SPECIFICATIONS

for
Diagnostic Ultrasound System

A Platform for Extended Pure Harmonic Detection

THE INNOVATOR IN ULTRASOUND


SV stands for . . . Scanning Method

Superb Value ・ Electronic Convex Sector


・ Electronic Linear
Superb Versatility
・ Electronic Phased Array Sector
Superb Vascular analysis ・ Mechanical Sector/Radial

Image Display Modes


The SSD-5500SV provides you with superb
・ B: gray-scale imaging
values through high image quality and high
・ D: Spectral Doppler (PW, HPRF PW, and CW)
performance. The technically integrated
・ Dual B
platform offers superb versatility and superb
・ B and M
vascular analysis functions.
・ B and D
・ B, M, and D
Advanced Technologies
・ B(Flow)
・ Extended Pure Harmonic Detection Flow: Color Doppler and PowerFlow imaging
It is possible to separately detect phase-shift ・ Dual B(Flow)
components, harmonic components, attenuation ・ M(Flow)
and backscattering components. By combining ・ B(Flow) and M(Flow)
these components, Extended Pure Harmonic ・ B(Flow) and D
Detection offers superior sensitivity for both Tissue ・ Triplex mode: B, Flow, and PW Doppler
and Contrast Harmonic Echo. simultaneous real-time display
・ Harmonic Echo TM ・ B(Flow), M(Flow) and D
The Harmonic Echo is the technology to construct ・ Dual Dynamic Display (DDD): B and B(Flow)
the image by using the 2nd harmonic components. simultaneous real-time display

The Harmonic Echo provides clearer edge ・ Intermittent trigger mode


definition, reduced side lobe artifacts, and less ・ KI (Kinetic imaging)* — included in the A-SMA
reverberation noise compared to the conventional function*

fundamental frequency imaging. ・ TDI (Tissue Doppler Imaging)*


Available on phased array probes.
・ HST (Hemispheric Sound Technology)
・ Volume Mode*
Forms ideal ultrasound beam shape with less side
* Option
lobe artifacts both in array direction and elevation
Request function: In multi-display mode, it is possible
direction.
to select one mode for full screen display.
・ Impedance Matching Technology
Prevents signal loss by total impedance matching Beamformer
from transducer to the front-end.
12-bit Digital Beam Former
・ New-generation Wide Dynamic Range Digital
Front-end System Dynamic Range
The 12-bit A/D converter provides superior
182 dB
contrast resolution compared to conventional
digital systems and analog systems.
Frame Rate
・ QFI (Quint Frequency Imaging)
Max. 1000 frames/sec or more*
You can select ultrasound frequencies from among
*Probe and system setting dependent
up to 5 selections with one probe. The QFI can
(1,493 frames/sec in the case of UST-5531)
provide better resolution with a higher frequency
and better penetration with a lower frequency.
(Selectable frequencies are probe dependent.)

2
B-mode Spectral Doppler:
・ Display: Power spectrum
・ Imaging Gray Scale: 256 levels
・ Scanning area: ・ Real-time Doppler Auto Trace
100% to 25%, continuously changeable ・ Doppler methods:
・ Image Zoom PW (Pulsed Wave) Doppler
Write zoom (magnification of real-time image): HPRF (High Pulse Repetition Frequency) PW
Up to X6 Doppler
Read zoom* (magnification of frozen image):
CW (Continuous Wave) Doppler (Probe
Varies in 16 steps at maximum
dependent)
・ Depth range selections:
2 to 24 cm*, changeable by minimum 1 cm ・ Reference frequency (Probe dependent):
step (* Probe dependent) PW: 2.14, 2.5, 3, 3.75, 5, 6, and 7.5 MHz
・ Image frequency select: 4 selections CW: 2.14 and 3 MHz
Resolution, Standard 1 & 2, Penetration ・ Pulse repetition frequency on PW Doppler:
・ Scanning line density: PW: 0.5 to 20 kHz
3 selections
・ Max. velocity range:
・ Gain*: 30 to 90 dB, continuously variable
・ STC (sensitivity time control) PW/HPRF PW: +5.6 or –5.6 m/sec
Gain versus depth curve control: 8 slide (2.14 MHz reference freq., 0 degree, with base
controls line shift)
・ Angle gain: CW: +15.92 or –15.92 m/sec
Gain versus angle curve control: 8 slide (2.14 MHz reference freq., 0 degree, with base
controls (Available on phased array sector
line shift)
image)
・ Base line shift: Available up to double velocity
・ Contrast: 16 steps
・ AGC—Suppression of brightness saturation: ・ Steerable CW Doppler: Possible on phased array
16 steps sector probes
・ Relief: 4 selections (Off + 3 levels) ・ Steered linear scanning: Possible with linear
・ Frame correlation: Continuously variable probes
・ Horizontal smoothing: 3 selections (Off + 2 levels) Steerable angle: Max. ±30 degrees , 1 or 5
・ Post Processing
degrees step changeable
Echo enhance curve: 3 kinds
Rejection: 64 steps ・ Spectrum inversion: Possible
・ View gamma: 5 kinds ・ Angle correction:
・ Image direction Available up to 80 degrees presetable
Longitudinal and lateral inversion ・ Sample volume size for PW Doppler:
90 degrees rotation 0.5 - 10 mm, changeable in 0.5 mm step
* Gain and image magnification (Read zoom) can be
・ Wall motion filter:
changed also after freezing.
Manual:
M-mode: 50, 100, 200, 400, 800 or 1600 Hz,
・ Display method: Moving bar PW and CW independently selectable
・ Seep speed*: 1, 1.5, 2, 3, 4, 6, or 8 sec/screen
Auto: varies in 12 steps for a velocity range
・ Gain*: B-gain ±30 dB, continuously variable
10 Hz – 1500 Hz,
・ Contrast: 16 steps
・ AGC—Suppression of brightness saturation: ・ Doppler gain: up to 50 dB, 1 dB step changeable
16 steps ・ Contrast: 16 steps, selectable
・ Relief: 4 selections (Off + 3 levels) ・ Black/white inversion: Possible
・ Smoothing processing: On/off selection ・ Audio output: 2 channels stereo
・ Window function—zoom in on the region of interest
・ Free Angular M-mode (FAM)— Up to 3 M-mode
cursors can be set omni-directionally on real-time
B-mode image. (Option)
* Gain and sweep speed (time/screen) can be
changed also after freezing.

