Professional Documents
Culture Documents
User Manual
PN 551026-06 Rev. A 2013.10
ii
DRAFT/PRE-RELEASE/CONFIDENTIAL
Copyrights
© 2013 Intuitive Surgical, Inc. All rights reserved.
MSim™ provided by Mimic Technologies, Inc.
© 2009-2012 Mimic Technologies, Inc. All rights reserved.
PhysX® Technology provided under license from NVIDIA Corporation.
© 2002-2012 NVIDIA Corporation. All rights reserved.
2 TERM. This license will begin on the date the Software is delivered to buyer, and will continue until the end of the useful
life of the System. Notwithstanding the foregoing, this license shall terminate immediately upon written notice to buyer
by Intuitive if buyer materially breach any term or condition of this license. Buyer agrees upon termination to promptly
discontinue all use of and destroy the Software and all copies thereof (whether in tangible form or as installed on buyer
equipment).
3 EXPORT LAW. The Software and related technology are subject to U.S. export control laws and may be subject to export
or import regulations in other countries. Buyer agrees to strictly comply with all such laws and regulations and
acknowledge that buyer have the responsibility to obtain such licenses to export, re-export or import as may be required.
4 U.S. GOVERNMENT BUYERS. The Software is a “commercial item” as that term is defined at 48 C.F.R. 2.101, consisting of
“commercial computer software” and “commercial computer software documentation” as such terms are used in 48 C.F.R.
12.212. Consistent with 48 C.F.R. 12.212 and 48 C.F.R. 227.7202-1 through 227.7202-4, all U.S. Government end users
acquire the Software with only those rights set forth therein.
_________________________________End of section______________________________
Contents
1 Introduction . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-1
•1.1 User Manual Organization . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-1
•1.2 Skills Simulator . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-2
• Surgeon Console Required . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-2
• Skills Simulation Software by Mimic, MSim and MScore. . . . . . . . . . . . . . . . . . . . . . . . . . 1-3
• Suture Skills Simulation Software by Simbionix. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-3
DRAFT/PRE-RELEASE/CONFIDENTIAL
1 Introduction
Note: The Skills Simulator is compatible with both the da Vinci Si and Si-e Surgical
Systems. The da Vinci Si-e is an optional configuration for the da Vinci Si System.
Note: This manual addresses the Skills Simulator specifically. For details regarding
the da Vinci Si (or Si-e) Surgical System, its components, instruments or accessories,
refer to the da Vinci Si Surgical System User Manual, PN 550650.
Note: For information specific to Skills Simulator Manager (Sim Manager) software,
see the Skills Simulator Manager User Manual, PN 551116.
Note: Before using the Skills Simulator with Sim Manager, verify that the
Sim Manager software has received the available upgrades to ensure compatibility.
This user manual is organized for quick reference and ease of use. A brief introduction,
overview and connection guide, power on, user interface instructions, and a system software
overview followed by troubleshooting guide, will get you up and running.
The manual consists of the following sections:
• Chapter 1 Introduction (the section you are reading now): Contains regulatory and safety
information that should be read by every user of the Skills Simulator. Includes general
precautions, power information and related information.
• Chapter 2 Connections, Startup, and Use: Explains how to connect power, system cables,
10/29/13
and optional devices to the Skills Simulator, as well as how to start up and begin using the
system, and how to navigate the user interface.
• Chapter 3 Onboard Simulator Administrator and Reports Guide: Provides information on
the Administrator function and the Simulator reports feature.
• Chapter 4 Simulation Software by Mimic Technologies: Contains software information
including user interface, exercise and scoring information.
• Chapter 5 Suturing Skills Simulation Software by Simbionix: Contains software
information for suture skills exercises including user interface, exercise and scoring
information.
• Chapter 6 Sleep Mode, Shut Down, Disconnection, Maintenance: Explains how to shut
down, store, clean and maintain the Skills Simulator.
• Appendix A: da Vinci® OnSite® for Skills Simulator: Contains information for Simulator
systems that have automatic updates enabled using da Vinci OnSite®.
• Appendix B: Troubleshooting: Provides remedies to setup issues.
• Appendix C: Specifications, Symbols, Glossary: Contains the reference material that is
indicated by its title.
• An Index at the end helps you find the information you want.
Intuitive Surgical and Simbionix collaboratively designed the suture skills exercise simulation
platform enabling realistic suturing scenarios.
The Skills Simulator mounts on the back of a da Vinci Si Surgeon Console, as shown in Figure
1.1. See section 2.1 Mount the Skills Simulator for mounting instructions.
10/29/13
CAUTION: The Skills Simulator is only designed to work with the Surgeon Console,
and it will not work with any other components of the da Vinci Si Surgical System.
Connecting the Skills Simulator to the Vision Cart will trigger the safety system, and
may impact system behavior during surgery, cause delay due to conversion, and
cause excess trauma to the patient from conversion.
the Skills Simulator. For Suturing Skills simulation software features and performance
evaluation details, refer to Chapter 5 Suturing Skills Simulation Software by Simbionix.
Contact Information
For Customer Service and Reporting of Complaints or Adverse Events
Use the following information for ordering, or for reporting complaints or adverse events; and
for general information regarding Intuitive Surgical or our products and services.
Intuitive Surgical, Inc. Intuitive Surgical Sàrl
1266 Kifer Road 1, chemin des Mûriers,
Sunnyvale, CA 94086 USA 1170 Aubonne
Toll free: 1.800.876.1310 Switzerland
Direct: 408.523.2100 Toll free: +800.0821.2020
Fax: 408.523.2377 Direct: +41.21.821.2020
Fax: +41.21.821.2021
Manufacturer
10/29/13
above statement.
The Skills Simulator, also referred to as the da Vinci Skills Simulator, is used in conjunction with
the Surgeon Console, Model SS3000.
The Surgeon Console, Model SS3000 also has been evaluated for compliance with standards
referenced above by a Nationally Recognized Test Laboratory (NRTL), Underwriters
Laboratories, Inc. (UL).
