You are on page 1of 48

NIDEK

GREEN LASER PHOTOCOAGULATOR Model GYC-1000

BINOCULAR INDIRECT OPHTHALMOSCOPE DELIVERY UNIT

[Keeler All Pupil II type]

OPERATOR’S MANUAL
(SUPPLEMENT TO GYC-1000 OPERATOR’S MANUAL)
NIDEK CO., LTD. : 34-14, Maehama, Hiroishi-cho, Gamagori, Aichi 443-0038, Japan
(Manufacturer) Telephone: (0533) 67-6611
Facsimile: (0533) 67-6610
NIDEK CO., LTD : 6th Floor, Takahashi Bldg., No.2, 3-chome, Kanda-jinboucho
(Tokyo Office) Chiyoda-ku, Tokyo 101-0051, Japan
Telephone: (03) 3288-0571
Facsimile: (03) 3288-0570
Telex: 2226647 NIDEK J
NIDEK INCORPORATED : 47651 Westinghouse Drive Fremont, California 94539, U. S. A.
(United States Agent) Telephone: (510) 226-5700
Facsimile: (510) 226-5750
NIDEK SOCIETE ANONYME : Europarc 13, rue Auguste Perret, 94042 CRETEIL, France
(Authorized Representative) Telephone: (01) 49 80 97 97
Facsimile: (01) 49 80 32 08
2005.3
17120-P902B
Printed in JAPAN
BEFORE USE OR MAINTENANCE OF THE PHOTOCOAGULATION
SYSTEM, READ THIS MANUAL AND OPERATOR’S MANUAL OF THE
MAIN BODY.

THIS OPERATOR’S MANUAL CONTAINS ONLY INFORMATION TO


UNDERSTAND THE OPERATING PROCEDURE AND MAINTENANCE
OF THE BINOCULAR INDIRECT OPHTHALMOSCOPE DELIVERY
UNIT (Keeler All Pupil II type).

The Operator’s Manual contains information necessary for the operation of the
photocoagulation system that is comprised when the NIDEK GREEN LASER
PHOTOCOAGULATOR Model GYC-1000 (main body) is connected with the
BINOCULAR INDIRECT OPHTHALMOSCOPE (hereafter called B.I.O.) DELIVERY
UNIT (Keeler All Pupil II TYPE). Refer to the applicable operating manual for the operating
procedures of photocoagulation using other devices.
This manual includes the operating procedures, cautions for safety, specifications and
maintenance instructions.
This manual complies with IEC 60601. This manual and the operator’s manual for the main
body are required to correctly use the photocoagulation system.
Especially, the cautions for safety and operating procedures must be thoroughly
understood before using the instrument. Keep this manual handy to verify use whenever
necessary.
Use of this system is limited to the treatment of eye disease by qualified physicians
only.
The physicians are responsible for the application of this photocoagulation system and
the treatment of various eye diseases.
If you encounter any problems or have questions about the instrument, contact NIDEK
or your authorized distributor.

[Note on display of the exposure time and the interval time (in the repeat mode)]
When the exposure time or the interval time (in the repeat mode) on the display of the control box of
the GYC-1000 is less than “1.00” or “1.0”, “0” before the decimal point is displayed small. (Example:
if the setting is “0.30”, the display shows “ ”.)
Table of Contents

§1 INTRODUCTION .......................................................................................................Page
1-1
1.1 Outline of the B.I.O. Delivery Unit .............................................................................. 1-1
1.2 Indications for Use ..................................................................................................... 1-1
1.3 Classifications of the B.I.O Delivery Unit ..................................................................... 1-2
1.4 Symbol Information .................................................................................................... 1-3
§2 SAFETY ......................................................................................................................... 2-1
2.1 Storage, Transport, and Installation Precautions ........................................................... 2-1
2.2 Wiring and Connection Precautions ............................................................................. 2-2
2.3 Usage Precautions ...................................................................................................... 2-3
2.4 After Use Precautions, Maintenance, and Checks ........................................................ 2-5
2.5 Disposal ..................................................................................................................... 2-5
2.6 Safety Devices ........................................................................................................... 2-6
2.7 Nominal Ocular Hazard Distance (NOHD) ................................................................. 2-6
2.8 Labels ........................................................................................................................ 2-7
§3 DEVICE DESCRIPTION ............................................................................................. 3-1
§4 INSTALLATION AND STORAGE ............................................................................. 4-1
4.1 Installing the System ................................................................................................... 4-1
4.2 Storing the System ...................................................................................................... 4-6
4.2.1 Storing the main body ........................................................................................ 4-6
4.2.2 Storing the B.I.O. delivery unit ......................................................................... 4-11
§5 OPERATING PROCEDURES ..................................................................................... 5-1
5.1 Staring the Photocoagulation System ........................................................................... 5-1
5.2 Preparing for Laser Emission ...................................................................................... 5-3
5.3 Laser Emission ........................................................................................................... 5-6
5.4 End of Operation ...................................................................................................... 5-11
5.5 Indications of Misoperation ....................................................................................... 5-12
5.6 Emergency Stop ....................................................................................................... 5-12
§6 MAINTENANCE .......................................................................................................... 6-1
6.1 Replacing the Illumination Lamp .................................................................................. 6-1
6.2 Cleaning ..................................................................................................................... 6-2
6.2.1 Cleaning the optical parts ................................................................................... 6-2
6.2.2 Cleaning other parts .......................................................................................... 6-2
6.2.3 Cleaning the fiber optic cable ............................................................................. 6-3
§7 SPECIFICATIONS AND CONFIGURATIONS ......................................................... 7-1
7.1 Specifications ............................................................................................................. 7-1
7.2 Standard Configurations ............................................................................................. 7-2
7.3 Options ...................................................................................................................... 7-2

Words in This Manual .................................................................................... End of this manual


§1 INTRODUCTION

1.1 Outline of the B.I.O. Delivery Unit


This unit is connected to the NIDEK GREEN LASER PHOTOCOAGULATOR, Model GYC-1000
(main body) to comprise the photocoagulation system and applied to the treatment of the affected area
using the B.I.O. (Keeler All Pupil II TYPE).
The comprised photocoagulation system makes it possible to switch between photocoagulation with
the green laser beam (wavelength: 532 nm) and observation with the B.I.O..

B.I.O Operator

Fiber-optic cable
antenna (option)

Fiber-optic cable

Laser emission

GYC-1000 main body


Aspheric B.I.O. lens
Patient

Foot switch

CAUTION
• United States Federal law restricts this device to sale by or on the order of a physician.

