Alarm list for DS 500 model Rev.

00

ALARM
ALARM TEXT DEVICE POSSIBLE CAUSES REMEDY
DESCRIPTION
1- General power supply
interrupted 1- Power supply restoration
Indicate a power failure
1 POWER FAILURE
at the restoration
6A1 2-
3-
Fuse 230V burnt
The voltage supply is lower than
2-
3-
Replace the fuse 2A 5X20
Control the voltage and current value
the voltage needed
15B3

1- Check and eventually replace the


The loading door is open 1- Door sensor 15B3 stuck or
LOADING DOOR defective sensor
2 OPEN
or/and unlocked during a
cycle
15B4
2-
defective
Sensor connection NOT OK
2- Check input I22 and relative wire
15.22

3 -------------------------- -------------------------------- ----------------------- ---------------------------------- ----------------------------------


15B3

1- Check and eventually replace the


1- Locking door micro switch 15B3 defective micro switch
15B4 stuck or defective 2- Check and eventually replace the
2- Door sensor 15B3 stuck or defective sensor
defective 3- Check input I23 and relative wire
LOADING DOOR The loading door is open 3- Micro switch connection NOT 15.23
4 FAILURE and locked (incongruity) OK 4- Check input I22 and relative wire
4- Sensor connection NOT OK 15.22
5- Gear motor 12M2 stuck or 5- Check and eventually replace the
defective defective gear motor
12M2
6- Gear motor connection NOT OK 6- Check output O13 of board 6A1 and
relative wire 12.13

5 -------------------------- -------------------------------- ----------------------- ---------------------------------- ----------------------------------

6 -------------------------- -------------------------------- ----------------------- ---------------------------------- ----------------------------------


15B3
1- Time set by P6.05 too low 1- Increase the diagnostics time P6.05
2- The door was open during the 2- Close door
locking door process 3- Check and eventually replace the
3- Unlocking door micro switch defective micro switch
The loading door is not
UNBLOCKED 15B3 stuck or defective 4- Check and eventually replace the
7 DOOR 1
locked after the start of
the cycle
12M2 4- Gear motor 12M2 stuck or defective gear motor
defective 5- Check input I22 and relative wire
5- Micro switch connection NOT 15.22
OK 6- Check output O13 of board 6A1 and
6- Gear motor connection NOT OK relative wire 12.13

8 -------------------------- -------------------------------- ----------------------- ---------------------------------- ----------------------------------


15B3
1- Increase the diagnostics time P6.14
1- Time set by P6.14 too low
2- Check and eventually replace the
2- Locking door micro switch 15B3
defective micro switch
The door locking stuck or defective
3- Check and eventually replace the
UNLOCKING opening of the loading 3- Gear motor 12M2 stuck or
9 DOOR 1 FAILURE side is not complete after 12M2 defective
4-
defective gear motor
Check input I22 and relative wire
the time set by P6.14 4- Micro switch connection NOT
15.22
OK
5- Check output O13 of board 6A1 and
5- Gear motor connection NOT OK
relative wire 12.13

10 -------------------------- -------------------------------- ----------------------- ---------------------------------- ----------------------------------

10YV2 1- Water supply interrupted 1- Verify if the water tap is open and if
2- Water filling solenoid valve there is no interruption on the loading
10YV2 defective or not pipe
The cold water filling is connected 2- Check and eventually replace the
11 NO COLD WATER not complete after the
time set by P6.32
3-
4-
Time set by P6.32 too low
Pressure switch 14B2 defective 3-
defective solenoid valve
Increase the diagnostics time P6.32
5- Solenoid valve connection NOT 4- Check and eventually replace the
OK defective pressure switch
14B2
6- Pressure switch connection NOT 5- Check output O5 of board 6A1 and

This document is a property of STEELCO S.p.A., it is a relevant value good and it can’t be transmitted or divulged to third party without a specific written
authorization of STEELCO direction.