3
Color Flow Imaging Cine memory
Quadruple split-screen display: possible
・ Represented parameters:
Memory capacity:
PowerFlow mode: Total power
・ Cine memory (B mode):
Color Doppler mode: Mean Doppler frequency shift
Max. 1024 images
・ Color area size: Variable from 100% to 5%
—loop display possible
・ Steered linear scanning: Possible with linear
—ECG time phase display possible
probes
・ Store memory (B and/or M, or D mode)
Steerable angle: Max. ±30 degrees, 1 or 5 degrees
Max. 100 images (The upper limit varies
step changeable
depending on ECG ON/OFF and other
・ Color enhance: 5 selections
conditions.)
Color line density: 3 selections
・ Scrolling Memory (M or D mode)
・ Image frequency select: 4 selections
Max. 1024 seconds
Resolution, Standard 1 & 2, Penetration
Note: The number of storable images depends on
・ Color smoothing: 7 selections
probe type, scanning angle and other conditions.
・ Flow filter: 6 selections
・ Frame correlation: 4 selections
・ Motion Reject: 4 selections
・ Sampling average rate: 3 steps selectable
・ Color coding:
Abdomen: 5 selections
PV: 5 selections
Cardiology: 6 selections
Variance: 1 kind
PowerFlow: 5 selections
PowerFlow
・ Gradation: 32 levels
・ Color coding: 5 selections
・ Background ON/OFF: Possible
・ Color resolution: 4 selections
Directional PowerFlow
Possible
Color Doppler
・ Reference frequency:
2.14, 2.5, 3, 3.75, 5, 6 and 7.5 MHz
・ Pulse repetition frequency:
0.5 to 10.0 kHz
・ Max. velocity range:
+1.23 or –1.23 m/sec (at 2.14 MHz reference
freq.)
・ Color base line shift:
Available up to double velocity (±31 steps)
・ Gradation:
±32 levels for velocity (red and blue)
16 levels for variance (green)
・ Color polarity invert: Possible

Spatial Flow*
・ Pseudo 3 dimensional display of blood flow image
* Optional

4
Measurement and Analysis: Aortic valve
Tricuspid valve
● Basic measurements
pulmonary valve
On B-mode image
Doppler mode
Distance
LV in-flow study
Area and Circumference by ellipse/by trace
LV ejection-flow study 1, 2
Angle
RV in-flow study
Volume
RV ejection-flow study 1, 2
Histogram
SV/CO
PSA density
Flow volume
Hip joint angle
On M-mode image
● Peripheral vessels analysis
Velocity
CCA (common carotid artery)
Time interval
ICA (internal carotid artery)
Heart rate
ECA (external carotid artery)
On spectral Doppler
Vertebral artery
Velocity
% Stenosis area
Acceleration
% Stenosis distance
Average flow velocity
S/D ratio
Time interval
Resistance index
● Report Functions
Pulsatility index
・ Obstetrical report
Pressure gradient
・ Cardiac report
Pressure half time
・ PV report
Heart rate
・ Doppler auto trace: Possible on real-time image
● Post measurement on VCR playback image:
Available with auto and manual calibration.
Others
Stenotic rate, ratio

