3 Release Notes
VPMC-13457B
Copyright © 1997-2014 Vital Images, Inc. All rights reserved. Date of Publication July 2014.
VitreaAdvanced® protected by U.S. Patents 5,986,662; 6,130,671; 6,219,059; 7,031,504; 7,136,064; 7,362,329; 7,574,029; 7,590,272;
7,660,481;7,929,748;7,991,210; 8,214756; 8,249,687 Other Patents Pending in the U.S. and other countries.
VitreaWorkstation™ protected by U.S. Patents 5,986,662; 6,130,671; 6,219,059; 7,031,504; 7,136,064; 7,362,329; 7,574,029; 7,590,272;
7,660,481;7,929,748;7,991,210; 8,214756; 8,249,687 Other Patents Pending in the U.S. and other countries.
VitreaCore protected by U.S. Patents 5,986,662; 6,130,671; 6,219,059; 7,039,723; 7,136,064; 7,362,329; RE42,952; Other Patents Pending in the U.S.
and other countries.
VitreaView® protected by U.S. Patents 6,130,671; 6,219,059; 7,136,064; and 7,362,329. Other Patents Pending in the U.S. and other countries.
MDSS GmbH
Schiffgraben 41
30175 Hannover, Germany
CELEO Pty Ltd. and Toshiba Australia Pty Ltd are authorized sponsors in Australia and act on behalf of Vital Images, Inc. in the communication of
safety-related incidents and regulatory matters with Therapeutic Goods Administration in Australia. Distributors are still the first line of communication
with their customers regarding service and complaints.
Manufactured by: Vital Images, Inc; 5850 Opus Parkway, Suite 300; Minnetonka, MN, USA; 55343; Phone 866.433.4624
CAUTION: For general Safety and Regulatory Considerations, refer to the appropriate Education and Reference Guide.
CAUTION: Federal law restricts this device to sale by or on the order of a physician, as directed by 21 CFR 801.109(b)(1).
Contact Us
For general, non-technical support questions, contact us through our Web site: www.vitalimages.com.
For customer technical support, contact us at:
o In the U.S., call the Customer Support line at 1.800.208.3005; or
o Outside the U.S., contact your Vital Images distributor; or
o Send an email to support@vitalimages.com.
To provide feedback about this document or other Vital Images product documentation, send an email to
feedback@vitalimages.com.
Contact Customer Support for a printed version of the release notes, education and reference guide, or installation guides.
As of
Products Description
Version
6.6.1 VitreaExtend There are known image quality issues when running VitreaExtend on Windows XP 32-bit and 64-bit
(SP2 and SP3) where images display tearing artifacts. To avoid any image quality issues, Vital
recommends you use Windows 7.
6.6.1 VitreaExtend Console Logout Restrictions
VIT-10705 Once any user logs into the VitreaExtend Host System, they must not log out if any Remote sessions
are active. If they do log out, then those sessions will become unresponsive. If there are no active
Remote sessions, a user can log out of the VitreaExtend Host System without issue. For more
information, refer to the VitreaExtend Install Guide.
6.6.1 VitreaExtend Remote sessions will log off after 15 minutes of inactivity. This amount of time is configured by your
System Administrator.
6.6.1 VitreaExtend If all remote sessions are in use, you will be notified that no additional sessions are available. The user
names using the open sessions will be listed.
6.6.1 VitreaExtend When loading very large datasets and/or using applications with high computation needs (such as Brain
Perfusion 4D) while using multiple concurrent remote sessions, you may notice longer loading and
processing times.
6.6.1 VitreaExtend VitreaExtend supports loading the same dataset on multiple sessions concurrently. Be aware that
software does not prevent multiple users viewing the same patient study at the same time.
6.6 VitreaAdvanced By default, both the Anticipatory Processing and MINT services are disabled. Only one of these two
VitreaWorkstation services should be enabled on a VIMS node with 8GB of RAM or less. If both services are needed, the
VIMS node should be upgraded to 16GB of RAM or more.
Anticipatory
Processing
6.6 VitreaAdvanced Some of the Anticipatory Processing scripts require customization in order to operate correctly in a
VitreaWorkstation particular environment. Before enabling the Anticipatory Processing services, all scripts should be
reviewed and/or updated by Vital Professional Services to ensure a proper configuration for the site.
Anticipatory
Processing
General Notes
As of
Products Description
Version
6.6 VitreaAdvanced Cardiac: Pediatric Functional CT protocol is no longer included with Multi-Chamber CFA license
VitreaWorkstation In earlier releases, the CT Cardiac Multi-Chamber Functional Analysis license provided access to the 3-
Cardiac: Pediatric Chamber Functional Analysis preset in two protocols: Cardiac Functional CT and Cardiac Pediatric
Function CT Functional CT. This license continues to provide the 3-chamber preset in the CT Cardiac Function CT
protocol. However, the Cardiac Pediatric Functional CT protocol will not be available. Snapshots created
in earlier releases using the Cardiac Pediatric Functional CT protocol will not be restorable.