1/9
OK relative wire 10.5
6- Check input I18 and relative wire
14.18

10YV1
1- Water supply interrupted 1- Verify if the water tap is open and if
2- Water filling solenoid valve there is no interruption on the loading
10YV2 defective or not pipe
connected 2- Check and eventually replace the
3- Time set by P6.32 too low defective solenoid valve
The hot water filling is
3- Increase the diagnostics time P6.32
12 NO HOT WATER not complete after the
time set by P6.32 14B2
4-
5-
Pressure switch 14B2 defective
Solenoid valve connection NOT 4- Check and eventually replace the
OK defective pressure switch
6- Pressure switch connection NOT 5- Check output O4 of board 6A1 and
OK relative wire 10.4
6- Check input I18 and relative wire
14.18

10YV3
1- Verify if the water tap is open and if
1- Water supply interrupted there is no interruption on the loading
2- Water filling solenoid valve pipe
10YV3 defective or not 2- Check and eventually replace the
The demineralised water connected defective solenoid valve
filling is not complete 3- Time set by P6.32 too low 3- Increase the diagnostics time P6.32
13 NO DEMI WATER
after the time set by 12B2 4- Pressure switch 12B2 defective 4- Check and eventually replace the
P6.32 5- Solenoid valve connection NOT defective pressure switch
OK 5- Check output O6 of board 6A1 and
6- Pressure switch connection NOT relative wire 10.6
OK 6- Check input I18 and relative wire
12.1

10YV2

1- Water supply interrupted 1- Verify if the water tap is open and if


2- Water filling solenoid valve there is no interruption on the loading
10YV2 or 10YV1 defective or not pipe
10YV1
connected 2- Check and eventually replace the
The level of the cold+hot
3- Time set by P6.32 too low defective solenoid valve
water filling has not be
NO COLD+HOT 3- Increase the diagnostics time P6.32
14 WATER
reached after the time
set by P6.34 (level
4-
5-
Pressure switch 14B2 defective
Solenoid valve connection NOT 4- Check and eventually replace the
OK defective pressure switch
control)
6- Pressure switch connection NOT 5- Check output O5 or O4 of board 6A1
OK and relative wire 10.5 or 10.4
14B2 6- Check input I18 and relative wire
14.18

10YV2

1- Water supply interrupted 1- Verify if the water tap is open and if


2- Water filling solenoid valve there is no interruption on the loading
10YV3 or 10YV2 defective or not pipe
10YV3
connected 2- Check and eventually replace the
The level of the
3- Time set by P6.32 too low defective solenoid valve
cold+demi water filling
NO COLD+DEMI 3- Increase the diagnostics time P6.32
15 WATER
has not be reached after
the time set by P6.34
4-
5-
Pressure switch 14B2 defective
Solenoid valve connection NOT 4- Check and eventually replace the
OK defective pressure switch
(level control)
6- Pressure switch connection NOT 5- Check output O6 or O5 of board 6A1
OK and relative wire 10.6 or 10.5
14B2 6- Check input I18 and relative wire
14.18

10YV1 7- Water supply interrupted 7- Verify if the water tap is open and if
8- Water filling solenoid valve there is no interruption on the loading
The level of the
10YV3 or 10YV1 defective or not pipe
hot+demi water filling
NO HOT+DEMI connected 8- Check and eventually replace the
16 WATER
has not be reached after
the time set by P6.34
9- Time set by P6.32 too low defective solenoid valve
10- Pressure switch 14B2 defective 9- Increase the diagnostics time P6.32
(level control)
11- Solenoid valve connection NOT 10- Check and eventually replace the
10YV3 OK defective pressure switch

This document is a property of STEELCO S.p.A., it is a relevant value good and it can’t be transmitted or divulged to third party without a specific written
authorization of STEELCO direction.