System Label
Note: This user manual identifies labels as they appear on systems configured for
destinations in the USA. The languages supplied on labels may vary by the country
or countries for which a system is configured.
Note: The unit identification label includes the serial number and date of
manufacture. It may be necessary for the reader to be as close as 6 in (15 cm) from
the label to read this information.
10/29/13
Space reserved
for unit identifiers
Power Requirements
The Skills Simulator requires electrical power independent of the Surgeon Console. To ensure
optimum performance, make sure the Skills Simulator is connected to a noise-free and
well-grounded AC power outlet. If the branch circuit supplying the Surgeon Console powers
no other loads, it is acceptable to power the Skills Simulator from the same branch circuit.
The Skills Simulator automatically adapts to 100VAC, 120VAC, or 230VAC. (This is called “Auto
Sense” in table 1-1 below.) Please refer to the electrical rating label on the Skills Simulator.
Table 1-1 Power Requirements
System Component Voltage Rating
10/29/13
100/120/230V~
Skills Simulator 50/60Hz 500VA Continuous
Auto Sense
Environmental Specifications
Table 1-2 Environmental Specifications
Environmental Conditions: Operating
Temperature: 10 to 30 °C (50 to 86 °F)
Humidity: 10 to 85% non-condensing
The Skills Simulator shall function properly in atmospheric pressures
ranging from 523mm HG (10,000 ft) to 774mm HG (-500 ft). For every
1000 feet above sea level, the 30 ºC operational temperature limit
Atmospheric Pressure:
specified above will be reduced by 1 ºC. (For example, the maximum
operating temperature at 5000 feet will be 25 ºC, and the maximum
operating temperature at 10,000 feet will be 20 ºC.)
The Skills Simulator is intended to be a practice tool that includes a variety of exercises and
scenarios to give a user the opportunity to improve his or her dexterity with the Surgeon
Console controls. The Skills Simulator is intended to augment, not replace, existing training
programs for the da Vinci Si Surgical System. Successful completion of the exercises is not
intended to convey readiness for surgery nor endorsement by Intuitive Surgical of a user's skill
level.
Illuminator on/off No
White balance No
Auto-calibration No
Up/down/left/right arrows (related to
No
calibration)
Scope selection No
2.3 Display Preferences
Digital Zoom No
TilePro No
Viewer mode Yes* (All navigation menus are 2D only)
Telestration eye No
Image depth Yes
3. Audio
Sound level Yes
Mute No
4. Utilities
4.1 Control preferences
Scaling Yes
Finger clutch enable Yes
10/29/13
TilePro QuickClick No
Haptic zoom No
Master associations No
4.2 Account Management Yes
4.3 Inventory management No
4.4 Event logs Yes
5. Lock Tools Yes
Video
• Brightness adjustments
• Advanced Video Adjustments
DRAFT/PRE-RELEASE/CONFIDENTIAL
• Brightness
• Contrast
• Red
• Yellow
• Edge enhancement
• Restore factory settings
• Display Preferences
• Digital zoom
• TilePro
• Viewer mode
• Telestration eye
• Image depth
Audio
• Sound level
• Mute
Utilities
10/29/13
• Control Preferences
• Scaling
• Finger clutch enable
• TilePro QuickClick
• Haptic zoom
• Account management
Laser Safety
Class 1 Laser Product: The Skills Simulator circuit boards may be equipped with optical
communication transmitters, which have been evaluated and found to be in compliance with
requirements found within 21 CFR (FDA-CDRH) and EN 60825-1 for Class 1 laser devices.
CAUTION: While Class 1 laser products are considered to be “eye safe” without the
need for additional protection, when working around the circuit boards, the
following general safety guidelines should be observed to reduce the risk of eye
injury:
• Do not look or stare at optical fiber ports or optical fibers that are connected to a light
source.
• Do not examine with optical instruments optical fiber ports or optical fibers that are
connected to a source. The use of optical instruments (for example, a magnifying glass)
may increase eye hazard.
• Adjustments and settings, maintenance, operational parameters and procedures other
than those specified and allowed herein and on the rating plate of the Skills Simulator
may result in increased eye hazard.
10/29/13
Note: The Skills Simulator weighs approximately 14 kg (33 lbs). To protect your back,
bend at the knees, and/or get help when picking up, moving, mounting, or
dismounting the Skills Simulator.
Note: Remove the Skills Simulator from the Surgeon Console if the Console is
transported through doorways that are less than 106.7 cm (42”) wide.
Note: It is not necessary to dismount the Skills Simulator from the Surgeon Console
during surgery, only to disconnect it as described above.
Note: Use the power switch to disconnect the Skills Simulator from mains power. To
completely isolate the Skills Simulator from AC mains, disconnect the power cord
10/29/13
In the event of suspected interference from other equipment, which prevents the proper
functioning of the Skills Simulator, contact Intuitive Surgical and/or discontinue use of the
system until the problem can be remedied.
Note: The Skills Simulator is to be used with the provided audio cable to connect to
the da Vinci Si Surgeon Console in order to ensure compliance with EMC
requirements. Use of cables or accessories other than those provided may result in
increased emissions or decreased immunity.
Note: The essential performance of the Skills Simulator during EMC testing was
defined using the following required tests:
• no component failures
• no noise in the image that makes it unusable
• no changes in programmable parameters
• no resets to factory defaults
• no change in operating mode
• no false alarms
• no initiation of any unintended operation
• no cessation or interruption of any intended operation
10/29/13
FCC Compliance
Note: This equipment has been tested and found to comply with the limits for a
Class A digital device, pursuant to Part 15 of the FCC Rules. These limits are designed
to provide reasonable protection against harmful interference when the equipment
is operated in a commercial environment. This equipment generates, uses, and can
radiate radio frequency energy, and if not installed and used in accordance with the
instruction manual, may cause harmful interference to radio communications.
