You are on page 1of 8

Original Article

Effects of screw and host factors on insertion torque and pullout strength
Ankit H. Shaha; Rolf G. Behrentsb; Ki Beom Kimc; Hee-Moon Kyungd; Peter H. Buschange

ABSTRACT
Objective: To experimentally study the effects of altering implant length, outer diameter, cortical
bone thickness, and cortical bone density on the primary stability of orthodontic miniscrew implants
(MSIs).
Materials and Methods: Maximum insertion torque (IT) and pullout strength (POS) of 216 MSIs
were measured in synthetic bone with different cortical densities (0.64 g/cc or 0.55 g/cc) and
cortical thicknesses (1 mm or 2 mm). Three MSIs were evaluated: 6-mm long/1.75-mm outer
diameter, 3-mm long/1.75-mm outer diameter, and 3-mm long/2.0-mm outer diameter. To test
POS, a vertical force was applied at the rate of 5 mm/min until failure occurred.
Results: The 6-mm MSIs displayed significantly (P , .001) higher IT and POS than the 3-mm
MSIs did. The 3-mm MSIs with 2.0-mm outer diameters showed significantly higher (P , .001) IT
and POS than the 3-mm MSIs with 1.75-mm outer diameters. The IT and POS were significantly
(P , .001) greater for the MSIs placed in thicker and denser cortical bone.
Conclusion: Both outer diameter and length affect the stability of MSIs. Increases in cortical bone
thickness and cortical bone density increase the primary stability of the MSIs. (Angle Orthod.
2012;82:603–610.)
KEY WORDS: Miniscrew implants; Insertion torque; Pullout strength; Cortex; Density

INTRODUCTION most important factors for maximizing pullout strength


(POS) of orthopedic screws.4,5 Longer orthopedic
Orthodontic miniscrew implants (MSIs) have become
screws demonstrate greater POS than shorter screws
an important tool in the orthodontists’ armamentarium;
do.6 The MSI literature pertaining to the effects of MSI
MSIs provide absolute anchorage, are affordable, can be
length remains controversial.7–9 Previous studies often
quickly placed at various sites, and are easily removed.1–3
compare MSIs that differ in more than one character-
However, MSIs have proven to be less stable than other
istic, which limits their ability to evaluate the effects of
endosseous implants that osseointegrate. Both screw
diameter and length.
and the host factors affect the stability of MSIs.
The host factors are related to the quantity (cortical
The screw factors are related to the screws’ design.
thickness) and quality (cortical density) of the bone.3,10
Implant diameter has been shown to be one of the
Because maximum stress occurs at the cortical level,11 it
has been recommended that cortical bone at MSI
a
Resident, Department of Orthodontics, Orthodontic Depart- insertion sites should be at least 1.0-mm thick.12
ment, St Louis University, St Louis, Mo.
b
Professor and Department Chair, Orthodontic Department, Increased cortical thickness has been associated with
St Louis University, St Louis, Mo. significantly greater MSI success.13 Bone mineral density
c
Assistant Professor, Orthodontic Department, St Louis is also important for ensuring the stability of endosseous
University, St Louis, Mo. implants.14 Bone density has been positively correlated
d
Professor and Department Chair, Orthodontic Department,
with both insertion torque (IT) and POS.15,16
Kyungpook National University, Daegu, Korea.
e
Professor, Department of Orthodontics, Baylor College of Few studies have simultaneously evaluated the IT
Dentistry, Dallas, Tex. and POS of MSIs. Both must be quantified because
Corresponding author: Dr Peter H. Buschang, Department of they provide different information about primary,
Orthodontics, Baylor College of Dentistry, 3302 Gaston Ave, mechanical stability. IT and POS do not respond in
Dallas, TX 75246
(e-mail: PHBuschang@bcd.tamhsc.edu) the same way to experimental effects.15 The ideal MSI
should minimize IT (less potential bone damage) and
Accepted: October 2011. Submitted: July 2011.
Published Online: December 8, 2011
maximize POS (greater holding power).
G 2012 by The EH Angle Education and Research Foundation, The primary aim of this study is to evaluate the
Inc. effects of altering implant length, outer diameter,