2/9
12- Pressure switch connection NOT 11- Check output O6 or O4 of board 6A1
OK and relative wire 10.6 or 10.4
12- Check input I18 and relative wire
14.18

14B2

J1
The chemical product
associated to dosing
pump 1 is used-up (if it is
1- Replace the chemical tank
set as an alarm by 1- Chemical product used-up
2- Check and eventually replace the
P3.27) 2- Dosing pump 11M1 defective
defective dosing pump
Diagnostics with dosing 11M1 3- Pressure switch connection NOT
3- Check output O11 and relative wire
pump enabled: OK
17 NO CHEMICAL 1
- Pressure switch
4-
11.11
Check and eventually replace the
state with pressure With flowmeter control:
defective flowmeter
switch presence; 4- Flowmeter J1 defective
5- Increase the time P6.07
- Lack of a new 5- Time set by P6.07 too low
6- Check input J1
impulse after the 6- Flowmeter connection NOT OK
time set by P6.07 if
a flowmeter is used

J2
The chemical product
associated to dosing
pump 2 is used-up (if it is
1- Chemical product used-up 1- Replace the chemical tank
set as an alarm by
2- Dosing pump 12M1 defective 2- Check and eventually replace the
P3.27)
3- Pressure switch connection NOT defective dosing pump
Diagnostics with dosing 12M1 OK 3- Check output O12 and relative wire
pump enabled:
18 NO CHEMICAL 2
- Pressure switch
With flowmeter control: 4-
12.12
Check and eventually replace the
state with pressure
4- Flowmeter J2 defective defective flowmeter
switch presence;
5- Time set by P6.07 too low 5- Increase the time P6.07
- Lack of a new
6- Flowmeter connection NOT OK 6- Check input J2
impulse after the
time set by P6.07 if
a flowmeter is used

J4
The chemical product
associated to dosing
pump 4 is used-up (if it is
1- Replace the chemical tank
set as an alarm by 1- Chemical product used-up
2- Check and eventually replace the
P3.27) 2- Dosing pump 26M1 defective
defective dosing pump
Diagnostics with dosing 26M1 3- Pressure switch connection NOT
3- Check output O6 and relative wire
pump enabled: OK
19 NO CHEMICAL 4
- Pressure switch
4-
26.6
Check and eventually replace the
state with pressure With flowmeter control:
defective flowmeter
switch presence; 4- Flowmeter J4 defective
5- Increase the time P6.07
- Lack of a new 5- Time set by P6.07 too low
6- Check input J4
impulse after the 6- Flowmeter connection NOT OK
time set by P6.07 if
a flowmeter is used

J3
The chemical product
associated to dosing
pump 3 is used-up (if it is
1- Chemical product used-up 1- Replace the chemical tank
set as an alarm by
2- Dosing pump 12M3 defective 2- Check and eventually replace the
P3.27)
3- Pressure switch connection NOT defective dosing pump
Diagnostics with dosing 12M3 OK 3- Check output O14 and relative wire
pump enabled:
20 NO CHEMICAL 3
- Pressure switch
With flowmeter control: 4-
12.14
Check and eventually replace the
state with pressure
4- Flowmeter J3 defective defective flowmeter
switch presence;
5- Time set by P6.07 too low 5- Increase the time P6.07
- Lack of a new
1- Flowmeter connection NOT OK 6- Check input J3
impulse after the
time set by P6.07 if
a flowmeter is used

The safety input is


1- Heating elements 1EH3 1- Check and eventually replace the
contradictory with the 1EH2-
1EH3
defective defective heating element
output of the heating
HEATING 2- Heating element connection 2- Check output O7 and relative wire
21 ELEMENT 1
element relay
(diagnostics on the
NOT OK 11.7
11KM1 3- Contactor 11KM1stuck or 3- Check and eventually replace the
heating element
defective defective contactor
contactor feedback)

This document is a property of STEELCO S.p.A., it is a relevant value good and it can’t be transmitted or divulged to third party without a specific written
authorization of STEELCO direction.

