You are on page 1of 6

Reprinted from CLINICAL CHEMISTRY, 39, 1069 (1993).

Copyright 1993 by the American Association of Clinical Chemistry and reprinted by permission of the copyright owner
CLIN.CHEM. 39/6, 1069-1074 (1993)

Analytical Evaluation of i-STAT Portable Clinical Analyzer and Use by Nonlaboratory


Health-Care Professionals
Ellis Jacobs, 1,2,4 Elizabeth Vadasdi, 2 Laszlo Sarkozi, 1,2 and Neville Coman 1–3

We evaluated the performance of the i-STAT Portable turnaround time (1, 6). Various mechanisms have been
Clinical Analyzer, a hand-held instrument that, with its used to meet these clinical needs, e.g., pneumatic tube
current cartridge, analyzes for electrolytes, urea nitrogen, systems to improve specimen transport to centralized
glucose, and hematocrit in ~60 µL of whole blood in ~90 laboratories, establishment of specialized “stat” labora-
s. Accuracy, imprecision, and linearity studies were per- tories, and use of point-of-care testing devices (glucose
formed with aqueous controls and standards and by meters, blood gas/electrolyte analyzers) (7–9).
split-sample analysis. Intrarun imprecision (CV) ranged One of the major impediments to the implementation
from 0.34% to 3.97%. Total imprecision over a 2-month of point-of-care testing, in addition to concerns about
period ranged from 0.42% to 4.83%, with urea nitrogen analytical accuracy and precision, are considerations of
and glucose analyses generating the higher values. Pa- the reliability of these systems in the hands of nontra-
tients’ results from the Portable Clinical Analyzer corre- ditionally trained health-care professionals. Until re-
lated well with those obtained for whole blood or plasma cently, technology has not been able to produce a point-
by the Nova Stat Profile 5, the Beckman Synchron CX3, of-care analytical system that was robust enough to
or the Technicon H1 Hematology Analyzer, with Sy x withstand the rigorous environment associated with
values <0.2 mmol/L for potassium; 1.5 mmol/L for dispersed locations (10, 11). To maintain the desirable
sodium, glucose, and urea nitrogen; <2.4 mmol/L for accuracy and precision, systems have had demanding
chloride; and <2.4% for hematocrit. We also ascertained operational and maintenance requirements. Thus, it
imprecision and accuracy of the system placed in a has been difficult for nurses, physicians, or other non-
cardiothoracic intensive-care unit and operated by laboratory health-care professionals to obtain and main-
nurses. There were no significant differences in either the tain the skills necessary for proper operations and
imprecision or accuracy of the system in this setting. We maintenance of the systems (12). Various technological
conclude that operator technique is not a factor in the innovations have extended the lifetime of electrochem-
analytical performance of the system and that it can be ical sensors, reduced maintenance, and led to the devel-
used by nonlaboratorians with a high degree of confi- opment of single-use and unit-use disposable sensors
dence that reliable results will be obtained. (13, 14). Additionally, in conjunction with reagent sta-
bilization through the use of biosensor technology, the
Indexing Terms: point of care testing • biosensors • electro- system must have built-in software capability for qual-
chemical sensors • emergency analyses • critical-care medi- ity-control (QC) activities and detection of instrument
cine • electroyles • glucose • urea nitrogen malfunctions.
Recently, a new point-of-care testing instrument, the
Changes in medical practice have intensified institu- Portable Clinical Analyzer (PCA; i-STAT Corp., Prince-
tional pressures to achieve clinical efficacy (1, 2). Thus ton, NJ), was introduced into the market (10, 15). This
hospitals are decreasing the admission of patients with system was designed for use by nonlaboratory personnel
nonacute conditions and increasing the proportion of in the critical-care environment of a hospital’s ICUs,
patients admitted for major therapeutic interventions, surgical units, and emergency department. In its cur-
e.g, transplants or open heart surgery, that require rent configuration, this hand-held instrument analyzes
more-intensive support systems (3–5). Additionally, for sodium, potassium, chloride, urea nitrogen, glucose,
there is a push to decrease patients’ overall length of and hematocrit in ~60 µL of whole blood in ~90 s.
stay as well as the use of high-cost resources such as We report here the results of our study of the i-STAT
intensive-care units (ICUs).5 PCA. Correlation studies were performed by comparison
In the laboratory, these various pressures translate with established laboratory methodologies. A system was
into requests to improve service by decreasing test placed in our Cardiothoracic ICU to evaluate its
reliability in the ICU, when operated and maintained
1 Department of Pathology, Mount Sinai & School of Medicine; by nonlaboratory health-care professionals. Precision
2 Center for Clinical Laboratories, The Mount Sinai Hospital, New
York, NY 10029. and accuracy of the system were assessed for utilization
3 Veterans Administration Medical Center, Bronx, NY 10468. by both laboratory technologists and ICU nurses.
4 Address for correspondence: The Mount Sinai Medical Center
STAT Laboratories, Box 1519, One Gustave L. Levy Place, New Materials and Methods
York, NY 10029-6574. System Description
5 Nonstandard abbreviations: PCA, Portable Clinical Analyzer
I C U, i n t e n s ive - c a re unit; QC, quality control; and NCCLS, N a- The PCA is a hand-held system that utilizes a single-
tional Committee for Clinical Laboratory Standards. use disposable cartridge (2.7 x 4.5 x 0.5 cm, w x l x h)
Received September 23,1992; accepted December 11, 1992. containing microfab ri c ated biosensors. Whole blood

