Professional Documents
Culture Documents
KENZA 240
Random Access Analyzer
Service Manual
Clinical Chemistry Automatic and Diagnostic Analyzers
The photographs contained in this manual have only an explanatory function; therefore, some non-functional features
may not correspond to the device in question.
KENZA 240
Service Manual
1. 1. Introduction pg. 3
1. 2. Readout system pg. 5
1. 3. Robotic system pg. 6
1. 4. Control system pg. 8
1. 5. Power supply unit pg. 12
1. 6. Hydraulic system pg. 12
1. 7. Technical specifications pg. 14
2 - Maintenance pg. 17
2. 1. Cuvette CK pg. 19
2. 2. Washing the cuvettes pg. 21
2. 3. Prime function pg. 21
2. 4. Setting the instrument pg. 21
2. 5. Sample/ Reagent level detector pg. 21
2. 6. Setting the barcode reader pg. 23
2. 7. Preventive maintenance pg. 24
1.1 Introduction
G KENZA 240 is an automatic analyser with microprocessor control, for clinical chemistry and
imunotur- bidimetry analysis, available in the table-top version. The “Random Access” operation
mode allows
rapidly running all the tests on a patient and printing the report, without having to wait the routine
to be completed. Moreover, you can add new patients, even with STAT priority, whilst running the
tests and, if necessary, repeat calibrations. KENZA analyser has been designed to achieve a
throughput of
240 tests/hour.
The graphical interface has been designed to make all the functions of this appliance simple and
intuiti-
ve, as well as to provide flexibility as required by every laboratory. KENZA 240 is composed of the
parts described in the following page.
3
1 KENZA 240
General Description
Pos. Name
1 Protection cover
2 Diluter module access panel
3 Distilled water and water drainage containers+level detector sockets unused 25-pin
con- nector junction nozzles
4 Halogen lamp housing access panel
5 power supply housing ventilation grid
6 Reagent cooling system forced ventilation socket
7 Switch+fuses+network filter unit
8 Voltage selector
9 Serial communication port
10 Sample and reagents tray
11 Cuvette tray
12 Cuvette washing arm
13 Sampling arm
4
KENZA 240
1
Service
Manual
This system is composed of a photometer equipped with a rotor containing a set of interference
filters;
an optic fibre connection and of a carousel containing 50 plastic optical cuvettes. It is also equipped
with
an automatic optical cuvette washing-rinsing and drying station for their next use.
The interference filters are installed on a rotor, which turns at a constant speed during the readout,
than-
ks to a DC motor. At the photometer outlet, a series of monochromatic pulses are transmitted via the
op-
tic fibre connection to the cuvette carousel, where the rays of light pass through each optical cuvette.
It is important to notice that each cuvette is crossed by a series of monochromatic light pulses,
which allows measuring transmittance at various wavelengths. These measurements are then
selected accor- ding to the test parameters.
Once the transmitted light reaches the solid-state detector located on the opposite side of the
selected cuvette, it is converted into an electrical signal, which is then amplified and sent to the
A/D Conversion circuit.
The KENZA 240 photometer provides a series of light pulses with different wavelengths, thanks to
a rotor onto which are installed a set of interference filters. The used light source is a halogen
quartz lamp po- wered at 5.8 V. The produced light is concentrated into a converging beam by a
convex and plain-convex lenses unit and passes through the filters, which turn at a constant
speed; the train of light pulses is then transmitted to the reading cuvettes via an optic fibre. The
rotation of the filter carousel occurs by means
of a DC motor, while the rotation control and the synchronisation for the readout of every single filter
are obtained via two pulse signals that result from the passage of the specially shaped filter wheel
through
a pair of optical couplers.
1. A halogen quartz lamp with an adjusted and stabilised power supply, which emits light in the
visible,
UV and near Infrared regions.
2. A filter carousel that contains a set of 9 narrow band interference filters. By continuously
rotating during the readout, this filter carousel emits a train of monochromatic light pulses. This
way, it is pos- sible to have a readout of the cuvettes with all the available wavelengths.
Moreover, the continuous rotation of the filter carousel allows dissipating the heat absorbed by
each filter, thus increasing their average life. The rotation phase and speed are electronically
controlled.
5
1 KENZA 240
General Description
Position 1 is also used for controlling the Offset i.e., the amplifier output voltage in the absence of
light
Position 1 (Offset) is closed in order to prevent the light from passing and allow the readout of
the rest current of the amplifier.
The train of light pulses deriving from the photometer is transmitted to the readout cuvette tray
via an optic fibre. This rotating tray contains 50 plastic cuvettes with a 0.6 cm optical path, in
which chemical reactions occur, after which the readout of the absorbencies related to the
intensity of the colours de- veloped during the reactions is performed. A resistive system with
electronic temperature control heats
the air around the cuvette tray, so that the induced temperature stabilises the cuvette temperature
at
37°C.