DOI: 10.2319/070111-427.1 603 Angle Orthodontist, Vol 82, No 4, 2012


604 SHAH, BEHRENTS, KIM, KYUNG, BUSCHANG

Figure 1. (A) Six-millimeter-long miniscrew implant (MSI) with an outer diameter of 1.75 mm. (B) Three-millimeter-long MSI with an outer
diameter of 1.75 mm. (C) Three-millimeter-long MSI with an outer diameter of 2 mm.

Angle Orthodontist, Vol 82, No 4, 2012


EFFECTS OF SCREW AND HOST FACTORS ON STABILITY 605

Figure 2. Synthetic bone cubes used showing (A) cortical and cancellous layers and (B) miniscrew implant.

cortical bone thickness, and cortical bone density on and the 3-mm MSIs were 5.5 mm and 2.5 mm,
the primary stability of orthodontic MSIs. Given the respectively.
potential root damage that can be caused by MSIs,17,18
it is especially important to determine the stability of Synthetic Bone
the shorter 3-mm MSIs that have been recently
introduced.19–22 The secondary aim was to determine Because of the variability of cadaver and animal
if cortical thickness and density interact with MSI bone,23 synthetic bone has become the standard for
length and diameter. evaluating the primary stability of endosseous implants
and MSIs because of its uniform material properties.5,24
MATERIALS AND METHODS Blocks of synthetic polyurethane bone (Sawbones,
Vashon, WA) were used to test the IT and POS of the
Three different MSIs were specifically fabricated by MSIs. The density of the cortical layer was 0.64 g/cc or
Dentos (Daegu, Korea). They were either 3-mm or 6- 0.55 g/cc; the cortical density of the human mandible
mm long; the 6-mm MSIs had an outer diameter of has been reported to be 0.64 g/cc.25 Density of the
1.75 mm; the 3-mm MSIs had outer diameters of either cancellous layer was 0.48 g/cc for all of the specimens.
1.75 mm or 2 mm. Other than length and outer The cortical layers were either 1-mm or 2-mm thick.
diameter, the MSIs were identical in terms of all other
design features (Figure 1); all inner diameters were
Testing Groups
1.5 mm, all had 0.5-mm pitch, the apical 1.5 mm of all
MSIs was tapered, all MSIs were self-drilling, and all The larger bone blocks were cut into 11-mm cubes
were self-tapping. The threaded portions of the 6-mm for testing purposes (Figure 2). The three MSIs were

Table 1. Groupings of the Three Miniscrew Implants According to Cortical Bone Thickness (CT) and Cortical Bone Density (CD)
Group 1 Group 2 Group 3 Group 4
6 mm/1.75 mm 6 mm/1.75 mm 6 mm/1.75 mm 6 mm/1.75 mm
CT 5 1 mm CT 5 1 mm CT 5 2 mm CT 5 2 mm
CD 5 0.56 g/cc CD 5 0.64 g/cc CD 5 0.56 g/cc CD 5 0.64 g/cc
Group 5 Group 6 Group 7 Group 8
3 mm/1.75 mm 3 mm/1.75 mm 3 mm/1.75 mm 3 mm/1.75 mm
CT 5 1 mm CT 5 1 mm CT 5 2 mm CT 5 2 mm
CD 5 0.56 g/cc CD 5 0.64 g/cc CD 5 0.56 g/cc CD 5 0.64 g/cc
Group 9 Group 10 Group 11 Group 12
3 mm/2 mm 3 mm/2 mm 3 mm/2 mm 3 mm/2 mm
CT 5 1 mm CT 5 1 mm CT 5 2 mm CT 5 2 mm
CD 5 0.56 g/cc CD 5 0.64 g/cc CD 5 0.56 g/cc CD 5 0.64 g/cc

