Professional Documents
Culture Documents
Effects of screw and host factors on insertion torque and pullout strength
Ankit H. Shaha; Rolf G. Behrentsb; Ki Beom Kimc; Hee-Moon Kyungd; Peter H. Buschange
ABSTRACT
Objective: To experimentally study the effects of altering implant length, outer diameter, cortical
bone thickness, and cortical bone density on the primary stability of orthodontic miniscrew implants
(MSIs).
Materials and Methods: Maximum insertion torque (IT) and pullout strength (POS) of 216 MSIs
were measured in synthetic bone with different cortical densities (0.64 g/cc or 0.55 g/cc) and
cortical thicknesses (1 mm or 2 mm). Three MSIs were evaluated: 6-mm long/1.75-mm outer
diameter, 3-mm long/1.75-mm outer diameter, and 3-mm long/2.0-mm outer diameter. To test
POS, a vertical force was applied at the rate of 5 mm/min until failure occurred.
Results: The 6-mm MSIs displayed significantly (P , .001) higher IT and POS than the 3-mm
MSIs did. The 3-mm MSIs with 2.0-mm outer diameters showed significantly higher (P , .001) IT
and POS than the 3-mm MSIs with 1.75-mm outer diameters. The IT and POS were significantly
(P , .001) greater for the MSIs placed in thicker and denser cortical bone.
Conclusion: Both outer diameter and length affect the stability of MSIs. Increases in cortical bone
thickness and cortical bone density increase the primary stability of the MSIs. (Angle Orthod.
2012;82:603–610.)
KEY WORDS: Miniscrew implants; Insertion torque; Pullout strength; Cortex; Density
Figure 1. (A) Six-millimeter-long miniscrew implant (MSI) with an outer diameter of 1.75 mm. (B) Three-millimeter-long MSI with an outer
diameter of 1.75 mm. (C) Three-millimeter-long MSI with an outer diameter of 2 mm.
Figure 2. Synthetic bone cubes used showing (A) cortical and cancellous layers and (B) miniscrew implant.
cortical bone thickness, and cortical bone density on and the 3-mm MSIs were 5.5 mm and 2.5 mm,
the primary stability of orthodontic MSIs. Given the respectively.
potential root damage that can be caused by MSIs,17,18
it is especially important to determine the stability of Synthetic Bone
the shorter 3-mm MSIs that have been recently
introduced.19–22 The secondary aim was to determine Because of the variability of cadaver and animal
if cortical thickness and density interact with MSI bone,23 synthetic bone has become the standard for
length and diameter. evaluating the primary stability of endosseous implants
and MSIs because of its uniform material properties.5,24
MATERIALS AND METHODS Blocks of synthetic polyurethane bone (Sawbones,
Vashon, WA) were used to test the IT and POS of the
Three different MSIs were specifically fabricated by MSIs. The density of the cortical layer was 0.64 g/cc or
Dentos (Daegu, Korea). They were either 3-mm or 6- 0.55 g/cc; the cortical density of the human mandible
mm long; the 6-mm MSIs had an outer diameter of has been reported to be 0.64 g/cc.25 Density of the
1.75 mm; the 3-mm MSIs had outer diameters of either cancellous layer was 0.48 g/cc for all of the specimens.
1.75 mm or 2 mm. Other than length and outer The cortical layers were either 1-mm or 2-mm thick.