3/9
1EH1

The safety input is


1- Heating elements 1EH3 1- Check and eventually replace the
contradictory with the 10KM2 defective defective heating element
output of the heating
HEATING 2- Heating element connection 2- Check output O8 and relative wire
22 ELEMENT 2
element relay
(diagnostics on the
NOT OK 11.8
3- Contactor 10KM2stuck or 3- Check and eventually replace the
heating element
defective defective contactor
contactor feedback)

After a set time from the


drain pneumatic valve
1- Remove the obstruction
activation, the water 1- Drain obstructed
11M2 2- Check and eventually replace the
level on the tank is still 2- Tank level pressure switch 14B2
defective pressure switch
active (tank not empty). 14B2 defective
3- Check and eventually replace the
3- Drain pump 11M2 defective
defective pump
23 DRAIN PROBLEM
The timeout is referred
4-
5-
Drain pump connection NOT OK
Pressure switch connection NOT
4- Controllare l’uscita O10 e relativo filo
to: 11.10
OK
5- Check input I18 and relative wire
6- Time set by P6.04 or P6.31 too
14.18
a) P6.04 simple drain low
6- Increase P6.04 or P6.31
b) P6.31 drain with
rinse water intake

1- Check and eventually replace the


With drying enable:
defective pressure switch
1M2 1- Drying pressure switch 17B1
2- Check and eventually replace the
defective
17B1 defective fan
2- Fan 1M2 defective
Pressure switch open 3- Check and eventually replace the
3- Air filter obstructed
with fan switched on at defective filter
4- Fan connection NOT OK
the maximum speed; 4- Check output O7 of board 6A2 and
24 FAN PROBLEM 5- Pressure switch connection NOT
OK
relative wire 26.7
Pressure switch closed 5- Check input I40 of board 6A1 and
with fan switched off relative wire 17.40
With drying disabled:
6- Check and eventually replace the
6- Drying pressure switch 17B1
defective pressure switch
defective
7- Check output O7 of board 6A2 and
7- Fan connection NOT OK
relative wire 26.7

20B1

1- Check and eventually replace the


1- Probe 20B1 defective
1EH4-1EH6 defective probe
2- Probe connection NOT OK
2- Check input I35-I36
The air temperature 3- Heating element 1EH4-1EH6
3- Check and eventually replace the
have never reached the defective
defective heating element
25 DRYING MIN °C minimum temperature
set by P7.13 during the
4- Air heating element contactor
28KM1 defective
4- Check and eventually replace the
defective contactor
drying phase 28KM1 5- Heating element connection
5- Check output O8 of board 6A2 and
NOT OK
relative wire 28.8
6- Time set by P7.13 too high
6- Decrease P7.13

1- Prewashing limit temperature


19B1 (P7.15) incorrect
The tank temperature is 1- Set P7.15
2- Program phases error
over the maximum 2- Correct working program
PREWASH MAX (prewashing phase execute
26 °C
allowed temperature
during a prewashing
immediately after an high
3- Check and eventually replace the
defective probe
temperature phase)
phase 4- Set the probe offset
3- Probe 19B1 defective
4- Probe off set 19B1 incorrect
19B1
1- Check and eventually replace the
1- Probe 19B1 defective
The tank temperature is defective probe
2- Probe off set 19B1 incorrect
TANK PROBE over 102°C 2- Set the probe offset
27 LIMIT °C (= max limit 95°C +
3-
4-
Probe connection NOT OK
Tank heating element contactor
3- Check input I32-I33
emergency 7°C) 4- Check and eventually replace the
stuck
defective contactor

This document is a property of STEELCO S.p.A., it is a relevant value good and it can’t be transmitted or divulged to third party without a specific written
authorization of STEELCO direction.