CLINICAL CHEMISTRY, Vol. 39, No. 6, 1993 1069


from a capillary fingerstick or a syringe is placed in the the system each day. Specimens were obtained in lith-
c a rt ri d ge, the cart ri d ge is then inserted into the ana- ium heparinized syringes from arterial lines and ana-
ly ze r, and operator and patient identification (numeri- lyzed by the PCA in the ICU before being sent, by
cal only) are entered into the system. Automatically a pneumatic tube, to the stat laboratory. Once in the
calibration solution is released and a one-point calibra- laboratory, the specimens were tested according to the
tion is performed just before sample analysis. The re- protocol previously described.
sponses of the various biosensors are monitored during
the rehydration, calibration, and sample analysis pro- Statistics
cesses. If any of the sensor responses fall outside prede- The National Committee for Clinical Laboratory
termined limits, then the controlling software sup- Standards (NCCLS) EP5 & EP6 protocols for the assess-
presses the output from that sensor. Additionally, ment of imprecision of an analytical system and for the
mechanical and user operational errors are identified assessment of linearity were utilized (16, 17). However,
through the controlling software and, if detected, cause instead of the standard EP5 protocol, i.e., duplicate runs
a suppression of result reporting. twice a day for 20 days, we used an extended protocol of
65 days.
Laboratory Study Protocol Student’s t-test and Demings linear-regression anal-
Syringe specimens, anticoagulated with dry lithium ysis (18) were used in the methodology and specimen-
heparin (3-mL preheparinized syringes; Concord-Por- type comparisons.
tex, Keene, NH) to a final concentration of 30 kIU/L,
Results and Discussion
were selected without conscious bias from those rou-
tinely received in the stat laboratory for analysis. Imprecision and Linearity Studies
These specimens were obtained from either the operat- The precision, or imprecision, of the i-STAT system
ing rooms, the ICU, or the emergency department. The was determined when the system was operated either by
samples were transported to the stat laboratory either traditionally trained laboratory personnel or by Car-
by a high-speed pneumatic tube system or by hand. diothoracic ICU nurses who do not have training to
Once in the laboratory, the samples were remixed by operate laboratory equipment. Within the laboratory,
hand. The whole-blood samples were tested without rotating the use of five PCAs and multiple lots of
delay on the PCA, the Stat Profile 5 (Nova Biomedical, cartridges and using our extended NCCLS EP5 protocol
Waltham, MA), and the H-1 Hematology analyzer yielded a within-run imprecision (CV) of 0.34–3.96%;
(Miles, Tarrytown, NY). Plasma was obtained by cen- the total imprecision ranged from 0.42% to 4.82% (Table
trifuging an aliquot of the whole-blood specimen for 2 1). Sodium, potassium, and chloride had total CVs <2%,
min at 2536 x g in a Microfuge 12 (Beckman Instru- whereas for urea nitrogen and glucose total imprecision
ments, Brea, CA) with a 13.2 fixed-angle rotor. The was between 3.8% and 4.8%. Because an aqueous con-
plasma obtained was then analyzed by the Synchron trol was used, the hematocrit data reflect the stability of
CX3 analyzer (Beckman) and by the PCA. All testing the conductivity electrode used in this measurement.
was done according to manufacturers’ recommended These values are comparable with those obtained by
procedures. All specimens were analyzed in duplicate established laboratory-based analytical systems, i.e.,
and testing was completed within 10 min of specimen 0.3–4.1%.
receipt. Five different PCA units were used during this Even though five PCA analyzers were used through-
study, one on each day of the week, to assess intermeter out the imprecision study, and >250 data points were
variability. Aqueous bi-level QC and linearity test collected per analyte, there were no statistical outliers.
solutions were provided by i-STAT Corp. However, some data points were suppressed by the
analyzer because of some malfunction identified by the
ICU Study Protocol controlling software. The rate of data suppression by
In the Cardiothoracic ICU, eight nurses were given 30 the PCA shown in Table 2 represents the results of
min of instructions on the use and operations of the testing both QC material and patients’ specimens with
PCA. The aqueous QC material was analyzed singly on > 1800 test cartridges, from seven different lots. One or