1.2 Indications for Use


The NIDEK Green Laser Photocoagulator Model GYC-1000 Binocular Indirect Ophthalmoscope is
intended to be used in ophthalmic surgical procedures, including retinal and macular photocoagulation.
1-2

1.3 Classifications of the B.I.O Delivery Unit


[Classification under the provision of 93/42/EEC (MDD)] Class IIb
As long as this delivery unit is connected to the main body, this delivery unit is classified as a
Class IIb unit.

[Class of the laser system] Class 4


As long as this delivery unit is connected to the main body, this unit is classified as a Class 4
system.
A Class 4 laser is capable of producing an injury to eyes with direct laser exposure or by
diffuse reflections (0.5 W or more).

[Protection method against electric shock] Class I


As long as this delivery unit is connected to the main body, this delivery unit is classified as a
Class I system.
A Class I system is a system in which the protection against electric shock does not rely on
basic insulation only, but which includes an additional safety precaution that provides for the
connection of accessible conductive parts to the protective (earth) conductor in the fixed
wiring of the installation in such a way that accessible conductive parts cannot become live in
the event of a failure of the basic insulation.

[Degree of protection against electric shock] Type B applied part


As long as this delivery unit is connected to the main body, this delivery unit is classified as a
system with a Type B applied part. A system with a Type B applied part provides an adequate
degree of protection against electric shock particularly regarding;
- allowable leakage currents
- reliability of the protective earth connection (if present)

[Degree of protection by the enclosure]


This delivery unit is classified as IP20.
The IP20 system is protected against an ingress of solid foreign objects, such as a finger
having a diameter of 12.5 mm or greater. However, it is an ordinary system without protection
against an ingress of liquids. Be careful not to get water on this delivery unit.

[Sterilization methods recommended by the manufacturer]


This delivery unit does not have any part to be sterilized or disinfected.

[Degree of safety in the presence of flammable anesthetics and/or flammable cleaning agents]
This delivery unit should be used in an environment where no flammable anesthetics and/or
flammable cleaning agents are present.

[Mode of operation]
As long as this delivery unit is connected to the main body, this unit is an continuous operating
system.

[Classification by transportability]
This delivery unit is classified as a transportable system.
1-3

1.4 Symbol Information


This symbol indicates that the degree of protection against electric shock is of a Type B
Applied Part.

This symbol indicates that important descriptions are contained in the operator’s manual and
that the operator must refer to the operator’s manual prior to operation or maintenance.
§2 SAFETY

In this manual, signal words are used to designate the degree or level of safety alerting. The
definitions are as follows.

WARNING: Indicates a potentially hazardous situation which, if not avoided,


could result in death or serious injury.
CAUTION: Indicates a potentially hazardous situation which, if not avoided,
may result in minor or moderate injury or in a property damage
accident.

Even situations indicated by WARNING and CAUTION may result in serious injury
under certain conditions. Safety precautions must be strictly followed at all times.

2.1 Storage, Transport, and Installation Precautions


CAUTION
• In storage, transport, and installation, verify that the following conditions are met:
- Not exposed to direct sunlight or ultraviolet rays
- Not exposed to rain or water
- No chemicals or organic solvents are present
- No poisonous gas, sulfur, salt, or large amount of dust is contained in the air
- Level and stable without vibration and shock
- The following are the specified environmental conditions for storage and transport (packed
condition), and installation (unpacked condition)
For storage and transport:
Temperature: 14 to 122 ºF (–10 to 50 ºC) / Humidity: 10 to 95 % (non-condensing)
For installation:
Temperature: 50 to 86 ºF (10 to 30 ºC) / Humidity: 30 to 85 % (non-condensing)

• In transport of the photocoagulation system, observe the following instructions:


- To avoid injury or malfunction, remove the delivery unit from the main body and store them
in the carrying case or shipping carton (keep the shipping carton used for delivery)
- To avoid shift of the optical axis, do not subject the main body or the delivery unit of the
photocoagulation system to physical shock even if they are packed in the carrying case or
shipping carton.
- To avoid condensation, keep the change in temperature as little as possible.
2-2

CAUTION
• In installation of the photocoagulation system, observe the following instructions:
- To avoid troubles from condensation, let the photocoagulation system sit until the condensation
has dissipated before installation.
- To avoid malfunction from change in temperature and condensation, do not install the
photocoagulation system where it is exposed to the direct flow of air conditioning.
- To avoid adverse effect on the lens or mirror, do not install the photocoagulation unit in a
high temperature, high humidity, or dusty environment.
- To let the photocoagulation system dissipate heat properly, install it so that the air vent on the
rear panel and the wall or the foreign object are more than 10 cm apart.

• Attach or remove the delivery unit to the main body with the key switch off ( ).
If the delivery unit is attached ore removed with the key switch on ( ), an error may occur.

2.2 Wiring and Connection Precautions


CAUTION
• In handling the power cord and cables, observe the following instructions:
- To avoid short circuit or fire from broken wire, perform the connection and disconnection
holding the plug, and do not coil the power cord and connecting cables forcefully in a short
curvature, or crush or pinch them with heavy objects.
- To avoid short circuit or fire, replace the broken wire with a new one.
- Never pull the power cord or a connecting cable to transport the system.

• In handling the power cord, observe the following instructions:


- To avoid malfunction or electric shock, use a grounded power outlet which meets the power
requirements specified in the label on the main body.
- To avoid malfunction or fire, do not overload the electrical outlet.

• In handling the illumination cable and the AC adapter cable, observe the following instructions:
- To avoid malfunction or failure of the system, connect the cable plugs to the connector at the
bottom of the wall pack securely.

• In handling the fiber optic cable, observe the following instructions:


- To avoid damage or deterioration in laser delivery performance, do not let any part of the
fiber-optic cable bend with a radius of 10 cm (4.5 in) or more.
- To avoid damage or deterioration in laser delivery performance, avoid impact (from falling
or hitting) on the fiber optic cable.
- To avoid deterioration in laser delivery performance, be careful not to soil or damage the tip
of the plug of the fiber optic cable, especially when inserting the plug into the FIBER connector.
- Connect the cable plug to the FIBER connector on the front panel of the main body.
2-3

2.3 Usage Precautions


WARNING
• Use of controls or adjustments, or performance of procedures other than those specified herein
may result in hazardous radiation exposure.

• In handling of the delivery unit, observe the following instructions:


- Only service technicians properly trained by Nidek may install and adjust the delivery unit.
Only qualified physicians may perform emission of the green laser for treatment.
- To avoid hazardous radiation exposure, do not perform operation which is not described or
different from the procedure specified in the operator’s manual.
- To avoid electric shock, do not modify nor touch the internal structure.
- To maintain the performance of laser emission, never soil or scratch the lens or mirror.

