You are on page 1of 70

EKG2000

SERVICE MANUAL
Version 1.0

www.ebionet.com
Service Manual of Cardio 2000

Revision History
Edition Date Page Comment

Ver 1.0 2019-10-18 Initial Release

C7SM-1.01
2/70
Service Manual of Cardio 2000

- TABLE OF CONTENTS –

Section 1. General Information ............................................................................................... 5

1-1. Introduction...................................................................................................................................................................... 5
1-2. Features of the Equipment ...................................................................................................................................... 5
1-3. Equipment Description .............................................................................................................................................. 6
1-4. System Installation .....................................................................................................................................................15
1-5. Terms of Warranty .......................................................................................................................................................16
1-6. Specification .................................................................................................................................................................16

Section 2. Troubleshooting .....................................................................................................18

2-1. Power .................................................................................................................................................................................18


2-2. Wave Printing ................................................................................................................................................................19
2-3. Diagnosis .........................................................................................................................................................................22
2-4. Filter ....................................................................................................................................................................................23
2-5. Trouble Shooting Guide ..........................................................................................................................................25

Section 3. Adjustment .............................................................................................................31

3-1. LCD Backlight................................................................................................................................................................31


3-2. Lead Fault ........................................................................................................................................................................32
3-3. Filter Setting ....................................................................................................................................................................33

Section 4. Block Diagram/Description ................................................................................34

4-1. Circuit Block Diagram ..............................................................................................................................................34


4-2. Equipment Overview ................................................................................................................................................35
4-3. Size and Weight ...........................................................................................................................................................38

Section 5. Disassembly and Assembly ................................................................................39

5-1. Module Description ...................................................................................................................................................39


5-2. Internal Block Diagram & Cable ........................................................................................................................46
5-3. Disassembly ...................................................................................................................................................................49
5-4. Assembly .........................................................................................................................................................................59

C7SM-1.01
3/70
Service Manual of Cardio 2000

Section 6. Replaceable Parts List and Accessory .............................................................60

6-1. Replaceable Part List ................................................................................................................................................60

Section 7. Connector Pin Assignment .................................................................................63

7-1. Patient Cable Port ......................................................................................................................................................63

Section 8. PCB Part Disposition ..............................................................................................64

8-1. Base (Power+Analog) Board................................................................................................................................64


8-2. CDUI Board .....................................................................................................................................................................65
8-3. CPU Board .......................................................................................................................................................................66

Section 9. Software Upgrade .................................................................................................67

9-1. Requirement of Upgrading....................................................................................................................................67


9-2 Procedure of Upgrading (Cardio7 UI screen, upgrade method is same) .................................68

C7SM-1.01
4/70
Service Manual of Cardio 2000

Section 1. General Information

1-1. Introduction

CardioCare2000 is an electrocardiogram (ECG) device capable of measuring and recording 12


channels of cardiac signals of patients. The device enables operators to record cardiac signals,
review them using various viewing modes, and perform automatic detection of cardiac conditions.
In addition, the device allows operators to enter patient information for it to be printed alongside
cardiac signals, so that operators can sort and manage charts conveniently. It is also possible to
manage the digital files by transferring the stored data to a networked PC.
Furthermore, user convenience has been further enhanced by addition of a functionality that
can initiate ECG recording & saving, filter enhancing, parameter sorting and automatically
detecting cardiac signal all at once by pressing one key.
A battery (optional) can be added to the device so that the device can be operated
conveniently during bedside visits or emergency conditions.

1-2. Features of the Equipment

- Print 12-channel (ch) cardiac signals with reports in: 3ch+1 rhythm, 6ch+1 rhythm, and 12ch
rhythm patterns on A4 or Letter size printing paper.
- Record and print 12 channel rhythms continuously and simultaneously.
- Calculate heart rate, P-R-T axis and PR, QRS, QT and QTc widths from recorded signals
automatically and print them on a report for use in rhythm analysis.
- Get diagnostic reports using automatic detection functionality.
- Able to modify filter setting, signal sensitivity, printing speed, channel view settings and rhythm
settings, and print on previously recorded EKG signals to aid data analysis.
- Able to attach a battery so that the device can become portable.
- More than 200 patient‟s data can be saved and transferred to a PC through LAN.
- It offers various protocols to make it possible to link with hospitals‟ computing networks. Also File
Databases are strengthened.

C7SM-1.01
5/70
Service Manual of Cardio 2000

1-3. Equipment Description

The CardioCare2000 system consists of the items below. Unpack the package and check the
items below are included. Also, be sure to check for any damage to the body and accessories.

Basic Components and Accessories

① CardioCare2000 Body (1 EA) – Dimension: 290(W) 300(D) 97.5(H)mm

② Patient Cable (1 EA) – Length: 3,700mm (Max)

③ Limb Electrodes (1 SET)

④ Chest Electrodes (1 SET)

⑤ ECG Paper (1 EA)

⑥ Power Cable (1 EA) – Length: 3,000mm (Max)

C7SM-1.01
6/70
Service Manual of Cardio 2000

Optional Components

① Battery (1 EA) – Replaceable and Rechargeable, Lithium ion, 11.1V, 2600mA

② Hanger (1 EA)

③ Cart (1 EA)

CAUTIONS

You may have distortion or signal noise when you use nonstandard or other brand
accessories. We strongly recommend you use only the authorized accessories which we
supply.

