Precision and accuracy in


ATP hygiene testing
ATP bioluminescence is a well established rapid method that gives instant results, and is used to
measure cleanliness and hygiene. Advances in solid-state technology have enabled new instruments
to be developed that deliver performance, convenience and robustness at low cost.

One such instrument is trend analysis (Sure Trend) with samples. All sampling devices
SystemSURE Plus (Hygiena pre-defined and customised have a simple snap and squeeze
International) which was report generation at the click of activation step with 12-months’
recognised in Business Excellence a button. shelf life and tolerance to ambient
Awards for Innovation in 2008, and ■ Other features include sampling temperature abuse. These devices
has the following unique features: plans, re-test tag and trace are slim and lightweight, thus
■ Low background that minimises function, on-screen statistical combining ease-of-use with
signal to noise ratio which review of results without the reduced environmental impact
maximises sensitivity and need to download to PC. because they contain fewer parts
enables reliable detections at and are 50–60% lighter than
low ATP levels, therefore comparable products.
reducing false positives. All sampling devices Low background noise
have a simple snap and
■ The output scale (Relative Light One of the unique features of the
Units, RLU) is quantitative and squeeze activation step
systems is the very low
1 RLU = 1 fmol ATP (higher with 12-months’ shelf background noise which means
RLU numbers do not mean life and tolerance to that the system can clearly
higher sensitivity). ambient temperature differentiate between results at
■ Self-calibration that significantly abuse very low RLU values unlike other
reduces service and systems. This low background
maintenance costs, and the low confers a high degree of precision
The sample collection and testing
voltage supply means that and sensitivity to the system at
devices are integrated all-in-one
simple AA pen batteries can be low ATP levels.
and ready-to-use, and contain a
used compared to expensive novel liquid-stable reagent Campden-BRI (the largest
dedicated mains/re-chargeable formulation. These sampling/ membership based food and drink
systems. testing devices are available for research centre in the world) was
■ Secure data download to PC by solid surfaces (Ultrasnap), and for commissioned to test the
either of two software options liquid samples (Aquasnap), such as systemSURE Plus Instrument.
including advanced, simplified CIP rinses and other water While Campden-BRI does not

Figure 1: Schematic description of Figure 2: Variation of ATP detection systems


accuracy and precision

food issue one | 2009


54 analysis and control

endorse specific products or Continuous Improvement industry, the application of


services. Its conclusions were that Aquasnap is mainly the detection
Luciferase reagent preparations
SystemSURE Plus produces of product residues in rinse water
and their delivery devices for ATP
accurate and repeatable results. samples. However, in certain
detection vary from supplier to
The ATP and the food residue situations it can be used to
supplier and are optimised for
detection experiments estimate microbial populations.
each system. Each reagent system
demonstrated the sensitivity and
is a balanced cocktail of enzyme, For industrial water samples eg
repeatability of the SystemSURE
co-factors, buffer and extractant. cooling towers, the monitoring and
Plus system to detect ATP over a
control of biofilm and biomass is
range of concentrations and for a The robustness and sensitivity of
important for process efficiency,
range of food groups with the ATP reagent preparations can
biocide dosage and the control of
reproducible results. be improved to meet the more
water borne risks such as
stringent test requirements from
Accuracy and precision Legionella. In these water
harsh samples for a variety
treatment systems, ATP from
Accuracy and precision are a different industrial applications.
organic sources is typically very
consequence of the variation of a Hygiena has developed a new
low such that the main source of
method (as shown in Figure 1). In formula with more robustness and
ATP is from microbial
an extensive study conducted by sensitivity that is incorporated into
contamination and biofilm.
an independent international the Aquasnap and Supersnap
Historically, the limit of detection
laboratory, accuracy and precision devices. Both devices have more
for microbes in water using a
of different ATP systems were resilience against samples at
direct ATP test on water was
studied. Figure 2 shows the results extremes of pH and chemical
10,000–100,000cfu/ml but
from 30 replicate samples of ATP interference, more light output per
Aquasnap and Supersnap can now
tested directly in five test systems unit of ATP, and better extraction
detect 1000cfu/ml (see Figure 4).
by the same analyst. Hygiena of ATP. These characteristics,
SystemSURE Plus was shown to combined with the low background Accordingly, SystemSURE Plus
be the most accurate and precise to minimise the signal to noise with its low background noise,
ATP test system because results ratio deliver enhanced detection of consistent and reproducible
have an acceptable standard error ATP (0.05fmols) in systemSURE reagent performance with
limit of +/– 10%. Other systems Plus which is at least tenfold excellent sample collection and
can vary up to 130%, which is better than comparable systems. recovery delivers the best
highly imprecise and very performance of repeatable,
Supersnap has a swab format for
inaccurate. accurate and precise results. ■
collecting surface samples for
In addition, the Hygiena hygiene monitoring and uses the
systemSURE recovered most ATP patented double-snap valve
Martin Easter
from the swab and with the least device. Aquasnap is a simple
General Manager
variation thus enabling reliable, convenience device designed to
Hygiena International Ltd
precise and accurate sample collect and test liquid samples www.hygiena.net
collection (see Figure 3). (0.1ml). For the food and beverage

Comp 1 = Competitor system 1 Comp 2 = Competitor system 2


Figure 3: Recovery of ATP from the swab by different Figure 4: Microbial detection in water using ATP
ATP detection systems Bioluminescence.