Table 1. PCA Within-Laboratory Precision Study: Extended NCCLS EP5 Protocol


Level 1 Level 2

Mean S IR(CV, %) STOT(CV, %) Mean SIR(CV, %) S TOT(CV, %)


Sodium, mmol/L 115.0 0.521 (0.45) 0.535 (0.47) 142.5 0.490 (0.34) 0.592 (0.42)
Potassium, mmol/L 5.73 0.079 (1.38) 0.088 (1.53) 2.84 0.046 (1.62) 0.055 (1.93)
Chloride, mmol/L 78.5 0.589 (0.75) 0.671 (0.85) 106.6 0.767 (0.72) 0.822 (0.77)
Urea Nitrogen, mmol/L 18.90 0.496 (2.63) 0.772 (4.08) 22.79 0.514 (2.26) 0.879 (3.85)
Glucose, mmol/L 10.63 0.421 (3.96) 0.513 (4.82) 3.03 0.087 (2.87) 0.146 (4.82)
Hematocrit, %PCV (–3.04)a 0.264 (–8.69) 0.264 (–8.69) ( 3 . 0 3 )a 0.205 (–6.75) 0.205 (–6.75)
SIR, intrarun standard deviation; S TOT, total standard deviation;PCV, packed cell volume.
a
These values represent artefacts from measuring aqueous control material rather than a biological specimen.

