You are on page 1of 5

Appendix: System Controls and Their Uses

Paul L. Allan

In addition to the basic choice of transducer type and imaging is normally chosen. The Doppler fre-
frequency for the examination in hand, there are many quency used by any transducer is often 1–2 MHz
other factors on a Doppler ultrasound system which below the imaging frequency, although modern
need to be adjusted. Despite the efforts of the manufac- equipment has a wide range of receive frequencies
turers to automate and simplify things, it is still neces- such as 5–14 MHz. In addition to the basic imaging
sary to adjust continually many of the scan and Doppler requirements, it should be remembered that the
parameters during the course of an examination. Is the deeper a vessel lies, the longer it takes a sound pulse
vessel superficial or deep? Is flow fast or slow, high vol- to travel there and back, so that the Nyquist limit
ume or low volume? Most systems now come with a becomes very relevant (Chapter 1), limits the
variety of preset programs for different situations: Doppler frequencies that can be used and therefore
peripheral arteries, veins, cerebrovascular, etc.; together the frequency shifts/velocities that can be recorded
with automatic image and spectral display optimisation. accurately.
Many also have the facility to allow users to save their
own program preferences, which is a convenient option
to store preferred settings once a satisfactory set up has B-MODE IMAGE
been achieved for a particular type of examination. For colour Doppler examinations this should be set
Whilst the manufacturer’s presets allow the basic appro- up with relatively low overall gain, so that the image
priate settings to be employed at the start of an exami- is a little on the dark side as the software tends to
nation, fine adjustments will still be required during the allocate colour to darker areas, rather than to areas
course of the examination to make the most of the avail- which contain echoes. See ‘Colour write priority’
able information. Familiarity with the ultrasound below.
system, together with experience, enable skilled opera-
tors to set up their system appropriately for the exam-
ination being performed. Different manufacturers
sometimes use different names for the same controls TRANSMIT POWER
or functions; it is not possible to give an exhaustive list The transmit power of the system should be set at
of all the possible options on the range of modern the lowest level consistent with an adequate exam-
equipment now available but the following notes ination, especially during obstetric and gynaecolo-
describe the basic controls, or parameters on most sys- gical examinations. It is better to start at a medium
tems that can be adjusted during the course of an exam- level and increase the power only after other mea-
ination to improve and maximise the information that is sures to improve system sensitivity have been tried,
obtained from the examination. such as adjusting colour gate size, removing filters,
adjusting the scale/pulse repetition rate. For low
General Principles mechanical index (MI) contrast studies (see
Chapter 17) the transmit power should be set as
TRANSDUCER FREQUENCY low as possible and the MI reading ideally should
The highest frequency which will achieve the highest be less than 0.4. Modern systems often have con-
resolution consistent with adequate penetration for trast-agent-specific programs already installed, or
370

Downloaded for Anil Kumar Mishra (mis1579@yahoo.co.in) at King George's Medical University from ClinicalKey.com by Elsevier on July 04, 2017.
For personal use only. No other uses without permission. Copyright ©2017. Elsevier Inc. All rights reserved.
Appendix: System Controls and Their Uses 371

