SOMATOM Sensation
CT

Replacements of Parts
syngo Aquisition Workplace

Instructions for replacement or installation of parts

This document is valid for:


SOMATOM Sensation 10/16/Cardiac
SOMATOM Sensation 40/ 64/ Cardiac 64/ Open
10351301
10351255
10130040
10017857
08874427
08872017
08377520
08098951
07711760
07711703
07543106
07543015
07394617
07393114
07130300

© Siemens AG 2003
The reproduction, transmission or use
of this document or its contents is not
permitted without express written
authority. Offenders will be liable for
damages. All rights, including rights
created by patent grant or registration
of a utility model or design, are
reserved.

Print No.: CT02-023.841.04.18.02 English


Replaces: CT02-023.841.04.17.02 Doc. Gen. Date: 10.07
2 Revision / Disclaimer
1Revision / Disclaimer

Document revision level


The document corresponds to the version/revision level effective at the time of system
delivery. Revisions to hardcopy documentation are not automatically distributed.
Please contact your local Siemens office to order current revision levels.

Disclaimer
The installation and service of equipment described herein is to be performed by qualified
personnel who are employed by Siemens or one of its affiliates or who are otherwise
authorized by Siemens or one of its affiliates to provide such services.
Assemblers and other persons who are not employed by or otherwise directly affiliated
with or authorized by Siemens or one of its affiliates are directed to contact one of the
local offices of Siemens or one of its affiliates before attempting installation or service pro-
cedures.

SOMATOM Sensation CT02-023.841.04.18.02 Page 2 of 76 Siemens AG


10.07 CS PS CT Medical Solutions
Table of Contents 3
0Table of Contents

1 _______ General ________________________________________________________ 7

Safety notes . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7
Notice for this document. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9
Overview. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10

2 _______ ICS___________________________________________________________ 11

Overview. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 11
ICS towers . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 11
Options of ICS. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 11
Field replaceable units of ICS . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 11
Prerequisites for replacements. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 12
Tools and auxiliary equipment . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 12
Documentation references . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 12
Tower 7 - Overview . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 13
Opening the casing . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 13
Closing the casing. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 14
Tower 8 - Overview . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 15
Opening the casing . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 15
Closing the casing. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 15
ICS tower . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 17
Preparation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 17
Installing . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 18
Final steps. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 21
CD/ DVD recorder (Tower 5, 6 and 7). . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 22
Prerequisites . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 22
Preliminary steps . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 22
Removing . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 23
Installing . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 24
Start up . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 25
Final steps. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 26
DVD recorder Tower 8 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 27
Prerequisites . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 27
Preliminary steps . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 27
Removing . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 28
Installing . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 28
Start up . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 29
Final steps. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 29
Battery on motherbord . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 31
Prerequisites . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 31
Preliminary steps . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 31
Removing . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 31
Installing . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 32
Start up . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 32
Final steps. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 32
Dongle . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 33
Preparation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 33

Siemens AG CT02-023.841.04.18.02 Page 3 of 76 SOMATOM Sensation


Medical Solutions 10.07 CS PS CT
4 Table of Contents

Installing . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 33
Final steps . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 33
Active cooling for console . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 34
Overview . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 34
Installing the cooling unit . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 34
Optional components . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 36
Ethernet switch . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 37

3 _______ Keyboard change _______________________________________________ 38

General . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 38
Keyboard change for Towers 1, 2, 3, 4 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 38
Keyboard change for Towers 5 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 38
Keyboard change for Towers 6 and 7 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 39
Keyboard change for Tower 8 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 39

4 _______ BIOS installation _______________________________________________ 40

General . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 40
Create the BIOS CD / CD, DVD Firmware update CD. . . . . . . . . . . . . . . . . . . . . . . . 40
Summary . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 42
Install BIOS . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 42
CD/DVD Firmware installation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 42
Correcting the drive letter of CD/DVD drive . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 43

5 _______ IRS 1 (7394617)_________________________________________________ 44

Overview . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 44
Computer equipment of IRS 1 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 44
Field replaceable units of IRS 1. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 44
Cabinet overview and labeling. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 44
Cabling . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 46
Prerequisites for replacements . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 48
Tools and auxiliary equipment . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 48
Opening covers . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 48
Acquisition, Image Reformatting and Recon Computer . . . . . . . . . . . . . . . . . . . . . . . . . 49
Preliminary steps . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 49
Removing . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 49
Installing . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 49
Final steps . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 50
100MBit-, 1GBit-Switch . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 52
News: 1GBit-Switch kit from January 2004 onwards. . . . . . . . . . . . . . . . . . . . . . . . . 52
Preliminary steps . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 52
Removing . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 53
Installing . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 53
Final steps . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 54
Console . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 55
Preliminary steps . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 55
Removing . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 55
Installing . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 55
Final steps . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 55

SOMATOM Sensation CT02-023.841.04.18.02 Page 4 of 76 Siemens AG


10.07 CS PS CT Medical Solutions
Table of Contents 5

KVM switch. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 56
Preliminary steps . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 56
Removing . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 56
Installing . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 56
Final steps. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 57

6 _______ IRS 2, IRS 2c and IRS 2F _________________________________________ 58

Overview. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 58
Computer equipment of IRS 2 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 58
Field replaceable units of IRS 2, IRS 2c and IRS2F . . . . . . . . . . . . . . . . . . . . . . . . . 58
Location of components and labeling . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 58
Prerequisites for replacements. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 59
Tools and auxiliary equipment . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 59
Opening covers . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 59
AQP -, TCR -, REC1/2 - Computer. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 60
Preparation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 60
Removing . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 60
Installing . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 60
Installing correct BIOS and CMOS- settings. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 60
Installing IRS Software, TCR tables and update GREC-FPGA version (if applicable)61
1GBit-Switch. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 62
Preparation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 62
Removing . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 62
Installing . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 62
Final steps. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 62
Service Console . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 63
Preparation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 63
Removing . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 63
Installing . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 63
Final steps. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 63
KVM switch. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 64
Preparation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 64
Removing . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 64
Installing . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 64
Final steps. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 64

7 _______ IRS 3 (83 77 785) _______________________________________________ 65

Overview. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 65
Computer equipment of IRS 3 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 65
Field replaceable units of IRS 3 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 65
Location of components and labeling . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 65
Prerequisites for replacements. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 66
Tools and auxiliary equipment . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 66
Opening covers . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 66
TCR -, REC1/2/3 - Computer . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 67
Preparation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 67
Removing . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 67
Installation. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 67

Siemens AG CT02-023.841.04.18.02 Page 5 of 76 SOMATOM Sensation


Medical Solutions 10.07 CS PS CT
6 Table of Contents

Installing correct BIOS and CMOS- settings (only for IRS REC 1, 2 and 3) . . . . . . . 67
Installing IRS Software (and TCR- tables if applicable). . . . . . . . . . . . . . . . . . . . . . . 68
1GBit-Switch . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 69
Preparation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 69
Removing . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 69
Installation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 69
Final steps . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 69
Service Console . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 70
Preparation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 70
Removing . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 70
Installation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 70
Final steps . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 70
KVM switch . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 71
Preparation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 71
Removing . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 71
Installation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 71
Final steps . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 71
Fan . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 72
Preparation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 72
Removing . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 72
Installation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 73
Final steps . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 73

8 _______ Changes to previous version _____________________________________ 74

9 _______ Index _________________________________________________________ 75

SOMATOM Sensation CT02-023.841.04.18.02 Page 6 of 76 Siemens AG


10.07 CS PS CT Medical Solutions
General 7
1-
Safety notes
1General

The following instructions contain important safety information as well as information


about tools to be used.

WARNING [ hz_serdoc_F13G01U12M02 ]
Avoid accident and injury or damage of parts.
Risk of accident and injury!
¹ Read and observe the “General safety notes” and the
“Product-specific safety notes”.

WARNING [ hz_serdoc_F13G01U05M08 ]
Dangerous voltages may be present. The ON/OFF switch does not
disconnect the ICS/IES from the line voltage.
Risk of electric shock!
¹ Remove the power plug from the grounded power outlet.

WARNING [ hz_serdoc_F12G02U07M02 ]
The UPS contains its own energy source (batteries). The parts
supplied with UPS power (ICS/IES/IRS) may still have power even
when the UPS is not connected to an AC supply (power off sta-
tus).
Risk of electrical shock from the UPS!
¹ Switch off UPS power by switching off the relevant circuit
brakers.
¹ Secure the system against unintended switch-on (e.g.
block breaker against switching on and/or mark breaker
against switch on).

CAUTION [ hz_serdoc_F13G08U01M01 ]
CE conformity of system will no longer be valid when using other
parts than itemized in the spare parts catalogue.
Risk of damage or data loss!
¹ Only components itemized as assemblies in the spare
parts catalogue will be used.

Siemens AG CT02-023.841.04.18.02 Page 7 of 76 SOMATOM Sensation


Medical Solutions 10.07 CS PS CT
8 General

WARNING [ hz_serdoc_F13G01U14M01 ]
Non-permissible or faulty manipulations or changes to the hard-
ware or software can cause system malfunctions. This can cause
injury and/or damage to the equipment.
Risk of accident and injury or damage!
¹ Do not manipulate or modify hardware.
¹ Use only original spare parts shown the spare parts cata-
logue.
¹ The use or installation of software not approved by Sie-
mens is strictly prohibited.

