Professional Documents
Culture Documents
TruControl 1000
Software manual
TruControl 1000
Edition 2014-10
Order Information Please specify when ordering this document:
Software manual
TruControl 1000
Edition 2014-10
Document number 10-06-80-A1-CR02
About the content of the The present software manual describes the operation and pro-
software manual gramming of laser devices using TruControl 1000. It is aimed at
users, programmers and maintenance personnel of the laser
devices by TRUMPF.
6 Symbols 1‐25
7 Abbreviations 1‐26
Chapter 3 Setup
Chapter 4 Programming
Chapter 5 Diagnosis
5 Messages 5‐18
5.1 Displaying messages 5‐19
5.2 Displaying the history of messages 5‐20
5.3 Displaying a message text 5‐21
5.4 Resetting messages 5‐22
5.5 Setting the message filter 5‐22
5.6 Searching for a message 5‐23
5.7 Adding a note 5‐23
9 Statistics 5‐29
9.1 Utilization of the laser 5‐29
Display the statistics of the MPI slots 5‐30
Showing the distribution over all MPI slots 5‐31
Resetting the counters for all MPI slots 5‐31
9.2 Use of the light paths 5‐32
Display of the light paths statistics 5‐33
Resetting the counters for all light paths 5‐33
Chapter 6 Maintenance
1 Overview 6‐2
Chapter 8 Service
2 Faults 8‐4
2 License 9‐3
2.1 Information on using the software licenses 9‐3
2.2 Display licenses 9‐4
Fundamentals of operation
6 Symbols 1‐25
7 Abbreviations 1‐26
EMERGENCY STOP The laser and the cooling unit are immediately switched off when
pushbutton the pushbutton EMERGENCY STOP (2) is pressed. In particular,
no more laser light is emitted.
The EMERGENCY STOP pushbutton can be integrated in the
Emergency Stop circuit of a system.
Press this pushbutton to cut off the entire system. More informa-
tion can be found in the interface description "EMERGENCY
STOP".
CONTROL ON button With the CONTROL ON (3) button, the control unit is switched
on.
CONTROL OFF button With the CONTROL OFF (4) button, the control unit is switched
off.
On a table
AW-00003
Fig. 1-2
Swing out the bracket on the rear side of the operating panel
and put the operating panel on the table.
Fig. 1-3
Insert the bracket on the rear side of the operating panel into the
guide on the laser device.
Fig. 1-4
Swing out the bracket on the rear side of the operating panel
and insert it into the guide on the laser device.
Fig. 1-5
AW-00001EN
Main screen Fig. 1-6
Tip
For the lasers of the TruDisk and TruFiber families, a dialog box
showing a diagram with the current laser power can be displayed
on the main screen.
Information regarding the activation key of this function can be
obtained from TRUMPF on request.
Main operations button By pressing the Main operations (1) button, a main operation is
activated. The Main operations button displays the symbol of the
main operation selected currently. Each main operation has sub-
ordinate secondary operations (functions).
Main and secondary operations correspond to menus and sub-
menus, as known, for example in Windows programs. The sec-
ondary operations are displayed in horizontal and vertical tabs as
well as in selection lists.
The section Main and secondary operations (see "Main and sec-
ondary activities", pg. 1‐22) shows an overview of all main and
secondary operations and the corresponding user levels.
Status line The status line (2) shows symbols indicating the current settings
of the laser device.
Select the status line to display a window. It contains all symbols
currently active and their meaning.
The Laser status tab shows the active status symbols and their
meaning.
Title bar The message line (3) displays the message code and the short
text of the message.
Select the message line to open the extended status dialog box.
The current message is displayed in the Message tab.
AW-00007EN
Symbol Meaning
Fault message
Tab. 1-2
User level There are seven different main user levels for the user. The
passwords for setting the user levels are available from
TRUMPF.
Press the "User level" symbol (4) to open the extended status
dialog box.
The User level tab displays the set user level which can be
changed there as well (see "Changing the user level", pg.
3‐5).
Equipment number The equipment number (5) is the unique number of the laser
device (laser device number).
Laser status The "Laser status" symbol (6) indicates the status of the laser.
Symbol Meaning
Laser is on.
Laser is off.
Tab. 1-3
TRUMPF logo Double-click the TRUMPF logo (7) to open the TruControl 1000
start dialog box.
Context bar The context bar (8) contains buttons used to control TruControl
1000. Depending on the set user level, the buttons are active or
inactive.
Button is active.
Button is inactive.
Tab. 1-4
Tab. 1-5
The user interface has different display fields with different back-
ground and font colors depending on their function:
AW-00008EN
Secondary activities in a selection list Fig. 1-10
User level ≥ 0
Select the selection arrow.
Open the selection menu with the secondary activities.
Tab. 1-8
Tabs
AW-00009EN
Graphical overviews
User level ≥ 0
Select Hand.
The dialog box with additional information is displayed.
Tab. 1-9
1 2
RP-00541
1 Turning 2 Pressing
Fig. 1-14
Input fields for e.g. power and pulse duration can be precisely
set using the turn-press button. By pressing the turn-press button
in an input field, the scroll bar opens. The position of the scroll
bar is set by turning the turn-press button to the right or to the
left. The value is entered in the input field when the turn-press
button is pressed.
AW-00013EN
Soft keyboard with buttons for letters, numbers and the following
functions:
Tab. 1-10
AW-00014EN
Letter magnifying glass Fig. 1-19
User level ≥ 70
Expert dialog boxes
Tab. 1-11
Note
If in the main activity Setup the option Display TRUMPF service
information is selected, the expert dialog boxes will be displayed
from user level ≥ 0.
≥ 40 Laser program ≥ 40 -
Pulse shape ≥ 40 -
PFO program ≥ 40 if PFO provided
CFO data ≥ 40 if CFO provided
≥ 60 Light path ≥ 60 -
Cutting ≥ 40 if CutAssist modules provided
Cutting of small contour ≥ 40 if CutAssist modules provided
Piercing ≥ 40 if CutAssist modules provided
≥ 0 Versions ≥ 0 -
License ≥ 0 -
Documentation ≥ 0 -
Request service ≥ 30 -
≥ 30 Maintenance ≥ 30 -
Lamp factor optimization ≥ 30 TruPulse
(LFO)
Pulse preselection coun- ≥ 50 for TruPulse, TruMicro
ter
Emptying the cooling cir- ≥ 30 only if cooling function is sup-
cuit ported
PPC: power measure- ≥ 30 if CFO provided
ment
PPC: energy measure- ≥ 30 if CFO provided
ment
Pump switch ≥ 30 if CMD board provided
Main operations and functions with corresponding user levels Tab. 1-12
All symbols and their meaning are listed in the Status symbols
tab of the main operation Help.
