Compact Spectrophotometer
Service Manual
Revision : D01-201407
Halo XB-10/VIS-20 Spectrophotometer
Service Manual
1
Content
1. Introduce ...................................................................................... 6
1.3 Structure....................................................................................... 6
2
3.1 Basic Optical Path ...................................................................... 13
3
5. Replacement and Maintenance of Lamp Source ASSY .............. 25
8. Warranty .................................................................................... 32
4
Preface
Thank you for your purchase of Halo XB-10/VIS-20 UV/VIS spectrophotometer.
Statement: Both common and difference of XB-10 and VIS-20 UV/VIS
spectrophotometer will be introduced in this service manual.
The service manual has been compiled for reference for operators to allow them
to grasp, diagnose and remedy the spectrophotometer trouble quickly in the event of
XB-10/VIS-20 UV/VIS spectrophotometer.
Before using the instrument, be sure to read through the manual and according to
the operation manual to use it. Correct operation can reduce trouble and slow down
lifetime.
Once XB-10/VIS-20 UV/VIS spectrophotometer is faulty, please stop working,
disconnect the power cable.
Please use this service manual combined with the operation manual.
5
1. Introduce
1.1 Outline
1.2 Principle
1.3 Structure
1.3.1 Shape
6
Keypad Board
Door
Screw
Cuvette
Shell
Screw
Optical Path
System
Bottom Panel
Screw
Fig.1-1 XB-10/VIS-20
7
1.3.2 Base unit
Switch power
Circuit Board Fan
Optical Path
Unit
Grating
Cuvette
PC
Switch power
Accessory
Plug
Power
Fan
Printer
Fuse
8
Remark: :
Name Function
Switch power Switch on/off the instrument.
Plug Input AC
Fan For VIS-20, lamp source is WI lamp, need fan to exhaust heat
For XB-10, no fan.
Fuse Avoid high voltage
PC
Printer Connect with 80-column thermal print to collect data
Accessory Communication joint for accessories
Power Output voltage
9
2. Operation Requirement
Environment:
Indoor temperature: 5℃~35℃ (20℃~25℃ is better)
Humidity: 45%~85% (30%~70% is better)
Storage temperature: -20℃~55℃
Size:
(W)400mm×(D)280mm×(H)160mm
Weight:
About 4kg
Printer:
80-column thermal printer
2.2.1 XB-10/BIOmaster
2.2.2 VIS-20/ENVIRONmaster
Please do select the connect wire in electric circuit which is specified in the
manual to minimize the interference of electromagnetic wave.
Operation of this instrument may cause electromagnetic interference on the
electronic equipment nearby such as TV Set or radio. If you suspect that this
instrument is responsible for any electromagnetic wave interference, turn off the
power of this instrument to check it by observing the signal received of TV or radio.
Remark: Electromagnetic waves generated by the instrument will not affect
people's normal life, and meets the nation’s standard. Under good operation
environment, you can use the instrument without worries.
If the interference of electromagnetic wave is confirmed, you can try the
following ways to improve.
· Make the troubled equipment such as TV or radio apart from the other site.
· Avoid using same power line.
Optical path system includes light source, lens, grating, sample cell and detector
board and so on. Optical path accuracy is related with instrument’s performance and
parameters, and only professional person can calibrate optical path system by tools
and equipment.
This manual describes XB-10 and VIS-20 Spectrophotometer’s optical path
system. And we suggest that only professional serviceman can align optical path.
Lens
Light source
Detector board
Grating
13
1-light source 2-entrance lens 3-grating 4-exit lens 5-sample cell 6-detector board
Fig.3-2 Optical path system diagram
Entrance lens of XB-10 is different from VIS-20’s, but its function is the same as
VIS-20’s. Light crosses the slit and display on the grating.
Entrance lens assembly includes entrance and exit slit, lens, lens holder and slit
holder.
Name Model No. Reference No. Part No.
Entrance lens XB-10 0006-XB10-0000 0027745
14
VIS-20 0006-VIS20-0000 0027701
Grating assembly includes grating, aperture, control board, sheathing and holder.
Remark: Only professional serviceman can remove the grating assembly.
Name Reference No. Part No.
Grating ASSY 0007-XB10-0000 0027771
XB-10’s exit lens assembly is different from VIS-20’s. But their functions are
same. Spectrum focuses on grating and cross the cuvette in the sample compartment.
Use filter to ensure light’s monochromaticity. Two filters (yellow one and white
one) are placed in the XB-10 Spectrophotometer to test high wavelength region, and
one black filter is placed in the VIS-20.
Exit lens assembly includes filter, holder, lens and exit slit.
Name Model No. Reference No. Part No.
Exit Lens XB-10 0005-XB10-0000 0027747
ASSY VIS-20 0005-VIS20-0000 0027703
Screw
Spring 2
Spring 1
Pressure panel
Cuvette holder
Base panel
3.2.6 Photocoupler
Detector board assembly includes lightproof film, detector board, lens and
photodiode.
As Fig.3-2 shown, photodiode converts optical signal to electric signal.
XB-10 spectrophotometer’s wavelength range is not same as VIS-20’s, so the
function of detector board is different. The detector boards of XB-10/VIS-20
spectrophotometer are also different.
Name Model No. Reference No. Part No.