• Before starting the photocoagulation system, observe the following instructions:


- To avoid ignition or explosion from the laser emission, verify that there is no flammable
anesthetic gas in the operating room.
- To protect the eyes of personnel in the operating room, instruct them to wear recommended
safety goggles (or the equivalent) before the operation and not to look at the green laser
beam directly during the operation.
Recommended goggles ...... Model YL-300 for frequency doubled Nd: YAG: D315-532
L8 YL DIN
(Produced by YAMAMOTO KOGAKU CO., LTD. Japan)
- To prevent accidents, perform the check following 5.3 Checks Before Use and 5.4
Function Checks (p. 5-3) in the operator’s manual of the main body and record each
result in the list (p. 5-5) when starting operation of the photocoagulation system.

• During operation of the photocoagulation system, observe the following instructions:


- To avoid unintended exposure to the laser beam, do not gaze at the aiming beam that is
emitted from the laser aperture of the delivery unit or direct it toward personnel. Always pay
attention to the direction in which it is emitted.
- To avoid accident caused by unauthorized personnel, do not leave the photocoagulation
system unattended while it is operational. If the operator has to be away from the system,
turn the key switch to the off position, remove the key, and store it in the customary place.

• During operation of the photocoagulation system, observe the following instructions:


- Do not use the photocoagulation system simultaneously with other electronic equipment to
avoid electromagnetic interference with the action of the system
- Do not use the photocoagulation system simultaneously in the same room with other equipment
such as a life-support equipment that has serious effects on the life of patient and the treatment,
or other measurement or treatment equipment that involves small signal.
- Do not use cables and accessories that are not specified for the photocoagulation system
because that may deteriorate the electromagnetic compatibility.
- Do not use the photocoagulation system simultaneously with portable radio frequency
communication systems because it may have an adverse effect on the system.
2-4

WARNING
• In use of the B.I.O. delivery unit, observe the following instructions:
- To avoid damage to the retina of the patient, only trained physician may perform observation
using the B.I.O delivery unit. Do not emit unnecessarily intense laser beam.
- To avoid the blue hazard, set the illumination for observation to the minimum level initially,
and then increase it until the desired intensity can be obtained. After the observation, lower
the illumination intensity to the minimum level again.

CAUTION
• For laser emission, observe the following instructions:
- When the green laser beam (wavelength: 532 nm) of the GYC-1000 is emitted into tissue,
following symptoms may occur. Pay attention to the direction of the aiming beam to avoid
emitting the laser beam into the eye or onto skin inadvertently.
- Skin symptoms ........... Damage to the cornea, etc. or blindness
- Eye symptoms ............ Pain or burn
- To avoid exposure to the reflected laser beam, verify that there is no reflective object in the
optical path before laser emission.
- To avoid excessive reaction, set the laser to a low output power initially, and then increase it
until the desired effect can be obtained. Always set the output power to the minimum after
the laser emission.
- To avoid exposure to the accidentally emitted laser beam, always place the laser system in
the condition in which the laser cannot be emitted (standby mode) except when emitting the
laser.
- Confirm that the photocoagulation system is in a proper condition for laser emission by the
following procedures: (if any abnormality is found in step 2, ask Nidek for check and
adjustment.
1. Project the aiming spot on a surface that is
not specular.
2. Verify that the intensity of the illumination
spot is even around the center as shown in
the figure on the right and that lowering of
the intensity or vignetting does not occur.
2-5

2.4 After Use Precaution, Maintenance, and Checks


CAUTION
• After use of the photocoagulation system, observe the following instructions:
- To maintain the performance of laser emission, when the system is not being used, turn off
the power and put dust cover over it.
- To avoid the tracking phenomenon, if the system will not be used for a long period of time,
disconnect the power cord from the grounded power outlet.

• For maintenance of the photocoagulation system, observe the following instructions:


- To avoid burn, replace the burned-out illumination lamp when it has been cooled enough.
- To avoid troubles, use the illumination lamp of the specified rating.
- To avoid reduction of life time of the lamp, do not touch the glass part of it with bare hands.
If the glass part is touched with bare hands, wipe the part with a clean cloth dampened with
alcohol.
- To avoid accidents caused by improper handling of the system, only service technicians
properly trained by NIDEK may repair the system.
- To protect the exterior or maintain the operability of the system, do not use organic or
abrasive solvents for cleaning.

• For check of the photocoagulation system, observe the following instructions:


- To maintain the performance of laser emission, ask NIDEK or your authorized distributor
for calibration of output power and exposure time of the laser beam, and for measurement
of the earth resistance and leakage current once a year.
- To avoid infection, wipe the surface of the delivery unit (especially the inside parts of the
headband and overband) with a clean cloth dampened with alcohol before returning it to
NIDEK for repair or maintenance.

2.5 Disposal
CAUTION
• When disposing of the B.I.O. delivery unit, follow the local governing ordinances and recycling
plans. (For details, ask NIDEK or your authorized distributor.)

• When disposing of packing materials, sort them by material and follow local governing ordinances
and recycling plans.
2-6

2.6 Safety Devices


[Protective filter]
To protect the operator’s eyes from the reflected light of the green laser beam, the protective
filter is built into the observation window of the B.I.O..

[Manual reset function]


After the photocoagulation system is stopped due to the stop signal from the remote connector,
a power outage, and an unintended cause, etc., the system does not restart automatically even
though all the problems are solved and the system is ready to be restarted. It is operator’s
decision whether to restart the system or not. To restart the system, turn the key switch to the
OFF ( ) position (manual reset), and then to the ON ( ) position again.

2.7 Nominal Ocular Hazard Distance (NOHD)


The Nominal Ocular Hazard Distance (NOHD) is the distance along the axis of the unobstructed beam
from the laser aperture where the exposure or irradiance falls below the applicable exposure limit.
The nominal ocular hazard distance (NOHD) of the laser beam of the B.I.O. delivery unit is as follows:
NOHD = 7.82 m (W.D. = 300 mm) - 23.47 m (W.D. = 700 mm)

* In this operator’s manual, W.D. (working distance) means the distance between the eyepiece lens
of the B.I.O. and the 20D aspheric B.I.O. lens.
2-7

2.8 Labels
In order to call the operator’s attention, the appropriate warning labels are attached to the specified
locations on the delivery unit.
2-8

This page is intentionally left blank.


§3 DEVICE DESCRIPTION
[Front of the B.I.O. delivery unit]

Illumination spot size selector Filter selector

Working distance control knob Angle fixing knob

Headband Mirror adjustment knob

Observation window

Headband Filter selector


Used to fit the delivery unit to the operator’s Used to insert the protective filter into the ob-
head. servation optical path. Pushing and pulling the
Knobs on the top and the back of the headband selector inserts the red-free filter or the wide angle
adjusts the height and tightness of it. diffuser.