User level ≥ 0
1. Select the main activity Help.
Note
A window is displayed by selecting the status line. It contains all
symbols currently active and their meaning (see "Layout of the
user interface", pg. 1‐10).
User level ≥ 0
1. Select the main activity Help.
Data backup
Tip
For the data transmission, the USB port of the control unit (CPX
and OP84) can also be used for TruControl Remote in case of a
local instance.
This is particularly useful if the USB ports of a Windows PC are
locked
Note
The PFO programs can be transferred from another device type.
Prior to using PFO programs they have to be opened and stored
in the PFO editor.
User level ≥ 40
1. Select the main activity Setup.
6. Select Type.
Before importing/exporting PFO or CFO programs, the
number of the focusing optics must be additionally
selected.
7. Select Next.
The Adjust program data: laser programs dialog box is
opened.
8. Tick the required data.
9. Select Export.
or
Select Import.
The selected data is displayed on the right side of the
dialog box.
10. Select Close.
The data has been imported/exported.
Tab. 2-1
Export and import of this general system data guarantee the fol-
lowing:
■ A backup is available at any time, e.g. to restore the system.
■ The information of the backup file can be used for the diag-
nostics of the laser device.
Note
For this purpose, the system data is saved as a Zip file. The
system data can be sent by e-mail to TRUMPF, if necessary.
This data include all the user's settings in connection with the
laser tasks on site, e.g. the task-specific laser programs, pulse
data.
User level ≥ 60
1. Select the main activity Setup.
3. Select parameter.
5. Select option.
6. Select the directory.
7. Select Next.
The system settings are exported.
For the import, the Import system settings dialog box is
displayed.
8. Select the file with the desired system settings for the
import.
9. The system settings are imported.
User level ≥ 0
1. Connect the USB stick to the USB interface of the panel
computer, the Windows PC or the operating panel.
The installation wizard is displayed on the user interface.
2. Select Start .
The software update is installed. After the installation
has been finished the installation wizard is closed auto-
matically.
Tab. 2-4
User level ≥ 0
1. Store the data of the CD-ROM on the hard disk.
2. Execute the installation file (TLCpxUpdate.exe).
The installation wizard is opened.
3. Select the laser device in a list.
4. Select Start .
The software update is installed. After the installation
has been finished the installation wizard is closed auto-
matically.
Tab. 2-5
TRUMPF provides you with a link that allows you to save the
desired version on your hard disk or an USB stick via the
TRUMPF download server. If requested, you get a CD-ROM
containing the data.
Setup
AW-00022EN
1 Number range 3 TRUMPF service information 5 Language
2 Context bar 4 User level
Fig. 3-1
Number range If several operating panels are connected to the control unit of
the laser device, the access to programs, pulse shaes and light
paths are organized by means of the number range function (1).
Only the programs or light paths within the set number range
can be edited using the current operating panel.
Context bar The context bar (2) contains buttons available in the secondary
activity Program setting.
TRUMPF service information If the option "Display TRUMPF service information" (3) is acti-
vated, the Expert dialog boxes and LCU component list are dis-
played with software versions independently of the user level.
User level Different user levels (4) can be set in TruControl 1000. Each
user level implies different rights for the user. In order to set a
user level with a higher authorization (bigger number), a pass-
word is necessary. The number of the set user level can be
seen in the status line of TruControl 1000.
The respective passwords are available from TRUMPF.
Language Different languages can be set using the selection field Lan-
guage (5) .
User level ≥ 0
1. Select the main activity Setup.
User level ≥ 0
1. Select the main activity Setup.
Tab. 3-3
User level ≥ 0
1. Select the main activity Setup.
Tab. 3-4
Note
The system parameters are set by TRUMPF. Almost all parame-
ters have a fundamental effect on the laser operation. Therefore
consult the after-sales service of TRUMPF for modifying the sys-
tem parameters.
4 1
3
AW-00023EN
Check box If the check box "Display only adjustable values" (1) is selected,
the list displays only the parameters that can be changed in the
set user level.
Parameter list The Parameter list (2) shows the parameters that can be
changed at the set user level.
Context bar The context bar (3) contains buttons available in the secondary
activity System setting.
Prerequisite:
■ Since almost all parameters basically interfere with the laser
operation, TRUMPF has been contacted before carrying out
any modification.
User level ≥ 10
1. Select the main activity Setup.
Tab. 3-5
Note
The TRUMPF peripheral bus is an optional system component of
TRUMPF LaserNetwork (TLN). It is based on real-time Ethernet
(see "Setting the network configuration of the Ethernet interface
(TRUMPF peripheral bus)", pg. 3‐28).
For more information on the TRUMPF Peripheral Bus (TPB), see
the interface description.
Table of MPI settings The MPI settings table (1) contains the following information:
Context bar The context bar (2) contains buttons available in the secondary
activity Multi PLC interface (MPI).
Note
The configuration of an assigned interface is described in the
corresponding interface description.
User level ≥ 60
1. Select the main activity Setup.
Tab. 3-7
User level ≥ 60
1. Select the main activity Setup.
Tab. 3-8
User level ≥ 60
1. Select the main activity Setup.
9. Select Save .
The parameter has been changed.
Tab. 3-9
User level ≥ 60
1. Select the main activity Setup.
Tab. 3-10
User level ≥ 60
1. Select the main activity Setup.
5. Select Yes.
The slot is deleted after the laser control has been
restarted.
Tab. 3-11
Fig. 3-4
All plug & play components are displayed in the function Plug &
Play . The components whose configuration has been changes
are marked by the Plug & Play symbol.
AW-00025EN
1 Expand / reduce 4 Plug & Play symbol 7 Serial no.
2 Configuration status 5 ID-NUM 8 Context bar
3 Action 6 Software version
Fig. 3-5
Expand / reduce The column Expand / Reduce (1) is used to expand the table to
the secondary components by pressing "+" and reduce to the
main components by pressing "-".
