You are on page 1of 26

MAG

ALIA GROUP INC.


Electr omagnetic Flow meter

Operation Manual
AMC2100 Series

AMC3100 Series
Index
1. Sensor - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - 1
1.1 AMF900 Flange Type - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - 1
1.2 AMF500 Wafer Type - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - 2
1.3 AMF300 - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - 2
1.4 AMF500 - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - 2
1.5 AMF601 - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - 3
1.6 AMF900 - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - 3
2.Installation- - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - 4
2.1 Install position - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - 4
2.2 Remove the interference of magnetic field - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - 4
2.3 Straight tube length - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - 4
2.4 Install method - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - 4
3. Converter AMC2100 and AMC3100 Operate Manual - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - 6
3.1 Converter AMC2100 Operate Manual- - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - 6
3.1.1 Power and signal output connected line (Compact and separate mode) - - - - - - - - - - - 6
3.1.2 Separate wiring - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - 7
3.1.3 AMC2100 Panel deploy - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - 7
3.2 Converter AMC3100 Operate Manual - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - 8
3.2.1 Power and signal output connected line (Compact and separate mode) - - - - - - - - - - - 8
3.2.2 Separate wiring - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - 9
3.2.3 AMC3100 Panel deploy - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - 9
3.3 Survey mode setting- - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - 10
3.4 Example- - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - 10
3.5 Totalizer Rest - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - 10
3.6 Auto zero Adjustment- - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - 11
3.7 Check for power off message- - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - 11
3.8 Operate flow chart - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - 12
3.9 User operate description - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - 14
3.10 System Mode- - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - 18
3.11 Advance Mode- - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - 19
3.12 Batch Control - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - 21
4. Common Alarm Code Indication - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - 23

1.1.6ENG
1. Sensor
EM Flowmeter composed by sensor and Converter, be compact version or separate version, there are several
specification of sensor as follow:
92 92

1.1 AMF900 Flange Type 56

Size: 6mm ~ 2000mm (1/4” ~ 80”)


Liner : Neoprene
Polyurethane
K
FEP
PTFE
PFA
Protection : IP68 H
Max.Temp.:180 Deg. C
Installation : Flange End

L D

Standard
Size Liner Material Dimensions (mm)
Pressure
(mm) Kg/cm2 FEP / PFA Neoprene Polyurethane PTFE L D H
6
90 145
10
15 95 155
20 150 105 160
25 115 166
40
32 140 180
40 150 190
50 165 201
65 200 185 220
80 200 235
100 220 254
250
125 16 250 284
150 300 285 314
200 350 340 369
250 400 395 430
300 445 480
450
350 505 540
400 500 565 600
450 550 615 640
10
500 670 700
600
600 780 800
700 700 895 910
800 800 1015 1020
900 900 1115 1120
1000 1000 1230 1230
1200 1200 1405 1450
1400 1400 1630 1560
1600 6 1600 1830 1770
1800 1800 2045 2040
2000 2000 2265 2250

1
1.2 AMF500 Wafer Type
Size: 25mm ~ 200mm (1” ~ 8”)
Liner: FEP
Protection: IP68
Max.Temp.: 180 Deg. C
Installation: Wafer

Size Dimensions (mm)


mm Inch L D H
25 1" 90 71 138
32 1-1/4" 80 147
100
40 1-1/2" 86 153
50 2" 115 100 167
H 65 2-1/2" 115 120 187
80 3" 130 131 198
100 4" 155 151 218

D 125 5" 155 181 248


L

150 6" 185 206 273


200 8" 215 261 328

1.3 AMF300 Installation

1.4 AMF500 Installation

Tie-rod

NUT

Packing

Flange

2
1.5 AMF601 Installation

1.6 AMF900 Installation

3
2. Installation
When you design the tube, please consider following items:
2.1 Install position
Please avoid the sunlight when you install the flowmeter, the ambient temperature between-25~60 Deg.C will be great.
2.2 Remove the interference of magnetic field
Please DO NOT install flowmeter near motor-driven machine, transformer, frequency transformer etc. for it will cause interference
of magnetic field..
2.3 Straight tube length
In order to guarantee the EM Flowmeter accuracy, upstream and downstream of the installation should satisfied the following
conditions(Picture).

T - Tube

ELBOW 90°

> 5D

> 5D

Grove Valve

< 7 . 5°

>10D

When the upside and downside is shrink tube,


the degreeθ should be smaller then 15∘

Generally tube installation, the flowmeter Upstream could try to stay 5D-10D ,Downstream 2D-5D.