The rotating movement of the loader is due to a stepper motor with a 360° rotation.
When in idle state, the cuvette tray is positioned so that cuvette 21 in front of the ray of light.
The KENZA 240 robotic system involves all the mechanical parts related to fluid sampling operations
and
to the placement of the reagents and samples necessary for running the tests. The readout
module is also integral part of this system.
6
KENZA 240
1
Service
Manual
KENZA 240 is an instrument that operates in SIP & DIP mode; therefore, is uses one single
sampling arm for the aspiration and dispensing of the sample and reagent. The arm is
equipped with two bipolar stepper motors and it can move vertically as well as rotating by 320° in
order to place itself on samples, internal and external reagents, on the readout cuvettes and on
the drain and washing sump. The arm is also provided with a capacitative level detector for
detecting reagents and samples.
The rotating movement is performed according to the number of steps from an initial position
detected
by an optical sensor and a metal disc with reference slit. The vertical movement of the arm is
performed
in the same manner, only in this case, the reference position is indicated by a flag on the conveyor
belt that interrupts the photo-coupler.
When the level detector does not detect any sample, the appliance emits a sound to warn
the operator that the sample is insufficient or missing.
The sample tray can operate in two different configurations. The tray contains 40 samples that
can be arranged in 3ml secondary cups or in 12x100 primary tubes. In both cases there are 10
further positions available for calibrators and control sera.
Thanks to a bipolar stepper motor, the tray rotates clockwise and anticlockwise by 360° and it moves for
a preset number of steps, starting from the reference position indicated by a photo-coupler and a
flag.
A barcode reader detects the samples arranged in the primary tubes. The tray can be removed
from its housing by loosening the two fixing knobs.
The reagent tray is provided with 30 positions for the insertion of as many plastic containers containing
the reagents necessary for running the tests. There are two different types of containers, the
single ones, which can contain 60 ml of reagent the double ones, with two separate
compartments, which can contain 17 ml and 40 ml respectively.
The reagent tray is provided with a cooling system, which is constantly in operation, even when
the appliance is off.
The stepper motor and the movement control system is the same as those used for the sample tray.
7
1 KENZA 240
General Description
The washing arm is a device located on the side of the readout module, which only performs a
vertical movement thanks to a stepper motor. The movement control is carried out via two
photo-couplers that detect the two extreme positions “high” and “low”.
On the head of the washing arm there are some tubes for introducing and sucking the washing
water,
as well as a teflon pad for drying the cuvette.
Most of the KENZA 240 electronic control system is located at the back of the appliance, while all the
robotics, temperature, conversion and digital signal and computer interface control boards are
located
on the upper part, beneath which is located the power supply unit. The micronotrollers that
are on some electronic boards can be identified by a yellow plate bearing the microcontroller
number and the software version.
Microcontroller 0 (PCB interface) communicates with all the other microcontroller in a
unidirectional manner, since it can require performing robotic movements or activating
solenoid valves and suction pumps. Only with micro 5 is there a bidirectional communication.
For an easy identification of these devices, please see the following table, which shows the
microcontrollers on each board and the associated command functions.
8
KENZA 240
1
Service
Manual
PCB MICROCONTROLLER
interface U22/micro 0
board unidirectional
920104 and bidirectio-
nal communi-
cation with the
other micro-
controllers
motor control U6/micro U7/micro 1 U12/micro 4 U13/micro 3 U14/micro 2
board 1 13 cuvette cuvette tray diluter + sole- sampling arm sampling arm
920098 wa- shing noid valve vertical horizontal
arm move- ment mo- vement
temperature
and level
de- tector
control
board
9
1 KENZA 240
General Description
This board is used for interfacing the instrument with an external management computer. The
instrument operates by means of three different managing software. The first and highest level is
represented by
the Windows management software, which includes the user interface. This software
communicates through the serial port with the interface board, which decodes the various
instructions via an interme- diate software and manages the motor and cycle control processes
of the instrument. The lowest level software is represented by the firmware configured in the
microcontrollers, which provide to the direct management of the electro-mechanical parts and
of each stepper motor. Micro 0 located on the Inter- face board directly controls some solenoid
valves, the suction pump, the drain sump and the vacuum pump.
Upon switching the instrument on, the management software receives all the settings
regarding the positioning of the robotic assemblies, as well as the various pre-set functions
stored in an EEprom of the interface board. The screen will briefly show the data reception and
then the main menu will appear. In case some of the settings have not been carried out, the
screen image below will appear, with the list of
all the instrument’s settings: those already made and those yet to be done.
10
KENZA 240
1
Service
Manual
This board allows amplifying the electric signal coming from the photodiode pre-amplifier located
in
the readout module and converting the analogue signal in a digital one, which will be processed
by the external management software.
The amplifier rest current gain level and offset level adjustments, as well as the filter carousel
motor ro- tation speed adjustment are carried out on this board. Moreover, this board also
houses some solenoid valve and discharge pump drivers.