WARNING [ hz_serdoc_F13G01U12M01 ]
Failure to observe the guidelines specified in this document may
lead to serious injury or loss of life.
Risk of accident and/or injury!
¹ The guidelines contained in this document have to be
strictly followed.

CAUTION [ hz_serdoc_F13G01U02M01 ]
Only approved service tools should be used during replacement
and troubleshooting.
Risk of injury!
¹ Use only approved tools.

SOMATOM Sensation CT02-023.841.04.18.02 Page 8 of 76 Siemens AG


10.07 CS PS CT Medical Solutions
General 9

Notice for this document 1.1

NOTE Depending on the software and hardware version, the illustrations


and figures may differ slightly from your installed system. Please
refer to the information within the software, labels and layout of
the component.

Siemens AG CT02-023.841.04.18.02 Page 9 of 76 SOMATOM Sensation


Medical Solutions 10.07 CS PS CT
10 General

Overview 1.2

The Navigator consists of the components ICS and IRS.

1 3 4

6
5

Fig. 2: IRS cabinet


Fig. 1: ICS components
Pos. 1 LCD monitor
Pos. 2 Mouse
Pos. 3 Switch box
Pos. 4 Power distribution
Pos. 5 Keyboard
Pos. 6 Control box
Pos. 7 ICS computer

This document only describes the replacement of major parts of the Navigator. Compo-
nents such as keyboard, control box, mouse and connection cables are not covered.
Separate documentation is available for:
• Adjustment and Quality Assurance of the Monitors
• Installation of Options

SOMATOM Sensation CT02-023.841.04.18.02 Page 10 of 76 Siemens AG


10.07 CS PS CT Medical Solutions
ICS 11
2-
Overview
2ICS

ICS towers 0

Depending on the system and its delivery date, the following ICS towers can be found in
the field:
• 71 30 300 - ICS Tower 1
• 77 11 703 - ICS Tower 2
• 77 11 844 - ICS Tower 5A
• 88 79 418 - ICS Tower 6A
• 88 79 616 - ICS Tower 6S
• 88 79 566 - ICS Tower 7A
• 10 14 35 62 - ICS Tower 8A

Options of ICS 0

Depending on the system, the following options can be found in the field:
• Additional monitors
• MOD Dicom
• Desktop MOD

NOTE The field replaceable unit “ICS tower” does not contain any
optional component. When replacing the tower, the options from a
defective ICS have to be reinstalled to the new ICS tower.

Field replaceable units of ICS 0

The field replaceable units of ICS comprise the following components:


• ICS tower - as a spare part delivered in the basic configuration without keyboard,
mouse, dongle or options
• Dongle
• CD recorder
• DVD recorder
• Battery on motherbord
• Fuse (1A) on CIB D31PCI
• Ethernet switch
• Active cooling for console
• Optional component

Siemens AG CT02-023.841.04.18.02 Page 11 of 76 SOMATOM Sensation


Medical Solutions 10.07 CS PS CT
12 ICS

Prerequisites for replacements 2.1

Tools and auxiliary equipment 0

• Standard service kit and case


• Anti-static wristband and grounded tabletop pad

Documentation references 0

Depending on the component to be replaced, the following supplementary documentation


is required:
• Installation of Options
• Installation of Software
• System Installation

SOMATOM Sensation CT02-023.841.04.18.02 Page 12 of 76 Siemens AG


10.07 CS PS CT Medical Solutions
ICS 13

Tower 7 - Overview 2.2

Opening the casing 0

NOTE Make sure, that the computer is switched off and that the power
plug and all other connections are removed.

1. Unscrew the two knurled screws (1) on the back of the casing.

Fig. 3: Screws of side cover


2. Slide the side cover approximately 2 cm (a) in the direction of the arrow (1), until the
stop.
Unhook the upper edge of the side cover from the casing.
Pull the side cover in the direction of the arrow (2).

Fig. 4: Open cover

Siemens AG CT02-023.841.04.18.02 Page 13 of 76 SOMATOM Sensation


Medical Solutions 10.07 CS PS CT
14 ICS

Closing the casing 0

1. Hook the side cover into the lower guide trail in the casing (1). Make sure that the side
cover is offset by approximately 2 cm (a).
Engage the top edge of the side cover on the casing.
Push the side cover as far as it will go in the direction of the arrow (2).

Fig. 5: Close cover


2. Press the side cover onto the casing (1) at the rear and tighten the two knurled screws
(2) .

Fig. 6: Screws of side cover

NOTE After replacing the parts and reattaching covers, make sure to
reconnect all ground cables!

SOMATOM Sensation CT02-023.841.04.18.02 Page 14 of 76 Siemens AG


10.07 CS PS CT Medical Solutions
ICS 15

Tower 8 - Overview 2.3

Opening the casing 0

NOTE Make sure, that the computer is switched off and that the power
plug and all other connections are removed.

1. Unscrew the locking bolt for the side cover on the rear panel of the casing.

Fig. 7: Opening the casing


2. Pull the locking lever (1).
The side panel is now free and can be detached.

Closing the casing 0

Fig. 8: Closing the casing


3. Hook the side panel into the cutouts in the upper and lower guide rails.

Siemens AG CT02-023.841.04.18.02 Page 15 of 76 SOMATOM Sensation


Medical Solutions 10.07 CS PS CT
16 ICS

4. Close the locking lever in the direction of the arrow (1).


5. Tighten the locking bolt for the side cover on the rear panel of the casing.

NOTE After replacing the parts and reattaching covers, make sure to
reconnect all ground cables!

SOMATOM Sensation CT02-023.841.04.18.02 Page 16 of 76 Siemens AG


10.07 CS PS CT Medical Solutions
ICS 17

ICS tower 2.4

Preparation 0

• Supplementary documentation:
As there are differences for each individual tower (e.g. for opening the covers), please
refer to the detailed instructions and figures listed in “Installation of Options”.
In order to generate the backup and to install the software, refer to “Installation of Soft-
ware” (if requested).
Detailed cabling information (interconnections) can be found in “System Installation”.
• If possible, clarify whether patient data have to be saved.
• If possible, create an up-to-date software backup.
• If possible note data, e.g. host name, static SRS routes, auto reports, service key,
regional settings.
• Shut down the software and switch off the tower.
• Disconnect all cable connections at the rear.
• Remove the tower from its position and find an adequate working place for the follow-
ing steps:
• Remove the dongle.
• If installed, remove the optional components (refer to “Installation of Options”).

Deinstallation of the main memory kit.


Only if Tower 6 or Tower 7 with software verison VB30 will be replaced:

NOTE With the rollout of software version VB30 an upgrade kit for tower
6 and 7 is available. This upgrade kit include a memory extension
for the main memory from 2GB to 3GB. The following working
steps are necessary before you install the new tower.

For Tower 6: To deinstall the main memory extension kit perform the following working
steps.
• Remove the side cover of the tower according to (Opening the casing / p. 13)

Siemens AG CT02-023.841.04.18.02 Page 17 of 76 SOMATOM Sensation


Medical Solutions 10.07 CS PS CT
18 ICS

• Remove the RAM memory (2 x 1GB modules) with the labeling 1GB. Both modules are
needed for the new tower.

Fig. 9: Main memory extension - 1GB for tower 6


Pos. 1 Labeling - 1GB

• Take the old memory modules at the bottom inside of the computer housing and install
these two modules where the memory extension kit was installed.
• Close the side cover according to (Closing the casing / p. 14)

For Tower 7: To deinstall the main memory extension kit perform the following working
steps.
• Remove the side cover of the tower according to (Opening the casing / p. 13)
• Remove the RAM memory (2 x 512MB modules) with labeling 512MB. Both modules
are needed for the new tower.

Fig. 10: Main memory extension - 512MB for tower 7


Pos. 1 Labeling - 512MB

• Close the side cover according to (Closing the casing / p. 14).

Installing 0

• Reinstall the optional components for the new tower.

SOMATOM Sensation CT02-023.841.04.18.02 Page 18 of 76 Siemens AG


10.07 CS PS CT Medical Solutions
ICS 19

For Tower 6: To install the main memory extension kit perform the following working steps.
• Only if all memory slots are equipped, remove two of the installed memory modules
(2x512 MB) starting from the left see (1 and 2/Fig. 11 / p. 19).

Fig. 11: Memory occupancy - Tower 6


Pos. 1 DIMM 1 memory slot
Pos. 2 DIMM 2 memory slot

• Plug in the two previously deinstalled 1GB memory modules next to the modules
already installed.
• Put the removed modules into anti-static packing and sticky-tape the bag to the bottom
inside of the computer housing. The modules have to be returned with the tower, if it is
exchanged in future.

Siemens AG CT02-023.841.04.18.02 Page 19 of 76 SOMATOM Sensation


Medical Solutions 10.07 CS PS CT
20 ICS

For Tower 7: To install the main memory extension kit perform the following working steps.
• Plug in the two deinstalled 512MB memory modules next to the modules already
installed (1 and 2/Fig. 12 / p. 20).

Fig. 12: Memory occupancy - Tower 7


Pos. 1 DIMM 5 memory slot
Pos. 2 DIMM 6 memory slot

• Put the removed modules into anti-static packing and sticky-tape the bag to the bottom
inside of the computer housing. The modules have to be returned with the tower, if it is
exchanged in future.

• Insert the dongle at the printer port.


• Position the new tower at the previous location.
• Connect the cable connectors in the back, ensuring that they match their appropriate
location.

NOTE After replacing the parts and reattaching all covers, make sure to
reconnect all ground cables!