2.4 Install method


a. Flowmeter can be horizontal, vertical or slanting, please make sure the tube is full of fluid no matter if it is
moving or not.
b. When fluid contained solid material, It’s suggested to try vertical installation(bottom to top), which can avoid the
solid material sink inside the tube.

c. Pipe design should make sure no air bubble in the flowmeter position.

Grove Valve

4
d. Electrode position should parallel with ground
EM Flowmeter installed by horizontal or slanting the electrode position(A.B) should match the 2 side( right/left)of
tube, converter(wiring box)should be top of the tube.

Exhaust valve

>300mm

Horizontal install, the electrode position A.B should on the right and left side.

e. Transportation
When you moving, please remember DO NOT use rope through the flowmeter tube, it may cause Liner inside
broken.
Please DO NOT use your hand or rope to hang on Converter or wiring box, if Flowmeter size is bigger than 80mm,
because the material of converter and wiring box is tender aluminum, it can’t stand huge weight.

f. Grounding measure

Grounding Ring

Grounding resistance < 10 Ω Grounding resistance < 10 Ω


General metal tube Non-metal tube (plastic tube Liner)

5
3. Converter AMC2100 and AMC3100 Operate Manual
3.1 Converter AMC2100 Operate Manual
3.1.1 Power and signal output connected line(compact and separate model)
Open the 4 screws of the converter, and you can see terminal, decide if you need to connect according to your
needs.

F
U
S
E

I+ P+ R+ A C X L N PE
I- P- R-
P- B Y

1 .4 ~ 20mA output +
2.4~ 20mA output -
3 . Pulse output + + -
4. Pulse output -
Cable from sensor Power supply
5 . RS485 output +
90 ~ 260V AC
6. RS485 output -
50/60Hz
or
24VDC

1 2 3 4 5 6
+ -
+ +
mA

Load Load

Output a load: 600Ω Max. Outer Voltage: 30VDC Modbus protocol

4~20mA Output 1.Pulse (Frequency)Output RS485 Output


( Hart protocol)
2. Batch Control

6
3.1.2 Separate Wiring
Converter
AMC2100
Series
A B C X Y

Electorde
Coil

Sensor
AMF300 / AMF500
AMF601 / AMF900
Series A B C X Y

3.1.3 AMC2100 Panel deploy

70 mm

Σ 123.456 m3
+0.000 m3/h

150 mm
150 mm

Key Button Normal status function Parameter setting status function


Name Sign

ENT Login parameter setting level Save presently settled parameter. Jump to next
Setting parameter setting

Selected one of 4 display line Change present digit, dot and survey unit.
Up Up/Down movement

Change the selected line on the display


Right
Move

7
3.2 Converter AMC3100 Operate Manual
3.2.1 Power and signal output connected line(compact and separate model)
Open the rear cover, and you can see terminal, according to your needs decide if you need to connect.

+ + +

-
-

-
1 . 4~20mA output +
2 . 4~20mA output -
3 . Pulse output +
4 . Pulse output -
5 . RS485 output +
6 . RS485 output -

+ -

Power supply
90~260VAC
50 / 60Hz
or
24VDC

I+ I- P+ P- R+ R-
+ -
+ +
mA
Load Load

Output a load: 600Ω Max. Outer Voltage: 30VDC Modbus protocol

1.Pulse (Frequency)Output RS485 Output


4~20mA Output
( Hart protocol) 2. Batch Control

8
3.2.2 Separate Wiring
Converter
AMC3100
Series
A B C X Y

Electorde
Coil

Sensor
AMF300 / AMF500
AMF601 / AMF900
Series A B C X Y

3.2.3 AMC3100 Panel deploy

190mm

>
>

20mm

Key Button Normal status function Parameter setting status function


Name Sign

ENT Login parameter setting level Save presently settled parameter. Jump to next
Setting parameter setting

Selected one of 4 display line Change present digit, dot and survey unit.
Up Up/Down movement

Change the selected line on the display


Right
Move

9
3.3 Survey mode setting
Flowmeter will enter to normal mode when electrified, every line has several choice as follow:

Position Display Unit sign Unit Description


Totalizer Σ
Line 1 Totalizer+ Σ+ L, M3, G , kg, T Σ = Σ+ (−) Σ−
Totalizer- Σ−
Velocity m/s When reverse flow, it will show”- ”
Flowrate m3/h When reverse flow, it will show”- ”
Line 2 % of Full Scale %FS Actual flowrate should match
flowrate%
mA output mA Actual mA output Value
value
Frequency Hz Full scale output : 5000Hz

3.4 Example
If you need Line 1 display from m3/h turn to %
Σ 123.456 m3
+0.0000m3/h

Σ 123.456 m3 In normal display press button until Line 2 is selected.