Configuration status The Configuration status column (2) indicates with the "check
mark" symbol that the component is operated with the current
configuration. If the "Plug & Play" symbol is displayed in that col-
umn, this means that the configuration has been changed and
that an adjustment must be carried out.
Action The Action column (3) shows the state of a component and may
contain the following information:
■ New component in the system (a new board or focusing
optics has been installed in the system).
■ Component changed (a board or focusing optics has been
replaced by a new one).
■ No Plug & Play support (the component has no Plug & Play
support).
■ Configuration changed (the setting data of a board has
changed).
Plug & Play symbol The fault message 0000 1490 in the message line indicates the
change in configuration of a component by the short text
"Changed control configuration". The "Plug & Play" symbol (4) is
displayed in the status line.
Software version The Software version column (6) displays the currently used soft-
ware of the component.
Serial no. The Serial no. column (7) displays the serial number of the
board.
Context bar The context bar (8) contains buttons available via the function
(secondary operation) Plug & Play overview.
User level ≥ 30
1. Select the main activity Setup.
7. Select Finish.
The adjustment is carried out.
Tab. 3-13
User level ≥ 30
1. Select the main activity Setup.
Tab. 3-14
Fig. 3-6
Prerequisites:
■ The laser device must be equipped with a central shutter
(VZ).
■ The safety circuit is open.
User level ≥ 60
1. Select the main activity Setup.
3. Select Activate.
or
Select Deactivate.
The simulation mode is activated when the indicator
lamp "Active" in the left column lights up in green. The
simulation mode is deactivated if the indicator lamp
remains gray.
The simulation mode can also be activated via the PLC.
In the simulation mode dialog box both indicator lamps
at the respective slot (workstation) light up green.
Tab. 3-15
Word number There are up to 61 input words and 62 output words. They are
provided with a word number, which is given in the Word no. (2)
column. Depending on the laser type TRUMPF has already
assigned variables to some of the output words.
Parameters The Parameters (3) column shows the names of the parameters.
Output word type The type of the output word (4) uses an abbreviation to identify,
whether the measured value preset by TRUMPF is assigned to
the measured value or whether the user has set another varia-
ble.
Context bar The context bar (5) contains buttons of the functions, which are
available in the secondary operation Interface parameters.
User level ≥ 60
1. Select the main activity Setup.
6. Select Apply.
The parameter has been changed.
Tab. 3-17
User level ≥ 60
1. Select the main activity Setup.
3. Select Calibrate.
The TRUMPF Logo is displayed with cross hairs.
Mark the center of the cross hairs with the finger four
times.
The touch screen is calibrated. The calibration becomes
active after a restart.
Tab. 3-18
User level ≥ 60
1. Select the main activity Setup.
Tab. 3-19
User level ≥ 60
1. Select the main activity Setup.
Tab. 3-20
AW-00027EN
1 Date setting 3 Time zone setting 4 Context bar
2 Time setting
Fig. 3-8
Prerequisite:
■ The time server is deactivated (see "Setting the time server
parameters", pg. 3‐26).
User level ≥ 60
1. Select the main activity Setup.
2. Select Date/time.
4. Select Save .
The time has been set.
Tab. 3-21
Prerequisite:
■ The time server is deactivated (see "Setting the time server
parameters", pg. 3‐26).
User level ≥ 60
1. Select the main activity Setup.
2. Select Date/time.
Tab. 3-22
User level ≥ 60
1. Select the main activity Setup.
2. Select Date/time.
Tab. 3-23
User level ≥ 60
1. Select the main activity Setup.
2. Select Date/time.
Tab. 3-24
Tab. 3-25
User level ≥ 50
1. Select the main activity Setup.
2. Select Network.
Tab. 3-26
Note
The Undo button restores the previously set configuration.
User level ≥ 60
1. Select the main activity Setup.
2. Select Firewall.
3. Select New.
4. Select IP address.
or
Select Network.
or
Select IP address range.
5. Activate the firewall for the following address.
or
Deactivate the firewall for the following address.
6. Enter the address(es).
7. Select Apply.
The network connection is set up. The setting is dis-
played in the list "Enabled connections".
Tab. 3-27
10-06-80-A1-CR02 2014-10 Firewall mode of the laser device in the data 3‐29
network
10.2 Changing the network connection
User level ≥ 60
1. Select the main activity Setup.
2. Select Firewall.
Tab. 3-28
User level ≥ 60
1. Select the main activity Setup.
2. Select Firewall.
5. Select Yes.
The network connection has been deleted.
Tab. 3-29
3‐30 Firewall mode of the laser device in the data 2014-10 10-06-80-A1-CR02
network
11. Switching functions
Prerequisite:
■ The PilotLaserMode must be set in the system settings (see
chapter "Diagnosis").
■ The shutter switch (VW) must be closed.
■ The safety circuit (SIK) must be closed.
■ The operating unit is not locked (signal Lock operating unit
not active).
User level ≥ 20
1. Select the main activity Setup.
User level ≥ 60
1. Select the main activity Setup.
User level ≥ 60
1. Select the main activity Setup.
Tab. 3-32
The control unit and the power supply of the laser device are
switched off and restarted by means of this function
Tab. 3-33
Note
To send e-mails, the laser device uses the SMTP protocol (engl.
Simple Mail Transfer Protocol). The e-mails are sent by the laser
device to port 25 of the transfer server.
The following filters are available in the Create e-mail dialog box:
■ Maintenance messages
■ Fault messages
■ Monitoring messages
■ Logbook entries
User level ≥ 60
1. Select the main activity Setup.
User level ≥ 60
1. Select the main activity Setup.
Tab. 3-36
User level ≥ 60
1. Select the main activity Setup.
5. Select Yes.
The selected e-mail address is deleted.
Tab. 3-37
The Transfer server tab displays the access data of the transfer
server.
■ E-mail transfer server:
IP address name of the e-mail transfer server.
■ Sender:
E-mail address of the sender.
■ Users
User name to access the e-mail transfer server.
■ Password:
The symbols •••• mean that a password has been saved for
the e-mail transfer server access.
Tab. 3-38
Note
This dialog box is only available with pulsed laser devices.