There are two motor control boards onto which the microcontrollers and the stepper motor
drivers are installed. These boards use two PBL 3717 bipolar motor drivers for each motor. Each
board is provided with a quartz oscillator circuit for producing the clock signal, which is necessary
for a correct microcon- troller operation.
This board includes all the electronic circuits that control the temperature in the various sections of
the instrument.
The control system used is always based on the same circuit configuration i.e., an operational
amplifier with a reference threshold connected to a temperature detection device. The control
actuation systems,
on the other hand, vary according to the used heating or cooling
elements. The temperature control functions of this board are carried out
as follows:
• a circuit controls the temperature of the readout module, i.e., of the cuvettes; in this case, an electri-
cal resistance powered with 30V AC is also used as a heating element.
• another part of the circuit controls the cooling system of the reagent tray; in this case, a Peltier cell
cools a metal tray onto which the reagent vials are arranged.
Finally, part of the electronic system controls the reagent pre-heating. Also in this case, a Pertier cell, in
contact with a teflon tube on a round support, heats the reagent during the sampling phase.
This board also houses the level detector control circuit to check the presence of the reagent and
the sample and, in the case of reagents, to measure their volume.
11
1 KENZA 240
General Description
The power supply unit is located under the board rack and it is composed of two switching
power sup- plies and one conventional stabilised power supply. The latter provides the
necessary voltage for powe- ring the Photodiode Amplifier and ADC boards, as well as the 30V AC
for the readout module temperature control.
One of the two switching power supply, which is always in operation once the cable is connected,
pro- vides the +24V DC voltage for the operation of the cooling module and of some fans.
The other switching power supply, on the other hand, provides the +5V, +12V, -12V e +24V DC
voltages
for all the other parts of the instrument.
• a sampling system
• a washing/drainage system
The sampling system is based on a dilution module with a 1ml syringe powered by a bipolar stepped
mo-
tor, a solenoid valve installed onto the module itself and a pre-heating device installed on the
sampling arm.
This sampling system operates in SIP & DIP mode. The sampling arm places itself on the
required rea- gent and the sampling probe lowers itself and withdraws the established volume of
reagent. The probe then it rises up, while the arm rotates and places itself on the sample tray
in correspondence of the sample to be analysed. The probe lowers itself and withdraws the
established volume of the sample and then it rises up again.
Once this stage has been completed, the sampling arm rotates and places itself on the readout
carousel
in the dispensing position. The probe descends in the cuvette and dispenses the sample with
the rea- gent and then, to make the solution even, it aspirates part of the liquid again to dispense
it in the readout cuvette.
The pre-heater located on the sampling arm, raises the temperature of the reagent up to
approximately
35°C, thus allowing the sample-reagent mixture to quickly reach 37°C.
After each procedure, a cleaning system washes the external surface sampling probe with a
mixture of water and air, preventing any contamination of the sampling system.
12
KENZA 240
1
Service
Manual
• Internal washing of the sampling probe and recovery of the waste liquid dispensed in the drain
sump
during the washing cycles;
• control of the water flow in the probe washing device for the external washing of the sampling probe
with the help of a pair of solenoid valves and a membrane suction pump.
The two solenoid valves are connected in parallel and are inactive during the washing phase
and activated when emptying the drain sump.
The readout cuvettes are washed by means of a mechanical arm that, when lowered, introduces
some
probes in the cuvettes that fill them with water and then aspirate it. The filling phase is carried out throu-
gh a water suction and dispensing cycle, which involves the use of a syringe with a metal piston control-
led by a plexiglass support and a solenoid valve. The drainage phase is based on the use of a
membrane suction pump and a solenoid valve that selects the cuvette to be emptied. A specially
shaped teflon pad, connected to the vacuum pump, dries the cuvettes that have been washed.
13
1 KENZA 240
General Description
14
KENZA 240
1
Service
Manual
15
Maintenance 2
KENZA 240
2
Service
Manual
T his section describes the various preventive and corrective maintenance operations for the
optimal operation of the analyser. The table at the end of this chapter includes a list of spare
parts subject
to wear, which must be replaced periodically.
The diagnostic routines available to the maintenance technician allow carrying out the
following operations:
2.1. Cuvette CK
This function of the maintenance menu fills the cuvettes with distilled water first and then
checks each test-tube, reading the optical density (OD) with each interference filter.
This function is very useful to verify the condition of the cuvettes and, indirectly, the state of
the photometer and of the washing system, since it carries out a voltage control in relation to
the cuvette transmittance. This value is expressed in volts for cuvette 1, while for the remaining
49 cuvettes it is expressed as the voltage differential compared to the voltage obtained through
cuvette 1.
While performing the cuvette CK, close the cuvette compartment cover in order to prevent
any influence from external light.