WARNING

How to replace battery: Please make sure you use the right battery we show here.

Otherwise we are not liable for any damages and/or explosion/fire caused by using the

wrong battery.

C7SM-1.01
7/70
Service Manual of Cardio 2000

Body Configuration

▣ Top View


② ③
① Handle

② Printer Cover

③ Printer Cover Switch

④ LCD

⑤ Control Panel

C7SM-1.01
8/70
Service Manual of Cardio 2000

▣ Front View

① Printer Cover Switch

▣ Rear View

① ②


④ ⑤

① Protective Ground Terminal

② Power Switch

③ AC Power Connection Port ( Fuse : 250V , 5A ( High breaking type ))

④ USB Port(Use for equipment upgreades only)

⑤ RJ45 LAN Port

C7SM-1.01
9/70
Service Manual of Cardio 2000

▣ Left Side View

① Handle

▣ Right Side View

① Patient cable connection port

WARNING

There is a risk of electric shock if the Rest stand of the equipment is damaged or can

not be fixed to the product. Do not use the product and immediately ask the

manufacturer and the seller for repair

NOTE

To avoid an expected electric shock, do not open the equipment cover or


disassemble the equipment. Refer servicing to Bionet, Inc.

C7SM-1.01
10/70
Service Manual of Cardio 2000

LCD Panel

LCD panel displays system setting status after indicating the version of the system and

manufacturer name for 2 seconds when power is turned on.

Displayed items on the LCD are as follows.

① ② ③ ④

1 0 1 2 . 5 O F F O F F

1 0 0 6 C H + 1 R H Y a V R

⑤ ⑥ ⑦

① Display ECG signal level out of 5, 10, 20, aut (Auto Gain)

② Display printing speed out of 12.5, 25, 50

③ Display base line filter setting on or off

④ Display EMG filter setting on or off

⑤ Display heart rate

⑥ Display channel form of the output report out of 3ch+1rhy, 6ch+1rhy, 12ch rhy, 60s 1rhy.

⑦ Display rhythm channel setting out of I, II, III, aVR, aVL, aVF, V1, V2, V3, V4, V5, V6..

C7SM-1.01
11/70
Service Manual of Cardio 2000

Indicator Lamp

I Indicate the lead connection status.

If LED is illuminated green, connection status is good .

If LED is off, connection status is bad.

In this case, check off which one of lead connection is in lead-off status

through monitor mode output.

Caution: Start printing when LED is illuminated green.

Indicate the battery capacity status: Green and Red colors are used

to display 2 different statuses.

The battery indicator is red when the device is operational without having

an A/C power supply. When an A/C power is supplied, the battery

indicator will be either red or green depending on the charge level. No

color will be shown, if there is no battery installed in the device.

Indicate AC power is operated if LED is illuminated green and not in use if

LED is off.

Control panel

Select signal level(mm/mV) out of 5, 10, 20, aut

(I-aVF:10, V1-V6:5 )

Select printing speed(mm/sec) out of 12.5, 25, 50

Select on or off whether or not to activate the filter that eliminates base line

drift

Select on or off whether or not to activate the filter that eliminates EMG.

ECG signal can be distorted by applying this filter.

This key is excepted to use when network function is added ahead.

C7SM-1.01
12/70
Service Manual of Cardio 2000

Select channel form of the output report out of 3ch+1rhy, 6ch+1rhy, 12ch

rhy, 60s 1rhy

Select rhythm channel

Call up the menu to enter patient ID, name, age, sex, height, weight

Display real-time ECGs from the patient continuously or monitoring the

patient ECGs for long time,

Record ECGs for 10 seconds by pressing key and prints in A4 sized report

form after calculating the measurement parameter of the recorded ECGs

Print the same report as previously printed one in record mode or prints

after changing system setting such as filter, signal level, output speed,

channel form, rhythm cannel of previously recorded ECGs data in record

mode.

Exit during printing or recording of ECGs.

Except that, call up the menu or select menu.

Move the selected focus to the left in menu mode

Move the selected focus to the right in menu mode

C7SM-1.01
13/70
Service Manual of Cardio 2000

Power

When AC power is supplied to the equipment, AC power indicator lamp at the front side is
illuminated green and battery is being charged automatically.

Battery Charge AC Power


Indicator Lamp Indicator Lamp

If no battery is installed, battery indicator lamp will be off. Once a battery is plugged in and

the device is powered on, it will change its power mode to supply power to the battery. If the

device is running on the power from the battery, battery charge indicator lamp will light up

red color. If the battery is running low on power, three continuous buzzer alarms will sound. If

this occurs, turn off the device and connect the A/C power before turning the device back

on.

- Required time for charging : more than 4 hours

- Continuous battery operating time : 1 hour

Caution

For the protection of environment, do NOT throw wasted battery away. Please inquire
at the biomedical engineering lab of the hospital and dispose according to the
appropriate procedure to the authorized place(according to national regulations).