● Cardiac analysis
Left ventricle analysis:
M mode
Pombo, Pombo (wall)
Teichholz, Teichholz (wall)
Gibson, Gibson (wall)
B mode
Pombo, Pombo (wall)
Teichholz, Teichholz (wall)
Gibson, Gibson (wall)
Single plane ellipsoid
Bi-plane ellipsoid
Bullet
Modified Simpson
Simpson
Measurements:
B mode
Long axis view
Long axis view with wall thickness
Short axis view
Apex view
M mode
Mitral valve

5
Optional Functions
Contrast Harmonic Echo Volume Mode*
In Contrast Harmonic Echo, by adding the ・ Scanning method:
phase-shift component, which is specifically Electronic convex sector & mechanical sector

produced by resonance and rupture of micro ・ Image processing


bubbles, to the harmonic component, the image of Pre-processing: Gain, Opacity, Contrast, Gamma

the contrast agent is enhanced and emerges from Post-processing: Brightness

the tissue image.


・ Image magnification (possible on frozen images):
From 0.5 to 2 times in 0.25 times increment
・ Monitor mode ・ Image rotation (possible on frozen images):
In the Monitor Mode, the system intermittently Possible by 45 degrees
transmits high-pressure ultrasound while
performing continuous transmission on lower 3D probes for Volume Mode*
sound pressure levels. ASU-1009 ASU-1002
・ Time-intensity curve display
Convex sector 60゚ 180゚
Possible (Optional eDMS is necessary.) Radius of 40 mmR 9 mmR
curvature
CQ (Cardiac Quantification) Maximum
volume rate 6 volumes/s 2 volumes/s
Displays in real time the line graph of the left Frequency 2-10 MHz 3-9MHz
range
ventricular volume, EF (Ejection Fraction) and each
* Volume Mode and 3D probes are optional.
rate of change.
A-SMA (Automated Segmental Motion Analysis):
・ On-line quantification of regional wall motion
Data Communication Function (ALK-3)*
・ Line graph display of FAC (Fractional Area Change)
It is possible to transmit patient data and OB/GYN,
and area.
cardiology, urology and peripheral vessels
・ Histogram display of FAC with KI (kinetic imaging)
measurement data obtained by Aloka diagnostic
KI (Kinetic Imaging)
ultrasound system to a personal computer.
Displays the movement of the endocardial walls
* Optional Isolation Unit model SIU-5500-1 is
with color gradation with time.
necessary. Please prepare personal computer,
WT (Wall thickness)
appropriate software, and connection cable at your
Displays in real time the line graph of the thickness
side. As for necessary conditions of such items,
of the myocardium.
please contact your local distributor or Aloka office.

TDI (Tissue Doppler Imaging) Direct Printout *


It is possible to printout report of OB/GYN and
Displays the movement of tissue with color.
urological measurement directly to an external
Velocity mode or Power mode selectable.
printer.
* Optional Isolation Unit model SIU-5500-1 is
FAM (Free Angular M-mode ) necessary. Please prepare printer and connection
・ Up to 3 M-mode cursors can be set cable at your side.
omni-directionally on real-time B-mode image. Compatible printers: An HP printer of PCL (Printer
Command Language) level 3 or higher
Physiological Signal Display Checked with HP Deskjet 890C
・ 3 channels of ECG, PCG, and Pulse wave
・ ECG lead ― second lead system
・ ECG synchronized display:
Available for one phase