diameter, the MSIs were identical in terms of all other
design features (Figure 1); all inner diameters were
Testing Groups
1.5 mm, all had 0.5-mm pitch, the apical 1.5 mm of all
MSIs was tapered, all MSIs were self-drilling, and all The larger bone blocks were cut into 11-mm cubes
were self-tapping. The threaded portions of the 6-mm for testing purposes (Figure 2). The three MSIs were
Table 1. Groupings of the Three Miniscrew Implants According to Cortical Bone Thickness (CT) and Cortical Bone Density (CD)
Group 1 Group 2 Group 3 Group 4
6 mm/1.75 mm 6 mm/1.75 mm 6 mm/1.75 mm 6 mm/1.75 mm
CT 5 1 mm CT 5 1 mm CT 5 2 mm CT 5 2 mm
CD 5 0.56 g/cc CD 5 0.64 g/cc CD 5 0.56 g/cc CD 5 0.64 g/cc
Group 5 Group 6 Group 7 Group 8
3 mm/1.75 mm 3 mm/1.75 mm 3 mm/1.75 mm 3 mm/1.75 mm
CT 5 1 mm CT 5 1 mm CT 5 2 mm CT 5 2 mm
CD 5 0.56 g/cc CD 5 0.64 g/cc CD 5 0.56 g/cc CD 5 0.64 g/cc
Group 9 Group 10 Group 11 Group 12
3 mm/2 mm 3 mm/2 mm 3 mm/2 mm 3 mm/2 mm
CT 5 1 mm CT 5 1 mm CT 5 2 mm CT 5 2 mm
CD 5 0.56 g/cc CD 5 0.64 g/cc CD 5 0.56 g/cc CD 5 0.64 g/cc
Table 2. Analysis of Variance (ANOVA) of Differences in Insertion Torque (Ncm), Measured at 100% of Insertion, of Miniscrew Implant (MSI)
#1 (6-mm Length, 1.75-mm Outer Diameter), MSI #2 (3-mm Length, 1.75-mm Outer Diameter), and MSI #3 (3-mm Length, 2-mm Outer
Diameter), With Post Hoc Pairwise Comparisons
Post Hoc Probabilities and Relative Differences
MSI #1, MSI #2, MSI #3, ANOVAs,
Density Thickness Mean 6 SD Mean 6 SD Mean 6 SD F Value/Prob MSI # 2 vs 1 MSI # 2 vs 3 MSI # 3 vs 1
0.56 g/cc 1 mm 6.43 6 0.66 4.64 6 0.54 5.44 6 0.72 34.80/,.001 ,.001; Q27.84% .001; Q14.71% ,.001; Q15.4%
0.56 g/cc 2 mm 10.19 6 0.75 7.22 6 0.66 8.24 6 0.88 69.91/,.001 ,.001; Q29.15% .001; Q12.38% ,.001; Q19.14%
0.64 g/cc 1 mm 9.21 6 1.07 6.14 6 0.36 6.46 6 0.66 90.10/,.001 ,.001; Q33.33% .437; Q4.95% ,.001; Q29.86%
0.64 g/cc 2 mm 11.26 6 0.98 8.37 6 0.62 9.18 6 1.3 39.38/,.001 ,.001; Q25.67% .048; Q8.82% ,.001; Q18.47%
inserted into four different types of bone (two densities MSI until failure occurred. Peak load at failure of the
and two cortical thicknesses), resulting in 12 test MSI was recorded in kilograms.
groups. There were a total of 216 MSIs, with 18 MSIs
randomly allocated to each group (Table 1). Statistical Analysis
Table 3. Analyses of Variance (ANOVAs) Evaluating Differences in Pullout Strength (kg) of Miniscrew Implant (MSI) #1 (6-mm Length,
1.75-mm Outer Diameter), MSI #2 (3-mm Length, 1.75-mm Outer Diameter), and MSI #3 (3-mm Length, 2-mm Outer Diameter) With Post Hoc
Pairwise Comparisons
Post Hoc Probabilities and Relative Differences
MSI #1, MSI #2, MSI #3, ANOVAs,
Density Thickness Mean 6 SD Mean 6 SD Mean 6 SD F Value/Prob MSI # 2 vs 1 MSI # 2 vs 3 MSI # 3 vs 1
0.56 g/cc 1 mm 23.79 6 1.45 6.67 6 0.79 8.47 6 1.07 1234.15/,.001 ,.001; Q71.96% ,.001; Q21.25% ,.001; Q64.40%
0.56 g/cc 2 mm 27.43 6 1.36 8.61 6 0.61 9.94 6 0.86 2017.36/,.001 ,.001; Q68.61% .001; Q13.38% ,.001; Q63.76%
0.64 g/cc 1 mm 26.02 6 1.78 8.15 6 0.76 10.01 6 0.78 1142.05/,.001 ,.001; Q68.68% ,.001; Q18.58% ,.001; Q61.53%
