Professional Documents
Culture Documents
Product Guide
August 2009
AT1442E, Rev. 06
Copyright 2009 by AVL List GmbH, Graz - Austria
The contents of this document may not be reproduced in any form or communicated to any third party without the
prior written consent of AVL. While every effort is made to ensure its correctness, AVL assumes no responsibility
neither for errors and omissions which may occur in this document nor for damage caused by them.
All mentioned trademarks or registered trademarks are owned by their respective owners.
Product Guide
4 Warnings and Safety Instructions
WARNING
WARNING
Product Guide
6 Grouped Safety Messages
NOTICE
NOTICE
Bear in mind that during mounting there is a risk of injury due to the heavy
weight, the center of gravity and the sharp edges of the case.
If the AVL Fuel Mass Flow Meter has been rigged for a short-term period without
any fastening, the device cover may not be opened under any circumstances –
the device would tip over.
During transportation of the device the transport lock of the sensor has to be
inserted, otherwise there is a risk of damage to the fuel lines!
Cables and hoses have to be professionally laid to the correct connections.
Make sure that they are not exposed to mechanical stress and nobody is
obstructed (e.g. use cable pits).
NOTICE
NOTICE
There absolutely must be an appliance that interrupts the fuel supply in case of
errors. If there is no corresponding mechanism available at the test bed, the
option shutoff valve has to be used.
Information
The device is not suitable for monitoring the fuel circuit for line breaks.
During work at any part of the fuel circuit / device, switch off the device and
disconnect all communication lines (digital I/O, COM1 and COM2).
Please note that the compressed air supply must be maintained for a short
period in case of emergency during switching off the power supply (delayed
switching-off).
The device must not be exposed to rain or permanent humidity.
Check the functioning of the float switch and the entire safety circuit (shutoff
valve) on a regular basis.
After the warning "Tank overflow", the overflow tank must be completely drained!
Information
Information
There are two types of the device AVL 735S, which differ in terms of fuel resis-
tance.
See Scope of Delivery on page 15.
Please absolutely make sure that only the appropriate spare parts are used for
the type 7351CME:
Type "alcohol-resistant" (M100).
Information
Please follow the instructions given in the section Scope of Application and
Intended Use on page 65.
Information
When disposing of the product or of parts of it, make sure that the legal regula-
tions in force in the country in which the device is operated are observed (e.g.
regulations for the disposal of electronic scrap). The device must always be
completely emptied of any fuel which might be present.
Please contact AVL for further information on the substances that are contained
in the device.
Product Guide
8 Grouped Safety Messages
Table of Contents
Warnings and Safety Instructions................................................................................................................ 3
2 Installation Instructions.................................................................................................. 15
2.1 General................................................................................................................................... 15
Product Guide
Table of Contents 10
Product Guide
Table of Contents 11
Product Guide
Table of Contents 12
Product Guide
What You Should Know 13
WARNING
CAUTION
Additional messages:
NOTICE
This text indicates situations or operation errors which could result in property
damage or data loss.
Product Guide
14 What You Should Know
Information
Lists:
1. Step-by-step procedures with a given sequence
2.
• One-step instructions
Unordered series of concepts, items or options
–
–
2 Installation Instructions
This section covers the following topics:
General on page 15
Scope of Delivery on page 15
Installation at the Test Bed on page 22
Hydraulic / Pneumatic Integration on page 30
Electrical Connection on page 40
Connecting the Options on page 57
First-Time Start-up on page 62
2.1 General
These installation instructions describe the prerequisite conditions as well as
each individual steps for a successful installation and commissioning of the AVL
Fuel Mass Flow Meter.
This documentation also relates to the fuel conditioning device AVL Fuel
Temperature Control – for further information refer to the corresponding Product
Guide.
Information
Fig. 1
Product Guide
16 Installation Instructions
Description Number
(min. 2 ×) Serial cable RS232 BV2431
(interface cable, 15 m)
(min. 1 ×) Analog cable BV2432
(BBY signal cable, 15 m)
(min. 1 ×) Digital I/O cable (15 m) BV2266
(min. 1 ×) Emergency stop cable (15 m) BV2573
(min. 2 ×) 25-pin panel jack EU0033
(min. 2 ×) 9-pin panel jack EU0930
(min. 2 ×) 9-pin panel plug EU1511
(min. 2 ×) 25-pin cap EU1145
(min. 2 ×) 9-pin cap EU1228
1 × 8-pin cable plug EU1458
1 × Product guide AT1442E
1 × Test protocol —
1 × Test Certificate —
1 × Checklist AT1670
1 × CD-ROM PC software XF0163
1 × CD-ROM documentation and software PA0713
Tab. 1
2.2.2 Options
This section contains a listing of the optional equipment for the operation with
AVL 735S.
Example 735S
9 3
1
4
7 6
Product Guide
18 Installation Instructions
Example
735S + 753C
15 2
12
17
4
13
8
16 9
14
5
18 10
11 6
Description Number
Calibration Unit TM0735CALA.02
Tab. 2
Description Number
Power supply 24 V/10 A, 115/230 V AC TM0735P2A.01
Tab. 3
Description Number
Shutoff valve (including valve, plug, 2 screw
fittings, cable, hose pipe)
for standard fuels TM0735AVA.01
alcohol-resistant TM0735AVM.01
Tab. 4
Description Number
Flame Filter TM0730FF1A.01
Tab. 5
Description Number
Fine filter 25 µ for standard fuels TM0730FE1A.01
alcohol-resistant TM0733F1MA.01
Tab. 6
Product Guide
20 Installation Instructions
Description Number
Coarse filter 60 µ for standard fuels TM0730FE2A.01
alcohol-resistant TM0733F2MA.01
Tab. 7
Description Number
Remote control AVL Fuel Mass Flow Meter / TM0735FA.02
AVL Fuel Temperature Control
Tab. 8
The following component is included in the scope of supply of the remote control
option:
Description Number
1 × Connecting cable (15 m) BV2191
Tab. 9
2.2.2.8 Trolley
This option allows one to mount the AVL Fuel Mass Flow Meter and AVL Fuel
Temperature Control on a movable trolley.
Description Number
Trolley TM0735TROA.01
Tab. 10
Description Number
Fuel filling pump module TMFILLPUMP.01
Tab. 11
Description Number
Venting valve BH0317
Tab. 12
Description Number
Application Notes AT0802E
"Fuel Consumption Measurement"
Tab. 13
Description Number
Fuel supply line for standard fuels BH0338
Fuel supply line alcohol-resistant BH0340
Tab. 14
Fuel supply and return line of type "standard" BH0336, type "alcohol-resistant"
BH0341
Coarse filter (60 µm) and fine filter (25 µm). This option is installed into the fuel
supply line and has a filter cross-section of 590 cm2.
Description Number
Fuel supply and return line for standard fuels BH0336
Fuel supply and return line alcohol-resistant BH0341
Tab. 15
Product Guide
22 Installation Instructions
Description Number
NPT connection set for standard fuels BH0339
Tab. 16
CAUTION
CAUTION
Information
770
667
324
683 342
530
770
Fig. 4
Transport The AVL Fuel Mass Flow Meter and the corresponding accessories will be deliv-
ered in an appropriate packing.
Information
Although the device has been drained, remainders of the test fuel can be in fuel
lines and containers.
1. Remove all sealing stoppers with which all connections are sealed.
2. Remove the transport lock of the flow sensor (see the figure below).
3. Position the transport safety guard at the output pressure gage correctly,
see Fig. 33 on page 59.
If the device has been stored below room temperature, let it be warmed up
to room temperature for several hours before commissioning the device.
The AVL Fuel Mass Flow Meter can be rigged at the test bed in various ways:
fixed wall mounting (preferred)
mobile on a movable trolley
In any case it can be used standalone as well as in combination with the AVL
Fuel Temperature Control.
Regardless of the mounting method the AVL Fuel Mass Flow Meter must be
installed free of vibrations – therefore it is recommended to mount it at a fixed
wall of the test bed.
Product Guide
24 Installation Instructions
NOTICE
Make sure that the bearing capacity of the wall is sufficient to allow for mounting
the device. The wall must be solid, fixed, even and totally perpendicular.
Allow 360 mm of free space above the mounting bracket for a fixed mounting at
a test bed (wall mounting).
After the AVL Fuel Mass Flow Meter is mounted, it has to be leveled with the
built-in box level. Thus, make sure that the mounting bracket is mounted as hori-
zontally as possible.
The AVL Fuel Mass Flow Meter may only be mounted and installed indoors.
In order to mount the device the mounting bracket has to be screwed off the
device.
1. Remove the upper fastening screw (1).
2. Loosen the securing screws (2) at the bottom of the device by positioning
the knurled lock nuts fully into the direction of the housing and by slightly
turning the hexagonal bolt outwards.
The following figure shows the fastening screws on the mounting bracket.
1 3
2 2
1 .... Upper fitting screws
2 .... Lower fastening screws (left and right)
3 .... Transport lock
Fig. 5
NOTICE
Components, hoses or lines of the AVL Fuel Mass Flow Meter must not be used
for lifting the device because of the risk of damage.
216,5
175,5
140,5
121,5
85
66,5
75
45
96
113 71,5
138 101,5
163,53
189,25 215
321,5
430
Fig. 6
Side view:
1
79,5
53,5
2
1 .....Assembly Drilling of the Mounting Bracket
2 .....Fuel Connections
Fig. 7
Product Guide
26 Installation Instructions
Information
Utilize armored hoses for flexible connections. If you apply teflon hoses, you
should only use technically conducting PTFE hoses with metal coating.
There are adaptor kits available for the connection of the systems AVL Fuel
Mass Flow Meter and AVL Fuel Temperature Control.
WARNING
Danger of explosion
The device must always be vented and filled up with fuel!
The device must not be drained after it has been operated with Otto fuels or
alcoholic mixtures (e.g. for a longer standstill period).
If draining is necessary, (e.g. for transportation or storage) make sure to observe
the following instructions:
1. Drain the device completely.
2. Fill up the device with inert-gas.
3. Seal the five hydraulic connectors (except connector K, compressed air)
hermetically.
These steps must only be followed if the device was filled up with Otto fuels or
alcoholic mixtures!
The AVL Fuel Mass Flow Meter is also involved in the provisions:
The fuel circuit in the AVL Fuel Mass Flow Meter has to be leak-tight due to
functionality reasons. The leak-tightness is checked during assembly.
Normally a surface temperature of 260 °C cannot be reached inside the device,
the isolation of the built-in electrical components would be destroyed much
earlier.
2.3.3 Combination of AVL Fuel Mass Flow Meter and AVL Fuel Temperature Control
Space requirement of the mounting brackets for the installation of AVL Fuel
Mass Flow Meter with AVL Fuel Temperature Control
360
~400 50 430 50 >350
50
494
150 250
Ø12
A B
165
1566,5
811
E
45,5
C 215 D
430
>345
530
Fig. 8
Product Guide
28 Installation Instructions
Information
The devices AVL Fuel Mass Flow Meter and AVL Fuel Temperature Control must
be installed in the specified vertical distance.
360
110 ~150 50 430 50 ~250
50
494
Ø12
120
430
530
Fig. 9
CAUTION
Information
The trolley has to be placed free of vibrations! Make sure that the trolley is
placed on an even base.
Fit the mounting bracket with the enclosed mounting material to the borings of
the trolley.
NOTICE
Components, hoses or lines of the AVL Fuel Mass Flow Meter must not be used
for lifting the device because of the risk of damage.
Information
A leveling of the device is only required if the option calibration unit is integrated.
Nevertheless it is recommended to level the device after mounting; see section
Leveling the AVL Fuel Mass Flow Meter on page 29.
2 3 4 2
1 .....Upper fitting screws
2 .....Lower fastening screws (left and right)
3 .....Tank foot
4 .....Box level
Fig. 10
Utilize the adjustment and fitting screws to level the system.
Product Guide
30 Installation Instructions
NOTICE
Information
Adhere to the local safety regulations (e.g. grounding of fuel lines …)!
Make sure that the fuel pipework and the system elements (filter,
pumps, heat exchanger) are abolutely leak-tight and sufficiently resis-
tant against the used fuels and pressures (recommended: 25 bar).
Keep the fuel volume in the measurement as small as possible.
Utilize absolutely pressure-proof pipes and system components in
order to keep the geometric volume constant.
Recommended pipe material:
– stainless steel
– metal-braided, electrically conducting teflon hoses for flexible connec-
tions
Product Guide
32 Installation Instructions
R>
Fig. 12
NOTICE
There must always be an appliance that interrupts the fuel supply in case of
errors.
If there is no corresponding mechanism available at the test bed, the shutoff
valve option has to be used.
The fuel line in has to have a fuel filter.
Draining the device via the drain cock as used in device type 7350 (see type
label) causes only a draining of the overflow tank with device type 7351.
A C D B
K E
Fig. 13
Product Guide
34 Installation Instructions
NOTICE
The maximum pressure may not exceed 80 kPa in the fuel supply line. Espe-
cially pay attention to the pressure if you are using a fill pump.
If the existing fill pressure is too low, please utilize the "option fuel fill pump". If a
different fill pump is utilized, the flow rate of the pump has to be within the range
of 150 … 250 l/h.
A currentless shutoff valve must be installed in the fuel supply line between the
supply tank and the AVL Fuel Mass Flow Meter (see section Connecting the
Shutoff Valve (X1) on page 47).
The fuel fill pressures of the following table refer to an open fill valve in front of
the shutoff valve at ~25 °C and diesel fuel.
Tab. 17
If there is still some fuel in the overflow tank after the error "Tank Overflow", you
have to drain the overflow tank with this valve (see Venting, Overflow on
page 35).
WARNING
Product Guide
36 Installation Instructions
AVL
Fuel Mass 2
Flow Meter
3 4
Tab. 18
NOTICE
The maximum pressure drop in the venting line may not exceed 0.5 mbar at 400
l
/h air throughput.
The pressure at the venting connector affects the measurement circuit and thus
influences possibly also the mixture former.
An excess pressure of more than 0.2 bar can cause damage to the AVL Fuel
Mass Flow Meter.
Pressure changes due to venting of more than 0.5 mbar affect the venting of the
device negatively.
WARNING
As, in case of fault, the venting line is also used as overflow line, it must be sized
appropriately.
At vertically arranged test beds the pipework must be provided such that the
overflow of a higher system does not inadvertently fill the lower venting lines!
The venting line must be implemented in every case technically leak-tight.
2.4.2.5 Combination of AVL Fuel Mass Flow Meter and AVL Fuel Temperature Control
If the AVL Fuel Mass Flow Meter is used in conjunction with the AVL Fuel
Temperature Control, a separate adaptor kit is available.
Product Guide
38 Installation Instructions
45
43
41
58
e b c a d 57
46 G
44
J A B
42
F E D C
H P
51
52 54
P P
60
53
48
F
E
49
50
Fig. 15
Fuel to Engine
Connect this outlet via the supplied hose with the connector fuel inlet at the
AVL Fuel Temperature Control.