6
eDMS (Extended Data Management Subsystem): ・ Network interface: 10 BASE/T or 100 BASE/TX,
Model DAS-5500-1Q/ DAS-5500-1VQ (1 of 2) automatically switched.)
Basic specifications ・ Maximum data transfer rate: 10 Mbps or 100 Mbps,
● General specifications depending on the network interface
1. Data acquisition and storage ・ Output file format:
1-1. Multiple-frame image MO disk: Bitmap, TIFF, JPEG, AVI, DICOM
RF (Radio frequency) data (Option) CD-R: DICOM
Line data Others: DICOM
Image data (Pallet, RGB or Monochrome) ・ Data attendant to the image (Automatically
1-2. Single-frame image transferred from ultrasound main unit):
Image data (Pallet, RGB or Monochrome) Conforms to DICOM standard
Compressed image data: RLE Run Length ・ Data compression: Lossless compression
Encoding) and JPEG ・ Image degradation: 0%
1-3. Measurement data ・ Film Box function ― thumbnail display of stored
Those measured on the diagnostic ultrasound images and print format editing
system main unit and eDMS ・ Print Queue function ― printout of stored images
1-4. Patient data according to the edited format
Displayed information*
Patient information: ID (up to 64 characters), 3. DICOM network communication
Name (up to 64 characters), Birthday, Sex ・ Conformity to DICOM standard:
Study information: Study ID, Age, Height, Service class user of storage, print, query/retrieve,
Weight, Accession, Referring physician, Study and basic worklist management.
description, Operator (For details, please refer to the DICOM
Series information: Application Conformance Statement issued by Aloka.)
* Conforms to DICOM 3.0 standard ・ Storage media:
1-5. Memory capacity for multiple-frame image: Built-in hard disk, FD, MO, CD-R
RF (16bit): Max.650 frames (100beams/frame) DICOM file server via Ethernet
RF (24bit): Max.325 frames (100beams/frame) ・ Print: Directly printout images with DICOM
Line (B): Max.12000 frames (100lines/frame) compatible printer
Work list management: Retrieval of patient and
Line (16bit): Max.6000 frames (100lines/frame)
reservation information from hospital information
Line (24bit): Max.3000 frames (100lines/frame) system (HIS)
Image (Palette): Max.1000 frames NOTE: The HIS needs to be compatible with DICOM
standard supplement 10. The HIS network and the
Image (RGB): Max.500 frames
DICOM network need to be linked.
Image (Monochrome): Max.1000 frames
・ Router setting
1-6. Acquisition mode:
・ MPPS (Modality Performed Procedure Step):
Post manual: user can decide acquisition start
Option (SOP-5500-26)
and stop timing
Pre manual: user can select acquisition data
4. Image View functions
after freezing
4-1 Image viewer
Post ECG: ECG synchronized data acquisition
Zoom
after pressing ACQUIRE switch (Max. 10 cardiac
Gain and contrast: changeable
cycles (R-R))
Storable image formats:
Pre ECG: ECG synchronized data acquisition
Still images: JPEG, Bitmap, and TIFF
before pressing ACQUIRE switch
Multiple-frame images: AVI
Post TIME: fixed-time data acquisition after
Frame rate: changeable
pressing ACQUIRE switch(Max. 9s)
Image rotation/inversion
Pre TIME: fixed time data acquisition before
Number of thumbnail images: 1 – 24 images
pressing ACQUIRE switch
4-2 Line Viewer
1-7. Hard disk capacity
Zoom
Total space for data: 36 GB
Gain and contrast: changeable
Color map selection: possible
2. Image Export
Image rotation, Image inversion: possible
・ Output medium:
Storable image formats:
Floppy disk
Still images: DICOM (RGB)
MO disk (128 Mbyte/230 Mbyte/540 Mbyte)
Multiple-frame images: DICOM (RGB)
CD-R (640 Mbyte)
Number of thumbnail images: 1 – 16 images
DICOM file server through Ethernet

7
eDMS (2 of 2) ● eTRACKING (Echo Tracking)
Optional analysis functions It is possible to precisely measure displacement of
● Contrast Echo analysis blood vessel to obtain indexes of stiffness of the
・ Image Subtraction vessels such as pressure-strain elastic modulus (Ep),
Fixed Reference: Subtraction of reference frame stiffness parameter (β), arterial compliance (AC),
from all frames one-point pulse wave velocity (PWVβ), and
Any 2 Frame: Subtraction between 2 selected augmentation index (AI).
frames FMD (Flow Mediated Dilatation) analysis is possible
Display modes: All images, arbitrary images by using the eTRACKING function.
Time-Intensity Curve display for subtraction
images: available ●BETA (Backscattered Energy Temporal Analysis)
・ Time-Intensity Curve Analysis
Series: Graphic display in frame sequence or ● FAM (Free Angular M-mode)
time sequence Omni-directional and arbitrary-position setting of
By Group: Graphic display with the time of one M-mode cursors (up to 3) on B-mode image
sequence of intermittent acquisition as the
horizontal scale (Graphs of multiple sequences ● KI (Kinetic Imaging)/A-SMA (Automated segmental
are overlapped.) motion analysis)
Display mode: Image, Graph Regional wall motion quantification
ROI type: Square, Draw, Arc, and Circle
* CSV output of analyzed data is possible. CSV is ● IMP (Index of Myocardial Performance)
a file format that can be taken into Excel file The IMP is an index for judging cardiac function in
directly. both systole and diastole.

