Diagnostic Ultrasound System
2
B-mode Spectral Doppler:
・ Display: Power spectrum
・ Imaging Gray Scale: 256 levels
・ Scanning area: ・ Real-time Doppler Auto Trace
100% to 25%, continuously changeable ・ Doppler methods:
・ Image Zoom PW (Pulsed Wave) Doppler
Write zoom (magnification of real-time image): HPRF (High Pulse Repetition Frequency) PW
Up to X6 Doppler
Read zoom* (magnification of frozen image):
CW (Continuous Wave) Doppler (Probe
Varies in 16 steps at maximum
dependent)
・ Depth range selections:
2 to 24 cm*, changeable by minimum 1 cm ・ Reference frequency (Probe dependent):
step (* Probe dependent) PW: 2.14, 2.5, 3, 3.75, 5, 6, and 7.5 MHz
・ Image frequency select: 4 selections CW: 2.14 and 3 MHz
Resolution, Standard 1 & 2, Penetration ・ Pulse repetition frequency on PW Doppler:
・ Scanning line density: PW: 0.5 to 20 kHz
3 selections
・ Max. velocity range:
・ Gain*: 30 to 90 dB, continuously variable
・ STC (sensitivity time control) PW/HPRF PW: +5.6 or –5.6 m/sec
Gain versus depth curve control: 8 slide (2.14 MHz reference freq., 0 degree, with base
controls line shift)
・ Angle gain: CW: +15.92 or –15.92 m/sec
Gain versus angle curve control: 8 slide (2.14 MHz reference freq., 0 degree, with base
controls (Available on phased array sector
line shift)
image)
・ Base line shift: Available up to double velocity
・ Contrast: 16 steps
・ AGC—Suppression of brightness saturation: ・ Steerable CW Doppler: Possible on phased array
16 steps sector probes
・ Relief: 4 selections (Off + 3 levels) ・ Steered linear scanning: Possible with linear
・ Frame correlation: Continuously variable probes
・ Horizontal smoothing: 3 selections (Off + 2 levels) Steerable angle: Max. ±30 degrees , 1 or 5
・ Post Processing
degrees step changeable
Echo enhance curve: 3 kinds
Rejection: 64 steps ・ Spectrum inversion: Possible
・ View gamma: 5 kinds ・ Angle correction:
・ Image direction Available up to 80 degrees presetable
Longitudinal and lateral inversion ・ Sample volume size for PW Doppler:
90 degrees rotation 0.5 - 10 mm, changeable in 0.5 mm step
* Gain and image magnification (Read zoom) can be
・ Wall motion filter:
changed also after freezing.
Manual:
M-mode: 50, 100, 200, 400, 800 or 1600 Hz,
・ Display method: Moving bar PW and CW independently selectable
・ Seep speed*: 1, 1.5, 2, 3, 4, 6, or 8 sec/screen
Auto: varies in 12 steps for a velocity range
・ Gain*: B-gain ±30 dB, continuously variable
10 Hz – 1500 Hz,
・ Contrast: 16 steps
・ AGC—Suppression of brightness saturation: ・ Doppler gain: up to 50 dB, 1 dB step changeable
16 steps ・ Contrast: 16 steps, selectable
・ Relief: 4 selections (Off + 3 levels) ・ Black/white inversion: Possible
・ Smoothing processing: On/off selection ・ Audio output: 2 channels stereo
・ Window function—zoom in on the region of interest
・ Free Angular M-mode (FAM)— Up to 3 M-mode
cursors can be set omni-directionally on real-time
B-mode image. (Option)
* Gain and sweep speed (time/screen) can be
changed also after freezing.
3
Color Flow Imaging Cine memory
Quadruple split-screen display: possible
・ Represented parameters:
Memory capacity:
PowerFlow mode: Total power
・ Cine memory (B mode):
Color Doppler mode: Mean Doppler frequency shift
Max. 1024 images
・ Color area size: Variable from 100% to 5%
—loop display possible
・ Steered linear scanning: Possible with linear
—ECG time phase display possible
probes
・ Store memory (B and/or M, or D mode)
Steerable angle: Max. ±30 degrees, 1 or 5 degrees
Max. 100 images (The upper limit varies
step changeable
depending on ECG ON/OFF and other
・ Color enhance: 5 selections
conditions.)