0.64 g/cc 2 mm 34.11 6 0.82 9.78 6 0.61 11.91 6 0.81 5780.34/,.001 ,.001; Q71.33% ,.001; Q17.88% ,.001; Q65.08%
Table 4. Analyses of Variance Evaluating the Effects of Cortical Bone Thickness (1 vs 2 mm) and Density (0.56 vs 0.64 g/cc) on Insertion
Torque for Each Miniscrew Implant (MSI)
Effect F Value Prob Differences in Ncm at 100% of Insertion
MSI #1 (6-mm length, 1.75-mm outer diameter)
Thickness 158.54 ,.001 1 mm was 2.91 Ncm (Q27.10%) less than 2 mm
Density 69.39 ,.001 0.56 g/cc was 1.92 Ncm (Q18.78%) less 0.64 g/cc
MSI #2 (3-mm length, 1.75-mm outer diameter)
Thickness 330.01 ,.001 1 mm was 2.40 Ncm (Q30.82%) less than 2 mm
Density 100.37 ,.001 0.56 g/cc was 1.33 Ncm (Q18.32%) less than 0.64 g/cc
MSI #3 (3-mm length, 2-mm outer diameter)
Thickness 163.64 ,.001 1 mm was 2.76 Ncm (Q31.73%) less than 2 mm
Density 20.60 ,.001 0.56 g/cc was 0.98 Ncm (Q12.53%) less than 0.64 g/cc
Bone Characteristics and the 3-mm long 2.0-mm wide MSIs (r 5 .65;
P , .001). After controlling for cortical thickness and
IT and POS were significantly (P , .001) greater for
density, none of the MSIs showed statistically signif-
MSIs placed in 2-mm- than in 1-mm-thick cortical bone
icant correlations between IT and POS.
(Table 4). Reducing cortical thickness by 50% (from
2 mm to 1 mm) decreased IT by 27.1%–31.7%.
Reducing density by 14% (from 0.64 g/cc to 0.56 g/cc) DISCUSSION
decreased IT by 12.5%–18.8%. While the 3-mm MSIs showed less IT and POS than
IT increased in a curvilinear fashion as the MSIs the 6-mm MSIs did, the values were consistently
were inserted, with the greatest increases generally above limits previously recommended for stability. IT in
occurring during the last 25% of MSI insertion the present study was consistently greater than the
(Figure 2). The effects of cortical thickness and density 4 Ncm needed to provide sufficient anchorage for
(q IT with q thickness and q density) were well MSIs.26 More importantly, the pullout forces of the
established after the MSI had been inserted 50% of its 3-mm MSIs, especially those 2-mm wide, were also all
length and increased thereafter. substantially above the ranges of orthodontic forces
POS was also significantly (P , .001) greater when typically applied for tooth movements (0.03–0.40 kg)
MSIs were placed in thicker, more dense, cortical bone and skeletal changes (0.5–1 kg).27,28 Importantly, MSIs
(Table 5). The POS of MSIs placed into the 1-mm- are not typically subjected to axial loads such as those
thick cortex were 1.7–5.9 kg or 15.4%–19.1% less
used in the present experiment to standardize testing
than when placed into 2-mm-thick cortical bone.
techniques; MSIs might be expected to fail at even
Compared with more dense cortical bone, the POS
lower loads when forces are delivered from other
was reduced 1.8–4.5 kg or 14.8%–16.1% when placed
directions. Even after reducing POS by 34%, as
in less dense bone.
suggested to more closely approximate the POSs of
the same screws tested in the tangential (cantilever)
Intercorrelations mode,29 the stability of the 3-mm MSIs is still more than
IT and POS were significantly and positively sufficient to withstand orthodontic loads (Figure 3).