the relevant settings can be obtained from the supplier larger areas take more processing power and time.
of the contrast agent. Larger sizes and greater angulations may therefore
reduce frame rates. Greater angulation also reduces
UPDATE/DUPLEX/TRIPLEX sensitivity, which can therefore be improved by reduc-
ing the degree of angulation of the colour box, for
In duplex ultrasound there is the ability to acquire and
instance when examining deeper segments of the fem-
display both real time imaging and spectral Doppler
oral vein in larger legs. Deeper boxes need slower
information either simultaneously, or alternately.
pulse repetition frequencies, therefore limiting the size
Simultaneous display results in degradation of both
of shift that can be measured without aliasing.
the image and the spectral display as the computer
has to process data from both sources. The update
COLOUR MAP
facility allows the operator to set the system to handle
either imaging data, or Doppler data. This results in a Choose a colour map that has good contrast for the
higher quality of display for the selected mode and it is identification of aliasing. Maps which grade to white
usually used for acquisition of the best-quality spectral at each end, or to very pale colours, do not give as much
display for analysis. Simultaneous duplex scanning is information about aliasing as those which have signifi-
of some value in the initial stages of an examination in cantly different colours, such as pale green and pale
order to position the sample volume in a specific area orange. Variance maps were said to register the amount
of interest. Triplex mode refers to the simultaneous of spectral broadening in the colour map, although this
acquisition, processing and display of colour Doppler, is not easy to appreciate and may, in fact, not be true as
spectral display and imaging information. As with this ‘variance’ seems to reflect the velocity, rather than
duplex scanning, this requires significant division of spectral broadening. Variance maps are most frequently
processing power and consequent compromise in used in cardiological examinations. In addition, colour
the quality of the display. Newer systems with more tags can be put into the scale so that velocities above or
powerful computers are less prone to these problems below the chosen range can be identified.
but there is still a division of processing power and
experienced operators will still tend to switch between COLOUR GAIN
imaging and spectral Doppler for optimal results. The colour gain is set to the optimal level. This can be
identified by turning up the gain until noise/colour
AUTOMATIC IMAGE OPTIMISATION speckle is seen in the colour box and then backing
off slightly on the overall colour gain. It should also
Many systems now have automatic image optimisation
be adjusted to remove any spill of the colour map over
controls which adjust factors such as time gain com-
the boundaries of the vessel seen on the B-mode
pensation, receiver gain and colour Doppler gain to
images; this is sometimes called colour bleeding. Col-
improve the image. In some cases this facility can also
our gain often needs to be increased if the colour box is
be applied to the spectral display. These can be useful
moved to a deeper location.
in getting a reasonable initial display but should not be
relied upon entirely and the operator needs to make
COLOUR SCALE
the final adjustments to obtain the best images.
The colour scale should be set to levels appropriate
for the range of velocities under investigation. It
Colour Doppler Controls should be remembered that colour Doppler only gives
information on the mean Doppler shift in a pixel;
COLOUR BOX the mean velocity in a pixel is only calculated when
This defines the area of the image over which colour an angle correction for that pixel is performed. The
Doppler data are acquired. It can be adjusted for size, colour scale is closely related to the pulse repetition
position and direction/angulation. It is better to keep frequency, or sampling rate; this has to be above the
the box as small and as superficial as practical because Nyquist limit for the frequency shift being measured.

Downloaded for Anil Kumar Mishra (mis1579@yahoo.co.in) at King George's Medical University from ClinicalKey.com by Elsevier on July 04, 2017.
For personal use only. No other uses without permission. Copyright ©2017. Elsevier Inc. All rights reserved.
372 Appendix: System Controls and Their Uses

If it is set too low then aliasing will occur; this can be COLOUR PERSISTENCE
rectified by increasing the colour scale range, also This reflects the amount of frame-averaging which
known as the pulse repetition frequency. If it is set occurs. A low level of persistence results in an image
too high then there will be poor colour sensitivity which has a fast temporal response but may have a
for slower velocities, resulting in inadequate colour poor signal-to-noise ratio. A high level of persistence
fill-in across the vessel. improves the signal-to-noise ratio by summating data
from several frames, but this results in an impaired
COLOUR INVERSION temporal response, so that pulsatile flow information
This changes the colour map display for mapping the is dampened.
direction of blood flow in relation to the transducer,
e.g. red for flow towards the transducer and blue for
away from the transducer, or vice versa. Power Doppler Controls
Although the power mode is intrinsically simpler than
COLOUR GATE velocity mode, there are still several controls and
This relates to the size of the colour pixels, a smaller options which influence the power image.
size gives better spatial resolution, whereas a larger
size provides better sensitivity at the expense of spatial POWER MAPS
and sometimes temporal resolution. Most systems offer a variety of colour maps from which
the operator can select according to personal prefer-
COLOUR BASELINE ence, usually a yellow- or magenta-based scale is used
The colour baseline can be adjusted to provide a wider for display of power Doppler information. The thresh-
range of shifts on one side or the other, depending on old or sensitivity can be varied; this results in changes
the characteristics of the blood flow in the vessel being in the transparency of the power box. When the power
examined and particularly if the flow is in only one map background is opaque the sensitivity is maximal.
direction. Various degrees of increasing translucency are avail-
able so that sensitivity can be traded for imaging and
COLOUR FILTERS anatomical information.
These help remove low-frequency noise and clutter
from the image. It is best to set filters at the lowest level POWER GAIN
compatible with an acceptable image as they also This is similar to colour gain and amplifies the received
remove low-frequency shift information from the Doppler signal. Excessive gain will result in unneces-
image. sary noise and artefact in the image.