19
2 KENZA 240
Maintenance
The relation with the obtained results will be displayed on screen; column 01 shows the absolute
vol-
tages read through cuvette 1, while columns 02-50 indicate the voltage differential of the
remaining cuvettes compared to cuvette 1.
The voltage values measured through cuvette 1 must be within 8.5V-3.5V. The difference
with the other cuvettes is usually below 0.100 V. For higher values, we recommend
washing the
cuvettes with the appropriate solutions.
The Cuvette CK function cannot be considered as an autozero function that will be later used
for calculating the results. It is only a function to verify the integrity of the readout cuvettes and of
the
entire measuring system. The autozero of each cuvette is always carried out during the
routine
before use.
20
KENZA 240
2
Service
Manual
Each time the readout cuvettes are replaced, we recommend performing at least 3 washing cycles
with Ipocleaning and Extracleaning solutions, in order to accurately clean the new cuvettes and
prevent the formation of air microbubbles that would affect the OD readout.
This function is considered as ordinary preventive maintenance and it allows washing the readout
cu- vettes. Please see Chapter 6 of the User Guide for detailed instructions.
This function allows carrying out a series of washing cycles using the dilution syringe.
We recommend carrying out a few Primes every day, before starting the routine, to fill the
hydraulic circuit completely.
Moreover, the Prime function can be useful for verifying the hydraulic sealing of the piston
also fol- lowing its replacement. The reason of the possible presence of air bubbles in the
syringe may be found by checking the elements composing the sampling circuit in the following
order:
syringe, sampling tube, pre-heater, dilutor solenoid valve, syringe plexigas
support. Please see Chapter 6 of the User Guide for detailed instructions.
You can access this function from the Service Menu by clicking on “Set Steps”. This function
allows carrying out a series of adjustments regarding the mechanical parts positioning. These
adjustments
are stored by the system and then used when counting the phases. These adjustments are
described in detail in the Chapter “Checks and Adjustments”.
21
2 KENZA 240
Maintenance
On the bottom right of the screen there are keys that allow activating the following functions:
The activation of the “Reagent detector” key allows measuring the volume of reagent inside the vial
and detecting the sample inside the cup. The subsequent activation of the “Automatic volume Ck”
verifies
the reagent volume before starting the routine. If the level detector is activated, it controls the
sample
and reagent levels during the patient routine and it emits a sound alarm in case these volumes
are insuf- ficient. The ”Volume Test” and “Volume Calibration” keys on the lower right of the
screen, are used to carry out the volume verification and the volume calibration, respectively.
Click on the key and follow the instructions that appear on the screen.
Do not use distilled water to calibrate the volume detector. Since it is a capacitative detector
and
the distilled water has a very low capacity, being poor in electrolytes, we recommend using
a
protein-based reagent.
22
KENZA 240
2
Service
Manual
The barcode reader is an optional accessory; therefore, its activation obviously depends on its
actual presence. The “Barcode Reader” key, located on the right of the “Set Step” menu, turns
green when enabled.
(Barcode Reader) Allows setting the barcode reader if present on the analyser. An
error message will appear if it is activated though not installed on the instrument.
By clicking on “Set Reader”, the menu for the barcode reader setup will appear.
Here you can select the type of barcode, according to the used label. For a correct operation, it is
im- portant to also set the Length ID and Position ID. Each code has its amount of digits, usually
more than
16, although KENZA actually uses only some of them. For this reason, it is important to indicate how
many digits of the entire barcode are actually used, as well as the position of the first digit read. In
short, the
ID length indicates the number of digits used to identify the samples, while the Position ID indicates
the position of the first digit used in the entire code.
23
2 KENZA 240
Maintenance
This table shows the maintenance interventions that need to be carried out periodically, according
to the frequency of replacement of the parts subject to wear.
Component
Pos. Operazioni di controllo, sostituzione, allineamento code 6 months 1 year 18 months
24
Replacing the 3
Components
KENZA 240 3
Service Manual
T he external cover needs to be removed to access the internal parts of the instrument, unless
other- wise specified.
When lifting the external cover, the safety switch cable located at the centre front of the
analyser must be disconnected.
27
3 KENZA 240
Replacing the components
28
KENZA 240 3
Service Manual
This operation should be carried out every six months, although this lapse of time can vary according
to the daily volume of work.
To replace the readout cuvettes, proceed as follows:
1. Remove the readout tray cover and move the sampling arm so that it
does not obstruct the removal of the cuvette carousel.
2. To remove the readout carousel, loosen the 4 fixing screws (red ar-
rows) in the centre and lift with caution. When removing, make
sure that the carousel does not come into contact with the
washing sy- stem probes.
3. Insert a new carousel with the relative cuvettes inside the container
and fix the four screws beck on. Make sure that each cuvette is
com- pletely lowered and fixed to the plastic wheel.
4. Make sure that the carousel can rotate freely before putting the cover
back on.