The pulse counter shows the number of pulses that were trig-
gered by the laser. A monitoring message is displayed when the
limit value set is exceeded. This indicates, that the flash lamps
have to be replaced.
For information on how to set the optimum limit value for your
application, please contact TRUMPF.
User level ≥ 50
1. Select the main activity Maintenance & Commissioning.
3. Enter the desired value for the pulse number in the field
Select limit value.
The limit value is set. A monitoring message requesting
the replacement of the lamp is displayed as soon as the
limit value is exceeded.
Tab. 3-39
User level ≥ 50
1. Select the main activity Maintenance & Commissioning.
User level ≥ 50
1. Select the main activity Maintenance & Commissioning.
3. Select Reset.
The pulse counter has been reset to zero.
Tab. 3-41
Prerequisite
■ For switching the laser on or for activation of the optimized
standby mode, the Power Board Tiny (PBT) must be availa-
ble.
User level ≥ 30
1. Select the main activity Setup.
4. Select Time.
5. Select Action.
Note
By pressing Reset all entries which have not yet been saved are
deleted from the list.
User level ≥ 30
1. Select the main activity Setup.
Note
Press Reset to reset all modifications to the last set time sched-
ules.
User level ≥ 30
1. Select the main activity Setup.
Note
Press Reset to reset all modifications in the list.
List of options The Manage options function shows all possible options for the
laser device and its status.
Option coding For all active options the Option coding (2) is displayed. This
hexadecimal number serves for easy transfer of the options.
Context bar The context bar (3) contains buttons, which are available in the
Manage options function.
Note
To receive an activation key for an option or to get more infor-
mation on the options, contact the After-sales department of
TRUMPF.
User level ≥ 60
1. Select the main activity Setup.
Note
Options which belong to the standard scope of delivery cannot
be deactivated.
User level ≥ 60
1. Select the main activity Setup.
4. Select Yes.
The selected option is deactivated.
In the Active column, the option is now displayed inac-
tive without a checkmark.
Tab. 3-48
Programming
Synchronization with External machine cycles can be synchronized during the execu-
external machine cycles tion of laser programs.
3
AW-00028EN
Program no. and program The program number and the name (1) of the laser program are
name above the program table.
Program table All parameters that can be specified in a program line are
entered into the program table (2)(see "Laser parameters", pg.
4‐12).
Context bar The context bar (3) contains buttons available in the Create laser
program dialog box.
User level ≥ 40
1. Select the main activity Programming.
Tab. 4-1
User level ≥ 40
1. Select the main activity Programming.
Tab. 4-2
User level ≥ 40
1. Select the main activity Programming.
7. Select Apply.
The line change has been set.
Tab. 4-3
User level ≥ 40
1. Select the main activity Programming.
8. Select Apply.
The maximum time monitoring is activated. The desired
maximum time is set.
Tab. 4-4
User level ≥ 40
1. Select the main activity Programming.
8. Select Apply.
Except when None has been selected in step 6, the
control of the pulse counter is now activated.
The desired number of pulses is set.
Tab. 4-5
User level ≥ 40
1. Select the main activity Programming.
9. Select Yes.
The line has been deleted.
Tab. 4-7
The laser programs in the Laser program tab are sorted by num-
bers. A laser program can be copied from one number to
another.
User level ≥ 40
1. Select the main activity Programming.
Tab. 4-8
Laser functions
The laser control unit can be to other devices, e.g. other control
units, through interfaces. In this way, laser programs can be
Wait functions The laser control runs the next program row only if the selected
query input of the specified signal level is activated.
Set functions When the program line is started, the selected user output is
placed on the specified signal level.
Delay functions The laser control executes this program line with a delay.
Loop functions The laser control repeats one or several program lines within a
loop. One to endless repetitions are possible.
Real-time start A real-time start can be triggered externally via the central pro-
cessor CPX.
Analog multiplier function The nominal value for the laser power can be multiplied by
means of this function with an analog input signal.
Measuring function The control unit records the average energy of the laser for proc-
ess monitoring.
User level ≥ 40
1. Select the main activity Programming.
7. Select Apply.
The modifications are displayed in the Create laser pro-
gram dialog box.
8. Select Save .
The changes are saved.
Tab. 4-19
Tab. 4-21
Prerequisite:
■ A continuous wave or rectangle is set for Pulse type.
User level ≥ 40
1. Select the main activity Programming.
Tab. 4-22
scalable
Template
Tab. 4-23
Static pulse shape The power course and the duration of static pulse shapes is
always as programmed in the pulse shape editor. For modifying
them they must be edited again in the pulse shape editor. Static
pulse shapes are programmed with absolute values.
Scalable pulse shape Scalable pulse shapes are created in the pulse shape editor, like
the static pulse shapes. They get the property "scalable". When
embedding them in parameter sets or laser programs the power
course and the duration can be linked with absolute values for
power and duration. This allows, e.g. a pulse shape to be
reduced to half of the original height (power) or to increase its
duration to the double value. Some marginal conditions like pulse
P
L100
Lmin
RP-00204
TPUmin TPUmax
t
Pulse shape as a reference Pulse shapes embedded in a parameter set or laser program
using a link, are saved as "global pulse shapes". Changing a
global pulse shape effects all places where the pulse shape is
used.
Pulse shape as a copy Pulse shapes embedded in a parameter set or laser program as
a copy, are saved as "local pulse shapes". When embedding
them in a laser program a copy is created that is transferred as
Pulse shape as a template Existing pulse shapes can be saved as templates. If a pulse
shape is stored as a template, the original pulse shape is kept
unchanged when the template is modified (see "Creating a pulse
shape as template", pg. 4‐27).
1 Title bar 3 Option to change the time differ- 5 Diagram pulse shape
2 Table pulse data ence
4 Context bar
Pulse shape editor Fig. 4-3
Pulse data table The pulse data table (2) contains the points from power [P] and
time [t], where a certain laser power is to be generated. The
points are shown in the diagram. Adjacent points are connected
with straight lines.
Time difference If the checkmark for the time difference option (3) is set, the
change of a time difference shifts automatically all other interpo-
lation points. Their time differences, however, remain preserved.
Context bar The context bar (4) contains buttons available in the Pulse shape
editor dialog box.
Pulse shape diagram In the pulse shape diagram (5) the power course over the dura-
tion of the pulse shape is displayed.