Color line density: 3 selections
・ Scrolling Memory (M or D mode)
・ Image frequency select: 4 selections
Max. 1024 seconds
Resolution, Standard 1 & 2, Penetration
Note: The number of storable images depends on
・ Color smoothing: 7 selections
probe type, scanning angle and other conditions.
・ Flow filter: 6 selections
・ Frame correlation: 4 selections
・ Motion Reject: 4 selections
・ Sampling average rate: 3 steps selectable
・ Color coding:
Abdomen: 5 selections
PV: 5 selections
Cardiology: 6 selections
Variance: 1 kind
PowerFlow: 5 selections
PowerFlow
・ Gradation: 32 levels
・ Color coding: 5 selections
・ Background ON/OFF: Possible
・ Color resolution: 4 selections
Directional PowerFlow
Possible
Color Doppler
・ Reference frequency:
2.14, 2.5, 3, 3.75, 5, 6 and 7.5 MHz
・ Pulse repetition frequency:
0.5 to 10.0 kHz
・ Max. velocity range:
+1.23 or –1.23 m/sec (at 2.14 MHz reference
freq.)
・ Color base line shift:
Available up to double velocity (±31 steps)
・ Gradation:
±32 levels for velocity (red and blue)
16 levels for variance (green)
・ Color polarity invert: Possible
Spatial Flow*
・ Pseudo 3 dimensional display of blood flow image
* Optional
4
Measurement and Analysis: Aortic valve
Tricuspid valve
● Basic measurements
pulmonary valve
On B-mode image
Doppler mode
Distance
LV in-flow study
Area and Circumference by ellipse/by trace
LV ejection-flow study 1, 2
Angle
RV in-flow study
Volume
RV ejection-flow study 1, 2
Histogram
SV/CO
PSA density
Flow volume
Hip joint angle
On M-mode image
● Peripheral vessels analysis
Velocity
CCA (common carotid artery)
Time interval
ICA (internal carotid artery)
Heart rate
ECA (external carotid artery)
On spectral Doppler
Vertebral artery
Velocity
% Stenosis area
Acceleration
% Stenosis distance
Average flow velocity
S/D ratio
Time interval
Resistance index
● Report Functions
Pulsatility index
・ Obstetrical report
Pressure gradient
・ Cardiac report
Pressure half time
・ PV report
Heart rate
・ Doppler auto trace: Possible on real-time image
● Post measurement on VCR playback image:
Available with auto and manual calibration.
Others
Stenotic rate, ratio
● Cardiac analysis
Left ventricle analysis:
M mode
Pombo, Pombo (wall)
Teichholz, Teichholz (wall)
Gibson, Gibson (wall)
B mode
Pombo, Pombo (wall)
Teichholz, Teichholz (wall)
Gibson, Gibson (wall)
Single plane ellipsoid
Bi-plane ellipsoid
Bullet
Modified Simpson
Simpson
Measurements:
B mode
Long axis view
Long axis view with wall thickness
Short axis view
Apex view
M mode
Mitral valve
5
Optional Functions
Contrast Harmonic Echo Volume Mode*
In Contrast Harmonic Echo, by adding the ・ Scanning method:
phase-shift component, which is specifically Electronic convex sector & mechanical sector
6
eDMS (Extended Data Management Subsystem): ・ Network interface: 10 BASE/T or 100 BASE/TX,
Model DAS-5500-1Q/ DAS-5500-1VQ (1 of 2) automatically switched.)
Basic specifications ・ Maximum data transfer rate: 10 Mbps or 100 Mbps,
● General specifications depending on the network interface
1. Data acquisition and storage ・ Output file format:
1-1. Multiple-frame image MO disk: Bitmap, TIFF, JPEG, AVI, DICOM
RF (Radio frequency) data (Option) CD-R: DICOM
Line data Others: DICOM
Image data (Pallet, RGB or Monochrome) ・ Data attendant to the image (Automatically
1-2. Single-frame image transferred from ultrasound main unit):
Image data (Pallet, RGB or Monochrome) Conforms to DICOM standard
Compressed image data: RLE Run Length ・ Data compression: Lossless compression
Encoding) and JPEG ・ Image degradation: 0%
1-3. Measurement data ・ Film Box function ― thumbnail display of stored
Those measured on the diagnostic ultrasound images and print format editing
system main unit and eDMS ・ Print Queue function ― printout of stored images
1-4. Patient data according to the edited format
Displayed information*
Patient information: ID (up to 64 characters), 3. DICOM network communication
Name (up to 64 characters), Birthday, Sex ・ Conformity to DICOM standard:
Study information: Study ID, Age, Height, Service class user of storage, print, query/retrieve,
Weight, Accession, Referring physician, Study and basic worklist management.
description, Operator (For details, please refer to the DICOM
Series information: Application Conformance Statement issued by Aloka.)
