You are on page 1of 6

Journal of Hospital Infection (2002) 5l: 79–84

doi:l0.l053/jhin.2002.l2l7, available online at http://www.idealibrary.com on

REVIEW

Environmental controls in operating theatres


S. Dharan and D. Pittet
Infection Control Programme, Department of Internal Medicine, University of Geneva Hospitals,
I2II Geneva I4, Switzerland

Summary: Surgical-site infection is the leading complication of surgery. Normal skin flora of patients or
healthcare workers causes more than half all infections following clean surgery, but the importance of air-
borne bacteria in this setting remains controversial. Modern operating theatres have conventional plenum
≤ µm are removed. For orthopaedic and other implant surgery,
ventilation with filtered air where particles 5
laminar-flow systems are used with high-efficiency particulate air (HEPA) filters where particles ≤0.S µm are
removed. The use of ultra-clean air has been shown to reduce infection rates significantly in orthopaedic
implant surgery. Few countries have set bacterial threshold limits for conventionally ventilated operating
rooms, although most recommend 20 air changes per hour to obtain 50–150 colony forming units/m S of air.
There are no standardized methods for bacterial air sampling or its frequency. With the use of HEPA filters
in operating theatre ventilation, there is a tendency to apply cleanroom technology standards used in
industry for hospitals. These are based on measuring the presence of particles of varying sizes and
numbers, and are better suited than bacterial sampling. Environmental bacterial sampling in operating
theatres should be limited to investigation of epidemics, validation of protocols, or changes made in
materials which could influence the microbial content.
© 2002 The Hospital Infection Society
Keywords: Surgical wound infection; operating rooms; environmental monitoring; quality control; reference
standards; air microbiology.

Introduction of operating theatres has been advocated. There is


Surgical-site infections are still a major problem in no international consensus on the methods, types
modern medicine. These infections may be deep or of sampling and tolerable limits of bioburden in
superficial. The superficial ones are easier to deal operating theatres. The main parameters
with, but the deep ones can be complicated, associated with environmental biocontamination in
leading to re-operation, or even be life threatening. operating theatres are discussed with a special
Factors causing surgical-site infection are emphasis on air quality and its control.
multifarious; type of operation, surgeon*s skill,
insertion of foreign material or implants, Source and transport of organisms
appropriateness of surgical preparation, adequacy
and timing of antimicrobial prophylaxis, the immune The control of surgical-site infection requires
state of the patient and contamination of the a knowledge of the source and transport of the
inanimate environment. To reduce these infections, cau- sative organisms. In today*s operating
bacteriological management environment, more than half of clean surgical-site
infection pathogens originate from normal skin flora
Author for correspondence: Professor Didier Pittet, of patients or staff.1–S Bacteria on skin squames, lint
Infection Control Programme, Department of Internal Medicine, and other dusts get into the air in the operating
University of Geneva Hospitals, 1211 Geneva 14, Switzerland; theatre and by turbulent air currents deposit on
Tel. (office): ‡ 41-22/SU2 98 28; Fax: ‡ 41-22/SU2 S9 8U. surfaces. They are also spread by direct contact
E-mail: didier.pittet@hcuge.ch
between

