ALARM
ALARM TEXT DEVICE POSSIBLE CAUSES REMEDY
DESCRIPTION
1- General power supply
interrupted 1- Power supply restoration
Indicate a power failure
1 POWER FAILURE
at the restoration
6A1 2-
3-
Fuse 230V burnt
The voltage supply is lower than
2-
3-
Replace the fuse 2A 5X20
Control the voltage and current value
the voltage needed
15B3
10YV2 1- Water supply interrupted 1- Verify if the water tap is open and if
2- Water filling solenoid valve there is no interruption on the loading
10YV2 defective or not pipe
The cold water filling is connected 2- Check and eventually replace the
11 NO COLD WATER not complete after the
time set by P6.32
3-
4-
Time set by P6.32 too low
Pressure switch 14B2 defective 3-
defective solenoid valve
Increase the diagnostics time P6.32
5- Solenoid valve connection NOT 4- Check and eventually replace the
OK defective pressure switch
14B2
6- Pressure switch connection NOT 5- Check output O5 of board 6A1 and
This document is a property of STEELCO S.p.A., it is a relevant value good and it can’t be transmitted or divulged to third party without a specific written
authorization of STEELCO direction.
1/9
OK relative wire 10.5
6- Check input I18 and relative wire
14.18
10YV1
1- Water supply interrupted 1- Verify if the water tap is open and if
2- Water filling solenoid valve there is no interruption on the loading
10YV2 defective or not pipe
connected 2- Check and eventually replace the
3- Time set by P6.32 too low defective solenoid valve
The hot water filling is
3- Increase the diagnostics time P6.32
12 NO HOT WATER not complete after the
time set by P6.32 14B2
4-
5-
Pressure switch 14B2 defective
Solenoid valve connection NOT 4- Check and eventually replace the
OK defective pressure switch
6- Pressure switch connection NOT 5- Check output O4 of board 6A1 and
OK relative wire 10.4
6- Check input I18 and relative wire
14.18
10YV3
1- Verify if the water tap is open and if
1- Water supply interrupted there is no interruption on the loading
2- Water filling solenoid valve pipe
10YV3 defective or not 2- Check and eventually replace the
The demineralised water connected defective solenoid valve
filling is not complete 3- Time set by P6.32 too low 3- Increase the diagnostics time P6.32
13 NO DEMI WATER
after the time set by 12B2 4- Pressure switch 12B2 defective 4- Check and eventually replace the
P6.32 5- Solenoid valve connection NOT defective pressure switch
OK 5- Check output O6 of board 6A1 and
6- Pressure switch connection NOT relative wire 10.6
OK 6- Check input I18 and relative wire
12.1
10YV2
10YV2
10YV1 7- Water supply interrupted 7- Verify if the water tap is open and if
8- Water filling solenoid valve there is no interruption on the loading
The level of the
10YV3 or 10YV1 defective or not pipe
hot+demi water filling
NO HOT+DEMI connected 8- Check and eventually replace the
16 WATER
has not be reached after
the time set by P6.34
9- Time set by P6.32 too low defective solenoid valve
10- Pressure switch 14B2 defective 9- Increase the diagnostics time P6.32
(level control)
11- Solenoid valve connection NOT 10- Check and eventually replace the
10YV3 OK defective pressure switch
This document is a property of STEELCO S.p.A., it is a relevant value good and it can’t be transmitted or divulged to third party without a specific written
authorization of STEELCO direction.