COLOUR WRITE PRIORITY SCALE


A pixel in the image can display either B-mode imaging The energy scale affects the sensitivity of the system for
information, or colour Doppler information. The col- signal intensities of varying strengths: lower scales are
our write priority facility allows the relative priorities more sensitive to lower-intensity signals, which will
for these two types of information to be defined. High also increase flash artefact. On some systems the
colour priority results in colour information being power Doppler scale is linked to the filter settings.
displayed in areas which might contain low-intensity
imaging information, for instance at the margins of DYNAMIC RANGE
vessels. Alternatively, high imaging priority results As in spectral Doppler, this affects the range of signal
in grey scale information displacing colour informa- intensities over which the available colour scale is
tion, such as might occur with reverberation artefacts spread and the appearance of the Doppler information
appearing within the vessel lumen amongst the colour on the screen: increasing the dynamic range will tend
Doppler information. to increase the amount of colour on the screen.

Downloaded for Anil Kumar Mishra (mis1579@yahoo.co.in) at King George's Medical University from ClinicalKey.com by Elsevier on July 04, 2017.
For personal use only. No other uses without permission. Copyright ©2017. Elsevier Inc. All rights reserved.
Appendix: System Controls and Their Uses 373

POWER BOX STEERING ANGLE spectral receiver gain, can result in artefactual broad-
As with the colour box, the power box can be adjusted ening of the spectra displayed. Further manipulation
for size, position and direction/angulation. Although of the spectral display can be performed by altering
the power display is much less dependent on angle the postprocessing algorithms on some systems in
than the velocity display, there may still be loss of order to emphasise, or suppress, particular frequency
the power signal at 90 because these low Doppler fre- shifts. In normal practice a simple linear allocation is
quencies fall below the motion discrimination filter most convenient.
cut-off level. In most situations, steering the power SPECTRAL SCALE
Doppler box does not have much effect on the display;
however, the loss of signal at 90 may be overcome by Altering the scale affects the pulse repetition frequency
having a degree of angulation of the transducer. and thus the range of shifts which can be registered
without aliasing. In practice the scale is adjusted so
POWER/COLOUR HYBRID IMAGES that the Doppler waveform is displayed without
wrap-around, which indicates aliasing. The scale
Many systems now allow a hybrid image containing
may be displayed as either a KHz scale for frequency
both power Doppler and colour Doppler data to be
shift, or a velocity scale. The use of velocity, rather than
displayed. This aims to combine the sensitivity of
frequency shift, allows some comparison of different
power Doppler with the directional and velocity infor-
examinations, performed with different transducer
mation from colour Doppler. On some systems, the
frequencies and with different angles of insonation.
relative contributions from the two sources can be
adjusted. SPECTRAL INVERSION
FILTERS This allows the operator to change the orientation of
the display. Many operators prefer to display both arte-
Motion discrimination filters in power Doppler are
rial and venous waveforms above the baseline, even if
used to filter out excessive signal noise from structures,
flow is away from the transducer. However, care is
other than blood, which are moving in the Doppler
therefore required when assessing the vertebral arter-
box. Low filter settings provide more sensitivity but
ies for reverse flow, or the leg veins for reflux, as errors
are more prone to flash artefacts, whereas higher filter
may occur if spectral inversion is not recognised. Some
settings reduce flash artefacts but will also filter out