29
3 KENZA 240
Replacing the components
The cuvette drying pad must be replaced every 12 months depending on the planned ordinary
maintenance. Should the pad not remove the water from the cuvette completely, it could result
in the sample dilution which is the cause of inaccuracy and low result reproducibility.
30
KENZA 240 3
Service Manual
31
3 KENZA 240
Replacing the components
The sampling probe washing device is subject to wear; therefore, it should be replaced at least
every six months according to the planned preventive maintenance. In case of a high number of
samples analysed daily, this lapse of time can be even lower. A poor sealing of the washing
device gasket causes an unsu- itable washing of the sample probe, which could lead to
contamination of the samples with an alteration
of the results and a possible loss of liquid during the sampling operations as a consequence.
1. use the hexagonal key to loosen the washing device grub screw.
2. Disconnect the two tubes from the probe washing device.
3. Connect the tubes to the new probe washing device.
4. Insert a new probe washing device from the bottom, making sure not to bend the sampling
probe.
The device can slightly force the probe, since it has a little O’ring inside.
5. Tighten the grub screw with a hexagonal key.
6. Carry out a few probe washing cycles through the special function in the Service Menu.
32
KENZA 240 3
Service Manual
The Photodiode Pre-amplifier board is located inside the readout carousel in correspondence with
the optical fibre. Usually, it does not need to be replaced; however, a leak from the cuvettes could
damage
the circuit. Since the replacement is quite difficult, to make sure that the board is acually faultes,
carry out the following verifications:
Carry out an ADC readout and if the voltages appear unstable or equal to
0V, check if the problem depends on the ADC system.
This verification can be made by connective the positive pole of a
1.5V standard battery to the blue cable and the negative pole to the
brown ca- ble of the connector shown in the figure.
This connector is part of the cabling and connects the amplifier photodio-
de board to the ADC board. It is located in the lower compartment
under the readout unit (red arrow).
The total absence or the instability of the signal despite the
insertion of the battery indicates a malfunction of the ADC board.
If, on the other hand, the fault is due to the photodiode pre-
amplifier board, the filter readouts would have all equal value and
be stable; therefore, the fault that was pointed out would disappear.
33
3 KENZA 240
Replacing the components
1. Turn the amplifier off, disconnect the power supply cable and remove
the external cover.
2. Remove the optical fibre from its housing by loosening the fixing grub
1
screw on the cylindrical support (fig.1).
3. Remove the four screws that fix the readout unit to the metal base of
the instrument (fig.2).
4. Lift the unit and disconnect the ground connection cable together with
the photocoupler connectors of the Amplifier Photodiode board and of
the heating element (fig.2). 2
5. Remove the eight screws that fix the cuvette tray support to the unit
and remove both the cuvettes and the support (fig.3).
6. Remove the collar and the cylindrical support of the optical fibre
(fig.4).
3
7. Remove the three screws that fix the external band to the base of the
readout unit and remove the band (fig.5).
8. Remove the three fixing screws and carefully lift the heating ring. For
an easier movement, remove the cabling fixing bands (fig.6).
9. Remove the two fixing screws and remove the lphotodiode pre-ampli-
fier board. (fig.7) 4
10. Insert the photodiode of the new board in its housing, making sure that
the surface is clean. If necessary, use a soft cloth to clean the lenses
(fig.7).
11. Reassemble the readout module, repeating the previous actions in re-
verse order.
5
12. Carry out the checks and adjustments provided for following this repla-
cements. For this purpose, please refer to the Checks & Adjustments
flow chart in Chapter 4.
34
KENZA 240 3
Service Manual
1. Turn the amplifier off and disconnect the power supply cable
1. Remove the external cover
2. Remove the 4 screws that fix the photometer unit to the frame
1
3. Remove the two optical sensors (red arrows fig.1)
4. Disconnect the connector from the filter wheel motor, remove the lamp
connection terminal board and the ground connection (yellow arrows
fig.1)
5. Once free, lay the photometer on the table 2
6. Remove the 3 screws that fix the filter wheel motor to the unit
(fig.2)
7. Remove the 4 screws that fix the optical unit to the filters (fig.3)
8. Remove the fan and the filter wheel cover (fig.3)
9. Remove the filter wheel fixing screw by locking the rotation axis 3
with a screwdriver (fig.4)
10. Extract the filter wheel, the rotation axis, the two bearings and the
spacer in this order (fig.5)
11. Replace the faulted bearings with new ones of the same type and reas- 4
semble the photometer by carrying out the same operations in
reverse order.
12. With the replacement of the bearings, we recommend checking
and, if necessary, replacing the filter wheel motor.
5
13. Once the bearings have been replaced, carry out the checks and
adjustments as in paragraph 4.13
35
3 KENZA 240
Replacing the components
Remove the external cover of the analyser to access the components located under the worktop.
Remove the fixing screws and lift the worktop fixing it with its special bracket.