Working distance control knob Angle fixing knob


Used to set the working distance and spot size. Used to fix the angle of the B.I.O. part of the
By rotating this knob, the working distance/spot delivery unit. Loosen this knob and adjust the
size can be continuously changed in the range angle to obtain the maximum visual field through
between 300 mm/185 µm and 700 mm/556 µm the B.I.O.. After the adjustment, tighten the knob
(when observed with the 20 D aspheric B.I.O. to fix the angle.
lens on a normal eye).
Mirror adjustment knob
Illumination spot size selector Used to raise or lower the direction in which the
Used to select the illumination spot size. There illumination and the laser beam are emitted. Turn-
are three sizes to select from by sliding the se- ing this knob changes the inclination of the illumi-
lector to left or right. nation mirror, and thus the direction of the illumi-
nation and the laser beam emission.
3-2
[Side of the B.I.O. delivery unit]

Height adjustment knob

Fiber-optic cable

Illumination cable

Circumference adjustment knob

Illumination control

Pupillary distance adjustment knob

Height adjustment knob Illumination control


Turned to adjust the height of the B.I.O. when Used to adjust the intensity of the illumination.
you put on the delivery unit on your head. The intensity can be changed continuously from
the extinction to the maximum.
Fiber-optic cable
Pupillary distance adjustment knob
Delivers the laser beam from the main body to
Used to adjust the pupillary distance of the eye-
the delivery unit. The plug at the end is connected
pieces. While observing the illumination spot,
to the fiber connector on the main body.
slide the lens holder to adjust the pupillary dis-
tance in the range of 47 to 75 mm.
Illumination cable
The plug at the end of this cable is connected to
the wall pack that hangs on the stand.

Circumference adjustment knob


Turned to adjust the circumference of the
headband when you put on the delivery unit on
your head.
3-3
[Other standard accessories]
Short plug for the B.I.O. delivery unit
Connected to the DELIVERY connector on the
front panel of the main body.

Aspheric B.I.O. lens (with a case)


Used in combination with the B.I.O. to examine
the patient’s eye. Face the side with a white line
on the rim toward the patient.

Carrying case
Store the B.I.O. delivery unit in this carrying case
for transfer or when it will not be used for a long
period of time.

Stand / Wall pack


* Hang the B.I.O.delivery unit on this part. The wall pack is a transformer for the illumina-
tion adjustment that doubles as a hanger. It is
hung on the hook of the stand.
When the B.I.O. delivery unit is not used, hang
the headband on the hanger.
Wall pack

Stand

[Option]

Laser goggles (option)


Have the personnel in the room except the
operator and patient wear this goggles during
operation of the photocoagulation system.
§4 INSTALLATION AND STORAGE

4.1 Installing the System


CAUTION
• Install the B.I.O. delivery unit with all the power source of the main body turned off.
Connecting the B.I.O. delivery unit to the main body while the power is turned on may
cause electric shock.

• Handle the fiber optic cable of the B.I.O. delivery unit with special care.
Not only coiling the fiber optic cable with a radius of 10 cm (4.5 in) or less, but also
physical shock may deteriorate the laser delivery performance.

1. Secure the power source and the installation place for the photocoagulation system.

1) Prepare a wagon or the equivalent that is at least 60cm wide and 40cm deep (recommended).
* Lock the casters if possible.

CAUTION
• Confirm that the surface on which the prepared wagon or the equivalent is placed is level and
stable.
Otherwise, the main body or the B.I.O. delivery unit on the stand may fall during use of the
photocoagulation system.

2) Confirm that the power outlet to be used is grounded and meets the power requirements labeled
on the main body.

CAUTION
• If the power outlet does not meet the power requirements labeled on the main body, the
system may not perform sufficiently, and be damaged. If the power outlet is not grounded
and current leakage occurs due to malfunction, electric shock may result.
4-2
2. Install the main body.

1) Take the main body out of the shipping carton.

2) Place the main body on the wagon or the equivalent prepared in step 1.
* Leave the room for the B.I.O. delivery unit.

CAUTION
• Install the main body so that its rear panel is at least 10cm apart from the wall.
Blocking the release of heat from the laser may damage the main body.

3. Prepare the B.I.O. delivery unit.

1) Take the stand and the AC adapter out of the carrying case and place them on the wagon or
another convenient place.

2) Take the B.I.O. delivery unit out of the carrying case.


View from under
the wall pack
3) Take the wall pack out of the carrying case.
Connect the plugs of the illumination cable and
the AC adapter to the wall pack. Then hang
the wall pack on the hook of the stand.
Plug of AC adapter Plug of illumination
cable

4) Hang the part around the height adjustment


knob of the headband on the hanger of the Step 4)
stand.
* When holding the B.I.O. delivery unit, hold
Step 3)
the headband with one hand while holding
the connecting cable and fiber-optic cable
with the other. Wall pack
(hanger)

B.I.O. delivery unit

Stand
4-3
4. Connect the cables to the rear panel of the main body.

1) Connect the power cord.

Confirm that the main switch on the rear panel


is off ( ). (If it is on ( ), set it to off ( ).)

Connect the female plug of the power cord


to the inlet.

Short plug

Power cord Cable of the foot switch

2) Connect the foot switch.

Place the foot switch at a convenient position on the floor.

Connect the cable plug of the foot switch to the FOOT SW connector on the rear panel.

3) Connect the short plug or plug of the external switch to the REMOTE connector.

Align the short plug or plug of the external switch with the REMOTE connector on the rear panel.

While holding both sides of the plug, insert the plug to the connector as far as it goes.

5. Connect the B.I.O. delivery unit to the main body.

1) Connect the short plug of the B.I.O. delivery Plug of the fiber-optic cable
unit to the DELIVERY connector on the control
panel.

2) Connect the plug of the fiber-optic cable to the


FIBER connector on the control panel.

Remove the cap from the end of the plug.

Align the pin of the plug with the notch of the


FIBER connector and insert the plug until a click
Short plug of the B.I.O. delivery unit
is felt.
4-4

CAUTION
• Pay attention to the handling of the fiber optic
cable so that it is not coiled in a radius of less
than 10 cm. 10 cm or more
The transmittancy of the laser beam may
be reduced or the optical fiber may be
damaged.

• Be careful not to soil or scratch the tip of the


plug of the fiber optic cable when
disconnecting the plug from the FIBER
connector.
The laser delivery performance may
deteriorate. Tip of the plug of the fiber optic cable

6. Connect the power cord and the AC adapter to the power outlet.
Connect the power cord of the main body and the AC adapter of the wall pack to the power outlet
verified in step 1.

7. Start the operation of the photocoagulation


system.

1) Turn on ( ) the master switch on the rear panel


of the main body.
The indication “ ” lights up on the
control panel.

2) Insert the key into the key switch and turn it to Master switch

the on ( ) position.
“ ” “ ”

After approximately 10 seconds of countdown,


the indications “ ” and “ ”
on the control panel lights up. Then the
photocoagulation system goes into the
STANDBY mode.

