Diagnostic Ultrasound System
ProSound 4000SV
A Platform for Pure Harmonic Detection
2
B-mode Spectral Doppler:
・ Display: Power spectrum
・ Display Gray Scale: 256 levels
・ Doppler methods:
・ Longitudinal and lateral inversion
PW (Pulsed Wave) Doppler
・ 90 degrees rotation
HPRF (High Pulse Repetition Frequency) PW
・ Scanning area:
Doppler
100% to 25%, continuously changeable
CW (Continuous Wave) Doppler*
・ Image Zoom
* CW Doppler is optional
Write zoom (magnification of real-time image):
・ Reference frequencies:
Max. 6 times
PW: 2.14, 2.5, 3, 3.75, 5, 6, and 7.5 MHz
Read zoom (magnification of frozen image):
CW: 2.14, 3, 3.75 MHz
2 times
・ Pulse repetition frequency on PW Doppler:
・ Depth range selections *:
PW: 0.5 to 20 kHz
2 to 24 cm
・ Base line shift: Available up to double velocity
* Probe dependent
・ Max. velocity range:
・ Scanning line density:
PW: –6.36 to 0 or 0 to +6.36 m/sec (2.14 MHz
Max. 3 selections
reference freq., 0 degree, with base line shift)
・ Gain: 30 to 90 dB
CW:–15.92 to 0 or 0 to +15.92 m/sec (2.14
・ STC (sensitivity time control) — gain versus depth
MHz reference freq., 0 degree with base line
curve control: 8 slide controls
・ Angle gain: shift)
Gain versus angle curve control: 8 slide ・ Steerable CW Doppler: Possible with some phased
controls (Available on phased array sector array sector probes
image) ・ Steered linear scanning: Max. ±15 degrees *
・ Contrast: 16 steps 5 degrees step changeable
・ AGC—Suppression of brightness saturation: * Probe dependent
16 steps ・ Spectrum inversion: Possible
・ Relief: 4 selections (Off + 3 levels) ・ Angle correction:
・ Frame correlation: 4 kinds (16 steps each) Available up to 80 degrees presetable
・ Horizontal smoothing: 3 selections (Off + 2 levels) ・ Sample volume size for PW Doppler:
・ Post Processing 0.5 - 10 mm, changeable in 0.5 mm step
Echo enhance curve: 4 kinds ・ Wall motion filter:
Rejection: 64 steps Manual: 50, 100, 200, 400, 800 or 1600 Hz,
・ View gamma: 5 kinds PW and CW independently selectable
・ Color Map (B): Gray, A, B, C, D Auto: 12 steps automatically changed
・ Doppler gain: 0 - 60 dB
3
Color Flow Imaging
RT-3D (Real-time 3D)*
・ Represented parameters:
Real-time 3D imaging is possible with trans-abdominal
Color Doppler mode: Mean Doppler frequency shift
and trans-vaginal 3D probes.
PowerFlow mode: Total power
• 3D image rotation: 180 degrees on X, Y, and Z
・ Color area size: Variable from 100% to 5%
axes)
・ Steered linear scanning:
• 3D image shift: Possible
Max. ±15 degrees *, 5 degrees step changeable
• Multi-planar display: It is possible to display 3
* Probe dependent
mutually orthogonal sections simultaneously with 3D
・ Reference frequency:
image
2.14, 2.5, 3, 3.75, 5, 6, and 7.5 MHz
• Scanning rate: Up to 6 volumes/sec
・ Pulse repetition frequency:
• 4 kinds of rendering selectable
0.5 to 10.0 kHz
• Detailed scanning in the ROI: Possible
・ Color base line shift:
• It is possible to remove unnecessary part of a 3D
Available up to double velocity (±31 steps)
image
Max. velocity range:
• Volume cine search function: Storable up to 71
–3.98 to 0 or 0 to +3.98 m/sec (at 2.14 MHz
volumes of 3D images (depends on various settings).