+0.0000m3/h

Σ 123.456 m3 Press ENT


button to change m3/h to %FS
+0.0000 %FS

Σ 123.456 m3 When alarm select” batch control”,It will display” ”


+0.0000 %FS

3.5 Totalizer Reset

Σ 123.456 m3 In general display, Press or button and select ,make sure enter zero window
+0.0000m3/h
ENT

Σ 123.456 m3 Zero window.


Σ1 123.450 m3 Σ current totalizer value Σ1 last totalizer value Σ2 the time before last time totalizer value
Σ2 123.400 m3 Please selected CLR and press enter,doing the zero operation
EXIT CLR

ENT

Clear Total Select YES for making zero operation and press ENT
to confirm

NO YES

10
3.6 Auto zero Adjustment
When the tube are full pipe, and no flow, flowmeter still show the flowrate, then you could use Auto Zero adjustment
to adjust your flowmeter to Zero, the method as follow:
Σ 123.456 m3 In the normal display, press ENT
button for 5 seconds, then it will go to next setting
+0.1234m3/h

ENT

OPERATIONS In this setting, Press or button till you selected ZERO and press ENT
to confirm

SET ZERO ESC

ENT

ZERO Select YES, and press ENT


to confirm .

NO YES

ENT

Zero Adjustment.
Zero Triming …

ENT

Zero Tirm End Zero Adjustment end. Select ESC, press to turn back to the operations select
ENT

-2.624mv (If you want to do Zero Adjustment again, Pls select ZERO )

Zero Esc
ENT

OPERATIONS Select ESC, press ENT


to confirm and turn back to the normal display

SET ZERO ESC

ENT

Σ 123.456 m3 the normal display


+0.0000m3/h

3.7 Check for power off message


Σ 123.456 m3 In general display, Press or button to select ,Press ENT to confirm, then check for the
+0.0000m3/h power off message window

ENT

Power off message window


ON 2008/04/23 ON: date & time when power on OFF: date & time when power off
13:30:55
Q: the flowrate when power off Σ: the totalizer when power off
OFF 2008/04/23
13:30:55
Example: 01 means the power off recored series No.(It could record 99 times of power off)
Q: 1.0234 m3/h Press PgUP or PgDN check the history record of power off.
Σ: 4.123 m3 Ps: Press PgDN till the last page, OFF option means current date, time, and flowrate.
01 PgUP PgDN ESC

11
3.8 Operate flow chart
Σ 123.456 m3
+0.0000m3/h

ENT

OPERATIONS

SET ZERO ESC

ENT

Password

ENT
0000

Language Empty Detect Empty Limit(%)

ENT ENT ENT

PV Units Communication

ENT ENT

URV
MODBUS Type HART Address
ENT

ENT ENT
Switch Func

ENT
Baudrate

ENT
Liter/Pulse Freq of URV (Hz)

ENT ENT
Data Bit

ENT
Pulse Width Alarm Select

ENT ENT
Parity

ENT
Pulse Level UPPER-Alarm
ENT
Stop Bit

ENT
Low Cutoff(%fs)

ENT
Modbus Address

ENT
Damping(s)

ENT
DATE (yy-mm-dd)

ENT
Hydro-State

ENT
TIME (hh-mm-ss)

Direction ENT

ENT

Indication

12
ENT

Password Password

1111 ****

Sensor NO. Sensor NO.

ENT
ENT

Diameter (mm) Diameter (mm)

ENT
ENT

Sensor Factor Sensor Factor

ENT
ENT

Output Check Output Check

ENT
ENT

Restore Shipping Loop Test Restore Shipping

ENT ENT
ENT

Pulse Test Convertor NO.

