EMS Piezon® built-in scaler kit w. or wo.

LED Ultrasonic generator P18


Generic AC or DC connection diagram - power on/off by power supply.

 Caution: Never connect AC + DC power supplies as this would damage the module.
Cut, insulate and tie the wires that are not used.

Connect pin 4 to
ground to enable
the light

Potentiometer
Type A: 2.2 kΩ
Type B 4.7 kΩ

0V
IMPORTANT NOTICE:
for AC installation +
Conect (12)DCI to (2)DCO +
for DC installation
remove (2)DCO OR
0V Control voltage:
+ Type A: 0 to 5V DC
Type B: 0 to 10V DC
1
2 gray
3 violet
4 pink
5 brown
6 white
Module P18

7 gray
8 brown/white
9 yellow
10 blue
11 blue
12 orange
13 black
optional air actuated
A white
B violet
electric switch
(or other switch)
1 green
2 If LED supply
3
4 red of module
not used
connect adapter
Option1
AC installation:
Supply voltage
24V AC (±10%)

OR
Option2
+ DC installation:
0V Supply voltage
24-32V DC (±10%)

Cold water supply


1.0 - 2.0 bars
clear tubing
(100 - 200 kPa)
(15 - 29 PSI)
Note: installation of a
ed. 2011/12

Adjustable flow rate


connector is recommended. 0-50ml/min ±10
(not supplied in kit)

Anti extension string


FB-519

(attach to frame when a


connector is not used)
© Copyright EMS SA

EMS Electro Medical Systems SA EMS worldwide offices

Ch. de la Vuarpillière 31 EMS Canada Inc. EMS France Sarl / 23, Av. Louis Bréguet EMS Gmbh EMS Italia S.p.a. EMS España SL EMS Corporation
1260 Nyon / Switzerland 5524 St. Patrick St., Suite 312 Immeuble Santos Dumont, Bât. D, Schatzbogen 86 Via Faravelli 5 Bernardino Obregón 14 bis, 11886 Greenville Ave. #120
Tel. +41 22 99 44 700 Montreal, QC H4E 1A8 F-78140 Vélizy Villacoublay D-81829 München I-20149 Milano E-28012 Madrid Dallas TX-75243, USA
Fax +41 22 99 44 701 Tel. +1 514 736 5066 Tél. +33 1 34 58 03 80 Tel. +49 89 42 71 610 Tel. +39 02 3453 8111 Tlf. +34 91 528 99 89 Tel. +1 972 690 83 82
e-mail: welcome@ems-ch.com Fax. +1 514 736 5065 Fax +33 1 34 58 03 90 Fax +49 89 42 71 6160 Fax +39 02 3453 2778 Fax +34 91 539 34 89 Fax +1 972 690 89 81
www.ems-swissquality.ch e-mail: info@ems-canada.ca e-mail: info@ems-france.fr e-mail: info@ems-ch.de e-mail: dental@ems-italia.it e-mail: ems-es@arrakis.es e-mail: emsusa@emsdent.com

FB-519-module P18 diagrams.indd 1 12.12.2011 16:21:41


EMS Piezon® built-in scaler kit w. or wo. LED Ultrasonic generator P18
Generic AC or DC connection diagram - remote power on/off - Module P18 Type B only.

 Caution: Never connect AC + DC power supplies as this would damage the module.
Cut, insulate and tie the wires that are not used.

Connect pin 4 to
ground to enable
the light

Potentiometer
Type B: 4.7 kΩ

0V
IMPORTANT NOTICE:
for AC installation +
Conect (12)DCI to (2)DCO +
for DC installation
remove (2)DCO OR
0V Control voltage:
+ Type B: 0 to 10V DC

1
gray
Module P18 Type B

2
3 violet
4 pink
5 brown
6 white
7 gray
8 brown/white
9
optional air actuated
yellow
10 blue electric switch
11 blue
12 orange
(or other switch)
13 black
A white
B violet
1 green
2 If LED supply
3
4 red of module
not used
connect adapter
Option1
AC installation:
Supply voltage
24V AC (±10%)
OR
Option2
DC installation:
+
Supply voltage
0V
24-32V DC (±10%)