Angle Orthodontist, Vol 82, No 4, 2012


606 SHAH, BEHRENTS, KIM, KYUNG, BUSCHANG

Table 2. Analysis of Variance (ANOVA) of Differences in Insertion Torque (Ncm), Measured at 100% of Insertion, of Miniscrew Implant (MSI)
#1 (6-mm Length, 1.75-mm Outer Diameter), MSI #2 (3-mm Length, 1.75-mm Outer Diameter), and MSI #3 (3-mm Length, 2-mm Outer
Diameter), With Post Hoc Pairwise Comparisons
Post Hoc Probabilities and Relative Differences
MSI #1, MSI #2, MSI #3, ANOVAs,
Density Thickness Mean 6 SD Mean 6 SD Mean 6 SD F Value/Prob MSI # 2 vs 1 MSI # 2 vs 3 MSI # 3 vs 1
0.56 g/cc 1 mm 6.43 6 0.66 4.64 6 0.54 5.44 6 0.72 34.80/,.001 ,.001; Q27.84% .001; Q14.71% ,.001; Q15.4%
0.56 g/cc 2 mm 10.19 6 0.75 7.22 6 0.66 8.24 6 0.88 69.91/,.001 ,.001; Q29.15% .001; Q12.38% ,.001; Q19.14%
0.64 g/cc 1 mm 9.21 6 1.07 6.14 6 0.36 6.46 6 0.66 90.10/,.001 ,.001; Q33.33% .437; Q4.95% ,.001; Q29.86%
0.64 g/cc 2 mm 11.26 6 0.98 8.37 6 0.62 9.18 6 1.3 39.38/,.001 ,.001; Q25.67% .048; Q8.82% ,.001; Q18.47%

inserted into four different types of bone (two densities MSI until failure occurred. Peak load at failure of the
and two cortical thicknesses), resulting in 12 test MSI was recorded in kilograms.
groups. There were a total of 216 MSIs, with 18 MSIs
randomly allocated to each group (Table 1). Statistical Analysis

Insertion Torque MSIs were randomly sorted to test IT and POS.


Skewness and kurtosis statistics showed that the
Each synthetic bone cube was placed in a base that variables were normally distributed. Because of signif-
secured five of its sides. The sixth side was secured icant interactions, separate analyses of variance of
with a jig attached to the base, which also served as a each bone type were performed to compare the three
guide for inserting the MSIs into the center of the bone MSIs, followed by Bonferroni post hoc tests. Separate
cube. The jig was attached to a motor, which rotated analyses of each MSI were also performed for evaluate
the bone cube and automatically inserted the MSIs at a the effects of cortical thickness and density.
constant speed of nine revolutions per minute.
A drill press was modified to secure a Mecmesin RESULTS
Advanced Force and Torque Indicator (Mecmesin Ltd,
West Sussex, UK), which measured IT. The Torque The IT of the MSIs ranged from 4.6 Ncm to 11.3 Ncm
Indicator and the MSI were lowered onto the rotating (Table 2). The 3-mm MSIs had significantly (P , .001)
bone cube with a 3-lb weight attached to the arm of the lower ITs than the 6-mm MSIs; the 3 mm MSIs with
drill press. A video camera recorded IT; the video was outer diameters of 1.75 mm and 2 mm had ITs that
used to determine IT at the point of initial engagement were 26%–33% and 15%–30% less than the 6-mm
and when 25%, 50%, 75%, and 100% of the MSIs’ MSIs. Except for the cortical bone that was 1-mm thick
threaded portions had been inserted. and had a density of 0.64 g/cc, the wider (2 mm) 3-mm
MSIs showed significantly higher IT than the narrower
(1.75 mm) 3-mm MSIs.
Pullout Strength
Depending on the MSI and material properties of the
To evaluate POS, each bone block, along with an bone, POS at failure ranged from 6.7–34.1 kg
embedded MSI, was placed in a custom metal base of (Table 3). POS was significantly (P , .001) lower for
approximately the same dimensions as the bone the 3-mm than for the 6-mm MSIs. The POSs of 3-mm
blocks and secured with a lid. Pullout was performed MSIs with an outer diameter of 1.75 mm and 2.0 mm
by attaching an adapter to the miniscrew head; the were 69%–72% and 62%–65% lower, respectively,
adapter was secured to an Instron machine model than the POS of the 6-mm MSIs. POS was also
1011 (Instron Corp, Canton, MA), which exerted a significantly lower (13%–21%) for the narrower
vertical force (5 mm/min) parallel to the long axis of the (1.75 mm) than wider (2 mm) 3-mm MSIs.

