Professional Documents
Culture Documents
Artwork and Signature File For: 9-500-0246 "MANUAL, AFFINITY OPERATOR"
Artwork and Signature File For: 9-500-0246 "MANUAL, AFFINITY OPERATOR"
Revision 006
Service Support
USA: 888-505-7910
Europe: +32.2.711.4690
All Other: 203-731-8320
Hologic and the Hologic Logo are registered trademarks of Hologic, Inc. Other trademarks registered or used by Hologic and its divisions and subsidiaries in the United States and
other countries include: Affinity, Affinity Platinum, Auto Film ID, “Clarity of Vision,” Dataport, HTC, LORAD, “LORAD A Hologic Company,” LORAD DSM, LORAD Elite, M-IV, M-IV
Platinum, MultiCare, Permagrid, ScoutMarc, SecurLook, SecurView, SecurView DX, SecurView RT, Selenia, SmartWindow, StereoLoc, and TechMate.
Table of Contents
Preface ................................................................................................................. xi
1.0 Overview ................................................................................................................................. xi
2.0 Intended Use ............................................................................................................................ xi
3.0 Product Complaints .................................................................................................................. xi
4.0 User Training ............................................................................................................................ xi
5.0 Quality Control ........................................................................................................................ xi
6.0 Terms and Definitions ............................................................................................................. xii
7.0 Warnings, Cautions and Notes ............................................................................................... xiii
8.0 International Symbols ............................................................................................................. xiii
9.0 Using the Operator’s Manual .................................................................................................. xiv
iv P/N 9-500-0246
Operator’s Manual
P/N 9-500-0246 v
Operator’s Manual
Index ................................................................................................................... 73
vi P/N 9-500-0246
Operator’s Manual
List of Figures
List of Figures
List of Tables
P/N 9-500-0246 ix
Operator’s Manual
List of Tables
x P/N 9-500-0246
Operator’s Manual
Preface
Overview
Preface
1.0 Overview
The Affinity is a dedicated mammography system that features automatic and manual
exposure modes, a compact Control Panel, and optional film labeling abilities.
P/N 9-500-0246 xi
Operator’s Manual
Preface
Terms and Definitions
WARNING! This alerts you to procedures that you must follow precisely to
avoid causing potentially serious or fatal injury to yourself or
others.
Warning! Warnings point out procedures that you must follow precisely to
avoid injury to yourself or others.
Caution: Cautions point out procedures that you must follow precisely to avoid
damage to equipment, loss of data, or corruption of files in software
applications
Chapter 1—Introduction
P/N 9-500-0246 1
Operator’s Manual
Chapter 1—Introduction
System Description
2 P/N 9-500-0246
Operator’s Manual
Chapter 1—Introduction
Safety
2.0 Safety
Read and understand this Operator’s manual completely before turning the system On.
Always follow all the instructions in this manual.
Hazardous Material Warning!
• The X-ray tube contains a beryllium window that should not be touched.
• Follow all local ordinances or regulations for disposal of the X-ray tube.
• The X-ray tube covers and the image receptor contain lead that should not be touched.
• Follow all local ordinances or regulations for disposal of these covers.
Note… The Affinity does not contain any mercury or latex in its construction.
P/N 9-500-0246 3
Operator’s Manual
Chapter 1—Introduction
Safety
Warning: For maximum protection from X-ray exposure, the operator must
keep their entire body behind the Operator Radiation Shield for the
duration of the exposure.
Warning: This X-ray system may be dangerous to patient and operator unless
you use safe exposure factors and follow operating instructions.
Observe all operating and safety instructions before making an X-
ray exposure.
WARNING! In the event of loss of power, always remove the patient from
the system before re-applying power.
Warning: Never leave the patient unattended during the exam, especially
when in contact with the mammography system.
Warning: Keep the patient’s hands away from any operator controls (for
example, all buttons and switches) at all times.
4 P/N 9-500-0246
Operator’s Manual
Chapter 1—Introduction
Safety
WARNING! Lethal voltages are present within the interior of this system.
Do not open any of the panels.
Caution: Use only recommended accessories with this equipment. Failure to heed
this warning may cause unexpected functions.
WARNING! To ensure the isolation integrity for the system, attach only
Hologic accessories or options to the system.
1.5m
P/N 9-500-0246 5
Operator’s Manual
Chapter 1—Introduction
Safety
Figure 1-4: Location of the Emergency Off Switches and the Circuit Breaker
6 P/N 9-500-0246
Operator’s Manual
Chapter 1—Introduction
Safety
2.6 Interlocks
In addition to the Emergency Off Switches described earlier, the Affinity has several other
safety interlocks.
