Pre-use integrity tests are performed to ensure that the effectiveness of sterilizing grade filters has not
been compromized during shipping, operator handling or sterilization prior to use. In applications where
high value fluid products are used, pre-use filter integrity testing can prevent the loss of costly product
batches by alerting operators to a failed filter prior to use.
A risk assessment should be performed to determine whether the filter must be tested prior to use. In
applications where the filter is used to sterilize product prior to filling, both EC and US GMP require
that the filter is integrity tested pre- and post-use(1), (2). In other applications where single use systems are
used, for example for preparing buffers for chromatography columns or media for fermenters, regulations
may not require a pre-use integrity test and the decision whether to perform a pre-use integrity test or
not should be based on economic considerations.
Where a pre-use filter integrity test is not required, the filter can be disconnected from the single use
system after processing is complete and simply integrity tested off-line.
(1)
The EC GMP guide for Aseptic Processing says that: “The Integrity of the sterilized filter should be verified before use and should be con-
firmed immediately after use by an appropriate method”
(2)
FDA Guidance – Sterile Drug Products Produced by Aseptic Processing
“Normally, integrity testing of the filter is performed prior to processing, after the filtration equipment has already been assembled and
2 sterilized.”
“It is important that integrity testing be conducted after filtration to detect any filter leaks or perforations that might have occurred during the
filtration.”
Figure 1 Palltronic® Flowstar unit testing a capsule off-line
REMOTE
VENT VALVE
• Procedure for wetting the filter and collection of the flush liquid
Most single use system applications are for liquid filtration. Vent filters are not required for systems with
bags, since the bags are empty and do not require venting. Air, nitrogen or vent filters can be required
for disposable bioreactor applications and could be integrity tested, following procedures outlined below.
• No need to wet the filter with water before testing and may still need to flush the filter to
reduce extractables
• No need to remove the integrity test fluid before processing the product.
In addition, when product is used as the wetting fluid for a post-use test, there is no need to flush the
product-wetted filter with water before testing.
3
When product wet integrity test values can be used, the system shown in Figure 2 could be employed
with the following protocol. For this example, the filter is attached to a disposable container (bag). The
process fluid is filtered and stored in the bag. The bag can be used for pre-use integrity test as well as for
storage of filtered fluid.
2. Flush filter with sufficient product to wet the filter for the integrity test.
3. Attach an automated integrity test instrument to the upstream side of the filter. Perform an integrity
test using the test limits recommended by the filter manufacturer for that filter, fluid and test gas
combination. During the stabilization phase, fluid remaining on the upstream side of the filter will
be forced through the filter and collected in the product bag.
Figure 2
PRODUCT
PRODUCT
CLAMP BAG
FILTER
4
Integrity Test Type
Filter integrity tests are performed using Forward Flow or Bubble Point tests. With either test, it is
important that steps be taken to evaluate the possible effects of filter failure because significant pressurization
of the product bag could result in a bag failure. A Forward Flow test is generally preferred because it
quantitatively measures the diffusive flow as well as flow through any open pores in a wetted membrane
filter and performs the test at a defined test pressure. For an intact filter, the gas flows are relatively small.
This enables the supplier to determine and design the test system for maximum system pressure and
downstream gas volumes.
If the filters are to be tested by Bubble Point test, the maximum system pressure will be unknown because
the test pressure will increase until the bubble point is detected. In addition, when the bubble point is
reached, very high gas flow passes through the filter which must be managed within the downstream sterile
environment.
Vent filters typically used in pharmaceutical processes are made from PVDF membranes (for example
Pall® Emflon® II filters) or PTFE membranes (for example Pall Emflon PFR filters). We recommend the
use of a vent filter with a gamma tolerant membrane, for example Pall Emflon II filters as PTFE is not
stable to gamma irradiation and therefore not suitable for use in a gamma irradiated pre-sterilized
disposable system.
Figure 3
PRODUCT
PRODUCT
CLAMP BAG
FILTER
VENT
FILTER
5
A Pre-Use Test Using Water or Solvent Wet Integrity Test Values
Where an integrity test is not performed using product as the wetting fluid, a flush bag on the downstream
side can be used to collect the fluid required to wet the filter for the integrity test.
The following outline of pre-use filter integrity test steps, which corresponds to Figure 4 below, provides
a simple solution to wetting the filter with fluid, which is not product, prior to processing.
1. Ensure that clamp 1 is fully closed. Open clamp 2 to the flush bag.
2. Flush filter with the wetting solution. The flush volume and wetting fluid may vary depending on
filter type and size. Collect used flush fluid in the flush bag.
3. Attach an automated integrity test instrument to the upstream side of the filter. Perform a Forward
Flow integrity test using the test limits recommended by the filter manufacturer. During stabilization
phase, fluid remaining on the upstream side of the filter will be forced through the filter and
collected in the flush bag.
4. Drain any residual wetting solution in “T” connector into the flush bag.
6. Open clamp 1 to process bag. System is ready to process product. (Note: If product dilution is a
concern, a small quantity of product can be directed to the flush bag before collecting product in
the product bag.)
Figure 4
CLAMP 1
“T”
PRODUCT
CONNECTOR
PROCESS
BAG
FILTER
FLUSH
BAG
CLAMP 2
6
Conclusion
The decision tree below provides a guide to choosing a simple method to integrity test a filter in a pre-
sterilized disposable system. As most systems are customized to meet individual process requirements,
there is no single method or solution for testing the filter. Further information and guidance can be given
by your local Pall office.
Figure 5
PRE-USE NO
INTEGRITY FIGURE 1
TEST
YES
NOT PRODUCT
FIGURE 4
7
2200 Northern Boulevard Visit us on the web at www.pall.com
East Hills, New York 11548-1289
Pall Corporation has offices and plants throughout the world in locations including:
800.717.7255 toll free Argentina, Australia, Austria, Belgium, Brazil, Canada, China, France, Germany, India,
516.484.5400 phone Indonesia, Ireland, Italy, Japan, Korea, Malaysia, Mexico, the Netherlands,
516.801.9548 fax New Zealand, Norway, Poland, Puerto Rico, Russia, Singapore, Spain, South Africa,
Sweden, Switzerland, Taiwan, Thailand, United Kingdom, the United States and Venezuela.
pharmafilter@pall.com e-mail Distributors are located in all major industrial areas of the world.
Europa House, Havant Street , Pall, Palltronic and Emflon are trademarks of Pall Corporation.
Portsmouth PO1 3PD, United Kingdom Filtration. Separation. Solution. is a service mark of Pall Corporation.
® indicates a trademark registered in the USA.
+44 (0)23 9230 3303 phone © 2004, Pall Europe Ltd.
+44 (0)23 9230 2506 fax
BioPharmUK@europe.pall.com e-mail
10.2004