4/9
1- Check and eventually replace the
1- Probe 20B1 defective
20B1 defective probe
2- Heating element connection
The air temperature is 2- Check input I35-I36
NOT OK
DRYING PROBE over 162°C 3- Check and eventually replace the
28 LIMIT °C (= max limit 150°C +
3- Air heating element contactor
stuck
defective contactor
emergency 12°C) 4- Check output O8 of board 6A2 and
4- Contactor connection NOT OK
relative wire 28.8
5- Probe off set 20B1 incorrect
5- Set the probe by P4.21
1- Check and eventually replace the
1- Probe 31B2 defective
31B2 defective probe
2- Heating element connection
The boiler temperature is 2- Check input I19-I20
NOT OK
BOILER PROBE over 100°C 3- Check and eventually replace the
29 LIMIT °C (= max limit 80°C +
3- Boiler heating element contactor
stuck
defective contactor
emergency 20°C) 4- Check output O11 of board 6A2 and
4- Contactor connection NOT OK
relative wire 28.2
5- Probe off set 31B2 incorrect
5- Set the probe by P4.21
19B1

Tank temperature probe 1- Probe 19B1 defective 1- Replace the defective probe
30 TANK PROBE
failure (working probe) 2- Probe connection NOT OK 2- Check input I32-I33

31B1

Tank temperature probe


1- Probe 31B1 defective 3- Replace the defective probe
31 TANK PROBE 2 failure (registration
probe)
2- Probe connection NOT OK 1- Check input I21-I22 of board 6A2

20B1

Air temperature probe 1- Probe 20B1 defective 4- Replace the defective probe
32 DRYING PROBE
failure 2- Probe connection NOT OK 1- Check input I35-I36

31B2

Boiler temperature probe 3- Probe 31B2 defective 5- Replace the defective probe
33 BOILER PROBE
failure 4- Probe connection NOT OK 2- Check input I19-I20

Only for the treatment


phase if: 19B1
a) The tank
temperature is
over P7.11
1- Incongruity parameter P.7.04 1- Set P7.04
b) The registration
incorrect 2- Set the probe by P4.19 e P4.25
probe is different
CHECK 2- Probes off set incorrect 3- Replace the defective probe
34 TEMPERATURE
from the working
probe of a value
31B1
3- Probe 19B1 defective 4- Replace the defective probe
4- Probe 31B1 defective 5- Check input I32-I33 and I29-I30 for
over P7.04,
5- Probes connection NOT OK board 6A2
c) Tank heating
element off
d) All the condition
a,b,c) still exist
form 30 s
1- Loading side plug cable
defective
The communication 1- Replace the plug cable
2- Control panel- main board
SERIAL between the main board 2- Check the connection
35 CONNECTION 1 and the control panel is
6A1
3-
connection NOTOK
Main board 6A1 defective
3- Replace the defective board
interrupt (loading side) 4- Replace the defective control panel
4- Loading side control panel
defective

36 -------------------------- -------------------------------- ----------------------- ---------------------------------- ----------------------------------

6A1 e 6A2

1- Board connection cable


The communication 1- Replace the plug cable
defective
CAN SERIAL between the main board 2- Check the connection
37 CONNECTION and slave board is
2-
3-
Board connection NOT OK
Main board 6A1 defective
3- Replace the defective main board
interrupt 4- Replace the defective slave board
4- Slave board 6A2 defective

38 -------------------------- -------------------------------- ----------------------- ---------------------------------- ----------------------------------

This document is a property of STEELCO S.p.A., it is a relevant value good and it can’t be transmitted or divulged to third party without a specific written
authorization of STEELCO direction.

5/9
19B1

1- Time set by P6.01 too low 1- Increase P6.01


2- Heating element contactor 2- Check and eventually replace the
During the tank electrical
10KM1 stuck defective contactor
heating the temperature
NO TANK 3- Contactor connection NOT OK 3- Check output O7 and relative wire
39 HEATING
is increased less than
1°C after the time set by
10KM1
4- Probe 19B1 defective 11.7
5- Probes connection NOT OK 4- Replace the defective probe
P6.01
6- Water filling solenoid valve 5- Check I32-I33
defective 6- Replace the defective solenoid valve

40 -------------------------- -------------------------------- ----------------------- ---------------------------------- ----------------------------------