P/N 9-500-0246 7
Operator’s Manual
Chapter 1—Introduction
Compliance
3.0 Compliance
This section describes the mammography system compliance requirements and the
manufacturer’s responsibilities.
8 P/N 9-500-0246
Operator’s Manual
Chapter 1—Introduction
Compliance
P/N 9-500-0246 9
Operator’s Manual
Chapter 1—Introduction
Compliance
10 P/N 9-500-0246
Operator’s Manual
Chapter 2—System Controls and Indicators
Affinity System Controls and Indicators
P/N 9-500-0246 11
Operator’s Manual
Chapter 2—System Controls and Indicators
The Control Panel Controls and Indicators
LCD Display Screen (Item 1): Displays system status, alerts, and system settings. The display
is described in detail in Chapter 3—The User Interface.
Function Keypad (Item 2): Contains buttons which scroll through on-screen fields, change
exposure settings, clear alerts, release compression, and initiate an X-ray exposure.
Table 2-1: The Function keypad Buttons and Indicators
7 Exposure Indicator The Exposure Indicator is on when the system generates X-rays.
12 P/N 9-500-0246
Operator’s Manual
Chapter 2—System Controls and Indicators
Gantry Controls and Indicators
4 Change Keys The left and right arrow Buttons (Change Keys):
• Increase or decrease the values presented.
• Toggle options on or off.
• Cycle through the available choices.
5 Reset Button The Reset Button:
• Alone, it acknowledges and clears an alert message.
• When used with the Up and Down arrows (Scroll keys) it
adjusts the display contrast.
• When used with the Change Keys, it gives access to the User
Default and Exposure Default modes.
3 Scroll Keys The up and down arrow buttons (Scroll Keys):
• Move the focus (highlight) through the fields.
• Adjust the display contrast, when used with the Reset Key.
P/N 9-500-0246 13
Operator’s Manual
Chapter 2—System Controls and Indicators
Gantry Controls and Indicators
14 P/N 9-500-0246
Operator’s Manual
Chapter 2—System Controls and Indicators
Compression Device Controls and Indicators
1 3 5 7 AUTO
7 AUTO
5
3 AEC
1
AEC
CM NEWTONS / LBS
Figure 2-5
Item Description Function
Item #
1 Compression Handwheels (2) Manually raises and lowers compression device.
Compression Thickness Indicator Indicates the compression thickness and
automatically compensates for the image receptor
2 installed.
AEC Sensor Position Indicator Displays the AEC sensor detent position in the
Image Receptor Support Device.
3
Compression Force Indicator This display shows the applied compression force.
4
P/N 9-500-0246 15
Operator’s Manual
Chapter 2—System Controls and Indicators
The C-arm Keypads
16 P/N 9-500-0246
Operator’s Manual
Chapter 2—System Controls and Indicators
Affinity Controls and Functional Tests
• Verify the Light Field Lamp illuminates, and the compression device moves
down while the button is held.
Compression Down Button
3
• Verify the compression thickness readout on the compression device
changes accordingly.
• Ensure the compression device stops when the button is released and the
compression brake is active (handwheel-up has resistance).
C-arm down Button • Verify the C-arm moves down while the button is held and it stops when the
4 button is released.
• Verify the C-arm can be rotated while the button is held, and can not be
C-arm Rotation Release
rotated when the button is released.
Button
5 • Verify that when the C-arm rotates, the rotation readouts on the Gantry
change accordingly.
• Verify that the C-arm rotation angle display shows zero degrees of rotation
when the C-arm is in the “CC” position.
C-arm Up Button • Verify the C-arm moves up while the button is held and it stops when the
6 button is released.
Compression Up Button • Verify the compression device moves up while the button is held and the
7 device stops when the button is released.
P/N 9-500-0246 17
Operator’s Manual
Chapter 2—System Controls and Indicators
Affinity Controls and Functional Tests
AEC Sensor Position The AEC Indicator on the Compression Device shows the position of the AEC
Sensor. Position 1 is closest to the chest wall.
• Move the AEC sensor handle and ensure the indicator changes.
Emergency Off Switches • Push the switch in. The system immediately shuts down.
• Rotate the switch one-quarter turn right. Toggle the circuit breaker off then
on. The system restarts.
18 P/N 9-500-0246
Operator’s Manual
Chapter 2—System Controls and Indicators
System Startup Procedures
Note… If during the internal tests, the system detects a fault condition, a message is
displayed, and startup is suspended until the problem is corrected.