Fuel Return Line
Connect this outlet via the supplied hose with the connector fuel return line
(venting) at the AVL Fuel Temperature Control.
Product Guide
40 Installation Instructions
NOTICE
Make sure that the device is permanently supplied with compressed air.
Please note that the compressed air supply must be maintained for a short
period in case of emergency during switching off the power supply (delayed
switching-off).
Operating voltage: 24 V DC ± 5 %
Current consumption: max. 10 A
If the AVL Fuel Mass Flow Meter is used in conjunction with the AVL Fuel
Temperature Control, the power supply has to be provided by the AVL Fuel
Temperature Control. An appropriate cable is part of the related adaptor kit.
If there is no AVL Fuel Temperature Control available, either the option power
supply or another power pack with secure power separating, which conforms to
the aforementioned specifications, can be used as power supply.
WARNING
2.5.1.1 Connecting the AVL Fuel Mass Flow Meter With the Option Power Supply
2
AVL
3
Fuel Mass
Flow Meter 4
1
WARNING
The AVL Fuel Mass Flow Meter can be integrated in two different ways:
Wiring without potential separation
(state of supplied device, jumper J17 in position 2-3)
Wiring with potential separation
(jumper J17 in position 1-2)
Product Guide
42 Installation Instructions
+24 V
F5
0.5 A X7
1
2
1
2
3
J17
X7 0.5 A
1
2
1
2
J17 3
3
1 2
1 .....Jumper J17
2 .....Device front side
Fig. 19
Information
The supplied cable (emergency stop cable, Material No. BV2573) is designed for
the wiring without potential separation. For an integration with potential separa-
tion the pin assignment has to be changed.
The figure below shows the emergency stop cable,
the view 1 shows the solder side of the socket.
S
+24 V
Sig.
1
22 1
33 3 1
Tab. 19
Product Guide
44 Installation Instructions
Information
The option venting valve cannot be used in conjunction with the AVL Fuel
Temperature Control.
2.5.4 Interfaces
10
11
Fig. 21
Information
Information
If the AVL Fuel Mass Flow Meter is combined with the AVL Fuel Temperature
Control, the system venting is controlled by the AVL Fuel Temperature Control –
therefore this input may not be used in this case.
Product Guide
46 Installation Instructions
NOTICE
Make sure that the media (fuel, compressed air) are available while an active
edge is sent at the input SET DEVICE OK (X5), because otherwise the measure-
ment device could be damaged (dry running of the fuel pump).
Information
Analog Output
The measurement value (depending on the selected domain mass, volume or
volume API) is available as analog voltage:
0 … 10 V correspond to 0 … 125 kg/h or 0 … 125 l/h
(default values; can be parameterized)
The permissible value range for 0 V is –50 … 200 kg/h or l/h,
the parameterized value for 10 V must be at least 10 kg/h or l/h higher than the
parameterized value for 0 V.
The data rate of the analog output is 20 Hz.
During connecting, note that the integration of the data sampling may only be
accomplished without grounding (differential), otherwise the measurement signal
can be disrupted.
Information
NOTICE
If there is a failure of the compressed air supply or an overflow of the tank inside
the device, it has to be ensured that the fuel supply is interrupted. Therefore a
shutoff valve has to be provided.
The activation is electrically triggered via the output EXT. VALVE (X1).
This output is switched with a currentless open relay. The internal activation of
the relay is done by a float switch.
This connection can be actuated with max. 60 VA (at max. 60 V).
For third party valves, an additional power supply has to be provided, as shown
in the following figure. If necessary, provide a free-wheeling diode to keep
voltage spikes away from the device board.
Socket EXT. VALVE (X1),
the socket is shown viewed from the plug.
7 8 6
3 1
5 2 4
Fig. 22
Product Guide
48 Installation Instructions
Tab. 20
DIGITAL
ANALOG COM2 COM1
I/O
1
2 3
X1 X2 X3 X4 X5
POWER COM 2 COM 1 EXTERN COM 0
! GRAZ
AUSTRIA
FOLLOW THE DIRECTIONS! DON´T
OPERATE THIS INSTRUMENT IN
Type S/No
EXPLOSIVE HAZARDOUS LOCATIONS!
THE OPENING OF THIS DEVICE IS Rev
PERMITTED ONLY BY AUTHORIZED
TRAINED PERSONNEL!
Interface Parameter
Baudrate 9600
Databits 8
Stopbits 1
Parity none
Handshake: none
Tab. 21
Pin Assignment
Pin 1 RxD
Pin 2 TxD
Pin 3 Ground
Pin 4 Ground Supply for remote control
Pin 8 +24 V
Tab. 22
Product Guide
50 Installation Instructions
Pin configuration of the RS232 Interface, the socket is shown viewed from the
plug.
7 8 6
3 1
5 2 4
Fig. 24
Pin Connection:
Tab. 23
NOTICE
Ensure the correct position of the switches S1/S2 according to your application
before wiring the digital inputs and outputs.
If the outputs are wired without potential separation (switch S1 in position), pin
2/7/8 of the connection digital I/O (X5) can have +24 V DC.
S2 / Pos 2
1; 4; 5
(+)
6
(-)
5 4 3 2
Fig. 25
S2 / Pos 1
1; 4; 5
6
+24 V
6 5 2 1
Fig. 26
Product Guide
52 Installation Instructions
S1 / Pos 2
2; 7; 8
....
3
5 4 3 2
Fig. 27
S1 / Pos 1
2; 7; 8
....
3
+24V
+24 V
6 5 2 1
Fig. 28
Logic assignment
A suitable cable (digital I/O cable, Material No. BV2266) is available for the
connection to the digital I/O socket.
Pin / color assignment of the digital I/O cable:
Tab. 24
7 8 6
3 1
5 2 4
Fig. 29
Tab. 25
Information
Make sure that none of the poles of the analog output is grounded, because this
could cause interference at the analog output.
Integrate the analog output differentially into your data aquisition.
Product Guide
54 Installation Instructions
Tab. 26
Information
Changes of the DIL switch settings only take effect after a reset of the devices.
Information
In the shipment configuration the DIL switches DIL2 through DIL8 of the DIL
switch block S3 are in the OFF position.
The DIL switches DIL1 through DIL8 of the DIL switch block S4 are in the OFF
position.
With the DIL switch block S3, the following device parameters can be set:
Tab. 27
The AVL Fuel Mass Flow Meter offers a simple method for achieving measure-
ment readiness / state STBY. Immediately after switching on the device tries to
get into the state STBY. The automatic procedure is called Auto-STBY.
Product Guide
56 Installation Instructions
Tab. 28
WARNING
Leaking of fuel
When utilizing the AVL Fuel Mass Flow Meter with the Auto-STBY function,
please ensure that the device gets into the state STBY via the Pause state after
the device is supplied with voltage. With that a fuel pressure is built up and the
fuel is conveyed.
Do not use this device function if it gives rise to safety risks (e.g. because of
leaking fuel).
Function Bypass
Tab. 29
Information
The device function Bypass can only be set from the state Pause and the corre-
sponding DIL switch setting. Compressed air must be available and the emer-
gency stop circuit must be closed.
The DIL switches DIL2/3/4/6/7 of the DIL switch block S3 have no documented
function, they must be set to position Off. All DIL switches of the DIL switch block
S4 also have to be set to position Off.
The filter at the analog output (X6) can be varied with the jumper J19.
Tab. 30
The jumpers J16 and CONT1 have no documented function and must be set to
position 1/2.
Tab. 31
Tab. 32
NOTICE
There must always be an appliance that interrupts the fuel supply in case of
errors.
If there is no corresponding mechanism available at the test bed, the shutoff
valve option has to be used.
If the test bed is equipped with a corresponding mechanism for interrupting the
fuel supply in case of failure, the output EXT. VALVE (X1) has to be connected
there.
Product Guide
58 Installation Instructions
2
8
7 6
3 1
4
1. Use the 8-pin cable plug EU1458 for the connection to EXT. VALVE (X1).
2. Wire the plug according to the following table.
Tab. 33
7 8 6
3 1
5 2 4
Fig. 32
2
1
Information
After a period of 18 months the calibration unit has to be sent to the relevant AVL
subsidiary in order to check the calibration.
Product Guide
60 Installation Instructions
1
1 .... Eccentric screws
Fig. 34
Information
To ensure a proper calibration procedure, the AVL Fuel Mass Flow Meter has to
be leveled (see Leveling the AVL Fuel Mass Flow Meter on page 29).
If the device is to be moved, the calibration unit has to be removed.
Information
Due to the exhalation of fuels it is also from time to time necessary, even after
the first-time installation, to ensure a sufficient venting of the calibration sensor
for a proper functioning of the calibration option.
With gasoline, the calibration unit has to be vented before each calibration
procedure (-point).
Before any transportation, the calibration unit has to be removed. We recom-
mend that the calibration unit is even removed before transportation with the
optional trolley. Make sure that the quick release couplers are centered when
plugging on the calibration unit.
Information
For consumers without return flow, this option may be omitted. In that case, the
connection "fuel return (C)" has to be closed.
Mounting:
1. Connect the venting valve option to the connections "fuel to engine" and
"fuel return" of the AVL Fuel Mass Flow Meters an (see the figure below).
2. Connect the pneumatic outlet of the pilot valve with the air pressure supply
by the means of a T-piece.
3. Lead the air pressure conductor through the rubber bushing in the bottom of
the device.
4. Lead the cable through the secong rubber bushing into the device and con-
nect it with the circuit board on the upper left side.
5. Check that installation is successful:
– Switch the AVL Fuel Mass Flow Meter to operating state PAUSE.
– Drain the AVL Fuel Mass Flow Meter.
– The connections 2 and 4 in the figure below must be connected when
the pilot valve is not supplied with power.
When the pilot valve is supplied with power (bypass), connections 3
and 4 have to be connected.
Information
Allow 120 mm of free space above the mounting bracket for a fixed mounting at
a test bed (wall mounting).
Product Guide
62 Installation Instructions
3 6
5
4
1 .... Connect with the connection "fuel to engine" at the AVL Fuel Mass
Flow Meter
2 .... Connect with the connection "fuel return" at the AVL Fuel Mass
Flow Meter
3 .... Fuel supply of the engine
4 .... Fuel return from the engine
5 .... Servo control valve (pre-assembled)
6 .... Electrical connection cable of the servo control valve
7 .... T-piece
Fig. 35
Information
Before each start-up, check whether all fuel lines which are required for oper-
ating are properly connected.
Make sure that the drain cock of the overflow tank is connected.
Check, whether the transport locks have been removed and are properly
aligned, see Installation at the Test Bed on page 22.
If the outputs are wired without potential separation (switch S1 in position), pins
2/6/7/8 of the connection digital I/O (X5) can have +24 V DC.
Information
The operation of the AVL Fuel Mass Flow Meter can be accomplished with a test
bed computer, the PC software, the remote control or a terminal program. In the
following comments these special circumstances will not be dealt with, but rather
the corresponding AK commands will be specified.
1. Make sure that the requirements of the section Installation at the Test Bed
on page 22 and Hydraulic / Pneumatic Integration on page 30 as well as the
related subsections are fulfilled.
2. If the AVL Fuel Mass Flow Meter is integrated with a hybrid interface, it is
recommended to temporarily connect the serial interface during the
first-time start-up and parameterization of the system.
3. If fuel and compressed air supplies are available, switch on the system.
It takes a maximum of 50 s until the device has completed the initialization
and the interfaces are ready for operation. After the initialization, the device
switches into the state Pause and performs an internal error check.
Now, you can check the communication with the device across the serial
interface (9600 Baud, 1 stopbit, no parity, 8 databits).
4. Enter the gravitational acceleration (g) for your location (can usually be
requested at the national calibration laboratories for force-measuring). (AK
command EKGP)
Warnings are indicated by slow flashing of the device. At the first-time
start-up (the device has not yet been filled) the error "Density too low"
should be indicated.
5. Start the first filling of the AVL Fuel Mass Flow Meter (AK command SFIL). If
the device has been parameterized to Auto-STBY, the device automatically
starts the first filling.
The device fills the internal tank completely with fuel. If due to excessively
filling pressure, the filling takes longer than 80 s, the error "Filling rate too
low" is set. In this case, recheck the pressure of the fuel inlet line and restart
the first filling.
After complete filling of the internal device tank, the device performs an
internal venting and remains in the state SFIL (permanent venting), until it is
transferred into another state.
Now, all warnings should have been reset.
Information
Ensure proper fuel supply when sending the command for the first filling.
The first filling can also be accomplished in the state STBY (AK command
STBY), but the venting has to be made manually. The state STBY may have to
be activated several times, because the device might be set into the state Pause
due to the warning "Density too low".
6. The filling of other fuel devices (e.g. fuel conditioning or engine circuit)
should be done during the internal venting in state SFIL. Always ensure a
good venting of all fuel lines before you switch the device into the state
STBY!
7. Now, you can switch the device into the state Pause (AK command SPAU).
Product Guide
64 Installation Instructions
8. If you want to adjust the output pressure regulator, switch the device into the
state STBY. Now, adjust the output pressure regulator of the AVL Fuel Mass
Flow Meter to the desired output pressure:
– Return line pressure in combination with a conditioning system
– Supply pressure in systems without return lines
If there is no calibration available for the used fuel, the warning "Adjustment
error" is shown, after the system switches into the state STBY for the first
time.
Make sure that the required minimum pressure of >0.05 bar is maintained!
The first-time start-up is now complete.
NOTICE
The device was subjected to a leak check prior to delivery. Caused by transpor-
tation or thermal stress, however, leakages may occur during the first operating
hours which could not be detected in a cooled-down system condition.
2.7.1.1 Drain
1. Switch the device into the state Pause or STBY.
2. Switch off the fuel supply (takes place automatically, if the option "shutoff
valve" is integrated).
3. Start the draining procedure (AK command SDRN) and ensure fuel con-
sumption (possibly by opening the drain lines in the AVL Fuel Temperature
Control).
After several minutes the device is drained and the device remains in this
operating state.
4. Now, drain any other devices (e.g. AVL Fuel Temperature Control).
Not all internal lines can be drained with this draining procedure:
return line inside device (~75 ml)
Drain these lines as well as the other fuel lines to and from the engine
through suitably positioned valves.
Information
The AVL Fuel Mass Flow Meter can also be drained by manually closing the fuel
inlet line in state STBY. The device will be drained, until the warning "Density too
low" appears.
3 User's Guide
NOTICE
Fuel resistance:
Standard: Otto and diesel fuels, alcoholic mixtures up to 20 %
(corresponds to the short term M20 and E20)
FlexFuel: Otto and diesel fuels, alcoholic mixtures up to 100 %
(corresponds to the short term M100 and E100)
Intended Use According to the intended use, the AVL Fuel Mass Flow Meter 735S is suitable
for the highly precise measurement of fuel consumptions according to the tech-
nical specifications, particularly in areas with the demand of a permanent
measurement capability.