2/9
12- Pressure switch connection NOT 11- Check output O6 or O4 of board 6A1
OK and relative wire 10.6 or 10.4
12- Check input I18 and relative wire
14.18
14B2
J1
The chemical product
associated to dosing
pump 1 is used-up (if it is
1- Replace the chemical tank
set as an alarm by 1- Chemical product used-up
2- Check and eventually replace the
P3.27) 2- Dosing pump 11M1 defective
defective dosing pump
Diagnostics with dosing 11M1 3- Pressure switch connection NOT
3- Check output O11 and relative wire
pump enabled: OK
17 NO CHEMICAL 1
- Pressure switch
4-
11.11
Check and eventually replace the
state with pressure With flowmeter control:
defective flowmeter
switch presence; 4- Flowmeter J1 defective
5- Increase the time P6.07
- Lack of a new 5- Time set by P6.07 too low
6- Check input J1
impulse after the 6- Flowmeter connection NOT OK
time set by P6.07 if
a flowmeter is used
J2
The chemical product
associated to dosing
pump 2 is used-up (if it is
1- Chemical product used-up 1- Replace the chemical tank
set as an alarm by
2- Dosing pump 12M1 defective 2- Check and eventually replace the
P3.27)
3- Pressure switch connection NOT defective dosing pump
Diagnostics with dosing 12M1 OK 3- Check output O12 and relative wire
pump enabled:
18 NO CHEMICAL 2
- Pressure switch
With flowmeter control: 4-
12.12
Check and eventually replace the
state with pressure
4- Flowmeter J2 defective defective flowmeter
switch presence;
5- Time set by P6.07 too low 5- Increase the time P6.07
- Lack of a new
6- Flowmeter connection NOT OK 6- Check input J2
impulse after the
time set by P6.07 if
a flowmeter is used
J4
The chemical product
associated to dosing
pump 4 is used-up (if it is
1- Replace the chemical tank
set as an alarm by 1- Chemical product used-up
2- Check and eventually replace the
P3.27) 2- Dosing pump 26M1 defective
defective dosing pump
Diagnostics with dosing 26M1 3- Pressure switch connection NOT
3- Check output O6 and relative wire
pump enabled: OK
19 NO CHEMICAL 4
- Pressure switch
4-
26.6
Check and eventually replace the
state with pressure With flowmeter control:
defective flowmeter
switch presence; 4- Flowmeter J4 defective
5- Increase the time P6.07
- Lack of a new 5- Time set by P6.07 too low
6- Check input J4
impulse after the 6- Flowmeter connection NOT OK
time set by P6.07 if
a flowmeter is used
J3
The chemical product
associated to dosing
pump 3 is used-up (if it is
1- Chemical product used-up 1- Replace the chemical tank
set as an alarm by
2- Dosing pump 12M3 defective 2- Check and eventually replace the
P3.27)
3- Pressure switch connection NOT defective dosing pump
Diagnostics with dosing 12M3 OK 3- Check output O14 and relative wire
pump enabled:
20 NO CHEMICAL 3
- Pressure switch
With flowmeter control: 4-
12.14
Check and eventually replace the
state with pressure
4- Flowmeter J3 defective defective flowmeter
switch presence;
5- Time set by P6.07 too low 5- Increase the time P6.07
- Lack of a new
1- Flowmeter connection NOT OK 6- Check input J3
impulse after the
time set by P6.07 if
a flowmeter is used
This document is a property of STEELCO S.p.A., it is a relevant value good and it can’t be transmitted or divulged to third party without a specific written
authorization of STEELCO direction.
3/9
1EH1
20B1
This document is a property of STEELCO S.p.A., it is a relevant value good and it can’t be transmitted or divulged to third party without a specific written
authorization of STEELCO direction.
4/9
1- Check and eventually replace the
1- Probe 20B1 defective
20B1 defective probe
2- Heating element connection
The air temperature is 2- Check input I35-I36
NOT OK
DRYING PROBE over 162°C 3- Check and eventually replace the
28 LIMIT °C (= max limit 150°C +
3- Air heating element contactor
stuck
defective contactor
emergency 12°C) 4- Check output O8 of board 6A2 and
4- Contactor connection NOT OK
relative wire 28.8
5- Probe off set 20B1 incorrect
5- Set the probe by P4.21
1- Check and eventually replace the
1- Probe 31B2 defective
31B2 defective probe
2- Heating element connection
The boiler temperature is 2- Check input I19-I20
NOT OK
BOILER PROBE over 100°C 3- Check and eventually replace the
29 LIMIT °C (= max limit 80°C +
3- Boiler heating element contactor
stuck
defective contactor
emergency 20°C) 4- Check output O11 of board 6A2 and
4- Contactor connection NOT OK
relative wire 28.2
5- Probe off set 31B2 incorrect
5- Set the probe by P4.21
19B1
Tank temperature probe 1- Probe 19B1 defective 1- Replace the defective probe
30 TANK PROBE
failure (working probe) 2- Probe connection NOT OK 2- Check input I32-I33
31B1
20B1
Air temperature probe 1- Probe 20B1 defective 4- Replace the defective probe
32 DRYING PROBE
failure 2- Probe connection NOT OK 1- Check input I35-I36
31B2
Boiler temperature probe 3- Probe 31B2 defective 5- Replace the defective probe
33 BOILER PROBE
failure 4- Probe connection NOT OK 2- Check input I19-I20
6A1 e 6A2
This document is a property of STEELCO S.p.A., it is a relevant value good and it can’t be transmitted or divulged to third party without a specific written
authorization of STEELCO direction.
5/9
19B1
16B1 - 16B2
Max steam condenser 1- Level switch 16B2 and 16B1 1- Replace the defective level switch
LEVEL STEAM level enabled and min defective 2- Check input I26 and I25 and relative
44 CONDENSER steam condenser level 2- Level switch connection NOT wire I16.25 of board 6A2 and relative
disabled OK wire 25.2
This document is a property of STEELCO S.p.A., it is a relevant value good and it can’t be transmitted or divulged to third party without a specific written
authorization of STEELCO direction.
