You are on page 1of 1

CMP 200 Switching Frequencies

The following plots show the CMP 200 inverter voltage and current at maximum
and minimum power. These were done with a 32 KW generator on a 208 VAC
line.

Resonant
current
250A/Div
212 A RMS

Inverter
voltage

Maximum power, 125 KV, 250 mA, 10 ms after start of exposure

Resonant
current
100 A/Div
69 A RMS

Inverter
voltage

Minimum power, 40 KV, 10 mA, 10 ms after start of exposure.

Use and disclosure is subject to the restrictions on the title page of this CPI document