Operation Card of BS-400 Chemistry Analyzer

1 Checking before Powering On software will be started automatically.

1 Check the power supply and make sure it can supply 4 Setting Parameters
proper voltage for the analyzer.
2 Check the connections among the analyzing unit, Before requesting tests, perform the following steps to
operation unit and printer. Make sure the connections finish the settings:
are right and secure. Check the power cords of the
analyzing unit, operation unit and printer, and make 9 Select ParametersÆTestÆBasics in succession,
sure they are well connected to the power sockets. and enter test parameters following the reagent
package insert.
3 Check and make sure sufficient printing paper is
prepared for the printer. 9 Select ParametersÆTestÆReference in
4 Check if concentrated wash solution A and B are succession, and set the reference ranges.
sufficient. If not, please refill. 9 Select ReagentsÆReagent Setup, and set up the
5 Check the reagent disk and ensure alkaline and acid reagent information.
wash solutions are placed respectively on position D1 9 Carryover: Select ParametersÆCarryover, and
and D2, and distilled water on position W. If not, set up reagent carryover.
please refill. 9 Select ParametersÆTest ÆCalibration, and set
6 Check the sample disk and ensure the following up calibrators and calibration rule.
solutions are placed correctly. 9 Select ParametersÆTest ÆQC, and set up
Position U: Urine diluent; controls and QC rule.
Position D1: ISE cleaning solution; 9 Select SetupÆHospital, and set up the hospital
information.
Position D2: Acid wash solution;
9 Select SetupÆPrint, and set up printer and print
Position D3: Alkaline wash solution; templates.
Position W: Distilled water 9 Select SetupÆLIS, and set up LIS host and
If not, please refill. communication mode.
7 If your analyzer is equipped with the ISE module, 9 Select SetupÆBar Code, and set up sample and
please make sure the reagent pack is installed with reagent bar code rules.
sufficient reagent.
8 Check the connection of deionized water tank and
5 Loading Reagents
turn on the inlet valve.
Load reagents to their assigned positions on the
9 Check the drainage of low-concentration waste. Make reagent disk, and then remove the bottle caps. Please
sure the tubing is not bent and the waste level is not select correct bottle type as configured so that the
higher than the waste outlet of the analyzer. reagent inventory can be checked accurately.
10 Check the connection of high-concentration waste
bucket and make sure it is emptied. 6 Programming Calibrators
11 Check if leakage or air bubbles exist on the sample Run calibrations when needed.
syringe and reagent syringe.
12 Check the sample probe and make sure the probe is Select CalibrationÆCalib. Request, and request
not contaminated or bent. If the probe is contaminated calibration run for desired tests.
or bent, clean or replace it.
After requisition is finished, place the calibrators in
13 Check the reagent probe and make sure the probe is
assigned positions on the sample disk.
not contaminated or bent. If the probe is contaminated
or bent, clean or replace it.
Select the button to start running the calibrators.
14 Check the sample mixer and reagent mixer and make
sure the mixers are not contaminated or bent. If the Select CalibrationÆResults and select a test from the
mixers are contaminated or bent, clean or replace Tests list to view the results.
them.
7 Programming Controls
2 Powering On
Select QCÆQC Request, and request quality controls
Connect the power cord of the system to the socket and then for desired tests.
switch on the powers in the sequence presented below:
After requisition is finished, place the control samples in
a. Turn on the water unit; assigned positions on the sample disk.
b. Place the Main Power to the ON position;
Select the button to start running the controls.
c. Place the Power to the ON position;
To view QC results, select QCÆReal-time (Daily/Day
d. Turn on the monitor of the computer; to Day).
e. Turn on the computer; 8 Programming Samples
f. Turn on the printer.
„ Request samples
3 Starting the Operating Software 1) Request samples manually (in any conditions)
Start the Windows operating system. The BS-400 operating Select SamplesÆSample Request, and request

1
Operation Card of BS-400 Chemistry Analyzer
samples with desired tests. button to download samples from the LIS host that are
identified successfully by the bar code reader.
After requisition is finished, place the samples in assigned
positions on the sample disk. Select SamplesÆSample Request, and select the
Scan button to scan bar code labels of the added
2) Download samples from LIS (when the system is connected samples.
to LIS)

Load the samples on the sample disk; Select the button to start running the samples.

Select SamplesÆSample Request, and select Download to „ Adding/Deleting tests


download samples from LIS that are configured or identified
Select SamplesÆSample Request, then select a
successfully by hand-held bar code reader.
sample and select desired tests that you want to add or
3) Scanning/Downloading samples (when the system is delete.
equipped with the fixed sample bar code reader and LIS)
12 Rerunning a Sample
Place the samples in any positions on the sample disk.
When analysis is finished, you can rerun desired
sample in two modes: Manual and Auto.
„ Select the button to start running the samples.
„ To view test results, select SamplesÆCurrent (or To rerun a sample manually, you can:
History).
„ Select SamplesÆCurrent, then select desired
9 Programming STAT Samples sample and click the Rerun button. Or

„ Requesting/running STAT samples „ Select StatusÆSample Disk, then select desired


sample on the disk graph and click the Rerun button.
1) Emergent samples
To rerun a sample automatically, you should:

Select and then request emergent samples on the Set up the rerun conditions on the System page of the
Request Emergent Samples dialog box. Setup screen. The system will rerun the samples
according to the conditions you have configured.
After entering the test information, place the samples on
assigned positions and then select . 13 Editing Test Results

2) Ordinary STAT samples When analysis is finished, you can edit the test results if
needed.
Select SamplesÆSample Request, then request samples
with desired tests and tick the check box to the left of STAT. ‹ Note: Test results can only be edited by
authorized personnel.
After requisition is finished, place the samples in assigned
positions on the sample disk. Select SamplesÆCurrent (or History), then select the
Edit button and edit desired test result on the popup
dialog box.
Select the button to start running the samples.
To make linear transform to the results of one or more
„ To view test results, select SamplesÆCurrent (or tests, select a test on the Current or History page(By
History). Test), then select a test and click Compen..
10 Programming Reagent Blanks 14 Printing/Sending Test Results
If needed, follow this procedure to run reagent blanks. Select SamplesÆCurrent (or History), then select
desired test results and click the Print/Send button.
Select ReagentsÆReagent Blank and request reagent
blanks. 15 Exiting the Operating Software

Select the button to start running the reagent blanks. When analysis is finished and the system is in Idle
status, select Shutdown to exit the operating software.
To view test result of reagent blank, select date range on the
View Blank Results window of the Reagent Blank page and 16 Powering Off
search for desired reagent blanks.
After exiting the Windows operating system, following
11 Adding/Deleting Samples and Tests this procedure to turn off the powers:

„ Adding new samples a. Turn off the printer;


b. Turn off the monitor of the computer;
1) Adding samples manually
c. Place the Power of the analyzing unit to OFF;
Select SamplesÆSample Request and request samples with d. Turn off the water unit.
desired tests.
17 Checking after Powering Off
After requisition is finished, place the samples in assigned
positions on the sample disk. 1 Cap every reagent bottle on the reagent disk.
2 Remove the calibrators, controls and samples
Select the button to start running the samples. from the sample disk.
2) Obtaining samples automatically 3 Check the surface of the analyzing unit for
stains, and if any, wipe them off with clean soft
Load the samples on the sample disk; cloth.
4 Check the high-concentration waste bucket
Select SamplesÆSample Request, and select the Download and empty it if necessary.