C7SM-1.01
14/70
Service Manual of Cardio 2000

1-4. System Installation

Precautions for Installation


While installing CardioCare2000, please pay attention to the following items:
- Use the equipment between the ambient temperature 5 to 40℃ and humidity 10 - 90%.
- Check the power cord is properly connected, and the probe carefully handled.
- Do not plug multiple cords in a power outlet.
- Install and operate unit on a flat surface.
- If you experience noise, ground the device.
- Do not use a power cord that may make a connection noise.
- Device settings will be recorded in the internal memory even when it is off.
- Prevent any shock or excessive force that may cause damage to the device.
- Place the device away from any dust or flammable materials.

Power Connection
The equipment needs electrical power to operate. Plug in one end of the power cable to wall
socket and the other to CardioCare2000.

Patient Cable Connection


- Connect the patient cable to the patient cable connection port on the right side of the body.
- Connect limb electrodes to RL (N), LL (F), RA(R), and LA (L) leads of the patient cable and chest
electrodes to V1 (C1), V2 (C2), V3 (C3), V4 (C4), V5 (C5), and V6 (C6), respectively.

Paper Installation
- Push the printer cover release switch to the right to open the printer door of the
CardioCare2000. Install EKG paper with the side to be recorded appearing on top.
Close the cover to finish the paper installation process.

WARNING

No modification of this equipment is allowed.

Do not modify this equipment without authorization of the manufacturer.

If this equipment is modified, appropriate inspection and testing must be conducted to

ensure continued safe use of equipment.

C7SM-1.01
15/70
Service Manual of Cardio 2000

1-5. Terms of Warranty

- This product was made through strict quality control and examination. The repair and

compensation standards follow the consumer damage compensation regulations of the Ministry of

Finance and Economy.

- The warranty for this product runs for 1 year. (2 years in EU) But the warranty for accessories runs for

6 months.

- This product is guaranteed against faulty workmanship for a period of 1 year. Under this warranty,

we undertake to repair free of charge at our service centers.

- When having troubles during this period, please provide us with the model name, serial number,

date of purchase and a description of the problem you are experiencing.

CAUTIONS

Federal law restricts this device to sale by or on the order of a physician

1-6. Specification

ECG Leads Simultaneous 12 Leads Resting ECG

Recording Channel 3CH+1RHY, 6CH+1RHY, 12CH,


1CH Long Time (1min)

Gain 5, 10, 20, Auto (I~aVF: 10, V1~V6: 5) mm/mV

Printing Speed 12.5, 25, 50 mm/sec

Sampling Rate Analysis Sampling Rate – 500sample/sec


Digital Sampling Rate - 8,000sample/sec

Filters AC (50/60Hz, -20dB or better),


Muscle (25~35Hz, -3dB or better),
Base Line Drift (off, 0.05, 0.1, 0.2Hz, -3dB or better),
Low Pass Filter(off, 40Hz, 100Hz, 150Hz)

C7SM-1.01
16/70
Service Manual of Cardio 2000

Display 2 x 16 Char LCD Display


Sensitivity, Speed, Filter Status, HR, Printing Form, Rhythm Lead

User Interface Short Key

Printer Resolution Thermal Print Head, Roll Paper


Report Paper
- Width: A4/Letter – 215mm (8.5”)
- Length: A4 – 297mm (11.7”)
Letter – 279mm (11”)
Resolution: 8dot/mm (0.125mm pitch)

Patient data ID, Name, Age, Gender, Height, Weight

Basic Measurement Heart Rate: 30~300bpm (±3bpm)


PR, QRS, QT/QTc, P-R-T axis

Electrical Internal Noise: 20uV(p-p)max


Input Impedance: More than 50MΩ
Input Voltage range: ±5mV
CMRR: Greater than 105dB
DC offset Voltage: ≥ ±400mV
Patient Leakage Current: < 10uA
Frequency Response: 0.05~150 with in –3dB
Isolated, Defibrillation and ESU Protected

Line Power Input: 100- 240Vac, 1.5-0.75A, 50-60Hz, Output: 15Vdc, 4.2A

Battery type Replaceable and Rechargeable, Lithium ion, 10.8V, 3350mA

Battery Capacity 360 minutes of continuous operation without recording or


200 ECGs in 12 channel format at 25mm/s and 10mm/mV
with a battery charged during approximately 3 hours from total
discharge (with display off)

Communication LAN

Safety Conformity Class I, Type CF Defibrillation-proof applied Part.

Environmental Operating Temperature: 5~40℃


Operating Humidity: 10~90%RH
Atmospheric Pressure: 70~106KPa

Dimensions 290(W) 300(D) 97.5(H)mm, Approx 3.0kg (Main Body)

Standard Accessory Patient Cable (1 EA), Limbs Electrodes (1 SET),


Chest Electrodes (1 SET), ECG Paper (1 EA),
Power Cord (1 EA)

Options Rechargeable Battery (1 EA),


Cart(1 EA), Hanger (1 EA)