● Stress Echo analysis ● TDI analysis


Image acquisition methods: B-mode
Temporal Velocity Profile
・ Arbitrary ECG synchronized images
Velocity, time, acceleration, ratio
・ Programmable Template: up to 96 matrix
Regional Velocity Profile
settings possible
Velocity, distance
・ Continuous: up to about 3 minutes, depending on M-mode
probe and other conditions Velocity trace
Edit function: Velocity, time, acceleration, ratio
Systole display (Arbitrary systole timing setting) Wall thickness
Elimination of images after systole: possible Distance, time, velocity
Playback speed: variable Velocity Profile
Image expand: possible Velocity, distance
Automatic registration: On/Off CSV output of analyzed data is possible. CSV is a
Scoring function: file format that can be taken into Excel file directly.
Playback speed: variable
Base line image comparison ● Flow profile
Image playback range selection Velocity profile display
Systole display (Arbitrary systole timing setting) Automatic boarder detection and tracking
Report function: Measurements:
Display styles: Blood vessel inner diameter
Chart/Stage overview/View overview Mean velocity
Protocol: Flow volume
・ Exercise stress protocols:
2 phases (pre-exercise/post-exercise) ● EFV (Extended Field of View)
4 phases (pre-exercise/peak Panoramic display of image in an extensive range
exercise/post-exercise/recovery) of body
・ Pharmacological stress protocols:
BLPR (Baseline/low dose/ peak dose/ recovery) ● RT-3D (Real-time 3D)
BLMP (Baseline/low dose/ mid dose/ peak dose) A 3D image and three free sections can be
・ User's protocol: The user can make a protocol in displayed at the same time (Multi-planar display).
which 8 stages/phases can be set at maximum. Omnidirectional rotation of the view point for the
・ Full disclosure: possible 3D image is possible

8
DMS (Data Management Subsystem):
Model DMS-5500C Built-in digital interface option:
* Option (Note: Only one of eDMS or DMS can be <DICOM network communication>*4
installed in a unit. ・ Conformity to DICOM standard:
Service class user of storage, print , and basic work
<Memory> list management. (For details, please refer to the
● Cine Memory
DICOM Conformance Statement issued by Aloka.)
・ Display format: ・ Storage: Directly transmits image with patient
Single (Full screen), Dual (2-split screen) and information to a DICOM file server
Quadruple (4-split screen) display ・ Print: Directly printout images with DICOM
・ Loop image playback: compatible printer
Possible independently on each split screen ・ Work list management: Retrieval of patient and
Playback speed changeable reservation information from hospital information
・ Play-through function―simultaneous display of live system (HIS)
NOTE: The HIS needs to be compatible with DICOM
image and cine memory image
・ ECG synchronized image acquisition: possible standard supplement 10. The HIS network and the
DICOM network need to be linked.
・ Cine Memory capacity:
NTSC ・ Router setting
Max. 256 images (B/W) or 128 images (color)
PAL <Image processing>
●3-D image processing
Max. 224 images (B/W) or 112 images (color)
● Non-volatile Memory
・ Volume Mode―photo-realistic display suitable for
・ Hard disk capacity: imaging fetus
Up to about 1000 images ・ Spatial Echo―pseudo 3 dimensional display of
・ Film Box function―thumbnail display of stored gray-scale image
images and print format editing ・ Spatial Flow―pseudo 3 dimensional display of
・ Print Queue function―printout of stored images blood flow image
directly with DICOM compatible printer ・ Spatial Color―pseudo 3 dimensional display of
gray-scale and blood flow images
<Digital image transmission> ・ 3D Slicing― display of planes viewed from the side
・ Data compression: Lossless compression and the bottom
●Kinetic Imaging
・ Image degradation: 0%
・ Output medium: Displays the movement of the endocardial wall with
MO disk (128 Mbyte/230 Mbyte/540 Mbyte) color gradation with time
Ether net (Network interface: 10 BASE/T or 100
BASE/TX, automatically switched.) <Stress Echo Package>
Max. data transfer rate: 10 Mbps or 100 Mbps, ・ Stress scoring*3
according to the network interface) ・ Quad-screen display
・ Output file format: BitMap and TIFF ・ Exercise stress protocols:
・ Data attendant to the image (Automatically 2 phases (pre-exercise/post-exercise)
transferred from ultrasound main unit): 4 phases (pre-exercise/peak
Patient information exercise/post-exercise/recovery)
Study information ・ Pharmacological stress protocols:
Display mode BLPR (Baseline/low dose/ peak dose/ recovery)
Scale factor BLMP (Baseline/low dose/ mid dose/ peak dose)
Baseline position ・ User's protocol: The user can make a protocol in
Look-up Table for color which 8 stages/phases can be set at maximum.

・ <Measurements and calculations> Necessary optional kits:


Those available on the main unit are also available
on the DMS. *1 PM-DMS-10
・ BETA (Backscattered Energy Temporal Analysis) *1 *2 PM-DMS-15
・ OB/GYN report data output*2 *3 PM-DMS-16
*4 SOP-DMS-1

9
System Configuration

Standard Components
Main unit with 15-inch color viewing monitor Standard components and optional items depend on the country.
1
Probe *1 * Aloka recommend UST-9126 for general abdominal imaging and
Accessories UST-52101 for general cardiac examination.