1070 CLINICAL CHEMISTRY, Vol.39, No. 6, 1993


tions, linear-regression analysis is performed and tested
Table 2. Data Suppression Rage of PCA Cartridges
for lack of fit to the linear model.
Results suppressed No. (%) of cartridges
To test the linearity of the PCA, we used five aqueous
All tests 24 (1.31)
solutions with concentrations distributed equally across
All tests except glucose 26 (1.42)
Chloride 1 (0.05)
the following ranges (mmol/L): sodium 102-167, potas-
Urea nitrogen 6 (0.32) sium 2.5-8.5, chloride 70-130, glucose 1.93-19.57, and
Glucose 4 (0.22) urea nitrogen 1.29-39.42 (Figure 1). There were no
Hematocrit 11 (0.60) outliers in the data sets. At a significance level of 0.05
and over the range being tested, there was equality of
variance for all the analytes except glucose. Even at P =
more test results were suppressed in 3.94% of the 0.01, glucose failed Cochran’s test of equality—primar-
c a rt ri d ges used. For the majority (69%) of the car- ily because of the increased imprecision of the glucose
t ri d ges that had tests suppressed, the internal moni- sensor at concentrations >11 mmol/L. Even though
toring blocked the reporting of either all the analytes, these variances are significant statistically, there is
or all except glucose. sufficient precision in the data to continue with the
Lot variations. A significant concern with unit-based linearity study.
testing systems, especially with biosensor technology as After performing the linear-regression calculations,
in the PCA, is the va ri ability of calibration and re- we compared the linear “significance” value (G) with
sponse, both within and across various manufacturing the critical F-value, at P = 0.1, for testing the hypoth-
lots of cartridges. By sorting and analyzing the QC data esis of a linear fit. Sodium, potassium, and glucose tests
by cartridge lots, one can assess calibration reproduc- had G-values lower than the critical F-value, and thus
ibility. At least 15 data points were obtained from six were determined to be linear over the ranges tested.
different manufacturing lots of cartridges. Of the with- Over the entire range tested, urea nitrogen was linear
in-lot variations (Table 3), the highest values were but at P = 0.01. Furthermore, visual inspection of the
obtained for glucose and urea nitrogen. In general, the data in Figure 1 reveals a definite break in urea
lot-to-lot variation was excellent for the analytes tested linearity at 23 mmol/L. Over the limited range of 1.3-23
(Table 3). CVs at high concentrations of urea nitrogen mmol/L, urea nitrogen testing was linear at P =
and glucose (CV of within-lot means >3.0%), were 0.1. Chloride had G-values significantly greater (6-13-
substantially greater than with the other analytes fold) than the critical F-values and thus failed the
(<0.6%). Nevertheless, the lot-to-lot variations for all the hypothesis of linear fit. However, for very precise
analytes were well within acceptable clinical limits assays, data with clinically acceptable linearity may be
for overall imprecision and would have had no impact on d e cl a red statistically nonlinear. Furt h e rm o re, upon
patient care. inspecting the plot of data in Figure 1, it is obvious that
Linearity. We assessed the linearity (or proportional- the system is linear over the range of values tested.
ity of response) of the chemistry tests on the PCA by This further proves the need for a visual inspection of
analyzing aqueous standards according to the NCCLS
EP6 protocol (17). This method utilizes linear-regres-
sion analysis as its basis for determination of linearity.
However, in this protocol, the data are subjected to two
verifications before regression analysis is performed.
First is determination of any outliers. Then Cochran’s C
test is applied, to test the hypothesis of equality of
variance across dilutions. After these two determina-

Table 3. Variation of QC Results due to Different Lots


of PCA Test Cartridges
Mean of CV of
Range of lot lot
Control means, Range of means, means,
Analyte level mmol/L lot CVs, % mmol/L %
Sodium 1 114.9–115.4 0.28–0.74 115.1 0.16
2 142.1–142.8 0.28–0.42 142.4 0.22
Potassium 1 5.70–5.75 0.96–2.12 5.73 0.29
2 2.82–2.86 1.59–2.21 2.84 0.58
Chloride 1 78.2–78.9 0.55–1.04 78.6 0.36
2 106.2–107.0 0.50–0.93 106.6 0.27
Urea Nitrogen 1 18.05–19.72 1.71–4.31 18.97 3.08 Fig.1.Linearity of i-STAT Portable Clinical Analyzer for (▲) chloride,
2 21.85–23.75 1.68–3.78 22.94 3.18 (●) sodium, (●) potassium, (▲) urea nitrogen, and (■) glucose
Glucose 1 9.96–10.89 3.36–6.22 10.54 3.14 assays
All testing was performed as described in the test. Each solution was analyzed
2 2.95–3.16 2.55–5.26 3.06 2.15
four times, according to the NCCLS EP6 protocol, and the averages repre-
sented here