centres do not allow spectral inversion because of the
some blood flow information.
potential for misinterpretation, particularly in relation
to examinations for venous insufficiency.
Spectral Controls
SPECTRAL SWEEP SPEED
SPECTRAL GAIN A medium speed is adequate for most arterial work,
This affects the receiver gain for the spectral display. As with a slower speed better for venous flow. A fast
in B-mode imaging, the level should be adjusted in sweep speed is useful for acceleration time measure-
order to give a balanced distribution of grey shades ment and waveform analysis, particularly if there is
across the displayed spectra. Excessive or inadequate tachycardia.
levels of gain will produce erroneous estimates of
Doppler shift/velocity. ANGLE CORRECTION
The measurement of the angle of insonation relative to
SPECTRAL DYNAMIC RANGE the direction of flow is required in spectral Doppler in
This can be varied to optimise the display of particular order to convert frequency shift information into
frequency shifts. A narrow dynamic range results in velocity information and also in colour Doppler to
the loss of low-intensity shifts above and below the convert mean pixel shifts into mean velocities. The
main shift frequencies. Conversely a wide dynamic main direction of flow may not necessarily be parallel
range, particularly if associated with a high level of to the vessel wall and colour Doppler is useful in

Downloaded for Anil Kumar Mishra (mis1579@yahoo.co.in) at King George's Medical University from ClinicalKey.com by Elsevier on July 04, 2017.
For personal use only. No other uses without permission. Copyright ©2017. Elsevier Inc. All rights reserved.
374 Appendix: System Controls and Their Uses

precise positioning of the angle-correction cursor FILTERS


along the line of the jet. The angle of insonation should Removal of low-frequency noise and clutter arising
be less than 60 or the errors in velocity calculation from the vessel wall and surrounding tissues contrib-
become significant. utes to a cleaner signal, but filters should be set as low
GATE SIZE as is practical, otherwise low-frequency shifts from
slow blood flow will be filtered out, which could result
This defines the range of depths from which Doppler in the mistaken impression of absent diastolic flow in
data are collected. The gate should be positioned arteries, or occlusion in veins. In practice it is best to
across the lumen of the vessel but clear of the walls keep the filters set at the lowest setting and only
in order to reduce wall thump artefact. The position increase filtration as required during an examination.
of the gate which corresponds to the maximum Dopp-
ler shift is located using a combination of the colour
map for initial positioning and then refining this using
the operator’s ears, which are the most sensitive and
Conclusions
efficient spectral analyser available; with experience, Modern systems allow the operator to preset many of
your ears will register when the peak frequency shift these parameters and create different personal pro-
is obtained. If an assessment of volume flow is being files/setups for different types of examination such as
made, the gate should be wide enough to encompass veins, carotids, renal, etc. However, it should be
the entire vessel width so that all the flow contributes remembered that ultrasound is a dynamic examination
to the signal and the time-averaged mean velocity will and, even with modern equipment, the best results will
be most representative. Smaller gates give a cleaner sig- be obtained if the system controls are adjusted to opti-
nal, especially with laminar flow, but at the expense of mum settings for the task in hand, rather than relying
a reduction in sensitivity. on the preset profiles alone.

Downloaded for Anil Kumar Mishra (mis1579@yahoo.co.in) at King George's Medical University from ClinicalKey.com by Elsevier on July 04, 2017.
For personal use only. No other uses without permission. Copyright ©2017. Elsevier Inc. All rights reserved.

You might also like