In particular, the device is suitable for the continuous measurement of flows of
liquid fuels (diesel fuel, Otto fuel, in the special version also for methanol and
ethanol as well as mixtures of them) to an emission location (consumer like
combustion engines, injection pumps and jets).
WARNING
Danger of explosion
According to the intended use, the AVL Fuel Mass Flow Meter 735S is not suit-
able for the measurement of flows of mixtures of two or more phases, such as
boiling fuels, immiscible liquids or precipitating particles or solids, even if these
are only formed by the physical conditions within the measurement circuit.
According to the intended use, the AVL Fuel Mass Flow Meter 735S is not suit-
able to measure flows of gaseous media. Such a use represents a high safety
risk.
According to the intended use, the AVL Fuel Mass Flow Meter 735S may not be
operated in an explosion-endangered environment. The minimum requirements
of monitoring of these indoor rooms must be fulfilled, see Measures to Prevent
Fire and Explosion on page 26.
Product Guide
66 User's Guide
According to the intended use, the AVL Fuel Mass Flow Meter 735S has to be
mounted stationary or on a mobile trolley indoors and is usually used at
steady-state engine and vehicle test beds.
Information
3.2.1 General
The AVL Fuel Mass Flow Meter is based on a Coriolis mass flow sensor –
allowing continuous measurements. The high sensitivity and precision is accom-
plished with a very small inside diameter of the sensor pipe, a pump in front of
the sensor produces the required pressure.
A pressure regulator in front of the sensor always adjusts the required pressure
for the corresponding fuel consumption, so that behind the sensor a constant
adjustable output pressure is available.
The following figure shows the AVL Fuel Mass Flow Meter in the state current-
less, compressed air available.
7
1 10
B
5 12
9 4
2
C
E 11 6
H
G
A
8
F
Fig. 36
Standard M 100
1 Tank level sensor
2 Fill valve
3 Float switch (overflow)
4 Fuel Pump
5 Stop valve
6 Pressure regulator
7 Flow sensor
8 Bypass venting valve
9 Drain cock overflow tank
10 Outlet pressure sensor
11 Inlet pressure sensor
12 Outlet pressure manometer
A Fuel in
B Fuel to engine
C Fuel return line
D Venting
E Drain overflow tank
F Shutoff Valve option TM0735AVA.01 TM0735AVM.01
G Venting valve option BH0317
H Calibration TM0375CALA.02
unit option
Product Guide
68 User's Guide
The following measured quantities can be provided by the AVL Fuel Mass Flow
Meter:
Mass domain
– Mass flow [kg/h]
Instantaneous value (10 Hz), time-averaged value during the set mea-
surement time, moving average on the number of measurements,
standard deviation of the mean time values of the set number of mea-
surements
– Accumulated mass flow per period = mass [kg]
Volume domain
– Volume flow [l/h]
Instantaneous value (10 Hz), time-averaged value during the set mea-
surement time, moving average on the number of measurements,
standard deviation of the mean time values of the set number of mea-
surements
– Accumulated volume flow per period = volume [l]
Volume domain/API
– Volume flow API (standard volume according to API-2540, standard-
ized at 15 °C) [l/h]
– Instantaneous value (10 Hz), time-averaged value during the set mea-
surement time, moving average on the number of measurements,
standard deviation of the mean time values of the set number of mea-
surements
– Accumulated volume flow API per period = volume [l]
In all domains
– Density [kg/m³]
– Fuel temperature in the AVL Fuel Mass Flow Meter
The following diagram shows the various operating states and their transitions.
In order to change from the state Pause to another state, the emergency stop
circuit must be closed. (You can only change to the state Service if the emer-
gency stop circuit is not closed.)
Power on
- AK: SRES
- Emergency stop
- No compressed air AK: SRES
Init Pause
From all states:
- AK: SPAU AK: SSRV
Pause Service
- System error
AK: SDRN
Pause AK: SRES
Draining
System Error
"Density too low"
- AK: STBY
- Emergency stop not interrupted + Reset
AK: SDRN
Setting On
S3 / DIL8
Bypass
Init STBY
AK: SVNT
STBY Venting
AK: STBY
AK: SATK
AK: SKCH
AK: STBY
AK: STBY
Accuracy
Adjusting
check
Fig. 37
Product Guide
70 User's Guide
Information
An active edge at the digital input SET DEVICE OK does not switch the AVL Fuel
Mass Flow Meter to the state Init Pause.
The following error conditions are checked during the state Init Pause:
15 Tank overflow
23 Error NV-RAM
26 Error flow sensor
27 Communication calibration sensor (if available)
28 Error EEPROM
29 General device error
30 Communication error flow sensor
31 Fan blocked
After completing the state Init Pause the AVL Fuel Mass Flow Meter changes to
the state Pause.
3.2.2.2 Pause
In the state Pause, the automatic fuel refill as well as the fuel pump is switched
off, therefore there is no outlet pressure available. The device does not generate
any measurement values.
The state Pause is always taken up statically. After switching on and exiting the
state Init Pause the device takes on at least temporarily the state Pause.
If DIL5 of switch block S3 is set to position "Off" (see also Auto-STBY on
page 72) or the emergency stop circuit is open, this state is static, i.e. the state
can only be changed by sending a related request to the device (e.g. AK
command STBY, closing the emergency stop circuit or an edge at input "Set
Device OK" at a statically closed emergency stop input).
If compressed air is available, an external venting is possible in this state, if not,
the related inputs are switched passive. The error "No compressed air" is not
activated in this state. The error is only activated when the change of state is
requested (e.g. AK command STBY via emergency stop).
Due to safety reasons the valves are in a state which allows a venting with an
external conditioning unit, but this is not recommended, because the internal
venting cannot be ensured.
In the state Pause the fuel circuit is switched pressureless via the internal tank.
Depending on the tank level, only the pressure of the column of the fuel is avail-
able.
The following errors are checked in the state Pause:
At entering the state Pause:
1 Maintenance limit pump
3 Maintenance limit fill valve
Every second:
9 No compressed air (set only at the attempt to change the
state)
11 Density too low
10 Hz cycle:
26 Error flow sensor
At occurrence:
15 Tank overflow
3.2.2.3 STBY
In the state STBY the AVL Fuel Mass Flow Meter provides all measurement
values that are available in the selected domain. The filling of the internal tank is
carried out automatically. The state is static and can only be left by transition to
another state.
A constant outlet pressure can be adjusted with the pressure regulator.
The state STBY can be called in the following ways (provided that: S3/DIL8 is in
position "Off"):
AK command STBY
Positive edge at input SET DEVICE OK, if the emergency stop circuit is
statically closed (from state Pause and Drain)
The following error conditions are checked during the state STBY:
In the transition phase after STBY:
7 Error adjustment
23 Error NV-RAM
28 Error EEPROM
Every second:
9 No compressed air
11 Density too low (will be reset 30 s after the start of the internal
venting)
13 Pump overload
14 Filling rate too low
10 Hz cycle:
26 Error flow sensor
Product Guide
72 User's Guide
Every hour:
1 Maintenance limit pump
3 Maintenance limit fill valve
At occurrence:
15 Tank overflow
3.2.2.4 Auto-STBY
If the Auto-STBY is activated (switch S3/DIL5 in position "On"), the AVL Fuel
Mass Flow Meter immediately after switching on attempts to change to the state
STBY.
Auto-STBY deactivated:
After switching on, the device changes to the state Pause.
STBY activated:
After switching on, the device changes to the state Pause and then to the
state STBY.
WARNING
leaking fuel
If the AVL Fuel Mass Flow Meter is used with the function Auto-STBY, please
make sure that the device changes to the state STBY via the state Pause as
soon as the device is supplied with voltage. This builds up a fuel pressure and
fuel is conveyed.
Do not use this device function if safety risks (e.g. from discharged fuel) could
arise.
3.2.2.5 Venting
In this state an internal venting of the device is carried out.
The state Venting can be reached from the states Pause and STBY.
Venting is a default procedural operating state, i.e. the device returns into
the original state after the set venting time. Here, only the AVL Fuel Mass
Flow Meter is vented internally.
In the state Venting the external fuel circuit (motor circle) is also vented with
the option venting valve.
If the venting is controlled via the AVL Fuel Temperature Control, the
external fuel circuit (motor circle) and the AVL Fuel Temperatur Control are
also vented.
Information
In the state Venting the AVL Fuel Mass Flow Meter is not able to provide the set
outlet pressure. The adjusted pressure at the outlet pressure regulator cannot be
maintained.
Make sure that no venting of the fuel circuit is carried out, if the vapor pressure of
the fuel is higher than the pressure in the overflow tank of the AVL Fuel Mass
Flow Meter (environment pressure). With highly volatile fuels it can result in an
overflow of the tank by an exhalation of the fuel even at only slightly increased
fuel temperatures.
10 Hz cycle:
26 Error flow sensor
At occurrence:
15 Tank overflow
Information
Due to the exhalation of fuels it is also from time to time necessary, even after
the first-time installation, to ensure a sufficient venting of the calibration sensor
for proper functioning of the option calibration unit.
The following conditions have to be met for taking up this state and proper func-
tioning:
The calibration option must be properly installed in the device.
The flow sensor must be fully rested in the relevant quick-release couplings
and must seal completely.
The connection cable to the calibration sensor must be fitted.
After the installation of the option calibration unit the device must have been
switched at least once to the state Pause.
The local gravitational acceleration (g) must have been entered at least
once.
A flow of >3 l/h must be available.
A sufficient venting of the calibration sensor must be ensured (see section
Calibration Unit on page 59).
The valve settings and the maintenance of the pressure in the measurement
circuit by controlled pumping corresponds to the state STBY during the accuracy
check.
The accuracy of the flow sensor will be checked in this state by means of an
additional measurement system consisting of the internal tank and a pressure
sensor.
The determined deviations between the flow sensor and the calibration unit are
used to calculate a mean deviation, which will be output as well as the mean
flow.
Product Guide
74 User's Guide
Mathematical Formula
Descriptions:
ρ Media density
g Local gravitational acceleration
A Cross-sectional area of the tank
hi Distance tank level membrane of the pressure sensor
ΔM K Mass determined by calibration
ΔM S Mass determined by flow sensor
m· ( t ) Instantaneous mass flow of the flow sensor
R Deviation in %
3.2.2.7 Adjustment
The state Adjusting is a temporary state (3 … 5 s), during which a fitted curve is
calculated by a completely performed multi-point calibration (see section Cali-
bration – Adjustment on page 75) in order to compensate the flow sensor.
In terms of valve position and system pressure (i.e. pump mode) and error
checks the state Adjusting is identical with the state STBY, but no measurement
points are output.
3.2.2.8 Bypass
The static state Bypass allows a pressureless fuel supply without operating the
fuel pump. A consumption measurement is not possible.
The state Bypass (signalized as STBY) can only be taken on from the state
Pause and when the DIL-switch is in the respective position (S3/DIL8 "On"). If
system errors 26, 27, 28 or 30 are active, the state Bypass can be selected to
still supply the engine with fuel.
Conditions Compressed air must be available.
The emergency stop circuit must be closed.
3.2.2.9 Draining
The state Draining corresponds to the state STBY, but without automatic fuel
refilling.
The error "Density too low" completes the state Draining – the device switches to
the state Pause.
3.2.2.10 Filling
The state Filling corresponds to the state Venting, but with prior fuel filling. The
filling of further fuel devices (e.g. fuel conditioning or engine circuit) should only
be carried out during the internal venting in the state SFIL.
3.2.4.1 Definitions
Calibration
Calibration is a comparative check with a reference device. The accuracy of the
reference device should be three times higher than the accuracy of the checked
system.
Single-point calibration
A single-point calibration is the calibration of one point within the measurement
range of the measurement device to be calibrated.
Multi-point calibration
A multi-point calibration is the calibration of several points within the measure-
ment range of the measurement device to be calibrated.
Product Guide
76 User's Guide
Adjustment
If a measurement device shows a systematic error, e.g. a systematic nonlin-
earity, this error can be reproduced by multiple calibration runs. Systematic
errors can be compensated with an error correction (compensation). This allows
to compensate for offset, gain or linearity errors of the measurement device.
3.2.4.2 Implementation of the Calibration in the AVL Fuel Mass Flow Meter
The AVL Fuel Mass Flow Meter offers two different methods of calibration:
An accuracy check at a special measurement point within the measurement
range.
A comprehensive multi-point calibration in the range between 0 kg/h and
125 kg/h.
After a multi-point calibration the sensor can be adjusted, if required. Usually it is
sufficient to check the sensor by carrying out a multi-point calibration. However,
the linearity error might be increased or the sensor might be maladjusted during
the adjustment. If you perform an adjustment, make sure that multi-point calibra-
tion is carried out very carefully.
Product Guide
78 User's Guide
Geometric conditions
Fig. 38
Fuel is sampled continuously from a tank at a constant flow rate. At a given time
t1, the tank has a defined cross-sectional area A and a fill height h1. At a later
time t2, the level has fallen to a fill height h2. A pressure sensor that is connected
to the lowest point of the tank determines the pressures p1 and p2 at the times t1
and t2. From these values, the following information is derived for the pressures
p1 and p2 with the fill heights h1 and h2, under consideration of the gravitational
acceleration g and the fuel density ρ:
(1)
(2)
(3)
If equation (3) is multiplied by the cross-sectional area A and if then the result is
divided by the gravitational acceleration g, the following is obtained for the mass
ΔM sampled in the time period Δt = (t2 - t1):
(4)
If equation (4) is divided by the time difference Δt, the mass flow of the reference
measurement at the time t2 is obtained:
(5)
For an arbitrary moment i, the mass flow of the calibration unit is as follows:
(6)
Calibration procedure
In the calibration procedure, the deviation of the mass flow of the Coriolis sensor
from the reference system (calibration unit) is determined at regular intervals
(~300 ms):
(7)
In addition, the mean value and the standard deviation of these deviations since
calibration start are calculated:
(8)
(9)
(10)
Product Guide
80 User's Guide
If the limit is not reached within seven cycles of emptying the device-internal
tank, the calibration will be terminated and the determined uncertainty will be
output.
The figure below shows the single deviation values, the mean value and the
measurement uncertainty interval. As can be seen, the measurement uncer-
tainty interval decreases with time while the number of measured values
increases.
Single deviations, mean deviation and measurement uncertainty of mean devia-
tion over time
Fig. 39
(11)
·
Here, M K is used for the mean flow of the calibration unit during the calibration
procedure. The parameters C1 and C2 can be adjusted with the AK command
ASFP/ESFP, whereby C1 = 0.16 % and C2 = 0.004 kg/h are selected as the
base settings.
If the percental deviation of the calibration procedure is smaller than the relevant
value from the evaluation curve, successful calibration is reported (query using
AK command AKAL).