CLINICAL CHEMISTRY, Vol.39, No. 2, 1993 1071


the data in addition to the application of mathematical data for the PCA and the Stat Profile 5 were not as
formulas. tight as the correlation with the CX3. Nevertheless,
except for sodium and chloride, the r values were >0.93
Method Comparison and, except for chloride, the slopes were >0.95. The
The analytical performance of the i-STAT system was lower r value for sodium is due to the limitation of the
compared with various established laboratory-based numerical range of the data, rather than a problem
systems. Initially, to ensure that there was no bias due with either testing system. This is seen from the
to sample type, both whole blood and plasma derived difference data, which show an average difference of 2.0
from the same specimen were analyzed by the i-STAT mmol/L in the range of 130-150 mmol/L, the standard
system. The data in Table 4 show the lack of significant deviation of the differences being 2.16. Despite an r
analytical bias due to sample type, with average differ- value of 0.977, a slope of 1.162, and an intercept of
ences ranging from 0.1 mmoVL for potassium to 0.44 -1.102 mmol/L, glucose results on the PCA were
mmoVL for glucose. Other than for glucose values >11 significantly higher than the Stat Profile 5 for samples
mmoVL, the standard deviations of the differences dem- with glucose concentrations >11 mmol/L. This is sim-
onstrate good precision. Demings regression analysis ilar to the results of the sample-type comparison study,
reveals excellent correlations, with slopes between 1.01 which showed decreased precision of the glucose bio-
and 1.06 and r >0.95. Only glucose showed significant sensor at concentrations > 11 mmol/L. The PCA hemat-
scatter about the regression line (Syx = 0.5216 mmol/L). ocrit methodology correlated well with the conductivity
The correlation of results from whole blood analyzed methodology on the Stat Profile 5 system: slope =
on the PCA with whole blood on the Stat Profile 5 or H1 1.119, S y x = 2.06%, and intercept = -3.759%.
analyzer or with plasma analyzed on the Synchron
CX3 is shown in Table 5. In general, there was good Reliability in Point-of-Care Settings
agreement between the PCA and the CX3, with slopes There was no significant difference in the imprecision
between 0.987 and 1.07. The lowest r value and the of the system when it was operated by the Cardiotho-
largest Sy x value were obtained with chloride testing. racic ICU nurses (CV 0.56-4.76%) (Table 6). Addition-
Chloride ion-selective electrode systems are sensitive ally, there was no significant difference in the rate of
to protein effects (19). The CX3 system uses a diluted data suppression by PCAs used by the nurses vs those
sample, which minimizes the impact of specimen pro- used by laboratory personnel. Thus operator technique
tein on the analysis. Glucose results by the PCA were is not a factor in the analytical performance of the
~7% greater than by the CX3 and showed greater i-STAT system.
scatter (Sy x ) than for the other analytes. The hematocrit When the same sample was first analyzed by the
measured by conductivity on the PCA correlated nurses in the Cardiothoracic ICU and then transported
well with the hematocrit calculated from erythrocyte and analyzed within 5 min in the stat laboratory on the
counts and indices obtained on the H1: slope = 1.143 i-STAT system, no significant difference was observed in
and Sy x = 2.32%. the difference and regression analysis (Table 6). The Syx
As would be expected, given the matrix problems values ranged from 0.0792 to 1.2528 mmol/L for the
associated with whole-blood testing, the regression chemistry tests, and was 2.0671% for hematocrit. This