P/N 9-500-0246 19
Operator’s Manual
Chapter 2—System Controls and Indicators
System Shutdown Procedures
20 P/N 9-500-0246
Operator’s Manual
Chapter 3—The User Interface
Introduction
1.0 Introduction
Three main screens comprise the Affinity user interface: the Run Mode screen, the User
Defaults screen, and the Exposure Defaults screen.
Run Mode, the main user screen, is The User Defaults screen changes The Exposure Defaults screen
displayed after startup. compression defaults and other system accesses exposure system defaults.
options.
Reminders… To access the User Defaults or Exposure Defaults screens, press and hold
the Reset Button on the keypad, then press and release one of the Change
Keys (Left/Right Arrows). Use this method to cycle forward and backward
through the three screens.
To select or change values, use the Scroll Keys (Up/Down Arrows) to
highlight the item, then use the Change Keys to make a selection.
P/N 9-500-0246 21
Operator’s Manual
Chapter 3—The User Interface
The Run Mode Screen
Note… Some fields are not available for every exposure. The fields enabled depend
on the exposure mode and the installed accessories.
22 P/N 9-500-0246
Operator’s Manual
Chapter 3—The User Interface
The Run Mode Screen
Note… All alert conditions must be corrected before the message can be cleared
using the Reset Key.
P/N 9-500-0246 23
Operator’s Manual
Chapter 3—The User Interface
The User Defaults Screen
24 P/N 9-500-0246
Operator’s Manual
Chapter 3—The User Interface
The Exposure Defaults Screen
3. Record the settings for this system on the worksheets in Table C-2, “Selected Exposure
Technique Defaults,” on p. 69 and Table C-3, “Selected Magnification Device Exposure
Technique Defaults,” on p. 70.
P/N 9-500-0246 25
Operator’s Manual
Chapter 3—The User Interface
The Exposure Defaults Screen
Note… For ease of explanation, inaccessible fields are shown in light gray in the
following screen illustrations. All fields are the same color on the actual
screens.
Item Options
Starting kV 25 or 26
Density -5 through +5
Screen-Film Film1/2/3
Auto-kV Window 200 mAs, Auto
Item Options
Starting kV 25 or 26
Density -5 through +5
Screen-Film Film1/2/3
Focus Large/Small
Auto-kV Window LG FS: 200/165/125/Auto
SM FS: 60/50/38/Auto
26 P/N 9-500-0246
Operator’s Manual
Chapter 3—The User Interface
The Exposure Defaults Screen
Item Options
kV 20-35 (Moly), 28-39 (RH)
Density -5 through +5
Screen-Film Film1/2/3
Focus Large/Small
Item Options
kV 20-35 (Moly), 28-39 (RH)
mAs 3-500 LG FS
3-150 SM FS
Screen-Film Film1/2/3
Focus Large/Small
P/N 9-500-0246 27
Operator’s Manual
Chapter 3—The User Interface
The Exposure Defaults Screen
Item Options
Starting kV 25 or 26
Density -5 through +5
Screen-Film Film1/2/3
Auto-kV Window 60/50/38/Auto
Item Options
kV 20-35 (Moly), 28-39 (RH)
Density -5 through +5
Screen-Film Film1/2/3
28 P/N 9-500-0246
Operator’s Manual
Chapter 3—The User Interface
The Exposure Defaults Screen
Item Options
kV 20-35 (Moly), 28-39 (RH)
mAs 3-150
Screen-Film Film1/2/3
EXPOSURE DEFAULTS
MODE : MANUAL
kV : 25 Table 3-11: Default Exposure Modes
mAs : 100
Density : ----- Item Options
Screen-Film : ----- Default Exposure Mode Manual
Focus : LARGE Auto-kV
Auto kV Win : ----- Auto-Time
Def Exp Mode : MANUAL Auto-Filter
Def MAG Mode : MANUAL
Default Mag Mode Mag Manual
Mag Auto-kV
Mag Auto-Time
SETUP
P/N 9-500-0246 29
Operator’s Manual
Chapter 3—The User Interface
The Exposure Defaults Screen
30 P/N 9-500-0246
Operator’s Manual
Chapter 4—Accessories
Introduction
Chapter 4—Accessories
1.0 Introduction
You can configure the system to perform screening and diagnostic applications, using exam-
specific accessories. This chapter describes the use of the accessories, accessory installation,
set up and removal. Note that the accessories installed depends on the option package
purchased with your system, and the type of clinical application. For a list of available
accessories for the Affinity, see Table C-4, “List of Available Accessories,” on p. 71.