User level ≥ 40
1. Select the main activity Programming.
8. Select Finish.
The Pulse shape editor is displayed. The individual
points of the pulse shape can be entered now. The first
two lines are already defined.
9. Select the upper line.
The line is highlighted in yellow.
Tab. 4-25
13. Define the next point in the same way. The point is
connected to the adjacent point in the diagram.
14. If required, set further points:
Insert line must be selected.
The new line is displayed in the pulse data table.
15. Define all the required points in this way.
16. Select Change name.
The Enter name dialog box is displayed.
Tab. 4-26
User level ≥ 40
1. Select the main activity Programming.
Tab. 4-27
Tab. 4-28
User level ≥ 40
1. Select the main activity Programming.
Tab. 4-29
User level ≥ 40
1. Select the main activity Programming.
5. Select Copy.
6. Select Next.
The Select memory location dialog box is displayed.
Tab. 4-30
Tab. 4-31
User level ≥ 40
1. Select the main activity Programming.
5. Select Move.
6. Select Next.
The Select memory location dialog box is displayed.
Tab. 4-32
User level ≥ 40
1. Select the main activity Programming.
5. Select Yes.
The pulse shape has been deleted.
Tab. 4-33
User level ≥ 40
1. Select the main activity Programming.
Tab. 4-34
Tab. 4-35
User level ≥ 40
1. Select the main activity Programming.
7. Select Yes.
The pulse shape type has been changed.
Tab. 4-36
The magnifying glass on the right side of the pulse shape dia-
gram in the Pulse shape editor and the Pulse shape details dia-
log boxes can be used to switch on or off the zooming in the
pulse shape diagram. The resolution of the X axis of the diagram
can be zoomed.
Prerequisite:
■ The Pulse shape editor dialog box or the Pulse shape details
dialog box is open.
User level ≥ 40
1. Select Magnifying glass.
Two buttons with upward and downward arrows are dis-
played. The upward arrow increases the resolution of the
X axis. The downward arrow reduces the resolution of
the X axis.
2. Select the upward or downward arrow.
The resolution of the X axis increases or is reduced.
Two buttons with left and right arrows are displayed.
These arrows are used to move the visible section of
the time range on the X axis.
Tab. 4-39
Diagnosis
5 Messages 5‐18
5.1 Displaying messages 5‐19
5.2 Displaying the history of messages 5‐20
5.3 Displaying a message text 5‐21
5.4 Resetting messages 5‐22
5.5 Setting the message filter 5‐22
5.6 Searching for a message 5‐23
5.7 Adding a note 5‐23
9 Statistics 5‐29
9.1 Utilization of the laser 5‐29
Display the statistics of the MPI slots 5‐30
Showing the distribution over all MPI slots 5‐31
Resetting the counters for all MPI slots 5‐31
9.2 Use of the light paths 5‐32
Display of the light paths statistics 5‐33
Resetting the counters for all light paths 5‐33
User level ≥ 0
1. Select the main activity Production.
User level ≥ 0
1. Select the main activity Production.
User level ≥ 50
1. Select the main activity Production.
Tab. 5-3
Now you have the option to invert the signals and edit the data
words.
Inverting signals
Prerequisite:
■ The simulation has been started (see "Start simulation", pg.
5‐6).
User level ≥ 50
1. Select the Signals tab.
2. Select the desired input signal from the "Input signal"
list.
or
Select the desired output signal from the "Output signal"
list.
The selected signal is highlighted in yellow.
3. Select Invert.
The selected signal has been inverted.
Tab. 5-4
Prerequisite:
■ The simulation has been started (see "Start simulation", pg.
5‐6).
User level ≥ 50
1. Select the Words tab.
2. Select the desired input word from the "Input word" list.
or
Select the desired output word from the "Output word"
list.
The selected signal is highlighted in yellow.
3. Select Input.
The dialog box for the value input is opened. The input
can be performed in a hexadecimal or decimal form.
If an input or output word 0/1 is selected, the Input but-
ton is used to open a dialog box for light path and pro-
gram number input.
4. Select Close.
The data words have been edited.
Tab. 5-5
This function is used to display the states of the inputs and out-
puts (I/O signals) of the selected slot.
User level ≥ 0
1. Select the main activity Production.
User level ≥ 0
1. Select the main activity Production.
Prerequisite:
■ The I/O signals are displayed (see "Displaying I/O signals",
pg. 5‐7).
User level ≥ 0
Select Previous slot.
or
Select Next slot.
The previous or next slot is displayed.
Tab. 5-8
Prerequisite:
■ The I/O signals are displayed (see "Displaying I/O signals",
pg. 5‐7).
■ A real-time interface has been assigned to the selected slot.
User level ≥ 0
Select Real-time interface.
The PLC interface details (signals) dialog box is dis-
played. The input and output signals of the real-time
interface are displayed.
Tab. 5-9
Prerequisite:
■ The I/O signals are displayed (see "Displaying I/O signals",
pg. 5‐7).
User level ≥ 0
1. Select Signal selection.
The PLC interface details (signals) dialog box is dis-
played.
2. Select Inputs tab.
or
Select the Outputs tab.
3. Select Activate all.
or
Select Deactivate all.
or
Tick or untick the desired input and outputs.
4. Select Apply.
Tab. 5-10
Prerequisite:
■ The I/O signals are displayed(see "Displaying I/O signals",
pg. 5‐7).
User level ≥ 0
1. Select the required input or output word from the input
or output word list.
The selected word is highlighted in yellow.
2. Select Details.
The PLC interface details (signals) dialog box for the
required input or output word is displayed.
Tab. 5-11
The light path list shows all light paths and the statuses of the
components in the light paths.
AW-00018EN
Light path overview Fig. 5-1
Symbol Meaning
Green: light path active
OK
Fault
Not connected
Tab. 5-13
Note
In the "LLK" column, information is displayed only if the light
path is operated with an ILN box or an LLKM board..
Information is only shown in the column "LPL" if this function is
set in the system settings. For further information, contact the
TRUMPF after-sales service.
User level ≥ 0
1. Select the main activity Production.
Prerequisite:
■ The light path overview is displayed (see "Displaying the light
path overview", pg. 5‐13).
User level ≥ 0
1. Select the button in the desired column.
The dialog box with the corresponding detail information
is displayed.