reference freq. with base line shift)
• Animated loop replay: Possible
・ Gradation:
• Measurements on each plane (B-mode): Possible
±32 levels for velocity (red and blue)
* Option
16 levels for variance (green)
・ Capture mode: Off, On
VCM (Video Cine Memory)
Capture time: 1 sec./2 sec./3 sec./Continuous
• The number of acquirable frames
・ Color map: 30 kinds (5 kinds x 6 groups)
EU-9083 EU-9111
・ Color Polarity: Normal, Invert Grayscale 217 frames 300 frames x4
・ Flow filter: 6 selections images planes
・ Image Select Color 108 frames 300 frames x 4
images planes
Resolution, Standard, Penetration
• In continuous acquisition mode
・ Average: 3 levels
EU-9083 EU-9111
・ Rejection: 16 steps
Grayscale - Approx. 3,300
・ Frame Rate Accelerator: Off, On images frames
・ Frame correlation: 8 selections Color - Approx. 1,600
・ Wall Motion Reduction: images frames
・ Cine search and loop display (B mode): In sweep mode (M, D): 3 ch
4
iDMS (Integrated Data Management System): standard supplement 10. The HIS/RIS network and the
DICOM network need to be linked.
Standard features: ・ Router setting
・ Hard Disk memory capacity: Up to 999 files.
Stored images remain even after powering off.
Thumbnail display: 4, 6, 12, 16, or 24 images
・ Patient data management eStressEcho *4
Bitmap, TIFF, JPEG, and DICOM static frame image Image acquisition method
5
Measurement and Analysis: Pombo (wall), Teichholz (wall)
Gibson (wall)
● General measurements
Mitral valve measurements
On B-mode image
LA/Ao measurements
Distance, Dist-trace
Tricuspid valve measurements
Area and Circumference (Trace, Ellipse, Circle)
Pulmonary valve measurements
Volume (Spheroidal, Prolate, Area-length,
IVC measurements
Simpson)
Doppler mode
B. Index,
LV outflow tract flow
Histogram
RV outflow tract flow (right ventricle outflow tract)
Hip joint, Angle
Trans-mitral flow, Pulmonary vein flow
On M-mode image
Regurgitant flow (AR, PR, MR, TR)
Velocity
Stenotic flow (AS, PS, MS, TS)
M. Length
IMP (Index of myocardial performance): option
Time interval
B(Flow)/D mode
Heart rate
Severity evaluation of regurgitant flow (PISA
M. Index
method)
On spectral Doppler
Time interval, Heart rate, Velocity, Acceleration, ● Vascular analysis
Resistance index, Pulsatility index Carotid artery:
Pressure half time CCA (common carotid artery)
D. Caliper, D. Index, ICA (internal carotid artery)
Mean flow velocity, Pressure gradient ECA (external carotid artery)
Stenotic Flow, Regurgitant Flow BIFUR (Bifurcation of carotid artery)
D. Trace VERT (Vertebral artery)
・ Doppler auto trace: Possible on frozen image Stenotic rate
On B/D mode image IMT
Flow volume, Flow volume (SV/CO) Max-IMT (Intima-media thickness)
On B(Flow) mode image Mean-IMT (mean IMT of CCA)
Flow profile (option) Measurements on arteries in extremities:
Lower extremity artery,
● Obstetrical measurements & calculations
Upper extremity artery
Gestational age
Stenotic rate
Fetal weight
Measurements on veins in extremities:
Fetal Doppler measurements
Lower extremity venous
Fetal cardiac function measurements
Upper extremity venous
AFI (Amniotic fluid index) and volume
● Urological measurements & calculations
Cervical length
Prostate volume (PSA volume, PRS Slice volume)
Growth analysis function (display of past
Seminal vesicle
measurement data)
Bladder volume
Compatible to multiple pregnancy
Testicle volume
● Gynecological measurements & calculations
Renal volume
Uterus measurements
Cortical thickness
Endometrial Thickness
Adrenal volume
Cervical measurements
Renal artery Doppler (pulsatility index, resistance
Ovary measurements
index)
Follicular measurements
Urinary bladder measurements
● Abdominal measurements
Gynecological Doppler measurements
B mode
Uterine artery, Ovarian artery measurements
Gall bladder
● Cardiac analysis (Optional SOP-ALPHA5-3 is necessary)
Common bile duct
B mode
Liver
Area-length, Bi-plane ellipse, Modified Simpson,
Pancreas
Bullet
Kidney
Others: AVA, MVA, LA/Ao, Ratio, RVD, LVMass (AL)
Spleen
IVC (inferior vena cava) measurements
SOL (Space Occupying Lesion)
M mode
Vessel diameter (aorta, portal vein)
6
Stenotic rate (diameter, area)
Doppler mode
Abdominal aortic flow
Renal flow
Pulmonary vein flow
Shunt flow
B/D mode
Flow volume
● Report Functions
・ Obstetrical report
・ Gynecological report
・ Cardiac function report
・ Vascular report
・ IMT report
・ Urological report
・ Abdominal measurements report
It is possible to recall past measurement report.