0l95–670l/02/060079 + 06 $35.00/0 © 2002 The Hospital Infection Society


80 S. Dharan and D.
Pittet

carrier and wound, but the importance of airborne Ventilation


bacteria as a source of infection remains a subject
of debate among professionals in infection Most modern operating theatres have conventional
control.S–6 The inanimate environment is probably plenum ventilation with filtered air, using filters with
one of the factors that can be managed efficiently an efficiency of 80–95% to remove airborne
by the infec- tion control team. ≤ particles 5 µm [DOP test25]. Laminar air-flow
Several studies, including experience at our systems with HEPA filters which remove
institution, have shown a reduced number of infec- airborne particles of
tions when orthopaedic surgery is performed in 0.S µm and above with 99.9U% efficiency are
operating theatres with ultra-clean air facilities. U–14 generally used for orthopaedic and other implant
Thus, it can be assumed that for this type of surgery. HEPA-filtered laminar air-flow can be
surgery, higher bacterial counts in the air correlates supplied to the operating area by ceiling-mounted
with a higher risk for surgical site infection. (vertical flow) or wall-mounted (horizontal flow)
To reduce prolonged morbidity and healthcare units. Both systems have their inconveniences and
costs associated with these infections, airborne the unidirectional flow of air can be disrupted. With
bac- teria and other sources of contamination must the vertical flow system, heat generated by surgical
be reduced to the minimum. This can be achieved lamps creates minor air turbulence. A horizontal
by the choice of staff theatre dress, patient laminar air-flow system was developed to overcome
preparation and draping, education of personnel the problems associated with vertical air-flow.
and design and controlled ventilation of operating However, this system has a major inconvenience in
theatres.15–2S The infection control team*s role that the operating team usually disrupts the uni-
should be that of a consultant where expertise in directional air-flow. Some studies have shown that
materials and methods in asepsis are shared vertical laminar air-flow generates less bacteria at
among the other professionals involved in the the operating site.8,26 To overcome the problems
design, layout and functioning of operating asso- ciated with both horizontal and vertical flow
theatres. systems, a combination of vertical and horizontal
flow has been developed, known as exponential
laminar air- flow with the form of an upside-down
Clothing and healthcare trumpet.2U,28 Few countries have set bacterial
worher education threshold limits
in conventionally-ventilated operating theatres,
Surgical drapes and the fronts and sleeves of although most recommend 20 air changes per hour
surgical gowns should be impregnated with a fluid in order to obtain 50–150 colony forming units
repellent and made of a non-woven or close-woven (cfu)/mS of air. In the United Kingdom, the limit is
fabric.18 In the European Medical Devices S5 cfu/mS for an empty operating theatre and in
Directive, these articles are considered as medical activity it should not exceed 180 cfu/mS for an
devices, and as such they should pose no infection average 5 min period. In an ultra-clean air
risk for the user and the patient. The requirements operating theatre the
for surgical drapes and gowns have been defined limit is set at c10 cfu/mS sampled within S0 cm of
(resistance to bacterial penetration wet and dry, the wound using conventional clothing. The limit
linting, tensile strength, etc.) and should be met by is set at c1 cfu/mS of air when total body exhaust
the manu- facturers and distributors. Surgical staff gowns are used (UK Department of Health docu-
should wear hoods and masks although the ment, Health Technical Memorandum 2025).
wearing of masks is controversial.19,21,24 Operating These limits are not always easy to achieve
theatre dress made of non-woven fabric or during an
impermeable material can be uncomfortable so operation as various factors influence the
climatic conditions in the operating theatre should bioburden in the air during the surgical procedure.
compensate for this. Staff should be educated in There are no standardized methods for air sam-
the maintenance of sterility of instruments by no- pling in operating theatres or for its frequency. In
touch behaviour, covering of sterile instruments addition, a variety of instruments are used in volu-
with a sterile drape until use, and movement and metric air sampling which renders correlation of
number of persons in the operating theatre should results difficult due to variability.
be kept to the minimum.20,22 With the use of HEPA filters in operating
theatre ventilation, there is a tendency to apply
clean- room technology standards used in industry
for hospitals. Most of the industrialized countries
have
Environmental controls in operating 8l
theatres