The following features, functionality and enhancements have been incorporated into VitreaAdvanced, VitreaExtend, and
VitreaWorkstation products:
Clinical Application Enhancements
1. Vessel Tracking Improvements - CT Angiography vessel tracking and segmentation results have been improved in this release
for high hounsfield unit exams. These improvements include the following:
Improved vascular algorithm performance
o The algorithm can process 80 kVp – 100 kVp source images
o We support 80 kVp – 100 kVp input images
o These features have been validated at high hounsfield unity thresholds
Improved vessel tracking, lumen segmentation, and centerline creation
Improved compatibility with scan acquisitions and iterative reconstruction methods
Key benefits are reduced editing time, workflow efficiency, and reproducibility
2. CT Liver Enhancements - Vital has added the following new features and enhancements to the Resection Planning protocol, part
of the CT Liver Analysis Application:
One click liver segmentation with improved bone (rib) removal
One click segmentation of Portal and Hepatic Veins with improved vessel grow
One click segmentation of Hepatic Artery with improved vessel grow
Vessel tree editing tool with “show” tree feature
Improved 3D image rotation for segmented anatomy and smaller regions
Better visual display of multiple series with triple volume 6-up layouts
Multi-Volume Fusion (provides ability to fuse arterial and venous structures into one 3D image)
New user HELP button to guide a user through the liver workflow
3. CT Lung Enhancements - Vital has added the following new features and enhancements to the CT Lung Analysis Application:
New dictation table captures clinically relevant findings for monitoring and tracking lung nodules such as:
Added “Fleischer Clinical Criteria” for assisting with follow-up and management of nodules
Automatic nodule measurements including RECIST and WHO (Max and Short Axis Diameter)
Ability to restore nodules from prior exams
Manual nodule probe tool enables ability to find, create and edit a nodule region
The Nodule Management table was increased in size to minimize scrolling for analyzing multiple nodules
4. CT Brain Perfusion Enhancements– we have added the following support for Perfusion Datasets
Created a more simplified product structure with the following name and licensing changes:
o 2 CT Brain Perfusion Licenses available – CT Brain Perfusion 2D and CT Brain Perfusion 4D
o CT Brain Perfusion is now CT Brain Perfusion 2D
o CT Brain Analysis 4D is now CT Brain Perfusion 4D
Support for Wide Volume Thin slice Brain Perfusion datasets from Toshiba Aquilion Prime Scanners
Support Thick Slice “Jog Shuttle” from Philips Scanners (2D Perfusion)
Summary maps are now available in all CT Brain Perfusion protocols without a separate license (except for VitreaLT)
5. Head and Neck Bone Subtraction - CT Angiograms of the head and neck contain bone and vessels together; the fundamental
workflow is to separate them to visualize the vessels optimally. We have added the following new features to the Vascular COW
CT and Vascular Carotid CT protocols:
Ability to load two identical series (non-contrast and post-contrast)
Two NEW icons in the Application Tab and VitreaAdvanced for a visual cue for the new feature
Automatic Registration and Subtraction Processing
o Head ONLY uses Rigid Registration
o Head/Neck uses Deformable Registration
Manual Registration and Subtraction buttons available if needed
Volume Navigation with renaming option
Dual Volume layout (default) with optional 6-up layout for Pre, Post, and Sub
Fusion of Pre and Sub series
2. Medis (QMass and QFlow) – We have updated to the latest released version of the Medis products into VitreaAdvanced. Key new
features include:
QMass 7.6 - MassK mode (automatic papillary muscle segmentation)
QFlow 5.6 - Support for smaller vessels
Integration – improved snapshot identification to enable display of snapshots on the report tab
3. TomTec Integration – We have updated to the latest released version of the TomTec product into VitreaAdvanced, TomTec-
Arena 1.1. This integration update addresses defects and issues within the software.
2. Reporting Templates
Six NEW report templates
o Portrait: 1 x 2, 1 x 3, 2-up, 3-up, and Portrait Run-off layout
o Landscape: 3-up, 4-up
Minimized “white spaces” and reduced print margins
Optional header/footer information
Thumbnail display for easy image sizing assessment
General Serviceability
The following “Serviceability,” in terms of installation and updates of Vital products have been implemented:
System Upgrade/Install Enhancements – “Serviceability” of Vital products represents a key area of focus for improvements. We
have made the following changes in an effort to simplify and reduce time during installation.
System Check Tool that validates settings before installation eliminating time-consuming trouble shooting and starting over.
Available for VIMS, VitreaAdvanced and VitreaCore Server
Upgrade Version Messaging. Dialog box pop-up detects version restrictions when upgrading to version 6.6
VitreaCore Installer support for Internet Explorer 10 and Windows 8
The following features, functionality and enhancements have been incorporated into VitreaAdvanced, VitreaExtend, and
VitreaWorkstation products:
Translated User Interface and documentation
For the 6.6.3 software release, the User Interface and the Education and Reference Guides will be translated in multiple languages.
Refer to the Help tab to view the translated documents.
VitreaExtend
The VitreaExtend software is available as of the 6.6.1 release
Known Limitations
As of
Products Description
Version
6.6.1 VitreaExtend Unsupported Monitor Resolution may result in degraded performance
The VitreaExtend software supports the following monitor setups:
a single 24" standard format monitor with a supported resolution of 1920x1200.
a single 30” monitor with a supported resolution of 2560x1600.
Performance may suffer if larger than 2560 X 1600 monitor resolution is used.
6.6.1 VitreaExtend VitreaExtend does not support multi-monitor configurations
6.6.1 VitreaExtend Unsupported functionality on remote sessions
The following functionalities are not supported on remote sessions:
Import from Media (Network/Desktop/CD/DVD)
Export to Media (Desktop/CD/DVD)
Save As feature for snapshot PNGs
Export to Media for non-DICOM data (CSV, EP files)
6.6.1 VitreaExtend Colon auto flythrough may cause image tearing on the flythrough screen
VIT-10485 This issue occurs when you launch a colon study using VitreaExtend, select the Colon protocol and
Flythrough preset, and select the Auto Flythrough option in the upper left corner of toolbar. As a
result, the image may produce tearing on the flythrough screen.