Examination data history of each report is possible.
Direct printout of each report is possible with optional
local printer.
Measurement values can be output as a CSV file.
● Hot Key function; It is possible to assign measuring
functions to the alphabet keys on the keyboard
● Measurement on VCR playback image: Possible
● User’s calculation: Possible
7
System Configuration
Standard Components
Standard components and optional items depend on the country.
Main unit with 15-inch color viewing monitor *1 Aloka recommend UST-9123 for general abdominal imaging and
Probe * 1 UST-5299 for general cardiac examination.
Independent Probe connection unit: EU-9094B NOTE: To install EU-9094B, UCW-4000B is necessary.
When EU-9094B is installed, SCU-4000B cannot be installed.
Mechanical radial probe connection unit: SCU-4000B
NTSC(not CE marked), 100-120V: PAL (CE marked), 220-240V: For details of print papers, please ask
CP900UM (Mitsubishi) CP900E (Mitsubishi) your local distributor of the printer.
B/W printer
UP-895MD/SYN, Worldwide (SONY), available in Japan Print paper: UPP-110HG
8
OPTIONAL PROBES
Electronic convex sector probes T.H.E.: Tissue Harmonic Echo, C.H.E.: Contrast Harmonic Echo
Application Ultrasound Frequencies (MHz) Scanning Radius of
(description)
Model number angle curvature Optional accessories
B and M modes Doppler/Flow (degrees) (mmR)
10
Electronic phased array sector probes H.E.: Harmonic Echo
Application Model number Ultrasound Frequencies (MHz) Scanning Optional
angle accessories
B and M Doppler & Flow
(degrees)
Adult heart UST-5299 2.1/2.5/3.0/3.8 PW: 2.1/2.5/3.0/3.8, CW: 2.1 90 —
(Harmonic Echo) H.E.: 1.9
Pediatric heart UST-5298 3.8/5.0/6.0/7.5 PW: 3.0/3.8/5.0/6.0, CW: 3.8 90 —
Rotary plane T.E.E. UST-5293-5 3.8/5.0/6.0/7.5 PW: 3.0/3.8/5.0/6.0, CW: 3.8 90 —
Burr hole UST-5268P-5 3.8/5.0/6.0/7.5 PW: 3.0/3.8/5.0/6.0 90 Puncture adapter
neurosurgery MP-2450 is included
as standard.
End-fire UST-52109 3.8/5.0/6.0/7.5 PW: 3.0/3.8/5.0/6.0 90
laparoscopic probe
* Optional UCW-4000B-2 is necessary for CW Doppler.
3D probes
NOTE: Optional Real-time 3D kit is necessary to connect the 3D probes.
Description Model number Ultrasound Frequencies (MHz) Scanning Radius of Optional accessories
angle curvature (Probe cover)
B and M Doppler & Flow
(degrees) (mm)
Transvaginal ASU-1003 3.8/5.0/6.0/7.5 3.0/3.8/5.0/6.0 163 9 RB-945BP-S (sterilized)*1
RB-945BP-NS
(nonsterilized)
Transabdominal ASU-1009 3.8/5.0/7.5/10.0 2.1/2.5/3.0/3.8 60 40 —
H.E.: 2.1P/2.1S
2.5R/2.5H
*1 Sterilized probe cover cannot be sold in EU member countries.
*1 PM-4000-10 is necessary.
Note: Available models differ according to the country. Please ask your local distributor for the details.
11