Fig. 40
Information
The measurement values of the integral measurement are lost upon switching
off the device or changing the domain (stop + reset).
Product Guide
82 User's Guide
Information
The device has to be switched to the state Pause to avoid errors in the long-term
metering of the consumption.
The values of the background totalizers can only be retrieved or changed (start,
stop, domain) during the state Pause. A change of the totalizer domain stops and
deletes the corresponding totalizer.
The values are output with one digit after the decimal point. The significance of
this digit after the decimal point is to be seen in connection with the specification.
Tab. 34
3.3.1 Application
Information
The AVL Fuel Mass Flow Meter has to be at the highest point of the fuel circuit to
avoid an overflow of the system.
The following applications are examples, which have to be adapted to the indi-
vidual requirements.
Pay attention to the instructions in the Application Notes "Fuel Consumption
Measurement" (Material No. AT0802E).
Fit drain cocks at the lowest positions of the lines to and from the engine to
simplify and facilitate the draining of the fuel lines.
3
4 11 5 6
7 8
10
Product Guide
84 User's Guide
3
4
5
2
6
7 5
8
3
2 6
3.3.2 Operation
For the various ways of integration of the AVL Fuel Mass Flow Meter please refer
to the section Electrical Connection on page 40.
The AVL Fuel Mass Flow Meter can be controlled in different ways:
Serial RS232 interfaces with AK-compliant protocols allow easy integration
of the most common test bed automation systems. In addition, for safety
reasons, digital control lines are required for proper operation.
The hybrid test bed integration is used in test bed automation systems
which do not allow integration via RS232 interfaces.
Then the system is operated via digital control lines and the data is output
via an analog interface.
PC software (included in the scope of delivery)
Remote Control (optional)
The operation of the AVL Fuel Mass Flow Meter should preferably be accom-
plished via serial communication with an automation system, the PC software or
the remote control.
3.4.1 General
The remote control is a universal control device for various AVL measurement
devices, including the AVL Fuel Mass Flow Meter, AVL Fuel Temperature Control
and AVL Fuel Balance. Depending on the firmware version, two of these devices
can be controlled with one remote control.
This section describes the basic operation of the AVL Fuel Mass Flow Meter by
means of the remote control. For detailed information about the various oper-
ating states and parameters of the AVL Fuel Mass Flow Meter please refer to the
related sections of this manual.
Fig. 44
The inverted field shows the device ID of the device and/or devices, for which
the remote control is configured. If the remote control has not been used, hori-
zontal dashes are shown instead and the remote control must be configured first
(see section Configuration on page 87).
For instructions about the connection of the remote control, please refer to
the section Serial Interfaces (X3/X4) on page 48.
The operation of the remote control is described in the section Operation on
page 86.
Product Guide
86 User's Guide
The setting options of the remote control are shown in the section Configu-
ration on page 87.
The sections Measurements on page 90 through Warnings and Errors on
page 97 illustrate the individual displays of the remote control.
Information
The shown display screens of the remote control are examples, which can differ
from the actually displayed information.
The firmware version of the remote control must be MT 1.25 or later.
Device IDs AVL devices can be identified with the internal device IDs in the menu of the
remote control:
Device Device ID
AVL Fuel Mass Flow Meter 735S
AVL Fuel Temperature Control 753C
AVL Smoke Meter 415S
AVL Opacimeter 439
Tab. 35
3.4.2 Operation
The upper left shows the title of the screen, the device ID and the software
version of the device is displayed to the right of this.
Fig. 45
In the second line, the operating mode and the device state is shown. The oper-
ating mode indicates where the device is controlled from:
You can only communicate with the device (e.g. set parameters), if the operating
mode is set to REMOTE.
In order to control a connected device with the remote control, the remote control
must take over control. This is accomplished by pressing the key READY.
Information
Settings which have not been stored are lost after leaving the screen.
3.4.3 Configuration
Fig. 46
If the remote control has not been configured, dashes appear in the black field
after switching on.
Product Guide
88 User's Guide
1. Select MENU to go to the main menu and use the cursor keys ( and )
to select the menu item Configuration.
Fig. 47
Fig. 48
3. Press SET, select the combination of connected devices with the cursor
keys ( and ), confirm your selection with ENTER.
4. Select an interface with the cursor keys, press SET and select the con-
nected device, confirm your selection with ENTER.
5. If necessary, use the cursor keys to go to the display of the baud rate, press
SET and select the desired value.
Information
Make sure that the setting corresponds to the setting of the respective device.
Fig. 49
6. Finally store the changes by pressing STORE and leave the screen with
MENU.
In the main menu of a configured remote control, the left side shows the
available devices, which can be selected, and on the right side the options
CONTRAST, LANGUAGE, PASSWORD NEW and CONFIGURATION are
shown.
Fig. 50
7. At the beginning, use the cursor keys to select the menu item CONTRAST
and press the key RUN.
8. Use the cursor keys to adjust the contrast for good readability.
9. Complete the setting with ENTER.
10. In the same way you can select the language.
At the menu option PASSWORD, a valid password can be entered to bypass the
password protection of the parameterization of the AVL Fuel Mass Flow Meter. A
bypassed password protection is indicated by the menu PASSWORD NEW. This
menu is used to view the current password or to enter a new password.
The password protection will only be activated after a restart (switching off
and switching on of the remote control).
If password protection is not desired, select the figure 0 as password.
Product Guide
90 User's Guide
3.4.4 Measurements
If the remote control is configured and only the AVL Fuel Mass Flow Meter is
connected, the title screen is shown for a few seconds after switching on and
then the main screen is shown (Measurement, see below).
1. If the remote control is configured for more than one device, you first have
to select the device in the main menu and go to the respective menu with
RUN.
Fig. 51
2. Press the key RUN in the 735 menu to activate the main screen (Measure-
ment).
Fig. 52
Fig. 53
Fig. 54
Product Guide
92 User's Guide
Fig. 55
Information
A start of the accuracy check is only possible, if the device is in the state STBY
(conditions see section Operating States on page 68.
The remaining time for the calibration is always set to zero.
The message "Accuracy check successful" is not shown on the remote control,
even if the check has been successful. Instead of this message, it always
displays "Accuracy check not passed".
3.4.6 Calibration
The following figure shows the initialization of a multi-point calibration.
Fig. 56
After editing the values and pressing SET, the device initializes the multi-point
calibration list. You can fill this list now by performing an accuracy check repeat-
edly. The results of the individual accuracy checks can be retrieved as summary
on this screen. With a filled multi-point calibration list the adjustment can be
started by pressing ADJST.
Now, the multi-point calibration is ready for the adjustment.
Fig. 57
3.4.7 Adjustment
Fig. 58
On this screen you can revise the adjustments of the current density range which
are already stored in the device and, if necessary, retrieve an adjustment and set
it on the screen Calibration. The currently active adjustment is checked.
Product Guide
94 User's Guide
3.4.8 Parameter
Fig. 59
Fig. 60
3.4.10 Information
Fig. 61
3.4.11 Maintenance
Fig. 62
Information
The fill valve switchings (cycles) are displayed in steps of 100 cycles.
Product Guide
96 User's Guide
3.4.12 Service
Fig. 63
Press VENT., in order to control the valve settings, or TEST, in order to test the
Overflow Float Switch.
Fig. 64
Fig. 65
Fig. 66
3.5.1 General
This section refers mainly to programmers, who have to integrate the AVL Fuel
Mass Flow Meter into a test bed automation system.
The communication via the RS232 interface follows the AK protocol, which is
very common in test bed technology. The protocol frame is defined in the AK
paper "V24/RS232-Schnittstelle – Grundsätzliches". The communication is
based on a master-slave principle.
For instructions on the interface parameters, please refer to section Serial Inter-
faces (X3/X4) on page 48.
Product Guide
98 User's Guide
Byte Function
1 <STX>
2 Don’t care byte
3…6 Function code
Sxxx … Control command
Exxx … Setting command
Axxx … Request command
Rxxx … Request command (repeated query)
7 Space (20 hex) The channel number (bytes 7 to 9) can be
8 K … ID for channel number omitted.
9 0 … Channel number
Space (only if there is a subsequent data block)
data block (depending on the function code)
The individual data are separated by spaces.
n. Byte <ETX> … Command level
Tab. 36
Byte Function
1 <STX>
2 Don’t care byte (identical with the sent command)
3…6 Function code (identical with the sent command)
7 Space (20 hex)
8 Error status
0 No error
1…9 One or more errors
(counts up cyclically, if the error screen changes)
Space (only if there is a subsequent data block)
data block (depending on the function code)
The individual data are separated by spaces.
n. Byte <ETX> … Command level
Tab. 37
Product Guide
100 User's Guide
3.5.2.1 AKEN
Device ID
Tab. 38
Note The device ID of the AVL Fuel Mass Flow Meter ("735S") consists of the field 1
(device ID) and the first character of field 2 ("S").
The firmware version is in field 2 (without the prefix "S").
3.5.2.2 SREM
Control mode – switches the interface to control mode, i.e. setting and control
commands are available via this interface.
Condition None, command is allows permitted.
3.5.2.3 SMAN
Monitoring mode – switches the interface to monitoring mode and sets user to
"none", i.e. this interface allows only request commands.
Condition None, command is allows permitted.
3.5.2.4 ASTF
Provides the number of the active errors (ascending); if there is no error the data
block is omitted.
The third table below shows the assignment of the errors displayed by PUMA 5.x
and PUMA Open (with version earlier than 1.2) to the error messages of the AVL
Fuel Mass Flow Meter.
Notifications
Error Number Description
1 Maintenance limit pump
2 No tank data
3 Maintenance limit fill valve
4 State Bypass
Tab. 39
Product Guide
102 User's Guide
Warnings
Error Number Description
7 Error adjustment
8 Gravity not entered
9 No compressed air
11 Density too low
13 Pump overload
14 Filling rate too low
15 Tank overflow
17 Emergency stop
18 Error fill valve
23 Error NV-RAM
Tab. 40
System errors Assignment of the errors that are displayed by PUMA 5.x and PUMA Open (with
version number below 1.2) to the error messages of the AVL Fuel Mass Flow
Meter.
Tab. 41
Tab. 42
Tab. 42
3.5.2.5 ASTZ
Device state
Tab. 43
3.5.2.6 SPAU
Stops running procedures (SFIL, SDRN, SSRV); switches the device to state
Pause.
3.5.2.7 STBY
Stops running procedures (calibration, adjustment, venting, etc.), switches the
device to state STBY (measurement running, engine can be supplied with fuel).
Condition Not allowed in SSRV and SDRN.
Product Guide
104 User's Guide
Tab. 44
3.5.2.9 ASTX
Error causes
ASTX n
n … Error number
If the existing error n can have more than one cause, the response contains a
binary number, the bits of which are assigned to the error causes according to
the following table, otherwise 0 is returned .
Tab. 45
The AK command ASTX 26 returns a series of digits (0 or 1) in 16 blocks of four
bits, enumerated from left to right.
Tab. 46
Product Guide
106 User's Guide
Tab. 46
3.5.2.10 ASIN
Cause for non-execution of a control command.
If the last control command could not be executed and therefore has been
acknowledged with BS or NA, ASIN returns a 12-bit binary number, the bits of
which are assigned to the causes according to the following table.
Bit Cause
0 Error status
1 Not in standby
2 Not in the necessary state
3 Venting AVL Fuel Temperature Control 753C is running
4 —
5 —
6 No calibration sensor installed
7 Wrong domain
8 Fuel alcohol
9 No acceleration due to gravity entered
10 Flow too low
11 No tank data entered
Tab. 47
Example
Initialization, error and request of a measurement value.
Command Acknowledgment
Initialization
Correct device connected?
<STX> AKEN<ETX> <STX> AKEN 3 735 S1.00<ETX>
3 … Error status
735S … AVL Fuel Mass Flow Meter
1.00 … Firmware version
Error status not equal 0 → request
current error numbers
<STX> ASTF<ETX> <STX> ASTF 3 1 9<ETX>
3 … Error status
1 … Error number (Maintenance limit main pump)
9… Error number (No compressed air)
Periodical requesting of the
measurement value (e.g. 1 Hz)
<STX> ASEM<ETX> <STX> ASEM 3 5 M M S 13.672 13.221 1.735 765.23 21.39<ETX>
If the error status changes, the error 3 … Error status
numbers have to be updated with 5 … Parameter status
the request command ASTF. M … User (the second serial RS232 interface is in control mode)
M … Device status: mass domain
S … Device function: standby
13.672 … Mean over measurement time
13.221 … Moving average
1.735 … Standard deviation
765.23 … Density
21.39 … Sensor temperature
Tab. 48
Tab. 49
Product Guide
108 User's Guide
Tab. 50
Note A change of the domain stops the working totalizer and resets it.
Tab. 51
Tab. 52
The volume flow is derived in the sensor from the mass flow based on the fuel
density.
Conditions Available only in volume domain.
Available only in STBY or SKCH.
Tab. 53
The volume flow is derived in the sensor from the mass flow based on the fuel
density at 15 °C (60 °F).
Conditions Available only in volume domain API.
Available only in STBY or SKCH.
Tab. 54
Product Guide
110 User's Guide
Tab. 55
Tab. 56
Tab. 57
Product Guide
112 User's Guide
Example
Setting of measurement parameters, domain, scaling the analog output and peri-
odical requesting of the measurement value.
Command Acknowledgment
Data acquisition
Control mode
<STX> SREM<ETX> <STX> SREM 0<ETX>
0 … Error status
Set measurement parameters:
1 s measurement time, averaging of
15 measurement values
<STX> EMSP 1 15<ETX> <STX> EMSP 0<ETX>
0 … Error status
Analog output:
scaling on volume
(0 … 60 l/h to 0 … 10 V)
<STX> EDSP 2 0 60<ETX> <STX> EDSP 0<ETX>
0 … Error status
Periodical requesting of the
measurement value (e.g. 1 Hz)
<STX> ASEV<ETX> <STX> ASEV 0 5 R V S 13.672 13.221 1.735 765.23 21.39<ETX>
If the error status changes, the error 0 … Error status
numbers have to be updated with 5 … Parameter status
the request command ASTF.
R … User (this serial RS232 interface is in control mode)
V … Device status: volume domain
S … Device function: standby
13.672 … Mean during measurement period
13.221 … Moving average
1.735 … Standard deviation
765.23 … Density
21.39 … Sensor temperature
Alternative: fast data acquisition for dynamic measurements, e.g. 10 Hz
(optional: change filter setting for the fast data acquisition: EFIP)
<STX> AMEV<ETX> <STX> AMEV 0 13.570<ETX>
If the error status changes, the error 0 … Error status
numbers have to be updated with 13.570 … Instantaneous value
the request command ASTF.
Tab. 58
3.5.4 Maintenance
3.5.4.1 SVNT
Activates venting.
Conditions Permissible only in the state STBY.
Warning 15 must not be active.