Table 3. Analyses of Variance (ANOVAs) Evaluating Differences in Pullout Strength (kg) of Miniscrew Implant (MSI) #1 (6-mm Length,
1.75-mm Outer Diameter), MSI #2 (3-mm Length, 1.75-mm Outer Diameter), and MSI #3 (3-mm Length, 2-mm Outer Diameter) With Post Hoc
Pairwise Comparisons
Post Hoc Probabilities and Relative Differences
MSI #1, MSI #2, MSI #3, ANOVAs,
Density Thickness Mean 6 SD Mean 6 SD Mean 6 SD F Value/Prob MSI # 2 vs 1 MSI # 2 vs 3 MSI # 3 vs 1
0.56 g/cc 1 mm 23.79 6 1.45 6.67 6 0.79 8.47 6 1.07 1234.15/,.001 ,.001; Q71.96% ,.001; Q21.25% ,.001; Q64.40%
0.56 g/cc 2 mm 27.43 6 1.36 8.61 6 0.61 9.94 6 0.86 2017.36/,.001 ,.001; Q68.61% .001; Q13.38% ,.001; Q63.76%
0.64 g/cc 1 mm 26.02 6 1.78 8.15 6 0.76 10.01 6 0.78 1142.05/,.001 ,.001; Q68.68% ,.001; Q18.58% ,.001; Q61.53%
0.64 g/cc 2 mm 34.11 6 0.82 9.78 6 0.61 11.91 6 0.81 5780.34/,.001 ,.001; Q71.33% ,.001; Q17.88% ,.001; Q65.08%

Angle Orthodontist, Vol 82, No 4, 2012


EFFECTS OF SCREW AND HOST FACTORS ON STABILITY 607

Table 4. Analyses of Variance Evaluating the Effects of Cortical Bone Thickness (1 vs 2 mm) and Density (0.56 vs 0.64 g/cc) on Insertion
Torque for Each Miniscrew Implant (MSI)
Effect F Value Prob Differences in Ncm at 100% of Insertion
MSI #1 (6-mm length, 1.75-mm outer diameter)
Thickness 158.54 ,.001 1 mm was 2.91 Ncm (Q27.10%) less than 2 mm
Density 69.39 ,.001 0.56 g/cc was 1.92 Ncm (Q18.78%) less 0.64 g/cc
MSI #2 (3-mm length, 1.75-mm outer diameter)
Thickness 330.01 ,.001 1 mm was 2.40 Ncm (Q30.82%) less than 2 mm
Density 100.37 ,.001 0.56 g/cc was 1.33 Ncm (Q18.32%) less than 0.64 g/cc
MSI #3 (3-mm length, 2-mm outer diameter)
Thickness 163.64 ,.001 1 mm was 2.76 Ncm (Q31.73%) less than 2 mm
Density 20.60 ,.001 0.56 g/cc was 0.98 Ncm (Q12.53%) less than 0.64 g/cc