Key switch

8. Perform the function checks of the photocoagulation system.

Perform the check referring to [5.4 Function Checks] (p. 5-3) in the Operator’s Manual of the
main body.
* Record each result in [5.5 Function Check List] (p. 5-6) in the same Operator’s Manual.
4-5
9. Hang the B.I.O. delivery unit on the stand.

Hold the headband of the B.I.O. delivery unit with one hand while holding the connecting cable and
the fiber optic cable with the other.

Hang the height adjustment knob for the headband on the hanger. Then place the connecting cable
and the fiber optic cable on the table while coiling them so that they are not strained.

10. Stop the operation of the photocoagulation system.

Illumination control
1) Turn the illumination control of the B.I.O.
delivery unit to the minimum (not off).
* Turning the illumination control in the direction
indicated with an arrow in the figure on the
right decreases the intensity of the illumination.

2) After approximately 15 seconds, turn the


illumination control fully counterclockwise to
turn off the illumination.

3) Turn the key switch of the main body to the off ( ) position.

4) Turn off ( ) the master switch of the main body.

11. Remove the key from the main body and store it in a customary place.

12. Put the dust cover over the main body.

13. Put the dust cover or the equivalent over the B.I.O. delivery unit.

4.2 Storing the System


When transporting the main body or B.I.O. delivery unit to other facilities or when they will not be used
for a long period of time, store them in the shipping carton or the carrying case (only for the delivery
unit).
4-6

4.2.1 Storing the main body


1. Disconnect the following plugs from the control panel of the main body.

1) Disconnect the fiber optic cable. Plug of the fiber optic cable

Hold the plug and disconnect the fiber optic


cable from the control panel.

Coil the fiber optic cable in an appropriate


radius and bundle it.

Put the rubber cap on the plug of the fiber optic


cable.

Place the plug temporarily where it will not be


an obstruction.
Rubber cap

CAUTION
10 cm or more
• Pay attention to the handling of the fiber optic
cable so that it is not coiled in a radius of less
than 10 cm.
The transmittancy of the laser beam may
be reduced or the optical fiber may be
damaged.

• Be careful not to soil or scratch the tip of the


plug of the fiber optic cable when inserting
the plug into the FIBER connector.
The laser delivery performance may
deteriorate.
Tip of the plug of the fiber optic cable

2) Disconnect the short plug of the B.I.O. delivery


unit. Short plug of
the B.I.O. delivery unit
Disconnect the short plug of the B.I.O. delivery
unit from the DELIVERY connector.

Place the it temporarily where it will not be an


obstruction.
4-7
2. Remove the following plugs from the rear panel of the main body.

Short plug

Power cord Cable of the footswitch

1) Remove the short plug or plug of the external switch signal line from the REMOTE connector.

Remove the short plug or plug of the external switch signal line from the REMOTE connector.

(When the signal line of the external switch is used, coil the cable in an appropriate diameter and
bind it.)

Place the short plug or plug of the external switch signal line temporarily where it will not be an
obstruction.

2) Remove the footswitch from the FOOT SW connector.

Hold the plug and remove the cable plug from the FOOT SW connector.

Coil the cable in a diameter a little larger than the width of the cover, and put it in the cover.

Place the footswitch temporarily where it will not be an obstruction.

3) Remove the plug of the power cord.

Hold the plug and remove the power cord from the grounded power outlet.

Remove the female plug of the power cord from the inlet.

Coil the power cord in an appropriate diameter and bind it with the attached strap.
4-8
3. Store the accessories in the accessory box. Divider

1) Put the divider in the accessory box.


* Place the divider on a side of the box. Accessory box

2) Store the footswitch in the larger compartment


of the two.

3) Store the dust cover, short plug (or external


switch signal line), key switch, and spare fuse
Special pad
in the smaller compartment.

4) Close the lids of the accessory box and seal it Control panel
with packing tape. Bubble bag

Special pad

4. Store the main body and the accessory box in


the internal box. Top frame

1) Place the divider on the bottom of the internal


box.
* Place the divider on a side of the box. System
main body
2) Store the accessory box in the room beside
the divider.
* Lay the plastic
3) Lay the bottom frame in the divider. sheet.
Bottom frame
4) Lay the plastic sheet over the bottom frame.

5) Place the main body with the control box


detached on the bottom frame and wrap it with Divider
the plastic sheet.
Accessory
6) Put the top frame on the main body wrapped box
in the plastic sheet.

7) Put the control box in the air cellular cushioning


bag. Place it on the top cushioning.
* Bind the control box with the special pads.
Then put them in the bubble bag.

8) Place the operator’s manual on the top frame.

9) Seal the lids of the box with packing tape.


Internal box
4-9
5. Store the internal box in the external box.

1) Lay the bottom pad on the bottom of the


external box. Corner cushioning

2) Place the corner pads at each corner on the


bottom pad.

3) Place the internal box delicately so that its


bottom 4 corners come inside the 4 corner
Internal box
pads.

4) Put the corner pads between the corners at the


top of the internal box and the external box. Corner pads

5) Close the lids of the external box and seal it


with packing tape.

Bottom pad

External box
4 - 10

4.2.2 Storing the B.I.O. delivery unit


1. Prepare the B.I.O. delivery unit for storage.

1) Disconnect the plug of the power cord of the AC adapter from the power outlet.

2) Tighten the circumference and the height adjustment knobs so that the diameter of the headband
becomes the minimum.

3) Disconnect the plugs of the illumination cable


View from under the wall pack
and the AC adapter from the wall pack.

Plug of the illumination


control cable

Plug of AC adapter
4 - 11
2. Open the lid of the carrying case and store the B.I.O. delivery unit.

1) Coil the illumination cable and the fiber optic cable and store them in the bottom of the part where
the B.I.O. delivery unit will be stored later.

2) Store the B.I.O. delivery unit over the cables stored in step 1).

3) Remove the wall pack from the stand. Then store the wall pack as illustrated below.

4) Store the stand in the place as illustrated below.

5) Store the AC adapter in the place as illustrated below.

6) Put the aspheric B.I.O. lens in the case and store it in the place as illustrated below.

7) Store the operator’s manual as necessary.

8) Close the lid and latch the carrying case.

AC adapter

Aspheric B.I.O. lens

Stand

Wall pack

B.I.O. delivery
Carrying case * Store the coiled cables beforehand.
§5 OPERATING PROCEDURES

5.1 Starting the Photocoagulation System


1. Remove the dust covers or the equivalent from the main body and the B.I.O. delivery unit.

CAUTION
• Pay attention not to apply load to the connecting cable and fiber-optic cable when removing
the dust cover or the equivalent.
The photocoagulation system may malfunction.