Operation of this equipment in a residential area is likely to cause harmful
interference in which case the user will be required to correct the interference at his
own expense.
EMC Tables
The following tables contain the Manufacturer’s declaration and additional information
10/29/13
required by IEC60601-1-2:2007.
Table 1: Manufacturer’s Declaration – Electromagnetic Emissions
The Skills Simulator is intended for use in the electromagnetic environment specified below. The customer or the user of
the Skills Simulator should assure that it is used in such an environment.
Emissions Test Compliance Electromagnetic Environment – Guidance
The Skills Simulator uses RF energy only for its internal
function. Therefore, its RF emissions are very low and
RF emissions CISPR 11 Group 1
are not likely to cause any interference in nearby
electronic environment.
RF emissions CISPR 11 Class A The Skills Simulator is suitable for use in all
Harmonic emissions IEC 61000-3-2 Class A establishments, other than domestic establishments
and those directly connected to the public low-voltage
Voltage fluctuations/ flicker power supply network that supplies buildings used for
Complies
emissions IEC 61000-3-3 domestic purposes.
an uninterruptible power
supply.
Power frequency magnetic
fields should be at levels
Power frequency (50/60
characteristic of a typical
Hz) magnetic field IEC 3 A/m 3 A/m
location in a typical
61000-4-8
commercial or hospital
environment.
Note: UT is the AC mains voltage before application of the test level.
Note 1: At 80 MHz and 800 MHz, the higher frequency range applies.
Note 2: These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and
10/29/13
150 kHz to 80 MHz 80 MHz to 800 MHz 800 MHz to 2.5 GHz
d = 1.2 *SQRT(P) d = 1.2 *SQRT(P) d = 1.2 *SQRT(P)
0.01 0.12 0.12 0.23
0.1 0.38 0.38 0.73
1 1.2 1.2 2.3
10 3.8 3.8 7.3
100 12 12 23
For transmitters rated at a maximum output power not listed above, the recommended separation distance d in meters
(m) can be estimated using the equation applicable to the frequency of the transmitter, where P is the maximum output
power rating of the transmitter in watts (W) according to the transmitter manufacturer.
Note 1: At 80 MHz to 800 MHz, the separation distance for the higher frequency range applies.
Note 2: These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and
reflection from structures, objects and people.
Battery Disposal
10/29/13
The Skills Simulator contains lithium batteries that are not user serviceable. These batteries
must be disposed of in accordance with local regulations. Contact your local Intuitive Surgical
representative for disposal information.
_________________________________End of section______________________________
Note: The Skills Simulator weighs approximately 14 kg (33 lbs). To protect your back,
bend at the knees and/or get help to pick up, move, mount or dismount the Skills
Simulator.
Figure 2.3 Position the Skills Simulator on the floor behind the Surgeon Console as shown,
then prepare to lift carefully using the side handles.
8. Lift the Skills Simulator to the Surgeon Console. Position it so that the mounting brackets
will slide into the lateral supports on the back of the Surgeon Console (Figure 2.4).
10. Inspect the mounting brackets to ensure they are firmly in place before proceeding
(Figure 2.5).
10/29/13
Figure 2.5 Ensure that both mounting brackets are secure between the lateral supports
AC outlet
Figure 2.7 Connecting the Skills Simulator short fiber cable to the Surgeon Console
3. Connect the other end of the short fiber cable to either of the two receptacles on the
Skills Simulator (Figure 2.8).
10/29/13
Power switch
(in OFF position)
Power cord
Note: The protective metal caps attached to each short fiber cable should remain
installed on cable ends at all times when not connected to the system.
Note: Once the Skills Simulator is connected and powered on, the short fiber cable
should not be unplugged until the system has completely powered down.
Note: The short fiber cable has a fiber-optic core. Care should be taken to avoid
bending the cable, as kinks can damage the cable and may prevent system
operation. The minimum safe bend radius is 2.54 cm (1 inch).
Audio Line In
connector
AC Power Cord
Connect the power cord from the Skills Simulator (Figure 2.8) to an AC power outlet in the
wall. Make sure adequate power is available at the wall outlet according to Table 2-1. (See the
da Vinci Si user manual for the Surgeon Console power requirements.)
Table 2-1 Power Cord and Power Requirements
System Component Cord Length Power Requirement
500VA Continuous
Skills Simulator 12 ft/3.6 m 4.2A at 115V~
2.1A at 230V~
Note: Use the power switch to disconnect the Skills Simulator from mains power. To
completely isolate the Skills Simulator from AC mains, disconnect the power cord
from the AC power outlet.
External Monitor
DRAFT/PRE-RELEASE/CONFIDENTIAL
If desired, connect an external monitor to the Surgeon Console by connecting the monitor
video in (DVI) cable to the Video Out L (left) connector on the back of the Surgeon Console
(Figure 2.10). Make sure the monitor is also plugged into an AC power outlet in the room.
Video Out L
connector
3. Locate the black power switch on the side panel of the Skills Simulator and set it to the
I (ON) position (Figure 2.11).
DRAFT/PRE-RELEASE/CONFIDENTIAL
Power switch
(in ON position)
STATUS LED
illuminated
Power button
illuminated amber
Getting Started
Once the system is connected and powered on, you can begin practicing the various skills
exercises offered. For detailed software information, see Chapter 4 Simulation Software by
Mimic Technologies. For detailed software information related to Suturing Skills, see Chapter 5
Suturing Skills Simulation Software by Simbionix.
DRAFT/PRE-RELEASE/CONFIDENTIAL
The master controllers are used to control the pointer on the screen. To control the pointer,
first place your head in the stereo viewer on the Surgeon Console. Then lightly grasp the
master controllers as shown on the Welcome screen. To unlock each master controller, close
and release the master controller grip. You may then move either hand to control the pointer.
You will have to repeat this process to take control of the pointer if you remove your head from
the stereo viewer.
