Reanimator F 120 Infant Transport System

GLOBETROTTER®
trained personal only

Pursuant to Medical Device Directive 93/42, the apparatus must only be operated by
trained personnel. Prerequisite is an exact knowledge of the operating instructions!

The F. STEPHAN GmbH will not be liable for damages caused by faulty operation due to
non-observance of the operating instructions or from use for which the apparatus is not
earmarked.
IMV 2 min
default mode at start suppression

Display of frequency
three „---“ in CPAP mode
blinks if out of possible LED for
range max. 120/min (IMV Stenosis Visual alarm if the
Mode) permanently installed pressure
Inspiration
of 12 mbar is not fallen below 2,2
0,25 - 2,00s
seconds after the set end of inspi-
Expiration
ration time (stenosis)
0,25 - 30,00s
O2 MONITOR
REANIMATOR F120 Disconnection Visual alarm
manual ventilation with
%
Hill-Rom Air-Shields
A HILLENBRAND INDUSTRY
R

ensues if the permanently installed


prefixed patterns PEEP CAL
I AUTO
pressure limit of 12 mbar is not
0
execeeded after the set expiration
IMV CPAP HEATER TEST

PLATEAU
21%
ON / OFF

manual inflation
O2 MAX TIMER MONITOR
time +max inspiration time (2.0s)
9 8 FREQUENCY
MAX 120
/min
2 min 20 30
+
10 STEPHAN 40 O2 Alarm Visual alarm in case of
®
®

O2 MIN INSPIRATION (sec) EXPIRATION DISCONNECT.


STENOSIS

2 3
O2 - ALARM 0 50
violation of the set limit values for
0 5 0 0 0 5 0 BAT.
-10
cmH2O
60
+ +
12V ext.
MAINS Überlastgrenze: 600 cmH2O
O2 concentration (monitor must be
switched ON)
Beatmungsdruck

O2 O2
15 O2 AIR

Battery Visual and acoustic alarm


4 6 100 200
fill bottle with Aqua Dest. 2 8

(sterile and demineralised 0


kPa x 100
10 0
kPa x 100
300
L/min
1.013 bar if internal battery voltage falls
10 10
bar bar

20°C
below 10,8V. Heater is off.
315

water) not over max. level!


STEPHAN STEPHAN
® ®

Every 30s warning tone occurs not


OFF

Reduce level, if flow is Flow


L/min 2 4 6 8 10
suppressable. 20 minutes remai-
O2
O2 AIR O2 AIR O2 AIR O2 AIR O2 AIR 5 5
higher than 10l/min!
Vol %
21 0 2 0 4 0 6 0 8 0 10

Water is sufficient for 4-6 h


30
40
0,25 1,75 0,5
0,5 1,5 1
3,5 0,75 5,25
3 1,5 4,5
1
2
7
6
1,25 8,75
2,5 7,5
ning then respirator is shut off
50 0,75 1,25 1,5 2,5 2,25 3,75 3 5 3,75 6,25 1 1
60 1 1 2 2 3 3 4 4 5 5
70 1,25 0,75 2,5 1,5 3,75 2,25 5 3 6,25 3,75

Oxygen flowmeter for aux 80


90
1,5 0,5
1,75 0,25 3,5
3 1
0,5 5,25 0,75
4,5 1,5 6
7
2
1
7,5
8,75 1,25
2,5
Med.
AIR

outlet for incubator 100 2 0 4 0 6 0 8 0 10 0

O2 pressure
NeoSID NOVA PLUS
cylinder/
content

O2 pressure
internal
APNEA - CARDIAC - SpO2 - MONITOR
SET
R

0
aux. O2 connector for POSITION OF MENUE

incubator
2 min
ELECTRODES

CABLE PATIENT MAINS BAT

O2 sensor in normal posi-


tion removable for use in
incubator. Close the socket
with replacement plug if cell
is outside (prevents high cable alarm 2 min alarm
leakage flow) blinks at failure of patient suppression menu key to Battery only
cable or fixation of elec- for NOVA move up or NOVA
trodes down
Turn OFF:
patient alarm mains alarm
Press key, blinks at cardiac or only NOVA
respiratory alarm
press additional key,

unhand red key

brief instruction F120 GT ver 1.0 1


23.08.02 JS
230 V by means of the apparatus plug and the attached
connecting cable. Main switch should always be ON so
additional socket for
internal battery can be charged 230V supply

O2 M
A
I
N
AIR
S
SENSOR POT.
RS 232 for
2x 2x
1 AT 10 AT O2 BONDING NOVA NOVA
POP OFF
Reception VALVE
coupling for 12V EXT.
h
00000000

O2 and air Potential


OPERATION
INLET / FILTER HOURS bonding
ALARM 1
12V
ex-
ternal
ALARM 2
supply

Pop off valve Inlet filter has to Alarm NOVA Alarm F120
mechanical safety be changed if
valve 45 mbar at 5l dirty
Flow

Placing the electrodes


1. red electrode below the right nipple.
2. yellow electrode below the left nipple.
3. black electrode beside the bellybutton.
4. Place the electrodes into patient cable socket according
to the colour code on it.
5. The far end of the patient cable (grey plug)

heart rate Respiratory


no value “---” frequency Time and date

alarm limits Apnea alarm limit Symbols for


affected limit is displayed lung symbol blinks in battery (low charge alarm)
inverted case of resp. activity mains
heart symbol blinks in clock and bell (suppression
case of cardiac activity for 2 min)

brief instruction F120 GT ver 1.0 2