• Power on.
• ICS tower 1 only: Perform the BIOS setup according to “Installation of Options”. Set
“Clear Chassis Intrusion” from [No] to [Yes].

SOMATOM Sensation CT02-023.841.04.18.02 Page 20 of 76 Siemens AG


10.07 CS PS CT Medical Solutions
ICS 21

NOTICE As a safety mechanism, the ICS tower 1 senses (and stores the
information) whether the covers have been opened as well as the
time when the computer was switched off.
¹ If you attempt to perform the procedure without the cov-
ers installed, the following message appears “Warn-
ing!! Your computer chassis has been opened”
and the bootup sequence is stoppe

• ICS tower 6 with VB30 only: Perform a BIOS installation according to chapter (BIOS
installation / p. 40)
• Reboot the system.
• Install the software according to the current “Installation of Software”.
• Restore the backup (this will also restore the available licenses)

Final steps 0

• Check that all application and customer options are available.


• Ensure that the system is operating properly.

Siemens AG CT02-023.841.04.18.02 Page 21 of 76 SOMATOM Sensation


Medical Solutions 10.07 CS PS CT
22 ICS

CD/ DVD recorder (Tower 5, 6 and 7) 2.5

NOTE The replacement of the CD recorder applies to Tower 5 and 6,


whereas the replacement of the DVD recorder to Tower 7.

Prerequisites 0

Time and Manpower


• 1 person
• Estimated repair time 1hour

Tools and auxilliary equipment


• Standard service kit

Preliminary steps 0

1. Shut down the software and switch off the tower.


2. Disconnect all cable connections at the rear.
3. Remove the tower from its position and find an adequate working place for the follow-
ing steps:
4. Remove the left side cover of the tower according to (Opening the casing / p. 13)
5. Remove the front panel

SOMATOM Sensation CT02-023.841.04.18.02 Page 22 of 76 Siemens AG


10.07 CS PS CT Medical Solutions
ICS 23

NOTE The cable length of the LCD panel is long enough so that you can
carefully place it to one side with the front panel. you do not need
to unplug the cable before removing the front panel.

Fig. 13: Removing front panel


- Detach the three locking tabs on the left side of the front panel (1).
- Open the left side of the front panel approximately 2 - 3 cm (a).
- Pull the front panel off in the direction of the arrow (2).

Removing 0

1. Unplug the cable connections on the CD/ DVD recorder.


2. Press the two levers on both sides of the drive and slide the CD/ DVD recorder in for-
ward direction out of the tower.

Fig. 14: Removing drive

Siemens AG CT02-023.841.04.18.02 Page 23 of 76 SOMATOM Sensation


Medical Solutions 10.07 CS PS CT
24 ICS

3. Loosen the screws (1).


4. Remove the rails from the drive (2).

Fig. 15: Removing rail

Installing 0

1. Check the jumpers of the new CD/ DVD drive. It must be jumpered as “Master”. Cor-
rect, if required.
2. Lay the rails onto the new CD/ DVD recorder such that the holes in them align with the
screw holes (1).
3. Fasten the rails with the screws (2).

Fig. 16: Attaching rails

SOMATOM Sensation CT02-023.841.04.18.02 Page 24 of 76 Siemens AG


10.07 CS PS CT Medical Solutions
ICS 25

4. Slide the new CD/ DVD recorder into the casing.

Fig. 17: Inserting the drive


5. Connect all cables on the CD/ DVD recorder

Start up 0

1. Attach the front panel

Fig. 18: Attaching front panel


- Engage the three hinges on the right side of the front panel (1).
- Close the front panel on the casing so that the three locking tabs on the left side of the
front panel engage (2).

Siemens AG CT02-023.841.04.18.02 Page 25 of 76 SOMATOM Sensation


Medical Solutions 10.07 CS PS CT
26 ICS

NOTE Make sure that the LCD cable is not pinched.

2. Close the right side cover according to (Closing the casing / p. 14)

Final steps 0

1. Power on.
2. Reboot the system.
3. Perform the firmware update according to (CD/DVD Firmware installation / p. 42) to
ensure that the DVD recorder has the current firmware version.
4. Execute Tune CD Burning:
- Go to Local Service -> Utilities -> Tune CD Burning.
- Select Tune CD Burning for current hardware and click Go.
¹ A message is displayed: “Tune CD Burning for current hardware was selected.
This may take some minutes.”
- Click Yes.
¹ After Tune CD Burning is finished the message “Driver has been successfully
installed” is displayed.
- Select Home and reboot the system.
5. Test the DVD recorder by storing and reading data within the application software.

SOMATOM Sensation CT02-023.841.04.18.02 Page 26 of 76 Siemens AG


10.07 CS PS CT Medical Solutions
ICS 27

DVD recorder Tower 8 2.6

Prerequisites 0

Time and Manpower


• 1 person
• Estimated repair time 1hour

Tools and auxilliary equipment


• Standard service kit

Preliminary steps 0

1. Shut down the software and switch off the tower.


2. Disconnect all cable connections at the rear.
3. Remove the tower from its position and find an adequate working place for the follow-
ing steps:
4. Remove the left side cover of the tower according to (Opening the casing / p. 13)
5. Open the front panel

NOTE The cable length of the LCD panel is long enough so that you can
carefully place it to one side with the front panel. you do not need
to unplug the cable before removing the front panel.

Fig. 19: Opening the front panel

Siemens AG CT02-023.841.04.18.02 Page 27 of 76 SOMATOM Sensation


Medical Solutions 10.07 CS PS CT
28 ICS

- Detach the three lockking tabs on the left side of the front panel (1).
- Fold open the front in the direction of the arrow (2).
- If necessary, detach the hinge on the right-hand side of the front panel from the cas-
ing and carefully remove the front panel.

Removing 0

1. Unplug the cable connections on the DVD recorder.

Fig. 20: Removing the drive


2. Loosen the screws (1).
3. Slide the drive out of the bay in the direction of the arrow (2) from behind.
4. Slide the drive out of the casing (3).

Installing 0

1. Check the jumpers of the new DVD. It must be jumpered as “Master”. Correct, if
required.

Fig. 21: Installing the new drive

SOMATOM Sensation CT02-023.841.04.18.02 Page 28 of 76 Siemens AG


10.07 CS PS CT Medical Solutions
ICS 29

1. Slide the new DVD recorder into the casing until it locks in position (1).
2. Connect all cables to the DVD recorder
3. Fasten the drive with the screws (2).

Start up 0

1. Attach the front panel

Fig. 22: Closing the front panel


- If necessary, hook in the hinge on the right-hand side of the front panel.
- Close the front panel in the direction of the arrow (1) so that the three locking tabs on
the left-hand side of the front panel engage (2).
2. Close the right side cover according to (Closing the casing / p. 14)

Final steps 0

1. Power on.
2. Reboot the system.
3. Perform the firmware update according to (CD/DVD Firmware installation / p. 42), to
make sure that the DVD recorder has the current firmware version.
4. Execute Tune CD Burning:
- Go to Local Service -> Utilities -> Tune CD Burning.
- Select Tune CD Burning for current hardware and click Go.
¹ A message pops up “Tune CD Burning for current hardware was selected. This
may take some minutes.”
- Click Yes.
¹ After Tune CD Burning is finished the message “Driver has been successfully
installed” appears.

Siemens AG CT02-023.841.04.18.02 Page 29 of 76 SOMATOM Sensation


Medical Solutions 10.07 CS PS CT
30 ICS

- Select Home and reboot the system.


5. Test the DVD recorder by storing and reading data within the application software.

SOMATOM Sensation CT02-023.841.04.18.02 Page 30 of 76 Siemens AG


10.07 CS PS CT Medical Solutions
ICS 31

Battery on motherbord 2.7

NOTE This replacement of the battery is valid for ICS Tower 7 upwards.

Prerequisites 0

Time and Manpower


• 1 person
• Estimated repair time 1hour

Tools and auxilliary equipment


• Standard service kit
• Anti-static wristband and grounded tabletop pad
• BIOS Setup CD

Preliminary steps 0

1. Shut down the software and switch off the tower.


2. Disconnect all cable connections at the rear.
3. Remove the tower from its position and find an adequate working place for the follow-
ing steps:
4. Remove the left side cover of the tower according (Opening the casing / p. 13).

Removing 0

NOTE The lithium battery holder exists in different designs that function
in the same way.

Siemens AG CT02-023.841.04.18.02 Page 31 of 76 SOMATOM Sensation


Medical Solutions 10.07 CS PS CT
32 ICS

Fig. 23: Replacing battery

1. Press the catch in the direction of the arrow (1).


¹ The battery jumps out of the holder.
2. Remove the battery (2).

Installing 0

• Push the new lithium battery into the holder (3) and press it downward until it engages
(4).

Start up 0

1. Power on.
2. Perform the BIOS installation according to (BIOS installation / p. 40)

Final steps 0

• Check the functionality of the system.


• Please make sure that the system date and the system time are correct.

SOMATOM Sensation CT02-023.841.04.18.02 Page 32 of 76 Siemens AG


10.07 CS PS CT Medical Solutions
ICS 33

Dongle 2.8

Preparation 0

• When replacing the dongle, the corresponding new licenses have to be created and
implemented as well.

NOTICE Escalation procedure and ordering process:


¹ A dongle defect should always be escalated to the Licens-
ing Center of CS HSC 1 in Erlangen, tel. +49 9191 18 8080.