Tab. 5-15
User level ≥ 60
1. Select the main activityTechnology.
8. Select Save .
The name of the light path is changed.
Tab. 5-17
User level ≥ 60
1. Select the main activityTechnology.
Symbol Meaning
Pilot laser is switched on.
Tab. 5-19
User level ≥ 10
1. Select the main activity Setup.
6. Select Apply.
The pilot laser mode has been set.
Tab. 5-20
Mode Meaning
0 Pilot laser constantly off.
1 Pilot laser is switched on if the shutter switch is closed.
2 The pilot laser is switched on if the safety circuit is inter-
rupted.
3 The pilot laser is active, when
■ its switching functions are activated
OR
■ External signal PILOT_LASER_ON is set.
■ AND
Shutter switch closed.
Note
The pilot laser is switched off irrespective of the set pilot laser
mode if the single channel safety circuit (ASV3, connector X24)
is used for internal functions (for ex. protection of the laser light
cables in the user room) AND if the single channel safety circuit
is interrupted.
In the light path overview, the user can switch the pilot laser for
the individual light paths on and off.
The pilot laser can be activated or deactivated in the secondary
activity Switching functions in order to switch on and off all exist-
ing light paths (see chapter "Diagnosis").
Note
The pilot laser is switched off irrespective of the set pilot laser
mode if the single channel safety circuit (ASV3, connector X24)
is used for internal functions (e.g. protection of the laser light
cables in the user room) AND if the single channel safety circuit
is interrupted.
User level ≥ 0
1. Select the main activity Production.
5. Select Close.
Tab. 5-23
Monitoring message For the majority of the operational values monitored by the laser
control, monitoring limit values have been defined. Within these
limits, the laser operates without restrictions. If an operational
value reaches the monitoring limit, a monitoring message is dis-
played in the message line. This message informs about the cur-
rent status of the laser device and may be the first hint as to a
possible fault. Monitoring messages can be displayed when the
laser device is switched on but the laser is not activated.
Fault message For the majority of the operational values monitored by the con-
trol unit, fault limit values have been defined. If an operational
value reaches the fault limit, the laser is automatically switched
off. The fault message is displayed in the message line. The
laser can be switched on again only after the fault has been
eliminated and the message has been reset.
Type of message The message types are marked by the following symbols:
Context bar The context bar (2) contains buttons for functions available in the
secondary operation Messages .
2. Select Messages.
The list of pending messages is displayed.
Tab. 5-25
Filter settings The filter settings are displayed in the area Active filter (1) .
Message list The message list (2) contains the last 100 messages when the
message stack is called for the first time.
User level ≥ 0
1. Select the main activity Diagnosis.
2. Select Messages.
The list of pending messages is displayed.
3. Select Message history.
The history of messages is displayed.
Tab. 5-26
Note
A specific measure belongs to each Cause. The effect describes
the effect of the reported state on the laser device. An effect
may have several causes.
User level ≥ 0
1. Select the main activity Diagnosis.
2. Select Messages.
Tab. 5-27
2. Select Messages.
Tab. 5-28
User level ≥ 0
1. Select the main activity Diagnosis.
Select Messages.
The list of pending messages is displayed.
2. Select Message history.
User level ≥ 0
1. Select the main activity Diagnosis.
Tab. 5-30
User level ≥ 0
1. Select the main activity Diagnosis.
2. Select Messages.
5. Select Note.
A screen keyboard is displayed.
6. Enter a note.
7. Select Apply.
The note is displayed in the message text below the
TRUMPF service information.
Tab. 5-31
Note
A message code consists of hexadecimal characters "aabbccdd",
e.g. "E401 D023". The code is composed as follows:
aa: the ID of the component.
Tab. 5-33
User level ≥ 0
1. Select the main activity Diagnosis.
User level ≥ 0
1. Select the main activity Diagnosis.
Note
All mentioned hour counters can be reset so that a new evalua-
tion can be started.
User level ≥ 50
1. Select the main activity Diagnosis.
2. Select Statistics.
Tip
User level ≥ 50
1. Select the main activity Diagnosis.
2. Select Statistics.
Tip
Information on the signification of the distribution can be called
by pressing .
User level ≥ 50
1. Select the main activity Diagnosis.
2. Select Statistics.
Select Yes .
All hour counters are reset. The statistical evaluation
starts from the beginning.
Tab. 5-38
This function is used to determine the service life per light path
and to prepare the graphical representation. For the calculation,
the active times of each light path are taken into consideration.
Note
All hour counters of the light paths can be reset so that a new
evaluation can be started.
User level ≥ 50
1. Select the main activity Diagnosis.
2. Select Statistics.
User level ≥ 50
1. Select the main activity Diagnosis.
2. Select Statistics.
Select Yes .
All hour counters are reset. The statistical evaluation
starts from the beginning.
Tab. 5-40
User level ≥ 0
1. Select the main activity Diagnosis.
It shows:
■ the voltage transmitted on the basis of an analog value from
the AE or AEF board to the control unit as default value.
■ the parameter resulting from the voltage.
Note
If the analog voltage defines nominal values that lie below or
above the allowed default values, a white flash on a red back-
ground is shown on the left edge of the field in question.
User level ≥ 0
1. Select the main activity Diagnosis.
1 2
AW-00017EN 4
1 "Parameter set or laser pro- 3 "Nominal values" display 4 "Actual values" display
gram" display
2 "Light path status" display
Fig. 5-7
"Parameter set or laser The "Parameter set or laser program" display (1) shows the cur-
program" display rently active laser program or the currently active parameter set.
"Light path status" display The "Light paths status" display (2) shows by means of symbols
the status of a light path:
Fault
Light path OK
Tab. 5-43
"Nominal values" display The "Nominal values" display (3) shows the values preset
through the parameter set or the laser program.
"Actual values" display The "Actual values" display (4) shows which values are being
actually achieved.
User level ≥ 0
1. Select the main activity Production.
The following actual values are displayed for pulsed laser devi-
ces:
■ Power [J]
■ average power [W]
■ Frequency [Hz]
■ Controller intervention [%]
User level ≥ 0
1. Select the main activity Diagnosis.
3. Selecttab Overview .
The pulse data are displayed.
