• SUBJECT
When switching on immediately after shutting down a MyLab15 or MyLab20 with PC Module Basic
installed (required for Just 4D), the start up sequence may, occasionally, not be completed.
• DESCRIPTION
The embedded optional PC Module Basic has an internal power protection. This prevents the PC Module
Basic from starting up directly after shutting down.
• ACTION
When switching off a MyLab15 or MyLab20 with a PC Module Basic installed, wait 5 seconds before
restarting the unit.
0344