6/9
12M1
11M2
11M2
Device 9R1 1- Conductivity limit parameter not 1- Fix parameter within treatment phase
correct parameter
2- Max. number of attempts 2- Fix parameter MAIN
parameter not correct PAGE -> SETUP -> SYSTEM DATA
Conductivity value 3- Conductivity transmitter 15A3 -> "Nr Attm.Contr. Conduc.""
56 CONDUCTIVITY superior to the value set
by P7.23
Device 9B7
4-
not calibrated
Conductivity transmitter
3- Calibrate conductivity transmitter
15A3
15A3 defective 4- Replace conductivity transmitter
5- Conductivity sensor defective or 15A3
exhausted 15R1 5- Replace conductivity sensor 15R1
This document is a property of STEELCO S.p.A., it is a relevant value good and it can’t be transmitted or divulged to third party without a specific written
authorization of STEELCO direction.
7/9
59 -------------------------- -------------------------------- ----------------------- ---------------------------------- ----------------------------------
WARNING
WARNING TEXT DEVICE POSSIBLE CAUSES REMEDY
DESCRIPTION
It is possible to start a
cycle during a stand-by
state.
With a double door
PRESS START
machine is necessary to / / /
open and close the door
one time after the end of a
cycle
The chemical product
associated to dosing
pump 1 is used-up (if it is 14B3
set as a warning by 1- Chemical 1 used up
P3.27) 2- Level sensor 14B3 chemical 1
1- Replace the chemical tank 1
Diagnostics with dosing dirty or blocked
2- Clean the level sensor chemical 1
NO CHEMICAL 1 pump enabled: 3- Level sensor 14B3 chemical 1
3- Replace the level sensor chemical 1
- Pressure switch state defective
4- Check input I20 and relative wire 14.20
with pressure switch 4- Level sensor connection NOT
presence; OK
Lack of a new impulse
after the time set by P6.07
if a flowmeter is used
The chemical product
associated to dosing
pump 2 is used-up (if it is 12B1
set as a warning by 1- Chemical 2 used up
P3.27) 2- Level sensor 14B1 chemical 2
1- Replace the chemical tank 2
Diagnostics with dosing dirty or blocked
2- Clean the level sensor chemical 2
NO CHEMICAL 2 pump enabled: 3- Level sensor 14B1 chemical 2
3- Replace the level sensor chemical 2
- Pressure switch state defective
4- Check input I17 and relative wire 14.17
with pressure switch 4- Level sensor connection NOT
presence; OK
Lack of a new impulse
after the time set by P6.07
if a flowmeter is used
This document is a property of STEELCO S.p.A., it is a relevant value good and it can’t be transmitted or divulged to third party without a specific written
authorization of STEELCO direction.
8/9
WARNING
WARNING TEXT DEVICE POSSIBLE CAUSES REMEDY
DESCRIPTION
The chemical product
associated to dosing
pump 4 is used-up (if it is 15B2
set as a warning by 1- Chemical 4 used up
P3.27) 2- Level sensor 15B2 chemical 4
1- Replace the chemical tank 4
Diagnostics with dosing dirty or blocked
2- Clean the level sensor chemical 4
NO CHEMICAL 4 pump enabled: 3- Level sensor 15B2 chemical 4
3- Replace the level sensor chemical 4
- Pressure switch state defective
4- Check input I24 and relative wire 15.24
with pressure switch 4- Level sensor connection NOT
presence; OK
Lack of a new impulse
after the time set by P6.07
if a flowmeter is used
The chemical product
associated to dosing
pump 3 is used-up (if it is 15B1
set as a warning by 1- Chemical 3 used up
P3.27) 2- Level sensor 15B1 chemical 3
1- Replace the chemical tank 3
Diagnostics with dosing dirty or blocked
2- Clean the level sensor chemical 3
NO CHEMICAL 3 pump enabled: 3- Level sensor 15B1 chemical 3
3- Replace the level sensor chemical 3
- Pressure switch state defective
4- Check input I21 and relative wire 15.21
with pressure switch 1- Level sensor connection NOT
presence; OK
Lack of a new impulse
after the time set by P6.07
if a flowmeter is used
Make a salt refill after the
cycle number set by
SALT LOADING
P7.28 (with P7.27 / / /
different form 10)
Cycle successful
OK
completed / / /
The cycle has been The cycle was interrupt before the Restart a cycle form the beginning
NO DISINFECTION
interrupted thermal disinfection process end
This document is a property of STEELCO S.p.A., it is a relevant value good and it can’t be transmitted or divulged to third party without a specific written
authorization of STEELCO direction.
9/9