Table 4. PCA Sample-Type Comparison: Whole Blood vs Plasma


Differenceb Regression statistics

Analytea Conc range n Mean SD Slope y-Intercept Sy|x r


Sodium 100–130 5 –0.200 0.447 1.0311 –3.8546 0.6419 0.9723
130–150 198 0.424 0.908
Potassium 2.0–3.0 4 0.125 0.126 1.0304 –0.0617 0.0758 0.9837
3.0–5.0 189 0.057 0.096
5.0–9.0 10 0.100 0.094
Chloride 70–90 1 –1.000 0.0— 1.0620 –5.9309 1.0306 0.9532
90–110 184 0.576 1.462
110–140 18 –0.056 1.259
Urea Nitrogen 0–8.9 145 –0.185 0.388 1.0061 –0.2029 0.4737 0.9971
8.9–21.4 42 0.187 0.789
21.4–35.7 11 0.844 1.154
35.7–53.6 5 –1.142 1.053
Glucose 1.1–5.6 16 0.132 0.156 1.0247 0.0239 0.5216 0.9880
5.6–11.1 122 0.225 0.374
11.1–16.7 49 0.432 1.012
16.7–25.0 18 0.293 1.487
a
Units as in Table 1.
b
Plasma value minus wholeblood value.

1072 CLINICAL CHEMISTRY, Vol.39, No. 6, 1993


Table 5. Method Comparison of the PCA
PCA vs Synchron CX3 PCA vs Stat Profile 5

Differenceb Regression statistics Differenceb Regression statistics

Analytea Conc range n Mean SD Slope y-Inter. S y|x r n Mean SD Slope y-Inter. Sy|x r
Sodium 100–130 4 –0.000 0.816 0.9969 –0.1089 1.1695 0.8650 2 0.500 3.536 0.9489 5.2559 1.5263 0.8377
130–150 185 –0.546 2.032 192 –1.948 2.163
Potassium 2.0–3.0 4 0.025 0.150 0.9695 0.0217 0.0764 0.9780 3 –0.100 0.000 0.9880 –0.0137 0.1124 0.9475
3.0–5.0 169 –0.100 0.095 177 –0.059 0.169
5.0–9.0 10 0.100 0.094 13 –0.108 0.246
Chloride 70–90 1 5.000 0.0— 0.9874 –0.8202 1.6538 0.8170 0 .— .— 0.9265 4.2629 2.3272 0.7516
90–110 145 –1.697 2.822 131 –2.740 2.929
110–140 18 –0.056 1.259 61 –5.541 3.264
Urea Nitrogen 0–8.9 137 –0.122 0.407 0.9756 0.0370 0.5258 0.9966 . —c .— .— .— .— .— .—
8.9–21.4 39 0.073 0.729
21.4–35.7 9 –1.349 1.208
35.7–53.6 4 –0.893 1.220
Glucose 1.1–5.6 14 0.135 0.144 1.0698 –0.3719 0.6206 0.9886 10 –0.039 0.200 1.1615 –1.1018 0.8934 0.9770
5.6–11.1 114 0.163 0.357 125 0.150 0.629
11.1–16.7 50 0.493 0.839 49 1.034 1.086
16.7–25.0 13 1.218 1.820 10 2.222 2.993
Hematocritd 15–30 55 1.418 2.291 1.1432 –2.597 2.3170 0.9102 34 1.418 2.291 1.1189 –3.7590 2.0634 0.9315
30–50 127 2.228 2.658 157 2.228 2.658
50–70 0 .— .— 1 2.000 .—
a
Units as in Table 1.
b
PCA value minus wholeblood value.
c
Urea nitrogen is not available on the Stat Profile 5
d
The comparison method was the H1 analyzer.

Table 6. Use of PCA by Cardiothoracic ICU Nurses and Laboratory Personnel Compared
Imprecision analysis