NOTE: This issue is seen during display only and captured images are not affected. This issue only
affects VitreaExtend Windows XP clients.
6.6.1 VitreaExtend Spin rotation may cause image tearing
VIT-11152 This issue occurs when you launch a Cardiac or Vascular study using VitreaExtend and perform a spin
rotation on the 3DVR image. As a result, the image may produce tearing.
NOTE: This issue is seen during display only and captured images are not affected. This issue only
affects VitreaExtend Windows XP clients.
6.6.1 VitreaExtend A 1 Gbit/s connection between the remote client and the server is recommended (you may
experience reduced performance if a lower connection speed is used)
6.6 VitreaWorkstation Using side clicks with a tilt-wheel mouse causes application to crash
VitreaAdvanced System crashes have been experienced with tilt-wheel mice performing a side click action. As a
SUSENG-4044 workaround, use a basic mouse.
6.6 VitreaExtend Internet Explorer 11 is not a supported Internet browser for VitreaExtend Host System
deployments
6.6 VitreaAdvanced Siemens scanners that acquire three slice CT Brain Perfusion are not supported
VitreaWorkstation
VitreaExtend
As of
Products Description
Version
6.6 VitreaAdvanced Circle of Willis (COW) Protocol: Jog Shuttle does not build one volume
VitreaWorkstation The shuttle jog builds two volumes instead of one volume. Jog Shuttle is not supported in the COW
VitreaExtend protocol.
6.6 VitreaAdvanced Lung Density Analysis gives an incorrect result for patient with only one lung
VitreaWorkstation When you use the Lung Density Analysis preset for a patient with only one lung, the lung is divided into
VitreaExtend a left and a right portion and the results are incorrect.
As of
Products Description
Version
6.6 VitreaAdvanced NAS path is permanent
VitreaWorkstation The Network-Attached-Storage (“NAS”) path is permanent. Customers that wish to replace or rename
the NAS path must keep the original hostname or reprocess all patient data.
6.6 VitreaAdvanced Vitrea Review Tab
The Vitrea Review tab is not available in VitreaAdvanced. This functionality is currently offered through
the ‘Report Tray’ in the Data Manager to support report distribution.
6.6 VitreaAdvanced Study Directory Access
The VitreaWorkstation Study Directory is not available from VitreaAdvanced. Key features and
functionality that are only available on the standalone workstation include – Study Viewer and Disk
Space Indicator.
6.6 VitreaAdvanced Sub-Volume Loading not available in VitreaAdvanced
The Vitrea Study Directory supports the ability to select discrete points within a study’s 3D volume for
loading into the viewer. This capability enables users to manage local resources by only loading a
subset of the data and is not available in VitreaAdvanced.
6.6 VitreaAdvanced Endovascular Planning: Exporting files to media may not save
This following occurs when in VitreaAdvanced on a server or on Vitrea Powerstation:
1. Insert CD into drive.
2. Load dataset into Cardiac: EP Planning protocol, 3D Analysis preset.
3. After segmentation is complete, click on Save 3D button in lower left hand corner and wait for
“Save Done” message to display in lower right hand corner.
4. Go to the Report page or study directory.
5. From the Report page, click on burn CD/DVD. From the Study Directory, right-click on a dataset
and choose Save to media.
6. The option box is grayed out and files are not found in patient directory.
The Endovascular Planning 3D Model is for standalone workstations only.
6.6 VitreaAdvanced 3D image appearance differences when rendered by GPU vs. CPU
VitreaWorkstation The appearance of 3D images can be different when rendered with a dedicated GPU versus CPU-
based rendering.
6.6 VitreaAdvanced Multi-volume fusion limitations with older GPU’s
VitreaWorkstation Multi-volume fusion of more than three volumes is not supported on older graphic cards (Examples:
Quadro FX3500 and ATI FirePro V5800). Vitrea will only allow three volume Multivolume fusion with the
older graphic cards and will allow up to four volumes with the newer graphic cards.
6.6 VitreaAdvanced System Administration issue: Duplicate patientID entered with slightly different name
If two studies with the same patient ID and different names (such as one with a middle name and the
other without) are sent to Vitrea cache and auto-delete rules specify that the two studies should be
deleted, only the newer study will be deleted. The other study is orphaned.
A system administrator can delete the orphaned study folder manually.
6.6 VitreaAdvanced Internet Explorer 11 is not a supported Internet browser for the 6.6 software release
VitreaWorkstation
VIM-1640
As of
Products Description
Version
6.6 VitreaWorkstation Shadows display in CTA/Aorta when Auto Bone or Remove Fragments are used with CPU
SUSENG-2880 Rendering
The Auto Bone or Remove Fragments options when used with CPU rendering can cause ghosted or
shadow remnant of ribs in some cases. As a workaround, use the Dilate feature in the Regions
Options menu to improve the 3D image quality.
6.6 VitreaAdvanced .NET 4.5 is not supported for the 6.6 software release
FRDB-1199 Do not install .NET 4.5 on a VIMS Server or it will close unexpectedly.
6.6 VitreaAdvanced Brain Perfusion - Summary Map options do not save as part of Modified Preset
VIT-6815 For Brain Perfusion studies loaded into VitreaAdvanced, Summary Map options or controls are not
saved as part of Modified Presets.