3.5.4.2 SFIL
Starts first-time filling and remains in the state permanent venting after filling.
The first-time filling is completed after the init state is left.
Conditions Permissible only in SPAU.
Only the warning 11 (density too low) may be active.
3.5.4.3 SDRN
Starts draining procedure and remains in that state after the draining has been
completed. The draining is completed after the init state is left.
Condition Only permissible in STBY or SPAU.
Tab. 59
3.5.4.5 ACFG
Device Configuration
Tab. 60
Product Guide
114 User's Guide
Tab. 61
Tab. 62
Tab. 63
Note If the limits of maintenance are exceeded, the corresponding notification (notifi-
cation 1 … 3) is output.
Tab. 64
3.5.4.10 AADC
ADC- channels (raw data)
Tab. 65
Tab. 66
Product Guide
116 User's Guide
3.5.5.2 AATW
Working totalizer
Tab. 67
Note The measurement values of the integral measurement are lost at switching off
the device or change of domain (stop + reset). If the STBY is left, the working
totalizer is automatically stopped.
Tab. 68
Tab. 69
3.5.5.5 AHTW
Main totalizers
Tab. 70
Note These are only available in the states SPAU and SSRV.
If the system is switched off in state SPAU, no measurement errors arise.
Product Guide
118 User's Guide
Command Acknowledgment
Data acquisition
Control mode
<STX> SREM<ETX> <STX> SREM 0<ETX>
0 … Error status
Switch device to state STBY:
<STX> STBY<ETX> <STX> STBY 0<ETX>
If the error status changes, the error 0 … Error status
numbers have to be updated with
the request command ASTF.
Measurement domain: select mass
<STX> EDMP 0<ETX> <STX> EDMP 0<ETX>
0 … Error status
Start working totalizer
<STX> EATS 1<ETX> <STX> EATS 0<ETX>
0 … Error status
Request general measurement
values or the summation values of
the consumption periodically (e.g. at
1 Hz)
<STX> ASEM<ETX> <STX> ASEM 0 5 R M T 13.672 13.221 1.735 765.23 21.39<ETX>
If the error status changes, the error 0 … Error status
numbers have to be updated with 5 … Parameter status
the request command ASTF.
R … User (this serial RS232 interface is in control mode)
M … Device status: mass domain
T … Device function: standby / working totalizer active
13.672 … Mean during measurement period
13.221 … Moving average
1.735 … Standard deviation
765.23 … Density
21.39 … Sensor temperature
Tab. 71
Command Acknowledgment
Query working totalizer
<STX> AATW<ETX> <STX> AATW 0 0 1 1.230 147.6 30.412<ETX>
If the error status changes, the error 0 … Error status
numbers have to be updated with 0 … Mass domain
the request command ASTF.
1 … Totalizer started
1.230 … Integral value
147.6 … Integration time
13.570 … Mean flow
Stop working totalizer
Tab. 72
Command Acknowledgment
<STX> EATS 0<ETX> <STX> EATS 0<ETX>
0 … Error status
After the working totalizer has been
stopped the final value can be read
with AATW.
Tab. 72
3.5.6.1 SKCH
Starts a single-point calibration as an accuracy check or during a multi-point cali-
bration.
Conditions Permissible only in the state STBY.
Mass domain
Calibration unit must be available
Gravitational acceleration must have been entered
(AK commands EGTD, EKGP)
Tank area must have been entered
(AK command EGTD)
Volume flow must be >3 l/h.
3.5.6.2 SATK
Start of the adjustment procedure
Conditions Permissible only in the state STBY.
A complete multi-point calibration must be available.
3.5.6.3 AKAL
Result of the last single-point calibration carried out
Tab. 73
Product Guide
120 User's Guide
Information
The status of the accuracy check is compared to the systematic error curve of
the sensor.
C2
R [ % ] = C1 + 100 ⋅ -------
m·
Tab. 74
3.5.6.5 AKAS
Status calibration procedure
Tab. 75
Tab. 76
Note In event of transportation to another location, this value has to be entered once –
there is no default value.
3.5.6.7 AGTD
Gravitational acceleration and tank data
Due to compatibility reasons, the AGTD provides four parameters, of which the
second and third have no function.
Tab. 77
Note Gravitational acceleration and tank data are requirements for carrying out a cali-
bration.
3.5.6.8 EGTD
Gravitational acceleration and tank data
Tab. 78
Note Gravitational acceleration and tank data are requirements for carrying out a cali-
bration.
3.5.6.9 AKLE
Result of adjustment
Tab. 79
Product Guide
122 User's Guide
3.5.6.10 ASGJ
Valid adjustments
In ASGJ the optional parameter n selects the desired adjustment – a call without
this parameter returns the current adjustment.
Tab. 80
3.5.6.11 ESGJ
Initialization of the current multi-point calibration with the data of an existing
adjustment.
Tab. 81
3.5.6.12 AMKS
Status multi-point calibration
Tab. 82
Note Provides values, if a multi-point calibration has been initialized (EMKP, ESGJ).
The command provides uncorrected deviations [kg/h] to the calibration unit or
set-point of the flow [kg/h] of still missing calibration points.
Tab. 83
Product Guide
124 User's Guide
3.5.6.14 ASEK
Status single-point calibration
ASEK n
n=1…8
Tab. 84
3.5.6.15 AKMW
Calibration measurement values
Tab. 85
Tab. 86
Command Acknowledgment
Accuracy Check
(only available in mass domain)
Control mode
<STX> SREM<ETX> <STX> SREM 0<ETX>
0 … Error status
Switch device to state STBY:
<STX> STBY<ETX> <STX> STBY 0<ETX>
If the error status changes, the error 0 … Error status
numbers have to be updated with
the request command ASTF.
Start accuracy check
<STX> SKCH<ETX> <STX> SKCH 0<ETX>
0 … Error status
Periodically acquire measurement
values or calibration (e.g. with a rate
of 1 Hz)
<STX> ASEM<ETX> <STX> ASEM 0 5 R M U 13.672 13.221 1.735 765.23 21.39<ETX>
If the error status changes, the error 0 … Error status
numbers have to be updated with 5 … Parameter status
the request command ASTF.
R … User (this serial RS232 interface is in control mode)
M … Device status: mass domain
U … Device function: calibrate. After completion of the calibration: S
13.672 … Mean during measurement period
13.221 … Moving average
1.735 … Standard deviation
765.23 … Density
21.39 … Sensor temperature
<STX> AKAS<ETX> <STX> AKAS 0 6 # 5637.5 15.241<ETX>
If the error status changes, the error 0 … Error status
numbers have to be updated with
6 … Calibration phase
the request command ASTF.
5637.5 … Pressure of the calibration sensor
15.241 … Instantaneous mass flow
Querying of the results after the
automatic completion of the accu-
racy check:
<STX> AKAL<ETX> <STX> AKAL 0 0.81 15.361 1<ETX>
0 … Error status
0.81 … Deviation from the calibration unit
15.361 … Mean mass flow during calibration
1 … Accuracy check successful
Tab. 87
Product Guide
126 User's Guide
Command Acknowledgment
Multi-point calibration with subse-
quent adjustment
(only available in mass domain)
control mode
<STX> SREM<ETX> <STX> SREM 0<ETX>
0 … Error status
Switch device to state STBY:
<STX> STBY<ETX> <STX> STBY 0<ETX>
If the error status changes, the error 0 … Error status
numbers have to be updated with
the request command ASTF.
Initialize multi-point calibration
<STX> EMKP 030130 WIESEL 3 <STX> EMKP 0<ETX>
17<ETX>
0 … Error status
Start accuracy check several times
(in this case twice)
<STX> SKCH<ETX> <STX> SKCH 0<ETX>
0 … Error status
Retrieve multi-point calibration state
(provides a rough overview of the
calibration points for the following
eight single-point calibrations)
<STX> AMKS<ETX> <STX> AMKS 0 0.7 0.5 #7 #9 #11 #13 #15 #17<ETX>
If the error status changes, the error 0 … Error status
numbers have to be updated with 0.7 0.5 … Deviation from the calibration unit
the request command ASTF.
#7 … #17 … Flow set-points, which still must be calibrated. Detailed
information can be requested with the request command ASEK n (n
= 1 … 8).
Tab. 88
The respective flow set-points are used to determine the deviations with the
accuracy check procedure. These deviations are automatically stored by the
AVL Fuel Mass Flow Meter. After completing the multi-point calibration, an
adjustment can be started.
Command Acknowledgment
<STX> SATK<ETX> <STX> SATK 0<ETX>
0 … Error status
A few seconds later you can request
the results of the adjustment:
Tab. 89
Command Acknowledgment
<STX> AKLE<ETX> <STX> AKLE 0 # 030130 DIESEL1 0.09<ETX>
If the error status changes, the error 0 … Error status
numbers have to be updated with # … Number of existing partial calibrations (invalid after an adjust-
the request command ASTF. ment)
020130 … Date of current adjustment
DIESEL1 … Id of the current adjustment
0.09 … Maximum deviation
Tab. 89
3.5.7 Service
3.5.7.1 SRES
Reset – stops all running procedures, switches the device to power-on state.
Condition None, command is allows permitted.
3.5.7.2 SSRV
Service mode – switches the device into the service mode.
Condition State SPAU
NOTICE
Product Guide
128 User's Guide
Tab. 90
Note "Active state" and "passive state" refer to the name of the output, rather than the
logic level.
Number Output
1 Pump
2 (no function)
3 Fill valve
4 Stop valve
5 (no function)
6 External venting valve
7 Shutoff valve
8 Venting disable
9 Venting request
10 Status LED
11 DIO out 0: Device OK
12 DIO out 1
13 DIO out 2
Tab. 91
Tab. 92
Tab. 93
3.5.7.6 AIDE
Internal digital inputs
Tab. 94
Product Guide
130 User's Guide
Note "Active state" and "passive state" refer to the name of the output, rather than the
logic level.
3.5.7.7 APMP
Pump parameter
Tab. 95
3.5.7.8 EPMP
Pump parameter
Tab. 96
Tab. 97
Tab. 98
3.5.7.11 SREE
Read the permanently stored variables out of the EEPROM.
Condition SPAU and error 23 (NV-RAM).
Product Guide
132 User's Guide
Tab. 99
3.5.7.13 ESCK
Start the test procedure for the float switch
Tab. 100
3.5.7.14 ASCK
Test float switch
Tab. 101
Tab. 102
Tab. 103
Product Guide
134 User's Guide
3.5.7.17 AMOD
Register contents of the flow sensor
(The detailed assignment of the addresses and the permitted value ranges are
described in the Micro-Motion Modbus-Manual.)
Query
Field Description Format
1 Data format ASCII character
A … ASCII string
C … Coil
F … Float register
H … Holding register
I … Input register
2 Address integer
Tab. 104
Response
Field Description Format
1 Value According to field 1 of the query
Tab. 105
3.5.7.18 EMOD
Set the registers of the flow sensor
(The detailed assignment of the addresses is described in the Micro-Motion
Modbus-Manual.)
Tab. 106
Example
Query of the device ID
If the terminal program prompts you with 735S>, enter the following command:
AKEN <Enter>
The device responds with the following output:
735 Sx.xx
735S>
The device is ready again and waits for the next input.
Product Guide
136 User's Guide
Information
Before any maintenance work, switch the device to the state Pause. Read the
manometer to check that the fuel pressure is dropped completely and disconnect
all communication lines which could change the operating state (digital I/O,
COM1 and COM2).
Check the safety devices regularly (according to the maintenance plan):
compressed air switch and overflow float switch.
At suitable intervals (weekly), perform an eyeball check of the leak tightness of
the internal and external fuel lines.
Manual switching of the pilot valve is not permitted.
Product Guide
138 Maintenance and Troubleshooting
WARNING
Danger of explosion
The device must always be vented and filled up with fuel!
The device must not be drained after it has been operated with Otto fuels or
alcoholic mixtures (e.g. for a longer standstill period).
If draining is necessary, (e.g. for transportation or storage) make sure to observe
the following instructions:
1. Drain the device completely.
2. Fill up the device with inert-gas.
3. Seal the five hydraulic connectors (except connector K, compressed air)
hermetically.
These steps must only be followed if the device was filled up with Otto fuels or
alcoholic mixtures!
Information
The AVL Fuel Mass Flow Meter can also be drained by manually closing the fuel
inlet line in state STBY. The device will be drained until the warning "Density too
low" appears.
After the complete filling of the internal device tank, the device performs an
internal venting and remains in the state SFIL (permanent venting), until it is
transferred into another state.
Now, all warnings should have been reset.
Information
Ensure proper fuel supply when sending the command for the first filling.
The first filling can also be accomplished in the state STBY (AK command
STBY), but the venting has to be done manually. The state STBY may have to
be activated several times, because the device the device might be set into the
state Pause due to the warning "Density too low".
2. The filling of other fuel devices (e.g. fuel conditioning or engine circuit)
should be done during the internal venting in state SFIL. Always ensure a
good venting of all fuel lines before you switch the device into the state
STBY!
3. Now, you can switch the device into the state Pause (AK command SPAU).
4. If you want to adjust the outlet pressure regulator, switch the device into the
state STBY. Now, adjust the output pressure regulator of the AVL Fuel Mass
Flow Meter to the desired output pressure:
– Return line pressure in combination with a conditioning system
– Supply pressure in systems without return lines
If there is no calibration available for the used fuel, the warning "Adjustment
error" is shown, after the system switches into the state STBY for the first
time.
Make sure that the required minimum pressure of >0.05 bar is maintained!
Product Guide
140 Maintenance and Troubleshooting
NOTICE
There are two types of the device AVL 735S, which differ in terms of fuel resis-
tance. See Scope of Delivery on page 15.
Please absolutely make sure that only the appropriate spare parts may be used
for the type 7351CME:
Type "alcohol-resistant" (M100).
The following section contains instructions for the replacement of the following
.
components:
Fuel Pump Standard M20 on page 141
Fuel Pump Alcohol-Resistant M100 on page 142
Pressure Regulator on page 143
Flow Sensor on page 144
Tank Level Sensor on page 145
Fill Valve on page 146
Controller Board on page 147
Filter Mat on page 147
1. Drain the AVL Fuel Mass Flow Meter and switch it off.
2. Disconnect the device from power supply
(socket SUPPLY 24 V).
3. Unscrew the pump cable.
4. Remove the pump pressure line.
At works at the line fittings, always counter with a second wrench at the fit-
ting bodies!
– Open the swivel nut of the pressure line.
– Pull out the line upwards.
5. Loosen the swivel nut of the pump suction line.
6. Pull out the pump upwards.
Information
7. Unscrew the FC sealing ring and the swivel nut from the pump suction
thread and loosen the screw fitting on the pump pressure side.
The screw fitting of the pump pressure side has to be mounted again on the
replacement pump, the gasket of fluor rubber has to be replaced with the
supplied gasket.