correlated for the 6-mm MSIs (r 5 .73; P , .001), There have been in vivo studies that support the
the 3-mm long 1.75-mm wide MSIs (r 5 .81; P , .001), clinical use of the 3-mm MSIs. Mortensen et al.19
Table 5. Analyses of Variance Evaluating the Effects of Cortical Bone Thickness (1 vs 2 mm) and Density (0.56 vs 0.64 g/cc) on Pullout
Strength for Each Miniscrew Implant (MSI)
Effect F Value Prob Differences in kg
MSI #1 (6-mm length, 1.75-mm outer diameter)
Thickness 190.23 ,.001 1 mm was 5.87 kg (Q19.07%) less than 2 mm
Density 109.82 ,.001 0.56 g/cc was 4.46 kg (Q14.82%) less than 0.64 g/cc
MSI #2 (3-mm length, 1.75-mm outer diameter)
Thickness 117.15 ,.001 1 mm was 1.78 kg (Q19.40%) less than 2 mm
Density 64.76 ,.001 0.56 g/cc was 1.33 kg (Q14.83%) less than 0.64 g/cc
MSI #3 (3-mm length, 2-mm outer diameter)
Thickness 65.08 ,.001 1 mm was 1.69 kg (Q15.43%) less than 2 mm
Density 70.78 ,.001 0.56 g/cc was 1.76 kg (Q16.06%) less than 0.64 g/cc
Figure 3. Temporal changes in insertion torque from the initial engagement through 25%, 50%, 75%, and 100% of engagement for
(A) 6-mm-long, 1.75-mm-wide miniscrew implants (MSIs); (B) 3-mm-long, 1.75-mm-wide MSIs; and (C) 3-mm-long, 2.0-mm-wide MSIs.
2. Wilmes B, Rademacher C, Olthoff G, Drescher D. Param- 19. Mortensen MG, Buschang PH, Oliver DR, Kyung HM,
eters affecting primary stability of orthodontic mini-implants. Behrents RG. Stability of immediately loaded 3- and 6-mm
J Orofac Orthop. 2006;67:162–174. miniscrew implants in beagle dogs—a pilot study.
3. Huja SS, Litsky AS, Beck FM, Johnson KA, Larsen PE. Pull- Am J Orthod Dentofacial Orthop. 2009;136:251–259.
out strength of monocortical screws placed in the maxillae 20. Liu SS, Kyung H, Buschang PH. Continuous forces are
and mandibles of dogs. Am J Orthod Dentofacial Orthop. more effective than intermittent forces in expanding sutures.
2005;127:307–313. Eur J Orthod. 2010;32:371–380.
4. Hughes AN, Jordan BA. The mechanical properties of 21. Liu SS, Opperman LA, Buschang PH. Effects of recombi-
surgical bone screws and some aspects of insertion nant human bone morphogenetic protein-2 on midsagittal
practice. Injury. 1972;4:25–38. sutural bone formation during expansion. Am J Orthod
5. DeCoster TA, Heetderks DB, Downey DJ, Ferries JS, Jones Dentofacial Orthop. 2009;136:768.e1–8.
W. Optimizing bone screw pullout force. J Orthop Trauma. 22. Liu SS, Opperman LA, Kyung HM, Buschang PH. Is there
1990;4:169–174. an optimal force level for sutural expansion? Am J Orthod
6. Hitchon PW, Brenton MD, Coppes JK, From AM, Torner JC. Dentofacial Orthop. 2011;139:446–455.
Factors affecting the pullout strength of self-drilling and self- 23. Schwartz-Dabney CL, Dechow PC. Edentulation alters
tapping anterior cervical screws. Spine. 2003;28:9–13. material properties of cortical bone in the human mandible.
7. Chen C, Chang C, Hsieh C, et al. The use of microimplants J Dent Res. 2002;81:613–617.
in orthodontic anchorage. J Oral Maxillofac Surg. 2006;64: 24. Carano A, Lonardo P, Velo S, Incorvati C. Mechanical
1209–1213. properties of three different commercially available mini-
8. Lim JK, Kim WS, Kim IK, Son CY, Byun HI. Three screws for skeletal anchorage. Prog Orthod. 2005;6:82–97.
dimensional finite element method for stress distribution on 25. Kido H, Schulz EE, Kumar A, Lozada J, Saha S. Implant
the length and diameter of orthodontic miniscrew and diameter and bone density: effect on initial stability and
cortical bone thickness. Korea J Orthod. 2003;33:11–20. pull-out resistance. J Oral Implantol. 1997;23:163–169.