Note: To click a button on the screen, use the master controllers to move the pointer
to the button, then close and release either master controller grip to click it.
Note: The master controllers may be clutched during menu navigation, just as they
may be clutched during use of the regular da Vinci Si Surgical System. Applying the
clutch, either with the clutch foot pedal or finger clutches, allows you to move the
master controllers without moving the pointer. Refer to the da Vinci Si Surgical
System user manual for more information on clutching the master controllers.
2. Click Overview Video for a brief introduction to the Skills Simulator software. When the
video ends you will return to the Welcome screen (Figure 2.13).
3. Click Log In. The User Log In screen appears (Figure 2.14).
Navigation list
2. On the selected skills category screen, click any exercise button to select it. The selected
exercise screen appears (Figure 2.18).
DRAFT/PRE-RELEASE/CONFIDENTIAL
matching grips.
Exiting an Exercise
After Completing Exercise
The Skills Simulator will automatically exit an exercise once you have completed it. At the
completion of an exercise, the Performance Evaluation screen appears with metric results of
10/29/13
the exercise you just completed. This screen will also appear if you exit an exercise
prematurely.
Note: It is important that you briefly flick the left-side pod ergonomic switch if you
chose to exit a simulation exercise. Hold the switch down if you want to adjust the
Surgeon Console ergonomic settings.
Note: Exiting an exercise before it is complete will result in an incomplete score.
Note: Restarting (by pressing the blue restart button ) a Suturing Skills exercise
before it is complete will result in a lower score.
10/29/13
Text
Virtual EndoWrist
instruments Virtual objects
• 3.1 Onboard Simulator Administrator Overview and Log In, on page 3-1
• 3.2 Onboard Simulator Administrator Menu, on page 3-4
• 3.3 Reports, on page 3-7
Administrator Overview
Use the administrator feature to manage user accounts, change the administrator PIN, and
generate and save user reports that include exercise results.
10/29/13
Administrator Log In
1. Verify that the Skills Simulator is powered on and the main Welcome screen is displayed
(Figure 3.2).
DRAFT/PRE-RELEASE/CONFIDENTIAL
using the Sim Manager. Refer to the Skills Simulator Manager User Manual (PN 551116) for
more information on the Simulator Manager.
1. Click Edit Users. The Edit Users screen appears.
• Edit the user. Edit the first name, last name, or PIN and then click Update. The update
user confirmation screen appears. Click OK.
DRAFT/PRE-RELEASE/CONFIDENTIAL
3.3 Reports
The Simulator provides you with two ways to generate reports. The Administrator Menu lets
you save and view reports for all Simulator users (see Save User Reports – Administrator
Menu). The Main Menu lets you generate a single completion report – your own (see Save
Report – Main Menu).
2. Insert a USB memory stick into any of the USB ports on the Skills Simulator panel. The
Simulator saves the reports to the USB memory stick. Existing data on the memory stick
will not be erased.
DRAFT/PRE-RELEASE/CONFIDENTIAL
1. Click Save Report. The Skills Simulator Report Generator screen appears.
DRAFT/PRE-RELEASE/CONFIDENTIAL
3. Click Done.
4. Use your computer’s operating system to find and open the USB drive. You will find the
reports, which can be opened by any HTML-compatible browser.
5. Open the appropriate Reports folder.
Note: The metrics provided in this chapter are estimates, and as such Intuitive
Surgical does not intend for these metrics to reflect expected performance in an
actual surgical setting. Metrics are provided by Mimic Technologies with the intent of
mastering a particular simulated exercise, and should not be used to determine
proficiency for surgical operation.
See the da Vinci Si Surgical System user manual for information regarding the
Surgeon Console or other system components. This chapter covers the following topics:
• 4.1 Exercise Types, on page 4-1
• 4.2 General Exercise Tips, on page 4-2
• 4.3 Simulation Exercise Overview, on page 4-2
• 4.4 Exercise Scoring, on page 4-20
• 4.5 Simulator Targets, on page 4-25
performance. The Skills Simulator Mimic simulation software includes two types of practice:
1. quiz exercises, 2. simulation exercises.
Quiz Exercises
On the Main Menu, click Surgeon Console Overview to review and take quiz exercises.
These quiz exercises present multiple choice questions. For each question, you will study the
text and images and select the correct option from the choices provided. See Surgeon
Console Overview on page 4-2 for more information.
stress on the instruments. Stress is indicated by the instruments turning red (see
Excessive Force below).
• Approach from above and rotate the instrument wrists when picking up or guiding
objects.
• To avoid losing sight of one or both instruments, do not use camera zoom in large
increments.
There are tips associated with some exercises that may not pertain to all. These additional tips
are listed in each individual exercise summary under the Tips heading.
Excessive Force
Excessive force is defined as too much stress (or force) applied to an instrument, or to a target.
This excessive force can be caused when instruments collide with each other, or when
instruments collide with other objects in the scene. Pay special attention during tasks such as
tissue retraction, driving a needle, etc. Excessive force on the instruments is indicated by the
instrument shaft turning red (Figure 4.1).
Note: The quiz exercises will test you on material covered in the da Vinci Si Surgical
System user manual. The exercise screen will direct you to the appropriate section
of the manual to review the content.
DRAFT/PRE-RELEASE/CONFIDENTIAL
Tips
10/29/13
• While holding down the camera pedal, move both master controllers to manipulate the
camera view.
• If the arrow on the crosshairs disappears, or the crosshairs themselves disappear, it is
likely the camera target is behind your camera. To reacquire the camera target, simply
zoom out until the camera target reappears.
• Use multiple small camera sweeps when navigating to each blue camera target.
• Avoid instrument collisions with the cavity wall.
Scaling
DRAFT/PRE-RELEASE/CONFIDENTIAL
Tips
• This exercise requires the user to change the instrument scaling setting on the Touchpad.
This change will remain in effect after the exercise is completed.
10/29/13
• To activate the third instrument, tap the arm swap (left kick-plate) located on the left-side
of the footswitch panel.