8. Clean the screw fitting at the thread.
9. Seal the screw fitting with a liquid, fuel-resistant sealant (e.g. Loctite 572).
NOTICE
The must never be clamped in its circumference – this would cause deforma-
tions of the pump housing, thus destroying the pump.
10. Assemble the new pump in reverse order of the above steps.
11. Reattach the pressure line.
12. Put the device into operation again.
13. Carry out an eyeball check for leak-tightness.
14. Reset the operating counter to 0.
Product Guide
142 Maintenance and Troubleshooting
1. Drain the AVL Fuel Mass Flow Meter and switch it off.
2. Disconnect the device from power supply
(socket SUPPLY 24 V).
3. Unplug the pump cable.
4. Loosen the angle bracket at the mounting plate: Screws with hexagon
5 mm.
5. Remove the pump pressure line.
At works at the line fittings, always counter with a second wrench at the fit-
ting bodies!
– Open the swivel nut of the pressure line.
– Pull out the line upwards.
6. Loosen the swivel nut of the pump suction line.
7. Pull out the pump upwards.
Information
NOTICE
The pump must never be clamped in its circumference – this would cause defor-
mations of the pump housing, thus destroying the pump.
10. Assemble the new pump in reverse order of the above steps.
11. Reattach the pressure line.
12. Put the device into operation again.
13. Carry out an eyeball check for leak-tightness.
14. Reset the operating counter to 0.
HINWEIS
4
A
1. Drain the AVL Fuel Mass Flow Meter and switch it off.
2. Disconnect the device (A) from power supply
(socket SUPPLY 24 V).
3. Unplug the plugs of both pressure sensors from the pressure regulator and
note the position for the later assembly.
4. Loosen the swivel nuts of all pipe lines with diameter of 8 mm (wrench
15/16 mm) and disconnect the pipe lines carefully from their screw fittings.
5. Loosen the swivel nuts of the pipe line with diameter of 10 mm (wrench
19/18 mm) and disconnect the pipe line carefully from their screw fittings.
6. Loosen the pressure regulator form the angle bracket at the mounting plate:
Screws with hexagon 5 mm.
7. Remove the pressure regulator. At the same time push the tank return line
slightly away.
Information
Product Guide
144 Maintenance and Troubleshooting
WARNUNG
Fuel overflow
The float switch must only be replaced by specially trained staff (e.g. AVL service
engineer).
Handling the float switch without special training may lead to personal injury and
physical damage.
WARNING
Fuel overflow
The float switch must only be replaced by specially trained staff (e.g. AVL service
engineer).
Handling the float switch without special training may lead to personal injury and
physical damage.
1. Drain the AVL Fuel Mass Flow Meter and switch it off.
2. Disconnect the device from power supply
(socket SUPPLY 24 V).
3. Remove the cover from the electronic housing.
4. Loosen the connectors of the sensor cable.
Refer to the table below for the pin configuration.
5. Loosen the die PG screw fitting and pull out the sensor cable.
6. Turn around the electronic housing and remove the ground cable with a
slotted-screw driver.
7. Loosen the connector pieces with the wrenches 18 mm and 19 mm.
8. Loosen the sensor mounting screws and replace the sensor (hexagon
wrench 5 mm and box wrench 10 mm).
9. The assembly is made in reverse order.
10. Finally carry out an eyeball check for leak-tightness.
Tab. 107
1 .....Valve
2 .....Calibration block
3 .....Sensor
Fig. 68
1. Drain the AVL Fuel Mass Flow Meter and switch it off.
2. Disconnect the device from power supply
(socket SUPPLY 24 V).
3. Loosen the screw fitting of the calibration block (hexagon wrench 5 mm)
and disassemble the calibration block.
Information
Product Guide
146 Maintenance and Troubleshooting
4. Remove the plug of the tank level sensor after loosening the screw fitting
(cross-recessed).
5. Disassemble the tank level sensor with wrench 22 mm.
6. Install a new sensor. Do not tighten the screw fitting too strongly, low hand
power is sufficient!
7. The remaining assembly is carried out in reverse order.
8. Finally carry out an eyeball check for leak-tightness.
1 .... Valve
2 .... Calibration block
3 .... Sensor
Fig. 69
1. Drain the AVL Fuel Mass Flow Meter and switch it off.
2. Disconnect the device from power supply
(socket SUPPLY 24 V).
3. Loosen the screw fitting of the calibration block (hexagon wrench 5 mm)
and disassemble the calibration block.
Information
4. Remove the plug of the tank level sensor after loosening the screw fitting
(cross-recessed).
5. Disassemble the tank level sensor with wrench 22 mm.
6. Loosen the plug of the fill valve cable from the valve (with a screw driver).
7. Disassemble the solenoid spool after removing the nut (SW 19 mm).
Information
Do not tighten the screw fitting of the fill level sensor too strongly, low hand
power is sufficient!
Product Guide
148 Maintenance and Troubleshooting
Information
4.5.1.1 Notifications
This section is only used as a short overview.
For a detailed description with instructions for the correction and spare part
numbers refer to section "Help for Troubleshooting" in the Online Help.
Tab. 108
Product Guide
150 Maintenance and Troubleshooting
4.5.1.2 Warnings
This section is only used as a short overview.
For a detailed description with instructions for the correction and spare part
numbers refer to section "Help for Troubleshooting" in the Online Help.
Tab. 109
4.5.1.3 Errors
This section is only used as a short overview.
For a detailed description with instructions for the correction and spare part
numbers refer to section "Help for Troubleshooting" in the Online Help.
Tab. 110
Product Guide
152 Maintenance and Troubleshooting
Component State
Tab. 111
Product Guide
154 Maintenance and Troubleshooting
Fig. 70
Tab. 112
Product Guide
156 Maintenance and Troubleshooting
Wearing parts
Tab. 113
Spare parts
NV-RAM x x EE0669 E
Tab. 114
Product Guide
158 Maintenance and Troubleshooting
5 4 1 2 3
7
8
9
10
11
12
13
14
15
16
17
ST
Fig. 71
LED functions on The descriptions in the column Plan refer to the numbering in the figure above.
the controller
The position designations refer to the figure in section General on page 66.
board .
Tab. 115
Product Guide
160 Maintenance and Troubleshooting
Tab. 116
EU1080
J8 / 5
J8 /7
J8 / 8
J8 / 2
J8 /3
J8 / 9
J8 /1
J8 / 4
J8 /6
2R2 10NF 8
+5V +5V +5V +5 V
Fig. 72
R75
R91
R93
R74 L7
68R
TL16C452 TL16C452
150R
22K C81 LT1180
BAS21 BAS21 0.3A X3 / 2
17
18
6 28 24 0 1 2 3 28 24
CTS0 RTS0 CTS0 RTS0 0.1uF
A c 7 25 25 VCC ON_OFF* 2
C83
C71
D43 D42 T19 TLC3702CD 31 DSR0 31 DSR0 0.1uF C1_P
DTR0 DTR0 L9
MMBT3904 5 29 26 29 26 3 4 C9
RLSD0 RLSD0 C1_N
33uF
33uF
BAS21 BAS21 SOUT0 SOUT0 VPOS
22K
b ACE1 ACE1 X3 / 3
R67
30 45 LT1180 30 45 C8 5
0.1uF
C82
C70
I31 RI0 INT0 RI0 INT0 C2_P
D39 D38 e 41 +5V +5V 41 7 6
0.1uF
0.1uF
+5V_2 e SIN0 SIN0 VNEG C2_N
17
18
4 32 HSI(0) 1 0.1uF 10 32 C1 C2 0.1uF
T18 12 CS0 VCC ON_OFF* CS0 uP RS232 X3 / 4
8 MMBT3906 0.1uF
2
C1_P
R92
68R
b 13 12 2 3 4 C114 13 12 11 TR2_ IN TR2_OUT 8
CTS1 RTS1 VPOS C1_N CTS1 RTS1
2 c 5 11 C102 5 5 11
1K2 DSR1 0.1uF C2_P DSR1
RS232(3:0)
J9 / 1 1R 10uF 10K 82K DTR1 0 DTR1 REC2 _OUT REC2_IN X3 / 5
1 LT1366 8 10 7 6 8 10 10 9
COM 1
R96 R85 R78 R72 RLSD1 SOUT1 VNEG C2_N RLSD1 SOUT1
C85 3 6 ACE2 60 C91 C103 0.1uF 6 ACE2 60
RI1 INT1 uP RS232 RI1 INT1 PORT2(2)
+5V TR1_ IN TR1_OUT
C92
12 15
MONITOR(2:0)
I32 62 SIN1 62 SIN1
TR2_IN TR2_OUT
0 X3 / 6
20 0 11 8
10uF
3 CS1 11 3 CS1
R77
4
2K7
+5V 13 REC1 _OUT REC1_IN 14
4N7
J9 / 2 + 2. 5V _1 REC2_OUT REC2_IN
1
0 14 53 0 1 10 9 0 14 53 0
C89
330R DB0 PD0 DB0 PD0 0.3A X3 / 7
EU1218
10K
R63
1 15 52 1 1 15 52 1 GND
R79 DB1 PD1 2 DB1 PD1
+ 5V 2 16 51 2 2 12 TR1_IN TR1_OUT 15 2 16 51 2 I3
PRESSURE TRANSDUCER
DB2 PD2 DB2 PD2 L8
16
10K
R14
0.1uF + 5V 3 17 50 3 3 17 50 3
DB3 PD3 3 DB3 PD3 X3 / 8
+5V_2 C80 HT2012 4 18 49 4 13 REC1_OUT REC1_IN 14 4 18 49 4
EINPRESSBOLZEN M3 DB4 PD4 DB4 PD4
+5V EU2187
+ 2. 5V _1 +2.5V_1 1 27 5 48 5 5 48 5
FPROG
DIL1(7:0)
8
B 14 6 20 47 6 GND 6 20 47 6
4K7 ORXD DB6 PD6 DB6 PD6
R94
R84
220K
150K
8 16 17 7 21 46 7 I39 7 21 46 7
SCHIRM R73 OTXA INRTS DB7 PD7 DB7 PD7
16
2 4 3
IRXA OCD PARALLEL PORT PARALLEL PORT 0.3A
DATA_BUS(7:0)
1 X4 / 1
AR1
AR1
AR1
AR1
AR1
AR1
AR1
AR1
10K
5 TLC3702CD 2 28 63 63
7 2K7 O19K2 CLOCK ERROR ERROR DIL8POL
4.7NF LT1366 3 26 65 59 65 59
J15 / 1 J15 / 2
R80 TEST0 SLCT INT2 SLCT INT2 0 L11
