ALS-R User's Manual: Amplified Spontaneous Emission Light Source
ALS-R User's Manual: Amplified Spontaneous Emission Light Source
Caution: The user must read this manual before operating the ALS-R unit.
Operations other than those described in this manual may result in
personal injury and damage to the unit.
Note that any attempt to open or fix the equipment without prior
approval by Amonics Ltd. voids the warranty.
Ver. 3.0
August 8, 2007
Revision History
This document is copyrighted property of Amonics Ltd. It may not be used in whole
or in part for manufacture, sale, or design of items without the written permission of
Amonics Ltd.
Table of Contents
1. General Information___________________________________________________1
1.1 Introduction______________________________________________________________1
1.2 Product Overview_________________________________________________________1
1.3 Features__________________________________________________________________1
1.4 User Safety_______________________________________________________________1
2. Operation____________________________________________________________2
2.1 Introduction______________________________________________________________2
2.2 Initial Inspection__________________________________________________________2
2.3 Controls__________________________________________________________________3
2.4 Main Control Screen_______________________________________________________5
2.5 Operation Instructions_____________________________________________________7
2.6 Warning Messages_________________________________________________________8
2.7 PC Connection Mode_______________________________________________________8
3. Troubleshooting_______________________________________________________9
4. Service and Support___________________________________________________10
4.1 Warranty________________________________________________________________10
4.2 Service and Calibration____________________________________________________10
4.3 Care of Fiber-optic Connectors_______________________________________________11
Amonics Limited
Unit 101, 1/F, Winning Centre, 29 Tai Yau Street, San Po Kong, Kowloon, Hong Kong
Tel: (852) 2428 9723 Fax: (852) 2428 9704 Email: contact@amonics.com
ALS-R User’s Manual Page 1 of 11
-
5/11/2021
1. General Information
1.1 Introduction
1.3 Features
1. The ALS-R unit emits high intensity invisible light from the optical
output receptacle. Avoid direct exposure to skin and eyes.
2. The equipment case is fully certified for EMS protection. The user
should never open the equipment case; any attempt will void the
warranty and may result in electric shock and EMS attack to
equipment in the vicinity.
Amonics Limited
Unit 101, 1/F, Winning Centre, 29 Tai Yau Street, San Po Kong, Kowloon, Hong Kong
Tel: (852) 2428 9723 Fax: (852) 2428 9704 Email: contact@amonics.com
ALS-R User’s Manual Page 2 of 11
-
5/11/2021
Amonics Limited
Unit 101, 1/F, Winning Centre, 29 Tai Yau Street, San Po Kong, Kowloon, Hong Kong
Tel: (852) 2428 9723 Fax: (852) 2428 9704 Email: contact@amonics.com
ALS-R User’s Manual Page 3 of 11
-
5/11/2021
2. Operation
2.1 Introduction
When you receive your ALS-R, verify that the following items are
included:
Amonics Limited
Unit 101, 1/F, Winning Centre, 29 Tai Yau Street, San Po Kong, Kowloon, Hong Kong
Tel: (852) 2428 9723 Fax: (852) 2428 9704 Email: contact@amonics.com
ALS-R User’s Manual Page 4 of 11
-
5/11/2021
2.3 Controls
5 4 3 2
1
Feature Function
Main Display The LCD display shows the control parameters, warning
messages and optical powers of the ALS unit.
Power Indicator This indicator lights up when the power to the ALS unit has
been turned on.
Knob Button The knob button allows the user to select and modify the
control parameters of the ALS unit. Turning the knob
clockwise moves the selection cursor down or increases
the selected parameter. Turning the knob counter-
clockwise moves the selection cursor up or decreases the
selected parameter. Pressing the knob selects and
deselects the highlighted control parameter for
modification. The ALS beeps once to acknowledge when
the knob is pressed.
Main Switch The key switch is front panel power switch for the unit. The
Main Switch on the rear panel has to be in the On position
for this switch to be operational.
Amonics Limited
Unit 101, 1/F, Winning Centre, 29 Tai Yau Street, San Po Kong, Kowloon, Hong Kong
Tel: (852) 2428 9723 Fax: (852) 2428 9704 Email: contact@amonics.com
ALS-R User’s Manual Page 5 of 11
-
5/11/2021
1 5 2 3 4 6
Feature Function
AC Power Socket The AC power socket is the input for the AC power source.
A three-pin standard power cord should be used to connect
this equipment to any 110 or 220 V main supply.
Fuse Socket The fuse socket accepts a 5 x 20mm mini fuse. A Quick-
acting fuse should be used for maximum protection.
RS232 Female For remote control and monitoring, the user can plug-in the
Socket RS232 direct cable provided, connecting the ALS unit to a
personal computer. For the control software, please refer to
the control software user guide.
Main Switch The main switch is the master AC power switch for the unit
including the optical output.