• It may be necessary to use the clutch and camera in order to move the instruments
between targets while also keeping the grips in the center of the master workspace.
• Under lower scaling ratios, such as Quick, move the instruments slowly and deliberately
when near the targets in order to accurately hit them.
• Under higher scaling ratios, such as Fine, actively use the clutch and camera to keep your
grips in the center of the workspace while moving the instruments between targets.
Tips
10/29/13
the target sphere, or you are applying the wrong energy source.
Energy Switching 2:
• If a red circle appears around a target, you are either not correctly touching or grasping
the target sphere, or you are applying the wrong energy source.
Energy Dissection
10/29/13
Needle Driving
DRAFT/PRE-RELEASE/CONFIDENTIAL
Tips
• Plan the needle throw so that the needle will rotate into the center of the yellow insertion
target and out through the black extraction target in one smooth rotation.
• Use the left or right instrument for driving the needle to practice ambidextrous control.
Note: When an instrument is flashing yellow, it must be used to insert the needle
into the target.
Tubes
10/29/13
Peg Board
DRAFT/PRE-RELEASE/CONFIDENTIAL
Tips
• Rotate the instrument wrists to transfer the rings and place the rings onto the floor peg.
• Do not pull on a ring if it is still hooked on a peg.
• Avoid instrument collisions with the pegs and the floor.
• Always grasp the flashing ring on the peg board with the flashing instrument.
Tips
Match Board 1:
• When picking up objects, approach them from above and rotate the instrument wrists.
• Use the instrument closest to each character to pick it up.
Match Board 2:
Same as Match Board 1, and:
• Observe the graphic images on the door panels that serve as guides to open them.
DRAFT/PRE-RELEASE/CONFIDENTIAL
Match Board 3:
Same as Match Board 1, and:
• Observe the graphic images on the door panels and slider panels that serve as guides for
opening them.
Tip
• Alternate instruments as needed to navigate the curves of the rail.
Tips
Hand off the needle between instruments to correctly orient it so that you can pass it through
the flashing eyelet.
10/29/13
Games
Stacking Challenge
10/29/13
Tips
• Plan a strategy for stacking the boxes and dominoes before you enter following mode,
which will start the timer.
• Keep track of your time as indicated in the lower left of your screen.
Falling Dominoes
DRAFT/PRE-RELEASE/CONFIDENTIAL
MScore™
MScore is a performance evaluation tool developed by Mimic Technologies that provides
assessment based on a wide variety of metrics that measure robotic surgery skills. In addition
to viewing exercise reports in detail, users can keep track of their progress over time.
MScore launches automatically when you complete or quit an exercise.
DRAFT/PRE-RELEASE/CONFIDENTIAL
History and Metric History tabs. The Exercise Report tab is displayed by default. You can exit
MScore any time by clicking the Exit button.
Exercise Report
Tab
DRAFT/PRE-RELEASE/CONFIDENTIAL
progress bar represents a score in the low range or less. Any metric with a 0% score is
considered to be a critical error. These metrics should not be used to determine proficiency for
surgical operation.
Detailed information can be viewed by clicking on the box of the desired metric. This view
includes the definition of the metric, text-based feedback, raw value, percentage score,
baseline values used for calculating the percentage score, and the weight of the metric on the
overall score. The color of the text corresponds to the color of the evaluation icon displayed in
the metric box.
Overall score
The Overall Score box displays the overall score calculated based on individual metrics. Raw
values, percentage scores, base line values and weights of each individual metric are listed in a
tabulated format to explain how each metric score is calculated, and how much each
contributes to the calculation of the exercise score.
Detailed report
Click Overall Score to view the Detailed Exercise Report. Click Close to exit.
Score chart or by selecting an attempt number from the Attempt drop-down menu. Report of
an incomplete attempt displays “Incomplete” in the Overall Score box (and incomplete marks
in metric boxes).
Report for a specific attempt of a specific exercise
To view the reports of any attempt, select the desired exercise from the Exercise drop-down
menu, and then select the desired attempt from the Attempt drop-down menu, or click the
score bar of the desired attempt in the Overall Score chart.
To view history of any exercise, select the desired exercise from the Exercise drop-down
menu.
Camera Targets
A light blue sphere appears somewhere in the scene. The focal point of the endoscope is
indicated by a dark blue circle in center of the screen.
10/29/13
Handoff Targets
Grasp targets may link to form a handoff target. This requires the user to grasp the object with
the indicated instrument and pass it to a second instrument, while maintaining control of the
object.
10/29/13
Energy Targets
Small spheres of different colors that require applied energy. Each color represents an energy
type.
DRAFT/PRE-RELEASE/CONFIDENTIAL
Pass Targets
Require the user to move an object past a given location. The object and target location flash
10/29/13
yellow.
complete after the needle is passed completely through the insertion and extraction points,
and both points have turned green.
The Simbionix simulation software provides skills scenarios designed for practice using the
Intuitive Surgical da Vinci Si Surgical System. In this set of exercises you will practice tying
suture knots over various anatomical representations. You can complete tasks in categories
that range in order of complexity: Vertical Defect Suturing, Horizontal Defect Suturing,
Continuous Suturing, and Interrupted Suturing. The graded Suturing Skills scenarios record
metrics that assess performance based on individual improvement only; they do not assess
surgical proficiency.
Note: The metrics provided are estimates, and as such Intuitive Surgical does not
intend for these metrics to reflect expected performance in an actual surgical
setting. Metrics are provided with the intent of mastering a particular simulated
exercise, and should not be used to determine proficiency for surgical operation.
See the da Vinci Si Surgical System user manual for information regarding the Surgeon
Console or other system components. This chapter covers the following topics:
• 5.1 Suturing Skills Exercise Instructions, on page 5-1
• 5.2 General Exercise Tips, on page 5-2
• 5.3 Suturing Skills Exercise Overview, on page 5-6
• 5.4 Scoring, on page 5-10
10/29/13
Suturing Constraints
Suturing Breakage Under Tension
The suture may break when placed under too much tension; or, when there is too much
tension when tightening a knot.