CONTROL SIGNALS 6 I31 13 15 66 58 2N7002 66 58 1 16 S1 1
TEST1 GND BUSY SLIN BUSY SLIN 0.3A X4 / 2
C84 5% 67 57 D 67 57 15 S2 2
I32 I35 PE INIT RES1 / 1 RES1 / 2 PE INIT 2
0 RESET 68 56 68 56 14 S3 3
4 ACK AFD RESET_IN ACK AFD 3 L10
4 S
1 RD 1 55 G 1 55 4 13 S4 4
LPTOE STB LPTOE STB X4 / 3
38 13 38 12 S5 5
Circuit Diagrams
44 0
43 1
42 2
40 19
39 18
38 17
37 16
36 15
41 3
R82 7.372MHZ 820R CTC Q5 IOW IOW X4 / 5
7 FLASH 9 6 1 37 26 1 37 +5V
R65 RTC Q6 IOR IOR
COM 2
33PF 12 14 0 39 39
IO28
IO27
IO26
IO25
IO24
IO23
8 RAM RES Q7 RESET CONTROL_BUS(3) 0 RESET
VDD_VCC
C90 Q2 4 I34 14 4 I33
13
C55
C56
1 34
IO_GCK2
IO_GCK1
IO_GTS1
DIL2(7:0)
C 9 NV-RAM Q8 CLK IO_GCK3 IO_GTS2 CLK X4 / 6
15 5 2 IO1 IO_GSR 33 13
33PF
33PF
10 UART1 Q9
AR2
AR2
AR2
AR2
AR2
AR2
AR2
AR2
10K
1 6 3 IO2 IO22 32 12
nicht Q11 CLOCK CLOCK DIL8POL
2M2
+5 V I28
R95
R97
120K
11 UART2 2 7 5 IO3 IO21 31 11
bestueckt Q12 J14 / 1 DIL2(7:0) 0 X4 / 7
12 UART3 3 8 6 IO4 XC9536 IO20 30 10 16 S1 1
Q13 1
+5 V
CONTROL_BUS(21:0)
9 9 15 S2 2
ADDRESS_BUS(15:0)
13 PARPORT1 7 IO5 IO19
29
I30 J14 / 2 2
10 8 28 8 14 S3 3
14 PARPORT2/UART4 IO6 IO18 3
+5V + 5V +5V 74HC74 7 13 4 X4 / 8
+5V 9 TDI IO17 27 S4
15 ADDR_15 CLK1 J14 / 7 4 EU2187
4 S 10 24 12 S5 5
TMS TDO 5
16 ADDR_16 3 5 + 5V 74HC74 6 11 S6 6
22PF
22PF
C1 11 TCK IO16 23 6
10K
10K
J14 / 9
RESET_IN
2 6 10 10 S7 7
R110
R111
17 MMO 1D S CLOCK 7
4K7
8K2
4K7
CONTROL_BUS(0)
1 Q1 11 9 9 S8 8
R112
R113
R101
ADDRESS_BUS(15:0)
C51
C52
J18 / 2
18 DAC R C1
IO10
IO11
IO12
IO13
IO14
IO15
IO7
IO8
IO9
TL7702 nicht 12 8 J14 / 4 CONT1 /1
19 LOAD DAC 1D S4
2 5 bestueckt I26
+5V 13
12
13
14
16
18
19
20
22
21
RESIN RESMIN 20MHZ R J14 / 6 CONT1 / 2
20 UART4 10uF/35V
21 PARPORT2 7 6 I26 +24V_1 EI0288
11 12 13 14 15 1 2 4 5
J18 / 1
SENIN RES 4K7 CONT1 /3 0.5A 25V/6A /VAR 50 V
C50 20 14 21 1 3
CT VREV R62
CONTROL_BUS(21:0) F1
3 1 I38
PORT0(7:0)
D
0.1uF
7 6 2 0 1 3 5 3 1
10K
RESET1
R114
L6
10uF
DATA_BUS(15:0) 2
0.1uF
+24V_1
9
8
7
6
5
4
3
2
1
68
67
66
65
64
63
62
61
C101
C113
C111
+5V
Vpp
1 VPP 1 VPP
Product Guide
I11 MMO(7:0)
st1
6 22 7 22 74HCT273
I19 CE CE
I13
I14
I18 1 1 0
+5VANA 74HCT573 24 24 1
OE OE R
F11
+5V +5V +5V +5V 1 31 2 31 17 11
0.5A
10uF/35V
ACH7_PO.7
ACH6_PO.6
ACH2_PO.2
ACH0_PO.0
ACH1_PO.1
ACH3_PO.3
NMI
EA*
VCC
VSS1
XTAL1
XTAL2
CLKOUT
BUSWIDTH
INST
ALE_ADV*
RD*
EN WE WE C1
5 10 60 0 3 11
J11 / 1 ACH5_PO.5 AD0_P3.0 C1
C40 4 11 59 1 0 12 0 12 628128 0 3 2 0
ACH4_PO.4 AD1_P3.1 A0 0 A0 0 1D
28F010_smd
28F010_smd
0.1uF 12 58 2 0 2 19 0 1 11 13 1 11 13 0 12 1 4 5 1
J11 / 2 C43
ANGND AD2_P3.2 3 A1 DQ0 A1 DQ0 A0
10K
10K
R21
R20
13 57 1 3 18 1 2 10 2 10 1 11 13 0 2 7 6 2
MAX487E VREF AD3_P3.3 A2 1 A2 1 A1 I_O0
4
4K7
R23
14 56 2 4 17 2 3 9 14 3 9 14 2 10 14 1 3 8 9 3
J11 / 3 VSS2 AD4_P3.4 5 A3 DQ1 A3 DQ1 A2 I_O1
8 VCC 1 15 55 3 5 16 3 4 8 4 8 3 9 15 2 4 13 12 4
R EXTINT_P2.2 AD5_P3.5 A4 2 A4 2 A3 I_O2
7 B 2 16 54 6 4 6 15 4 5 7 15 5 7 15 4 8 17 3 5 14 15 5
J11 / 4 B RESET* AD6_P3.6 7 A5 DQ2 A5 DQ2 A4 I_O3
6 A 3 17 80C196KC 53 5 7 14 5 6 6 6 6 5 7 18 4 6 17 16 6
RXD_P2.1 AD7_P3.7 A6 3 A6 3 A5 I_O4
EU1219 6 5 7 7
FLOW SENSOR
5 4 +5V + 5V 18 52 8 8 13 6 7 5 17 7 5 17 6 6 19 18 19
D TXD_P2.0 AD8_P4.0 A7 DQ3 A7 DQ3 A6 I_O5
A GND 7 6
2 0 19 51 9 9 12 7 8 27 4 8 27 4 7 5 20
P1.0 AD9_P4.1 A8 A8 A7 I_O6
E EINPRESSBOLZEN M3 I9 1 20 50 10 9 26 18 9 26 18 8 27 21 7
D9
P1.1 AD10_P4.2 A9 DQ4 A9 DQ4 A8 I_O7
FPROG
4K7
R22
2 21 49 11 74HCT573 10 23 5 10 23 5 9 26
P1.2 AD11_P4.3 A10 A10 A9
R9
10K
+5V +5V 3 22 48 12 1 11 25 19 11 25 19 10 23 +5V 12 2 19 7
BAS86
PWM1_P1.3 AD12_P4.4 EN A11 DQ5 A11 DQ5 A10 23 A12 D7
4 23 47 13 11 12 4 6 12 4 6 11 25 11
SCHIRM PMW2_P1.4 AD13_P4.5 C1 A12 A12 A11 A11
0 24 46 14 13 28 20 13 28 20 12 4 22 8 10 21 18 6
HSI0 AD14_P4.6 A13 DQ6 A13 DQ6 A12 CS1* 24 A10 D6
+5V 1 25 45 15 8 2 19 8 14 29 14 29 13 28 30 9
HSI1 AD15_P4.7 A14 7 A14 7 A13 CS2 A9
10K 25 17
10K
10K
9 1 5
R31
R30
X24C08 3 2 26 44 3 18 9 3 21 15 3 21 14 3 24 8
R8 HSI2_HSO4 T2CLK_P2.3 A15 DQ7 A15 DQ7 A14 OE* 3 A8 D5
1 8 10 4 17 10 2 16 2 31 29 2 7
A0 VCC A16 A16 15 A15 WE* A7
2 7 11 5 16 11 30 30 2 1 6 4 16 4
A1 TEST 6 7 NC NC A16 NC A6 D4
3 6 6 12 6 15 12 5 5
A2 SCL 16 A5
4 13 7 14 I27 4 6 15 3
10NF
5 7 13
D8
GND SDA A4 D3
14 7
HSI3_HSO5
HSO0
HSO1
P1.5_BREQ*
P1.6_HLDA*
P1.7_HOLD
P2.6_T2UP-DN
HSO2
HSO3
VSS
VPP
P2.7_T2_CAPT
PWM_P2.5
WRL*_WR*
WRH*_BHE*
T2RST_P2.4
READY
8 13 3
1NF
1NF
14 A3
I10
PROG
15 9 12 8 13 2
C7
15 2
BAS86
A2 D2
27
28
29
30
31
32
33
34
35
36
37
38
39
40
41
42
43
3 1 9
C16
C15
I16 10 A1 12
0 1
ADDRESS_BUS(15:0) A0 D1
0 1
2 27 11 0
HSO(3:0) +5V 22 WE D0
2 1
3 20 OE
F 9
HSI(3:0) 4 CE
4 5 6 7 1
NC1
I20
2 26 PM1
PORT1(7:0) NC2
BQ4010
28 14 PM2
6 7 5 4 3 VCC1 GND PM3
+5V
PORT2(7:0)
Bearb
ZM1105
SZT-0481 1 2 10.02 W.M. H.E.S./S.H. CIRCUIT DIAGRAM
DIGITAL CAD Z.Nr.
SZT-0481 1 1 07.02 LIE Gepr K.GRUBER ID.NR. Blatt 1
1 2 3 4 5 6 7 8
+5V +5 V +5V
+24V_ 1 V5_POS1
0.47uF 0.47uF 0.05A 0.3A
+5VANA +5V_1 +15V X2 / 1
SPR3 +5V OUT
Fig. 73
+5V C58 C46
78L05 MMO(7:0) BSP78 F6 L3
470R
470R
1 8
R119
R120
OUT IN
100K 100K 5 1 IN X2 / 3
D 2
R58 R41 GND1-4 +1 5V VENT. REQUEST
+5VANA
F13
A
0.05A
L5
8K2
+15V
R42
7
6
3
2
D34
J10 / 1 8
ROT
+ 5VANA
GELB
3
10uF
0.1uF
ULN2003
LY_T670
LS_T670
8 82K S
9 COMMON
BAS21
T5
C28
SEN1
C18
LED10
LED11
R43 PORT0_2 3 4
4 14 3
J10 / 2 47K 6 47R 1 BSP78
8
R54 R33 AD706 GND 0.3A
7 33K 2 4 X2 / 4
AD706 2 ULN2003 1 IN D 2
I8
SYNC
D32
J10 / 3 PORT0(2) I17 L2
3 13 4
I22
R53
220R
4 8 3
BAS21
I21 GND
10NF
S
4 +5VANA I37 T4
C30
11
10K
-1 2V
R99
-12V 74HC14
PC357 J507 0.3A X2 / 5
4 10 11 4 1 K A
22N 22N HSI(2) +24V IN
+5VANA D3
I40 L1
C47 C26 3 2 BAS86 X2 / 6
12
47N 820K 47N EU2187 STATUS VENT.
14 D4
R36 OP491 I2 L4
+5V C57 C31 13 +5 V +5V
+5V_1 +5V +5V
I21
47K 1M 47K
B 0.3A
R52 R35 R37 11 0.05A X5 / 7
+15V REL5 OUT1:DEVICE OK
R16
470R
470R
1
R115
F4
14
L19
D33
J10 / 5 + 5VANA
470R
R124
8 39K 6
BAS21
SEN2 R51 PORT0_1
4
47K
15005B
J10 / 6 2 +2 4V _1
GRUEN
GRUEN
ULN2003
STATUS
R48 1 2M2 13
AD706 6 COMMON 9
GRUEN
LED6
LED15
R40
LG_T670
9
8
7
3 OP491 7 5 12
D31
J10 / 7 PORT0(1) HSO(1)
I22 5 8 GND
+5V +5V
BAS21
I21 I8 0.3A
0.05A X5 / 2
F10
0.05A
4 OUT2:TRIGGER
11
Maintenance and Troubleshooting
J12 / 1 F2 L14
-12V REL3
1
14
2N7002
470R
22N 22N R125
+5VANA D 6
C37 C35 T21
47N 47N
820K
15005B
6 G S
C R45 STATUS ULN2003
+5V C38 C36 COMMON 9 13
GRUEN
LED16
PORT1(7:0)
LG_T670
+5V_1
9
8
7
7 10
47K 1M 47K HSO(2)
ULN2003 8 GND
R38 R47 R46 74HC14
COMMON 9 +5V +5V
J12 / 2 I8 0.05A 0.3A X5 / 8
+15V 7 9 8 2 15
D28
J10 / 9 + 5VANA EU1770 OUT3
I40 8 GND
8 39K F3 L18
REL4
1
BAS21
SEN3
14
R44 PORT0_0 I8
470R
R126
4
J10 / 10 47K 6 6 +24V_1
R34 7 2M2
AD706 9
R39
15005B
5 OP491 8 6 1 1
ULN2003 0.3A
D27
J10 / 11 PORT0(0) +5V +5V +5V 13 5 2 X5 / 3
10 COMMON 9
GRUEN
LED17
EU1907
9
8
7
4 13 4 3
BAS21
I21 HSO(3) 2 L12
4 8 GND
11 S1
DIGITAL I/O
I8
-12V
470R
470R
470R
R123
R122
R121
+ 24 V_ 2 +24V_1
J3 / 1 3K3 0.05A
+5V +5V +5V
R17
6 1 1
GELB
GELB
GELB
D MV1 ULN2003 F7
LY_T670
LY_T670
LY_T670
9 COMMON 5 2
MMO(7:0) 0.3A
LED14
LED13
LED12
BSP78 X5 / 6
D18
ROT
12 5 4 3
LED1
BAS21
FILL VALVE J3 /2 0 2 COMMON IN
10K
10K
10K
2 D IN 1 8
R104
R102
R103
GND L17
74HC14 S2
I37 PC357 J507 0.3A X5 / 1
4 3 4 1 K A
+ 24 V_ 1 PORT0(5) IN1:RESET
3 +5V ANA ULN2003 I40 D14
J3 / 3 3K3 S BAS86 L16
9 COMMON 3 2
R25 10uF
MV2 T8 11 6 D15
C13 74HC14 I6
GND
8
BSP78 PC357 J507 0.3A X5 / 4
D20
ROT
TURN OFF 4 1
LED2
0.1uF 6 5 K A
BAS21
J3 /4 1 I37 IN2:TRIGGER
2 D IN 1 I40 D12
PORT0(6) BAS86 L15
C12 3 2
ULN2003
DATA_BUS(15:0)
CONTROL_BUS(21:0)
21 32 26 9 COMMON D13
+ 24 V_ 1 74HC14 I5
3 +2.5V 10 7
D23
ROT
19 19 27
LED3
D11
BAS21
VENT.BYPASS J3 /6 2 LDAC SREFH J19 / 1 J19 / 2 0.1uF I4
2 D IN 1
EU2061
C17
0 16 29 C11 A_OUT
D0 SREFL
1 15
D1 220K 22NF 0.3A
0.1uF
+ 24 V_ 1 3 2 14 30 X6 / 1
J2 / 1 3K3 D2 REFL R11
S 3 13 C3
MV4 R7 D3 L20 X6 / 2
T9 4 12 DAC7641YB
D4 10K 22K 22K
5 11
RETFLOW D5 R12 R3 R2
D7
BSP78 X6 / 3
ROT
6 10 I12
LED4
BAS21
J2 /2 3 D6 47PF
2 D IN 1 7 9 +15V +5V
(EXT) D7 0.1uF X6 / 4
8 8 C4
D8 10K
9 7 22 C5 +1 5V
D9 OUT R13 8 X6 / 5
+ 24 V_ 2 3 10 6
J1 /2 3K3 D10 8 FREQ
4K7
S 11
R15
5 23
R1 D11 SOUT X6 / 7
MV5 12 4 2
T3 D12 10K J16 / 1 J16 / 2
ANALOG OUT
13 3 1 6
SEC.VALVE D13 AD706
D1
BSP78 R10 7 47R X6 / 8
ROT
14 3
LED5
2 AD706
BAS21
(EXT) J1 / 3 6 D14 R4
F 2 D IN 1 15 1 5 2N7002
EU2058
D15 I1 0.05A 0.3A X6 / 6
D
AGND VSS DGND I1 EU2187
T6
4 0.1uF F8 L21
3 25 24 31 4 G S
S HSO(0)
C6
-12V -1 2V
T1
OTHER PURPOSE WITHOUT THE PRIOR WRITTEN PREMISSION OF AVL LIST GMBH.