C7SM-1.01
17/70
Service Manual of Cardio 2000

Section 2. Troubleshooting
2-1. Power

Problem Solution
< General Checking >
1. Check the power cable is connected with AC Power connection at the
back of the system.
2. Check the power source of the multi-tab with connection to the power
cable. (If power is connected, AC power indicator is lighted.)
Case1. 3. Check the power switch.
The EKG2000 4. Check any of two fuses on the Power Module (SMPS) and on fuse on the
open will turn on Base board is cut.
and off If it is, replace Power board.
frequently. Case1:
Check if the soldering of J1 on the Base board (SMPS + Base board
Connector) is poor.  Re-solder the cable.
Case2.
Power will not ● When the EKG2000 is turned on, the pin no. 8‟s voltage has to be always
turn on. 14.6V.
1) Is the pin no. 1/2‟s voltage is 14.6 Yes  replace the main board.
2) Is the pin no 1/2‟s voltage is 14.6 No  replace the power board.
● When the REC key is pressed,(during printer is working)
1) Pin no. 12‟s voltage has to be changed from 0V to 24V
If not, faulty power board, replace the power board.
2) Pin no. 15‟s voltage has to be changed from 3V to 0V.
If not, faulty CPU, replace the main board.

C7SM-1.01
18/70
Service Manual of Cardio 2000

2-2. Wave Printing

Problem Solution
1. If the normal wave patterns show up to the
right of chart as shown in problem1, 2,
diagnosis output print too early.
2. It takes some time to show the normal wave
patterns.
3. The normal wave patterns will show if output
is requested after the system is stabilized
by showing the HR variance of ±5 on the LCD
display.
4. Check connection of the limbs electrode
and the chest electrode to the body parts.
5. If used on the dry skin or in the winter
season, connectivity may be degraded. In
that case, used water or ECG gel. The
electrode must be washed after using ECG
1-1. The wave pattern changes from the flat line gel. Otherwise, the electrode will be rusted
on the left to the wave forms on the right. and corroded.

* Connection location of the limbs electrode


and the chest electrode

1-2. Some wave patterns are not shown in V1~6.


1. Press the menu key, showing the overlapped
circle and triangle, at the bottom of the
Keypad.
2. Press the right arrow button twice to select
PRN and then press the menu key.
3. Press the left arrow button once to select
DIAG and then press the menu key.
4. Press ESC key from the menu to enter the
basic display and then press '0' to printout.

2. The diagnosis is not shown in print.

C7SM-1.01
19/70
Service Manual of Cardio 2000

1. This problem occurs when the filter value is


not corrected.
2. Press the menu key, showing the overlapped
circle and triangle, at the bottom of the
Keypad.
3. Press the right arrow key once to select FLT
and then press the menu key.
4. Check if the filter value shows BASE: ON,
MUSC: OFF, AC: 60 Hz, LDF: 150Hz and, if not,
correct it.
5. Since the noise occurs as in problems 2 and
3 if the AC filter is turned off or set at
3-1. There are too many noises in the wave
different frequency from the power source,
pattern.
the system must be set at 50Hz or 60Hz to
correspond to the input power. (It is different
each country) The AC filter status is shown at
the left bottom of the printout as
'CUT OFF (ON) AC'.
6. Problems 3.4 occurs when the BASE is set at
OFF. Change it to ON.

3-2. The noises periodically occur throughout


the wave patterns.

3-3. The thick noise is made on the wave


patterns.

3-4. The wave output looks like the noise and


are not properly shaped.

C7SM-1.01
20/70
Service Manual of Cardio 2000

1. This problem occurs when the chest


electrode is not properly in contact with the
body.
2. Check with simulator there is any cut on the
patient cable.
3. Check there is any damage to the chest
electrode.
4. Test it again after applying alcohol, water or
ECG gel to the electrode.
* If the problem can not be solved by the
above procedures, please contact us.

4. The wave output looks like the noise and are


not properly shaped.
1. The noise occurs as there is no signal from
the limbs electrode.
2. If the polar plate in the limbs electrode is
blacked or copper colored, or if ECG gel
residue remains, the signal may not be
detected. Replace it.
3. If discoloring is too noticeable in the limbs
electrode, replace it.

5. Heavy noise on the wave forms


1. If some wave patterns are too small as
shown in the problem 6, the problem is most
likely to be with the patient cable, limbs
electrode or chest electrode.
2. If the problem is not with the accessories,
then the main board is the problem.
3. The main board must be shipped to us for
repair since the condenser of the main
board must be replaced.

6. The wave pattern shows the straight line


among V1~6.

7-1. When diagnosis output key is pressed, the 1. Replace the main board.
measurement goes on past 10S. * If the above procedure does not solve the
7-2. Operation stops at processing, print after problem, please contact us
10S measurement.
8. The backlight on the screen is on, but there is 1. Check the Key cable which connects Key
no display. board and the main board and replace the
main board.

C7SM-1.01
21/70
Service Manual of Cardio 2000

2-3. Diagnosis

Problem Resolution
1. If disease is diagnosed, abnormal ECG always
appears.

2. LVH occurs when the sum of one of V1 or V2


and one of V5 or V6 is 3.5 or higher.

3. Upgrade to the latest version of the software.

1. The diagnosis shows LVH.

1. If MI is diagnosed, it is likely that the print button


was pressed before the wave patterns are
stabilized.