DRAFT/PRE-RELEASE/CONFIDENTIAL
Figure 5.4 Suture breaks under tension | suture breaks when tightening knot
Collision Detection
The Simulator provides enhanced collision detection and feedback for when instruments
collide; or when instruments collide with the needle.
General Information
Restart Button
Move an instrument to touch the restart button ( ) at the upper corner of the screen to
receive a new suture at any time.
10/29/13
• Suturing Practice
Suturing Skills
Interrupted Suturing
A module that presents an environment to practice techniques involved with an interrupted
running suture.
10/29/13
Note: If you are unable to complete the three surgeon’s knots with a single piece of
suture, touch the Additional Suture button ( ) to request additional suture.
Suturing Practice
Represents an environment to practice Vertical Defect Suturing, Horizontal Defect Suturing,
DRAFT/PRE-RELEASE/CONFIDENTIAL
The Suturing Skills metrics are grouped into three categories: Knot-Tying, Needle Handling,
and Time Efficiency. To see the details for each metric, click the category grouping, then click
the specific metric to see how the performance was determined (Figure 5.19).
Achievement
10/29/13
Remember that Simbionix Suturing Skills exercises are based on improvement. To set a
benchmark, the exercise must be performed at least once. Please see the table below for a
detailed description of Suturing Skills exercise metrics. You should adhere to the
recommendations listed in the table to practice towards a passing score.
Below Benchmark/ Below Benchmark/ Below Benchmark/
Metric Description
Above Average at Average Below Average
Time to Complete Total time to complete task. N/A N/A N/A
Off-screen Needle Number of times a grasped You improved your average Keep your instruments in Keep your instruments in
Carries needle is outside the field of score and can still do better. view when grasping the view when grasping the
view. Continue to watch that you needle. needle.
don't carry your needle
off-screen.
Exercise Restarts Number of times the You improved your average You restarted the exercise You restarted the exercise
exercise is restarted via the score and can still do better. too many times. too many times.
blue restart button. Try to complete the exercise
without restarting.
Needle Passages The total number of times You improved your average Try to complete each Try to complete each
the needle passes score and can still do better. needle target with a single needle target with a single
completely through tissue Try to complete each pass without pulling the pass without pulling the
(inserted in one point and needle target with a single suture through. suture through.
extracted through another pass without pulling the
point). suture through.
_________________________________End of section________________________________
USB ports
Fiber cable
Power switch
(in OFF position)
Disconnect Cables
Once the Skills Simulator is shut down, proceed as follows:
1. Disconnect the optional keyboard and mouse, if present, from the Skills Simulator side
panel USB ports (Figure 6.2).
2. Disconnect the optional monitor DVI cable, if present, from the Video Out L connector
on the back of the Surgeon Console (Figure 6.3).
Video Out L
Audio in
cable
DRAFT/PRE-RELEASE/CONFIDENTIAL
Fiber cable
DRAFT/PRE-RELEASE/CONFIDENTIAL
before use.
Note: The system cables and their receptacles contain fiber-optics. Only when
requested by Intuitive Surgical personnel, clean both cable and receptacles by
blowing the fiber ends with a generic dry and oil-free canned air duster.
_________________________________End of section______________________________
10/29/13
Purpose
The purpose of this appendix is to provide the user with an overview and use instructions for
the da Vinci OnSite for Skills Simulator system networking solution. da Vinci OnSite provides
these basic capabilities and benefits:
1. Automated software update retrieval, where the Skills Simulator downloads scheduled
updates from an ISI server.
2. Install new updates to Skills Simulator training exercises.
3. Install updates to firmware.
General Precautions
da Vinci OnSite for Skills Simulator must be used in accordance with this manual, and only by
personnel who have been trained in its use by an authorized Intuitive Surgical, Inc. (ISI)
representative. Read all instructions carefully. For information specific to da Vinci OnSite for Si,
and networking infrastructure details, see the da Vinci Si Surgical System User Manual
(PN 550650).
Note: Ethernet networks (both wired and wireless) are subject to losses of
10/29/13
CAUTION: da Vinci OnSite for Skills Simulator components may be installed and
serviced only by Intuitive Surgical personnel. Do not attempt to install or service
equipment without Intuitive Surgical personnel.
Network Connections
DRAFT/PRE-RELEASE/CONFIDENTIAL
OnSite for Simulator requires a wired RJ45 Ethernet 10bT/100bT/1000bT and/or wireless
802.11 network connection with Internet access in the OR where the Skills Simulator is to be
used.
A.2 Introduction
When Intuitive Surgical field service personnel install and configure system software and
hardware to enable OnSite functionality, the Skills Simulator can connect to the ISI dVMS
Server and download all software updates.
OnSite is a proprietary Intuitive Surgical product leveraging the existing Intuitive Surgical Field
Service staff and applications.
An OnSite-equipped Skills Simulator sends and retrieves software status and updates
regardless of system state and mode.
System Overview
OnSite for Simulator consists of three major components, that is the Skills Simulator backpack
networking components, the dVMS Server Infrastructure, and the remote user
(Intuitive Surgical Field Service personnel).
To implement remote capabilities, the Skills Simulator must have access to the Internet. OnSite
10/29/13
is designed to accomplish this using existing hospital networks. It therefore conforms to most
hospital IT policies and security standards.
Note: Once the Ethernet cable is connected, it can take up to two minutes to detect
the da Vinci network and update the status on the Welcome screen.
The wireless adapter contains the following Transmitter Module: FCC ID: XU8TEW648UBM.
Specifications
• The Wireless Connectivity Option utilizes the IEEE 802.11 wireless standard using either
802.11b or 802.11g at 2.4 GHz Industrial, Scientific, and Medical (ISM) band.
• The Wireless Access Point must be located within 75 feet of the Skills Simulator.