ENT ENT

Coil Test Convertor Factor

ENT

Coil Current (A)

ENT

PGA

ENT

Coil Freq (Hz)

ENT

Zero Factor

ENT

4mA Trim

ENT

20mA Trim

13
3.9 User operate description
Σ 123.456 m3 In the normal display, press ENT
button for 5 seconds, then it will go to next setting
+0.0000m3/h

ENT

OPERATIONS In this setting, Press or button till you selected SET

SET ZERO ESC

ENT

Password Password login window


Input the password
0000 Password: 0000

ENT

Language Language
Press or to selected the language
简体 繁體 ENG

ENT

PV Units Select Flow Engineer unit


m3/h L/h GPH Press or to select the Engineer unit
Kg/h t/h GPM 0: m3/h 1: L/h 2: GPH 3: Kg/h 4: t/h 5: GPM
L/min m3/min 6: L/min 7: m3/min

ENT

URV m3/h Flow Range


Setting Flow Range, that’s mean 4-20mA and Pulse, frequency output max. Value
Press button to move, Press to change the Value
70.0000

ENT

Switch Func Contact output


Press or button to selected the output
Default Pulse 1. Default 2. Pulse 3. Frequency 4. Alarm
Freq Alarm

ENT
Batch Control When selected “Alarm”, It will display “ Batch Control”
Please read page 22

Liter/Pulse (L/P) Pulse output


Each pulse match the Liter value (Liter/Pulse)
Press button to move, Press to changed the Value
0.1000

ENT

Pulse Width Pulse width


Press or to selected Pulse width
1ms 10ms 20ms
100ms 1s 50%
ENT

14
Pulse Level Pulse Level
Press or to selected Pulse Level
Active L Active H

ENT

Low Cutoff (%fs) Function for cutoff the low flowrate


Press button to move, press button to change the Value
Setting Range: 0.0% ~ 9.9%
0.0

ENT

Damping (s) The flowrate response time setting,


Press button to move, Press button the Value
Setting Range:01 ~ 99 Seconds, Factory Standard set value is 2 Seconds
01

ENT

Hydro – State Hydro-State


Press or button to selected the Hydro-State
Normal Waved Normal: for general application
Waved: For fluid in wave states

ENT

Max Limit (%fs) Response speed adjustment


Press button to move, Press button to change the value
Range: 00~30%, 1% means max 1% flowrate value change for each Limit Time
05

ENT

Limit Time (s) Response time


Press button to move, Press to change the value
Response time range: 00~99, Adjust flowmeter response time.
05

ENT

Direction Direction
Press or button to select the Hydro-state
Fwd. Rev. Bid. 1. Forward 2. Reverse 3. Bidirection

ENT

Indication Indication
Press or button to select the Indication
Forward Reverse 1. Forward 2. Reverse

ENT

Empty Detect Empty pipe detect


Press or button to select if it needs empty pipe detect
NO YES

ENT

Empty Limit (%) Empty pipe detect the value


Press button to move, Press to change the value

50
ENT

15
Communication Communication
Press or button to select if need to use communication function
OFF ON

ENT
HART Address Please read page 18

00

MODBUS Type MODBUS Mode


Press or button to select MODBUS mode
RTU ASCII

ENT

Baudrate RS485 Speed


1200 2400 Press or button to select communication speed.
4800 9600
19200 38400

ENT

Data Bit Data Bit


Press or button to select the Data Bit
7 8 1. 7 2. 8 (When MODBUS model selects RTU, the Data Bit is 8 which is fixed)

ENT

Parity Parity Type


Press or button to select the parity type
NONE ODD EVEN 1. None 2. Odd 3. Even

ENT

Stop Bit Stop Bit


Press or button to select the Stop Bit
1 2 1. 1 2. 2

ENT

MODBUS Address MODBUS address


Press button to move, Press button to change the digit
001

ENT

DATE (yy-mm-dd) Date Setting


Press button to move, Press button to change the digit.
08-01-10

ENT

TIME (hh:mm:ss) Time Setting


Press button to move, Press button to change the digit.
Pls press button for 5 seconds, then it will go back to the normal display.
ENT

17:09:49

16
HART operate description

HART Address HART Address


Press or button to select HART Address
00

ENT

DATE (yy-mm-dd) Date Setting


Press button to move, Press button to change the digit.
08-01-10

ENT

TIME (hh:mm:ss) Time Setting


Press button to move, Press button to change the digit.
Pls press ENT
button for 5 seconds, then it will go back to the normal display.
17:09:49

17
3.10 System Mode

OPERATIONS In the normal display, press button for 5 seconds, then it will go to next setting
ENT

In this setting, Press or button till you selected SET


SET ZERO ESC

ENT

Password Password login window


Input the system password:
1111 Password: 1111

ENT

Sensor NO. Sensor No.