Warning: For exposure other than magnification studies, always ensure that
the Face Shield is attached. Remember to check its condition prior
to use each day. The Face Shield does not offer protection from
exposure through the acrylic.
P/N 9-500-0246 31
Operator’s Manual
Chapter 4—Accessories
Bucky Devices
Note… Due to the non-standardization of DIN cassettes, some cassettes may move
slightly as the C-arm is rotated.
1
2
1
2
Note… The top surface of the Bucky has two indicator lights (green and amber) to
indicate the presence or absence of a cassette, and to indicate exposure status.
GREEN—Ready status: The cassette is installed and ready for exposure.
AMBER—Standby status: Change the cassette (post-exposure), or the
cassette is not installed.
32 P/N 9-500-0246
Operator’s Manual
Chapter 4—Accessories
Bucky Devices
P/N 9-500-0246 33
Operator’s Manual
Chapter 4—Accessories
Cassette Holder and Magnification Platform
The right set of markers should be placed so that the markers clear the lead blocker
intended for patient information. Once the placement is set, tighten the adjustment
knob to secure the marker in place.
4.1 Installing and Loading the Cassette Holder for Magnification Studies
1. Remove the Face Shield.
2. Raise the Compression Device above the slots in the C-arm. This provides the space
needed to safely install the Magnification Tray.
3. Remove the Bucky device.
4. Align the slide rails on the bottom of the 18 x 24 cm cassette holder with the edges of the
IRSD. Slide the cassette holder onto the IRSD until it stops.
5. Orient the loaded cassette so that it is directly over the cassette holder. Tilt the tube side
edge of the cassette down, then slip it under the front lip of the cassette holder.
6. Push down on the chest wall edge of the cassette until it snaps securely into the cassette
holder.
34 P/N 9-500-0246
Operator’s Manual
Chapter 4—Accessories
Cassette Holder and Magnification Platform
P/N 9-500-0246 35
Operator’s Manual
Chapter 4—Accessories
Magnification Table
Note… As soon as the exposure has been made (indicated by a change from Green
to Amber light), remove the cassette from the Bucky device and keep it
separate from unexposed film.
36 P/N 9-500-0246
Operator’s Manual
Chapter 4—Accessories
Magnification Table
P/N 9-500-0246 37
Operator’s Manual
Chapter 4—Accessories
Compression Paddles
38 P/N 9-500-0246
Operator’s Manual
Chapter 4—Accessories
Compression Paddles
P/N 9-500-0246 39
Operator’s Manual
Chapter 4—Accessories
Compression Paddles
Note… When a Localization Paddle is installed, the system turns off the automatic
Compression Release features, and disables the Compression Release
Buttons. Patient release is available through the Compression Up functions
(buttons or footswitch).
All Compression Paddles are mounted to the Compression Device in the same manner.
1. Align the mounting pins on the rear paddle frame with the mounting holes in the
Compression Device.
2. Push the Compression Paddle firmly into the Compression Device until it stops.
3. Gently pull the paddle out approximately 1/8" until you feel it click into place.
4. To remove the Compression Paddle, grasp the metal paddle frame with both hands, then
pull it straight out of the mounting holes.
Caution: Removing the compression paddle by grasping the plastic may cause
damage to the plastic.
40 P/N 9-500-0246
Operator’s Manual
Chapter 4—Accessories
Compression Paddles
P/N 9-500-0246 41
Operator’s Manual
Chapter 4—Accessories
Apertures
7.0 Apertures
Apertures are used to collimate the X-ray field to an appropriate size for the clinical
application being performed. The Affinity automatically detects the type of aperture
installed. If the system detects an aperture that is not compatible with the image receptor, the
system remains in a “standby” condition and prohibits exposures.
42 P/N 9-500-0246
Operator’s Manual
Chapter 4—Accessories
Apertures
Caution: Always use care when handling apertures to prevent damage to the
assembly. Always store apertures in a safe location when they are not in
use. A mishandled aperture (for example, dropped, crushed) must be
checked by a authorized service engineer to ensure that it complies with
regulations.
P/N 9-500-0246 43
Operator’s Manual
Chapter 4—Accessories
Crosshair Device (optional)
Note… Before performing the following adjustment, ensure the localization paddle
is adjusted properly with respect to the edge of the image receptor, see
Figure 4-9, p. 41.