Installing 0

• Shut down the software and power off the tower

NOTICE Risk of damage


Removing the dongle with switched on power results in damage
¹ Make sure that the power is switched off when changing
electronic parts.

• Move the tower to access it from the back.


• Take off the two fastening screws and remove the dongle from the printer port.
• Insert the new dongle into the printer port and attach it with the screws.
• Position the tower at its previous location and ensure that all cables are connected
properly.
• Power on.
• Install the new licenses according to “Installation of Options”
• The replaced dongle has to be returned.

Final steps 0

• Verify that all licenses and customer options are available.


• The replaced dongle has to be returned.

Siemens AG CT02-023.841.04.18.02 Page 33 of 76 SOMATOM Sensation


Medical Solutions 10.07 CS PS CT
34 ICS

Active cooling for console 2.9

Overview 0

For Tower 7 onward in combination with the console furniture an active cooling is neces-
sary because of the heat development.

Installing the cooling unit 0

Preparation
• Shut down and switch off the computers.
• Remove the metal covers.
• Deinstall the spacers. Take care not to drop the nut and washer when doing this.

Fig. 24: Console without metal covers


Pos. 1 Console
Pos. 2 Spacer

Installing the cooling unit


• Install the cooling unit using the delivered mounting material and the spacers.
• Connect the power cable to the multiple socket strip.

SOMATOM Sensation CT02-023.841.04.18.02 Page 34 of 76 Siemens AG


10.07 CS PS CT Medical Solutions
ICS 35

• Connect the PE wire to an unused ground bolt at the multiple socket strip´s cover.

Fig. 25: Console with mounted cooling unit


Pos. 1 Console
Pos. 2 Spacer
Pos. 3 Cooling unit

Checking the cooling unit


• Test the cooling unit.
The fans have to suck the warm air out of the container.

Concluding steps
• Install the metal covers.

Siemens AG CT02-023.841.04.18.02 Page 35 of 76 SOMATOM Sensation


Medical Solutions 10.07 CS PS CT
36 ICS

Optional components 2.10

• For information about replacing optional components, refer to “Installation of Options”.

SOMATOM Sensation CT02-023.841.04.18.02 Page 36 of 76 Siemens AG


10.07 CS PS CT Medical Solutions
ICS 37

Ethernet switch 2.11

• For information about replacing the Ethernet switch, refer to “System Installation”
and/or to “Installation of Options”.

Siemens AG CT02-023.841.04.18.02 Page 37 of 76 SOMATOM Sensation


Medical Solutions 10.07 CS PS CT
38 Keyboard change
3-
General
3Keyboard change

There are four types of spareparts:


PS 2-keyboards:
• Keyboards for all ICS and IES Tower 1,2,3,4 (e.g. ICS 4a, 4b, ..., IES4)
• Keyboard kits for all ICS and IES Tower 5 (e.g., ICS 5a, 5b,..., IES5)
• Keyboards for all ICS and IES Tower 6 and Tower 7.
• Keyboards for all ICS and IES Tower 8 and higher
USB-keyboards:
• Keyboards for all ICS and IES Tower 8 and higher

NOTE If a keyboard extension is installed, make sure to keep this. The


sparepart is delivered without extension.

Keyboard change for Towers 1, 2, 3, 4 0

Replacing the Keyboard


• Switch off the system
• Disconnect the old keyboard
• Connect the new keyboard.

Keyboard change for Towers 5 0

Replacing the Keyboard


• Switch off the system

NOTE The sparepart keyboard requires BIOS version 1.07 or higher.


The BIOS CD is not delivered as a spare part. A BIOS CD has to be
created to perform the update, see (Create the BIOS CD / CD, DVD
Firmware update CD / p. 40) to create the BIOS CD and (Install
BIOS / p. 42) to install the BIOS.
If BIOS 1.07 is already installed and the BIOS installation is
repeated, there is no problem.

• Short summary of BIOS installation process: Insert CD- ROM, power up the PC, wait
some minutes.
For a detailed description, see section “Installation of BIOS”
• Disconnect the old keyboard
• Connect the new keyboard.

SOMATOM Sensation CT02-023.841.04.18.02 Page 38 of 76 Siemens AG


10.07 CS PS CT Medical Solutions
Keyboard change 39

Keyboard change for Towers 6 and 7 0

Replacing the Keyboard


• Switch off the system
• Disconnect the old keyboard
• Connect the new keyboard.

Keyboard change for Tower 8 0

Replacing the Keyboard


• Disconnect the old USB-keyboard.
• Connect the new USB-keyboard.

Siemens AG CT02-023.841.04.18.02 Page 39 of 76 SOMATOM Sensation


Medical Solutions 10.07 CS PS CT
40 BIOS installation
4-
General
4BIOS installation

The BIOS installation applies to the following Tower types:


• all ICS and IES Tower 5 (e.g. ICS 5a, 5b, ..., IES5)

NOTE For Tower 5, the keyboard kit 08873841(German), 08873858


(English), 08873866 (French) and 08873874 (Spanish) require the
BIOS version 1.07 or higher.
Earlier keyboard kits still contain a BIOS CD, which should be dis-
posed of to ensure that the newest BIOS version is installed.

• all ICS and IES Tower 6


• all ICS and IES Tower 7
• all ICS and IES Tower 8

Create the BIOS CD / CD, DVD Firmware update CD 0

Download the necessary BIOS file from the intranet. Navigate to “http://cs.med.sie-
mens.de/cms/en/home/default.asp“ and follow path “For Service > Product Information >
CT Systems > Software download“.
(Software download, https://cs.med.siemens.de/cms/productinfo/default.asp?start-
folder=8bd95977e70107fde60d2a5aeebc0049)
Download the BIOS / CD, DVD Firmware file:
• In the “CT-Software Update Page“, left-click the file to be downloaded.
• Save the file to the hard drive c:\temp. When finished, close the window.

Extract the zipped file:


• Go to c:\temp in the Windows Explorer.
• Right-click the BIOS file you just saved and select "WinZip > Extract to..."
• Select c:\temp in the window that appears, and click Extract.
¹ The file will be extracted to c:\temp

Label a blank CD-R/FD with the file name of the downloaded file:
• For ICS and IES Tower 5, e.g., “10141460 VA01A Tower5 CS”
• For ICS and IES Tower 6, e.g., “10142438 VA01A Tower6 CS”
• For ICS and IES Tower 7, e.g., “10142439 VA01B Tower7 CS“
• For ICS and IES Tower 8, e.g., “10143567 VA01D Tower8 CS“

SOMATOM Sensation CT02-023.841.04.18.02 Page 40 of 76 Siemens AG


10.07 CS PS CT Medical Solutions
BIOS installation 41

• CD/DVD Firmware, e.g., “10304223 VA01B PlextorUpdate“


Burn the BIOS / CD, DVD Firmware CD:
Write the files to CD using a standard burning software, e.g. Nero, WinOn CD (do not use
the XP burn SW, it can create problems when installing the update).
For the CD, it is recommended that you have to burn the Iso.file as an Image CD. As an
example, here are the work steps for creating the CD using the Nero Burn Software:

• Insert the labeled CD into the CD drive.


• Open Nero Burning Rom.
• Close the "New compilation" window that pops up
• Select Records -> Burn Image... (in some cases select: File-> Burn Image).
¹ an Explorer Window is displayed.
• Select the file that was extracted previously under:
c:\temp\
• Click Open
¹ the "Burn Compilation" window is displayed.
• on the tab card "Burn", under "Action", select the following two actions:
- write
- finalize CD
• Click Burn (in some cases Write, depending on the version).
¹ The Burn process starts.
• When finished an information window is displayed: "Burning process completed suc-
cessfully".
• Click OK (in some cases Done or Discard, depending on the version).
¹ The CD will be ejected automatically.
• Remove the Update CD-R from the CD drive.

NOTE Once the files are extracted to the CD, they must not be copied to
another CD for distribution (e.g. no CD-to-CD copies, as it cannot
be verified that the SW is virus-free).

Virus Check
After the CD is burned successfully, perform a virus scan of the CD after having updated
the virus scanning software with the latest versions of the virus patterns.
After you have performed the virus scan without any problems, the CD is ready and the
Bios update can be performed on the customer system.

Siemens AG CT02-023.841.04.18.02 Page 41 of 76 SOMATOM Sensation


Medical Solutions 10.07 CS PS CT
42 BIOS installation

Summary 0

The following steps are automatically performed:


1. System BIOS flash
2. Set System BIOS settings
3. Set SCSI BIOS settings

Install BIOS 0

Worksteps:
• Insert the BIOS setup- CD, power the computer off, the power it up to reboot from CD.
Make sure to use the correct CD, see above for partnumbers.
• The computer boots from the setup-CD.

NOTE If the computer does not boot from CD, check the boot sequence in
the BIOS.

NOTE If the CD inserted is not designed for the computer type (e.g., a
Tower 5 CD is inserted into a Tower 6 model), the message “CD
not designed for this computer” appears.

• First, the screen shows the booting process of Windows 98, next the progress of the
hardware detection is displayed and then the BIOS is flashed.
If the system BIOS is already the current version, the installer will directly go to loading
the BIOS settings.
¹ The message "BIOS xxx flashed” indicates the success (xxx = version of the
BIOS, e.g.,”107” for “BIOS version 1.07”)
• The computer reboots from CDROM.
Next, the System BIOS Settings are loaded.
¹ The message “Operation completed successfully” indicates the
success.
• Then, the SCSI BIOS Settings are set.
¹ The message “Update of ***5* successful” indicates the success.
¹ The BIOS update is complete.

