You are on page 1of 1

Technical note MyLab15 & MyLab20 CLASSIFICATION

Part PC Module Basic Mandatory


Document # 21A35X601 Recommended
Training level N/A • Performance
Page 1 of 1 Documentation
Date of issue September 13, 2006 Installation
Issued by Perry van Haaren
Checked by Marco Lagustena

• SUBJECT
When switching on immediately after shutting down a MyLab15 or MyLab20 with PC Module Basic
installed (required for Just 4D), the start up sequence may, occasionally, not be completed.

• DESCRIPTION
The embedded optional PC Module Basic has an internal power protection. This prevents the PC Module
Basic from starting up directly after shutting down.

• ACTION
When switching off a MyLab15 or MyLab20 with a PC Module Basic installed, wait 5 seconds before
restarting the unit.