Professional Documents
Culture Documents
Contact: Phone:
E-mail:
Department: New Department Dept control #:
Room: Lithoskopija Room control #:
Content
Tested equipment
Fluoroskop I Radiograf
Name: Lithoskop Model: Serial number:
Manufacturer: Siemens
Measurements
Set text
Protect the receptor with lead sheet
Waveforms
Analysis
Summary
Tube voltage accuracy: Pass
Result: Pass
kVp Accuracy
Tested equipment
Fluoroskop I Radiograf
Name: Lithoskop Model: Serial number:
Manufacturer: Siemens
Measurements
# Set kV_x00Phantom tTube voltakVp diff % Exposure Set time_ Exposure time_x000D_(ms)
1 63 5 63.90168 1.431244 0.014463 3622
2 64.5 10 64.54411 0.01395 3711
3 68 15 70.26175 3.326102 0.01084 3788
4 110 20 110.8 0.727273 0.007125 3677
5 120 25 121 0.833333 0.005433 3620
Waveforms
Analysis
Summary
Tube voltage accuracy: Pass
Result: Pass
Input Doserate
Tested equipment
Fluoroskop I Radiograf
Name: Lithoskop Model: Serial number:
Manufacturer: Siemens
Measurements
SDD_x000DSSD_x000DDose Mode
40 FLUORO
Waveforms
Analysis
Exposure rate Min/Max check
Result: Pass
Reproducibility
Tested equipment
Fluoroskop I Radiograf
Name: Lithoskop Model: Serial number:
Manufacturer: Siemens
Measurements
Waveforms
Analysis
Summary
Tube voltage reproducibility: ---
Exposure reproducibility: ---
Exposure time reproducibility: Pass
Result: ---
Exposure reproducibility
Result: ---
Result: Pass
HVL
Tested equipment
Fluoroskop I Radiograf
Name: Lithoskop Model: Serial number:
Manufacturer: Siemens
Measurements
Waveforms
Analysis
Summary
Half Value Layer: ---
Result: ---
HVL is --- mm Al
HVL limit: minimum 1,260 mm Al
ТајмерПерформанс
Tested equipment
Fluoroskop I Radiograf
Name: Lithoskop Model: Serial number:
Manufacturer: Siemens
Measurements
Analysis
Summary
Tube mAs reproducibility: (Calculation problem)
Exposure time reproducibility: ---
mA Linearity
Tested equipment
Measurements
Analysis
Summary
mA linearity: ---
mA linearity
Result: ---
Maximum difference in mGy/mAs between adjacent stations: --- % (limit 10,0 %)
Tested equipment
Fluoroskop I Radiograf
Name: Lithoskop Model: Serial number:
Manufacturer: Siemens
Measurements
Analysis
Summary
mA linearity: (Calculation problem)
mA linearity
Column(s) are missing:
Set mA
Set time
Rad.Izlez _x HVL_x000D_(mmTotal filt Set time_ Set text
0.022516586 2.7599167824 4.611658 7.3 DR
0.022340178 2.7644836903 4.686225 7.3 DR
0.022217039 2.712454319 4.491378 7.3 DR
0.021861389 2.6729803085 4.337195 7.3 DR
0.02026087 2.6 4.5 7.3 Fluoro
0.020530435 2.58 4.5 7.3 Fluoro
0.020621739 2.59 4.6 7.3 Fluoro
0.020382609 2.61 4.7 7.3 Fluoro
Set mA_x000DExposure_x000Set text 1.121
7.9 0.0366221427 Fluoro 28.6138 1.280
7.9 0.0398214024 Fluoro 29.3169 1.358
7.9 0.0432270227 Fluoro 29.9252 1.445
7.9 0.05443 Fluoro 29.0483 1.874
7.9 0.0623 Fluoro 28.598 2.178