* Conforms to DICOM 3.0 standard ・ Storage media:
1-5. Memory capacity for multiple-frame image: Built-in hard disk, FD, MO, CD-R
RF (16bit): Max.650 frames (100beams/frame) DICOM file server via Ethernet
RF (24bit): Max.325 frames (100beams/frame) ・ Print: Directly printout images with DICOM
Line (B): Max.12000 frames (100lines/frame) compatible printer
Work list management: Retrieval of patient and
Line (16bit): Max.6000 frames (100lines/frame)
reservation information from hospital information
Line (24bit): Max.3000 frames (100lines/frame) system (HIS)
Image (Palette): Max.1000 frames NOTE: The HIS needs to be compatible with DICOM
standard supplement 10. The HIS network and the
Image (RGB): Max.500 frames
DICOM network need to be linked.
Image (Monochrome): Max.1000 frames
・ Router setting
1-6. Acquisition mode:
・ MPPS (Modality Performed Procedure Step):
Post manual: user can decide acquisition start
Option (SOP-5500-26)
and stop timing
Pre manual: user can select acquisition data
4. Image View functions
after freezing
4-1 Image viewer
Post ECG: ECG synchronized data acquisition
Zoom
after pressing ACQUIRE switch (Max. 10 cardiac
Gain and contrast: changeable
cycles (R-R))
Storable image formats:
Pre ECG: ECG synchronized data acquisition
Still images: JPEG, Bitmap, and TIFF
before pressing ACQUIRE switch
Multiple-frame images: AVI
Post TIME: fixed-time data acquisition after
Frame rate: changeable
pressing ACQUIRE switch(Max. 9s)
Image rotation/inversion
Pre TIME: fixed time data acquisition before
Number of thumbnail images: 1 – 24 images
pressing ACQUIRE switch
4-2 Line Viewer
1-7. Hard disk capacity
Zoom
Total space for data: 36 GB
Gain and contrast: changeable
Color map selection: possible
2. Image Export
Image rotation, Image inversion: possible
・ Output medium:
Storable image formats:
Floppy disk
Still images: DICOM (RGB)
MO disk (128 Mbyte/230 Mbyte/540 Mbyte)
Multiple-frame images: DICOM (RGB)
CD-R (640 Mbyte)
Number of thumbnail images: 1 – 16 images
DICOM file server through Ethernet
7
eDMS (2 of 2) ● eTRACKING (Echo Tracking)
Optional analysis functions It is possible to precisely measure displacement of
● Contrast Echo analysis blood vessel to obtain indexes of stiffness of the
・ Image Subtraction vessels such as pressure-strain elastic modulus (Ep),
Fixed Reference: Subtraction of reference frame stiffness parameter (β), arterial compliance (AC),
from all frames one-point pulse wave velocity (PWVβ), and
Any 2 Frame: Subtraction between 2 selected augmentation index (AI).
frames FMD (Flow Mediated Dilatation) analysis is possible
Display modes: All images, arbitrary images by using the eTRACKING function.
Time-Intensity Curve display for subtraction
images: available ●BETA (Backscattered Energy Temporal Analysis)
・ Time-Intensity Curve Analysis
Series: Graphic display in frame sequence or ● FAM (Free Angular M-mode)
time sequence Omni-directional and arbitrary-position setting of
By Group: Graphic display with the time of one M-mode cursors (up to 3) on B-mode image
sequence of intermittent acquisition as the
horizontal scale (Graphs of multiple sequences ● KI (Kinetic Imaging)/A-SMA (Automated segmental
are overlapped.) motion analysis)
Display mode: Image, Graph Regional wall motion quantification
ROI type: Square, Draw, Arc, and Circle
* CSV output of analyzed data is possible. CSV is ● IMP (Index of Myocardial Performance)
a file format that can be taken into Excel file The IMP is an index for judging cardiac function in
directly. both systole and diastole.