Tab. 5-45
Control
■ Mode: mode set in the energy control
■ Value [J]: value set for the pulse energy in the energy control
■ Tolerance [J]: tolerance set in the energy control
Statistics
■ E 1 [J]: energy of the first pulse in Joule
■ E av [J]: average pulse energy in Joule
■ E max [J]: maximum pulse energy in Joule
■ E min [J]: minimum pulse energy in Joule
■ Pulses: number of pulses used for averaging
The number of pulses used for averaging is limited to 65000.
If this value is reached, the statistical values are no longer
updated. The displayed values refer in this case only to the
65000 pulses recorded.
■ E av + [%]: deviation in percent of the maximum pulse
energy from the average pulse energy
■ E av - [%]: deviation in percent of the minimum pulse energy
from the average pulse energy
Several LMP boards can be connected. For switching between
several LMP boards, you can use the buttons Previous LMP and
Next LMP.
Using the Energy table button, you can call the last 500 pulses
triggered and have their measured values represented in a
graph.
User level ≥ 0
1. Select the main activity Diagnosis.
Tab. 5-46
User level ≥ 0
1. Select the main activity Diagnosis.
Tab. 5-47
1
2
1
2
3
4
5
AW-00034
1 Pulse 3 Energy progress until the capac- 5 Time axis in ms
2 Capacitor charge between two itor has been completely
pulses charged
4 Utilization axis
Pulsed laser: Time for the capacitor to reach the fully-charged state Fig. 5-8
User level ≥ 0
1. Select the main activity Diagnosis.
User level ≥ 0
1. Select the main activity Diagnosis.
Color State
Blue Lamp OK
Red Lamp defective
Yellow Lamp lit
Tab. 5-51
User level ≥ 0
1. Select the main activity Diagnosis.
2. Select Cavity.
An overview graph of the cavity is displayed.
Tab. 5-52
4 2
AW-00035EN
Data table The data table (1) displays the recorded data:
■ ID: identification of the data record
■ DG: datagram number
■ WP: workplace
■ Date/time
■ I0 to I15: input word 0 to 15
■ O0 to O63: output word 0 to 63
Table Output signals The table Output signals(2) shows the status of the output sig-
nals:
Tab. 5-53
Context bar The context bar (3) contains buttons available in the secondary
operation PlcView.
Table Input signals The output signals table (4) shows the status of the output sig-
nals:
Tab. 5-54
User level ≥ 60
1. Select the main activity Diagnosis.
2. Select PlcView.
3. Select Display.
The dialog with the PlcView details is displayed.
Tab. 5-55
User level ≥ 60
1. Select the main activity Diagnosis.
3. Select Display.
The dialog with the PlcView details is displayed.
User level ≥ 60
1. Select the main activity Diagnosis.
2. Select PlcView.
3. Select Display.
The dialog with the PlcView details is displayed.
User level ≥ 60
1. Select the main activity Diagnosis.
2. Select PlcView.
3. Select Display.
The dialog with the PlcView details is displayed.
4. Select IO filter.
The IO filter dialog box is displayed.
8. Select Apply.
The changes are displayed in the IO filter dialog box.
9. Select Apply.
The changes are displayed in the data table.
Tab. 5-58
Prerequisite:
■ The TVar variable has been set to TRUE in the system set-
tings.
User level ≥ 60
1. Select the main activity Diagnosis.
Prerequisite:
■ The XVar variable has been set to TRUE in the system set-
tings.
User level ≥ 60
1. Select the main activity Diagnosis.
User level ≥ 60
1. Select the main activity Diagnosis.
Maintenance
1 Overview 6‐2
Symbol Meaning
No maintenance work required.
Tab. 6-1
Tab. 6-2
User level ≥ 30
1. Select the main activity Maintenance & Commissioning.
2. Select Maintenance.
Tab. 6-3
Manual mode
Note
Due to the dangers caused by laser light, it is possible that the
user level for the manual mode is set to a higher level than level
20. Change the user level (see chapter "Setup") or contact the
TRUMPF after-sales service.
If you operate a pulsed laser device with the focusing optics
MFO (Manual Focusing Optics) the MFO tab is displayed addi-
tionally. In the tab MFO, an MFO program can be selected. Fur-
ther information can be found in the operating instructions MFO.
If you use a laser workstation TruLaser Station 5004, the addi-
tional tabs CNC axes and Manual axes are displayed. Further
information to this subject can be found in the operating instruc-
tions TruLaser Station 5004.
Setting the parameters Select a laser program in the tab laser program. The laser pro-
gram editor can be used to create laser programs. Information
on this topic can be found in the chapter ”Programming"
In the Pulse data tab, a parameter set can be created or
selected. Laser parameters can be entered as default values, for
Selecting a light path The optical arrangement of the laser device mostly permits
selection between several laser paths through which the laser
light can be conducted to the machining point. The light path of
the laser beam is programmed in the laser program or selected
in the parameter set.
Activating the laser The Pulse data tab and the Laser program tab contain buttons to
activate and trigger the laser.
3
9
4
AW-00020EN
1 Parameter set selection field 4 Pulse duration input field 7 Light path selection field
2 Pulse type selection field 5 Number of pulses input field 8 Context bar
3 Power input field 6 Frequency input field 9 Parameter set control values
Pulse data tab Fig. 7-1
Parameter set selection field Parameter sets are combinations of values arranged for a laser
application in the manual mode, for example pulse power, pulse
duration, pulse number etc. Parameter sets are the simplest ele-
ments to condition a laser for an application. Other elements are
pulse shapes and laser programs (see chapter "Programming").
Existing parameter sets can be selected in the parameter sets
selection field (1). New parameter sets can be created by means
of the Manage button.
Pulse data The pulse data pulse type (2), power (3), pulse duration (4),
number (5), frequency (6) and light path (7) are saved automati-
cally in the selected parameter set during input.
Pulse type selection field In the pulse type selection field (2) pulse shapes like e.g. square
pulse can be selected. Pulse types are defined by means of the
pulse type editor in the >Programming >Pulse type dialog box.
Light path selection field The light path for the laser light to be guided is selected in the
Light path selection field(7). If the light path 0 is selected, the
central shutter is closed and the laser beam is guided into the
Context bar The context bar (8) contains buttons available in the Pulse data
tab.
Parameter set control values The values in the fields of the parameter set control values (9)
have the following meaning:
■ Pulsed laser devices:
− Energy: energy of a laser pulse.
