You are on page 1of 10

ACUSON S Family

Software
System
Service Patch Installation (VC25C through VC25F)

This document is valid for:


ACUSON S1000 ultrasound system
ACUSON S2000 ultrasound system
ACUSON S3000 ultrasound system
2 Document Version / Disclaimer / Copyright

Document Version
Siemens Healthcare GmbH reserves the right to change its products and services at any
time.
In addition, manuals are subject to change without notice. The hardcopy documents cor‐
respond to the version at the time of system delivery and/or printout. Versions to hard‐
copy documentation are not automatically distributed.
Please contact your local Siemens Healthineers office to order a current version or refer to
our website http://www.siemens-healthineers.com.

Disclaimer
Siemens Healthcare GmbH provides this documentation “as is” without the assumption of
any liability under any theory of law.
The content described herein requires superior understanding of our equipment and may
only be performed by qualified personnel who are specially trained for such installation
and/or service.

Copyright
“© Siemens Healthineers, 2018” refers to the copyright of a Siemens Healthineers entity
such as:
Siemens Healthcare GmbH - Germany
Siemens Shenzhen Magnetic Resonance Ltd. - China
Siemens Shanghai Medical Equipment Ltd. - China
Siemens Healthcare Private Ltd. - India
Siemens Medical Solutions USA Inc. - USA
Siemens Healthcare Diagnostics Inc. - USA and/or
Siemens Healthcare Diagnostics Products GmbH - Germany

ACUSON S Family US15-118.816.01.10.02 Page 2 of 10 © Siemens Healthineers, 2018


08.18
Table of Contents 3

1 Service Patch Installation (VC25C through VC25F) 4

1.1 Scope. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4
1.1.1 Supported Versions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4
1.1.2 Material Requirements . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4
1.1.3 Reference Documents . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4
1.1.4 Time Required . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4
1.1.5 System Preparation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4
1.2 Verification and Installation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5
1.2.1 Verify the Current Software Revisions. . . . . . . . . . . . . . . . . . . . . . 5
1.2.2 Install the Software Patch . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5
1.2.3 Administration. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7

2 Changes to Previous Versions 8

2.1 Changes to Previous Versions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8

3 List of Hazard IDs 9

© Siemens Healthineers, 2018 US15-118.816.01.10.02 Page 3 of 10 ACUSON S Family


08.18
4 1 Service Patch Installation (VC25C through VC25F)

1.1 Scope
The ACUSON S Family ultrasound systems periodically require software patch updates to
ensure optimal performance. This document provides software patch installation instruc‐
tions.

1.1.1 Supported Versions


Tab. 1 VC25 Software Upgrade Table

Software Up‐ From From From From From


date Version VC25A VC25B VC25C VC25D VC25E
VC25C Yes Yes No No No
VC25D Yes Yes Yes No No
VC25E Yes Yes Yes Yes No
VC25F Yes Yes Yes Yes Yes

1.1.2 Material Requirements


n DVD, S Family Update SW

1.1.3 Reference Documents


Hardware/Software Configuration Guide, Appendix A: Service SW and SW Patch Matrix

1.1.4 Time Required


30 minutes to perform the SW upgrade; additional 30 minutes to perform the FPD Opti‐
mization procedure, if required.

1.1.5 System Preparation

1.1.5.1 Prior to arriving at the site:


To save time, Siemens recommends calling the customer and asking them to archive the
patient database, preferably to a workstation, prior to the site visit. In addition, obsolete
patient records should be deleted.

ACUSON S Family US15-118.816.01.10.02 Page 4 of 10 © Siemens Healthineers, 2018


08.18
Service Patch Installation (VC25C through VC25F) 1 5

1.2 Verification and Installation


The Verification and Installation instructions include the following sections:
n Verify the current software revisions
n Install the software patch
n Administration

1.2.1 Verify the Current Software Revisions


1. Power on the system.
- Plug the power cable into a power receptacle rated for the system.
- Switch the Main circuit breaker to the On position.
- Press the Control Panel Power button, and boot the system to Imaging Mode.
2. Press the System Config (F4) key, and then click Service.
3. Verify the Product Version (Software Version) under System Information.
Fig. 1: System Software Version

a) If the desired software version is installed, no further action is required.

1.2.2 Install the Software Patch

1.2.2.1 Software Patch Installation Instructions


The installation of the software takes approximately 20 minutes.

© Siemens Healthineers, 2018 US15-118.816.01.10.02 Page 5 of 10 ACUSON S Family


08.18
6 1 Service Patch Installation (VC25C through VC25F)

 The network cable does not need to be removed because a network connection does not
affect the software patch installation.