6.6 VitreaAdvanced SQL Express database size limits the total DICOM data storage utilized by the system
VitreaWorkstation Any system where SQL Express 2008 or older is used and where more than 1TB of storage is used,
VIM-1685 you are limited to using 1TB or less of storage for your DICOM data.
As a workaround, upgrade your Windows Operating System to Windows 7 for a VitreaWorkstation or
upgrade to Windows Server 2008 R2 for a VES system. Then, SQL Server 2008 R2 will be used and
you can utilize up to 2TB of storage.
6.6 VitreaAdvanced Coronary Subtraction – Studies with multiple subtraction series and CTA series do not display
VitreaWorkstation correctly in VES
VIT-10900 You are unable to launch a sub series and corresponding CTA series into the subtraction application
using the Series tab on VES. Fortunately; you can launch into the Coronary Subtraction application
using the select Application data option from the Application tab on VES. VitreaWorkstation has no
issues with the data and launches and groups properly.
As of
Products Description
Version
6.6.3 VitreaAdvanced Custom template with custom logos or icons are lost during an upgrade
VitreaWorkstation This issue occurs when you complete the following workflow:
SUSENG-4077 1. Create a custom template with logos or icons for a specific application
2. Upgrade site from previous version to later Vital version
3. Load patient data and attempt to create a report with the previously generated template
As a result, the custom template, institution logos, or icons may not be retained after an upgrade. As a
workaround, Vital Customer Support should backup custom logos and icons before an upgrade and
use the report template backup to restore customer reports. An alternative is to recreate the custom
reports manually.
6.6.2 VitreaAdvanced TAVR Annulus Size checkbox may be grayed out
VitreaWorkstation This issue occurs when you complete the following workflow:
SUSENG-4043 1. Load a TAVR series into Vitrea
2. Click “Annulus Size” in the measurements section
3. Double click within the axial view to automatically select the Annulus
As a result the checkbox for the Annulus Size is grayed out and not selectable. As a workaround, right
click the Annulus contour in the axial viewer and select Annulus Size. The checkbox is now active and
can be selected.
6.6.2 VitreaAdvanced Restoring a Myocardial CT snapshot does not restore the heart rate value
VitreaWorkstation Snapshots taken within Myocardial CT do not restore the Heart Rate value in the viewer. As a
SUSENG-11745 workaround, edit any of the restored measurements followed by a re-compute or undo the edit. This
will recalculate the heart rate and display the value.
6.6.1 VitreaExtend Brain Perfusion Summary Map measurement values do not display on 2D perfusion
Brain Perfusion Brain Perfusion Summary Map volumes cannot be computed when the series contains only a single
VIT-11456 slice.
6.6.1 VitreaExtend VitreaExtend closes unexpectedly when you load Coronary Subtraction datasets that include
VIT-11485 secondary images
This issue occurs when you select a study that includes a secondary image (e.g., batch or report) in the
Study Directory and click the Load Volume button. As a result, VitreaExtend closes unexpectedly.
As a workaround, load the CTA volume and sub-volume. Do not select and attempt to load the
evidence with the CTA volume and sub-volume.
6.6.1 VitreaExtend User cannot mark TAVR measurements as done
VIT-11256 Double-clicking in an axial image creates an ROI. However, the name selected in the measurement list
does not become active so the user cannot mark the measurement as done.
As a workaround, right-click the measurement and assign the name.
6.6.1 VitreaExtend For remote sessions, movie previews in Report Findings tray are being cut short
VIT-11243 The Windows Media Player will close earlier than expected when being run on Remote Session.
As a workaround, the complete movie file can be previewed from the console or you can set the
Windows Media Player to loop.
6.6 VitreaExtend Compression fails for CD/DVD burning
VIT-3632 If you try to burn a CD/DVD from the Report page, the compression fails. Vital has disabled DICOM
compression for CD/DVD burning from the Report page.
6.6 VitreaExtend ROI tool is disabled for single volume in Calcium Scoring
VIT-10735 This issue occurs when you load a single volume into the Calcium Scoring CT: 2D Vscore protocol. The
ROI and Ellipse tools are disabled. If this happens, load the study into a generic protocol and perform
an ROI.
As of
Products Description
Version
6.6 VitreaWorkstation ALI Compression fails for CD/DVD burning
VIT-3632 If you try to burn a CD/DVD from the VitreaWorkstation Report page, the ALI compression fails. For the
6.6 release, Vital has disabled DICOM compression for CD/DVD burning from the VitreaWorkstation
Report page.
6.6 VitreaAdvanced VitreaCore did not open when you logged into McKesson 11.6 as an advanced user and tried to
VitreaWorkstation launch VitreaAdvanced Server when none was available
VIC-965 This issue occurred when you completed the following workflow:
1. Log into McKesson as an advanced user when all VitreaAdvanced servers are already in use.
2. In the McKesson client, right-click a study that is already on the VIMS, and select Open Study.
Images display.
3. Right-click an image and select VES.
4. Select Load Series in Vitrea WEB. The Data Manager displays.
5. Select the Series tab. On a series thumbnail, click the VitreaAdvanced button.
As a result, a message displays stating none of the VitreaAdvanced Servers are available. However,
VitreaCore did not open. As a workaround, right-click the series thumbnail and select Load.
As of
Products Description
Version
6.6 VitreaAdvanced Single Brain Perfusion image doesn’t display in 4-up layout
VIT-3879 This issue occurs when you load a single slice perfusion dataset into the generic protocol using a 4-up
layout. The axial image is not visible until you increase the slice thickness, then it displays as expected.