set their own standards, usually modifying the between sampling should be decided by each insti-
American Federal Standard 209E to their local tution based on the means available.
needs (Table I).29 These standards are based on We are not proponents of routine bacteriological
measuring the presence of particles of varying surveillance of air and surfaces in conventionally-
sizes and number [British standard 5295, Table ventilated operating theatres due to the fact that
II(a); German VDI 208S, Table II(b)]. S0,S1 Many of results obtained are valid only for the moment and
these are presently being amended to align with location where they were obtained. Different
the International Standards Organization (ISO) factors influence the results and we cannot
14644 [Table II(c)].S2 presume that results will be the same an hour or a
The European Standard, ³Cleanroom technol- day later. Fur- thermore, in a clinical setting not just
ogy*, is presently in its draft form. This document the quantity of bacteria is important, but also its
covers methods of analysing and measuring quality. Taking these factors into consideration, it
aero- biocontamination in zones at risk, classified would be a waste of resources to routinely take
accord- ing to risk categories 1 to 4. This bacterial samples on a regular, i.e., monthly basis.
document does not set limits in bioburden or Rather, we recommend annual maintenance of
frequency in sampling; the decision is left to the ventilation systems by the engineering department.
individual institution. Systems (ultra-clean air) Bacteriological sampling has its place in the
should be maintained to standards that will fulfil investigation of epidemics, validation of changes in
their desired functions. However, to measure products and procedures in the maintenance of
these, controls have to be instituted. At present, operating theatres (cleaning, disinfection, and
there is no international consensus on the ventilation) and education. Volumetric air sampling,
methods, types of samples (settle plates vs. either particle or bacterial counts, should be carried
volumetric air sampling), frequency of sampling, out in an empty operating theatre after any main-
and tolerable limits of bioburden in operating tenance work carried out in the ventilation system
theatres.26,SS–S8 Based on experience at our and should meet previously set target limits. The
institution, we consider that a standard limits set in Table III are not based on scientific
established on measured particle size and evidence of the relation between contamination
number, similar to the American Federal of the environment and surgical-site infection
Standard 209E (Table I) would be better suited risk, but on local evaluation of contamination in
than bacterial sampling. It is less demanding, empty operating theatres. When surgical
results are immediate, and it can be used for on- techniques
site teaching. The interval

Table I Limits of air particle contents according to standardized norms (adapted from FS209 E◆)

Superior limits in measured particle size (particle per volume unit) (equal to, or greater than stated size)

Class †
0.l µm volume unit 0.2 µm volume unit 0.3 µm volume unit 0.5 µm volume unit 5 µm volume unit

SI FS 209 m3 ft3 m3 ft3 m3 ft3 m3 ft3 m3 ft3

Ml 350 9.9l 75.7 2.l4 30.9 0.875 l0 0.283 – –


Ml.5 l l240 35.0 265 7.50 l06 3.00 35.3 l.0 – –
M2 3500 99.l 757 2l.4 309 8.75 l00 2.83 – –
M2.5 l0 l2 400 350 2650 75.0 l060 30.0 353 l0.0 – –
M3 35 000 99l 7570 2l4 3090 87.5 l000 28.3 – –
M3.5 l00 26 500 750 l0 600 300 3530 l00 – –
M4 75 700 2l40 30 900 875 l0 000 283 – –
M4.5 l000 35 300 l000 247 7.0
M5 l 00 000 2830 6l8 l7.5
M5.5 l0 000 3 53 000 l0 000 2470 70
M6 l0 00 000 28 300 6l80 l75
M6.5 l 00 000 35 30 000 l 00 000 24 700 700
M7 l 00 00 2 83 000 6l 800 l750
000
◆ Reference 29.

Refers to the engineering classification of cleanrooms. SI: International system (metric); column l is the metric equivalent
of the US standard FS209 (column 2).
82 S. Dharan and D.
Pittet

Table II Limits of air particle contents according to different international standards

(a) British Standard 5295◆: classification and requirements of cleanrooms

Superior limits in measured particle size per m 3 (equal to, or greater than stated size)

Class† 0.3 µm 0.5 µm 5 µm l0 µm 25 µm

C l00 35 0 NS NS
D l000 350 0 NS NS
E l0 000 3500 0 NS NS
F NS 3500 0 NS NS
G l 00 000 35 000 200 0 NS
H NS 35 000 200 0 NS
J NS 350 000 2000 450 0
K NS 3 500 000 20 000 4500 500
L NS NS 2 00 000 45 000 5000
M NS NS NS 4 50 000 50 000

NS = Not specified.
◆ Reference 30.