44 P/N 9-500-0246
Operator’s Manual
Chapter 4—Accessories
Dual-Function Footswitch(es)
COMPR
M ES SION
C-AR
P/N 9-500-0246 45
Operator’s Manual
Chapter 4—Accessories
Dual-Function Footswitch(es)
46 P/N 9-500-0246
Operator’s Manual
Chapter 5—Clinical Procedures
Introduction
1.0 Introduction
The Affinity is capable of performing routine screening, diagnostic studies, magnification
studies, and localization procedures. Each of these procedures requires a different
combination of accessories, but all are performed in the same general sequence of steps.
This chapter provides a suggested sequence of operation to take patient films with the
Affinity. It should be noted that these steps are a recommended route of proceeding through
the examination process, and that they are not intended as substitutes for conventional
clinical training and methodology.
P/N 9-500-0246 47
Operator’s Manual
Chapter 5—Clinical Procedures
Exposure Sequence
4. Make final height adjustments to bring the breast platform to the same plane as the
patient’s breast.
5. Place the patient’s breast under compression.
• Whenever possible, use the footswitch controls to provide hands-free compression
control and C-arm height adjustment.
• Use the Light Field Lamp as necessary to visualize the X-ray field. Position the
patient’s breast within the illuminated field to ensure total coverage.
6. Place the patient’s arm or hand on the patient handle. Inform the patient of the locations
of all push button controls and give instructions to avoid contact with any of these
system controls.
7. During motorized compression, pay attention to smooth any skin folds. Use positioning
techniques to maximize chest wall imaging.
8. Apply compression slowly. As necessary, stop and make positioning adjustments. Use
the Handwheels to make final compression adjustments.
1. For automatic exposure mode, position the AEC sensor beneath the densest portion of
the breast.
2. Step behind the Operator Radiation Shield, and check that all exposure factors are set as
desired. If necessary, make any final technique changes.
3. Verify that the status area on the Run Mode screen reads: READY. If the system does not
enter a “ready” status within 20 seconds, verify that all the necessary accessories are
installed and are compatible with the type of exam being conducted. Refer to Appendix
C, Aperture Compatibility, p. 72 if necessary.
4. Press and hold the X-ray Button and footswitch (if installed) for the entire exposure
duration—until the tone stops.
5. If the compression release method is set to manual, raise the compression device, then
assist the patient away from the C-arm. If the automatic release feature is set, the
compression device raises after the exposure.
6. Verify that the Run Mode Status field indicates: STANDBY. Check that after 30 seconds
the system status returns to READY.
7. Evaluate the post-exposure mAs value (automatic modes) for subsequent technique
adjustments.
48 P/N 9-500-0246
Operator’s Manual
Chapter 5—Clinical Procedures
Magnification, and Localization Procedures
Note… The Crosshair device can be moved out of the beam path, when not in use,
by rotating it in either direction. It locks into place when the device is
centered in the beam path.
4. Rotate the Crosshair device out of the field before taking the scout image.
5. Take the scout image using AEC or appropriate exposure mode.
P/N 9-500-0246 49
Operator’s Manual
Chapter 5—Clinical Procedures
Labeling the Film (Option)
Note… When the exam is complete, perform the care, cleaning, and disinfection
procedures described in Chapter 6—Care and Maintenance, before using
the system on another patient.
50 P/N 9-500-0246
Operator’s Manual
Chapter 6—Care and Maintenance
Care and Cleaning
1.2 Recommendations
For general cleaning, use a lint-free cloth dampened with water, or a lukewarm, diluted
aqueous solution of household dishwashing liquid.
The following cleaning solutions have been found to be compatible with polycarbonate
compression paddles and accessories:
• 10% chlorine bleach & water solution
• 70% (by volume) isopropyl solution
• 90% maximum concentration of Hydrogen Peroxide solution
Note… After cleaning with the above solutions, the paddles must be washed with
soap and warm water.
P/N 9-500-0246 51
Operator’s Manual
Chapter 6—Care and Maintenance
Maintenance Schedule
52 P/N 9-500-0246
Operator’s Manual
Appendix A—Specifications
System Requirements and Dimensions
Appendix A—Specifications
P/N 9-500-0246 53
Operator’s Manual
Appendix A—Specifications
System Requirements and Dimensions
54 P/N 9-500-0246
Operator’s Manual
Appendix A—Specifications
C-arm, Compression, and IRSD
2.2 Compression
Manual Compression Force Limited to a maximum 300 N, +0/-88 N (67.4 lb, +0/-20
lb) between -150° through 0° to +150° of C-arm
rotation. It is not less than 150 N (33.7 lb) at all other
angles.