2. Check HR before printout and print it again.

3. Upgrade to the latest version of the software.

2. Diagnosis shows MI.

C7SM-1.01
22/70
Service Manual of Cardio 2000

2-4. Filter

Base line Filter

Base line drift is caused by a patient‟s breathing. ECG signal is overlapped on a large parabola. If
you want to set Base line drift, select „ON‟. Otherwise, select OFF. The set value appears on the
bottom part of the printed form indicating 0.1HZ- when set to ON, and 0Hz- when set to OFF.

<BASE FILTER OFF> <BASE FILTER ON>

AC Filter

In addition to ECG, other signals such as power noise, base line drift by respiration and
electromyogram may be recorded in the output signals.
There are three types of the noise filter; 60Hz, 50Hz and OFF. When set to OFF, noise is not
eliminated. Set to 50Hz, it indicates to filter 50Hz power noise. Set to 60Hz, it means filtering 60Hz
power noise.
When the system is powered by battery, power noise is not generated, so you may have clear
ECG signal if set to OFF. In Europe where the power noise is 50Hz, the system is set to 50Hz. In other
nations including Korea and the U.S., it is set to 60Hz.
The status of the set filter can be checked on the bottom part of the printout, indicating AC 60Hz,
AC 50Hz or AC OFF.

<AC FILTER OFF>

C7SM-1.01
23/70
Service Manual of Cardio 2000

Low Pass Filter

When ECG is not clear even after all three filters are set to ON, the use of low frequency filter will
produce clear ECG signal. The low frequency filter can be set to 40Hz, 100Hz, and 150Hz and OFF.
Set to 40Hz, it eliminates all signals over 40Hz. The set value is indicated on the lower part of the
printed form. If you set the filter to OFF, -250Hz is indicated. If set to 40Hz, -40Hz is indicated. Set to
100 Hz, -100Hz is indicated. If set to 150Hz, -150Hz is indicated.

40HZ100HZ 150HZ 250HZ FREQUENCY

C7SM-1.01
24/70
Service Manual of Cardio 2000

2-5. Trouble Shooting Guide

1) Power is off

The power is off

Does the
power input Change the power
Yes A
cable break input cable
down?

No
A
Does the
power switch Change the power
Yes A
cable short- switch cable
circuit?
Normal
Yes Use
No operation

Does the main B


power Change the main No
Yes A
connector power connector
break down?

No

Is the fuse Again blown


Yes Change the fuse No
blown? fuse
A

No Yes

Does the main power Repair or change


module produce the No the main power A
power properly? module

Yes

Ask the
manufacturer
about it, then take
action

C7SM-1.01
25/70
Service Manual of Cardio 2000

2) LCD is blank although the power is on

LCD is blank
although the
power is on
B

Check the
Does the LCD cable
condition of the
have a poor condition
Yes connection and A
of connection or short-
then change if
circuit? A
it's not good

No

Broken 20 pin Change the 20 Normal


Yes A Yes Use
shield cable? pin shield cable operation

No No

Is the LCD B
Change the
module Yes A
LCD Module
unsatisfactory?

No

Ask the
manufacturer
about it, then take
action

C7SM-1.01
26/70
Service Manual of Cardio 2000

3) Printer is not working properly

When the printer


is not working
properly
B

Does you Make the paper flat and


place on the place it facing up in
No
printer order to let printer
correctly? cover shut?

Does the
Does the
paper come A
No motor work No Change the motor
out while
properly?
printing?

Yes

Have you
Connet the motor A
connected the
No power cable
motor power
correctly
cable correctly?

Yes Yes
A

Is the motor
Change the motor A
power cable No
power cable
satisfactory?

Normal
Yes Use
operation
Does the
Yes Have you
signal come Connect the THP A
connected the
out on the No No power cable
TPH powor
paper while correctly
cable correctly
printing?
B
Yes

Is the TPH
Change the TPH A
power cable is No
Yes power cable
satisfactory?

Yes

Ask the
manufacturer
about it, then take
action

C7SM-1.01
27/70
Service Manual of Cardio 2000

4) EKG wave does not come out or it's very small

EKG wave does


not come out or
small
B

Is the patient
Reconnect the
cable
No patient cable A
connecter
carefully
correctly?

A
Is the
electrode Cleanse the
connected to electrode with
Yes A
the patient alcohol or water,
cable then it use
correctly?
Normal
No Yes Use
operation
Is the
electrode Reconnected the
connected to electrode to the
No A
the patient patient cable
cable properly
correctly?
B
Yes

Is the patient
Change the
cable short- Yes A
patient cable
circuited?

No

Control the
Is the EKG
volume by setting
signal adjusted No A
a gain on the
properly?
main board

Yes
Ask the
manufacturer
about it, then take
action

C7SM-1.01
28/70
Service Manual of Cardio 2000

5) USB memory does not working properly

EKG files do not write


to USB memory
properly

Format USB
Is USB memory formatted in FAT
Yes memory in FAT
file system?
file system

No

Remove all files


Is USB memory space full ? Yes
in USB memory

No

Check the condition of


Does the USB sig cable have a
Yes the connection and Re-
poor condition of connection?
connect the cable.