2. Instruct all personnel present in the operation room except the operator and patient to wear
safety goggles (option) for the treatment beam.

CAUTION
• The safety goggles vary according to the laser type and its purpose. Be sure to use the
recommended goggles.
Recommended goggles ...... Model YL-300 for frequency doubled Nd: YAG: D315-532
L8 YL DIN (Produced by YAMAMOTO KOGAKU CO.,
LTD. Japan)

3. Start the operation of the photocoagulation system.

1) Turn on ( ) the master switch on the rear panel


of the main body.
The indication “ ” beside “LINE” lights up on
the control panel.

Master switch

2) Insert the key into the key switch and turn it to


“ ” “ ”
the on ( ) position.

In approximately 10 seconds of countdown,


the indications “ ” and
“ ” on the control panel lights up.
Then the photocoagulation system goes into
the STANDBY mode.

Key switch
5-2
4. Perform the function checks.

Perform the check referring to [5.4 Function Checks] (p. 5-3) in the operator’s manual of the main
body.
* Record each result in [5.5 Function Check List] (p. 5-6) in the same operator’s manual.
5-3

5.2 Preparing for Laser Emission


1. Turn off the aiming beam.

Keep pressing the AIMING switch on


the control panel to turn off ( ) the aiming beam.

AIMING switch

2. Put the B.I.O. delivery unit on the operator’s Height adjustment knob
head.
Circumference
1) Adjust the height of the headband with the adjustment knob
height adjustment knob.

2) Adjust the tightness of the headband with


the circumference adjustment knob.

3. Adjust the visual field through the B.I.O. so


that it becomes the maximum.

1) Slightly loosen the angle fixing knob to adjust


the inclination of the main part of the B.I.O..

2) Tighten the angle fixing knob to fix the B.I.O.


at the position where the visual field through
Angle fixing knob
the B.I.O. becomes the maximum.

4. Adjust the pupillary distance.

Make an adjustment of the pupillary distance


precisely by sliding the right and left pupillary
distance adjustment knobs separately so that an
unobstructed field of view can be seen.

NOTE Pupillary distance


• Make a precise adjustment especially when adjustment knob

observing patients with microcoria.


5-4
5. Instruct the patient to lie on the bed.

Have the patient lie in a way suited for observation of and laser emission to the patient’s eyes while he/
she is on his/her back.
[Cautions regarding emission of the illumination light]
• Set the intensity of the illumination to the minimum level (not off) at the start of the observation.
Then gradually raise the level as necessary. Make a habit of resetting the intensity to the
minimum level every time an observation of a patient is finished.
Raising the intensity suddenly dazzles the patient at the start of the observation. In addition,
it may cause a thermal and photochemical damage to the retina of the patient (blue light
hazard).

• Never fail to pay attention to the following points when operating the B.I.O. delivery unit.
- Keep the intensity of the illumination as low as possible.
- Set the range of the illumination as small as possible.
- Use a color filter as necessary.

• Pay attention when emitting the illumination to the infant or patients with aphakic eye
or eye disease.

6. Observe the fundus before the laser emission.

1) Adjust the intensity of the illumination spot.

Turn the illumination control of the B.I.O.


delivery unit to adjust the illumination spot to
the proper intensity.

NOTE
• Be sure to set the intensity of the illumination
spot to the minimum (not off) level at the
beginning, and increase it gradually as Illumination control
necessary.
5-5
2) Select the size of illumination spot that is Illumination spot size selector
appropriate for observation and laser emission.

Slide the illumination spot size selector to the


right and left to select the appropriate spot size.

3) Insert an appropriate protective filter for


Filter selector Filter selector
observation into the observation optical path.

Insert the red-free filter or the wide angle diffuser


into the observation optical path by pushing and
pulling the filter selector.

4) Observe the affected part on the fundus.

Move the aspheric B.I.O. lens, which is held approximately 5 cm away from the patient’s eye,
back and forth while the operator’s head remains still.
Perform observation under the condition where the fundus can be seen fully within the
aspheric lens. At this time, the fundus image is seen upside-down.

NOTE
• Use the aspheric B.I.O. lens facing the side with a white line on the rim toward the patient.

Fiber optic cable antenna


B.I.O Operator
(option)
approximately 300mm
Fiber optic cable
GYC-1000 main body
approximately 50mm

Patient

Wagon with
casters

Foot switch
5-6

5.3 Laser Emission


1. Adjust the aiming beam.

1) Adjust the intensity of the aiming beam to


an appropriate level with the AIMING
switch , on the control panel.

AIMING switch

2) Turn the laser mirror adjustment knob so that


the aiming beam and the illumination spot come
to the center of the field of view.

Laser mirror adjustment knob


5-7
2. Adjust the working distance*2 (W.D.).

While fixing the positions of the aspheric B.I.O. lens and operator’s head, move the working distance*2
control knob to the right or left, and stop it at the position where the spot size of the aiming beam
becomes the minimum. Perform the same operation when the patient’s eye is an aphakic eye or has
been treated with fluid-air exchange.
[Examples of working distance (W.D.) and spot size (when 20D aspheric lens is used.)]
W.D.
300 mm 400 mm 500 mm 600 mm 700 mm
Condition of eye
Normal eye 185 μm 278 μm 371 μm 464 μm 556 μm

Aphakic eye
215 μm 324 μm 433 μm 541 μm 649 μm
(Anterior chamber: Water)

Phakic & fluid-gas exchanged eye


(Anterior chamber: Water, Posterior 226 μm 272 μm _ _ _
chamber: Air)

Phakic & fluid-gas exchanged eye


176 μm 220 μm 264 μm _ _
(Anterior and posterior chambers: Air)

Aphakic & fluid-gas exchanged eye


_ 362 μm 544 μm 726 μm 909 μm
(Anterior and posterior chambers: Air)

[Examples of W.D. and spot size (when 28D aspheric lens is used.)]
W.D. 300 mm 400 mm 500 mm 600 mm 700 mm

Normal eye 259 μm 389 μm 519 μm 650 μm 778 μm

[Examples of W.D. and spot size (when 14D aspheric lens is used.)]
W.D. 300 mm 400 mm 500 mm 600 mm 700 mm

Normal eye 130 μm 195 μm 260 μm 325 μm 389 μm

NOTE
• The actual spot size depends on positions of the B.I.O., aspheric B.I.O. lens, and patient’s
eye. The spot size is not indicated on the control panel. The spot size display on the control
panel shows “bio”.

When the working distance control knob is rotated


counterclockwise seen from the side it is attached,
Distance from the eye
the working distance becomes longer and the spot to the fingertip
size becomes larger.
As the working distance depends on the length of
the operator’s extended arm, the distance between
the operator’s eyes and the fingertips is a guideline.
In the actual operation, keep the most comfortable
distance for the operator to observe the affected
part, and adjust the working distance control knob
to minimize the spot size of the aiming beam.
*2 The working distance described in this manual is the distance between the eyepiece of the B.I.O. and
the 20D aspheric B.I.O. lens.
5-8
3. Set each photocoagulation condition.