Optional Functions
CQ (Cardiac Quantification) A-SMA (Automated Segmental Motion Analysis) & KI (Kinetic Imaging): EU-9078CQ
(Physiological signal display unit PEU-5500Q is also necessary.)

Contrast Harmonic Echo *2: CHM-5500


*2 Possible with selected probes.
3D probe: ASU-1009
Volume Mode :EU-9079DQ
3D probe: ASU-1002 Probe cover*3:
(Transvaginal type) RB-945BP-NS
RB-945BP-S
*3 NS: Non-sterilized, S: Sterilized
NOTE: Sterilized probe cover cannot be
sold in EU member countries.

FAM (Free-angular M-mode): PM-5500-17


3-point footswitch MP-2614B
Flow Profile and TDI measurements (B TDI Profile,
TDI (Tissue Doppler Imaging):
MF Profile, MF Trace, M TDI Wall Thickness): Ultrasound gel (300 g):GEL-SCAN-PA
PM-5500-16
PM-5500-18 Ultrasound gel (5 lit.):GEL-SCAN-P
Index of Myocardial Performance: PM-5500-50 NOTE: Ultrasound gel cannot be sold in
EU member countries.
SpatialFlow: SOP-5500-18

Optional Module/Unit
Physiological signal display unit (ECG, PCG, and Pulse wave): PEU-5500Q Pulse wave transducer:
Includes ECG lead cords (for clip type electrode and disposable electrode) and a PCG microphone. TY-307A

FAM (Free Angular M-mode): SOP-5500-7C


Serial I/O isolation unit: SIU-5500-1
KI/A-SMA: SOP-5500-11C
eDMS (Extended Data Management Subsystem): IMP (Index of Myocardial Performance): SOP-5500-12C
DAS-5500-1Q (100-120V)
BETA: SOP-5500-13C
DAS-5500-1VQ (200-240V)
4
Contrast Echo analysis* : SOP-5500-14C
EFV (Extended Field of View): SOP-5500-22 4
* Needs also CHM-5500 for main unit).
Note: Possible with UST-5545, -5410, -5712, -9126, ,
and -9115 Stress Echo analysis: SOP-5500-15C
RT-3D (Real-time 3D): Echo Tracking: EU-9089Q + SOP-5500-16C
Volume Mode unit (EU-9079CQ or DQ), FMD: EU-9089Q + SOP-5500-25
EU-9093, and a Volume Mode probe Arm holder MP-AH0001, Probe Fixing unit MP-PH0001
(ASU-1009 or ASU-1002) are necessary.
5
TDI analysis* : SOP-5500-20C
DVD-RAM drive: PM-DAS-12 5
* Needs also PM-5500-16 (TDI kit for main unit)
Flow profile analysis: SOP-5500-21C
MPPS (Modality Performed Procedure Step): SOP-5500-26

NOTE: Either one of eDMS or DMS can be installed. SOP-DMS-1: DICOM network communication
Both cannot be installed at the same time.
PM-DMS-10: BETA (Backscattered Energy Temporal Analysis)
DMS (Data Management Subsystem): PM-DMS-15: OB/GYN report data management
DMS-5500C
PM-DMS-16: Stress Echo scoring and report data management

Optional Recorders
MO Disk Drive

Video Cassette Recorder (MITSUBISHI):


NTSC: HS-MD3000U
PAL: HS-MD3000E

B/W printer, worldwide spec. (available in Japan): UP-895MD/SYN Print paper: UPP-110HG
B/W printer (available outside Japan)
NTSC (100-120V): UP-895MD (SONY)
PAL (200-240V): UP-895CE (SONY) Color NTSC/PAL (not CE marked), Worldwide voltage:
printer UP-21MD(UC) (SONY) Color print pack:
RGB cable: L-CABLE-342-20 UPC-21S (small)
NTSC/PAL (CE marked), Worldwide voltage: UPC-21L (larege)
Slide adapter: UPA-21SA UP-21MD(CED) (SONY)

10
Optional Probes
HST (Hemispheric Sound Technology) probes
Convex sector probes *1
B: B-mode
PW: Pulsed Wave Doppler
Application Ultrasound Scanning Radius of Model F: Flow (Color flow)
frequency (MHz) *1 angle curvature BF: Flow in B-mode
(Feature) CW: Continuous wave Doppler
H.E.: Harmonic Echo
General abdomen, B: 3.0/3.75/5.0/6.0 60° 60 mmR UST-9126
OB/GYN PW/F: 2.14/2.5/3.0/3.75 Puncture adapter:
T.H.E.: 2.14 MP-2473
C.H.E.: B, BF: 1.88/2.14
General abdomen, B: 3.75/5.0/6.0/7.5 60° 60 mmR UST-9115-5
OB/GYN (High resolution) PW/F:3.0/3.75/5.0/6.0
Puncture adapter:
Endocavity B: 3.75/5.0/6.0/7.5 MP-2748-SET
180° 9 mmR UST-9118
(Transvaginal) PW/F: 3.0/3.75/5.0/6.0 Probe cover*2:
RB-945BP-NS, RB-945BP-S
Small part, Neonatal B: 5.0/6.0/7.5/10.0 70° 20 mmR UST-9120
head PW/F:5.0/6.0 Puncture adapter: MP-2458