36
Checks and 4
Adjustments
KENZA 240 4
Service
Manual
R emove the external cover to access the internal parts of the analyser.
Check the power supply voltage values and stability before carrying out any adjustment.
If necessary, use the extension boards p/n 920085 to carry out electronic adjustments and tests.
1. Motor driver 1
2. Motor driver 2
3. Interface board (between an external PC and the analyser)
4. ADC board
5. Temperature and level detector control board
To access the power supply compartment, disconnect the flat cable (red
arrow) from the interface board and remove the worktop fixing screws.
39
4 KENZA 240
Checks and adjustments
Checks and alignment flow chart to be carried out following removal or replacement of a
component or
a unit.
Cuvette tray Wash w ith Carry out ADC Filt ers out of range?
replacement Adjust Gain
Ipo+Extra readout par. 4.14
par. 3.3
Syringe
Carry out some
replacement
Prime cycles
par. 3.5
40
KENZA 240 4
Service
Manual
Connect a voltmeter to the ground frame and the points indicated in the picture, making sure that
the measured voltages coincide with those in table.
Position Voltage
1 Dddd
Ground
2 + 5 V ± 0.2V
3 + 12 V ± 0.2V
4 - 12 V ± 0.2V
5 + 24 V ± 0.3V
To adjust the lamp voltage, connect the two voltmeter probes in parallel
to the lamp cables. Adjust the voltage by turning the
potentiometer shown in the picture.
All adjustments must comply with the specifications contained in the following table:
41
4 KENZA 240
Checks and adjustments
This adjustment determines the reference voltage of the stepper motors used to move the
mechanical units.
To check the reference voltage, connect the digital voltmeter between ground and the voltage
regulator heatsink.
Whenever enabled, the level detector function is active during both the sample and the
reagent aspiration. The control circuit of this function is located in part in the reagent pre-heater
and in part in
the temperature board.
All checks and adjustments must be carried out according to the specifications contained in the
table below:
42
KENZA 240 4
Service
Manual
The type of alignment to be performed will appear on top left of the display.
The two keys at the bottom, indicated with + and – allow moving the mechanical unit one step at a
time. Once the alignment has been completed and stored via the “Save” key, you can proceed
with the next alignment.
This menu allows you to perform two further alignments: the height of the sampling probe inside
the sample cup and the probe stop position following sample aspiration. Two keys indicated as
“Ck Up”
and “Ck Down” move the sampling probe in the required position to verify the alignments. The
numbers inside the boxes represent the number of steps the motors have to make to reach
the set positions. These numbers are to be increased or decreased according to the
specifications.
The distance between the base of the probe washing device and the tip of
the sampling probe must be ~ 1mm (Tolerance +0.5mm ; -0mm)
The distance A between the base of the sample cup and the tip of the
sampling probe must be ~ 2mm (Tolerance +0.5mm ; -0mm)
43
4 KENZA 240
Checks and adjustments
1. Disconnect the power cord and remove the external cover of the in-
strument; lift the worktop and fix it with its special support (see fig. a)
2. Connect the power cord and turn the instrument on.
3. Perform a mechanical reset of the instrument.
The following picture shows a detail of the sampling arm unit
for performing a vertical movement (red arrow). When in
idle state, the position of the probe in the probe washing
device is determi- ned by the position of the flag on the
transmission belt.
There are 24V in the board containing the optical detector and
the robotic cable; therefore, we recommend turning the
instrument off
during alignment operations to prevent short circuits.
4. Loosen the two locking screws and move the flag up or down to lift or
to lower the sampling probe.
5. Tighten the flag screws and carry out a mechanical reset.
6. Make sure that the position of the probe in the probe washing device
meets the specifications shown in the figure. If not, repeat the
align- ment procedure.
7. Lower the instrument worktop and mount the external cover back on.
The top right part of the mechanical set-up menu allows aligning the
cu- vette in front of the light beam. Should the number of steps inside
the box
be modified, the cuvette position will also change. The alignment consists
in the centering of the cuvette in front of the light beam produced by
the photometer and brought into place through the optical fibre (red
arrow). The two screws used to fix the fibre and its support (yellow
arrows) can
be taken as a reference, or you can carry out a direct visual check of
the cuvette through the optical fibre duct.
44
KENZA 240 4
Service
Manual
Access the service menu; then, to set the correct washing arm position, proceed as follows:
1. Remove the washing arm cover and loosen fixing screws of the probe-
pad support (red arrows)and those of the pad vertical movement
(yel- low arrows).
2. Move the pad so that it is positioned slightly higher than the probes.
3. Carry out a mechanical reset to reposition the cuvette tray.
4. Lower the washing arm manually and move the probe support so that
the pad enters in the cuvette. Centre the four sides of the pad
inside the cuvette, considering that the pad perimeter is slightly
less than that of the cuvette.
5. Centre the two probes in the respective cuvettes and tighten the two
support fixing screws (yellow arrows).