− Average power: product of energy and frequency.
− Utilization: percentual value referred to the rated power of
the laser device.
■ CW laser devices:
− Power: laser power.
− Energy: product of power and frequency.
− Average power: product of energy and frequency.
■ Short-pulse laser devices:
− Power: laser power.
− Energy: product of power and frequency.
− Short-pulse energy if short-pulse parameter = on: product
of power and short-pulse duration.
− Average power if short-pulse parameter = off: product of
energy and frequency.
If a pulse is triggered, the actually measured values are dis-
played in the fields.
If on the left side of a field the symbol "white flash on a red
background" is displayed, no laser pulsed are generated
because the corresponding limit value has been exceeded.
User level ≥ 20
1. Select the main activity Manual mode.
User level ≥ 20
1. Select the main activity Manual mode.
Tab. 7-2
User level ≥ 20
1. Select the main activity Manual mode.
Tab. 7-3
User level ≥ 20
1. Select the main activity Manual mode.
6. Select Yes.
The parameter set has been deleted.
Tab. 7-4
User level ≥ 20
1. Select the main activity Manual mode.
Tab. 7-5
Prerequisites:
■ Main switch and laser control are switched on.
■ Laser is switched off.
■ There are no active fault messages.
User level ≥ 20
1. Select the main activity Manual mode.
4. Select Laser .
After approx. 15 seconds, the laser is ready to create
laser light. Yet no laser light is generated and no pulses
are triggered. The central shutter (VZ) is closed. The
Laser button changes its color from gray to yellow. The
laser is switched on.
Tab. 7-6
Use the Shutter button in the parameter set to open and close
the central shutter (VZ).
With the central shutter (VZ) open, laser light can be transmitted
to the optical components via the selected light path. With the
central shutter (VZ) closed, no laser light can be emitted from
the laser device.
User level ≥ 20
1. Select the main activity Manual mode.
Prerequisites:
■ There are no active monitoring and fault messages.
■ The laser has been switched on (see "Switching on or off the
laser in the parameter set", pg. 7‐9).
■ The central shutter (VZ) is open (see "Open or close the
central shutter (VZ) in the parameter set.", pg. 7‐10)
■ The correct light path has been selected.
■ The pulse data have been set (see "Creating the parameter
set", pg. 7‐7).
User level ≥ 20
1. Select the main activity Manual mode.
Prerequisite:
■ There is an active monitoring or fault message.
User level ≥ 20
1. Select the main activity Manual mode.
3
AW-00021EN
1 "Laser program" selection field 3 Context bar 4 Laser program control values
2 "Laser program lines" display
Laser program tab Fig. 7-2
Laser program selection The Laser program selection field (1) is used to select laser pro-
field grams. Laser programs are created in the laser program editor
(see chapter "Programming").
Laser program lines display The "Laser program line" display (2) shows the single lines of
the selected laser program.
Context bar The context bar (3) contains buttons available in the Laser pro-
gram tab.
The control values for energy and average power can be moni-
tored in two different ways:
■ by controlling the energy of the pulses in the laser program
measured by the LMP board to be monitored in a laser pro-
gram. This energy control is set via the laser program, see
chapter "Programming".
■ by limiting the values for the maximum energy, average
power and for the actual power. This light path limitation (if
released) can be activated per light path, see chapter "Diag-
nostics".
Prerequisite:
■ At least one laser program is provided (see chapter "Pro-
gramming").
User level ≥ 20
1. Select the main activity Manual mode.
Prerequisites:
■ Main switch and laser control are switched on.
■ There are no active fault messages.
■ A laser program is provided.
User level ≥ 20
1. Select the main activity Manual mode.
4. Select Laser .
After approx. 15 seconds, the laser is ready to create
laser light. Yet no laser light is generated and no pulses
are triggered. The Laser button changes its color from
gray to yellow. The laser is switched on.
5. Select Laser . to deactivate the light generation.
The button changes its color from yellow to gray.
The light generation is deactivated.
Tab. 7-11
Prerequisites:
■ The laser has been switched on (see "Switching the laser on
and off", pg. 7‐14).
■ There are no active monitoring and fault messages.
Prerequisite:
■ There is an active monitoring or fault message.
User level ≥ 20
1. Select the main activity Manual mode.
Service
2 Faults 8‐4
Prerequisites:
■ The network configuration of the Ethernet interface in the
laser control is set correctly.
■ The laser control is connected to the Internet.
■ The firewall of the company network or the router enables
the following IP address and ports:
− IP address of the TRUMPF TelePresence Portal:
217.6.238.164
− Port IPSec (UDP 500)
− Port IPSec+NAT (UDP 4500)
User level ≥ 30
1. Select the main activity Help.
Note
The connection can be separated when the service technician
has finished the diagnosis.
User level ≥ 30
1. Select the main activity Help.
Tab. 8-3
Appendix
2 License 9‐3
2.1 Information on using the software licenses 9‐3
2.2 Display licenses 9‐4
User level ≥ 0
1. Select the main activity Help.
Tab. 9-1
The License dialog box in the main activity Help contains a list
of all program licenses used for programming the TruControl
1000.
Exclusion of liability The laser device has been designed exclusively for use of the
TruControl 1000 software in the form supplied. The user is solely
responsible for any modification of the Open Source compo-
nents. Any warranty claims shall be void in this case. With
respect to the copyright holders, the following disclaimer shall
apply to the GPL components:
"This program is distributed in the hope that it will be useful, but
WITHOUT ANY WARRANTY; without even the implied warranty
of MERCHANTABILITY or FITNESS FOR A PARTICULAR PUR-
POSE. See the GNU General Public License for more details."
For the other Open Source components we refer to the disclaim-
ers of the copyright holders in the enclosed source codes.
Source codes The source codes of the Open Source programs used can be
found on the enclosed CD together with the corresponding
license texts. The CD is enclosed for reasons of licensing but is
not required for the operation. If this CD has not been supplied
with the goods or if you need an updated source code after you
have performed an update, you can order the CD by specifying
the TruControl 1000 version number (see "Display versions", pg.
9‐2) at the following address:
TRUMPF Laser GmbH + Co. KG
TLS560 - Software Development Department
Aichhalder Strasse 39
D-78713 Schramberg
User level ≥ 0
1. Select the main activity Help.
Tab. 9-2