Professional Documents
Culture Documents
Charpy impact tests provide an easier alternative to LEFM tests and are used to grade
the deterioration of the material.
Related terms:
Figure 4.2.76. Failure mode under Charpy impact test: (a) brittle failure for soldered
specimen, and (b) ductile failure for bulk Sn–Ag–Cu specimen.
Example 2.23
A series of Charpy impact tests on uPVC specimens with a range of crack depths gave
the following results
Crack length 1 2 3 4 5
(mm)
Fracture Ener- 100 62 46.5 37 31
gy (mJ)
Solution
Since B = D = 10 mm and using the values of from Table 2.3 we may obtain the
following information.
A graph of U against BD is given in Fig. 2.84. The slope of this gives Gc = 1.33 kJ/m2.
Alloy 263
A. Di Gianfrancesco, in Materials for Ultra-Supercritical and Advanced Ultra-Super-
critical Power Plants, 2017
Table 9.6. Range of loading rate and crack initiation time tf for different loading
techniques.
Recall Fig. 9.8 when discussing the dynamic crack initiation toughness of materials,
and Fig. 9.30 and Fig. 9.31 when discussing the dynamic crack arrest toughness
of materials; those results were obtained by Ravi-Chandar and Knauss (1984a) by
using the electromagnetic loading technology with adjustable load amplitude and
duration. The principle of this electromagnetic loading technology is schematically
illustrated in Fig. 9.36.
A flat copper strip is folded back on itself and the space between the two layers is
filled with an insulating strip. This assembly is then introduced into the crack of
single-edge crack specimen (Fig. 9.36a). When a pulse current flows through the
copper loop, each leg generates a magnetic field. The current vector in each leg
interacts with the magnetic field of the other leg to produce an electromagnetic
repulsion that forces the conductors apart. The two legs press upon the top and
bottom surfaces of the crack with a uniform pressure (Fig. 9.36b). The trapezoidal
pulse duration can be adjusted within the range of 150μs with a rising time in about
25μs. The pressure on the crack surface can be adjusted in the range of 1–20 MPa,
and the loading rate is in the order of . The major advantages of electromagnetic
loading are: (i) it provides good repeatable loading that makes experiments easy; (ii)
the experiment has been finished before the stress wave reflected from the specimen
boundary back to the crack tip and can be modeled as a pressurized semiinfinite
crack in an unbounded medium, so that is easy to analyze. Moreover, the dynamic
crack initiation toughness of the material (Fig. 9.8) and the dynamic crack arrest
toughness of the material (Fig. 9.30 and Fig. 9.31) can be carried out in the same
experiment.
Projectile impact is a typical method to generate a high loading rate. Recall the
plate impact experiment discussed in Chapter 4 “dynamic experimental study of
state equation of solids under high pressure”, in which a flyer launched by gas gun
impacts onto a target (specimen), and recall the Hopkinson bar technique discussed
in Chapter 7 “dynamic experimental study of material distortion law”, both belong
to the projectile impact technology list in Table 9.6. Different from the drop-weight
technique which directly impacts specimen by falling mass, the projectile impact
loading technique is essentially loaded by stress waves.
Hopkinson bar technique has been adopted to conduct experimental research on dy-
namic fracture toughness of materials by more and more researchers (e.g., Kalthoff,
1986; Jiang and Vecchio, 2009), which has become more and more popular and
mature.
The Hopkinson bar experiment technique has already introduced in detail in chapter
7 of this book. Here, we focus on its application in the study of dynamic fracture
toughness of materials.
Costin et al. (1977) first used the Hopkinson bar principle to study the dynamic crack
initiation toughness of materials, as schematically shown in Fig. 9.38. The specimen
is a long round rod with a circumferential precrack. An explosive charge is detonated
at one end of the specimen, generating a tensile pulse applied on the specimen. Two
strain gauges are mounted to both sides of the precrack of the long rod specimen.