Level 1 Level 2 Differenceb Regression statisticsc


Conc
Analytea Mean S TOT (CV, %) Mean S TOT (CV,%) range n Mean SD Slope y-Inter. Sy|x r
Sodium 115.0 0.556 (0.48) 142.7 0.795 (0.56) 100–130 3 0.333 0.577 0.9483 7.2761 0.7087 0.9671
130–150 96 0.125 1.059
Potassium 5.69 0.065 (1.14) 2.82 0.047 (1.65) 2.0–3.0 1 0.000 .— 0.9651 0.1365 0.0792 0.9772
3.0–5.0 93 –0.005 0.103
5.0–9.0 93 –0.020 0.130
Chloride 78.5 0.504 (0.64) 106.7 0.779 (0.73) 70–90 1 0.000 .— 1.0080 –1.1825 1.2528 0.9431
90–110 88 –0.205 1.627
110–140 10 –1.700 1.829
Urea Nitrogen 18.34 0.708 (3.86) 22.25 1.059 (4.76) 0–8.9 71 –0.055 0.411 1.0098 –0.1031 0.4023 0.9983
8.9–21.4 21 0.068 0.645
21.4–35.7 2 0.068 0.645
35.7–53.6 4 –0.357 1.237
Glucose 10.54 0.440 (4.17) 2.93 0.111 (3.77) 1.1–5.6 4 0.194 0.147 1.0274 0.2400 0.5223 0.9836
5.6–11.1 60 –0.029 0.392
11.1–16.7 31 0.208 0.947
16.7–25.0 4 –0.444 1.807
Hematocrit (–3.0) 0.000 (0.00) (–3.0) 0.000 (0.00) 15–30 11 0.818 1.328 0.9719 1.1637 2.0671 0.8754
30–50 86 0.081 2.879
STOT, total standard deviation.
a
Units as in Table 1.
b
ICU PCA value minus stat lab PCA value.
c
x = stat lab PCA value, y = ICU PCA value.

further proves that the system is fairly technique- Quality Control


independent and can be used by nonlaboratorians in
expectations, with a high degree of confidence, that Quality assurance deals with ensuring the reliability of
reliable results will be obtained. a system via various checks and balances. In our