1. Insert the update disk into the CD-R/DVD tray.


2. Press the System Config (F4) key, and then click Service.
3. Click Install Software.
The system displays a dialog box with the following message: This is a diag-
nostic\firmware installation. The software will close and then
the diagnostics will be installed. You must archive and back-up
data before doing the install. Do you want to continue install?

 This update procedure should not change or delete system presets or patient studies. It
is still recommended to perform a data backup.

4. Click OK. The system will reboot to the software update disk.
5. During system restart, a dialog box will pop up with the following message: Soft-
ware version <current version> has been verified. Press OK to
start the update to <Update Version>. Click OK to start the upgrade in‐
stall process.

If the current software version cannot be upgraded, a dialog box will pop up with follow‐


ing message: Software version <Current Version> is not supported
for update to <Update Version>. If this case happens, force the system to pow‐
er down by pressing and holding the power button.

IMPORTANT! Once the SW patch installation has started, do NOT press any key or shut

 down the system.


The whole SW patch installation process will take about 20 minutes.

6. The system will load the software patch and the system will reboot or shutdown when
completed.
7. Press the Power Button to restart the system, if the system shuts down.

1.2.2.2 Verify Software is Updated to Correct Version


Verify that the software version was updated.

1. Press the System Config (F4) key, and then click Service.
2. Verify the Product Version (Software Version) under System Information.

ACUSON S Family US15-118.816.01.10.02 Page 6 of 10 © Siemens Healthineers, 2018


08.18
Service Patch Installation (VC25C through VC25F) 1 7

Fig. 2: System Software Version (Example Only)

1.2.3 Administration

1.2.3.1 Final Work Steps

1. Eject the software patch installation disk.


2. Update the system documentation. Update the revision level, the user documents (if
applicable) and the technical documentation.
3. Save the new software disk together with the system software disk.
4. MANDATORY: Always update the install database to show the newly installed software
version.

© Siemens Healthineers, 2018 US15-118.816.01.10.02 Page 7 of 10 ACUSON S Family


08.18
8 2 Changes to Previous Versions

2.1 Changes to Previous Versions


Change(s) Description
Revision 09 Updated title and document to include SW
Version VC25F.
Revision 08 Updated title and document to include SW
Version VC25E. Title was Service Patch Instal‐
lation (VC25C and VC25D) and now is Service
Patch Installation (VC25C through VC25E).
Updated reference to VC25E within docu‐
ment.
Revision 07 Updated title and document to include SW
Version VC25D. Title was Service Patch Instal‐
lation (from VC25A/B to VC25C) and now is
Service Patch Installation (VC25C and
VC25D).
Revision 06 Changed document to include all S Family.
Change title from Service Update Installation
(Version VB21A to VB21B) to Service Patch In‐
stallation (from VC25A/B to VC25C)
Removed Deactivate/Reactivate System Man‐
agement sections.
Removed reference to performing FPD Opti‐
mization procedures.
Revison 05 Revised sections
Revision 04 Changed the document to upgrade SW Ver‐
sion from VB21A to VB21B. Changed referen‐
ces to this throughout the document.
Originally this dccument was for a SW Patch
Installation for SW Versions VB10D/VB10E.
Revision 03 Added note before SW Patch installation
steps:
“The network cable does not to need be re‐
moved because a network connection does
not affect the software patch installation.”
Revision 02 Changed document title to include VB10D
and VB10E.
Updated software version references to in‐
clude VB10E
Revision 01 Initial release.

ACUSON S Family US15-118.816.01.10.02 Page 8 of 10 © Siemens Healthineers, 2018


08.18
List of Hazard IDs 3 9

There are no Hazard IDs in this document.

© Siemens Healthineers, 2018 US15-118.816.01.10.02 Page 9 of 10 ACUSON S Family


08.18
All documents may only be used for
rendering services on Siemens
Healthcare Products. Any document
in electronic form may be printed
once. Copy and distribution of elec‐
tronic documents and hardcopies is
prohibited. Offenders will be liable
for damages. All other rights are re‐
served.

siemens-healthineers.com

Siemens Healthineers Headquarters


Siemens Healthcare GmbH
Henkestr. 127
91052 Erlangen
Germany
Telephone: +49 9131 84-0
siemens-healthineers.com

Print No.: US15-118.816.01.10.02 | Replaces: US15-118.816.01.09.02


Doc. Gen. Date: 08.18 | Language: English
© Siemens Healthineers, 2018

You might also like