Professional Documents
Culture Documents
Chapter 3 UV-Vis
Chapter 3 UV-Vis
Ultraviolet and visible spectrometers have been in general use for the last 35 years
and over this period have become the most important analytical instrument in the
modern day laboratory. In many applications other techniques could be employed but
none rival UV-Visible spectrometry for its simplicity, versatility, speed, accuracy and
cost-effectiveness.
The figure illustrates the general pattern of electronic energy levels, of a molecule
and the fact that the transitions are brought about by the absorption of different
amounts of energy. Light in the UV-VIS part of the spectrum is used to promote
electrons from the ground state to various excited states.
Examples of transitions and resulting λmax
Molecule Transition λmax
CH3-CH3 - * 135
CH3-OH - * 150
n - * 183
CH2=CH2 - * 175
C6H6 - * 245
Measurement of Absorbance:
General terms and their relations :
PSolution P
T
PSolvent P0
P0
A - logT log
P
A - log(%T) 2
In the form of Lamberts Beers law:
Path length / cm 0 0.2 0.4 0.6 0.8 1.0
A bC %T
Absorbance
100
0
50
0.3
25
0.6
12.5
0.9
6.25
1.2
3.125
1.5
Lamberts Beers Law: Absorptivity and molar absorptivity:
o High concentration of some other colorless salts adversely affects the absorption
of light. Similar electronic interaction with analyte may occur.
o Change of refractive index of the medium do not adhere the Beer’s law. If the
concentration change causes significant deviation of refractive index departure
in Beers law will be originated.
A correction can be made by substituting ε with,
= true x h/( h² + 2)²
Apparent Chemical Deviations
• Caused by association/dissociation/reaction with solvent…..}.
A common example of this behavior is found with acid/base indicators.
Deviations arising from chemical factors can only be observed when
concentrations are changed.
430 570
HIn H+ + In- HIn 6.30 X 102 7.12 X 103
Color 1 Color 2 (Ka = 1.42 X10-5)
In- 2.06 X 104 9.61 X 102
If we calculate absorbance data for
the total indicator concentration
K a 1.42 10 5
If, [In] total 2 10 - 5 M
Gives, [In-] 1.12 10 -5 M and [HIn] 0.88 10 -5 CHIn, M [HInN] [IN-] A430 A570
0.88x10- 1.12x10-
2x10-5 5 5 0.236 0.073
2.22x10- 1.78x10-
A 430 (2.06 10 4 1 1.12 10 5 ) (6.30 10 2 1 0.88 10 5 ) 0.236 4x10-5 5 5 0.318 0.175
5.27x10- 2.73x10-
5 5
A 570 (9.61 10 1 1.12 10 ) (7.12 10 1 0.88 10 ) 0.073
2 3 8x10-5 5 5 0.596 0.401
8.52x10- 3.48x10-
12x10-5 5 5 0.771 0.640
11.9x10- 4.11x10-
16x10-5 5 5 0.922 0.887
Beer’s law is not obeyed at both wavelength. The direction of fluctuation of absorbance from Beer’s law is opposite at two wavelength.
Apparent Instrumental Deviations with Polychromatic Radiation
A good picture of the effect of polychromatic radiation on Beer’s law can be
presented as follows.
If we considered just the two and λ”, and assuming that Beer’s law applies at each of
these individual wavelengths,
The absorbance at λ‘
P0′
A = log ′ = 𝜀′bc P = Po 10-bc
P
The absorbance at λ‘‘
P′′0
A = log ′ = 𝜀′′bc P” = P”o 10-”bc
P′
The Overall Absorbance
(P0' P0'' )
A m log ' , m measured
(P P ) ''
In the very special case where ’ = ’’, the above equation simplifies to Beer’s law. Am = ’bc
Instrumental Deviations in the Presence of
Stray Radiation
(P0 PS )
A' log ................ (19)
(P PS )
Components are,
(1) sources, (2) wavelengths selectors (3) sample containers,
(4) radiation detectors, (5) signal processors and readout devices.
Light Sources
For the purpose of molecular absorption a continuous source is
required with constant power over a considerable range of
wavelengths.
Deuterium and Hydrogen Lamps
Deuterium or hydrogen between the two electrodes will be
excited at low pressure and gives continuum UV spectrum.
The excited molecule dissociates into atomic species and UV
photons.
These lamps provide a continuous source of radiation form 375 nm down to about 160 nm
hence quartz windows must be used with these types of lamps since glass absorbs strongly
at wavelengths less than about 350 nm
•Xenon arc lamp:
•Radiation is produced by passing current through xenon atmosphere.
•It is applicable for 200-1000nm, has intense peak around 500nm.
•Wavelength selector:
Are the device/s which restricts radiation to a narrow band, that is absorbed or
emitted from analyte.
Determine sensitivity and selectivity of the method
.These are filters, monochromators, spectrograph etc.
Radiation filters: Filter absorb all but a restricted band of radiation from a
continuum source. Interference filters (Febry-Perot filters):\
White radiation
Glass plate
Metal film
Dielectric layer
1 2 3 4
A
4'
1' 2' 3'
B
nl = 2th
Monochromators generally have a diffraction grating disperse the radiation into its
component wavelengths. contains: entrance slit, collimating lens, mirrors, focusinglens,
a dispersion medium (prism and grating) and an exit slit.
Mechanism of diffraction from an
echellette-type grating
nλ = d(sin i + sin r)
Use geometry to derive this relation
The beam from the source is split into two beams by a beam splitter which is a
semipermeable mirror.