6.6 VitreaAdvanced Cardiac CTA – After restoring a snapshot, the heart rate is not displayed
VIT-10819 This issue occurs when you complete the following workflow:
1. Load one or two volumes of a cardiac CTA study that has heart rate in the DICOM tags.
2. Select the Cardiac Myocardial CT protocol and either the Myocardial Perfusion or Myocardial
Analysis preset.
3. Wait for segmentation and computing to be complete. Make sure the heart rate is displayed.
4. Take a snapshot.
5. Restore the snapshot from the Study Directory or Report page, and wait for segmentation and
computing to be complete.
As a result, the heart rate is not displayed. As a workaround, if you do an edit and then click compute
again, the heart rate will reappear.
As of
Products Description
Version
6.6 VitreaAdvanced Media Manager import fails with extremely large DICOM datasets
VIT-6268 This issue occurs when you attempt to import extremely large .dcm files.
6.6 VitreaAdvanced Liver Analysis – Preset selector options for fusion not working
VIT-2206 This issue occurs when you follow these workflow steps:
1. Load multiple series in resection planning preset.
2. Segment the liver, portal vein (Portal or Venous phase), and hepatic artery (Arterial phase).
3. Go to portal phase and select the rigid fusion option.
4. Toggle between various preset selector options in portal phase.
5. Select fusion phase and toggle between various preset selector options
As a result, you see no difference between the liver solid option and liver vessels, and you can no
longer select the inverted MiP option.
6.6 VitreaAdvanced Olea Sphere Cannot Perform Computations on Compressed DICOM Transfer Syntax
VIM-1961 If data is OleaSphere studies are sent to the VIMS in Compressed format, the software fails to perform
as Olea does not support JPEG lossless compressed data. To avoid this issue, send the data to Vitrea
using non-JPEG compressed transfer syntax and the data sent can be opened within Olea Sphere.
6.6 VitreaAdvanced This issue occurs with the following workflow steps:
VIM-1949 1. Load a series into Vitrea.
2. Select Generic CT and 3D preset.
3. Apply a Denoising preset.
4. Select the Save New Series option.
As a result, the new series is available in the study directory, but the volume is not built.
6.6 VitreaAdvanced Anticipatory Processing: MPR Autobatch does not skip old priors for batching
PLT-5064 This issue occurs with the following workflow steps:
1. Set up to pre-fetch the last three prior exams from archives that are older than 5 years.
2. Push the current dated study to VitreaAdvanced and let it pre-fetch the old prior.
3. Make sure the old studies have dates older than defined data script.
1. As a result, the AP batches every prior study even though the date is outside the desired max age of
studies to be batched. As a workaround, check the study date time in the script.
6.6 VitreaAdvanced VG3 Server “Flutters” due to use of local machine time stamps
PLT-5108 The VG3 servers (vitrea-license, vitrea-home, etc.) register themselves as online automatically. The
PLT-4455 local time stamp is used during registration. The servers are also marked offline automatically if they
have not "checked in" in a reasonable time. However, if the time stamp on different machines is off by
more than a minute, the servers will be accidentally marked offline.
This behavior causes "fluttering" of the services because they are marked offline and online very
frequently. In one instance, the services were being erroneously marked offline 41 times a minute.
This behavior will fill the logs and audit table with erroneous data quickly.
The problem is that the system uses the online status for some of the services. If the license server is
"down", Vitrea View will not tell eViz the URL, and then eViz will raise a "License Error" to the user. To
the user, it appears the system is down.
The solution is it to use the central DB timestamp instead of the local machines.
6.6 VitreaAdvanced Multi-Chamber Cardiac Functional Analysis displays blue box in VR viewer
VIT-6090 The issue occurs when you complete the following workflow steps:
VIT-3850 1. Load a multiple phase cardiac dataset.
2. Select Cardiac Function CT Protocol and 3-chamber preset.
3. Wait for segmentation to complete and then right-click on the LV region and select Show Volume.
As a result, a blue box displays on the image. This extra blue box is a rendering artifact and does not
inhibit the viewing of the anatomy. As a workaround, you could trim the affected part of the image
where the artifact displays.
As of
Products Description
Version
6.6 VitreaAdvanced MINT becomes unresponsive and requires a restart
SUSENG-3779 MINT becomes unresponsive and started having I/O Socket errors which then caused deadlock, and
time out to other VIMS services.
SUSENG-3767
6.6 VitreaAdvanced License Availability Report displays a gray screen
VIM-1867 Run a License availability report searching for the last day, and the results of the query returns a gray
screen. This is expected behavior if there are no events where there is a lack of license events. The
report should display a graph with values of the query if the systems detects that “lack of license
events” have occurred.
6.6 VitreaAdvanced Snapshots display with Red X on report page
VitreaWorkstation This issue occurs when you complete the following steps:
1. Setup your NAS sharename to be patients.
VIT-9828 2. Install VIMS using the NAS share name you created in step 1. NAS path should look something
SUSENG-3408 like this \\VIMSHOSTNAME_or_NASHOST\patients\.
3. Install VitreaCore. You can also install an advanced server, but not required to reproduce.
However, and advanced server will give you better error reporting on this issue.
4. Log into VitreaCore.
5. Load into VitreaCore viewer or VitreaAdvanced viewer.
6. Take a snapshot or batch.
7. Click on the Report tab.
The snapshots do not display in the Report page, and there is a red X in its place. The snapshot does
display within the snapshot tray in VitreaCore. Select the Report page tab and see the following error,
"System.Net.WebException: The remote server returned an error: (404) Not Found..."