32B2

1- Time set by P6.01 too low 1- Increase P6.01


2- Heating element contactor 2- Check and eventually replace the
During the boiler
10KM1 stuck defective contactor
electrical heating the
NO BOILER 3- Contactor connection NOT OK 3- Check output O11 and relative wire
41 HEATING
temperature is increased
less than 1°C after the
10KM1
4- Probe 32B2 defective 28.11
5- Probes connection NOT OK 4- Replace the defective probe
time set by P6.03
6- Water filling solenoid valve 5- Check I32-I33
defective 6- Replace the defective solenoid valve

42 -------------------------- -------------------------------- ----------------------- ---------------------------------- ----------------------------------

43 -------------------------- -------------------------------- ----------------------- ---------------------------------- ----------------------------------

16B1 - 16B2
Max steam condenser 1- Level switch 16B2 and 16B1 1- Replace the defective level switch
LEVEL STEAM level enabled and min defective 2- Check input I26 and I25 and relative
44 CONDENSER steam condenser level 2- Level switch connection NOT wire I16.25 of board 6A2 and relative
disabled OK wire 25.2

1- Replace the defective steam


Max and min steam condenser drain pump
1- Drain pump 23M2 defective
PUMP STEAM condenser level enabled: 23M2 2- Check output O9 of board 6A2 and
45 CONDENSER probable drain pump
2-
3-
Drain pump connection NOT OK
Pipe obstructed
relative wire 28.9
failure 3- Check the drain pipe and the pump-
steam condenser connection
1- Verify the cycle parameters and/or
verify the drain valve leak
2- Verify the cycle parameters (chemical
12M4 Pump 12M4 on:
quantity + P8.17)
1- Water quantity not sufficient
3- Check and eventually replace the
2- High Chemical quantity
defective pump
3- Pump 12M4 defective
4- Check and eventually replace the
4- Pressure switch 17B2 defective
defective pressure switch
5- Pressure switch connection NOT
5- Check input I41 and relative wire
Pump pressure switch 17B2 OK
17.41
46 PUMP closed with pump on, or
open with pump off
6- Pump connection NOT OK
6- Check output O15 and relative wire
12.15
Pump 12M4 off:
7- Check and eventually replace the
7- Pump connector stuck
defective contactor
8- Contactor connection NOT OK
8- Check output O15 and relative wire
9- Pressure switch 17B2 defective
12.15
10- Pressure switch connection NOT
9- Check and eventually replace the
OK
defective pressure switch
10- Check input I41 and relative wire
17.41
11M1

The chemical1 flowmeter


1- Dosing pump1 11M1 defective 1- Replace the dosing pump 1
signal an impulse
FLOWMETER 2- P7.21 incorrect 2- Check and control P7.21
47 FAILURE 1
number superior to value
set by P7.21 with the
3- Flowmeter J1 defective 3- Replace the defective flowmeter
4- Flowmeter connection NOT OK 4- Check input J1
dosing pump off J1

This document is a property of STEELCO S.p.A., it is a relevant value good and it can’t be transmitted or divulged to third party without a specific written
authorization of STEELCO direction.

6/9
12M1

The chemical2 flowmeter


1- Dosing pump2 12M1 defective 1- Replace the dosing pump 2
signal an impulse
FLOWMETER 2- P7.21 incorrect 2- Check and control P7.21
48 FAILURE 2
number superior to value
set by P7.21 with the
3- Flowmeter J2 defective 3- Replace the defective flowmeter
4- Flowmeter connection NOT OK 4- Check input J2
dosing pump off J2

11M2

The chemical3 flowmeter


1- Dosing pump3 11M2 defective 1- Replace the dosing pump 3
signal an impulse
FLOWMETER 2- P7.21 incorrect 2- Check and control P7.21
49 FAILURE 3
number superior to value
set by P7.21 with the
3- Flowmeter J3 defective 3- Replace the defective flowmeter
4- Flowmeter connection NOT OK 4- Check input J3
dosing pump off J3

11M2

The chemical4 flowmeter


1- Dosing pump3 11M2 defective 1- Replace the dosing pump 3
signal an impulse
FLOWMETER 2- P7.21 incorrect 2- Check and control P7.21
50 FAILURE 4
number superior to value
set by P7.21 with the
3- Flowmeter J3 defective 3- Replace the defective flowmeter
4- Flowmeter connection NOT OK 4- Check input J3
dosing pump off J3