Detector Board XB-10 0008-XB10-0000 0027739
Assembly VIS-20 0008-VIS20-0000 0027700
16
4. Electric Circuit System
As Fig.4-1 shown, and refer to the Fig.5-11 on the service manual when
maintaining the instrument.
17
Main control board
SD-card board
Detector board
Optical coupler
Wavelength motor
Fig.4-2 XB-10
USB
J1
J5
J2
J6
J3
J4 J7
J1 of touch screen
J1 of keypad board
19
Description of touch screen and keypad board joints
Joint No. Connection Working State Function
J1 of touch screen 40-pin cable J7
J1 of keypad board control board 4-pin cable J9
Joint
Function: Check beam signal and convert optical signal to electric signal. Detector
boards of XB-10 and VIS-20 Spectrophotometer are different, because
optical signal is not same.
20
4.5 Xenon lamp
J5 joint
Aperture
4.6 WI lamp
WI lamp
J2
J1
Description of connector
J1: input AC 22V
J2: output DC
1、J4: DC 24V, for control board and fan of VIS-20
2、J5: DC 5V, for control board
3、J5: DC 5V, for fan of XB-10
Name Reference No. Part No.
Power Board 0003-XB10-0008 0028499
22
4.8 SD-card board
Motor ASSY
23
4.10 Electromagnet ASSY
Joint
Electromagnet
24
5. Replacement and Maintenance of Lamp Source ASSY
VIS-20 Spectrophotometer
: Switch off the power (OFF), and let the lamp cools down.
: Loose off the screw and remove the cover, as Fig.5-1 shown.
Fig.5-1
: Loose off the two screws on the cover, as Fig.5-1 (1) shown, and keep the
cover vertically upward, as Fig.5-2 shown.
Fig.5-2
25
3
Fig.5-3
: Loose off the two screws on the WI lamp holder (as Fig.5-2(2) shown). Hold
the old WI lamp by forefinger and thumb (as Fig.5-3(3) shown) gently pull it up, and
remove it from the lamp holder.
: Check and replace the new WI lamp. The key is that the new WI lamp
filament and pin must be same as the old lamp ones. If there is difference, please
correct to prevent displacement of the optical path.
: When check or replace the new WI lamp, please wear soft gloves or hold the
new WI lamp by soft cloth. Don’t leave the fingerprint on the new WI lamp.
: When installing the new WI lamp, push down the pin in the lamp holder.
: Reverse the removal procedures to install.
26
1
2
Fig.5-4
: Pick up the xenon lamp module and xenon lamp assembly vertically upward.
: Loose the two screws (Fig.5-5(1)) and remove the xenon lamp.
: Reverse the removal Steppers to install the new lamp.
27
5.4 Clean of cover
Please don’t place the chemical reagent or sample on the cover of the instrument.
In case chemical reagent or sample is spilled on the cover, remove the chemical
by a piece of soft cloth or whip firstly, clean the cover with water or mild detergent
and then dry it as soon as possible to avoid corrosion.
▲ Attention:
1. Don’t spill out liquid on the LCD panel and keypad assembly.
2. Please unplug the power cable before cleaning the cover.
After measurement of the sample, rinse the cell with distilled water thoroughly
and wipe the cell with soft cloth.
Please place the cell in the sealed box and put the desiccant in it when storing the
cell. You also can place the cell in the box filled with distilled water such as beaker,
cover the cap. Please take note that the storage place of the box must be clean.
Clean the transparent side of the cell with Alcohol/Ether in the ratio of 1 to 1 as
required.
Replace the new cuvette if there is nick or stain.
Serious dirty and poor environment will contaminate the sample compartment,
which effects photometry accuracy and stability. So clean it if necessary. Before
wiping, clean the hands. As following steps to clean the lens.
Steps
(1) Turn off the instrument and disconnect the power cable.
(2) Open the cover of the sample compartment and remove the cell holder.
(3) Wipe off the dirty with a piece of soft paper or cloth soaked with
Alcohol/Ether (mixed in the ratio of 1 to 1) and dry it.
(4) Install the cell holder.
△ Cautions
(1) When cleaning the lens, it only needs to wipe off the side facing to the
sample compartment. The other side is sealed and it is difficult to be contaminated.
(2) Do not remove the lens or lens assembly to prevent the displacement of
optical path.
(3) Do not leave the fingerprints on the lens.
28
6. Routine Maintenance and Trouble Shooting
Correct use and maintenance ensures that the instrument works and extends the
lifetime of the instrument.
1) Operate the instrument according to the instruction manual strictly. Adopt
containment to keep the instrument work normally.
2) Compile operation rule and regulate user action.
3) Prohibit the excess liquid sample placed in the cuvette.
4) Know well about function and instruction of auxiliary equipment and keep the
replaced spare parts in a safe place.
5) Wipe the cover by warm water and soft cloth, and do not use alcohol, ether
and acetone. Please use dust cover when the instrument does not work.
6) Check the instrument regularly.
30
7. Accessories List
31
8 . Quality Guarantee
(1)Guarantee Scope (a) Free maintenance to instrument noumenon for errors caused by
fabrication fault.
(b) When maintenance, some substitute may be used and some damaged
parts might be replaced with identical parts instead of on-site service.
(c) Sometimes the same products like computer and printer used in
the instrument are not able to be supplied, because they are
eliminated in the market.
(d) Free maintenance is not provided for deserted maintenance or
those resold instruments and parts without our permission even
during the guarantee time.