SUPPLY_15V
+24V_1 BAS86 +15 V BAS86
+5V
+15V +15V
D47 D41
0.1uF
Fig. 74
MBR745 EI0288 7815CT 7805CT +15V
J6 / 1 3.15A/T 1A/T 25V/6A/VAR50V 10R 22R
1 3 KK-EM014 8 C77
R98 IN KK-EM0148 OUT R86 IN OUT SUPPLY_5V
R71
1W
82R
+24V D46 5W 5W
A F15 F16 GND1 GND1 8
LT1027ECS8
0.1uF
SPR1 SPR2 8
47uF
C94
33uF
C86
33uF
C61
33uF
C75
D45
D37
IN 3
2 L31 C63 4 T16 3
OUT 4K7
F14
1
R68
SUPPLY 1 1 BC817-40
30VAC
0.1uF
C95
C73
C74
6.3A/T
C93
NR AD706 R69
TVS328
MPTE-5
2
MPTE-15
0.1uF
0.1uF
0.1uF
D48
2
1000uF/40V
C104
2 3
J6 / 3 GND TRIM I29
C76
I25
0V 4
REF2
10K
R56
+5V
47PF
EI0288
J6 / 2 25V/6A/VAR50V
1 3
EU0235
+5VANA
470R
+5V
10uF
R116
GRUEN ULN2003
MBR745
9 COMMON +15V
LG_T670
D44
2 L29 5533_ER0207 16 1
C62
8
1 LED7 GND
8
1N4007 TEST REF1
10K
4
R105
74HC14 I37
D40 7 5
BAS21 J507 PC356 7 47R
TVS328 EI0210 EI0288 A1 A K 1 4 13 12
B 25V/6A/VAR50V HSI(1) AD706 R70 +2.5V
4 1 1 3 2 D16
6
D17 I40
D30
CB B 2 3
D36 5
I29
24V
3 CG PSG 2 8
C64
BAS21
I7
10K
R61
B1 4
L27 +24V_1 3
SOCKEL=ER0208=
2 L28
47PF
6
+15V 9
0.1uF
BEFESTIGUNG
C59 +5V +5V REL2
5
TEST_PUMP
8
+24V_1
PORT0(3)
4
0.3A
D35
0.5A X7 / 1
2
BAS86
4K7 F5 L24
1
LT1366 0.3A
10K R50 X7 / 2
1 3 +5V
INA168NA OUT R60
C42
I24 L22
+5V
I23 J17 / 1 0.3A X7 / 3
0.47uF
C 4
J17 / 2 EU21 86
R59
330K
470R
3
R117
0.1uF
ULN2003 ROT J17 / 3 L23
GND
COMMON 9 LS_T670 +5V
C60
2 15
BAS21
2
D25
8 GND LED8
KK-EM0148 0.5A J4 / 1
I37
10K
R106 F12
74HC14 74HC14
10K PC356 J507 J4 / 2
c 12 13 2 1 4 1 K A
T14 T10 PORT1(5) R107 EU2057
S G
PRESSED AIR
R27
1K
10K
b
R18
3 2
D e +24V_1 ER0116
T15 I15
680NF
10R
S G 1
DJA-5001 DJA-5001 IRF4905 R26 +24V_2
C109
5
J5 / 1 0.01R 204uH/5A e
5 1 5 204uH/5A 1 D 9
R49
2W 6 2 6 2 b 2 13
IRF4905 MMBT3906
1K2
R19
7 3 7 3 +5V 6
T11
D19
8 4 8 4 c 10
R55
24V
3K3
204uH/5A 204uH/5A 3
BAS21
D
L25 L26 7 14
PUMP
470R
SOCKEL=ER0192=
11
Product Guide
D29
R118
ULN2003 ROT
MBR745
c +5V
COMMON 9 LS_T670 4
KK-EM0148
T12
BAS21
10K 3 14 8
GELB
MMBT3904
C39
C72
R28
LY_T670
PORT2(5)
1000uF/40V
1000uF/40V
BAS21
b
D21
D22
8 GND LED9 12 2c
ERDE
LED18
J5 / 3
e I37
EU205 8
10K
REL1
R109
74HC14 74HC14
+5V +5V
BAS21
10 11 8 9
+24V_1
PORT0(4) R100 R108
I41 I41 D50
D6
+12V
D49
0.5A 10K
R5
J13 / 1 J7 / 2
5K6
+15V
BAS21
74HC14 R32
680NF
680NF
EU1682
BAS86
F9 ULN2003
C99
J13 / 2 5K6
C110
3 4 78M12AC 9 COMMON
TANK OVERFLOW
R6 HSI(3)
I41 1 3 16 1
J13 / 3 IN OUT MMO(7) X1 / 1
S G 8
GN D
D5
EU1223 GND1 T13
I8 X1 / 2
2 SPR4 D
BAS21
E VPP
VP0808M
1uF/16V
C20
+24V_2 BSP452
2N2 Vpp X1 / 3
+5V 4 VBB OUT 1
C105
8 X1 / 4
D2
V12_NEG
R29
10K
22K 120K 120K
ULN2003 3 IN 2
R87 R88 R89 6
MMO(6) X1 / 5
GND
9 COMMON
1uF/16V
BAS21
C19
ANA_GND1 7
0.1uF +15V 11 6 LT1366
0.3A 2R2 T2 X1 / 6
SEC. VALVE
8 5
R76 GN D
ANA_GND2 DIG_GND1
C106
L30 I8 I24 X1 / 7
-12V ANA_GND3 DIG_GND2
LT1054 c
T17 4
3 VPOS 14
X1 / 8
MMBT3904
FB_SD b EU2187
33uF 4
CPOS
C96 6 +5VANA +15V -12V
VOUT 11 e
CNEG
12 +5V
C88
VREF
13 5
OSC GND
R90
0.1uF
3K3
33uF
F
33uF
I36
0.1uF
0.1uF
0.1uF
0.1uF
0.1uF
0.1uF
0.1uF
0.1uF
0.1uF
0.1uF
0.1uF
C27
C32
C49
C34
C48
C33
C87
0.1uF
0.1uF
0.1uF
0.1uF
0.1uF
0.1uF
0.1uF
0.1uF
0.1uF
0.1uF
0.1uF
0.1uF
0.1uF
0.1uF
0.1uF
0.1uF
0.1uF
0.1uF
0.1uF
0.1uF
C97
C98
C66
C65
C10
C68
C23
C25
C67
C54
C53
C24
C29
C79
C78
C41
C45
C69
C44
C14
C21
C22
C100
C112
C107
C108
Bearb
ZM1105 +
SZT-0481 1 2 10.02. W.M. H.E.S./S.H. CIRCUIT DIAGRAM
POWER
SZT-0481 1 1 07.02. LIE Gepr K.GRUBER CAD ID.NR. Z.Nr. Blatt 3
OTHER PURPOSE WITHOUT THE PRIOR WRITTEN PREMISSION OF AVL LIST GMBH.
Fig. 75
Fig. 76
Product Guide
166 Maintenance and Troubleshooting
5 Technical Data
Installation location
Only for indoor use
Rel. air humidity
max. 80 % up to 30 °C, above this range linearly decreasing until a rel. air
humidity of 50 % at 40 °C is reached
Protection class by housing
IP24
Transportation and storage conditions
Temperature: -25 ... +55 °C
Relative air humidity: max. 80 %, no rain fall
Recommended measurement range
0 … 125 kg/h
Negative flows can be measured.
Measurement uncertainty of mass flow
AVL fuel consumption meter devices are tested before shipment in a compre-
hensive manufacturing test on a special test bed. Thereby, all systematic and
random components of the measurement uncertainty are determined with a
confidence level of 95 %.
The values of the measurement uncertainty stated herein are limits for the manu-
facturing test, which in practice are significantly lower most of the time.
Systematic measurement uncertainty (linearization error):
us ≤ 0.12 %
Stochastic measurement uncertainty uZ = 2σ (scattering) with a damping
constant of 0.1 s without flow:
Tab. 117
The total measurement uncertainty pursuant to DIN 1319 comprises the compo-
nents uS, uZ and the measurement value (MV) as follows:
⎛ kg 2
u z ------ ⎞
2 ⎜ h ⎟
u ges [ % ] ≤ u s [ % ] + ⎜ 100 ⋅ -----------------------⎟
⎜ MW ------ ⎟⎠
kg
⎝ h
Product Guide
168 Technical Data
Um max
… measurement uncertainty at max. measured consumption [kg/h]
m … average consumption [kg/h]
m ss … average of consumption at Start and Stop [kg/h]
m … consumed mass [kg]
Fuels
Standard implementation = Type 7351CST (ST=Standard)
(Otto and diesel fuels, alcoholic mixtures up to 20 %)
– Otto fuels, complying to the minimum requirements as described in
EN228
– Diesel fuels, complying to the minimum requirements as described in
EN590
– Alcoholic additives up to a maximum of 20 % volume (methyl and ethyl
alcohol)
– Biofuel/biodiesel additives up to a maximum of 20 % volume according
to the minimum requirements as decribed in EN14214
The model "Standard" M20 = Type 7351CST is conditionally perma-
nent for biofuel additives specified according to EN14214. For standstill
times that last longer than 24 hours, the device needs to be emptied
completely, rinsed with customary Otto or diesel fuel and left in a filled
up state.
Model FlexFuel = Typ 7351CME (ME=methyl/ethyl alcohol)
(Otto and diesel fuels, alcoholic mixtures up to 100 % and biofuels)
– Otto fuels, complying to the minimum requirements as decribed in
EN228
– Diesel fuels, complying to the minimum requirements as decribed in
EN590
– Alcoholic additives of up to 100 % volume (methyl, ethyl alcohol)
– Biofuel/biodiesel additives of up to 100 % volume complying to the
minimum requirements as decribed in EN14214
Output pressure
Adjustable, depending on media and flow
10 … 50 kPa (typical at zero consumption)
Maximum 100 kPa
Filling pressure
Depending on the maximum flow
max. 80 kPa
The following table shows reference values for the required filling pressures. In
every case, the filling rate should be 25 l/h higher than the maximum consump-
tion.
The fuel filling pressures of the following table refer to an open fill valve in front of
the shutoff valve at ~25 °C and diesel fuel as medium.
Tab. 118
Product Guide
170 Technical Data
Compressed air
550 ... 800 kPa (filtered)
Air consumption: Only when changing an operating state to switch a pneu-
matic valve (e.g. when switching from PAUSE to STBY), otherwise no air is
consumed.
Electrical connections
Operating voltage 24 V DC ± 5 %
Current consump- Max. 10 A
tion
Interfaces Shutoff valve
Two RS232 interfaces with AK compliant
protocols
Synchronization interface for the connection of
AVL Fuel Temperature Control
Digital I/O (with jumper setting w/ or w/o
potential separation)
Analog interface 0 … 10 V with digital I/Os
Analog output Data rate: 20 Hz
Resolution: 16 bit
Linearization error: < 6 mV
Max. output current (short circuit-proof):
15 mA
Dimensions
Basic device: c. 770 × 667 × 34 mm (W × H × D)
Weight
62 kg (including wall bracket, without options, not filled)
5.1 CE-Compliance
Compliant to the 89/336/EEC Directive of electromagnetic compatibility by
fulfilling the requirements of the following standard:
– EN 61326:97/A1:98/A2:01 Electrical equipment for measurement, con-
trol and laboratory use EMC requirements
The conformity of the pressure equipment to Directive 97/23/EC is ensured,
because an evaluation of the product / component did not result in any
exceeding of the ranges pursuant to "Article 3 / paragraph (3)".
6 Appendix
The figures below illustrate the design of the internal fuel tank.
A B
H
1 Overflow tank
4 Tank block
A Fuel in
B Fuel to engine
E Drain overflow tank
H Calibration unit option
Fig. 77
Product Guide
172 Appendix
6
5
4 Tank block
5 Fill valve
6 Tank level sensor
A Fuel in
E Drain overflow tank
H Calibration unit option
Fig. 78
H
4
D
A
C
1 Overflow tank
2 Tank cover
3 Tank
4 Tank block
A Fuel in
B Fuel to engine
C Fuel return line
D Venting
E Drain overflow tank
H Calibration unit option
Fig. 79
Product Guide
174 Appendix
Index
A B
AADC … 115 Bypass … 74
AAKN … 132
AALC … 113 C
AATS … 115 Calibration – Adjustment … 75
AATW … 116 Calibration Unit … 19, 59
ABSZ … 114 Venting … 60
Accessories … 15
Calibration
Accuracy Check … 73 unit option … 59
ACFG … 113 CE-Compliance … 170
ADJP … 114 Circuit Diagrams … 161
Adjusting … 74 Coarse Filter … 20
ADMP … 108 Connecting Compressed Air … 40
ADSP … 111 Connector … 57
AFIP … 111 Consumption Counter … 81
AGSZ … 128 Consumption Totalizers … 81
AGTD … 121 Control Lines
AHTP … 116 Digital … 45
AHTS … 117 Controller Board
AHTW … 117
Assembly Diagram … 158
AIDA … 129
Cooler
AIDE … 129
in the fuel return line … 31
AIMW … 131
AK Protocol … 46 Counter … 81
AKAL … 119
AKAS … 120 D
AKEN … 101 Delivery rate … 33
AKGP … 120 Device configuration … 54
AKLE … 121 Device dimensions … 23
AKMW … 124 Device Integration … 44
AKUG … 124 Device Variation … 15
AMEM … 108 Dimensions … 23
AMEV … 108 Drain … 64
AMKP … 123 Drain Overflow … 34
AMKS … 123 Drainage … 34
AMOD … 134 Draining … 74
AMSP … 107
Analog Interface … 53 E
APMP … 130 EAKN … 132
APMW … 114 EALC … 113
Application Notes … 21 EATS … 115
ARDR … 133 EBSZ … 114
ARMW … 131 EDJP … 114
ASCK … 132 EDMP … 108
ASEK … 124 EDSP … 111
ASEM … 109 EFIP … 111
ASEP … 109 EGSZ … 128
ASEV … 109 EGTD … 121
ASEX … 110 EHTP … 116
ASFP … 120 EHTS … 117
ASGJ … 122 EIDA … 129
ASIN … 106 EKUG … 124
ASRV … 133 Emergency Stop (X7) … 41
Assembly … 23 Emergency Stop Relay … 140
ASTF … 101 EMKP … 123
ASTX … 104 EMOD … 134
ASTZ … 103 EMSP … 107
Auto-STBY … 72 EPMP … 130
AVL Fuel Temperature Control … 27 ERDR … 133
AVNP, EVNP … 104 ESCK … 132
AWGP, EWGP … 114 ESFP … 120
Product Guide
176
ESGJ … 122 P
Explosion … 26 Pause … 70
Pin Assignment … 54
F Pipe Material … 30
Filling … 75, 138 Potential Isolation … 41
Filling pressure … 33 Power Supply … 19
Filling pump … 33 Power supply … 40
Fine Filter … 19 Pressure Regulator … 140
Fire … 26 Pump states … 152
First-Time Start-up … 62
Flame Filter … 19 R
Fuel Direction … 31 Remote Control … 20
Fuel Drain … 34 Replace Components … 140
Fuel filling pump module … 20 Replacement of Components … 140
Fuel Filter … 31 Rinse … 64
Fuel Lines … 39
S
G
SATK … 119
Gas Sensor … 26 Scope of Supply … 15
SDAA … 128
H SDAE … 128
Hydraulic Connections … 39 SDRN … 113
Serial Interfaces … 48
I SFIL … 113
Init Pause … 70 Shutoff Valve … 19
Inputs Shutoff valve … 34, 47, 57
Digital … 51 Shutoff Valve option … 34, 57
Installation … 15 SKCH … 119
Installation Guidelines … 30 SMAN … 101
Spare Parts … 154
Installation Instructions … 15
Interfaces … 44 SPAU … 103
Serial … 48
SREE … 131
SREM … 101
SRES … 127
L SSRV … 127
Leak-tightness of the system STBY … 71, 103
Test pressure … 30 SVNT … 113
LED Functions … 159
Level … 52 T
Levelling … 29 Tank overflow … 34-35
Technical Data … 167, 171
M Terminal mode … 135
Maintenance limit … 82 Transport … 23
Maintenance Schedule … 137 Trolley … 20, 28
Measured Quantities
Overview … 68 V
Measurement Error Valve settings … 152
Frequent Causes … 32 Venting … 35
Mounting Bracket … 23, 28-29 Flow Rate … 31
System … 31
O venting … 72
Operating States Venting valve … 20, 61
Overview … 68
Operating-hour Counter … 82 W
Option Power Supply … 40 Wearing Parts … 154
Option Venting Valve … 61
Options … 15, 57
Optocoupler … 52
Z
Outputs Zero Consumption Measurement … 139
Digital … 51
Overflow … 35
Product Guide
AVL List GmbH
Hans-List-Platz 1, A-8020 Graz, Austria
Phone: +43 316 787-0, Fax: +43 316 787-400
http://www.avl.com