No

Exchange the
Is USB sig cable broken? Yes
USB sig cable

No

Ask the
manufacturer
about it, then
take action

C7SM-1.01
29/70
Service Manual of Cardio 2000

6) USB barcode reader does not working properly

Does USB Barcode


reader not work
properly?

Check the condition of


Does the USB sig cable have a
Yes the connection and Re-
poor condition of connection?
connect the cable.

No

Exchange the
Is USB sig cable broken? Yes
USB sig cable

No

Ask the
manufacturer
about it, then
take action

C7SM-1.01
30/70
Service Manual of Cardio 2000

Section 3. Adjustment
3-1. LCD Backlight

The LED backlight is powered by two resistors as shown in the following figure.

C7SM-1.01
31/70
Service Manual of Cardio 2000

3-2. Lead Fault

NOTE

Ready LED is turned off when the lead fault occurs.

C7SM-1.01
32/70
Service Manual of Cardio 2000

3-3. Filter Setting

AC power noise, base line drift caused by breathing , EMG signal can be recorded along with

the output ECG signal. To set the AC noise filter, move to “FLT” menu to call up “AC” menu and

press key. key toggles between off, 50hz, 60hz. Off indicates that filter does not

suppress ac interference. As for 50Hz or 60Hz, filter suppresses 50Hz or 60Hz AC power noise. In case

of battery power, ECG signal is clear only with Off setting because AC noise interference hardly

exist. The selection is 50Hz for most of Europe and 60Hz for USA and other countries. The selected

setting is shown on the bottom line of the output form indicated AC 60Hz, AC 50Hz or AC OFF

Base line drift is caused by patient breath. Base line drift filter can be set by pressing key.

BASE key switches on or off. off, 0.05, 0.1 or 0.2Hz indicates base lien filter is applied or not.

The selected setting is shown on the bottom line of the output form indicated by „0.1hz-„ as for filter

on and „0hz-„ as for filter off.

EMG signal is induced from muscle and internal organs of the patient.
For clear ECG signal output, it is necessary to eliminate EMG signal interference.
EMG filter can be set by pressing MUSC key. MUSC key switches on or off.
On or off indicates EMG filter is applied or not. The selected setting is shown on the bottom line of
the output form indicated „EMG:On‟ or „EMG:Off
Low pass filter can be selected to optimize ECG signal after applying above three filters and if

ECG signal is still interfered. To set the low pass filter, move the menu to “FLT” to call up “LPF” menu

and press key. key toggles between off, 40hz, 100hz, 150hz. As for 40Hz, it indicates

filter suppresses higher frequency than 40Hz. The selected setting is shown on the bottom line of the

output form indicated by „Off„ as for off, by‟-40hz‟ as for 40hz, by‟-100hz‟ as for 100hz, by ‟-150hz‟ as

for 150hz.

To apply filter could result in ECG distortion. So, filter should be appropriately applied to reduce

the fluctuation without affecting the ECG signal when output signal is badly interfered only.

For your reference, base line file and ac filter should be always active and EMG filter can be

applied appropriately. It is recommended for LPF to set 150Hz during diagnosis.

C7SM-1.01
33/70
Service Manual of Cardio 2000

Section 4. Block Diagram/Description


4-1. Circuit Block Diagram

Patient Connector Analog Module


Character LCD

PACE
Shield
ECG AFE Detec
Driver
Character LCD Pin Header -tor

User Interface Module Character LCD Pin Header


3.3V, 5V
Power MCU

Ready LED
Membrane Keys Isolator Isolator
AC / Battery
Power Indicator LED
Battery Supply
Unit User Interface Module Connector Analog Module Connector

User Interface Signal Cable Analog Board Cable


Battery Cable PSU Cable

Basic Module Analog Module Connector


User Interface Module Connector
Battery Connector PSU Connector

Touch
Charger Power MCU CPU Module
controller

Backlight
Power
PMIC
System Power
Switch
DDR3
2G x 2
CPU
USB 1 Port 5V, 3.3V,
4.2V EMMC
Power 4G
Printer
Uart RS232 MCU
port Transceiver USB OTG
Connector

Paper
24V TPH Signal Motor
Ethernet Sensor
LAN PORT(RJ45) Printer Power Transceiver Driver
Transceiver Driver
JTAG
Coin Battery Buzzer
Connector

Printer Connector

Printer Cable

Thermal
Step Paper
Printer
Motor Sensor
Header

PRINTER ASS’Y

C7SM-1.01
34/70
Service Manual of Cardio 2000

4-2. Equipment Overview

1) Top

A
C

A : Printer Assembly
B : LCD Assembly
C : Control Panel

2) Bottom

A : Battery Cover

C7SM-1.01
35/70
Service Manual of Cardio 2000

3) Front

A B

A : Printer Cover Switch


B : Top Cover Assembly
C : Bottom Cover Assembly

4) Rear

A B

C D E

A : Protective Ground Terminal


B : Power Switch
C : AC Power Connection Port
D : USB Port (Use for equipment upgrade only)
E : RJ45 LAN Port

C7SM-1.01
36/70
Service Manual of Cardio 2000

5) Right Side

A : Patient Cable Connection Port

6) Left Side

A: Handle

C7SM-1.01
37/70
Service Manual of Cardio 2000

4-3. Size and Weight

 Size: 290(W) X 300(H) X 89.5(D) mm


 Weight: approx. 3 Kg

C7SM-1.01
38/70
Service Manual of Cardio 2000

Section 5. Disassembly and Assembly


5-1. Module Description

UI Board

Power Board
Base Board
(SMPS)

CPU Board

C7SM-1.01
39/70
Service Manual of Cardio 2000

1) UI Board

<Top > <Bottom >

A C

B D

A : Main Cable (Input Voltage)


B : Membrane Keys Connector
C : LCD Connector (LCD)
D : FFC Cable (40 pin)

It is linked to the Base Board with a Cable and consists of Membrane keys, Power Indicator LED, Battery
Indicator LED, Ready LED.