Security
• Wireless Network Infrastructure
• Intuitive has tested the Wireless Connectivity Option in WPA2 pre-shared key mode
with AES encryption, and recommends that the Wireless Connectivity Option is
integrated into the hospital network using this security configuration.
• da Vinci OnSite Software Application
• The OnSite software application uses a Secure Socket Layer (SSL) session based on
unique certificates on the Skills Simulator and the server.
Note: It is important to note that if the wireless network is modified or updated after
the Wireless Connectivity Option is installed, its suitability to support the wireless
applications should be re-assessed. In particular, contact Technical Support if any of
the following changes are planned or have occurred.
• If the Wireless Access Point or Skills Simulator is moved from the location where it resided
during installation.
• If the Wireless Access Point is replaced with a new make or model.
7.6 meters
or 25 feet
10/29/13
7.6 meters
or 25 feet
Signal
Path
Wireless Adapter
• If a microwave oven causes interference, use the wired backup to correct the problem.
• The RFID reader not operate in frequency hopping mode if it is being used in the same
room as the Wireless Connectivity Option.
• A separation of a least 4 MHz exists between the frequency range of the channel being
used by the wireless adapter/WAP and the operating frequency of the RFID reader. For
example, a wireless adapter/WAP operating on channel 10 spans 2.446 – 2.468 GHz;
therefore, to avoid disturbances from the RFID reader, its operating frequency should be
less than or equal to 2.442 GHz or greater than or equal to 2.472 GHz.
Valid RFID Reader Valid RFID Reader
DRAFT/PRE-RELEASE/CONFIDENTIAL
If there is a need to disable all network connectivity for the Skills Simulator, disconnect the
150 Mbps Micro Wireless N USB Adapter.
Note: To re-establish network connectivity, you must re-connect the indicated
150 Mbps Micro Wireless N USB Adapter.
Electromagnetic Compatibility
The essential performance for Skills Simulator wireless connectivity during EMC testing was
defined as follows during any of the required tests:
• No component failures
• Video quality exceeded pre-defined metric demonstrating that the video quality was not
impacted
• Audio script test passed demonstrating that the audio link was not impacted
• No changes in programmable parameters
• No resets to factory defaults
• No change in operating mode
• No false alarms
• No initiation of any unintended operation
Compliance Information
DRAFT/PRE-RELEASE/CONFIDENTIAL
CAN/CSA-C22.2 No 60950-1-03
Network Router:
ANSI/UL Std No 60950-1 1st Ed.
802.11b: up to 11 Mbps
Wireless Data Rate
802.11g: up to 54 Mbps
wireless
DRAFT/PRE-RELEASE/CONFIDENTIAL
• One-touch wireless connection with external Wi-Fi Protected Setup (WPS) button
• Backwards compatible with IEEE 802.11g and IEEE 802.11b devices
• Maximum reliability, throughput, and connectivity with automatic data rate switching
• Supports 64/128-bit WEP, WPA/WPA2-RADIUS, and WPA-PSK/WPA2-PSK
• Wi-Fi Multimedia (WMM) Quality of Service (QoS) data prioritization
• Easy user setup and intuitive diagnostic utility
• Coverage of up to 30 meters indoor (98feet) and 50 meters outdoor (164 feet) *
• 3-year limited warranty
*Maximum wireless signal rates are referenced from IEEE 802.11 theoretical specifications. Actual data throughput and coverage will vary depending on
interference, network traffic, building materials and other conditions.
TEW-648UBM
rev:09.30.2010
SPECIFICATIONS
Hardware
Interface • USB 2.0
DRAFT/PRE-RELEASE/CONFIDENTIAL
Standards • IEEE 802.11b, IEEE 802.11g, and IEEE 802.11n (draft 2.0)
Button • WPS button: enables quick wireless connection with WPS function
Power Consumption • Receive mode: 100mA (max.) • Transmit mode : 210mA (max)
Supported OS • Windows 7 (32/64-bit), Vista (32/64-bit), XP(32/64-bit)
Dimensions (LxWxH) • 20 x 15 x 7 mm (0.8 x 0.6 x 0.3 in.)
Weight • 1 g (0.03 oz)
Temperature • Operating: 0° ~ 40° C (32° ~ 104° F) • Storage: -10° ~ 70° C (14° ~ 158° F)
Humidity • Max. 95% (non-condensing)
Certifications • CE, FCC
Wireless
Modulation • OFDM, DSSS
Antenna • Built-in on board antenna
Frequency • 2.412 ~ 2.484 GHz
Data Rate (auto fallback) • 802.11b: up to 11Mbps • 802.11g: up to 54Mbps • 802.11n: up to 150Mbps
Output Power • 802.11b: 14dBm (typical) • 802.11g: 11dBm (typical) • 802.11n: 10dBm (typical)
Receiving Sensitivity • 802.11b: -84dBm (typical) at 11mpbs • 802.11g: -69dBm (typical) at 54Mbps • 802.11n: -66dBm (typical) at 150Mbps
Encryption • 64/128-bit WEP, WPA/WPA2-RADIUS, WPA-PSK/WPA2-PSK
Channels • 1~11 (FCC), 1~13 (ETSI)
RELATED PRODUCTS
TEW-651BR 150Mbps Wireless N Home Router
TEW-650AP 150Mbps Wireless N Access Point
TEW-647GA Wireless N Gaming Adapter
CONTACT INFORMATION
TRENDnet is a registered trademark. Other Brands and product names are trademarks of their respective holders. Information provided in this document pertain to TRENDnet
products and is subject to change at any time, without notice. For the most recent product information please visit http://www.trendnet.com.
Copyright © TRENDnet. All Rights Reserved.
This is a feature to ensure that a user’s session is not interrupted, should he or she
happen to be completing exercises on the Skills Simulator at the planned time of
download.
2. At the time specified on the Welcome screen, and when the Skills Simulator is in sleep
mode and not in use, the system will automatically start the installation process.