CLINICAL CHEMISTRY, Vol.39, No. 6, 1993 1073


operations of the PCA system over 6 months and utiliz- Carolyn Lo, Nursing Education Specialist, and the nurses of the
Cardiothoracic ICU for their cooperation with this study.
ing > 1800 cartridges from various production lots, there
were no analytical outliers, neither in the precision data References
nor in the method-comparison data. All data obtained 1. Rock RC. Why testing is being moved to the site of patient care.
Med Lab Observ 1991,23:2 5.
were included in our analysis. However, 3.94% of the 2. Civetta JM, Hudson-Civeta JA. Maintaining quality of care
cartridges had one or more tests suppressed by the while reducing charges in the lCU. Ann Surg 1985,202:523-32.
analyzer. All the tests, or all except glucose, were 3. Jacobs P, Noseworthy TW. National estimates of intensive care
suppressed in 2.73% of the cartridges. Thus, in our utilization and costs: Canada and the United States. Crit Care Med
1990;18:1282–6.
experience, the software in the system that monitors 4. Hines R, Mylott L, Barash PG. The evolving needs of critical
and analyzes the rehydration and calibration responses care. Med Lab Observ 1991,23:6-9.
of the electrodes is apparently able to adequately detect 5. Zaloga GP, Hill TR, Strickland RA, Kennedy D, Visser M, Ford
K, et al. Bedside blood gas and electrolyte monitoring in critically
and suppress analytical malfunctions. If this observa- ill patients. Crit Care Med 1989,17:920-5.
tion holds true after data from multiple studies are 6. Hall JR. Critical care medicine and the acute care laboratory.
combined (n in the thousands), it will have a significant Clin Chem 1990;36(Suppl):1552B-6B.
impact on how quality control is performed on this 7. Hardwick DF, Morrison JI, Cassidy PA. Managing lab costs and
utilization: balance supply and demand. Dimens Health Serv
system. 1985;62: 18-22.
An effective QC program tests both the mechanical/ 8. Fleisher M, S ch wa rtz MK Strat egies of orga n i z ation and ser-
electronic component of the system, i.e., the analyzer, vice for the critical care laboratory. Clin Chem 1990;36(Suppl):
and the analytical component (the disposable test car- 1557B-61B.
9. Jacobs E, Sarkozi L, Colman N. A centralized critical care (stat)
tridge). With the PCA system, an electronic cartridge laboratory: the Mount Sinai experience. Crit Care Rep 1991;2:397–
s i mu l ator is used to ensure that electrical ch a ra c t e ri s- 405.
tics of the analyzer are within the manufacturer’s spec- 10. Wilding P. New approaches to point-of-care testing. Adv Hosp
Technol 1990;2:38-42.
ifications. This test is done on each analyzer every 24 h. 11. Claremont DJ. Biosensors: clinical requirements and scientific
Mechanical characteristics of the analyzer are checked promise. J Med Eng Technol 1987;11:51.
on each cartridge by monitoring fluid flow. However, as 12. Nani AA Poon R, Hinberg I. Quality of laboratory test results
a departure from traditional laboratory QC concepts, it obtained by non-technucal personnel in a decentralized setting. Am
J Clin Pathol 1988;89:797-801.
is not necessary to assay liquid QC material with each 13. Burritt MF. Current analytical approaches to measuring blood
analyzer each day. Instead, a randomized sampling, in analytes. Clin Chem 1990;36(Suppl):1562B-6B.
accordance with various manufacturing QC standards 14. Meyerhoff ME. New in vitro analytical approaches for clinical
chemistry measurements in critical care. Clin Chem 1990
(20, 21), of the test cartridges in stock should be ana- 36(Suppl): 1567B-72B.
lyzed each day to ensure that they have been stored 15. Wilding P, E ri ckson K Eva l u ation of a hand-held micro fab-
correctly and are still functioning properly. ricated sensor system, the i-STAT lOOTM, for rapid assay of
electrolytes, glucose, urea nitrogen and hematocrit [Abstract].
Clin Chem 1990;36:1206.
In conclusion, the i-STAT PCA is an analytical sys- 16. NCCLS Te n t at ive Guideline EP5-T. User eva l u ation of pre c i-
tem for the testing of whole blood, as well as serum or sion performance of clinical chemistry devices. Villanova, PA:
heparinized plasma, that has accuracy and precision National Committee for Clinical Laboratory Standards, June
equivalent to, if not better than, established laboratory- 1984.
17. NCCLS Proposed Guideline EP6-P. Evaluation of the linearity
based systems. It is simple to operate and gives reliable of quantitative analytical methods. Villanova, PA: National Com-
results when used by either laboratory personnel or ICU mittee for Clinical Laboratory Standards, September 1986.
nurses, i.e., nonlaboratory-trained individuals. Because 18. We i s b rot IM. Statistics for the clinical lab o rat o ry. Philadel-
phia: JB Lippincott Co., 1985:36-7.
of the high level of sophistication in the controlling 19. Corimer A, Vitiello J. Chloride ion concentration by ISE’s. In:
software, a cost-effective QC program can be based on Aizawa M, Kuwa K eds. Methodology and clinical applications of
s t a n d a rd industrial manu fa c t u ring sampling tech- blood gas, pH, electrolyte and sensor technology. Vol. 13, Proceed-
niques, rather than on the traditional laboratory QC ings of an international symposium. Copenhagen, Denmark: Int
Fed Clin Chem, 1992:85-93.
concept of testing each analyzer/cartridge combination 20. American national standard for sampling procedures and
on each day of use. tables for inspection by variablefi for percent nonconforming.
ANSVASQC Zl.9-1980. Milwaukee, Wl: American Society for Quality
Control, 1980.
We acknowledge and thank i-STAT Corp. for supporting this 21. American national standard for sampling procedures and
study through a grant. In particular, we thank Robert Martin, from tables for inspection by attributes. ANSVASQC Z1.4-1981. Milwaukee,
i-STAT, for his aid in the data reduction. We are grateful to W I :A m e rican Society for Quality Control, 1981.

1074 CLINICAL CHEMISTRY, Vol.39, No. 6, 1993

You might also like