA signal, for both falls [1] UN, "Population Ageing 2009," Population Division, Department of
and ADL. Using a dot-product projection between the TA Economic and Social Affairs, United Nations, 2009.
[2] P. E. Cotter, S. Timmons, M. O'Connor, C. Twomey, and D.
signal and a low-pass filtered, delayed version of the TA
O'Mahony, "The financial implications of falls in older people for an
signal, accurate vertical acceleration signals were obtained acute hospital," Ir J Med Sci, vol. 175, pp. 11-3, Apr-Jun 2006.
and validated using an optical MA system; thus producing [3] Recent demographic developments in Europe: Council of Europe
an optimum gravitational vector estimate. Publishing, 2004.
[4] P. Bonato, "Wearable sensors and systems. From enabling technology
to clinical applications," IEEE Eng Med Biol Mag, vol. 29, pp. 25-36,
May-Jun.
[5] A. Godfrey, R. Conway, D. Meagher, and O. L. G, "Direct
measurement of human movement by accelerometry," Med Eng Phys,
vol. 30, pp. 1364-86, Dec 2008.
[6] N. Noury, P. Rumeau, A. K. Bourke, G. ÓLaighin, and J. E. Lundy, "A
proposal for the classification and evaluation of fall detectors," IRBM,
vol. 29, pp. 340-349, Dec 2008.
[7] K. J. O'Donovan, R. Kamnik, D. T. O'Keeffe, and G. M. Lyons, "An
inertial and magnetic sensor based technique for joint angle
measurement," J Biomech, vol. 40, pp. 2604-11, 2007.
[8] M. J. Mathie, "Monitoring and interpreting human movement patterns
using a triaxial accelerometer," in Faculty of Engineering Sydney: The
University of New South Wales, 2003.
[9] C. V. Bouten, K. T. Koekkoek, M. Verduin, R. Kodde, and J. D.
Janssen, "A triaxial accelerometer and portable data processing unit for
Fig. 3. Vertical acceleration profiles from both the tri-axial accelerometer the assessment of daily physical activity," IEEE Trans Biomed Eng,
sensor (black line) and motion analysis system (red line) for both falls and vol. 44, pp. 136-47, Mar 1997.
ADL; (A) sitting on a chair, (B) object picking, (C) lying on the floor, (D) [10] J. Fahrenberg, F. Foerster, M. Smeja, and W. Muller, "Assessment of
walking, (E) forward fall with knee flexion and (F) a side-fall right with posture and motion by multichannel piezoresistive accelerometer
knee flexion. The gravity vector was obtained with fc=1.85Hz and Oc=2. recordings," Psychophysiology, vol. 34, pp. 607-12, Sep 1997.
Measurement of the gravitational vector can be achieved [11] F. Foerster and J. Fahrenberg, "Motion pattern and posture: correctly
with the addition of a tri-axial rate gyroscope sensor [18]. assessed by calibrated accelerometers," Behav Res Methods Instrum
Comput, vol. 32, pp. 450-7, Aug 2000.
However, the aim of this study was to achieve gravitational [12] D. M. Karantonis, M. R. Narayanan, M. Mathie, N. H. Lovell, and B.
vector estimation using only a tri-axial accelerometer as a G. Celler, "Implementation of a real-time human movement classifier
number of human movement monitoring system only using a triaxial accelerometer for ambulatory monitoring," IEEE Trans
employ this sensor due to it cost and power efficiency. Inf Technol Biomed, vol. 10, pp. 156-67, Jan 2006.
[13] M. Mathie, "Monitoring and Interpreting Human Movement Patterns
An alternative to method presented, is to high-pass filter Using a Triaxial Accelerometer," in Faculty of Engineering. vol.
the TA signal, account for the delay, and subtract if from the Doctor of Philosophy Sydney: The University of New South Wales,
original TA signal. This was also attempted however inferior 2003, p. 512.
[14] F. Ferraris, U. Grimaldi, and M. Parvis, "Procedure for Effortless In-
results were obtained. Field Calibration of Three-Axis Rate gyros and Accelerometers.,"
The filter cut-off frequencies presented here are higher Sensors and Materials, vol. 7, pp. 311-330, 1995.
than those presented in previous literature where separation [15] J. C. Lötters, J. Schipper, P. H. Veltink, W. Olthuis, and P. Bergveld,
"Procedure for in-use calibration of triaxial accelerometers in medical
of the effect of gravity from the TA signal was performed. applications," Sensors and Actuators A: Physical, vol. 68, pp. 221-228,
The filter cut-off frequencies recommended in those studies 1998.
were also examined here, however higher RMS error values [16] A. K. Bourke, K. J. O'Donovan, and G. Olaighin, "The identification of
vertical velocity profiles using an inertial sensor to investigate pre-
and lower CMC values were produced. impact detection of falls," Med Eng Phys, Jan 31 2008.
To produce an acceleration signal, without the influence [17] K. Manal and W. Rose, "A general solution for the time delay
of gravity from a TA, the gravitational vector profiles can introduced by a low-pass Butterworth digital filter: An application to
musculoskeletal modeling," J Biomech, vol. 40, pp. 678-81, 2007.
now be accurately, delayed and subtracted from the original [18] D. Titterton and J. Weston, Strapdown inertial navigation technology,
TA signals and the Root-sum-of-squares calculated. 2nd ed., 2004.
[19] M. N. Boulos, R. C. Lou, A. Anastasiou, C. D. Nugent, J.
VI. CONCLUSION Alexandersson, G. Zimmermann, U. Cortes, and R. Casas,
"Connectivity for healthcare and well-being management: examples
This study determined an accurate method, including filter from six European projects.," Int J Environ Res Public Health, vol. 6,
cut-off frequencies and order, for extracting and estimating pp. 1947-71, 2009.
7899