Diagnostic Ultrasound System
ProSound 4000
Model SSD-4000
A Platform for Pure Harmonic Detection
Beamformer
256 channels
Frame Rate
2
B-mode Spectral Doppler:
・ Display: Power spectrum
・ Display Gray Scale: 256 levels
・ Doppler methods:
・ Longitudinal and lateral inversion
PW (Pulsed Wave) Doppler
・ 90 degrees rotation
HPRF (High Pulse Repetition Frequency) PW
・ Scanning area:
Doppler
100% to 25%, continuously changeable
CW (Continuous Wave) Doppler*
・ Image Zoom
* CW Doppler is optional
Write zoom (magnification of real-time image):
・ Reference frequencies:
Max. 6 times
PW: 2.14, 2.5, 3, 3.75, 5, 6, and 7.5 MHz
Read zoom (magnification of frozen image):
CW: 2.14, 3, 3.75 MHz
2 times
・ Pulse repetition frequency on PW Doppler:
・ Depth range selections *:
PW: 0.5 to 20 kHz
2 to 24 cm
・ Base line shift: Available up to double velocity
* Probe dependent
・ Max. velocity range:
・ Image Select (B): 3 selections
PW: –6.36 to 0 or 0 to +6.36 m/sec (2.14 MHz
Resolution, Standard, Penetration
reference freq., 0 degree, with base line shift)
・ Scanning line density:
CW:–15.92 to 0 or 0 to +15.92 m/sec (2.14 MHz
Max. 3 selections
reference freq., 0 degree with base line shift)
・ Gain: 30 to 90 dB
・ Steerable CW Doppler: Possible with some phased
・ STC (sensitivity time control) — gain versus depth
array sector probes
curve control: 8 slide controls
・ Steered linear scanning: Max. ±15 degrees *
・ Contrast: 16 steps
5 degrees step changeable
・ AGC—Suppression of brightness saturation:
* Probe dependent
16 steps
・ Spectrum inversion: Possible
・ Relief: 4 selections (Off + 3 levels)
・ Angle correction:
・ Frame correlation: 4 kinds (16 steps each)
Available up to 80 degrees presetable
・ Horizontal smoothing: 3 selections (Off + 2 levels)
・ Sample volume size for PW Doppler:
・ Post Processing
0.5 - 10 mm, changeable in 0.5 mm step
Echo enhance curve: 4 kinds
・ Wall motion filter:
Rejection: 64 steps
Manual: 50, 100, 200, 400, 800 or 1600 Hz,
・ View gamma: 5 kinds
PW and CW independently selectable
・ Color Map (B): Gray, A, B, C, D
Auto: 12 steps automatically changed
・ Doppler gain: 0 - 60 dB
M-mode: ・ Contrast: 16 steps, selectable
・ Display method: Scroll or moving bar ・ Black/white inversion: Possible
・ Scroll speed: 1, 1.5, 2, 3, 4, 6, or 8 sec/plane ・ Audio output: 2 channels stereo
・ Gain: B-gain ±30 dB ・ Scroll speed: 2, 3, 4, 6, or 8 sec/plane
・ Contrast: 16 steps ・ Image Select (D): 3 selections
・ AGC—Suppression of brightness saturation: Resolution, Standard, Penetration
3
Color Flow Imaging Volume Mode*
・ Scanning method: Convex sector & mechanical
・ Represented parameters:
sector
Color Doppler mode: Mean Doppler frequency shift
・ Image processing
PowerFlow mode: Total power
Pre-processing: Opacity, Contrast, Gamma
・ Color area size: Variable from 100% to 5%
Post-processing: Brightness
・ Steered linear scanning:
・ Image magnification (possible on frozen images):
Max. ±15 degrees *, 5 degrees step changeable
From 0.75 to 2 times in 0.25 times increment
* Probe dependent
・ Image rotation (possible on frozen images):
・ Reference frequency:
Possible by 45 degrees step
2.14, 2.5, 3, 3.75, 5, 6, and 7.5 MHz
Volume Mode probes:
・ Pulse repetition frequency:
ASU-1001
0.5 to 10.0 kHz
・ Scanning angle:
・ Color base line shift:
Convex sector: 90゜
Available up to double velocity (±31 steps)
Mechanical sector: Max. 60゜
・ Max. velocity range:
・ Scanning depth selection:
–3.98 to 0 or 0 to +3.98 m/sec (at 2.14 MHz
4 - 19 cm selectable in 9 steps
reference freq. with base line shift)
・ Frame rate: Max. 4 frames/sec
・ Gradation:
ASU-1003
±32 levels for velocity (red and blue)
・ Scanning angle:
16 levels for variance (green)
Convex sector: 163゜
・ Capture mode: Off, On
Mechanical sector: Max. 90゜
Capture time: 1 sec./2 sec./3 sec./Continuous
・ Scanning depth selection:
・ Color map: 30 kinds (5 kinds x 6 groups)
3 - 17 cm selectable in 9 steps
・ Color Polarity: Normal, Invert
・ Frame rate: Max. 2 frames/sec
・ Flow filter: 6 selections
* Option
・ Image Select
Resolution, Standard, Penetration
Physiological Signal Display*
・ Average: 3 levels
・ Rejection: 16 steps ・ ECG, PCG, and pulse wave
・ Frame Rate Accelerator: Off, On ・ ECG lead ― second lead system
・ Frame correlation: 8 selections ・ ECG synchronized display:
・ Wall Motion Reduction: Available for one phase
4
iDMS (Integrated Data Management System):
<Standard features>
・ Store Memory
Hard Disk memory capacity: Max. 999 images*
* Still images
Stored images remain even after powering off.
