The World’s Most Reliable Electrolyte Analyzer

Software Overview Quick Reference Guide


SERVICE PRINT QC/STD/DIALYSATE OPERATOR INSTRUMENT SERVICE PERFORM
MENU? MAINTENANCE?
REPORTS? URINE SAMPLE? SETTINGS? SETTINGS? MENU? CALIBRATION?

SAMPLE CONTROL VERIFY ISE PASSWORD CHECK PERFORM


CLEAN?
REPORTS? LEVEL 1? PACK? XXX? ELECTRODES? CALIBRATION?
LIFT
DOOR PRINT CAL CONTROL ACTIVATE SETUP CONTROL CHECK SAMPLE
CONDITION?
REPORT? LEVEL 2? STANDBY? LEVEL 1 RANGES? SENSOR?

MEASUREMENT PRINT QC CONTROL PRIME SETUP CONTROL CHECK FLUID


SERUM / PLASMA AND STATS? LEVEL 3? FLUIDS? LEVEL 2 RANGES? PACK SENSOR?
BLOOD?
PRINT ALL STANDARD TIME/DATE SETUP CONTROL CHECK RFID
RANGES? SAMPLE? SETTINGS? LEVEL 3 RANGES? SENSOR?

PRINT ERROR BICARBONATE CHOOSE SETUP NORMAL CHECK


MESSAGES? SAMPLE? PARAMETER? RANGES? LANGUAGE?

PRINT CORREL ACETATE RESET SAMPLE SETUP CORREL. CHECK


FACTORS? SAMPLE? COUNTER? FACTORS? SAMPLE DOOR?

URINE PATIENT DATA SETUP BICARB. CHECK


SAMPLE? ENTRY ENABLE? CORREL. FACTORS? PUMP?

ACTIVATE ENTER ACETATE CHECK PINCH


SHUTDOWN? CORREL. FACTORS? VALVES?

VET MODE SETUP URINE CHECK


ON / OFF? CORREL. FACTORS? INTERFACE?

DAILY AUTO CAL CHECK


SETTINGS? AMPLIFIER?

SETUP SETUP
PRINTER? CONFIG?

SETUP
INTERFACE?

SELECT
LANGUAGE?

The Smart Lab Solution. ECO# 8446 SOP05-5028F REV04 Effective date: 04/19/17
The Smart Lab Solution.
The World’s Most Reliable Electrolyte Analyzer

Measurement Quick Reference Guide

H-READY NA-K-Cl
1 2 3
Always wear gloves when
handling blood or parts
potentially contaminated
BLOOD / SERUM LIFT DOOR LOAD SAMPLE FROM WIPE PROBE with blood.
PLASMA SAMPLE LOAD SAMPLE TUBE / SYRINGE / CUP SHUT SAMPLE DOOR

Na K Cl
00.1 1.69 0.309

H-READY NA-K-Cl QC/STD/DIALYSATE SELECT SAMPLE


URINE SAMPLE? TYPE

- CONTROL LEVEL
1,2,3 SAMPLE

4 5 6 - STD SAMPLE
- BICARBONATE/ACETATE
- URINE SAMPLE
7 Keep the main door
closed during
RESULTS QC, STD. DIALYSATE PRESS YES TO QC / STD SELECT TYPE AND measurement.
PRINTED & DISPLAYED URINE SAMPLE: PRESS NO DIALYSATE / URINE SAMPLE PROCEED WITH STEP 1

Maintenance Troubleshooting
Daily • Perform cleaning cycle (5 or more samples/daily) • Open front cover and check if all electrodes and the right side sample
• Perform conditioning cycle sensor have O-rings.
• Check printer paper supply • Make sure maintenance has been performed in accordance with
• Perform cleaning cycle (5 or more samples/daily) the maintenance schedule.
Weekly
• Clean sample probe and fill port • Check the FLUID PACK level by going to menu [VERIFY PACK?]
• Clean exterior analyzer surfaces under [OPERATOR SETTINGS?]
• Check for clots and crystals that may be forming in the tubing
Monthly • Clean the reference electrode housing
or in the measuring chamber.
(see instructions for Use - chapter 5)
• Pull pump windings from pump roller and check for kinked tubing
Semi-annually • Replace peristaltic pump tubing and tears in the tubing.
Annually • Replace main tubing harness • Start a calibration by going to [CALIBRATION?] and pressing YES