Bone Characteristics and the 3-mm long 2.0-mm wide MSIs (r 5 .65;
P , .001). After controlling for cortical thickness and
IT and POS were significantly (P , .001) greater for
density, none of the MSIs showed statistically signif-
MSIs placed in 2-mm- than in 1-mm-thick cortical bone
icant correlations between IT and POS.
(Table 4). Reducing cortical thickness by 50% (from
2 mm to 1 mm) decreased IT by 27.1%–31.7%.
Reducing density by 14% (from 0.64 g/cc to 0.56 g/cc) DISCUSSION
decreased IT by 12.5%–18.8%. While the 3-mm MSIs showed less IT and POS than
IT increased in a curvilinear fashion as the MSIs the 6-mm MSIs did, the values were consistently
were inserted, with the greatest increases generally above limits previously recommended for stability. IT in
occurring during the last 25% of MSI insertion the present study was consistently greater than the
(Figure 2). The effects of cortical thickness and density 4 Ncm needed to provide sufficient anchorage for
(q IT with q thickness and q density) were well MSIs.26 More importantly, the pullout forces of the
established after the MSI had been inserted 50% of its 3-mm MSIs, especially those 2-mm wide, were also all
length and increased thereafter. substantially above the ranges of orthodontic forces
POS was also significantly (P , .001) greater when typically applied for tooth movements (0.03–0.40 kg)
MSIs were placed in thicker, more dense, cortical bone and skeletal changes (0.5–1 kg).27,28 Importantly, MSIs
(Table 5). The POS of MSIs placed into the 1-mm- are not typically subjected to axial loads such as those
thick cortex were 1.7–5.9 kg or 15.4%–19.1% less
used in the present experiment to standardize testing
than when placed into 2-mm-thick cortical bone.
techniques; MSIs might be expected to fail at even
Compared with more dense cortical bone, the POS
lower loads when forces are delivered from other
was reduced 1.8–4.5 kg or 14.8%–16.1% when placed
directions. Even after reducing POS by 34%, as
in less dense bone.
suggested to more closely approximate the POSs of
the same screws tested in the tangential (cantilever)
Intercorrelations mode,29 the stability of the 3-mm MSIs is still more than
IT and POS were significantly and positively sufficient to withstand orthodontic loads (Figure 3).
correlated for the 6-mm MSIs (r 5 .73; P , .001), There have been in vivo studies that support the
the 3-mm long 1.75-mm wide MSIs (r 5 .81; P , .001), clinical use of the 3-mm MSIs. Mortensen et al.19

Table 5. Analyses of Variance Evaluating the Effects of Cortical Bone Thickness (1 vs 2 mm) and Density (0.56 vs 0.64 g/cc) on Pullout
Strength for Each Miniscrew Implant (MSI)
Effect F Value Prob Differences in kg
MSI #1 (6-mm length, 1.75-mm outer diameter)
Thickness 190.23 ,.001 1 mm was 5.87 kg (Q19.07%) less than 2 mm
Density 109.82 ,.001 0.56 g/cc was 4.46 kg (Q14.82%) less than 0.64 g/cc
MSI #2 (3-mm length, 1.75-mm outer diameter)
Thickness 117.15 ,.001 1 mm was 1.78 kg (Q19.40%) less than 2 mm
Density 64.76 ,.001 0.56 g/cc was 1.33 kg (Q14.83%) less than 0.64 g/cc
MSI #3 (3-mm length, 2-mm outer diameter)
Thickness 65.08 ,.001 1 mm was 1.69 kg (Q15.43%) less than 2 mm
Density 70.78 ,.001 0.56 g/cc was 1.76 kg (Q16.06%) less than 0.64 g/cc

Angle Orthodontist, Vol 82, No 4, 2012


608 SHAH, BEHRENTS, KIM, KYUNG, BUSCHANG

Figure 3. Temporal changes in insertion torque from the initial engagement through 25%, 50%, 75%, and 100% of engagement for
(A) 6-mm-long, 1.75-mm-wide miniscrew implants (MSIs); (B) 3-mm-long, 1.75-mm-wide MSIs; and (C) 3-mm-long, 2.0-mm-wide MSIs.