CD/DVD Firmware installation 0

Worksteps:
• Insert the Firmware update CD, power the computer off, then power it up to reboot from
CD. To create the CD refer to (Create the BIOS CD / CD, DVD Firmware update
CD / p. 40).
Make sure to use the correct CD, see above for part numbers.

SOMATOM Sensation CT02-023.841.04.18.02 Page 42 of 76 Siemens AG


10.07 CS PS CT Medical Solutions
BIOS installation 43

NOTE If the CD contains a newer version, it displays “New Firmware


Version for Plextor <type> available”.

• A menu appears:
1 Leave Version<old#>
2 Update to version <new#> -Please check drive letter (see
"replacement of parts")
Select “1 ↵ ” to keep the prior firmware on the drive
or select “2 ↵ ” to update.

NOTE If the CD/DVD drive FW is updated, the drive letter has to be cor-
rected after the reboot.
“Windows plug and play” tries to assign the next free drive letter
instead of the letter R
See the following instructions.

Correcting the drive letter of CD/DVD drive 0

Perform only if drive FW was updated in addition to the tower BIOS


• Select System -> End -> Restart Application
Click yes and press the left Shift key until the login window appears
Log in as administrator.
• Select Start -> Settings -> Control Panel
• Double-click Administrative Tools
• Double-click Computer Management
• Double-click Disk Management
• Right-click CD-ROM to select "change drive letter"
• Click Change
• Change drive letter to R
• Click Yes
• Click the Start button on the task bar and perform Shutdown -> Restart.
• After restart, check that the CD/DVD drive letter is R.

Siemens AG CT02-023.841.04.18.02 Page 43 of 76 SOMATOM Sensation


Medical Solutions 10.07 CS PS CT
44 IRS 1 (7394617)
5-
Overview
5IRS 1 (7394617)

Computer equipment of IRS 1 0

The IRS 1 (73 94 617) cabinet comprises the following single computers
• 73 95 259 - AQP - Acquisition computer -CPU 1
• 73 95 275 - TCR - Image reformatting computer - CPU 2
• 73 95 267 - REC1,2 - Recon computer 1 and 2 - CPU 3 and CPU 4

NOTE The field replacable units of the IRS1 can be used with Windows
NT based SW- Versions (e.g., VA70) as well as with Windows XP
based SW- Versions (e.g., VB10) without modifying the BIOS of
the indivitual Computers (That step is exclusive for IRS2/IRS2c).

Field replaceable units of IRS 1 0

The field replaceable units of the IRS 1are as follows:


• Acquisition computer
• Image reformatting computer
• Recon computers 1 & 2
• 1 Gbit switches
• 100 Mbit switch
• Service console
• KVM switch
• Current limiting & contactors
• Data cables
• Airflow filters
• Covers
For material numbers, refer to the Spare Parts Catalog.

Cabinet overview and labeling 0

The cabinet is organized in rows from A to K and in columns from Mk to LA3. A schematic
drawing of the layout is located on the back of the front door.
According to the labeling, connectors refer to the physical location (refer to example
A-LA3 in (Fig. 26 / p. 45))

SOMATOM Sensation CT02-023.841.04.18.02 Page 44 of 76 Siemens AG


10.07 CS PS CT Medical Solutions
IRS 1 (7394617) 45

A-LA3
1

10
3

8 9

Fig. 26: Front view


Pos. 1 Recon Computer 2 - IRSREC2
Pos. 2 Recon Computer 1 - IRSREC2
Pos. 3 Service Console
Pos. 4 Image Reformatting Computer - IRSTCR
Pos. 5 Acquisition Computer - IRSAQP
Pos. 6 ACQ2RECON, 1GBit switch
Pos. 7 RECON2REFORM, 1GBit switch
Pos. 8 Control, 100MBit switch
Pos. 9 KVM switch
Pos. 10 Power LED

Siemens AG CT02-023.841.04.18.02 Page 45 of 76 SOMATOM Sensation


Medical Solutions 10.07 CS PS CT
46 IRS 1 (7394617)

1
5

6
2

Fig. 27: Back view, bottom rack


Pos. 1 Image Reformatting Computer
Pos. 2 Acquisition Computer
Pos. 3 1 GBit switches
Pos. 4 100 MBit switch
Pos. 5 Current limiting relays, K1 and K2
Pos. 6 Current limiting, R2

Cabling 0

Tab. 1 LAN connections

From To
Location Component Location Switch
A-LA1 Recon Comp. 2 Row J Control, Port 1
B-LA1 Recon Comp. 1 Row J Control, Port 2
E-LA1 Image Reform. Comp. Row J Control, Port 3
F-LA1 Acquisition Comp. Row J Control, Port 4
E-LA3 Image Reform. Comp. Row J Control, Port 5
ICS Row J Control, Port 8
F-LA2 Acquisition Comp. Row G ACQ2RECON, Port 1
A-LA2 Recon Comp. 2 Row G ACQ2RECON, Port 2
B-LA2 Recon Comp. 1 Row G ACQ2RECON, Port 3
E-LA2 Image Reform. Comp. Row H RECON2REFORM, Port 1
A-LA3 Recon Comp. 2 Row H RECON2REFORM, Port 2
B-LA3 Recon Comp. 1 Row H RECON2REFORM, Port 3

SOMATOM Sensation CT02-023.841.04.18.02 Page 46 of 76 Siemens AG


10.07 CS PS CT Medical Solutions
IRS 1 (7394617) 47
Tab. 2 KVM connections

From To
Location Component Location Component
A-KVM Recon Comp. 2 Row K KVM Switch, CPU 4
B-KVM Recon Comp. 1 Row K KVM Switch, CPU 3
E-KVM Image Reform. Comp. Row K KVM Switch, CPU 2
F-KVM Acquisition Comp. Row K KVM Switch, CPU 1
K-KVM KVM switch Row D Service Console

Siemens AG CT02-023.841.04.18.02 Page 47 of 76 SOMATOM Sensation


Medical Solutions 10.07 CS PS CT
48 IRS 1 (7394617)

Prerequisites for replacements 5.1

Tools and auxiliary equipment 0

• Standard tool kit


• Recon computer only: Configuration floppy (delivered with the spare part)
• IRS Software

Opening covers 0

Most service activities can be performed from the front side of the cabinet.
Only the components of the current limiting are located at the back of the cabinet.
In some cases, opening back and/or side covers may ensure better access to parts and
components. It is especially useful to open the side cover when replacing cables.

NOTICE Risk of accident and injury


Tipping over of IRS cabinet
¹ Secure IRS again to the wall or to the floor according to
the “System installation instruction” upon completion of
work (e.g. after removing it for performing service).

Front and back door


Both the front and back doors are locked with a common key also used for the PDC and
WCS.
The cabinet is mounted on wheels. The mechanical brakes at the front wheels have to be
unlocked before moving. This allows the cabinet to be easily moved to a service position.

NOTICE Risk of damage


IRS may get damaged because of uncontrolled movement
¹ Lock mechanical brakes at IRS cabinet

Side covers
The side covers can be opened by pressing the red lever down (at the top) and up (at the
bottom) with a screwdriver.

NOTE Please note the ground connection inside the side covers.

SOMATOM Sensation CT02-023.841.04.18.02 Page 48 of 76 Siemens AG


10.07 CS PS CT Medical Solutions
IRS 1 (7394617) 49

Acquisition, Image Reformatting and Recon Computer 5.2

Preliminary steps 0

1. Shut down the software application for Wizard and Navigator.


Click SYSTEM and select END in the main user menu.
Wait until the following message appears “It’s now safe to turn off the com-
puter.”

NOTICE Shut down sequence:


¹ Shut down the Wizard first, and then the Navigator.

2. Switch off system at the control box.


3. Disconnect all cables on the front side of the corresponding computer rack.

Removing 0

1. Remove the four Phillips-head screws (shown in (1/Fig. 28 / p. 49) for Recon Com-
puter 2) and pull out computer system to provide access to the power plug in the back.

1 1

Fig. 28: IRSREC1 & 2


2. Disconnect the power cable.
3. Remove the computer system from the cabinet.

Installing 0

1. Place the new computer system onto the rack.


2. Insert the power cable at the back of the computer.
3. Guide the computer completely into the rack.
4. Secure the computer system with four Phillips-head screws.
5. Connect all cables at the front side of the computer system.

Siemens AG CT02-023.841.04.18.02 Page 49 of 76 SOMATOM Sensation


Medical Solutions 10.07 CS PS CT
50 IRS 1 (7394617)

Final steps 0

NOTICE Important information regarding Recon Computer:


¹ IRSREC1 and IRSREC2 have identical hardware and
therefore have to be identified as system 1 or 2. The
Recon computer, delivered as a spare part, is fac-
tory-configured as IRSREC1. This means you have to
configure the computer only if you use it as IRSREC2.