Endocavity B: 3.75/5.0/6.0/7.5 180° 9 mmR UST-675P Probe cover*2:


(Transrectal) PW/F: 3.0/3.75/5.0/6.0 RB-665P-NS (Non-sterilized)
RB-665P-S (Sterilized)
Puncture adapter MP-2452 is attached as standard.

*2 Sterilized probe cover cannot be


Linear probes Steered linear scanning is possible. sold in EU member countries.

Application Ultrasound Scanning Model


(Feature) frequency (MHz) *1 width

Small parts, peripheral B: 5.0/7.5/10.0/13.0 34 mm UST-5410


vascular PW/F: 5.0/6.0
T.H.E.:5.0
Small parts, peripheral B: 5.0/7.5/10.0/13.0 34 mm UST-5545
vascular PW/F: 5.0/6.0
T.H.E.:5.0
Superficial tissue B: 7.5/10.0/13.0 34 mm UST-5543
PW/F: 5.0/6.0
Small parts B: 5.0/6.0/7.5/10.0 60 mm UST-5712 Puncture adapter: MP-2456
PW/F: 5.0/6.0 Echo coupler (stand-off ):
Intraoperative B: 5.0/6.0/7.5/10.0 60 mm UST-5713T MP-2463
PW/F: 5.0/6.0
Microsurgery B: 7.5/10.0/12.0/13.0 10 mm UST-533 Handling tool T type: MP-2749
PW/F: 5.0/6.0/7.5 Handling tool I type: MP-2750

*3 To connect UST-52104
Phased array sector probes connector adapter JB-266
is necessary.
Scanning
Application (Feature) Ultrasound frequency (MHz) *1 angle Model
Cardiology (Motorized B: 3.75/5.0/6.0/7.5, T.H.E.:3.0/3.75 90° UST-52104 *3
Transesophageal probe) PW/F: 3.0/3.75/5.0/6.0, CW: 3.75
Cardiology B: 2.5/3.0/3.75/5.0, 90° UST-52101
PW/F: 2.14/2.5/3.0/3.75, CW: 2.14
T.H.E.: 1.88
C.H.E.: B:1.88/2.14, BF:1.88,2.5
Pediatric/neonatal cardiology B: 3.75/5.0/6.0/7.5 90°
PW/F: 3.0/3.75/5.0/6.0, CW: 3.75 UST-5294-5
Transesophageal exam B: 3.75/5.0/6.0/7.5 90°
PW/F: 3.0/3.75/5.0/6.0, CW: 3.75 UST-5293-5

11
W-SHD (Wide-band Super High Density) probes and SHD probes

Convex sector probes


Application Ultrasound frequency (MHz) Scanning Radius of Model
angle curvature
Abdomen B: 3.0/3.75/5.0/6.0
PW/F:2.5/3.0/3.75/5.0 120° 14 mmR UST-9128 Puncture adapter: MP-2474
(intercostal scanning)
T.H.E:2.14
C.H.E.:B: 1.88/2.14,
BF: 1.88/2.5
Intraoperative B: 5.0/6.0/7.5/10.0
PW/F: 5.0/6.0 65° 20 mmR UST-995-7.5
(finger grip, I type)
Intraoperative B: 5.0/6.0/7.5/10.0
PW/F: 5.0/6.0 65° 20 mmR UST-MC11-8731
(finger grip, T type)
Small part, Neonatal B: 5.0/6.0/7.5/10.0
PW/F: 5.0/6.0 65° 20 mmR UST-987-7.5 Puncture adapter:
head MP-2458

Linear probes
Scanning
Application Ultrasound frequency (MHz) Model
width
Peripheral vessels*4 B: 3.75/5.0/6.0/7.5 42 mm UST-5548
(Harmonic Echo) PW/F: 3.0/3.75/5.0/6.0 Echo coupler (Flat) : ECM-12
T.H.E.: 3.75
Intraoperative *4 B: 5.0/6.0/7.5/10.0 42 mm UST-5534T-7.5
PW/F: 5.0/6.0
Laparoscopic probe B: 5.0/6.0/7.5/10.0 33 mm UST-5536-7.5 Trocar (adaptable size: 12 mm inner dia.)*
5
(Flexible type) PW/F: 5.0/6.0

*4 Steered linear scanning is possible.