51 -------------------------- -------------------------------- ----------------------- ---------------------------------- ----------------------------------

52 -------------------------- -------------------------------- ----------------------- ---------------------------------- ----------------------------------

53 -------------------------- -------------------------------- ----------------------- ---------------------------------- ----------------------------------

54 -------------------------- -------------------------------- ----------------------- ---------------------------------- ----------------------------------


Device 9R1 1- Conductivity limit parameter not 1- Fix parameter within treatment phase
correct parameter
2- Max. number of attempts 2- Fix parameter MAIN
parameter not correct PAGE -> SETUP -> SYSTEM DATA
3- Conductivity transmitter 9B7 not -> "Nr Attm.Contr. Conduc.""
CONDUCTIVITY Conductivity probe
55 PROBE failure Device 9B7
4-
calibrated
Conductivity transmitter
3-
4-
Calibrate conductivity transmitter 9B7
Replace conductivity transmitter 9B7
9A3 defective 5- Replace conductivity sensor 9R1
5- Conductivity sensor defective or
exhausted 9R1

Device 9R1 1- Conductivity limit parameter not 1- Fix parameter within treatment phase
correct parameter
2- Max. number of attempts 2- Fix parameter MAIN
parameter not correct PAGE -> SETUP -> SYSTEM DATA
Conductivity value 3- Conductivity transmitter 15A3 -> "Nr Attm.Contr. Conduc.""
56 CONDUCTIVITY superior to the value set
by P7.23
Device 9B7
4-
not calibrated
Conductivity transmitter
3- Calibrate conductivity transmitter
15A3
15A3 defective 4- Replace conductivity transmitter
5- Conductivity sensor defective or 15A3
exhausted 15R1 5- Replace conductivity sensor 15R1

57 -------------------------- -------------------------------- ----------------------- ---------------------------------- ----------------------------------

58 -------------------------- -------------------------------- ----------------------- ---------------------------------- ----------------------------------

This document is a property of STEELCO S.p.A., it is a relevant value good and it can’t be transmitted or divulged to third party without a specific written
authorization of STEELCO direction.

7/9
59 -------------------------- -------------------------------- ----------------------- ---------------------------------- ----------------------------------

During the treatment


phase the thermal-
regulating timeout
1- Excessive chimney suction 1- Decrease the chimney suction
expired (30min) (start to
60 TIME
count the first time that / 2- Wrong loading chemical
temperature
2- Check and set the chemical loading
temperature
the tank temperature
reach the setpoint
temperature +0.5 °C

61 -------------------------- -------------------------------- ----------------------- ---------------------------------- ----------------------------------

62 -------------------------- -------------------------------- ----------------------- ---------------------------------- ----------------------------------

63 -------------------------- -------------------------------- ----------------------- ---------------------------------- ----------------------------------

64 -------------------------- -------------------------------- ----------------------- ---------------------------------- ----------------------------------

65 -------------------------- -------------------------------- ----------------------- ---------------------------------- ----------------------------------

66 -------------------------- -------------------------------- ----------------------- ---------------------------------- ----------------------------------

67 -------------------------- -------------------------------- ----------------------- ---------------------------------- ----------------------------------

68 -------------------------- -------------------------------- ----------------------- ---------------------------------- ----------------------------------

69 -------------------------- -------------------------------- ----------------------- ---------------------------------- ----------------------------------

70 -------------------------- -------------------------------- ----------------------- ---------------------------------- ----------------------------------

71 -------------------------- -------------------------------- ----------------------- ---------------------------------- ----------------------------------