Settings of all the photocoagulation conditions except the spot size setting are performed on the control
panel. All the photocoagulation conditions are shown on the control panel.
Aiming level indicator INTERVAL display
COUNTER display POWER display Emergency stop switch

POWER switches TIME display

AIMING switches TIME switches


Reset switch INTERVAL switch

[a. Output power (laser mirror)]


The output power of the treatment beam is set with the POWER switches . The output
power can be set between 50 and 500 mW in 10mW increments and between 500 and 1700 mW
in 50 mW increments.

[b. Exposure time]


The exposure time of the treatment beam is set with the TIME switches .
The exposure time can be set in the range between 0.02 and 0.10 seconds in 0.01 second increments,
0.10 and 0.50 seconds in 0.05 second increments, 0.50 and 1.00 seconds in 0.10 second increments,
and 1.00 and 3.00 seconds in 1.00 second increments.

[c. Single/Repetition mode]


The single or repetition mode is selected by pressing the INTERVAL switch as necessary.
In the repetition mode, the interval time (between 1.0 and 0.1 seconds in 0.1 second decrements) is
indicated on the INTERVAL display on the control panel. Nothing is indicated in the single mode.

[d. Aiming level]


The intensity of the aiming beam is increased with the AIMING switch, and decreased with
the AIMING switch.

[e. Counter reset]


The indication on the counter is reset to “0” by pressing the RESET switch.
5-9
4. Place the photocoagulation system in the READY mode.
“ ”

Press the STATUS switch to place the system in


the READY mode. When the photocoagulation
system is placed in the READY mode,
“ ” indication lights up on the control
panel.
* Change from the STANBY mode to the READY
mode takes approximately 2 seconds.
* When the aiming beam is off “ ”, the
photocoagulation system does not change to the STATUS switch

READY mode.

CAUTION
• Always place the photocoagulation system in the STANDBY mode by pressing the STATUS
switch whenever the treatment laser beam is not going to be emitted.
This prevents an accidental emission of the laser beam even though the foot switch
is pressed inadvertently.

5. Emit the treatment beam by pressing the footswitch.

1) Adjust the position of the aspheric B.I.O.


lens and manipulate the working distance
control knob.

2) When the emission position is determined,


press the footswitch.
The treatment laser beam is emitted, and the
number of emissions is indicated on the
COUNTER display. COUNTER display

NOTE
• If pressing the footswitch produces a series of beeps but does not emit the treatment laser
beam, check whether the system is placed in the READY mode and the aiming beam is
emitted.
The system is designed to emit the treatment laser beam only when both of the conditions
described above are met.

• If pressing the footswitch does not produce any beep or does not emit the treatment laser
beam, check whether the footswitch is connected to the main body.
The treatment laser beam cannot be emitted if the footswitch is not connected to the main
body.
5 - 10
6. When the laser emission is completed, place the photocoagulation system in the STANDBY
mode.
STATUS switch
1) Press the STATUS switch to place the system “ ”
in the STANDBY mode.
The “ ” indication lights up on the
control panel.

2) Keep pressing the AIMING switch to


turn off the aiming beam.
The “ ” indication on the left of the AIMING
level indicator lights up.
AIMING switch

7. Set the intensity of the illumination to the minimum.

Turn the illumination control of the B.I.O. delivery


unit to set the intensity of the illumination to the
minimum (not off).

Illumination control

8. Release the patient from the bed.

After the postoperative treatment, release the patient from the bed.

9. When performing the laser emission on the next patient, go back to step 5 of “5.2 Preparation”
(p. 5-4) while the photocoagulation system is in the STANDBY mode.
5 - 11

5.4 End of operation


1. Stop the operation of the photocoagulation
system.

1) Turn the key switch on the control panel to the


off ( ) position.
Key switch
2) Turn off ( ) the master switch on the rear panel.

3) Remove the key and store it in a customary


place.

Master switch

2. Instruct all personnel present in the operating room to take off the safety goggles.

3. Put away the photocoagulation system.

1) Hold the headband of the B.I.O. with one hand


while holding the connecting cable and fiber-
optic cable with the other hand.

2) Hang the part around the height adjustment


knob of the headband on the hanger of the
stand.

3) Put the dust cover or the equivalent over the


main body and B.I.O. delivery unit.

4) Put the aspheric B.I.O. lens in the case and


store it in the customary place.
5 - 12

5.5 Indications of Misoperation


Abbreviated indication
The GYC-1000 produces a series of beeps to let
you know your misoperation.
In such a case, one of the following abbreviated
indications appear on the TIME display on the
control panel.
To correct the misoperation, follow the suggestions
below.

Indications Descriptions Suggestions

The STATUS switch is pressed to place


Press the AIMING switch ▲ to emit
"A.P." the system in the READY mode even
the aiming beam.
though the aiming beam is turned OFF.

The AIMING switch ▲ is pressed to emit Connect the cable plug of the fiber-
"F.S." the aiming beam even though the fiber- optic cable to the fiber connector of
optic cable is not connected. the main body.

The foot switch is pressed to emit the


Press the STATUS switch to place
"S.S." treatment laser beam even though the
the system in the READY mode.
system is placed in the STANDBY mode.

5.6 Emergency Stop


When any trouble occurs with the patient or the
photocoagulation system, press the emergency stop
switch (“STOP” is indicated) to stop the Emergency stop switch
(“STOP”)
photocoagulation system immediately during
operation.

NOTE
• To restart the photocoagulation system, turn
the key switch to the off ( ) position, and
turn it to the on ( ) position again.
§6 MAINTENANCE

6.1 Replacing the Illumination lamp


1. Verify that the power of the photocoagulation system is turned off.

If the power is on, turn off the key switch and master switch.

2. Pull out the burned-out illumination lamp from the B.I.O. delivery unit.

CAUTION
• The illumination lamp is very hot for some time after it has burned out, and touching it may
cause burn.
Remove the lamp after it has cooled enough.

Illumination lamp
Slit
Protrusion

Remove

Attach

Illumination lamp

3. Replace the illumination lamp with a new one.

Aligning the protrusion of the illumination lamp with the slit on the lamp housing. Then insert the lamp
as far as it goes. (Model of the lamp: AP II-7003 (2 pieces) manufactured by Keeler & Y.NA Ltd.)

CAUTION
• Never touch the lamp glass part with bare hands. If you touch the lamp glass accidentally,
wipe it gently with a clean cloth dampened with alcohol.
If the lamp glass part gets soiled with oil or dirt from fingers, the lifetime o it may be
reduced.