Cold water supply


1.0 - 2.0 bars
clear tubing
(100 - 200 kPa)
(15 - 29 PSI)
ed. 2011/12

Note: installation of a Adjustable flow rate


connector is recommended. 0-50ml/min ±10
(not supplied in kit)
FB-519

Anti extension string


(attach to frame when a
connector is not used)
© Copyright EMS SA

EMS Electro Medical Systems SA EMS worldwide offices

Ch. de la Vuarpillière 31 EMS Canada Inc. EMS France Sarl / 23, Av. Louis Bréguet EMS Gmbh EMS Italia S.p.a. EMS España SL EMS Corporation
1260 Nyon / Switzerland 5524 St. Patrick St., Suite 312 Immeuble Santos Dumont, Bât. D, Schatzbogen 86 Via Faravelli 5 Bernardino Obregón 14 bis, 11886 Greenville Ave. #120
Tel. +41 22 99 44 700 Montreal, QC H4E 1A8 F-78140 Vélizy Villacoublay D-81829 München I-20149 Milano E-28012 Madrid Dallas TX-75243, USA
Fax +41 22 99 44 701 Tel. +1 514 736 5066 Tél. +33 1 34 58 03 80 Tel. +49 89 42 71 610 Tel. +39 02 3453 8111 Tlf. +34 91 528 99 89 Tel. +1 972 690 83 82
e-mail: welcome@ems-ch.com Fax. +1 514 736 5065 Fax +33 1 34 58 03 90 Fax +49 89 42 71 6160 Fax +39 02 3453 2778 Fax +34 91 539 34 89 Fax +1 972 690 89 81
www.ems-swissquality.ch e-mail: info@ems-canada.ca e-mail: info@ems-france.fr e-mail: info@ems-ch.de e-mail: dental@ems-italia.it e-mail: ems-es@arrakis.es e-mail: emsusa@emsdent.com

FB-519-module P18 diagrams.indd 2 12.12.2011 16:21:42


EMS Piezon® built-in scaler kit w. or wo. LED Ultrasonic generator P18
Generic AC or DC connection diagram - remote power on/off - Module P18 Type A only.

 Caution: Never connect AC + DC power supplies as this would damage the module.
Cut, insulate and tie the wires that are not used.

Connect pin 4 to
ground to enable
the light

Potentiometer
Type A: 2.2 kΩ

0V
IMPORTANT NOTICE:
for AC installation +
Conect (12)DCI to (2)DCO +
for DC installation
remove (2)DCO OR
0V Control voltage:
+ Type A: 0 to 5V DC

1
Type A

2 gray
3 violet
4 pink
5 brown
6 white
7 gray
8 brown/white Power:
9 yellow ON : 0VL / open
10 blue
OFF: 5V
Module P18

11 blue
12 orange
13 black
A white
B violet
1 green
2 If LED supply
3
4 red of module
not used
connect adapter
Option1

AC installation:
Supply voltage
24V AC (±10%)

OR
Option2
+ DC installation:
0V Supply voltage
24-32V DC (±10%)

Cold water supply


1.0 - 2.0 bars
clear tubing
(100 - 200 kPa)
(15 - 29 PSI)
Note: installation of a
ed. 2011/12

Adjustable flow rate


connector is recommended. 0-50ml/min ±10
(not supplied in kit)

Anti extension string


FB-519

(attach to frame when a


connector is not used)
© Copyright EMS SA

EMS Electro Medical Systems SA EMS worldwide offices

Ch. de la Vuarpillière 31 EMS Canada Inc. EMS France Sarl / 23, Av. Louis Bréguet EMS Gmbh EMS Italia S.p.a. EMS España SL EMS Corporation
1260 Nyon / Switzerland 5524 St. Patrick St., Suite 312 Immeuble Santos Dumont, Bât. D, Schatzbogen 86 Via Faravelli 5 Bernardino Obregón 14 bis, 11886 Greenville Ave. #120
Tel. +41 22 99 44 700 Montreal, QC H4E 1A8 F-78140 Vélizy Villacoublay D-81829 München I-20149 Milano E-28012 Madrid Dallas TX-75243, USA
Fax +41 22 99 44 701 Tel. +1 514 736 5066 Tél. +33 1 34 58 03 80 Tel. +49 89 42 71 610 Tel. +39 02 3453 8111 Tlf. +34 91 528 99 89 Tel. +1 972 690 83 82
e-mail: welcome@ems-ch.com Fax. +1 514 736 5065 Fax +33 1 34 58 03 90 Fax +49 89 42 71 6160 Fax +39 02 3453 2778 Fax +34 91 539 34 89 Fax +1 972 690 89 81
www.ems-swissquality.ch e-mail: info@ems-canada.ca e-mail: info@ems-france.fr e-mail: info@ems-ch.de e-mail: dental@ems-italia.it e-mail: ems-es@arrakis.es e-mail: emsusa@emsdent.com

FB-519-module P18 diagrams.indd 3 12.12.2011 16:21:42


FB-519-module P18 diagrams.indd 4 12.12.2011 16:21:42

You might also like