1. Switch on system at the control box.


2. Configuration procedure for IRSREC2 only:
Bring service console into service position: pull console out, lift LCD monitor, and switch
it on.
Switch IRSREC2 onto service console: select computer 4 on the KVM switch using the
+/- buttons.
There are two methods of booting the configuration floppy delivered with the spare part:
- First method:
Switch off the IRSREC2 by pushing the on/off button on the dashboard
(3/Fig. 28 / p. 49)
Insert configuration floppy into IRSREC 2.
Switch on IRSREC2.
The floppy now boots automatically.
- Second method:
Leave computer system on.
Insert configuration floppy into IRSREC2.
Press the Reset button on the dash board of IRSREC2 (3/Fig. 28 / p. 49).
The floppy now boots automatically.
Select Load DMI File in the AWARD DMI Configuration Utility mask using the arrow
keys (2 x right).
Press ↵ .
Select IRSREC2 using the arrow keys (2 x down).
Press ↵ .
Answer the question with Y ↵ .
Now the configuration is performed.
Press Esc and Y ↵ to quit.
Remove floppy.
Reboot IRSREC2.
3. Install IRS software of the corresponding computer according to “Installation of
Somaris/5 VA50”, and higher versions, chapter “SW Installation of IRS”.
4. Switch off system at the control box.
5. Switch on the system again to synchronize ICS with IRS.

SOMATOM Sensation CT02-023.841.04.18.02 Page 50 of 76 Siemens AG


10.07 CS PS CT Medical Solutions
IRS 1 (7394617) 51

6. For Reformatting Computer (IRSTCR) only:


Perform a restore of the IRS tables according to “Installation of SOMARIS/5 VA50”, or
higher version.
7. Perform test scans.

Siemens AG CT02-023.841.04.18.02 Page 51 of 76 SOMATOM Sensation


Medical Solutions 10.07 CS PS CT
52 IRS 1 (7394617)

100MBit-, 1GBit-Switch 5.3

News: 1GBit-Switch kit from January 2004 onwards 0

From approx. January 2004 onwards the 1GBit-Switch, part no. 7395283, will not be avail-
able any more.
The 1GBit-Switch, part no. 7395283, will be replaced by the 1GBit-Switch kit, part no.
7541829.

Fig. 29: 1 GBit-Switch kit


(The 1GBit-Switch kit contains two 1GBit switches, part no. 7394690.)
In case one of the old 1GBit-Switch, part no. 7395283, is defective both 1GBit-Switches
must be replaced by the new 1GBit-Switch kit.
Cable connections at the front side of the 1GBit-Switch kit see table (Tab. 1 / p. 46).
For reimbursement: The old 1GBit-Switch assembly must be send back as usual (the box
of the 1GBit-Switch kit can be used).

Preliminary steps 0

1. Shut down the software application for Wizard and Navigator.


Click SYSTEM and select END in the main user menu.
Wait until the following message appears “It’s now safe to turn off the com-
puter”.

NOTICE Shut down sequence:


¹ Shut down the Wizard first, and then the Navigator.

2. Switch off system using the on-site ON/OFF switch or F5 within the PDC.
3. Disconnect all cables at the front side of the corresponding switch.

SOMATOM Sensation CT02-023.841.04.18.02 Page 52 of 76 Siemens AG


10.07 CS PS CT Medical Solutions
IRS 1 (7394617) 53

Removing 0

1. Remove the Phillips-head screws ((1/Fig. 30 / p. 53) for the 1GBit switch and
(2/Fig. 30 / p. 53) for the 100MBit switch) and pull out the switch to obtain access to the
power cable.

NOTE In case you have to replace one of the 1GBit switches, you have to
remove both switches.

1
1

2 2

Fig. 30: Switches


2. Disconnect power cables at the back side of the corresponding switch.
3. Remove the switch from cabinet.
4. Exchange angle brackets from the old switch to the new one.

Installing 0

1. Move new switch into the rack to obtain access to the power connection at the back side
of the switch.
2. Insert the power cable.
3. Move switch completely into rack.
4. Secure switch with the Phillips-head screws.

Siemens AG CT02-023.841.04.18.02 Page 53 of 76 SOMATOM Sensation


Medical Solutions 10.07 CS PS CT
54 IRS 1 (7394617)

NOTE 100MBit switch only:


Make sure that the switch is centered at the correct height. Air slot
1 between the 100MBit and 1GBit switches has to be the same size
as slot 2 between the 100MBit and KVM switches (refer to
(Fig. 31 / p. 54)).

Fig. 31: Air slots


5. Connect all cables at the front side of the switch according to table (Tab. 1 / p. 46).

Final steps 0

1. Switch on circuit breaker F5 in the PDC and/or on-site power.


2. Switch on the computer system at the control box.
3. Start application and perform test scans.

SOMATOM Sensation CT02-023.841.04.18.02 Page 54 of 76 Siemens AG


10.07 CS PS CT Medical Solutions
IRS 1 (7394617) 55

Console 5.4

Preliminary steps 0

1. Shut down the software application for Wizard and Navigator.


Click SYSTEM and select END in the main user menu.
Wait until the following message appears “It’s now safe to turn off the com-
puter”.

NOTICE Shut down sequence:


¹ Shut down the Wizard first, and then the Navigator.

2. Switch off system using the on-site ON/OFF switch or F5 within the PDC.

Removing 0

1. Pull out the console.


2. Disconnect the power and monitor cables.
3. Open lower left side cover after first removing front or back door.
4. Disconnect and dismount mouse and keyboard cable at the left side of the IRS cabinet.
Note the cable layout.
5. Remove the six Phillips-head screws (three at each side) at the guides and remove
console with mouse and keyboard cable.

Installing 0

1. Secure the new console with six Phillips-head screws at the guides.
2. Connect the monitor and power cables at the console.
3. Guide the keyboard and mouse cables back into position and connect them at the left
side of the IRS cabinet. Secure cables with cable ties.
4. Push console back into the cabinet.

Final steps 0

1. Switch on circuit breaker F5 in the PDC and/or on-site power.


2. Check functionality of console.

Siemens AG CT02-023.841.04.18.02 Page 55 of 76 SOMATOM Sensation


Medical Solutions 10.07 CS PS CT
56 IRS 1 (7394617)

KVM switch 5.5

Preliminary steps 0

1. Shut down application software for Wizard and Navigator.


Click SYSTEM and select END in the main user menu.
Wait until the following message appears “It’s now safe to turn off the com-
puter”.

NOTICE Shut down sequence:


¹ Shut down the Wizard first, and then the Navigator.

2. Switch off system using the on-site ON/OFF switch or F5 within the PDC.

Removing 0

1. Remove the four Phillips-head screws (1/Fig. 32 / p. 56) and pull out the KVM switch to
provide access to the power cable in the back.

1 1

Fig. 32: Switches


2. Disconnect power cable.
3. Remove the KVM switch from cabinet.

Installing 0

1. Check the power-on switch at the back side of the KVM switch. Make sure that it is
switched to the ON position.
2. Move the new KVM switch into the rack to obtain access to the power connection at the
back side of the switch.
3. Insert the power cable.
4. Move the KVM switch completely into the rack.

SOMATOM Sensation CT02-023.841.04.18.02 Page 56 of 76 Siemens AG


10.07 CS PS CT Medical Solutions
IRS 1 (7394617) 57

5. Secure the KVM switch with four Phillips-head screws (1/Fig. 32 / p. 56).

Final steps 0

1. Switch on circuit breaker F5 in the PDC and/or on-site power.


2. Switch on the computer system at the control box.
3. Test functionality of the KVM switch.

Siemens AG CT02-023.841.04.18.02 Page 57 of 76 SOMATOM Sensation


Medical Solutions 10.07 CS PS CT
58 IRS 2, IRS 2c and IRS 2F
6-
Overview
6IRS 2, IRS 2c and IRS 2F

Computer equipment of IRS 2 0

The IRS 2 (73 94 641) and IRS 2c (7542603) cabinet comprises the following single com-
puters:
• 73 94 658 - AQP - Acquisition computer - CPU 1
• 73 94 682 - TCR - Image reformatting computer - CPU 2
• 73 94 666 - REC1,2 - Recon computer 1/2 - CPU 3 and CPU 4

NOTE The IRS2c differs from the IRS2 only in the number of REC comput-
ers: IRS2 has 2 RECs, the IRS2c is only used for newer Sensation
10 and has 1 REC computer.
If the system has an IRS2 with 2 RECs, it is not possible to run it
with only 1 REC.

The IRS 2F cabinet (83 78 353) comprises the following single computers
• 83 78 361 - AQP
• 83 78 387 - TCR
• 83 78 379 - REC 1, REC 2

Field replaceable units of IRS 2, IRS 2c and IRS2F 0

The field replaceable units of the IRS 2, 2c and 2F are as follows:


• AQP - Acquisition computer
• TCR - Image reformatting computer
• REC1,2 - Recon computer 1 and 2 (spare parts delivered are configured as REC1). The
IRS 2c has only one recon computer (REC1)
• 1GBit switch
• Service console
• KVM switch
• Current limiting & contactors
• Complete set of data cables
• Airflow filters
• Covers
For material numbers, refer to the Spare Parts Catalog.

Location of components and labeling 0

The cabinet is organized in rows from A to K and in columns from Mk to LA3. Find a sche-
matic drawing of the layout on the open front door.
All data cables are labeled with the connection points.

SOMATOM Sensation CT02-023.841.04.18.02 Page 58 of 76 Siemens AG


10.07 CS PS CT Medical Solutions
IRS 2, IRS 2c and IRS 2F 59

Prerequisites for replacements 6.1

Tools and auxiliary equipment 0

• Standard tool kit


• SW Configuration CD (delivered with the spare part)
• IRS Software

Opening covers 0

Most service activities can be performed from the front side of the cabinet.
Only the components of the current limiting are located at the back of the cabinet.
In some cases, opening back and/or side covers may ensure better access to parts and
components. It is especially useful to open the side cover when replacing cables.