9. Kuroda S, Sugawara Y, Deguchi T, Kyung H, Takano- 26. Motoyoshi M, Uemura M, Ono A, et al. Factors affecting the
Yamamoto T. Clinical use of miniscrew implants as ortho- long-term stability of orthodontic mini-implants. Am J Orthod
Dentofacial Orthop. 2010;137:588.e1–5.
dontic anchorage: success rates and postoperative discom-
27. Ren Y, Maltha JC, Kuijpers-Jagtman AM. Optimum force
fort. Am J Orthod Dentofacial Orthop. 2007;131:9–15.
magnitude for orthodontic tooth movement: a systematic
10. Salmória KK, Tanaka OM, Guariza-Filho O, et al. Insertional
literature review. Angle Orthod. 2003;73:86–92.
torque and axial pull-out strength of mini-implants in
28. Proffit W. Contemporary Orthodontics. 4th ed. St Louis, Mo:
mandibles of dogs. Am J Orthod Dentofacial Orthop. 2008;
Mosby; 2007.
133:790.e15–22.
29. Pierce WA, Sucato DJ, Young S, Picetti G, Morgan DM.
11. Dalstra M, Cattaneo PM, Melsen B. Load transfer of miniscrews
Axial and tangential pullout strength of uni-cortical and bi-
for orthodontic anchorage. J Orthod. 2004;1:53–62.
cortical anterior instrumentation screws. Trans Orthop Res
12. Motoyoshi M, Inaba M, Ono A, Ueno S, Shimizu N. The effect Soc. 2003;28:abstract 543.
of cortical bone thickness on the stability of orthodontic 30. Morarend C, Qian F, Marshall SD, et al. Effect of screw
mini-implants and on the stress distribution in surrounding diameter on orthodontic skeletal anchorage. Am J Orthod
bone. Int J Oral Maxillofac Surg. 2009;38:13–18. Dentofacial Orthop. 2009;136:224–229.
13. Motoyoshi M, Yoshida T, Ono A, Shimizu N. Effect of 31. Lim S, Cha J, Hwang C. Insertion torque of orthodontic
cortical bone thickness and implant placement torque on miniscrews according to changes in shape, diameter and
stability of orthodontic mini-implants. Int J Oral Maxillofac length. Angle Orthod. 2008;78:234–240.
Implants. 2007;22:779–784. 32. Park J, Cho HJ. Three-dimensional evaluation of interradic-
14. Beer A, Gahleitner A, Holm A, Tschabitscher M, Homolka P. ular spaces and cortical bone thickness for the placement
Correlation of insertion torques with bone mineral density and initial stability of microimplants in adults. Am J Orthod
from dental quantitative CT in the mandible. Clin Oral Dentofacial Orthop. 2009;136:314.e1–12.
Implants Res. 2003;14:616–620. 33. Poggio PM, Incorvati C, Velo S, Carano A. ‘‘Safe zones’’: a
15. Hung E, Oliver D, Kim KB, Kyung H, Buschang PH. Effects guide for miniscrew positioning in the maxillary and
of pilot hole size and bone density on miniscrew implants’ mandibular arch. Angle Orthod. 2006;76:191–197.
stability. Clin Implant Dent Relat Res. In press. 34. Ono A, Motoyoshi M, Shimizu N. Cortical bone thickness in
16. Wang Z, Zhao Z, Xue J, et al. Pullout strength of miniscrews the buccal posterior region for orthodontic mini-implants.
placed in anterior mandibles of adult and adolescent dogs: Int J Oral Maxillofac Surg. 2008;37:334–340.
a microcomputed tomographic analysis. Am J Orthod 35. Farnsworth D, Rossouw PE, Ceen RF, Buschang PH.
Dentofacial Orthop. 2010;137:100–107. Cortical bone thickness at common mini-screw implant
17. Brisceno CE, Rossouw PE, Carrillo R, Spears R, Buschang placement sites. Am J Orthod Dentofac Orthop. 2011;139:
PH. Healing of the roots and surrounding structures after 495–503.
intentional damage with miniscrew implants. Am J Orthod 36. Friberg B, Sennerby L, Gröndahl K, et al. On cutting torque
Dentofacial Orthop. 2009;135:292–301. measurements during implant placement: a 3-year clinical
18. Hembree M, Buschang PH, Carrillo R, Spears R, Rossouw prospective study. Clin Implant Dent Relat Res. 1999;1:75–83.
PE. Effects of intentional damage of the roots and 37. Cha J, Kil J, Yoon T, Hwang C. Miniscrew stability evaluated
surrounding structures with miniscrew implants. with computerized tomography scanning. Am J Orthod
Am J Orthod Dentofacial Orthop. 2009;135:280.e1. Dentofacial Orthop. 2010;137:73–79.