8
DMS (Data Management Subsystem):
Model DMS-5500C Built-in digital interface option:
* Option (Note: Only one of eDMS or DMS can be <DICOM network communication>*4
installed in a unit. ・ Conformity to DICOM standard:
Service class user of storage, print , and basic work
<Memory> list management. (For details, please refer to the
● Cine Memory
DICOM Conformance Statement issued by Aloka.)
・ Display format: ・ Storage: Directly transmits image with patient
Single (Full screen), Dual (2-split screen) and information to a DICOM file server
Quadruple (4-split screen) display ・ Print: Directly printout images with DICOM
・ Loop image playback: compatible printer
Possible independently on each split screen ・ Work list management: Retrieval of patient and
Playback speed changeable reservation information from hospital information
・ Play-through function―simultaneous display of live system (HIS)
NOTE: The HIS needs to be compatible with DICOM
image and cine memory image
・ ECG synchronized image acquisition: possible standard supplement 10. The HIS network and the
DICOM network need to be linked.
・ Cine Memory capacity:
NTSC ・ Router setting
Max. 256 images (B/W) or 128 images (color)
PAL <Image processing>
●3-D image processing
Max. 224 images (B/W) or 112 images (color)
● Non-volatile Memory
・ Volume Mode―photo-realistic display suitable for
・ Hard disk capacity: imaging fetus
Up to about 1000 images ・ Spatial Echo―pseudo 3 dimensional display of
・ Film Box function―thumbnail display of stored gray-scale image
images and print format editing ・ Spatial Flow―pseudo 3 dimensional display of
・ Print Queue function―printout of stored images blood flow image
directly with DICOM compatible printer ・ Spatial Color―pseudo 3 dimensional display of
gray-scale and blood flow images
<Digital image transmission> ・ 3D Slicing― display of planes viewed from the side
・ Data compression: Lossless compression and the bottom
●Kinetic Imaging
・ Image degradation: 0%
・ Output medium: Displays the movement of the endocardial wall with
MO disk (128 Mbyte/230 Mbyte/540 Mbyte) color gradation with time
Ether net (Network interface: 10 BASE/T or 100
BASE/TX, automatically switched.) <Stress Echo Package>
Max. data transfer rate: 10 Mbps or 100 Mbps, ・ Stress scoring*3
according to the network interface) ・ Quad-screen display
・ Output file format: BitMap and TIFF ・ Exercise stress protocols:
・ Data attendant to the image (Automatically 2 phases (pre-exercise/post-exercise)
transferred from ultrasound main unit): 4 phases (pre-exercise/peak
Patient information exercise/post-exercise/recovery)
Study information ・ Pharmacological stress protocols:
Display mode BLPR (Baseline/low dose/ peak dose/ recovery)
Scale factor BLMP (Baseline/low dose/ mid dose/ peak dose)
Baseline position ・ User's protocol: The user can make a protocol in
Look-up Table for color which 8 stages/phases can be set at maximum.
9
System Configuration
Standard Components
Main unit with 15-inch color viewing monitor Standard components and optional items depend on the country.
1
Probe *1 * Aloka recommend UST-9126 for general abdominal imaging and
Accessories UST-52101 for general cardiac examination.
Optional Functions
CQ (Cardiac Quantification) A-SMA (Automated Segmental Motion Analysis) & KI (Kinetic Imaging): EU-9078CQ
(Physiological signal display unit PEU-5500Q is also necessary.)
Optional Module/Unit
Physiological signal display unit (ECG, PCG, and Pulse wave): PEU-5500Q Pulse wave transducer:
Includes ECG lead cords (for clip type electrode and disposable electrode) and a PCG microphone. TY-307A
NOTE: Either one of eDMS or DMS can be installed. SOP-DMS-1: DICOM network communication
Both cannot be installed at the same time.