6. Carry out a mechanical reset.
7. Lower the washing arm using the instrument diagnostic system and
make sure the two probes are slightly lifted.
8. Without moving the arm, lower the pad until it touches the bottom of
the cuvette and then tighten the two locking screws.
9. Carry out a mechanical reset and then lower and lift the washing arm
several times, verifying the following operative conditions:
a. No contact between pad and cuvette
b. The probes are centred in their respective cuvettes
c. The probes are lifted when the arm is lowered
10. Carry out a cuvette washing cycle with water; then remove and check
the cuvette tray to verify the succesful alignment. The cuvettes
must be dry, with no water drops on their walls.
11. Put the washing arm cover back on.
45
4 KENZA 240
Checks and adjustments
The ADC board controls the correct photometer operation, as well as the signal conversion in
digital format via micro 5. Micro 0 of the interface board communicates in a bidirectional manner with
micro 5
of the ADC board.
1. Micro 5
2. Gain
3. Offset
4. -15 V
5. +15 V
6. Filt. Mot. Speed
7. TP+15 V
8. TP-15 V
This adjustment regards the two voltages that power the photodiode pre-amplification circuit.
These
two AC voltages supplied by the toroidal transformer and rectified in the lamp power supply
board,
are adjusted in the ADC board. Adjust P6 and P8 according to the specifications contained in the
table below.
46
KENZA 240 4
Service
Manual
IC U3 Pin 1 / TP 40 ms ± 0.1 ms
Pcb 920096 P10
RESET-R-FilterS
47
4 KENZA 240
Checks and adjustments
*Gain must be set taking the filter with the highest voltage as a reference
48
KENZA 240 4
Service
Manual
To set the cuvette tray temperature, add 500 µl of distilled water to some equidistant readout
cuvettes
and cover the tray with its lid. To guarantee an even heating of the cuvette tray, select the main
menu (this way the cuvette tray starts turning and the heat is better distributed over time). After 30
minutes, access the service menu to stop the cuvette tray and submerge the temperature
probe inside each cuvette. Adjust the temperature according to the specifications contained in
the following table.
Cuvette N.10
Cuvette N.20
920102 P3 Cuvette N.30 37°C + 0.5 – 0 °C
Cuvette N.40
Cuvette N.50
49
4 KENZA 240
Checks and adjustments
First of all, it is important to stress that the KENZA reagent cooling system cannot be
considered as a true refrigerator, but simply as a device to reduce the reagent temperature
approximately 10 degrees below the ambient temperature; therefore, for a better reagent
preservation, we recommend installing
the instrument in a conditioned environment. To carry out temperature checks and adjustments,
proceed
as follows:
1. Remove the sample and the reagent trays from their housing
2. Pour a few drops of water onto the metal tray in the area delimited by the 4 screws (the
Peltier
element is located underneath), leave to cool for a couple of minutes and then submerge the
tip of the temperature-measuring probe and read the temperature.
3. Adjust the temperature according to the specifications contained in the table below
To allow the reagent-sample solution to quickly reach the incubation temperature of 37 ° C, during
the sampling phase the reagent is heated at a temperature of approximately 35 °C. This check can
be carried
out by simulating a routine analysis on some samples (distilled water can be used as a
reagent). To simulate the right operative conditions, it is important that the temperature of the
used reagent is the same as that of the reagents left to cool a whole night.
Simulate a routine analysis on a few samples and then, immediately after the sampling arm has
dispensed water in the readout cuvettes, dip the temperature probe in the cuvette and check
the temperature. Adjust P6 according to the specifications contained in the table below.
50
KENZA 240 4
Service
Manual
In case of faults, use the instrument diagnostic system and the following table to identify the faulty
elec- tronic board or device.
51
4 KENZA 240
Checks and adjustments
52
KENZA 240 4
Service
Manual
53
4 KENZA 240
Checks and adjustments
At times, some letters (flags) may appear next to the results, to communicate important details
concer- ning the test being run to the operator. Here is a list of flags and their meaning.
It means that the result is below the set reference va- lue lower
LO Low value limit
It means that the result is above the set reference va- lue
HI High value higher limit
It means that the result is above the set linearity limit – We
FL Out of linearity limit recommend repeating with dilution
In case of tests with multistandard calibrations, it me- ans
FC that the result is above the maximum or below the minimum
Out of calibration curve calibration point
<B Below blank It means that the result is below the reagent blank
54
KENZA 240 4
Service
Manual
Micro 0 constantly communicates in a unidirectional and bidirectional (micro 5) manner with the
other microcontrollers. Should a robotic movement fail following a mechanical, electronical or
detection pro- blem, a message with the indication of the microcontroller and the type of
movement will be displayed.