NOTICE Risk of accident and injury


Tipping over of IRS cabinet
¹ Secure IRS again to the wall or to the floor according to
the “System installation instruction” upon completion of
work (e.g. after removing it for performing service).

Front and back door


Both the front and back doors are locked with a common key also used for the PDC and
WCS.
The cabinet is mounted on wheels. The mechanical brakes at the front wheels have to be
unlocked before moving. This allows the cabinet to be easily moved to a service position.

NOTICE Risk of damage


IRS may get damaged because of uncontrolled movement
¹ Lock mechanical braked at IRS cabinet

Side covers
The side covers can be opened by pressing the red lever down (at the top) and up (at the
bottom) with a screwdriver.

NOTE Please note the ground connection inside the side covers.

Siemens AG CT02-023.841.04.18.02 Page 59 of 76 SOMATOM Sensation


Medical Solutions 10.07 CS PS CT
60 IRS 2, IRS 2c and IRS 2F

AQP -, TCR -, REC1/2 - Computer 6.2

Preparation 0

1. Shut down the software application for Wizard and Navigator.


Click SYSTEM and select END in the main user menu.
Wait until the following message appears “It’s now safe to turn off the com-
puter.”

NOTE Shut down sequence:


First, shut down Wizard, then Navigator!

2. Switch off the system at the control box.


3. Disconnect all cables on the front side of the corresponding computer rack.

Removing 0

1. Remove the four Phillips-head screws and pull out the computer system to access the
rear.
2. Disconnect the power cable.
3. Remove the computer system from the cabinet.

Installing 0

1. Place the new computer system onto the rack.


2. Insert the power cable in back of the computer.
3. Guide the computer completely into the rack.
4. Secure the computer system with four Phillips-head screws.
5. Connect all cables at the front side of the computer system.

Installing correct BIOS and CMOS- settings 0

1. Switch on the system at the control box.


2. Bring service console into service position: pull console out, lift LCD monitor, and switch
it on.
Select the replaced computer via the KVM switch to see it on the display of the service
console (CT02-023.840.04 / Selecting a computer)
3. Insert the SW Configuration CD which is delivered with the spare part.

SOMATOM Sensation CT02-023.841.04.18.02 Page 60 of 76 Siemens AG


10.07 CS PS CT Medical Solutions
IRS 2, IRS 2c and IRS 2F 61

4. To restart the computer:


Click Start on Windows Task Bar
Select Shut down-> Restart.
5. The boot menu displays the selectable options.
6. Select the correct option for the IRS computer type that was changed. Make sure that
NumLock is not active (indicated by an LED)
(The option to select depends on the exchanged computer, e.g.,
“A” for AQP, “B” for TCR, “C” for REC1 and “D” for REC2.)
Press any key to continue (or CTRL + C and reboot, if selection was wrong)
Press any key to continue
¹ the Installation starts
¹ After installing the Bios the computer reboots automatically
7. During the reboot, remove the CD and then continue with the next step.

Installing IRS Software, TCR tables and update GREC-FPGA version (if applica-
ble) 0

1. Install the IRS software of the corresponding computer according to “Installation of Soft-
ware”.
2. For IRS 2F and IRS AQP with SW version VB20 only:
Install GREC-FPGA version 3.7:
a) Insert the update CD into the CD drive on the IRS AQP.
b) An autostart is performed. Wait until the Command window pops up.
c) Press any key to continue
d) Wait until the update is finished
e) Remove update CD from the drive
3. Switch off the system at the control box.
4. Switch on the system again to synchronize ICS with IRS.
5. For Reformatting Computer TCR only:
Perform a restore of the IRS tables according to “Installation of Software”.
6. Perform test scans.

Siemens AG CT02-023.841.04.18.02 Page 61 of 76 SOMATOM Sensation


Medical Solutions 10.07 CS PS CT
62 IRS 2, IRS 2c and IRS 2F

1GBit-Switch 6.3

Preparation 0

1. Shut down the software application for the Wizard and the Navigator.
Click SYSTEM and select END in the main user menu.
Wait until the following message appears “It’s now safe to turn off the com-
puter”.

NOTICE Shut down the sequence:


¹ Shut down the Wizard first, and then the Navigator.

2. Switch off the system using the on-site ON/OFF switch or F5 within the PDC.
3. Disconnect all cables at the front side of the corresponding switch.

Removing 0

1. Remove the Phillips-head screws and pull out the switch to obtain access to the rear.
2. Disconnect the power cables on the back side of the switch.
3. Remove the switch from the cabinet.
4. Exchange the angle brackets from the old switch with the new ones.

Installing 0

1. Move the new switch into the rack to obtain access to the power connection on the back
side of the switch.
2. Insert the power cable.
3. Move switch completely into rack.
4. Secure switch with the Phillips-head screws.
5. Connect all cables at the front side of the switch according to the labels.

Final steps 0

1. Switch on circuit breaker F5 in the PDC and/or on-site power.


2. Switch on the computer system at the control box.
3. Start the application and perform test scans.

SOMATOM Sensation CT02-023.841.04.18.02 Page 62 of 76 Siemens AG


10.07 CS PS CT Medical Solutions
IRS 2, IRS 2c and IRS 2F 63

Service Console 6.4

Preparation 0

1. Shut down the software application for the Wizard and the Navigator.
Click SYSTEM and select END in the main user menu.
Wait until the following message appears “It’s now safe to turn off the com-
puter”.

NOTICE Shut down the sequence:


¹ Shut down the Wizard first, and then the Navigator.

2. Switch off the system using the on-site ON/OFF switch or F5 within the PDC.

Removing 0

1. Pull out the console.


2. Disconnect the power and monitor cables.
3. Open the lower left side cover after first removing the front or back door.
4. Disconnect and dismount mouse and keyboard cable at the left side of the IRS cabinet.
5. Remove the six Phillips-head screws (three at each side) at the guides and remove the
console with the mouse and keyboard cable.

Installing 0

1. Secure the new console with six Phillips-head screws at the guides.
2. Connect the monitor and power cables at the console.
3. Guide the keyboard and mouse cables back into position and connect them at the left
side of the IRS cabinet. Secure the cables with cable ties.
4. Push the console back into the cabinet.

Final steps 0

1. Switch on circuit breaker F5 in the PDC and/or on-site power.


2. Check the functionality of the console.

Siemens AG CT02-023.841.04.18.02 Page 63 of 76 SOMATOM Sensation


Medical Solutions 10.07 CS PS CT
64 IRS 2, IRS 2c and IRS 2F

KVM switch 6.5

Preparation 0

1. Shut down the software application for the Wizard and the Navigator.
Click SYSTEM and select END in the main user menu.
Wait until the following message appears “It’s now safe to turn off the com-
puter”.

NOTICE Shut down the sequence:


¹ Shut down the Wizard first, and then the Navigator.

2. Switch off the system using the on-site ON/OFF switch or F5 within the PDC.

Removing 0

1. Remove the four Phillips-head screws and pull out the KVM switch to access the rear.
2. Disconnect the power cable.
3. Remove the KVM switch from the cabinet.

Installing 0

1. Check the power-on switch at the back side of the KVM switch. Make sure that it is
switched to the ON position.
2. Move the new KVM switch into the rack to obtain access to the power connection on the
back side of the switch.
3. Insert the power cable.
4. Move the KVM switch completely into the rack.
5. Secure the KVM switch with four Phillips-head screws.

Final steps 0

1. Switch on circuit breaker F5 in the PDC and/or on-site power.


2. Switch on the computer system at the control box.
3. Test the functionality of the KVM switch.

SOMATOM Sensation CT02-023.841.04.18.02 Page 64 of 76 Siemens AG


10.07 CS PS CT Medical Solutions
IRS 3 (83 77 785) 65
7-
Overview
7IRS 3 (83 77 785)

Computer equipment of IRS 3 0

The IRS 3 (83 77 785) cabinet comprises the following single computers:
• 83 77 793 - TCR - Image reformatting computer
• 83 77 694 - REC1,2,3 - Recon computer 1, 2 and 3

Field replaceable units of IRS 3 0

The field replaceable units of the IRS 3 are as follows:


• TCR - Image reformatting computer which contains also the aquisition functionality
• REC1,2,3 - Recon computer 1, 2 and 3 (spare parts delivered are configured as REC1).
• 1GBit switch
• Service console
• KVM switch
• Current limiting & contactors
• Complete set of LAN cables
• SCI cable 1.6 m
• Set of fans
• Set of airflow filters
• Covers
For material numbers, refer to the Spare Parts Catalog.

Location of components and labeling 0

The cabinet is organized in rows from A to K and in columns from Mk to LA3. Find a sche-
matic drawing of the layout on the blanking plate.
All data cables are labeled with the connection points.

Siemens AG CT02-023.841.04.18.02 Page 65 of 76 SOMATOM Sensation


Medical Solutions 10.07 CS PS CT
66 IRS 3 (83 77 785)

Prerequisites for replacements 7.1

Tools and auxiliary equipment 0

• Standard tool kit


• Screwdriver Torx TX10 (for replacement of fans)
• SW Configuration CD (delivered with the spare part)
• IRS Software

Opening covers 0

Most service activities can be performed from the front side of the cabinet.
Only the components of the current limiting are located in back of the cabinet.
In some cases, opening back and/or side covers may ensure better access to parts and
components. It is especially useful to open the side cover when replacing cables.