Angle Orthodontist, Vol 82, No 4, 2012


EFFECTS OF SCREW AND HOST FACTORS ON STABILITY 609

bone degeneration in the furcation area, ankylosis, or a


total lack of healing associated with inflammatory
infiltrate or pulpal invasion.17,18 The small increase in
the outer diameter of a 3-mm MSI substantially
compensates for the reduction in its length, providing
an alternative that potentially avoids damage to the
teeth and surrounding structures.
Significant increases of IT and POS occurred when
MSIs were inserted into thicker cortical thickness.
Decreasing cortical thickness by 50% (2 mm vs 1 mm)
reduced IT and POS by 27%–32% and 15%–19%,
Figure 4. Pullout strengths (means 6 2 standard deviations) of respectively. Motoyoshi et al.13 and Huja et al.3 previously
6-mm and 3-mm miniscrew implants. showed decreases in IT and POS with decreases in
cortical thickness. Because the effects of cortical
recently reported success rates of 60% for 3-mm long, thickness were relatively greater on IT than POS, care
1.75-mm wide, mandibular MSIs loaded with 900 or must be taken when MSIs are placed in excessively thick
600 g of force in dogs. After they eliminated the screws cortical bone. For example, the cortical bone in the
whose tips had sheared off during insertion and all the mandibular posterior region can be up to 3-mm thick33–35;
MSIs from one unusually active dog that regularly thicker bone might be expected to increase strains and
chewed its bowl and cage bars, the net success rates microfractures during insertion, which could affect
increased to 95.2%. In a series of experiments using healing and compromise secondary stability.
rabbits, Liu et al. loaded 170 MSIs that were 3-mm The IT of MSIs placed in high-density cortical bone
long, 1.75-mm wide with continuous forces of up 200 g was 13%–19% greater than IT of screws placed in
and had an overall success rate of 91.1%.20–22 low-density cortical bone; POS was 15%–16% greater.
The 3-mm MSIs with 2-mm outer diameters provid- Increases in IT with greater bone density are well
ed greater primary stability than those with a 1.75-mm established in the prosthodontic and orthodontic litera-
outer diameter. When inserted into bone with a cortical ture.15,16,36,37 Greater bone density implies greater bone
density of 0.56 g/cc, the 0.25-mm difference in width quantity, which requires higher torsional forces to
resulted in a 12.4%–14.7% difference in IT. The positive advance the MSIs during insertion.24 Greater amounts
effects that the outer diameter has on primary stability of of bone also increase the amount of bone-to-implant
MSIs have been previously emphasized.5,30,31 Because contact and greater engagement of bone by MSI threads,
the wider outer diameter displaces more bone during both of which contribute to increases in the POS.36
insertion, it produces greater friction at the bone-screw
interface, leading to greater IT. Interestingly, the same
screws showed much smaller (5%–8.8%) differences in CONCLUSIONS
IT when they were inserted into denser cortical bone. N The shorter 3-mm MSIs had ITs and POSs that were
This indicates that increases in purchase power 26%–33% and 69%–72% lower, respectively, than 6-
associated with increases in the outer diameter of MSIs mm MSIs.
have less of an effect on denser bone. N Increasing the outer diameter by 0.25 mm signif-
Importantly, differences in POS related to MSI icantly increased the primary stability; the 3 mm
diameter were greater than the differences in IT MSIs that were 1.75-mm wide had ITs and POSs
(Figure 4). This indicates that the positive effects of that were 12%–14% and 13%–21% lower, respec-
the wider MSIs (ie, greater resistance to pullout) tively, than the 2.0-mm-wide MSIs.
outweigh the potentially negative effects (ie, increased N Decreasing cortical thickness by 50% produced IT
IT). These differences emphasize the importance of values that were 27%–32% lower and POS values
measuring both insertion torque and pullout in exper- that were 15%–19% lower.
iments performed to optimize MSI designs.
N Decreasing cortical density from 0.64 g/cc to 0.56 g/cc
The results of this study indicate that the wider 3-mm
(12.5%) decreased IT and POS by 12%–19% and 5%–
MSIs provide a feasible alternative to the longer MSIs
16%, respectively.
typically used by orthodontists. The drawbacks of
longer screws pertain to the surrounding anatomic
limitations.32,33 When MSIs are inserted into the REFERENCES
periosteum or teeth, abnormal or no healing often 1. Kyung H, Park H, Bae S, Sung J, Kim I. Development of
takes place; it is not uncommon to find a lack of orthodontic micro-implants for intraoral anchorage. J Clin
periodontal ligament or lack of bone regeneration, Orthod. 2003;37:321–328.