Warning list for DS 500 model Rev. 00

WARNING
WARNING TEXT DEVICE POSSIBLE CAUSES REMEDY
DESCRIPTION
It is possible to start a
cycle during a stand-by
state.
With a double door
PRESS START
machine is necessary to / / /
open and close the door
one time after the end of a
cycle
The chemical product
associated to dosing
pump 1 is used-up (if it is 14B3
set as a warning by 1- Chemical 1 used up
P3.27) 2- Level sensor 14B3 chemical 1
1- Replace the chemical tank 1
Diagnostics with dosing dirty or blocked
2- Clean the level sensor chemical 1
NO CHEMICAL 1 pump enabled: 3- Level sensor 14B3 chemical 1
3- Replace the level sensor chemical 1
- Pressure switch state defective
4- Check input I20 and relative wire 14.20
with pressure switch 4- Level sensor connection NOT
presence; OK
Lack of a new impulse
after the time set by P6.07
if a flowmeter is used
The chemical product
associated to dosing
pump 2 is used-up (if it is 12B1
set as a warning by 1- Chemical 2 used up
P3.27) 2- Level sensor 14B1 chemical 2
1- Replace the chemical tank 2
Diagnostics with dosing dirty or blocked
2- Clean the level sensor chemical 2
NO CHEMICAL 2 pump enabled: 3- Level sensor 14B1 chemical 2
3- Replace the level sensor chemical 2
- Pressure switch state defective
4- Check input I17 and relative wire 14.17
with pressure switch 4- Level sensor connection NOT
presence; OK
Lack of a new impulse
after the time set by P6.07
if a flowmeter is used

This document is a property of STEELCO S.p.A., it is a relevant value good and it can’t be transmitted or divulged to third party without a specific written
authorization of STEELCO direction.

8/9
WARNING
WARNING TEXT DEVICE POSSIBLE CAUSES REMEDY
DESCRIPTION
The chemical product
associated to dosing
pump 4 is used-up (if it is 15B2
set as a warning by 1- Chemical 4 used up
P3.27) 2- Level sensor 15B2 chemical 4
1- Replace the chemical tank 4
Diagnostics with dosing dirty or blocked
2- Clean the level sensor chemical 4
NO CHEMICAL 4 pump enabled: 3- Level sensor 15B2 chemical 4
3- Replace the level sensor chemical 4
- Pressure switch state defective
4- Check input I24 and relative wire 15.24
with pressure switch 4- Level sensor connection NOT
presence; OK
Lack of a new impulse
after the time set by P6.07
if a flowmeter is used
The chemical product
associated to dosing
pump 3 is used-up (if it is 15B1
set as a warning by 1- Chemical 3 used up
P3.27) 2- Level sensor 15B1 chemical 3
1- Replace the chemical tank 3
Diagnostics with dosing dirty or blocked
2- Clean the level sensor chemical 3
NO CHEMICAL 3 pump enabled: 3- Level sensor 15B1 chemical 3
3- Replace the level sensor chemical 3
- Pressure switch state defective
4- Check input I21 and relative wire 15.21
with pressure switch 1- Level sensor connection NOT
presence; OK
Lack of a new impulse
after the time set by P6.07
if a flowmeter is used
Make a salt refill after the
cycle number set by
SALT LOADING
P7.28 (with P7.27 / / /
different form 10)

---------------------------------- ---------------------------------- ----------------------- ---------------------------------- ----------------------------------

Inform that one door is


- OPEN DOOR -
open / / /
Generic warn that inform
WAIT to wait before to do a new / / /
action

---------------------------------- ---------------------------------- ----------------------- ---------------------------------- ----------------------------------

---------------------------------- ---------------------------------- ----------------------- ---------------------------------- ----------------------------------

---------------------------------- ---------------------------------- ----------------------- ---------------------------------- ----------------------------------

---------------------------------- ---------------------------------- ----------------------- ---------------------------------- ----------------------------------

---------------------------------- ---------------------------------- ----------------------- ---------------------------------- ----------------------------------

Cycle successful
OK
completed / / /
The cycle has been The cycle was interrupt before the Restart a cycle form the beginning
NO DISINFECTION
interrupted thermal disinfection process end

This document is a property of STEELCO S.p.A., it is a relevant value good and it can’t be transmitted or divulged to third party without a specific written
authorization of STEELCO direction.