PM-DMS-10: BETA (Backscattered Energy Temporal Analysis)
DMS (Data Management Subsystem): PM-DMS-15: OB/GYN report data management
DMS-5500C
PM-DMS-16: Stress Echo scoring and report data management
Optional Recorders
MO Disk Drive
B/W printer, worldwide spec. (available in Japan): UP-895MD/SYN Print paper: UPP-110HG
B/W printer (available outside Japan)
NTSC (100-120V): UP-895MD (SONY)
PAL (200-240V): UP-895CE (SONY) Color NTSC/PAL (not CE marked), Worldwide voltage:
printer UP-21MD(UC) (SONY) Color print pack:
RGB cable: L-CABLE-342-20 UPC-21S (small)
NTSC/PAL (CE marked), Worldwide voltage: UPC-21L (larege)
Slide adapter: UPA-21SA UP-21MD(CED) (SONY)
10
Optional Probes
HST (Hemispheric Sound Technology) probes
Convex sector probes *1
B: B-mode
PW: Pulsed Wave Doppler
Application Ultrasound Scanning Radius of Model F: Flow (Color flow)
frequency (MHz) *1 angle curvature BF: Flow in B-mode
(Feature) CW: Continuous wave Doppler
H.E.: Harmonic Echo
General abdomen, B: 3.0/3.75/5.0/6.0 60° 60 mmR UST-9126
OB/GYN PW/F: 2.14/2.5/3.0/3.75 Puncture adapter:
T.H.E.: 2.14 MP-2473
C.H.E.: B, BF: 1.88/2.14
General abdomen, B: 3.75/5.0/6.0/7.5 60° 60 mmR UST-9115-5
OB/GYN (High resolution) PW/F:3.0/3.75/5.0/6.0
Puncture adapter:
Endocavity B: 3.75/5.0/6.0/7.5 MP-2748-SET
180° 9 mmR UST-9118
(Transvaginal) PW/F: 3.0/3.75/5.0/6.0 Probe cover*2:
RB-945BP-NS, RB-945BP-S
Small part, Neonatal B: 5.0/6.0/7.5/10.0 70° 20 mmR UST-9120
head PW/F:5.0/6.0 Puncture adapter: MP-2458
*3 To connect UST-52104
Phased array sector probes connector adapter JB-266
is necessary.
Scanning
Application (Feature) Ultrasound frequency (MHz) *1 angle Model
Cardiology (Motorized B: 3.75/5.0/6.0/7.5, T.H.E.:3.0/3.75 90° UST-52104 *3
Transesophageal probe) PW/F: 3.0/3.75/5.0/6.0, CW: 3.75
Cardiology B: 2.5/3.0/3.75/5.0, 90° UST-52101
PW/F: 2.14/2.5/3.0/3.75, CW: 2.14
T.H.E.: 1.88
C.H.E.: B:1.88/2.14, BF:1.88,2.5
Pediatric/neonatal cardiology B: 3.75/5.0/6.0/7.5 90°
PW/F: 3.0/3.75/5.0/6.0, CW: 3.75 UST-5294-5
Transesophageal exam B: 3.75/5.0/6.0/7.5 90°
PW/F: 3.0/3.75/5.0/6.0, CW: 3.75 UST-5293-5
11
W-SHD (Wide-band Super High Density) probes and SHD probes
Linear probes
Scanning
Application Ultrasound frequency (MHz) Model
width
Peripheral vessels*4 B: 3.75/5.0/6.0/7.5 42 mm UST-5548
(Harmonic Echo) PW/F: 3.0/3.75/5.0/6.0 Echo coupler (Flat) : ECM-12
T.H.E.: 3.75
Intraoperative *4 B: 5.0/6.0/7.5/10.0 42 mm UST-5534T-7.5
PW/F: 5.0/6.0
Laparoscopic probe B: 5.0/6.0/7.5/10.0 33 mm UST-5536-7.5 Trocar (adaptable size: 12 mm inner dia.)*
5
(Flexible type) PW/F: 5.0/6.0
3D probes
Description Model Ultrasound Frequencies (MHz) Convex Sector Mechanica
l sector Optional accessories
number B and M Doppler & Flow Scanning Radius of Scanning
angle curvature angle
(degrees) (mmR) (degrees)
Transvaginal ASU-1002 3.75/5.0/6.0/7.5 3.0/3.75/5.0/6.0 180 9 90 Probe cover:
RB-945BP-S (sterilized)*
RB-945BP-NS (nonsterilized)
Transabdomi ASU-1009 3.75/5.0/7.5/10.0 2.1/2.5/3.0/3.75 60 40 60 —
nal H.E.: 2.14
* Sterilized probe cover cannot be sold in EU member countries.
12
SHD (Super High Density) probes
13
General Specifications
Acoustic Power
Viewing Monitor
・ 0 to 100%, changeable ・ 15-inch diagonal flicker-free non-interlaced monitor
Preset Function ・ Tilt, swivel, height adjustment are possible.
・ Composite, RGB, and Y/C separated signals, 1 ・ 60 cm (W) × 98 cm (D) × 132 –157 cm (H)
14
15
・ The specifications are subject to change without notice.