As a workaround, recreate a new NAS share without using patients as the sharename. Remove VES,
start with a new blank database, reinstall VES, and re-push all DICOM data back to VIMS.
Resolved Issues
Fixed
Products Description
Version
6.6.3 VitreaAdvanced Custom reports are replaced with original report after IIS or system restart
VitreaWorkstation This issue occurred when a custom report template was created. After an IIS restart or reboot of the
VIT-12334 system, the custom report page would no longer be available. This issue has been resolved.
Fixed
Products Description
Version
6.6 VitreaAdvanced Carotid vessel contours could not be edited after probing vessels in another study when
VitreaWorkstation SUREPlaque was not licensed
SUSENG-2738 Previously, this issue occurred when you completed the following workflow with a license that did not
include SUREPlaque:
VIT-10362
1. Loaded an abdomen or a peripheral study; probed a vessel, and closed the study.
2. Loaded a carotid study, probed a vessel, and edited the contours. When applying the contour
changes, the contours would do not display or the first/last contours would display only.
This issue is resolved.
6.6 VitreaAdvanced Using “Aortic Neck Angle” measurement listed within the Measurement findings list did not
VitreaWorkstation properly update after centerline edit
VIT-10030 Previously, this issue occurred when you completed the following workflow:
VIT-10090 1. Launched a study using the TAVR or Stent Planning protocol and preset.
2. Created an angle measurement in the longitudinal inset, based on a grayed out angle
VIT-10052 measurement in the list.
VIT-10051 3. Edited the centerline and clicked OK.
4. Viewed the angle measurement listed at the bottom of the inset.
As a result, the aortic measurement result did not update. This issue is resolved.
6.6 VitreaAdvanced VI RollingFile Logging failed to roll over if the log was being actively written to
SUSENG-3743 This issue occurred when you actively logged to any log file using the RollingFileStream and waited
PLT-4879 until the log reached the rollover limit that was specified in MB by maxFileLength. As a result, the File |
Move operation failed and produced a runtime exception from the Log command that was currently
executing. This issue is resolved.
6.6 VitreaAdvanced Zoom buttons perspective projection were not functioning and have been fixed
VitreaWorkstation This issue occurred when you rendered in 3D in perspective mode and clicked the Zoom button in the
VIT-2763 3D image. As a result, the 3D image did not display in the Viewport. This issue is resolved.
6.6 VitreaAdvanced Vessel names did not display in the MPR views
VitreaWorkstation This issue occurred when you used the vessel probe tool. Previously, the vessel name would not
SUSENG-3248 display in the MPR views. This issue is resolved. The vessel name will now display in the MPR views
VIT-9801 for all of the vessels.
6.6 VitreaAdvanced CT Multi-Chamber Cardiac - undo short-axis edit computed Polar Map incorrectly
VitreaWorkstation This issue occurred when the polar map was re-computed when the Undo button was clicked. For the
6.6 release, the Undo button is disabled for the 3-Chamber preset. It remains enabled for all other
VIT-10000 protocols.
VIT-10053 1. Load multi-phase cardiac dataset in the 3-chamber Cardiac Functional Analysis application
2. Select the 5-up viewer with polar maps
3. Rotate the cardiac sectors in the short-axis viewer using the LV Axis edit tool
4. Then select the undo button at the bottom of the screen.
This issue is resolved by disabling Undo in the 3-chamber preset.
NOTE: Due to a correction in the way myocardial contours are sampled for calculations, the numbers
and colors displayed on the polar maps of a restored snapshot may differ from the way they were
displayed when the snapshot was taken. The degree to which the calculated results can differ is
roughly equivalent to a 1-2 mm adjustment of the mitral valve plane in either end-systolic or end-
diastolic phases.
6.6 VitreaAdvanced CT Liver Resection Planning Base region was visible after segmenting the liver with the Pick
VitreaWorkstation tool
VIT-6956 This issue occurred any time the liver region was created using the liver pick tool. After segmentation,
the Show check box remained selected for the base region and the liver could not be seen well. This
VIT-8571 issue is resolved.
Fixed
Products Description
Version
6.6 VitreaWorkstation EP Planning - Save 3D button was present but did not save EP files on Powerstation
VIT-8474 Previously, the Save 3D button would display on a Powerstation, however, the Save 3D feature is
specific to VitreaWorkstations. For the 6.6 release, the Save 3D button no longer displays when using a
Powerstation, as this functionality is not supported on a Powerstation.
6.6 VitreaAdvanced Airway Vessel choice in the Region Color Selector menu did not function
VitreaWorkstation Previously this issue occurred when you performed the following workflow:
VIT-7007 1. Loaded a study into Generic CT/3D.
2. In the Anatomy List, clicked on the Options dropdown menu in the Base region and selected
Change Colors.
3. In the Region Color Selector popup menu, scrolled to the bottom and selected the Airway
Vessel option.
As a result, the Airway Vessel option in the Region Color Selector menu did not function.
Specifically, the space the button occupied on the user interface did not function. This issue is resolved.
6.6 VitreaAdvanced Crosshair on MPR images while scrolling with mouse did not release the full crosshairs when
VIT-4129 finished scrolling
This issue occurred when scrolling through MPR images by spinning the wheel on the mouse. When
you scroll using the wheel it is expected that the crosshairs on the image stay filled in unless you stop
scrolling or click somewhere in the image. When you stop scrolling the crosshairs should not be filled
in. This issue is resolved.