Motorized Compression Force User Selectable for either Full or Pre Compression
modes in User Defaults. 66 N to 178 N (15 lb to 40 lb),
user-selected within this range. Factory default of 132 N
(30 lb).
Full Compression User adjustable with a nominal range of 111 N
minimum to 178 N maximum (25 lbs minimum to 40
lbs maximum).
Pre Compression User adjustable with a nominal range of 66 N minimum
to 132 N maximum (15lbs minimum to 30 lbs
maximum).
Motorized Compression Motorized Up/Down controls on both sides of the C-
arm and on the dual function footswitch.
Manual Release Manual Up/Down handwheel controls on both sides of
the Compression Device.
P/N 9-500-0246 55
Operator’s Manual
Appendix A—Specifications
C-arm, Compression, and IRSD
56 P/N 9-500-0246
Operator’s Manual
Appendix A—Specifications
X-ray System
Mode kV mAs
Auto-Time 20 to 39 kV 5 to 500 mAs
Auto-kV 25 to 30 kV 5 to 500 mAs
Auto-Filter 25 to 32 kV 5 to 500 mAs
P/N 9-500-0246 57
Operator’s Manual
Appendix A—Specifications
X-ray System
58 P/N 9-500-0246
Operator’s Manual
Appendix A—Specifications
X-ray System
P/N 9-500-0246 59
Operator’s Manual
Appendix A—Specifications
X-ray System
60 P/N 9-500-0246
Operator’s Manual
Appendix A—Specifications
X-ray System
P/N 9-500-0246 61
Operator’s Manual
Appendix A—Specifications
X-ray System
X-ray Collimation The system detects the attached image receptor and full
field collimation is provided by an auto-aperture
mechanism with additional fixed apertures for
magnification and reduced field operation. An aperture
detection system ensures that the correct aperture is
used with the installed image receptor. Exposures are
inhibited when improper combinations are detected.
The collimator has a beam-limiting device that allows
the entire chest wall edge of the X-ray field to extend
5 mm, +2.5 mm, beyond the edge of the Image
Receptor and provides a means to ensure the X-ray field
does not extend beyond any chest wall edge of the
Image Receptor of more than 2% of the SID.
Light Field to X-ray Field Congruency The light field that passes through the X-ray beam
limitation device is aligned with the X-ray field so that
the total of any misalignment of the edges of the light
field and the X-ray field along either the length or the
width of the visually defined field at the plain of the
breast support surface does not exceed 2% of the SID.
Light Field Lamp The lamp illuminates for 30 seconds ±5 seconds upon
pressing a Light Field Lamp Button on the C-arm, or by
pressing a Compression Down Button. The lamp
extinguishes automatically when X-ray exposure is
initiated. A shatter shield is provided.
Light Field Illuminance The light localizer has a minimum illuminance of 160
lux and meets all of the requirements of 21CFR, section
1020.31. The lamp voltage is the least amount required
to assure 160 lux at low line. A shatter shield surrounds
the lamp. Lamp is service adjustable to provide
alignment of the light field to the X-ray field.
Light Field Mirror The light field mirror is motor driven out of the X-ray
field within 1.0 sec of X-ray initiation signal. Position
sensors provide for X-ray interlock. The lamp position is
adjustable between large focal spot and small focal spot
to provide proper light field to X-ray field congruency
required for a bi-angle target X-ray tube.
62 P/N 9-500-0246
Operator’s Manual
Appendix A—Specifications
Accessories
4.0 Accessories
4.1 Compression Paddles
Compression Paddles The compression paddles are transparent and the full
field paddles are marked with the location of the AEC
sensor positions. These markings are not detectable on
film when imaged with 1 cm of acrylic attenuator at 22
kV to an Optical Density of 1.2 OD. The paddle
attenuation is less than 20% (reduction of mR/mAs) at
28 kV. The full field paddles are adjustable to provide
focal spot, compression device and image receptor
alignment requirements.
The compression paddles including the frame and
compression device maintain a parallel compression
and do not deflect by more than 1 cm difference from
any surface providing compression under maximum
motorized compression force (except for FAST paddles).