Angle Orthodontist, Vol 82, No 4, 2012


610 SHAH, BEHRENTS, KIM, KYUNG, BUSCHANG

2. Wilmes B, Rademacher C, Olthoff G, Drescher D. Param- 19. Mortensen MG, Buschang PH, Oliver DR, Kyung HM,
eters affecting primary stability of orthodontic mini-implants. Behrents RG. Stability of immediately loaded 3- and 6-mm
J Orofac Orthop. 2006;67:162–174. miniscrew implants in beagle dogs—a pilot study.
3. Huja SS, Litsky AS, Beck FM, Johnson KA, Larsen PE. Pull- Am J Orthod Dentofacial Orthop. 2009;136:251–259.
out strength of monocortical screws placed in the maxillae 20. Liu SS, Kyung H, Buschang PH. Continuous forces are
and mandibles of dogs. Am J Orthod Dentofacial Orthop. more effective than intermittent forces in expanding sutures.
2005;127:307–313. Eur J Orthod. 2010;32:371–380.
4. Hughes AN, Jordan BA. The mechanical properties of 21. Liu SS, Opperman LA, Buschang PH. Effects of recombi-
surgical bone screws and some aspects of insertion nant human bone morphogenetic protein-2 on midsagittal
practice. Injury. 1972;4:25–38. sutural bone formation during expansion. Am J Orthod
5. DeCoster TA, Heetderks DB, Downey DJ, Ferries JS, Jones Dentofacial Orthop. 2009;136:768.e1–8.
W. Optimizing bone screw pullout force. J Orthop Trauma. 22. Liu SS, Opperman LA, Kyung HM, Buschang PH. Is there
1990;4:169–174. an optimal force level for sutural expansion? Am J Orthod
6. Hitchon PW, Brenton MD, Coppes JK, From AM, Torner JC. Dentofacial Orthop. 2011;139:446–455.
Factors affecting the pullout strength of self-drilling and self- 23. Schwartz-Dabney CL, Dechow PC. Edentulation alters
tapping anterior cervical screws. Spine. 2003;28:9–13. material properties of cortical bone in the human mandible.
7. Chen C, Chang C, Hsieh C, et al. The use of microimplants J Dent Res. 2002;81:613–617.
in orthodontic anchorage. J Oral Maxillofac Surg. 2006;64: 24. Carano A, Lonardo P, Velo S, Incorvati C. Mechanical
1209–1213. properties of three different commercially available mini-
8. Lim JK, Kim WS, Kim IK, Son CY, Byun HI. Three screws for skeletal anchorage. Prog Orthod. 2005;6:82–97.
dimensional finite element method for stress distribution on 25. Kido H, Schulz EE, Kumar A, Lozada J, Saha S. Implant
the length and diameter of orthodontic miniscrew and diameter and bone density: effect on initial stability and
cortical bone thickness. Korea J Orthod. 2003;33:11–20. pull-out resistance. J Oral Implantol. 1997;23:163–169.
9. Kuroda S, Sugawara Y, Deguchi T, Kyung H, Takano- 26. Motoyoshi M, Uemura M, Ono A, et al. Factors affecting the
Yamamoto T. Clinical use of miniscrew implants as ortho- long-term stability of orthodontic mini-implants. Am J Orthod
Dentofacial Orthop. 2010;137:588.e1–5.
dontic anchorage: success rates and postoperative discom-
27. Ren Y, Maltha JC, Kuijpers-Jagtman AM. Optimum force
fort. Am J Orthod Dentofacial Orthop. 2007;131:9–15.
magnitude for orthodontic tooth movement: a systematic
10. Salmória KK, Tanaka OM, Guariza-Filho O, et al. Insertional
literature review. Angle Orthod. 2003;73:86–92.
torque and axial pull-out strength of mini-implants in
28. Proffit W. Contemporary Orthodontics. 4th ed. St Louis, Mo:
mandibles of dogs. Am J Orthod Dentofacial Orthop. 2008;