6.6 VitreaAdvanced Users may have assumed volumes and EF percentages were calculated based on the contoured
SUSENG-2483 region rather than the tinted region
Previously, it may have been unclear to users that the LV volume was based on the tinted LV region.
This issue has been addressed by turning the region tint back on, to illustrate that the tinted region
does not necessarily match the contour. An informational icon has also been added to remind users to
verify the tinted region because this is the volume used to calculate the Ejection Fraction and Stroke
Volume.
6.6 VitreaAdvanced Vitrea Administration Tool User Audit Report did not show VitreaCore current sessions
VIM-1919 This issue occurred when you loaded a series in VitreaCore, launch the Vitrea Administration Tool, and
selected the User Audit Report page. As a result, the current session was not listed. It displayed once
the study was closed. This issue is resolved.
6.6 VitreaAdvanced Error seen printing Calcium Scoring graphs and table images
VIT-6348 Previously this issue occurred when you performed the following workflow:
1. Scored one of the vessels and took a snapshot.
2. Selected the Report tab. Clicked on an image template that was a 4:1 and replaced the current
calcium scoring template displayed.
3. Dragged the graph, table and snapshot to the image template on the right.
4. Clicked DICOM print.
Instead of printing as expected, an error displays and the page does not print. This issue is resolved.
6.6 VitreaAdvanced Logging into VitreaCore with an LDAP account took 30 seconds
VIC-942 This issue occurred when you logged into VitreaCore using an LDAP account, you were delayed by up
to 30 seconds before the Study List displayed.
Previously, Vital used to get all Groups that a user was associated with and the search was based at
the Root of the Domain Controller. For the 6.5.5 release, Vital only tests for Group memberships we are
interested in for the user and the search is based at the User node. This issue is resolved.
6.6 VitreaAdvanced Times in the VitreaCore User Auditing page were not in local time, but in GMT
VIC-902 This issue occurred when you loaded a study in VitreaCore, noted the local time or computer time, then
went to the VitreaCore Administration page, selected the User Auditing page, and made a query using
the current date. As a result, the loaded study was listed under GMT time. This issue is resolved.
Fixed
Products Description
Version
6.6 VitreaAdvanced Created a batch with Russian series description and export, the series description showed ???
VitreaWorkstation characters in the Study directory
VIT-6483 Previously this issue occurred when you performed the following workflow:
1. Loaded a dataset and selected the 3D Preset.
2. Selected the Batch tab, clicked on any MPR, scrolled to a different slice, and clicked on the End
button.
3. Entered a series description with Russian characters.
4. Selected the Show cover image with label checkbox.
5. Entered a Label that contained 56 characters with Russian Characters and special characters and
numbers such as &, *, (, ), $, @, !, etc.)
6. Clicked on the Batch button.
7. Selected the Study directory and Exported to another Vitrea/VES destination.
As a result, the series description showed ??????? in place of the Russian characters. This issue is
resolved. With the 6.6 release, Vital does not allow any non-ASCII characters in the series description.
6.6 VitreaAdvanced Report page was cut off with a client resolution of 1280 x 1024
SUSENG-3708 Previously this issue occurred when you launched an Advanced session on a VES 6.6 system, set the
client resolution to 1280 x 1024 (the documented minimum client resolution), and accessed the Report
page. As a result, the bottom portion of the Report page was cut off and buttons were missing and
could not be selected. This issue is resolved.
6.6 VitreaAdvanced Exported Olea color maps were highly pixilated when viewed on PACS
VitreaWorkstation Previously the exported batches were not exported in the same resolution as they were displayed
SUSENG-3339 creating images that looked pixelated. This issue is resolved.
6.6 VitreaAdvanced Error occurred when loading Cardiac CFA data in multi-chamber
SUSENG-3627 This issue occurred when you loaded a multiphase Cardiac dataset, selected desired the phases,
selected the Cardiac Functional CT protocol, and picked 3-Chamber Functional Analysis. As a result, a
“MYOCARDIUM SEGMENTATION FAIL” error displayed prior to the segmentation being completed.
This issue is resolved.
6.6 VitreaAdvanced Auto Export could not send images as a single series
McKesson This issue occurred when you launch through McKesson Integration into a patient, created snapshots
SUSENG-3505 and batches, and clicked the Back to PACS button. As a result, each snapshot image were sent as its
own series and no longer stacked on the PACS. This issue is resolved.
6.6 VitreaAdvanced Remove Fragments caused Vitrea to close unexpectedly
VitreaWorkstation Previously this issue occurred when you performed the following workflow:
SUSENG-3734 1. Loaded a carotid study and selected the Vascular: Carotid CT protocol and 3D Analysis.
2. Selected the Use modified presets check box and cleared the Automatic bone removal check
box.
3. Trimmed the front and sides of study so that carotids were readily visible.
4. Adjusted window/level to about 125/275.
5. Clicked on Remove Fragments.
6. Clicked AddTo: Recycle Bin.
As a result, Vitrea closed unexpectedly. This issue is resolved.
6.6 VitreaAdvanced Different Mirada Saved sessions were listed on the VitreaAdvanced Servers
Oncology Fusion This issue would appear random to the end user. If the first file encountered in the file directory was a
SUSENG-3563 non-DICOM file, the DICOM related file used to restore the saved session was not processed correctly
and Mirada would not display or restore the saved session. If the first files encountered were Mirada
saved sessions, the issue would not occur and the series was processed correctly. This issue is
resolved.