Refers to the engineering classification of cleanrooms.

(b) German VDI 2083◆: classification and requirements of cleanrooms

Superior limits in measured particle size per m3 (equal to, or greater than stated size)

Class† 0.l µm 0.2 µm 0.3 µm 0.5 µm l.0 µm: 5 µm l0 µm

0§ l50 33 l4 – l0 0

l l500 330 l40 45 l0l


2 l5 000 3300 l400 450 l02
3 – 33 000 l4 000 4500 l03
4 – – – 45 000 l04 300
5 – – – 4 50 000 l05 3000
6 – – – 45 00 000 l06 30 000
7 – – – – l07 30 000 70
000
◆ Reference 3l.

Refers to the engineering classification of cleanrooms.
:
Size chosen for classification.
§
This class should be considered when cleanroom is unoccupied.

(c) International norm (ISO I4644–I◆: cleanrooms and associated controlled environments, classification and requirements)

Superior limits in measured particle size per m3 (equal to, or greater than, stated size)

Class† ISO 0.l µm 0.2 µm 0.3 µm 0.5 µm l.0 µm 5 µm

l l0 2
2 l00 24 l0 4
3 l000 237 l02 35 8
4 l0 000 2370 l020 352 83
5 l00 000 23 700 l0 200 3520 832 29
6 l0 00 000 2 37 000 l 02 000 35 200 8320 293
7 3 52 000 83 200 2930
8 35 20 000 8 32 000 29 300
9 3 52 00 000 83 20 000 2 93 000

◆ Reference 32.

Refers to the engineering classification of cleanrooms.
Environmental controls in operating 83
theatres

Table III Classification of operating theatre zones according to risk


experience and the increasing use of alcohol-
categories and limits in airborne particles and bacteria◆, University of
Geneva Hospitals, Switzerland based solutions for surgical hand disinfection and
double- gloving, we strongly believe that the use of
Particle sterile water for surgical hand disinfection is not a
size (Particles per m3) Bacterial
counts necessity.
Class Level of risk ≤0.5 µm ≤ 5 µm (cfu/m3)

Conclusion
l. Very high risk, within l0 0 cl
laminar air-flow area There is presently no common standard method
with HEPA filtration used to determine bacterial threshold limits in the
2. High risk, exterior 353 l0 5
laminar air-flow,
operating theatre setting. In this situation, it is
but within the practically impossible to compare results between
operating theatre countries. We recommend the use of cleanroom
3. Medium risk, 3 530 25 25 technology standards based on the measurement
conventionally of the presence of air particles as routine
ventilated operating
theatre with air
procedure. Our opinion is that environmental
filtered through bacterial sampling should be restricted to the
terminal filters with an investigation of epidemics and validation of
efficiency changes in products and main- tenance procedure
of 95% and above NS NS NS of the operating theatre. We suggest that there is a
4. Low risk, areas with
uncontrolled ventilation
need to reach a consensus on the method of
evaluation and also, to differentiate
◆ Sampling is performed in the operating theatre between cleanroom technology developed for
whilst at rest. cfu, Colony forming unit; NS, no specific industry and that used in hospitals.
limit.