Compression Paddle Sizes Available paddles:
18 x 24 cm Full Field Contact
24 x 30 cm Full Field Contact
7.5 cm Spot Contact/Magnification
10 cm Contact/Magnification
10 cm Open (Rectangular) Localization
10 cm Perforated Localization
15 cm Contact/Magnification
15 cm Open (Rectangular) Localization
15 cm Perforated Localization
18 x 24 Full Field FAST Paddles
24 x 30 cm Full Field FAST Paddles
Ultrasound Paddle
Small Breast Paddle
The chest wall edge of the compression paddle does not
extend beyond the chest wall edge of the image
receptor by more than 1% of the SID when tested with
the compression paddle placed above the breast
support surface at a distance equivalent to standard
breast thickness. The shadow of the vertical edge of the
compression paddle is not visible on the image.
P/N 9-500-0246 63
Operator’s Manual
Appendix A—Specifications
Accessories
4.2 Apertures
There are five apertures for all combinations of paddles and receptors.
Auto-Aperture
7.5 cm Spot Contact
7.5 cm Spot Magnification
10 cm Coned
Magnification Full Field
4.3 Magnification
Magnification Factor 1.8X (2.25 cm above the surface) for platform and table
devices
Material of breast support surface
Platform: Polycarbonate
Table: Carbon Fiber
Size of breast support surface 12 x 16.5 cm minimum to provide full 18 x 24 cm full
field magnification at 1.8X.
64 P/N 9-500-0246
Operator’s Manual
Appendix B—Alert Codes and Messages
Alert Codes and Messages
P/N 9-500-0246 65
Operator’s Manual
Appendix B—Alert Codes and Messages
Alert Codes and Messages
66 P/N 9-500-0246
Operator’s Manual
Appendix C—Forms and References
Introduction
1.0 Introduction
This section provides the user with reproducible pages that they can copy and have available
for daily use.
❑ 8. Reset the Emergency Off Switch and turn the Circuit Breaker Off then On
❑ Verify—Aperature is installed
P/N 9-500-0246 67
Operator’s Manual
Appendix C—Forms and References
System Default Record Sheets
User Defaults
Field Available Choices Comments
Compression Release Manual
Automatic
Full Compression 10% 20% 30% 40% 50% 25-40 lb of compression
60% 70% 80% 90% 100%
Pre Compression 10% 20% 30% 40% 50% 15-30 lb of compression
60% 70% 80% 90% 100%
Compression Mode Full
Pre
Film ID None Printer Auto-ID Rapid ID
Force Units Newtons lb
Cassette Sense On Off
Date mm/dd/yyyy mm/dd/yyyy US format
dd.mm.yyyy International format
dd.mm.yyyy
Time hh:mm AM 12 h Example: 01:30 PM (12 hour clock) is the same
as 1330 (24 hour clock)
hhmm 24 h
68 P/N 9-500-0246
Operator’s Manual
Appendix C—Forms and References
System Default Record Sheets
Exposure Defaults
Technique Manual Auto-Time Auto-kV Auto-Filter
kV
kV (Mo)
kV (Mo)
Starting kV (Mo)
Starting kV (Mo)
mAs
Large Focal Spot
Not Selectable Not Selectable Not Selectable
P/N 9-500-0246 69
Operator’s Manual
Appendix C—Forms and References
System Default Record Sheets
mAs
Small Focal Spot
Not Selectable Not Selectable
70 P/N 9-500-0246
Operator’s Manual
Appendix C—Forms and References
Available Accessories for the Affinity
P/N 9-500-0246 71
Operator’s Manual
Appendix C—Forms and References
Aperture Compatibility
Hologic
Other
Hologic 24 x 30
Hologic Hologic (receptors not
Hologic 18 x 24 Linear Linear
Magnification 18 x 24 manufactured by
IR Device and and
Device Cassette Hologic)
HTC Bucky HTC
Bucky
Focal Spot S L S L S L S L S L
No Paddle Auto Auto Magnification Auto Auto
Full Field
18 x 24 cm Paddle Auto Auto Magnification Magnification Auto
Full Field Full Field
24 x 30 cm Paddle Auto Auto
10 cm Contact/ 10 cm Magnification Magnification Magnification Magnification
Magnification Coned/ Full Field Full Field Full Field Full Field
Paddle Auto
10 cm Localization 10 cm Magnification Magnification Magnification Magnification
Paddle Coned/Auto Full Field Full Field Full Field Full Field
72 P/N 9-500-0246
Operator’s Manual
Index
Index
A C
AEC 48 Care and Cleaning 51
AEC sensor markings 39 C-arm 1