Fixed
Products Description
Version
6.6 VitreaAdvanced Silent VIMS install on Windows Server 2008 with SQLEXPRESS database instance failed to
SUSENG-3415 complete
This issue occurred when you attempted to install VIMS silently using Windows Deployment Server
running Microsoft Deployment Toolkit when the SQL instance was VIMSSQLEXPRESS. As a result,
the installation failed. This issue is resolved. For more information, refer to the VIMS Install Guide.
6.6 VitreaAdvanced Snapshots and batches not sent in order selected when exported from Report page
VitreaWorkstation Previously this issue occurred when you performed the following workflow:
SUSENG-3713 1. Loaded any study in any protocol.
2. Took a series of snapshots and batches.
3. Selected the Report page.
4. Clicked and then CTRL+clicked the snapshots and batches in the order you wanted them sent to
PACS.
5. Clicked on the Export button. The images displayed in the export window in the order selected.
6. Selected Export as single series.
7. Exported to PACS
As a result, images were not sent in order selected on the Report page. This issue is resolved.
6.6 VitreaAdvanced Dynamic MR license was limited to one specific VitreaAdvanced Server
SUSENG-3592 Previously this issue occurred when you performed the following workflow:
1. Licensed Dynamic MR for one license and added two VitreaAdvanced Runtime licenses on a
system with two available VitreaAdvanced Servers with Dynamic MR installed.
2. Logged into VitreaCore.
3. Clicked on a Dynamic MR study.
4. Selected the Applications tab.
5. Double- clicked on the Visia Dynamic Review icon to load into an advanced session.
6. Exited this session.
7. Opened another Visia Dynamic Review session.
As a result, the Dynamic MR license was not checked out on the VIMS Server. This issue is resolved.
6.6 VitreaWorkstation Query time constraint dropdown did not work on 6.5.1
SUSENG-3409 This issue occurred when you attempted to query a PACS for a valid patient ID with both old and new
studies, selected last week from the Time Constraint dropdown, and clicked Query. As a result, the
results would only display if you selected any time from the Time Constraint dropdown. This issue is
resolved.
6.6 VitreaAdvanced Vital Start menu Links started an install in a load balanced environment
SUSENG-3050 This issue occurred when you upgraded a 6.3 load balanced system to 6.4.1, used the VIP as the
VIMS hostname, and used the Start menu links to access the Application Editor and the VIMS Monitor.
As a result, a repair was started. This issue is resolved.
6.6 VitreaWorsktation Study level query results showed the wrong number of images in VitreaWorkstation
SUSENG-3344 Previously when you queried a system that did not support “number of images in the study” in response
to a cFind query, the image count displayed on the workstation was incorrect. However, if you queried
a different system that supported number of images in the study, the correct image count was
displayed. This issue is resolved.
6.6 VitreaAdvanced Medis Contours saving to all sequences even when contours were edited
SUSENG-2908 This issue occurred when you loaded any Cardiac MR study for Medis QMass/QFlow, double-clicked
PIN-1420 the QMass icon, created contours, and either clicked the Save button or selected File | Save to Vitrea.
As a result, it saved the contours to all sequences and not just the sequence in which the contours
were created. This issue is resolved.
6.6 VitreaAdvanced Medis Quantitative Flow Velocity values display gray and are not easily seen on graph
SUSENG-3589 This issue occurred when you calculated the quantitative flow velocity using Medis software, and
generated the peak velocity graph. The line on the graph is gray and poorly visible on the flow curve
graph. This issue is resolved.
Fixed
Products Description
Version
6.6 VitreaAdvanced Wrong MTT CBF maps on specific Vein Input Function location
VIT-11018 Previously, if there was a temporary surge in contrast of the (automatically or manually) selected
VIT-10852 vein, the perfusion calculation may have incorrectly omitted all the data after that surge, which
SUSENG-3773 resulted in incorrect perfusion maps. The temporary surge was visible as a small peak in the vein
curve that preceded the primary peak.
Shown below is an example of the graph in which the phenomena occurred.
This issue is resolved. The perfusion miscalculation no longer occurs. For additional information,
please contact Vital Customer Support.
6.6 VitreaAdvanced Cardiac Pediatric CFA application is not available unless licensed
VitreaWorkstation The pediatric multi-chamber CFA application will not be available in the Application tab unless it is
VIC-956 licensed.
6.6 VitreaAdvanced WW/WL values missing after doing a batch in VitreaCore using Enhanced DICOM
SUSENG-3647 This issue occurs when you create a batch in VitreaCore using enhanced DICOM. The results are that
the WW and WL values are not in the DICOM tags of the batch. This issue is resolved.
6.6 VitreaAdvanced Closing the Batch dialog box is slow
SUSENG-3646 This issue occurred when you create a batch in VitreaCore and the uploading files window takes a long
time to close. This issue is resolved.
6.6 VitreaAdvanced Missing 3D Anatomy with Vessel Probe trim & Vessel Grow tools
VIT-9978 This issue occurred when you probed a vessel, then extended the vessel using the trim or vessel grow
SUSENG-3367 tool. In the 3D image, there are areas that are not filled in where the vessel was extended, even though
the vessel tint showed that area was selected. This issue is resolved.
6.6 VitreaAdvanced Repeated log messages during ingestion of large studies caused system slowness
VitreaWorkstation Unneeded log messages were being inserted into the database during the down load of large studies.
SUSENG-2886 These have been eliminated and this issue is resolved.