Achnowledgement
or theatre dress are to be evaluated, both volu-
metric air sampling and settle plates may be used The authors thank Rosemary Sudan for editorial
together. Plates should be placed as near as assistance.
possible to the operating team and on the
instrument trolley in order to obtain any meaningful
results. How- ever, this may create practical References
problems in some circumstances. 1. Bitkover CY, Marcusson E, Ransjo U. Spread of
coagulase-negative staphylococci during cardiac
operations in a modern operating room. Æww
Water supply Thorac Gurg 2000; ð9: 1110–1115.
2. Whyte W, Hambraeus A, Laurell G, Hoborn J. The
Another area of environmental control in operating relative importance of routes and sources of wound
theatres is the bacteriological quality of water used contamination during general surgery. I. Non-
for surgical handscrubs for which there are no stan- airborne. ƒ Hosp Iwfect 1991; 18: 9S–10U.
dardized limits at present. Some institutions use S. Whyte W, Hambraeus A, Laurell G, Hoborn J. The
sterile water for this procedure. At our institution, relative importance of routes and sources of wound
contamination during general surgery. II. Airborne.
we use drinking quality water, c200 bacteria/mL, ƒ Hosp Iwfect 1992; SS: 41–54.
with bacterial counts carried out twice a year on 4. Whyte W, Hodgson R, Tinkler J. The importance of
these water outlets. Counts have never exceeded airborne contamination of wounds. ƒ Hosp Iwfect
400 bacteria/mL with most being below 100 1982; 3: 12S–1S5.
(unpublished data). The exception to this was when 5. Ayliffe GAJ. Role of the environment of the operating
suite in surgical wound infection. Rev Iwfect Dis
a new operating theatre was built and warm water 1991; 13 (Suppl. 10): S800–804.
(S0–S5○C) was provided via a closed circuit for the 6. Humphreys H. Infection control team in the operating
comfort of surgeons. However, the problem was room: separating aspiration from reality! ƒ Hosp
rapidly rectified by lowering the temperature to 20– Iwfect 1999; 4S: 265–26U.
25○C. Organisms recovered from these samples U. Charnley J, Eftekhar N. Postoperative infection in
total prosthetic replacement arthroplasty of hip joint.
were mainly environmental, Gram-negative bacilli Br ƒ Gurg 1969; 5ð: 641–649.
and atypical mycobacteria. Taking into account our
84 S. Dharan and D.
Pittet