55
Spare Parts 5
KENZA 240 5
Service
Manual
5.1. Spare parts
59
5 KENZA 240
Spare Parts
60
KENZA 240 5
Service
Manual
61
5 KENZA 240
Spare Parts
62
KENZA 240 5
Service
Manual
63
5 KENZA 240
Spare Parts
64
KENZA 240 5
Service
Manual
65
5 KENZA 240
Spare Parts
66
KENZA 240 5
Service
Manual
67
5 KENZA 240
Spare Parts
68
KENZA 240 5
Service
Manual
69
5 KENZA 240
Spare Parts
70
KENZA 240 5
5.2. Hydraulic system Service Manual
71
Communication 6
Protocol
KENZA 240
6
Service Manual
S ince KENZA 240 needs to be interconnected to the laboratory management system, the
computer requires a further COM 2 serial port as COM 1 is reservesd to the connection with the
analyser. The serial connection cable is made according to the diagram below:
2 3 Recept. /Transmiss. 3 2
5 7 Bas 5 7
e
Datum Baud Rate Nr. of bits Stop bit Parity bit Handshakin Input Len
g
Val 19200 8 1 no 0 1
.
BCC ETX
Data strings are composed of printable ASCII digits only, whose codes range between ASCII 32
and
255.
BCC consits of 2 ASCII digits that represent the sum (module 256) of all the digits in the data string
(con- sidered in low case), except for STX, BCC and ETX digits. The sum must be expressed in
hexadecimal notation (HEX).
The BCC datum is represented by the last 2 ASCII low case digits on the right of the string
obtained by linking the digits “0” to the value Hex(i), where i is the sum (module 256) of all the
ASCII values of the data string digits (except for STX, BCC and ETX digits). E.g.:
Hex(i) =
AB
“0” & AB => “0AB”
BCC = LCase [ Right (“0AB”,2)] = “ab”
75
6 KENZA 240
Communication protocol
6.1.3. Protocol
Should the management computer want to send data to the Analyser, it must send an ENQ. If the
Analy-
ser is in Reception mode, it will respond with an ACK, then the Management computer can start
sending the data strings.
Should the Analyser want to send the result string to the Management computer, first of all it must
send
an ENQ to establish the serial connection. The management computer should respond with an
ACK. If no
ACK is received, the analyser sends an ENQ every 10 sec, until time-out.
Here is the transmission mode communication diagram (Man. comp. Analyser):
76
KENZA 240
6
Service Manual
77
6 KENZA 240
Communication protocol
Here below are the details regarding the string from and towards the Analyser, indicating the name
and the length of the fields that compose them.
No separators are used in the data strings.
There are two cases according to the length of the Patient ID field:
78
KENZA 240
6
Service Manual
NOTE: No separation digit is provided for between the fields composing the strings. The
result
is coded in the following format: xxxx.xxxx (9 digits including comma/period).
79
6 KENZA 240
Communication protocol
80
KENZA 240
6
Service Manual
Spare parts 7
List
81
MODULE Code ENGLISH / ANGLAIS FRENCH / Français
READING 000179 60°C Thermostat (for cuvette tray) Thermostat 60°C (pour portoir cuvettes)
MA000084 Cuvettes reading assembly Bloc de lecture des cuvettes (assemblage)
13-0027 Cuvettes tray optical sensor Senseur optique portoir cuvettes
LECTURE M000409 Filter locking ring anneau de blocage filtre
13-0004 Filter wheel motor assembly moteur de roue à filtre (assemblage)
000966 Filter wheel’s bearing Roulement roue à filtre
000116 Interference filter 340nm Filtre interférentiel 340nm
OEM000725 Interference filter 380nm Filtre interférentiel 380nm
000117 Interference filter 405nm Filtre interférentiel 405nm
000900 Interference filter 450nm Filtre interférentiel 450nm
000119 Interference filter 505nm Filtre interférentiel 505nm
000120 Interference filter 546nm Filtre interférentiel 546nm
000121 Interference filter 570nm Filtre interférentiel 570nm
000128 Interference filter 620nm Filtre interférentiel 620nm
OEM000124 Interference filter 700nm Filtre interférentiel 700nm
13-0003 Lamp assembly Lampe (assemblage)
000998 Optical fiber Fibre optique
13-0022 Photometer assembly Photometre (assemblage)
9-0033 Pre-filter attenuators set Set d'atténuateurs de filtres
13-0055 Tray with 50 cuvettes Portoir avec 50 cuvettes
MAINTENANCE MKIT1-240 Maintenance Kit 1 (6 months) for Kenza Kit de Maintenance (6 mois) pour Kenza 240CUV50-
240CUV50-240PCLEAN-240SPROB-240 240PCLEAN-240SPROB-240
KITS
KITS DE MKIT2-240 Maintenance Kit 2 (12 months) for Kenza 240 Kit de Maintenance 2 (12 mois) pour Kenza 240
LAMP-240 LAMP-240
MAINTENANCE
SYGLAS-240 SYGLAS-240
SYPIS-240 SYPIS-240
WSDRY-240 WSDRY-240