Press button to move, Press button to change the digit.
A7082106

ENT

Diameter (mm) Diameter


Press button to move, Press button to change the diameter
0065

ENT

Sensor Factor Sensor Factor KS


KS is the original factory rectified factor, please don’t change it unless professional or technical
staff Press button to move, Press button to change the value
00.94100

ENT

Output Check Output Calibration


Press or button to choose YES select if needs calibration
NO YES Choose NO jump to Restore Shipping

ENT

Loop Test 4-20 mA simulated output


Press or button to select the simulated output value
4 8 12 4 . 8 . 12 . 16 . 20
16 20

ENT

Pulse Test Pulse Simulated output


Press or button to select the calibration type
1 10 100 1 . 10 . 100 . 1000 . 2000 . 5000
1000 2000 5000

ENT

Coil Test Excit voltage calibration


Check the coil X to Y or Y to X excit signal are correct or have fail
ON XJY YJX Press or button to select the calibration type
Choose X->Y, the Voltage will be 24VDC, Choose Y->X, then Y to X will be 24 VDC

ENT

Restore Shipping Restore the parameter to second memory


Press or button to select if it needs to restore the factory setting
NO YES Pls press button for 5 seconds, then it will go back to the normal display.
ENT

18
3.11 Advance Mode

OPERATIONS In the normal display, press button for 5 seconds, then it will go to next setting
ENT

In this setting, Press or button till you selected SET


SET ZERO ESC

ENT

Password Password login window


Input the system password:
**** Password: * * * *

ENT

Sensor NO. Sensor No.


Press button to move, Press button to change the value

A7082106

ENT

Diameter (mm) Diameter


Press button to move, Press button to change the diameter
0065

ENT

Sensor Factor Sensor Factor KS


KS is the original factory rectified factor, please don’t change it unless professional or technical
staff
00.94100 Press button to move, Press button to change the value

ENT

Output Check Output Calibration


Press or button to choose YES select if needs calibration
NO YES Choose NO jump to Restore Shipping

ENT

Loop Test 4-20 mA simulated output


Press or button to select the calibration type
4 8 12 4 . 8 . 12 . 16 . 20
16 20

ENT

Pulse Test Pulse Simulated output


Press or button to select the calibration type
1 10 100 1 . 10 . 100 . 1000 . 2000 . 5000
1000 2000 5000

ENT

Coil Test Excit voltage calibration


Check the coil X to Y or Y to X excit signal are correct or have fail
ON XJY YJX Press or button to select the calibration type
Choose X->Y, the Voltage will be 24VDC, Choose Y->X, then Y to X will be 24 VDC

ENT

Restore Shipping Restore the parameter to second memory


Press or button to select if it needs to restore the factory setting
NO YES

19
ENT

Converter NO. Converter No.


Press button to move, Press button to change the value

B0000000

ENT

Converter Factor Converter Factor KC

Press button to move, Press button to change the value


01.05988

ENT

Coil Current (A) Coil Current IC(A)


Ic is the tested current of the converter when out of factory, please don’t change it unless
agent or professional technical staff.
0.1654 Press button to move, Press button to change the value

ENT

Coil Freq (Hz) Coil Frequency (Hz)


Press or button to select the coil frequency
3.125 6.25 33 Original setting Value: 6.25

ENT

PGA PGA
Press or button to select the gain type
x 1. x 2 . x 4. AUTO Original setting “AUTO”

ENT

Zero Factor Zero Factor manual setting


Press button to move, Press to change the value

0.0

ENT

4mA Trim 4mA output calibration


Press button to move, Press to change the Value

4.000

ENT

20mA Trim 20mA output calibration


Press button to move, Press to change the value
Pls press button for 5 seconds, then it will go back to the normal display.
ENT

20.000

20
3.12 Batch Control

Σ 123.456 m3 In the normal display, press ENT


button for 5 seconds, then it will go to next setting
+0.0000m3/h

ENT

OPERATIONS In this setting, Press or button till you selected SET

SET ZERO ESC

ENT

Password Password login window


Input the password
0000 Password: 0000

ENT

Language Language
Press or to selected the language
简体 繁體 ENG

ENT

PV Units Select Flow Engineer unit


m3/h L/h GPH Press or to select the Engineer unit
Kg/h t/h GPM 0: m3/h 1: L/h 2: GPH 3: Kg/h 4: t/h 5: GPM
L/min m3/min 6: L/min 7: m3/min
ENT