NOTICE Risk of accident and injury


Tipping over of IRS cabinet
¹ Secure IRS again to the wall or to the floor according to
the “System Installation Instructions” upon completion
of work (e.g. after removing it for performing service).

Front and back door


Both the front and back doors are locked with a common key also used for the PDC and
WCS.
The cabinet is mounted on wheels. The mechanical brakes at the front wheels have to be
unlocked before moving. This allows the cabinet to be easily moved to a service position.

NOTICE Risk of damage


The IRS may get damaged because of uncontrolled movement
¹ Lock mechanical brakes at the IRS cabinet

Side covers
The side covers can be opened with a coin or a large screwdriver.

NOTE Please note the ground connection inside the side covers.

SOMATOM Sensation CT02-023.841.04.18.02 Page 66 of 76 Siemens AG


10.07 CS PS CT Medical Solutions
IRS 3 (83 77 785) 67

TCR -, REC1/2/3 - Computer 7.2

Preparation 0

1. Shut down the software application for the Wizard and the Navigator.
Click SYSTEM and select END in the main user menu.
Wait until the following message appears “It’s now safe to turn off the com-
puter.”

NOTE Shut down sequence:


First shut down the Wizard and then the Navigator!

2. Switch off the system at the control box.


3. Disconnect all cables on the front side of the corresponding computer rack.

Removing 0

1. Remove the four Phillips head screws and pull out the computer system to access the
back.
2. Disconnect the power cable.
3. Remove the computer system from the cabinet.

Installation 0

1. Place the new computer system onto the rack.


2. Insert the power cable in back of the computer.
3. Guide the computer completely into the rack.
4. Secure the computer system with four Phillips head screws.
5. Connect all cables at the front side of the computer system.

Installing correct BIOS and CMOS- settings (only for IRS REC 1, 2 and 3) 0

1. Switch on the system at the control box.


2. Bring service console into service position: pull console out, lift LCD monitor, and switch
it on.
Select the replaced computer via the KVM switch to see it on the display of the service
console (CT02-023.840.04 / Selecting a computer)
3. Insert the SW Configuration CD which is delivered with the spare part.

Siemens AG CT02-023.841.04.18.02 Page 67 of 76 SOMATOM Sensation


Medical Solutions 10.07 CS PS CT
68 IRS 3 (83 77 785)

4. To restart the computer:


Click Start on Windows Task Bar
Select Shut down-> Restart.
5. The boot menu displays the selectable options.
6. Select the option for the correct IRS computer type. Make sure that NumLock is not
active (indicated by an LED)
(The option to select depends on the exchanged computer, e.g.,
“A” for TCR, “B” for REC1, “C” for REC2, “D” for REC3)
Press any key to continue (or CTRL + C and reboot, if selection was wrong)
Press any key to continue
¹ the Installation starts
¹ After installing the Bios the computer reboots automatically
7. During the reboot, remove the CD and then continue with the next step.

Installing IRS Software (and TCR- tables if applicable) 0

1. Install the IRS software of the corresponding computer according to the “Installation of
Software.”
2. Switch off the system at the control box.
3. Switch on the system again to synchronize ICS with IRS.
4. For Reformatting Computer TCR only:
Perform a restore of the IRS tables according to the “Installation of Software.”
5. Perform test scans.

SOMATOM Sensation CT02-023.841.04.18.02 Page 68 of 76 Siemens AG


10.07 CS PS CT Medical Solutions
IRS 3 (83 77 785) 69

1GBit-Switch 7.3

Preparation 0

1. Shut down the software application for the Wizard and the Navigator.
Click SYSTEM and select END in the main user menu.
Wait until the following message appears “It’s now safe to turn off the com-
puter”.

NOTICE Shut down the sequence:


¹ Shut down the Wizard first, and then the Navigator.

2. Switch off the system at the control box.


3. Disconnect all cables at the front side of the corresponding switch.

Removing 0

1. Remove the Phillips-head screws and pull out the switch to obtain access to the rear.
2. Disconnect the power cables in back of the switch.
3. Remove the switch from the cabinet.
4. Replace the angle brackets from the old switch with the new ones.

Installation 0

1. Move the new switch into the rack to obtain access to the power connection in back of
the switch.
2. Insert the power cable.
3. Move switch completely into rack.
4. Secure switch with the Phillips-head screws.
5. Connect all cables at the front side of the switch according to the labels.

Final steps 0

1. Switch on the computer system at the control box.


2. Start the application and perform test scans.

Siemens AG CT02-023.841.04.18.02 Page 69 of 76 SOMATOM Sensation


Medical Solutions 10.07 CS PS CT
70 IRS 3 (83 77 785)

Service Console 7.4

Preparation 0

1. Shut down the software application for the Wizard and the Navigator.
Click SYSTEM and select END in the main user menu.
Wait until the following message appears “It’s now safe to turn off the com-
puter”.

NOTICE Shut down the sequence:


¹ Shut down the Wizard first, and then the Navigator.

2. Switch off the system at the control box.

Removing 0

1. Pull out the console.


2. Disconnect the power and monitor cables.
3. Open the lower left side cover after first removing the front or back door.
4. Disconnect and dismount mouse and keyboard cable at the left side of the IRS cabinet.
5. Remove the six Phillips head screws (three at each side) at the guides and remove the
console with the mouse and keyboard cable.

Installation 0

1. Secure the new console with six Phillips-head screws at the guides.
2. Connect the monitor and power cables at the console.
3. Guide the keyboard and mouse cables back into position and connect them at the left
side of the IRS cabinet. Secure the cables with cable ties.
4. Push the console back into the cabinet.

Final steps 0

1. Switch on the system at the control box.


2. Check the functionality of the console.

SOMATOM Sensation CT02-023.841.04.18.02 Page 70 of 76 Siemens AG


10.07 CS PS CT Medical Solutions
IRS 3 (83 77 785) 71

KVM switch 7.5

Preparation 0

1. Shut down the software application for the Wizard and the Navigator.
Click SYSTEM and select END in the main user menu.
Wait until the following message appears “It’s now safe to turn off the com-
puter”.

NOTICE Shut down the sequence:


¹ Shut down the Wizard first, and then the Navigator.

2. Switch off the system at the control box.

Removing 0

1. Remove the four Phillips-head screws and pull out the KVM switch to access the rear.
2. Disconnect the power cable.
3. Remove the KVM switch from the cabinet.

Installation 0

1. Check the power-on switch at the back side of the KVM switch. Make sure that it is
switched to the ON position.
2. Move the new KVM switch into the rack to obtain access to the power connection in
back of the switch.
3. Insert the power cable.
4. Move the KVM switch completely into the rack.
5. Secure the KVM switch with four Phillips-head screws.

Final steps 0

1. Switch on the computer system at the control box.


2. Test the functionality of the KVM switch.

Siemens AG CT02-023.841.04.18.02 Page 71 of 76 SOMATOM Sensation


Medical Solutions 10.07 CS PS CT
72 IRS 3 (83 77 785)

Fan 7.6

Preparation 0

NOTE A complete fan set is delivered as a spare part. It includes two


fans 90 mm, two fans 120 mm with filters in the front fans, and
installation material.

1. Shut down the software application for the Wizard and the Navigator.
Click SYSTEM and select END in the main user menu.
Wait until the following message appears “It’s now safe to turn off the com-
puter”.

NOTICE Shut down the sequence:


¹ Shut down the Wizard first, and then the Navigator.

2. Switch off the system at the control box.

Removing 0

1. Loosen the small Torx screw (size TX10).

Fig. 33: Fan


2. Push the housing with the fans downward and pull it to the front.
3. Disconnect the power plug.

NOTE In the near future, each fan is attached with two screws.

SOMATOM Sensation CT02-023.841.04.18.02 Page 72 of 76 Siemens AG


10.07 CS PS CT Medical Solutions
IRS 3 (83 77 785) 73

Installation 0

NOTE The air flow is from the front to the back. Make sure that the cor-
rect fan is installed (refer to the labeling on the fan for air flow
direction).

1. Connect the power plug.


2. Insert the housing with the fan to the computer rack and push it upward.
3. Fix the fan with the Torx screw.

Final steps 0

1. Switch on the computer system at the control box.


2. Test the functionality of the fans

Siemens AG CT02-023.841.04.18.02 Page 73 of 76 SOMATOM Sensation


Medical Solutions 10.07 CS PS CT
74 Changes to previous version
8Changes to previous version 8-

Chapter Change Reason


3 Keyboard exchange for tower 5 adapted. Ä511227
4 Workflow for bios installation changed. Ä511227

SOMATOM Sensation CT02-023.841.04.18.02 Page 74 of 76 Siemens AG


10.07 CS PS CT Medical Solutions
Index 75

H
9Index

hz_serdoc_F12G02U07M02 . . . . . . . . .7
hz_serdoc_F13G01U02M01 . . . . . . . . .8
hz_serdoc_F13G01U05M08 . . . . . . . . .7
hz_serdoc_F13G01U12M01 . . . . . . . . .8
hz_serdoc_F13G01U12M02 . . . . . . . . .7
hz_serdoc_F13G01U14M01 . . . . . . . . .8
hz_serdoc_F13G08U01M01 . . . . . . . . .7

Siemens AG CT02-023.841.04.18.02 Page 75 of 76 SOMATOM Sensation


Medical Solutions 10.07 CS PS CT
76 Index