- Membrane Keys
- Power Indicator LED : Show the state of connection with AC adapter and a green light indicator
when plugging AC adapter into device.
- Battery Indicator LED : A light indicator lets you know that there is battery and show your current
battery charge strength. A red light indicator lets you know that the battery is charging and a
green light indicator lets you know the batter is fully charged.
- Ready LED : It Indicate the lead connection status.

C7SM-1.01
40/70
Service Manual of Cardio 2000

2) Base (Power + Analog) Board

<TOP> <BOTTOM>
I K
J
A C E D
F

H
B G

A : USB Port
B : LAN Port
C : Switch Connector
D : Input DC Connector (SMPS Output)
E : Voltage Regulator (3 ea)
F : LCD Data Connector
G : Motor Driver
H : Main Connector
I : Transformer
J : Isolation IC (2 ea)
K : Patient Cable Connector
L : MCU (3 ea)

The digital part consists of MCU, DC/DC Converter, DC/AC Controller and chip, which are linked
to other external modules.

C7SM-1.01
41/70
Service Manual of Cardio 2000

3) CPU Board

<TOP>

<BOTTOM>

A : MCU
B : DDR3 SDRAM
C : Power Management
D : eMMC Flash

C7SM-1.01
42/70
Service Manual of Cardio 2000

4) Power Module (MDS-065APS15-BA)

<Top>

<Bottom>

It receives 100-240V AC, 50-60Hz signal or battery and supplies 15V to the Base board.

5) Printer

It consists of the TPH (Thermal Printer Head), motor, and paper sensor. And it is linked to the Base
board with 3 pin, 6 pin, and 14 pin cable and approves 24V AC power which will be approved only
if the printer assembly is working.

C7SM-1.01
43/70
Service Manual of Cardio 2000

6) Battery

It is linked to the Digital board with a 3 pin cable. It is 10.8V, 3350mA/h and consists of three 4.2V,
3350mA/h Li-ion Rechargeable battery. It is packed in parallel with 2 sets which are connected with
3 cells in series and 1 spacer.

The cables of battery consist of 3 pins connector which are VCC, GND, GND (Battery Check) as
follow picture. GND (Battery Check) cable should be connected to the board, because the cable
is for checking the status of battery.

GND
GND (Battery Check)
VCC

C7SM-1.01
44/70
Service Manual of Cardio 2000

7) LCD

Character LCD : It connects to CDUI Board LCD with 14 pin Pin Header. It consists of LCD.

C7SM-1.01
45/70
Service Manual of Cardio 2000

5-2. Internal Block Diagram & Cable

Printer
MCU
Printer
Power
Control

24V
USB
VSYS VSYS_SW Printer Power
(15V, 9V~12.6V) (15V, 9V~12.6V)
VLINE
(15V)
Power System
Battery
Supply Power CPU
Charger
Unit Switch
VCPU
SYS_5V
VBAT System
PMIC
(9V~12.6V) 5V Regulator
System Power
Control

Lithium Ion PM System


Analog Isolator
Battery 3.3V LDO 3.3V Regulator
SYS_3.3V

Battery
Level Charger VPM_3.3V LCD
State/Control Power
Battery
Connection Power
Managing Power
State
MCU Signals
Battery
User
Temp.
interface

Battery Battery AC
Power
Orange Green Green
Button
LED LED LED

C7SM-1.01
46/70
Service Manual of Cardio 2000

2) Cable

LCD Assembly

40Pin FFC Cable

Battery Cable Assembly

C7SM-1.01
47/70
Service Manual of Cardio 2000

QN-TPH-Cable

Motor Assembly

Power Switch Cable Assembly

C7SM-1.01
48/70
Service Manual of Cardio 2000

5-3. Disassembly

Step1. Removing the Top cover

a. Removing the 5 screws from the bottom cover.

C7SM-1.01
49/70
Service Manual of Cardio 2000

b. Removing the Top cover.

1) Open the printer cover by pushing slide knob.


2) Disconnect 40 pin FFC cable and Main Cable.
3) Remove Top cover as shown below.

• Pull 40 pin FFC Cable from the connector of Base board.

C7SM-1.01
50/70
Service Manual of Cardio 2000

<Bottom Cover Assembly>

<Top Cover Assembly>

C7SM-1.01
51/70
Service Manual of Cardio 2000

Step2. Replacing the Printer Assembly

a. Remove the 4 screws from the bottom cover.

b. Disconnect the 3 printer cables from the Base Board.