• The Skills Simulator backpack does not need to be connected to the Surgeon Console
for the updates to take place.
Before starting the update, the system will check for the following parameters:
• No dual console connected
• No patient cart connected
• No vision cart connected
3. Press the power button on top of the Skills Simulator (Figure A.6). The system starts and
the welcome screen appears in the stereo viewer on the Surgeon Console. At log in, the
10/29/13
screen displays the software version and date of last upgrade (Figure A.7).
If the update is unsuccessful, a message appears on the Welcome screen (Figure A.8).
DRAFT/PRE-RELEASE/CONFIDENTIAL
A.4 IT Requirements
Internet Access
The wireless adapter device that will be integrated into the Skills Simulator requires Internet
access to contact servers at Intuitive Surgical.
ISI requires a wired RJ45 Ethernet 10bT/100bT/1000bT and/or wireless 802.11 network
10/29/13
connection with Internet access in the OR where the Skills Simulator is to be used. If your Skills
Simulator is used in multiple locations, then ISI requests that network access be made
available in each location.
da Vinci OnSite is compatible with both DHCP and Static networking addresses.
Proxy Server
All IP proxy servers are compatible with OnSite for Simulator.
Firewall
OnSite for Simulator requires outbound port 443 open.
Network Topology
OnSite for Simulator requires a minimum amount of bandwidth to post log files (generally less
than 1 MB per day).
________________________________End of section______________________________
B Appendix B: Troubleshooting
This appendix provides information to help troubleshoot setup problems.
issues, you should perform a reboot of the system, see section B.2 Reboot the Skills Simulator.
If symptoms persist after following the suggested remedies and/or rebooting the system,
please contact Intuitive Surgical For Technical Support. In the U.S., call 1-800-876-1310, where
phones are staffed 24 hours a day, seven days a week. In Europe, call +41.21.821.2020.
nonrecoverable fault red triangle a Surgeon Console with a different has compatible software.
appears on-screen. software version. • To update a Surgeon Console with
compatible software, contact
Intuitive Surgical Technical
Support.
• Turn Skills Simulator off, verify all
Nonrecoverable Fault: Red triangle Short blue fiber cable became
connections, and reboot the
appears on-screen. disconnected during use.
system.
• Turn the Skills Simulator off, verify
Recoverable Fault: Yellow triangle
all connections, and reboot the
appears on-screen.
Blue fiber cable is loose, or it may be system.
dirty. • Follow maintenance instructions
Error Message: “Fiber cable connectors
on page 6-5: Cleaning the Skills
require cleaning”.
Simulator.
Audio problems: Video has no sound; • Turn Skills Simulator off, verify all
The audio cable is not hooked up
other features (such as energy connections, and reboot the
correctly.
instruments) have no sound. system.
• Turn Skills Simulator off, verify all
Video problems: Skills Simulator and
Blue fiber cable is unattached. connections, and reboot the
Surgeon Console will not connect.
system.
3. Set the Skills Simulator power switch to the OFF position. All LED will turn off at this point.
4. Set the power switch on the Simulator back to ON. Wait until the power buttons on the
Skills Simulator and the Surgeon Console illuminate steady amber.
5. Press the power button on the Skills Simulator or the Surgeon Console. The system
reboot is complete once the power buttons illuminate steady blue.
Note: The Skills Simulator may be moved from one Surgeon Console to another; but,
the software version of the Skills Simulator must be compatible with the software
version of the Surgeon Console it is connected to.
Note: The software may load on the Surgeon Console touchpad before it is finished
loading on the Skills Simulator (the stereo viewer and optional additional monitor
may be affected by this). During this momentary delay, you may see a blank screen;
this is expected and normal.
________________________________End of section______________________________
10/29/13
components.
Note: Storage and transport of the system are not affected by atmospheric pressure.
3D Three-dimensional.
The informal term applied to the Skills Simulator, which resembles a backpack
Backpack
when mounted to the Surgeon Console.
10/29/13
The Default User, included with the Simulator, can run all of the exercises on
Default user
the Simulator with no restrictions.
Any one of several tools used to effect the procedure in the patient once
attached to an instrument arm and inserted into the patient. Instruments
Instrument
include, for example, Large Needle Drivers, DeBakey Forceps and Round Tip
Scissors. Instruments exist only in a virtual sense on the Skills Simulator.
Metrics Measures of user performance that are generated and saved by the Simulator.
Controls on the left end of the Surgeon Console arm rest that provide
Left-side pod
ergonomic adjustment. Also used in the Simulator to exit an exercise.
Performance evaluation A screen that contains the metric results of a completed exercise.
Exercise report A comprehensive report based on user metrics that is generated by MScore.
Scope Endoscope.
Simulator assembly Refers to the Skills Simulator hardware, informally known as the “backpack.”
A shorter name for the da Vinci Skills Simulator (see below). Also referred to as
Skills Simulator
the Simulator.
The full name for the Skills Simulator (see above), referred to more briefly as
10/29/13
The viewing system that comprises the upper portion of the Surgeon Console,
Stereo Viewer or 3D Viewer
where the surgeon looks to see the 3D image.
_________________________________End of section______________________________
Labels
Connections System 1-4
Cables 2-4 Laser Safety 1-9
Simulator (Side Panel) 2-5 Laying out the system cables 2-4
Contact Information 1-3
Contraindications 1-8 M
Crate Dimensions C-1 Manufacturer’s Declaration
Customer Service Electromagnetic Emissions 1-12
Contact Information 1-3 Electromagnetic Immunity 1-13
Match Board 4-14
D Metric History tab 4-23
Degree of Protection 1-3 Metrics 4-24
Dimensions Mimic 1-3
Crate C-1 Mode of Operation 1-3
Disposal Information MScore 4-20
Battery 1-15
Dots and Needles 4-10, 5-7 N
Needle Targeting 4-16
E
Electrical Rating 1-5 O
Electromagnetic Operation