・ Patient data management
・ Bitmap and TIFF static frame image storage:
MO disk*, Floppy disk
* MO disk drive is optional.
5
Measurement and Analysis: Measurements:
M mode
● Basic measurements
Mitral valve
On B-mode image LA/Ao
Distance Tricuspid valve
Area and Circumference (Trace, Ellipse, Circle) Pulmonary valve
Volume (Biplane, Single plane) Doppler mode
B. Index LV outflow tract flow
Histogram RVOT flow (right ventricle outflow tract)
On M-mode image Trans-mitral flow
Velocity Regurgitant flow (AR, PR, MR, TR)
M. Length Stenotic flow (AS, PS, MS, TS)
Time interval ● Peripheral vessels analysis
Heart rate
Carotid artery:
M. Index
CCA (common carotid artery)
On spectral Doppler
ICA (internal carotid artery)
Velocity
ECA (external carotid artery)
Acceleration
BIFUR (Bifurcation of carotid artery)
Mean flow velocity
VERT (Vertebral artery)
Resistance index
% Stenosis area
Pulsatility index
% Stenosis diameter
Pressure half time
Measurements on arteries in extremities:
Heart rate
Lower extremity artery,
D. Caliper
Upper extremity artery
D. Index
% Stenosis area
Time interval
% Stenosis diameter
Stenotic Flow
Measurements on veins in extremities:
Regurgitant Flow
Lower extremity venous
D. Trace
Upper extremity venous
・ Doppler auto trace: Possible on frozen image
On B/D mode image
Flow volume
Stroke volume
● Report Functions
● Obstetrical measurements & calculations
・ Obstetrical report
Gestational age
・ Gynecological report
Fetal weight
・ Cardiac function report
Fetal Doppler measurements
・ OB/GYN examination history
Fetal cardiac function measurements
Amniotic fluid index and volume
Cervical length
● Gynecological measurements & calculations
Uterus measurements
Ovary measurements
Follicular measurements
Urinary bladder measurements
Gynecological Doppler measurements
● Cardiac analysis
Left ventricle analysis:
M mode
Pombo (wall), Teichholz (wall)
Gibson (wall)
B mode
Area-length, Simplified Simpson, Bi-plane ellipse,
Bi-plane Simpson, Modified Simpson, Bullet
Others: AVA, MVA, LA/Ao, Ratio, RVD
6
Viewing Monitor
Acoustic Power ・ 15-inch diagonal flicker-free non-interlaced monitor
・ 0 to 100%, changeable ・ Tilt, swivel, height adjustment are possible.
Preset Function
・ 15 separate programs for specific clinical Safety Regulation
applications and/or users ・ Complies with IEC 60601-1 Class Ⅰ, Type BF
・ User programmable and/or factory default settings
Environmental Requirements
・ Preset contents storable in a floppy disk
In Operation
Characters and graphic displays ・ Temperature: +10 to +40 degrees C
・ Relative Humidity: 30 to 75%
・ Character input area:
(non condensing)
ID, name, age, sex, retained text ・ Atmospheric pressure: 700 to 1060 hPa
・ Automatic Annotation Labeling: In Storage/transportation
100 words/dictionary x 6 dictionaries ・ Temperature: -10 to +50 degrees C
(0 to +50 degrees C for mechanical probes)
・ Body mark: Default setting: 53 kinds
・ Relative Humidity: 10 to 90%
User edition: 24 kinds (non condensing)
・ Atmospheric pressure: 700 to 1060 hPa
Probe Connectors
・ For electronic scanning probes: 3 connectors
Power Requirement
・ Mechanical scanning probe: 1 connector (option)
・ 115/ 200 to 240V ±10%, 50 or 60 Hz,
Max. 1100VA
Video Signals
Input: Dimensions
Color composite, Y/C separated signals, Audio L/R, ・ 49 cm (W) × 86 cm (D) × 131 - 156 cm (H)
7
System Configuration
Standard Components
Standard components and optional items depend on the country.
Main unit with 15-inch color viewing monitor *1 Aloka recommend UST-9123 for general abdominal imaging and
Probe * 1 UST-5299 for general cardiac examination.
VCR (Panasonic)
Video cable: AG-C70A Remote controller: AG-A600
NTSC: AG-7350
PAL: AG-7350E
NTSC/PAL (not CE marked), Worldwide voltage: PAL (CE marked), 220-240V: *3 Color print pack: UPC-2010
UP-2950MD (SONY) *3 UP-2850P (SONY)*3
CP700E (Mitsubishi)*4 *4 Paper: CK700,
NTSC, 100-120V: Ink sheet: PK700S or PK700L
CP700UM (Mitsubishi)*4
B/W printer
SP-4000-V1.1-00
Optional Probes (See next page onward.)
8
OPTIONAL PROBES
9
Electronic linear probes
Application Model number Ultrasound Frequencies (MHz) Scanning Optional accessories
Doppler, Flow width
(description) B and M modes
modes (mm)
10
Volume Mode probes
NOTE: Volume Mode unit EU-9084 is necessary to connect Volume Mode probes
Application Ultrasound Frequencies (MHz) Scanning Radius of
Model number angle curvature Optional accessories
(description) B and M modes Doppler/Flow modes (degrees) (mmR)
11
・ The specifications are subject to change without notice.
We care, Ultrasound@Aloka