Amonics Limited
Unit 101, 1/F, Winning Centre, 29 Tai Yau Street, San Po Kong, Kowloon, Hong Kong
Tel: (852) 2428 9723 Fax: (852) 2428 9704 Email: contact@amonics.com
ALS-R User’s Manual Page 6 of 11
-
5/11/2021
The Main Control screen allows the user to control the Enable/
Disable status and driving current of the pump lasers and monitor
the output optical power of the unit. To switch between control of
the pump lasers, move the selection cursor to highlight the Laser
No. by turning the control knob button clockwise or counter-
clockwise and press the control knob button. (Note: Some models
have only one pump laser)
Parameter Description
Amonics Limited
Unit 101, 1/F, Winning Centre, 29 Tai Yau Street, San Po Kong, Kowloon, Hong Kong
Tel: (852) 2428 9723 Fax: (852) 2428 9704 Email: contact@amonics.com
ALS-R User’s Manual Page 7 of 11
-
5/11/2021
Laser No. (Note: Some models have only one pump laser)
1. Highlight the number after Laser No. parameter with the selection
cursor by turning the knob in the clockwise or counter-clockwise
direction.
2. Press select button once to select the next pump laser. The unit will
beep once in acknowledgement. The displayed parameter values will
now correspond to the selected pump laser.
Drv
1. Highlight the set of numbers after the Drv parameter with the
selection cursor by turning the knob in the clockwise or counter-
clockwise direction.
2. Press knob button once to select this parameter for editing. The unit
will beep once in acknowledgement
3. Turn the knob clockwise to increase the driving current to the
selected pump laser, increasing the optical output when the Status is
Enable.
4. Turn the knob counter-clockwise to decrease the driving current to
the selected pump laser, decreasing the optical output when the
Status is Enable.
5. Press knob button to exit from editing this parameter. The unit will
beep once in acknowledgement.
Status
Amonics Limited
Unit 101, 1/F, Winning Centre, 29 Tai Yau Street, San Po Kong, Kowloon, Hong Kong
Tel: (852) 2428 9723 Fax: (852) 2428 9704 Email: contact@amonics.com
ALS-R User’s Manual Page 8 of 11
-
5/11/2021
Start-up Procedure
1. Flip the main switch on the back to the On position to enable electrical
power to the unit.
2. Turn the main key switch clockwise to the On position to enable
electrical power to the unit. The power indicator will light up and the
LCD display will show a logo.
3. Wait for the LCD display to switch to the Main Control screen.
4. Select pump laser 0, change the Status to Enable, and then increase
the driving current Drv, for optical power.
5. Select pump laser 1, change the Status to Enable, and then increase
the driving current Drv, for more optical power. (Note: Some models
have only one pump laser)
Over-temperature Procedure
1. When the Temp Overheat warning appears the pump lasers and
optical output will shutdown automatically and the system will freeze
up.
2. Flip the main switch to the Off position.
3. Restart the unit using the Start-up Procedure described above.
Amonics Limited
Unit 101, 1/F, Winning Centre, 29 Tai Yau Street, San Po Kong, Kowloon, Hong Kong
Tel: (852) 2428 9723 Fax: (852) 2428 9704 Email: contact@amonics.com
ALS-R User’s Manual Page 9 of 11
-
5/11/2021
There are two warning messages that may appear on the LCD
screen: Temp. WARNING and Temp. Overheat.
Temp. WARNING
When the pump laser temperature approaches the upper limit, the
Temp. WARNING message appears on the LCD screen to warn
the user. The system will continue to operate normally, as long as
the temperature does not go over the limit.
Temp. Overheat
The unit begins the PC connection mode when the RS232 port is
connected to a PC and the control software is initialized. In this
mode there is no response to the controls on the front panel. The
user can only control the unit through the control software on the
PC. Once the software program stops execution, the unit will
return to local control. If the RS232 direct cable is disconnected,
the unit will automatically restore to local control.
Amonics Limited
Unit 101, 1/F, Winning Centre, 29 Tai Yau Street, San Po Kong, Kowloon, Hong Kong
Tel: (852) 2428 9723 Fax: (852) 2428 9704 Email: contact@amonics.com
ALS-R User’s Manual Page 10 of 11-
10 -
5/11/2021
Amonics Limited
Unit 101, 1/F, Winning Centre, 29 Tai Yau Street, San Po Kong, Kowloon, Hong Kong
Tel: (852) 2428 9723 Fax: (852) 2428 9704 Email: contact@amonics.com
ALS-R User’s Manual Page 11 of 11-
11 -
5/11/2021
3. Troubleshooting
Amonics Limited
Unit 101, 1/F, Winning Centre, 29 Tai Yau Street, San Po Kong, Kowloon, Hong Kong
Tel: (852) 2428 9723 Fax: (852) 2428 9704 Email: contact@amonics.com
ALS-R User’s Manual Page 12 of 11-
12 -
5/11/2021
4.1 Warranty
Amonics Ltd.
Unit 101, 1/F, Winning Centre,
29 Tai Yau Street,
San Po Kong, Kowloon
Hong Kong
Amonics Limited
Unit 101, 1/F, Winning Centre, 29 Tai Yau Street, San Po Kong, Kowloon, Hong Kong
Tel: (852) 2428 9723 Fax: (852) 2428 9704 Email: contact@amonics.com
ALS-R User’s Manual Page 13 of 11-
13 -
5/11/2021
1. Clean the connector with optical tissue soaked with alcohol that
is at least 90% pure.
2. Clean the connector again with dry optical tissue.
Reasons:
Amonics Limited
Unit 101, 1/F, Winning Centre, 29 Tai Yau Street, San Po Kong, Kowloon, Hong Kong
Tel: (852) 2428 9723 Fax: (852) 2428 9704 Email: contact@amonics.com