The measured incident waves, reflected waves, and transmitted wave are used to
determine the dynamic crack initiation toughness of the material. Crack initiation
occurs in the range of 20–25μs, and the loading rate reaches 106 MPa·m1/2·s−1.
However, because the long rod specimen itself plays the role of both Hopkinson
incident bar and transmission bar at the same time, it consumes a lot of material
and requires high processing technology.
In addition to directly apply the tensile stress wave loading to the precracked
specimen, the specimen can also be loaded through the Hopkinson compressive
bar. In such case, a tensile stress wave can be generated by reflecting the incident
compressive stress wave from the free surface of the short specimen (Stroppe et al.,
1992), as schematically shown in Fig. 9.39. It is assumed that the compressive wave
first passing through the specimen has no effect on the crack. Some researchers
(e.g., Lee et al., 2002) added a free sleeve to the short-cracked specimen, which was
fixed between the Hopkinson incident bar and the transmitting bar (see Fig. 7.7 in
Chapter 7), so that the incident compressive wave in the incident bar propagates
into the transmitted bar through the sleeve, while the tensile wave reflected at the
free end of the transmitted bar then loads the cracked specimen.
Figure 9.39. Cracked specimen loaded by the reflected tensile pulse using Hopkin-
son pressure bar.
The above is to load the cracked specimen by tensile stress. On the other side, based
on the split Hopkinson pressure bar (SHPB) experimental technique, researchers
have also developed a variety of ways to use compressive stress wave to load cracked
samples.
The SHPB apparatus is also especially suitable for dynamic three-point bending test
to various cracked specimen by using compression pulses, which is convenient for
the measurement of dynamic fracture toughness of materials.
Fig. 9.42 schematically shows a single-edge cracked specimen loaded in the
one-point bend (1 PB) mode by compression pulse propagating through the Hop-
kinson incident bar (see e.g., Ruiz and Mines, 1985). If the single-edge cracked
specimen and the loading mode are the same as that of the usual Charpy test, it can
be regarded as an improved Charpy test that can take into account the stress wave
effect and thus can reliably determine the dynamic initiation toughness of materials
(Weisbrod and Rittel, 2000).
In addition to the one-point bend (1 PB) test mentioned above, researchers have
also developed a variety of dynamic experimental techniques for three-point bend
(3 PB) specimens using Hopkinson pressure bar (Jiang and Vecchio, 2009) to study
the dynamic fracture toughness of materials, as shown in Fig. 9.43 (see e.g., Mines
and Ruiz, 1985), Fig. 9.44 (see e.g., Jiang and Vecchio, 2007), and Fig. 9.45 (see e.g.,
Yokoyama and Kishida, 1989).
Figure 9.44. Three-point bend (3 PB) cracked specimen placed between the incident
bar and the transmission bar (2bar/3 PB).
Figure 9.45. Three-point bend (3 PB) cracked specimen placed between the incident
bar and double transmission bars (3bar/3 PB).
Fig. 9.43 shows that compression pulse loading is directly applied to the three-point
bending crack sample by using a single incident bar.
Fig. 9.44 shows that the compression pulse loading is applied to the three-point
bending cracked specimens sandwiched between the incident bar and the transmis-
sion bar.
Fig. 9.45 shows that the three-point bending cracked specimen sandwiched between
the incident bar and the double transmission bar is subjected to compression pulse
loading.
Mainly the Mode-I cracked specimens are discussed above. Similarly, the Hopkinson
pressure bar technique can also be developed to study dynamic fracture toughness
on Mode-II cracked specimen. As shown in Fig. 9.46, Dong et al. (1998) adopted
Hopkinson pressure bar apparatus to load a single-edge parallel double-cracks
specimen to study the interaction between crack initiation and adiabatic shear
for Mode-II crack. In the test, the incident bar is arranged in contact with the
specimen at the position between the parallel double cracks; a compression pulse is
transmitted to the specimen through the incident bar, resulting in a Mode-II fracture
(see Chapter 10.1.5 “Interaction between adiabatic shear bands and cracks”).