DATA VERIFIKASI

SALINAN HASIL KALIBRASI ALAT STANDAR DATA ALAT

Standar : Termometer Digital + PT 100 Nama Alat


Merk : Graphtec Merk
Tipe : GL 840 Tipe
No. Seri : 05660 No. Seri
Resolusi 0.01 Resolusi
Ketertelusuran : LK-133-IDN Tanggal

DATA SERTIFIKAT
Set Poin Standar Value Uut Value Error U95
y x
-30 -29.936 -31.3 1.364 0.11
0 -0.038 -0.79 0.752 0.11
150 150.014 152.15 -2.136 0.3
300 299.963 305.02 -5.057 0.3
450 449.935 458.05 -8.115 0.3
600 600 611.3 -11.3 0.3

Rata-rata
Set Point Reading Standar Standar Display Unit Koreksi Set Point Untuk
Koreksi
(°C) (°C) Terkoreksi (°C) (°C) (°C) Penggunaan (°C)
(°C)
200.80 198 205 -7
200 200.30 197 204 -7 -7 207
199.70 197 203 -6
399.60 393 400 -7
400 400.30 393 400 -7 -7 407
400.40 393 400 -7
607.80 597 600 -3
600 607.80 597 600 -3 -3 603
607.90 597 600 -3
DATA ALAT

: Furnace Nabertherm
: Nabertherm
:L5/11/B410
: 375453
1
25-Feb-20
SUMMARY OUTPUT
mi ( g ) = a + bxi
a 0.798103655 Regression Statistics
b 0.980406829 Multiple R 1
R Square 1
Adjusted R Square 1
Standard Error 0.1016319
Observations 6

ANOVA
df SS MS F
Regression 1 315700 315700 30564300
Residual 4 0.0413162 0.010329
Total 5 315700.05

Coefficients
Standard Error t Stat P-value
Intercept 0.7981037 0.0606008 13.169847 0.000192
X Variable 1 0.9804068 0.0001773 5528.4989 6.423E-15
Significance F
6.423E-15

Lower 95%Upper 95%Lower 95,0%


Upper 95,0%
0.6298488 0.9663585 0.6298488 0.9663585
0.9799145 0.9808992 0.9799145 0.9808992