AEC Sensor Position 18 Down button 17
AEC Sensor Position Indicator 15 Keypads 16
Affinity Unit Description 1 Rotation Angle Display 14
Alert Codes and Messages 65 Rotation Release key 17
Aligning Up Key 17
Crosshair to Localization Paddle 44 C-arm Movement Interlocks 7
Angle of the C-arm 14 cassette
Annual Maintenance 52 eject lever 33, 37
Aperture Compatibility 43, 72 holder 34
Aperture Interlock 7 removing 33
Apertures 42, 71 Cassette Sense Interlock 7
available types 42 Change Keys 13
Installing and Removing 43 Checklist
Arrow Buttons 13 Power Up and Clinical Procedure 67
AutoFilm I.D. 32, 50 Circuit Breaker 19
Auto-Filter mode 26 Cleaning 51
exposures 47 Compliance 8
Auto-kV mode 26 Label Locations 9
exposures 47 Requirements 8
Automatic Compression Release Interlock 7 Statements 8
Auto-Time mode 27 Compression Device
exposures 47 Controls and Indicators 15
Available Accessories 71 Compression Down Key 17
Compression Force Indicator 15
B Compression Handwheels 15
Bucky Compression Paddle 38
release lever 32 AEC sensor markings 39
removing 33 Applications 39
Bucky devices 32 Installing and Removing 40
Button Compression Release
C-arm down Key 17 button 12
C-arm Rotation Release 17 function keypad button 17
C-arm Up Key 17 gantry button 17
Change 13, 21 Compression sequence 48
Compression Down 17 Compression Thickness Indicator 15
Compression Release 12, 17 Compression Up Key 17
Compression Up 17 connecting Dual-Function Footswitch 45
Footswitch Controls 18 Control Panel 1
Light Field Lamp 17 Display Screens 21
Reset 13, 21 Function Keypad 12
Scroll 13, 21 Crosshair Devices 44
X-Ray Exposure 13
9-500-0246 73
Operator’s Manual
Index
D Footswitch
Controls 14, 18
Daily Maintenance 52 detaching 45
Default Exposure Modes installing 45
Setting 29 Footswitch Controls 14
detaching the Footswitch 45 Function Keypad 12
diagram of Affinity 2
DIN-style cassettes 32 G
Disinfection 51
Gantry 1
Display
Gantry Controls and Indicators 13
C-Arm Angle 14
General cleaning 51
screen 12, 21
Dual-Function Footswitch(es) 45 I
E installing
accessories 47
Early Release Interlock 7
apertures 43
ejecting cassette 33
Bucky device 32
Electrical Safety 5
cassette holder 34
Emergency Off switch 6, 14, 18
compression paddles 40
Location 6
Face Shield 31
Exposure
footswitch 45
Auto-Filter mode 47
Magnification Platform 35
Auto-kV mode 47
Magnification Table 36
Auto-Time mode 47
Integrated Bucky Markers 33
Manual mode 47
Intended Use xi
Exposure Defaults 69
Interface
Mag Mode 70
Three main screens 21
screen 21, 25
Interlocks 7
Two sections 25
Automatic Compression Release 7
Screen, Overview 25
C-arm Movement 7
Exposure footswitch 14
Cassette Sense 7
Exposure Indicator 12
Early Release 7
Exposure Mode 26
Mirror and Filter 7
Auto-Filter 26
Internal start up tests 19
Auto-kV 26
Auto-Time 27 K
Mag Auto-kV 28
Mag Auto-Time 28 Keypads 16
Mag Manual 29
Manual 27 L
Exposure Sequence 48 Label Locations, Compliance 9
Exposure System Displays 23 Labeling the Film 50
Exposure Techniques 47 Light Field Lamp 17
F loading
Bucky device 32–33
Face Shield cassette holder 34
installing 31 locking clips 35
FAST Paddle Use 39
Film labeling options 50 M
Filter Indicator 17 Mag Auto-kV Mode 28
Filter Type LED Indicators 16
74 9-500-0246
Operator’s Manual
Index
Paddles 71 T
Positioning the Patient 47
Techniques
Post Startup Tests 19
Exposure 47
Pre-Startup Tests 19
Technologist and Patient 4
R Turning On the System 19
Radiation Safety 3 U
READY 23
Unit Description 1
removing
Unit Measurements 54
aperture 43
Up and Down buttons 13
Bucky 34
User Defaults
cassette 37
screen 21, 24
cassette and cassette holder 35
Settings 68
Compression paddles 40
User Interface 21
Face Shield 31
Using the Operator’s Manual xiv
Magnification Platform 35
Reset button 13, 21 X
Rotation Angle Display 11
Run Mode screen 21–22 X-RAY 23
Alert Codes 65 X-Ray Exposure Button 13
9-500-0246 75
Operator’s Manual
Index
76 9-500-0246