Professional Documents
Culture Documents
nice 5010
High-ow blender with
(30+30) cascade ow meter (optional)
nice 5005
Air Oxygen Blender for
Infant 'T' Piece Resuscitator & Humidier
To Provide continuous supply of Air and
Oxygen Blender Gas for Bubble CPAP
and T Piece Resuscitator.
Unique Features
nice 5005 and nice 5010 have gas savings ON/OFF bleed feature.
Available in very low to high ow capability.
Suited to specialty and general needs.
Mounting Solutions for all situations.
All blender are equipped with an audible alarm.
Applications
OxyM Medical Air/O2 blenders combine compressed Medical Air and Oxygen to deliver blended pressurized gas at a
precise oxygen concentration (FIO2) rm by the user.
All Device equipped with a unique gas bleed ON/OFF switch to increase accuracy when needed, conserve gas,
save money and time.
These blenders are suitable for respiratory applications including routine therapy, ventilator gas and supply, Bubble CPAP,
SiPAP, nCPCP Resuscitator and critically -limited NICU procedures.
nice Neotech offer different models with multiple Outlet ports that deliver the same highly accurate selected FiO2.
Most models can be customized to include the owmeter attachment with a variety of ow rates available.
The blenders contains an audible alarm which warns the user if either of the gas sources changes by more than
20 PSI from the other.
Air and Oxygen blenders require a Bleed for good accuracy at low ow range less than 15 lpm. The Bleed ON
a blender is activated by the knob, which is placed on the right port. This facility avoids the blended gas being
into the air when the blender is not in use.
(0 to 3 lpm)
88-00-139
Flow meter
(optional)
O2 Pressure Regulator Air Pressure Regulator
Air Inlet Hose
(optional) (optional)
50-05-131 50-05-130 (0 to 15 lpm) (optional) 50-20-171
88-00-105
Specications
nice 5005 nice 5010
Oxygen Concentration Range 21-100% 21-100%
FiO2 Accuracy _
+3% of full scale _
+3% of full scale
Alarm Activation bypass _ PSI
When supply pressure differs 20+2 _ PSI
When supply pressure differs 20+2
Alarm Type Reed alarm, 80db minimum at 1 foot Reed alarm, 80db minimum at 1 foot
Primary Outlet Flow Range (left
Up to 30 LPM Up to 120 LPM
side Outlet)
Auxiliary Outlet Flow Range (Right
side Outlet) 0-30 LPM 2-120 LPM
2
Bleed Flow at 50 PSI (3.5kg/cm ) 2.5 to 3.5 LPM 10 to 12 LPM
Inlet Gas Supply Pressure 30 to 75 PSIG (2.1 to 5.3kg/cm2) 30 to 75 PSIG (2.1 to 5.3kg/cm2)
DISS (Standard) DISS (Standard)
Gas Inlet Fittings
NIST (Optional) NIST (Optional)
Operating Temperature 15 to 40°C 15 to 40°C
Storage Temperature -20 to 60°C -20 to 60°C
Operating Humidity 15 to 95% noncondensing 15 to 95% noncondensing
Storage Humidity 0 to 95 % noncondensing 0 to 95 % noncondensing
MDD - Product Classication Class llb Class llb
Quality Management System ISO 13485 ISO 13485
Device Specic standard ISO 11195 ISO 11195
Dimension (Excluding ttings) 9cm (H) X 6cm (W) X 9.5cm (D) 9cm (H) X 6cm (W) X 9.5cm (D)
Weight 1.25 kg 1.25 kg
nice 5020
Number of ports Two Two
T - Piece Resuscitator
with Blender
74-00-021, Rev 03