8. Lidwell OM, Lowbury EJL, Whyte W, Blowers R, counts using disposable head and mask and
Stanley SJ, Lowe D. The effect of ultraclean air in helmet aspirator system, but dramatic increase
operating rooms on deep sepsis in the joint after hip upon omission of head-gear: an experimental study
or knee replacement: a randomised study. Br Med ƒ in horizontal laminar air-flow. ƒ Hosp Iwfect 2001;
1982; S85: 10–14. 47: 110–115.
9. Gosden PE, MacGowan AP, Bannister GC. Impor- 24. Mitchell NJ, Hunt S. Surgical face masks in modern
tance of air quality and related factors in the operating rooms—a costly and unnecessary ritual?
prevention of infection in orthopaedic implant ƒ Hosp Iwfect 1991; 18: 2S9–242.
surgery. ƒ Hosp Iwfect 1998; 39: 1US–180. 25. Military Standard 282. Military stawdard: filter uwits,
10. Friberg B, Friberg S, Burman LG. Correlation protective clothiwg, gas-mas§ compowewts awd
between surface and air counts of particles carrying related productiows: performawce-test methods. United
aerobic bacteria in operating rooms with turbulent States Department of Defence, 1956.
ventilation: an experimental study. ƒ Hosp Iwfect 26. Whyte W, Lidwell OM, Lowbury EJ, Blower R.
1999; 4S: 61–68. Suggested bacteriological standards for air in
11. Laufman H. Airflow effects in surgery. Ærch Gurg ultraclean operating rooms. ƒ Hosp Iwfect 198S; 4:
19U9; 114: 826–8S0. 1SS–1S9.
12. Schwan A, Bengstsson S, Hambraeus A, Laurell G. 2U. Friberg BE, Friberg S, Burman LG. Zoned vertical
Airborne contamination and post operative infection ultraclean operating room ventilation. A novel
after total hip replacement. Æcta Orthop Gcawd concept making long side walls unnecessary. Æcta
19UU; 48: 86–94. Orthop Gcawd 1996; ð: 5U8–582.
1S. Vasey H, Ducel G, Allgower D. Zones prote´ge´es en 28. Johnston IDA, Hunter AR (Eds). The Desigw awd
salle d*ope´ration. Evaluation d*une nouvelle enceinte Utilisatiow of Operatiow Theatres. London: Edward
de Charnley. Med Hyg 198U; 45; 1646–1651. Arnold 1984.
14. Lew DP, Pittet D, Waldvogel FA. Infections that 29. Federal Standard 209E. Æirborwe particulate
complicate the insertion of prosthetic devices. In: cleawliwess classes iw cleaw rooms awd cleaw sowes.
Mayhall CG, Ed. Hospital Epidemiology awd United States General Service Administration 1992.
Iwfectiow Cowtrol. Baltimore: Williams and Wilkins S0. British Standard 5295-1. Classificatiow awd require-
1995; US1–U48. mewts of cleawrooms. British Standards Institution
15. Mitchell NJ, Evans DS, Kerr A. Reduction of skin 1989.
bacteria in theatre air with comfortable, non-woven S1. Vereinigte Deutsche Ingenieure (VDI) 208S.
disposable clothing for operating theatre staff. Br Reiwraumtechwi§—Gruwdlagew, Defiwitiowew uwd
Med ƒ 19U8; 1: 696–698. Festleguwgew der Reiwheits§lassew 1995.
16. Fitzgerald RH, Jr. Microbiological environment of the S2. ISO 14644-1. Cleawrooms awd associated
conventional operating room. Ærch Gurg 19U9; 114: cowtrolled ewvirowmewts, Part l: Classificatiow of air
UU2–UU5. cleawliwess. International Organization for
1U. Ritter MA, Eitzen HE, Hart JB, French ML. The Standardization Technical Committee (ISO TC 209)
surgeon*s garb. Cliw Orthop 1980; 153: 204–209. 1999.
18. Berman BR, Hoborn J, Nachemson AL. Patient SS. Whyte W. In support of settle plates. PDÆ ƒ Pharm
draping and staff clothing in the operating theatre: Gci Techwol 1996; 50: 201–204.
a microbiological study. Gcawd ƒ Iwfect Dis 1985; S4. Gasper C, Calvente MJ, Fernandez C, Fereres J. Air
17: 421–426. microbiology monitoring in plenum-ventilated oper-
19. Hubble MJ, Weale AE, Perez JV, Bowker KE, ating rooms. Proposed standards. Ewferm Iwfecc
MacGowan AP, Bannister GC. Clothing in laminar- Microbiol Cliw 199U; 15: 250–254.
flow operating theatres. ƒ Hosp Iwfect 1996; 3S: 1– S5. Friberg B, Friberg S, Burman LG. Inconsistent
U. correlation between aerobic bacterial surface and air
20. Chosky SA, Modha D, Taylor GJ. Optimisation of counts in operating rooms with ultra clean laminar
ultraclean air. The role of instrument preparation. air-flows: proposal of a new bacteriological standard
ƒ Bowe ƒoiwt Gurg Br 1996; 78: 8S5–8SU. for surface contamination. ƒ Hosp Iwfect 1999; 4S:
21. Verkkala K, Eklund A, Ojajarvi J, Tiittanen L, 28U–29S.
Hoborn J, Makela P. The conventionally ventilated S6. Pasquarella C, Pitzurra O, Savino A. The index of
operating theatre and air contamination control microbial air contamination. ƒ Hosp Iwfect 2000; 4ð:
during cardiac surgery—bacteriological and 241–256.
particulate matter control garment options for low SU. Seal DV, Clark RP. Electronic particle counting for
level con- tamination. Eur ƒ Cardiothor Gurg 1998; evaluating the quality of air in operating theatres:
14: 206–210. a potential basis for standards. ƒ Æppl Bacteriol
22. Pittet D, Ducel G. Infectious risk factors related to 1990; ð8: 225–2S0.
operating rooms. Iwfect Cowtrol Hosp Epidemiol S8. Whyte W, Shaw BH, Barnes R. A bacteriological
1994; 15: 456–462. evaluation of laminar-flow systems for orthopaedic
2S. Friberg B, Friberg S, O¨ stensson R, Burman surgery. ƒ Hyg Camb 19US; 71: 559–564.
LG. Surgical area contamination—comparable
bacterial

You might also like