Technical note MyLab15/20 CLASSIFICATION

Part Backend software Mandatory


Document # 21A35X401 Recommended
Training level N/A • Performance
Page 1 of 2 Documentation
Date of issue March 24, 2006 Installation
Issued by Steven Romberg
Checked by Arno Dabekaussen

• SUBJECT
Video printer problems.

• DESCRIPTION
Sometimes the synchronization of the video signal is bad which results in a double image print. A new
backend software release (C: 3.1.0#1) has been issued to resolve this problem.

Examples:

• ACTION

To check the software version and backend firmware versions, switch on the unit and press the MENU key on
the keyboard. Now select SYSTEM ADMINISTRATION.
The following screen is displayed: (see next page)

Pie Medical Equipment 0344


Technical note MyLab15/20 CLASSIFICATION
Part Backend software Mandatory
Document # 21A35X401 Recommended
Training level N/A • Performance
Page 2 of 2 Documentation
Date of issue March 24,2006 Installation

The correct backend software version is C:3.1.0#1

If the software level of the involved MyLab is lower than C:3.0, update all files. For further installation
instructions, please consult software news document number 21A27SN05.

In case the software level of the involved MyLab is C:3.0, but the backend software version is C:3.0.0#1
update the backend software only. The last version is available from the Maastricht Product Support site.
Backend_Software_C3_1.ZIP contains the required file ‘backend.bin’. Unzip the file to a flash drive and install
it to the MyLab15 or MyLab20. Detailed installation instructions are described in software news MyLab15
document number 21A27SN05.

For more information or assistance contact the Esaote Pie Medical helpdesk.
helpdesk@esaote-piemedical.com
tel. +31 43 3814200
fax +31 43 3824601

Pie Medical Equipment 0344