Mosby; 2007.
133:790.e15–22.
29. Pierce WA, Sucato DJ, Young S, Picetti G, Morgan DM.
11. Dalstra M, Cattaneo PM, Melsen B. Load transfer of miniscrews
Axial and tangential pullout strength of uni-cortical and bi-
for orthodontic anchorage. J Orthod. 2004;1:53–62.
cortical anterior instrumentation screws. Trans Orthop Res
12. Motoyoshi M, Inaba M, Ono A, Ueno S, Shimizu N. The effect Soc. 2003;28:abstract 543.
of cortical bone thickness on the stability of orthodontic 30. Morarend C, Qian F, Marshall SD, et al. Effect of screw
mini-implants and on the stress distribution in surrounding diameter on orthodontic skeletal anchorage. Am J Orthod
bone. Int J Oral Maxillofac Surg. 2009;38:13–18. Dentofacial Orthop. 2009;136:224–229.
13. Motoyoshi M, Yoshida T, Ono A, Shimizu N. Effect of 31. Lim S, Cha J, Hwang C. Insertion torque of orthodontic
cortical bone thickness and implant placement torque on miniscrews according to changes in shape, diameter and
stability of orthodontic mini-implants. Int J Oral Maxillofac length. Angle Orthod. 2008;78:234–240.
Implants. 2007;22:779–784. 32. Park J, Cho HJ. Three-dimensional evaluation of interradic-
14. Beer A, Gahleitner A, Holm A, Tschabitscher M, Homolka P. ular spaces and cortical bone thickness for the placement
Correlation of insertion torques with bone mineral density and initial stability of microimplants in adults. Am J Orthod
from dental quantitative CT in the mandible. Clin Oral Dentofacial Orthop. 2009;136:314.e1–12.
Implants Res. 2003;14:616–620. 33. Poggio PM, Incorvati C, Velo S, Carano A. ‘‘Safe zones’’: a
15. Hung E, Oliver D, Kim KB, Kyung H, Buschang PH. Effects guide for miniscrew positioning in the maxillary and
of pilot hole size and bone density on miniscrew implants’ mandibular arch. Angle Orthod. 2006;76:191–197.
stability. Clin Implant Dent Relat Res. In press. 34. Ono A, Motoyoshi M, Shimizu N. Cortical bone thickness in
16. Wang Z, Zhao Z, Xue J, et al. Pullout strength of miniscrews the buccal posterior region for orthodontic mini-implants.
placed in anterior mandibles of adult and adolescent dogs: Int J Oral Maxillofac Surg. 2008;37:334–340.
a microcomputed tomographic analysis. Am J Orthod 35. Farnsworth D, Rossouw PE, Ceen RF, Buschang PH.
Dentofacial Orthop. 2010;137:100–107. Cortical bone thickness at common mini-screw implant
17. Brisceno CE, Rossouw PE, Carrillo R, Spears R, Buschang placement sites. Am J Orthod Dentofac Orthop. 2011;139:
PH. Healing of the roots and surrounding structures after 495–503.
intentional damage with miniscrew implants. Am J Orthod 36. Friberg B, Sennerby L, Gröndahl K, et al. On cutting torque
Dentofacial Orthop. 2009;135:292–301. measurements during implant placement: a 3-year clinical
18. Hembree M, Buschang PH, Carrillo R, Spears R, Rossouw prospective study. Clin Implant Dent Relat Res. 1999;1:75–83.
PE. Effects of intentional damage of the roots and 37. Cha J, Kil J, Yoon T, Hwang C. Miniscrew stability evaluated
surrounding structures with miniscrew implants. with computerized tomography scanning. Am J Orthod
Am J Orthod Dentofacial Orthop. 2009;135:280.e1. Dentofacial Orthop. 2010;137:73–79.

Angle Orthodontist, Vol 82, No 4, 2012

You might also like