URV m3/h Flow Range


Setting Flow Range, that’s mean 4-20mA and Pulse, frequency output max. Value
Press button to move, Press to change the Value
70.0000

ENT

Switch Func Contact output


Press or button to selected the output, Pls choose the “Alarm”
Default Pulse 1. Default 2. Pulse 3. Frequency 4. Alarm
Freq Alarm

ENT

Alarm Select Alarm Select


Press or button to “BAT”
MAX MIN 1. Max(Upper-Alarm) 2. Min(Lower-Alarm) 3. Batch 4. Direction
BAT DIR

ENT

Low Cutoff (%fs) Function for cutoff the low flowrate


Press button to move, press button to change the Value
Setting Range: 0.0% ~ 9.9%
0.0 Pls press button for 5 seconds, then it will go back to the normal display.
ENT

ENT

21
Σ 123.456 m3 In general display, Press or button and select ,make sure enter batch control window
+0.0000m3/h

ENT

Batch Control Batch Control


Press or button to SET
SET AUTO
MANUAL EXIT
ENT

Batch Value(m3) Batch Value(m3)


Press button to move, press button to change the Value

1.00000

ENT

Batch Control Batch Control


Press or button to AUTO or MANUAL
SET AUTO
MANUAL EXIT

ENT

SET 1.000m3 Press or button to START ,To start batch control


RUN 0.000m3 If choose CLR ,The actual flow value will be zero

START CLR EXIT

ENT

SET 1.000m3 In this setting, Press ENT


button , The flow metering will be pause
RUN 0.000m3

PAUSE

ENT

SET 1.000m3 At this time, If choose START,It will continue the next flow
RUN 0.000m3 If choose EXIT, It will quit the batch control

START CLR EXIT

ENT

Batch Control Batch Control


Press or button to EXIT
SET AUTO Press button. then it will go back to the normal display.
ENT

MANUAL EXIT

22
4. Common Alarm Code Indication

AMC2100/AMC3100 Alarm Table

Code Contents Meaning Solution

8 OV. Flow Over range Increase range

16 OV. Flow Exceeding AD value of flowrate Slow fluid velocity

Over range Increase range


24 OV. Flow
Exceeding AD value of flowrate Slow fluid velocity

32 OV. Freq Exceeding Fmax range Increase the frequency setting range

Exceeding the upper limit value of Turn off alarm output or raise alarm upper limit
64 UPPER
alarm settings value
Exceeding the upper limit value of Increase range
72 OV. Flow ; UPPER alarm settings Turn off alarm output or raise alarm upper limit
Over range value
Under the lower limit value of alarm Turn off alarm output or decrease the alarm
128 LOWER
setting lower limit value
Over range Increase range
136 OV. Flow ; LOWER Under the lower limit value of alarm Turn off alarm output or decrease the alarm
setting lower limit value

256 EMPTY Empty pipe alarm Check the pipe & full the pipe

512 Zero Zero mv value>99mv Re-zero when fluid is static.

1024 Coil Exciting Current<40% Check the XY wiring fall off or not

Over range Increase range


1032 OV. Flow ; Coil
Exciting Current <40% Check the XY wiring fall off or not
Exceeding the upper limit value of Turn off alarm output or raise alarm upper limit
1088 UPPER ; Coil alarm settings value
Exciting Current <40% Check the XY wiring fall off or not
Under the lower limit value of alarm Turn off alarm output or decrease the alarm
1152 LOWER ; Coil setting lower limit value
Exciting Current <40% Check the XY wiring fall off or not
Empty pipe alarm Check the pipe to full the pipe
1280 EMPTY ; Coil
Exciting Current <40% Check the XY wiring fall off or not
Zero mv value>99mv Re-zero when fluid is static.
1536 Zero; Coil
Exciting Current <40% Check the XY wiring fall off or not

23
Quality we care !

ALIAMAG ALIAPANEL ALIASONIC

ALIADP ALIAPT ALIAVA ALIAVTX

Tel: +1-302- 213- 0106


Fax: + 1- 302- 213- 0107
URL: www.alia-inc.com
Email: alia@alia-inc.com

113 Barksdale Professional Center,Newark,DE 19711,USA

24