4. Verify that the illumination lamp lights up.

1) Turn on the power of the photocoagulation system.


After turning on ( | ) the master switch, turn on ( ) the key switch.

2) Turn the illumination control on the B.I.O. delivery unit, and verify that the illumination lamp lights
up.
6-2

6.2 Cleaning

6.2.1 Cleaning the optical parts


1. Blow off dust and dirt from the optical parts with a blower.

2. Clean the optical parts with alcohol.

Gently wipe the optical parts with a Keeler’s cleaning cloth or with a lint free wiper dampened with a
little rubbing alcohol. (Do not expose the system to the water at this time. However, the optical parts
can be wiped with a clean cloth dampened with water and wrung out.)
The parts to be cleaned are:
• Observation window • Eyepiece • Aspheric lens

CAUTION
• Never wipe the optical parts while they are dirty or wipe them strongly.
The optical parts may be scratched.

6.2.2 Cleaning other parts


1. Clean the inside of the headband.

Clean the part that contacts the patient’s hair or forehead with a clean cloth dampened with rubbing
alcohol or neutral detergent.

2. Wipe the part that the operator’s finger touches with rubbing alcohol.

Clean the following parts that the operator’s finger touches with rubbing alcohol:
• Head band height /circumference adjustment knobs • Forehead bar adjustment knob
• Angle fixing knob • Filter selector • Illumination spot size selector
• Mirror adjustment knob • Illumination control • Pupillary distance adjustment knob
6-3

6.2.3 Cleaning the fiber optic cable


If the tip of the plug of the fiber optic cable contact other object in connection or disconnection of it, be
sure to clean it according to the procedure described below. Using the fiber optic cable with the tip of
it is soiled may damage it and make it unusable.

1. Soak clean gauze in rubbing alcohol or distilled water.

2. While touching the tip of the plug of the fiber optic cable to the gauze, clean the tip by
rotating the fiber optic cable lightly.

CAUTION
• Do not apply too much stress on or scratch
the tip while cleaning it.
The fiber optic cable inside the plug may
be broken.

3. Touch another part of the tip to the gauze and clean the probe a few times in the same
manner as in step 2 to make sure that the tip is properly cleaned.

4. Put the rubber cap on the plug of the fiber optic cable.

1) Start operation of the photocoagulation system


in a normal procedure.

2) Project the aiming spot on a surface that is not


specular and check the intensity of the whole
spot.
* If the intensity of the aiming spot is even
around the center and no decrease in intensity
or vignetting is found as figure on the right,
the condition of the tip of the fiber optic cable
plug is proper.

NOTE
• If lowering of the intensity or vignetting is found as a result of the check above, contact
NIDEK or your authorized distributor.
§7 SPECIFICATIONS AND CONFIGURATIONS

7.1 Specifications
1. Binocular indirect ophthalmoscope
1-1. Connectable photocoagulator: GYC-1000
1-2. Treatment laser beam power output:
50 to 1700 mW
<Tolerance>
50 to 90 mW: ± 20 %
100 to 1700 mW: ± 15 %
1-3. Aiming beam power output: Maximum; Red (635 nm) 0.1 to 0.3 mW
Minimum; one-eighth of the maximum
1-4. Working distance: 300 mm (SHORT) to 700 mm (LONG)
1-5. Spot size on the fundus: 185 µm (W.D. 300 mm)
278 µm (W.D. 400 mm)
556 µm (W.D. 700 mm)
* When using the 20 D aspheric B.I.O. lens on the fundus of a
normal eye
1-6. Beam divergence: 0.0372 rad (W.D.: 300 mm) to 0.0124 rad (W.D.: 700 mm)
<Tolerance>
± 20 %
1-7. Guard wavelength
of protective filter: 532 nm wavelength
1-8. Light source of illumination: 6 V, 1.7 A
1-9. Light source of halogen lamp: 0 to more than 3000 Lux
1-10. Dimension: 240 (W) × 280 to 330 (D) × 145 (H) mm
1-11. Weight: 520 g (without fiber-optic cable)
1-12. Fiber-optic cable: Length approximately 3 m

2. Aspheric B.I.O. lens


2-1. Type: 20 D Aspheric B.I.O. lens for diagnosis/laser
(with an anti-reflecting coating for 808nm wavelength)
2-2. Diameter: 50 mm (effective diameter 48.0 mm)
2-3. Focal distance: f = 50 mm
2-4. Center thickness: 15.0 mm
2-5. Field of view: 45º
2-6. Weight: 55 g

3. Environmental conditions
3-1. Transport / Storage: Temperature: 14 to 122 ºF (–10 to 50 ºC)
Humidity: 10 to 95 % (non-condensing)
3-2. In use: Temperature: 50 to 86 ºF (10 to 30 ºC)
Humidity: 80 % or less (non-condensing)
3-3. other conditions: little dust

NOTE
• The actual spot size depends on the working distance and positions of the aspheric B.I.O.
lens and patient’s eye.

• Specifications are subject to change without notice for improvement.


7-2

7.2 Standard Configurations


• B.I.O. delivery unit (Keeler, All Pupil II type) .............................................. 1
* including the wall pack and the AC adapter
• Fiber-optic cable ........................................................................................ 1
• 20 D Aspheric B.I.O. lens for diagnosis/laser (with a leather case) ............... 1
• Stand ......................................................................................................... 1
• Illumination lamp ......................................................................................... 1
• Short plug for the B.I.O. delivery unit .......................................................... 1
• Operator’s manual ...................................................................................... 1
• Carrying case ............................................................................................. 1

7.3 Options
• Safety goggles for treatment beam (Model YL-300 for laser diode 2: DI 800-910 L4 produced
by YAMAMOTO KOGAKU CO., LTD.)
• Aspheric B.I.O. lens 28D (with a case)
Words in This Manual
To understand the contents of this manual easily, the following words are used.
Before reading this manual, confirm the meaning of the following words.

Aiming beam ............................. Red laser beam for aiming. The aiming beam is “OFF” when it is not
being emitted.

Exposure time .......................... Length of time that the treatment beam is emitted. (Unit: sec.)

Green laser beam ..................... Green laser beam for treatment.

Laser beam ............................... Treatment beam and aiming beam.

Main body ................................ Main body of the green laser photocoagulation system GYC-1000.

Power output ............................. Power output of the green laser beam (at the aperture of the delivery
unit). (Unit: mW)

Single mode/Repetition mode


.................................... In the single mode, 1 shot of the treatment beam is emitted in
the set exposure time when the foot switch is pressed. In the
repetition mode, the treatment beam is repeatedly emitted in
the set exposure time at certain intervals while the foot switch
is pressed.

Spot size ................................... Diameter of the spot of the laser beam. (Unit: µm)

You might also like