The detector is usually a pair of photomultiplier tubes connected to a difference amplifier
and the wavelength selector is a grating or prism monochromators.
ABSORBING SPECIES
The absorption of UV/Vis. radiation by a molecular/atomic/ionic
species M can be considered to be a two-step process,
* Orbital v3
v2
v1
E0 v0
n bonding Orbital has the shape of p or hybridised orbital.
1. σ σ* trastion:
It is the transition between extreme position
Large energy is required; (vacuum UV) l max < 185nm .
CH4, l max = 125nm, C2H6, l max = 135nm
The strength of C-C > C-H bond (due to hybridization/overlapping)
l absorption Falls below cut-off point of most solvents [200-700]
Anti bonding orbital
C C *
*
C C
Bonding orbital
C C
1. Alkane:
Fig. Electronic trastion: σ σ* trastion:
2. n σ* trastion:
# A saturated organic compound must contain hetero atom/ with lone pair.
# The energy of transition, 150 nm < l max < 250nm
# maximum are below 200nm
# ranges between 100-3000 Lcm-1 mol-1 .less intense peak
# l for transtion depends on the nature of the bond and not on the structure.
H2O (l max= 167nm), CH3OH (l max=184 nm),
(CH3)2O (lmax=184 nm), CH3I (l max=258 nm),
# Absorbance maxima will shift to short wavelength by polar solvents ?? and is less
accessible for UV-vis absorption (below cut-off point) .
. n => *
C N * CN
C N
n(sp3)
C N -OH , -NH2= l175-200nm)
C N CN
C C
Alkene = l75nm)
C C
-Alkyne= l70nm)
3. Alkene, alkyne:
Fig. Electronic trastion: * trastion:
Organic chromophores and their approximate absorption maxima:
Note: This lmax is affected by solvent and structure details and also due to vibrational precise
determination of peak maxima is difficult. Also it is better to calculate max from cal. Curve
than the tabulated value.
Absorption spectra of acetone.
• λmax is the wavelength where maximum
absorption occurred
Chromophores:
• A chromophore (literally color-bearing) group is a functional group, not
conjugated with another group, which exhibits a characteristic absorption
spectrum in the ultraviolet or visible region. Like C=C, C=O, N=N, C=S
{n ⟶ 𝜎* , n ⟶ π* , π ⟶ π* }
• If the structural change occurs in the chromophore, the exact energy
and intensity of absorption are expected to be changed.
• When these groups are/is present the absorption pattern of the
compound give rough guidelines in identify the functional group.
• If any of the simple chromophores is conjugated with another (of the
same type or different type) a multiple chromophore is formed having a
new absorption band which is more intense and at a longer wavelength
that the strong bands of the simple chromophores.
Auxochrome:
• Functional group (or substituent’s) that do not absorb itself in
UV/vis region (above 220nm – 800 nm) but shift the
chromophore to long wavelength and also increase the intensity
of absorption.
• Auxochrome should contain at least a lone pair e- , which stabilizes
*excited state. e.g. Phenolate is strong auxochrome than phenol (anion has
extra pair of e-)
• Aniline is auxochrome but converting it to anilium cation the
effect is lost.
• Some auxochromes are –OH, -NH2
Shift Effect
• With increase in conjugation the HOMO-LUMO gap will decrease, the absorption
maxima for n * and *transtion shifts to longer wavelength (batho)
• Absorption by multifunctional chromophores in a single organic molecule is
approximately additive if are separated from one another by more than one single
bond.
• The conjugation of chromophore not only produce bathochromic shift but also
increase the intensity.
Alkene lmax
Ethylene 175 15000
1,3 butene 217 21000
1,3,5 hexatriene 258 35000
b carotene(4 double 465 125000
bond)
ketones
Acetone *189 300
n * 280 12
213
3-butene-2-one * 7100
320 27
n *
Lower wave
length limit
chemical structures involved.
Solvents
• Polar molecules when interacted with polar solvents show
dramatic effect.
• Interaction between solute and solvent leads to Acetonitrile 190
absorption band broadening and a consequent reduction Chloroform 240
in structural resolution and ε max. Cyclohexane 195
• Figure illustrates the effect of iso-octane and ethanol on 1-4 dioxane 215
the spectrum of phenol, a change from hydrocarbon to
95% ethanol 205
hydroxylic solvent.
N hexane 201
• The loss of fine structure in the latter is due to broad band
h-bonded solvent-solute complexes replacing the fine Methanol 205
structure present in the iso-octane. Isooctane 195
• The fine structure in the latter solvent illustrates the Water 190
principle that non-solvating or nonchelating solvents Trimethyl 210
produce a spectrum much closer to that obtained in the phosphate
gaseous state.
Spectra of Phenol in Iso-octane and in Ethanol
Energy
dxz dyz
Trithiocarbamate-500nm
APPLICATION OF UV/vis Spectroscopy
Application to absorbing species:
Use table:
• The determination of organic compound containing one or more of
these chromophores is feasible (many examples are found in
literature)
• Many inorganic species are susceptible for quantitative determination.
Like transition metal ion, nitrate ion, nitrite, chromate ion, osmium and
ruthenium tetroxides, molecular iodine, ozone
Procedural detail:
i. Selection of wavelength.
ii. The variables which can alter absorbance should be controlled.
iii. Cleaning and handling of cell/ cuvettes.
iv. Plotting the calibration curve for standard.
v. Dealing with mixtures of absorbing species.
References:
-
-
-
-