C7SM-1.01
52/70
Service Manual of Cardio 2000

<Printer Assembly>

Step3. Replacing the Base Board

a. Remove the 9 screws from the bottom cover.

C7SM-1.01
53/70
Service Manual of Cardio 2000

b. Disconnect the power connector and the battery cable.

c. Disconnect the power switch cable and ground cable.

C7SM-1.01
54/70
Service Manual of Cardio 2000

Step5. Replacing the Power Module (SMPS)

a. Disassemble Ground cable of Power Board (SMPS) and AC Inlet Ground cable from
Protection Grounding Terminal.

b. Remove the 4 screws from SMPS Bracket.

C7SM-1.01
55/70
Service Manual of Cardio 2000

Step6. Replacing the Key Board

a. Remove the 4 screws from Inside of Top cover

C7SM-1.01
56/70
Service Manual of Cardio 2000

b. Remove the 4 screws and disconnect the 40Pin SMD FFC cable, Main cable, Membrane
Keys cable.

Step7. Replacing the LCD

a. Remove the 4 screws from LCD bracket

C7SM-1.01
57/70
Service Manual of Cardio 2000

Step8. Replacing the Battery

a. Remove the 2 screws from Bottom cover

b. Take out Battery from Bottom cover

c. Remove the Battery cable connector

C7SM-1.01
58/70
Service Manual of Cardio 2000

5-4. Assembly

Assemble the reverse way from Step.8 to Step.1

* Note
Please check whether two parts which is circled in red (rubber handle & printer lock) are
connected correctly.

C7SM-1.01
59/70
Service Manual of Cardio 2000

Section 6. Replaceable Parts List and Accessory

6-1. Replaceable Part List

< Top Cover Ass’y>

C7SM-1.01
60/70
Service Manual of Cardio 2000

<Bottom Cover Ass’y>

<Printer Module Ass’y>


C7SM-1.01
61/70
Service Manual of Cardio 2000

<EKG2000 Ass’y>

<Battery Ass’y>

C7SM-1.01
62/70
Service Manual of Cardio 2000

Section 7. Connector Pin Assignment


7-1. Patient Cable Port

Pin Description
Pin Name
1 V2
2 V3
3 V4
4 V5
5 V6
6 GND
7 N.C
8 N.C
9 RA
10 LA
11 LL
12 V1
13 N.C
14 RL
15 N.C

C7SM-1.01
63/70
Service Manual of Cardio 2000

Section 8. PCB Part Disposition


8-1. Base (Power+Analog) Board

<Top>

<Bottom>

C7SM-1.01
64/70
Service Manual of Cardio 2000

8-2. CDUI Board

<Top>

<Bottom>

C7SM-1.01
65/70
Service Manual of Cardio 2000

8-3. CPU Board

<Top>

< Bottom>

C7SM-1.01
66/70
Service Manual of Cardio 2000

Section 9. Software Upgrade


9-1. Requirement of Upgrading

[ AdvancePreparation ]
1) USB memory (*Note 1)
2) Unzip the bionet.zip file and copy to USB memory the bionet.zip file.
3) Copy to bionet_ekg2000\packages folder the bup file.
* Note 1: File system of USB memory must be the FAT32 or FAT .

C7SM-1.01
67/70
Service Manual of Cardio 2000

9-2 Procedure of Upgrading


(Cardio7 UI screen, upgrade method is same)

1) Connect to the usb port that upgrade software files are copied to a usb memory.

Setup Path

- key → SYS → UPGRADE

2-1) Enter the Password


2-2) Run Upgrade Program that check the password.(*Note 1)
* Note 1 :PASSWORD: Products that display the current date (MMDD) .

- key → SYS → DATE


- ex) Basic → Date : Aug, 25 → Password : 0825

C7SM-1.01
68/70
Service Manual of Cardio 2000

3-1) Select the upgrade file.

3-2) Press key.

4-1) Select the APP(version) → Press key.


4-2) Run to upgrade.

5) When the upgrade is complete, turn off and on.

C7SM-1.01
69/70
Service Manual of Cardio 2000

International Sales & service


Bionet Co., Ltd. :
5F, Shinsegae I&C Digital Center 61 Digital-ro 31 gil,
Guro-gu, SEOUL 08375, REPUBLIC OF KOREA
Tel : +82-2-6300-6410 / Fax : +82-2-6499-7789 / e-mail: sales@ebionet.com
Website: www.ebionet.com

U.S.A sales & service representative


Bionet America, Inc. :
2691, Dow Ave, Suite B
Tustin, CA92780U.S.A.
Toll Free: 1-877-924-6638 / Fax: 1-714-734-1761 / e-mail: support@bionetus.com
Website: www.bionetUS.com

European sales & service representative


MGB Endoskopische Gerä te GmbH Berlin :
Schwarzschildstraße 6
D-12489 Berlin, Germany
Tel: +49(0)-30-6392-7000 / Fax: +49(0)-30-6392-7011 / e-mail: sales@mgb-berlin.de
Website: www.mgb-berlin.de

Bionet Co., Ltd


Model Name: EKG2000

C7SM-1.01
70/70