*5 Trocars of some manufacturers and models may not be physically adaptable to this probe. Trocars of some manufacturers and models
may damage the probe. For adaptability of the trocar, please contact your local distributor or Aloka office.

Convex/Linear probe Puncture adapter MP-2451


Probe cover:
Endocavity Convex sector: 120 deg./9 mmR Linear: 60 mm BL-664-S (Sterilized)
(Transrectal B: 3.75/5.0/6.0/7.5 B: 5.0/6.0/7.5/10.0 UST-672-5/7.5 BL-664-NS (Nonsterilized)
bi-plane) PW/F: 3.0/3.75/5.0/6.0 PW/F: 5.0/6.0 Elastic band: FS5/16
NOTE: Sterilized probe cover cannot be Stepping device and needle
sold in EU member countries. guidance template for
brachytherapy

Phased Array probes


Application Ultrasound frequency (MHz) Scanning angle Model
Neuro surgery B: 3.75/5.0/6.0/7.5 *1 Includes puncture adapter
90° UST-5268P-5*1
PW/F: 3.0/3.75/5.0/6.0 MP-2450.
Laproscoic surgery B: 3.75/5.0/6.0/7.5 90° UST-52109
(End-fire type) PW/F: 3.0/3.75/5.0/6.0

3D probes
Description Model Ultrasound Frequencies (MHz) Convex Sector Mechanica
l sector Optional accessories
number B and M Doppler & Flow Scanning Radius of Scanning
angle curvature angle
(degrees) (mmR) (degrees)
Transvaginal ASU-1002 3.75/5.0/6.0/7.5 3.0/3.75/5.0/6.0 180 9 90 Probe cover:
RB-945BP-S (sterilized)*
RB-945BP-NS (nonsterilized)
Transabdomi ASU-1009 3.75/5.0/7.5/10.0 2.1/2.5/3.0/3.75 60 40 60 —
nal H.E.: 2.14
* Sterilized probe cover cannot be sold in EU member countries.

12
SHD (Super High Density) probes

Independent probe connection kit:


EU-3040

Indepedent Probes (EU-3040 is necessary to connect the independent


Application Ultrasound frequency Model
CW Doppler (for heart) 2.14 MHz UST-2265-2
CW Doppler (for peripheral vessels) 5.0 MHz UST-2266-5

Mechanical probe connection unit:


SCU-5500Q

MECHANICAL RADIAL PROBE


Transrectal (prostate) B: 7.5/10.0 MHz 360° ASU-67 Puncture adaptor: MP-2438

13
General Specifications

Acoustic Power
Viewing Monitor
・ 0 to 100%, changeable ・ 15-inch diagonal flicker-free non-interlaced monitor
Preset Function ・ Tilt, swivel, height adjustment are possible.

・ 15 separate programs for specific clinical applications


and/or users Safety Regulation
・ User programmable and/or factory default settings ・ Complies with IEC 60601-1 Class , Type BF
・ Preset contents storable in a floppy disk
Environmental Requirements
Characters and graphic displays In Operation
・ Character input area: ・ Temperature: +10 to +40 degrees C
・ Relative Humidity: 30 to 85%
ID, name, age, sex, retained text
(non condensing)
・ Automatic Annotation Labeling: ・ Atmospheric pressure: 700 to 1060 hPa
Up to 120 words registerable to each application In Storage
・ Body mark: Default setting: 53 kinds ・ Temperature: -10 to +50 degrees C
・ Relative Humidity: 10 to 90%
User edition: 24 kinds
(non condensing)
Menu Control ・ Atmospheric pressure: 700 to 1060 hPa
・ Color LCD (liquid crystal display) touch control panel
Power Requirement
Probe Connectors ・ 115/ 200 to 240V ±10%, 50 or 60 Hz,
・ Electronic probes: 3 separate connectors Max. 950VA for main unit
・ One dummy connector for rest Max. 350VA for optional units
Total: Max. 1300VA

Video Input/Output (for VCR) Dimensions

・ Composite, RGB, and Y/C separated signals, 1 ・ 60 cm (W) × 98 cm (D) × 132 –157 cm (H)

channel each Weight


・ Approx. 180 kg (main unit only)

14
15
・ The specifications are subject to change without notice.

・ The standard components and optional items depend on the country.


Please contact your local ALOKA distributors for details.

・ ProSound is a registered trademark of ALOKA CO., LTD.

・ Harmonic Echo is a trademark of ALOKA CO., LTD.

We care, Ultrasound@Aloka

6-22-1 Mure, Mitaka-shi, Tokyo 181-8622 Japan


Telephone: +81 422 45 6049, Facsimile: +81 422 45 4058
URL: http://www.aloka.com/

Printed in Japan, 2005-03


SP-5500SV-V6.3.1-1.0 (IEC 2-37)